You are on page 1of 64

VetLyte ®

Electrolyte Analyzer
Operator's Manual

06-02305-00
TRADEMARKS

VetTest, VetLyte and IDEXX VetLab are trademarks of IDEXX Laboratories, Inc. in the United States
and/or other countries.
Preface

Welcome

Your VetLyte Electrolyte Analyzer is a powerful tool designed to help you quickly, accurately
and efficiently conduct basic electrolyte testing in the convenience of your own laboratory.

This manual will help guide you through setting up your analyzer and will help you start running
samples. As you become familiar with the operation of the unit, you may use the manual as a
reference for day-to-day routines and as a guide for maintenance and troubleshooting.

How to Use This Manual

If you have an analyzer that is not yet set up, you should begin by reading Sections A and B.
Information on analyzer operation and maintenance is contained in Sections C and D. Operator
functions are described in Section E. Detailed service information and operating principles can
be found in Sections F and H. Self-help procedures are outlined in Section G.


Table of Contents

A. Getting to Know Your VetLyte


Important Safety Instructions................................... A-1
Analyzer Components.............................................. A-3

B. Installation
Before you Begin..................................................... B-1
Setting Up . .............................................................. B-2
Electrodes and Measuring Chamber........................ B-2
Preparing the Analyzer for Operation...................... B-5
Connecting the VetLyte to the VetTest..................... B-6
Initial Electrode Maintenance.................................. B-8

C. Operation
Sample Preparation.................................................. C-1
Sample Storage Guidelines...................................... C-1
Operational Characteristics...................................... C-2
Operating the VetLyte Individually.......................... C-3

D. Maintenance
Scheduled Maintenance........................................... D-1
Unscheduled Maintenance....................................... D-6

E. Operator Functions
Checking Reagant Fluid Level ............................... E-1
Standby Mode ......................................................... E-2
Setting Date and Time.............................................. E-2
View Calibration Data.............................................. E-2
Change Baud Rate .................................................. E-2
Take Out of Operation ............................................ E-2

F. Service Functions
Test Electrodes..........................................................F-1
Test Sample Sensor...................................................F-1
Test Sample Door......................................................F-2
Test Peristaltic Pump.................................................F-2
Test Pinch Valves......................................................F-2
Test Amplifier ..........................................................F-3
Enter Service Codes..................................................F-3

G. Self-Help Guide
Status: Not Calibrated.............................................. G-1
Standard A Not Detected.......................................... G-1
Standard B Not Detected......................................... G-1
Check Sample Sensor.............................................. G-2
ii Check Reference Housing........................................ G-2
Please Close Sample Door....................................... G-2
No Sample................................................................ G-2
Cleaning Fluid Not Detected................................... G-3
Conditioning Fluid Not Detected............................. G-3
Check Electrodes..................................................... G-3
Clog in Sample Path................................................ G-4
↓↓↓↓ ↑↑↑↑............................................................. G-4
ERR......................................................................... G-4

H. Principles of Operation
The Measurement Principle..................................... H-1
Physical Principle..................................................... H-1
Sodium Electrode Specifications............................. H-4
Potassium Electrode Specifications......................... H-5
Chloride Electrode Specifications............................ H-6
Reference Electrode Assembly Specifications......... H-7

Appendices
Technical Specifications.............................................I-1
Program Flow Chart.................................................. J-1
Maintenance Log Master......................................... K-1
Warranty................................................................... L-1

iii
Section A - Getting To Know Your VetLyte

Important Safety Instructions


Before you begin installing your VetLyte, carefully read the overview information in this
section.

For your own safety and the proper operation of your equipment, always follow these
precautions when working with your VetLyte:

• Keep the analyzer away from all sources of liquids such as sinks and wash basins.
• Don’t use ammonia-based or alcohol-based cleaners, which can chemically react with
plastic, on or around the analyzer.
• Always handle blood samples and collection devices with care.
• Use approved protective gloves to avoid direct contact with sample.
• Aseptic procedures are required when cleaning the sampling probe to avoid
contamination.
• Dispose of ISE Fluid Pack according to local regulations.

A-1
VetLyte Operator’s Manual – Getting To Know Your VetLyte
Control Panel Display Control Panel Keypad

Front Door

Sample Probe Mechanism

ISE Fluid Pack

A-1. VetLyte Analyzer Major Components (external)

A-2
Getting To Know Your VetLyte – VetLyte Operator’s Manual
Analyzer Components
The VetLyte Electrolyte Analyzer is a fully automatic,
microprocessor-controlled medical instrument that
measures:

YES
NO Na+ : Sodium
K+ : Potassium
Cl¯ : Chloride

The analyzer consists of several major components that


are important for you to know and understand while
becoming familiar with the unit.
A-2. Control panel keypad See illustration A-1.

You communicate with the analyzer through a keypad


with YES and NO keys. With these keys you can
perform all analyzer functions, including: sample
measurement, data input and quality control testing.
See illustration A-2.
"/"-:;&4".1-&
 The analyzer communicates to you through a dot matrix
display. This two line, alphanumeric readout allows
up to 16 characters per line; displaying the activities
of the analyzer, sample results and other programmed
information.
See illustration A-3.
A-3. Control panel display
Inside the unit are other components which are
accessible by opening the main door.
See illustrations A-4 and A-5.

A-4. Opening the main door

A-3
VetLyte Operator’s Manual – Getting To Know Your VetLyte
Measuring Chamber

Peristaltic Pump

Sample Probe
Mechanism

Fluid Control
Valves

A-5. VetLyte Analyzer Major Components (internal)

A-4
Getting To Know Your VetLyte – VetLyte Operator’s Manual
The measuring chamber consists of the movable left
locking device that holds the electrodes in place, the
electrodes, the right electrode holder with sample sensor
connector, and the measuring chamber base. Electrodes
are labeled:

Ref : Reference
REF Na +
K +
Cl -

Na+ : Sodium
K+ : Potassium
Cl¯ : Chloride

See illustration A-6.


A-6. Measuring Chamber
A peristaltic pump is used to transport all liquids within
the analyzer.
See illustration A-7.

Valves control the movement of the liquid within the


analyzer.
See illustration A-8.

The sample probe mechanism is located behind the small


door at the front of the unit.
See illustration A-9.

A-7. Peristaltic Pump

R
V

A-8. Fluid control valves A-9. Sample probe mechanism

A-5
VetLyte Operator’s Manual – Getting To Know Your VetLyte
The self-contained ISE Fluid Pack uses an integral
checkvalve to ensure that waste cannot spill out of the
package.
See illustration A-10.

The model and serial number is located on an


identification plate above the probe mechanism.
See illustration A-11.

The rear panel of the unit contains a serial number plate,


as well as the power switch/power receptacle module
and an RS232 interface port.
See illustration A-12.
A-10. ISE Fluid Pack
Congratulations!

You’ve just learned the basic components of the analyzer


and are now ready to install your system.

A-11. Model - Serial No. Plate

VetLyte

A-12. Rear panel

A-6
Getting To Know Your VetLyte – VetLyte Operator’s Manual
Section B - Installation

Location is important for trouble-free operation of your analyzer. Before you begin setup,
choose a site that is convenient for your sampling needs and meets the following physical
requirements of the unit:

• Grounded electrical outlet, not shared with centrifuges, refrigerators, or microscopes.


• Away from direct sunlight
• Room temperature between 60°F and 90°F
• Maximum relative humidity of 85%
• Ample room to allow air to circulate freely around the unit. See illustration B-1.
• Away from strong electromagnetic fields, such as those created by electric motors and
x-ray equipment.
• Away from explosive gases or vapors.

Now it’s time to unpack your VetLyte Electrolyte Analyzer. Carefully remove the unit from
the box. DO NOT lift the analyzer by the foam packaging materials, which are provided for
shipping only.

Before you begin installing your system, take a moment to look over the contents to ensure
that you have everything you need to get your analyzer up and running.
Check for these items:

VetLyte Electrolyte Analyzer


Accessory Kit:

Operator’s Manual
Power cord
Electrodes
ISE Fluid Pack
Cleaning Solution A
Electrode Conditioning Solution
Cable

Trouble-Shooting Kit: 1 12cc service syringe


1 stylet
4 fuses
2 o-rings

13' 3/8"

You will also need a supply of lint-free tissues and


disposable sample cups, which should be kept in a
location convenient to the analyzer.

13' 3/4" 13"

B-1. Space Requirements

B-1
VetLyte Operator’s Manual – Installation
Setting Up

Now you’re ready to get your VetLyte Electrolyte


Analyzer prepared to operate. Prior to beginning the
actual installation, it is a good idea to completely read
V through this section to develop an understanding of the
procedures that are required.

Begin by placing the analyzer on a secure table top that


allows plenty of working space and is convenient to a
power connection.
A
Open the analyzer main door. Locate and carefully
B-2. Removing relief clamps remove the four red relief clamps from the valves by
sliding out the clamps.

Note: Save the clamps for re-use to prevent damaging


the tubes in the event the analyzer is later shut down for
any reason. Store clamps in your troubleshooting kit.

See illustration B-2.

Slip the pump tubings around the analyzer pump rollers,


being careful not to twist or cross them.
See illustration B-3.

Electrodes And Measuring Chamber


B-3. Installing the pump tubings The next procedure involves preparing and installing
the electrodes in the measuring chamber. Remove the
reference housing and electrodes from their protective
boxes and place them on a soft, clean surface. Check
that each electrode has an o-ring in the left side of the
electrode.
See illustration B-4.

B-4. Electrode o-ring

B-2
Installation – VetLyte Operator’s Manual
Unscrew the red transport housing from the reference
electrode and check that the o-ring on the electrode is
properly seated.
See illustration B-5.

Note: Save the transport housing for storage of the


reference electrode during monthly maintenance and
in the event the analyzer is taken out of service for any
reason. Discard any reference solution remaining in the
transport housing and store in your troubleshooting kit.

Carefully screw the reference electrode into the


reference electrode housing and place it with the other
B-5. Transport housing removal
electrodes.
See illustration B-6.
Reference Reference
electrode
housing
electrode Slide the measuring chamber forward until it locks in
the front position. Unclamp the left electrode holder by
moving the lever toward you.
See illustrations B-7 and B-8.

Red transport
housing

B-6. Installing reference electrode

B-7. Slide chamber forward B-8. Unclamping electrode holder

B-3
VetLyte Operator’s Manual – Installation
Locate the sample sensor cable and ensure that it is
securely inserted into the receptacle above the measuring
chamber. Check that an o-ring is present in the right
electrode holder.
See illustration B-9.

Now install the electrodes in the measuring chamber,


beginning on the right and working to the left (the
reference electrode will be installed last). Check to
make sure that the lettering on the measuring chamber
matches the lettering on the electrode. Also, note that
all electrodes have a lip on the bottom that rests on the
flat edge of the measuring chamber to aid in proper
B-9. Sample sensor cable and o-ring positioning.
See illustration B-10.

Close the left electrode holder by moving the lever away


from you until it locks in the back position, and ensure
that the electrodes are properly seated.
See illustration B-11.

Slide the measuring chamber back until it snaps into


position.
See illustration B-12.

B-10. Installing electrodes

B-11. Closing electrode clamp B-12. Slide chamber backward

B-4
Installation – VetLyte Operator’s Manual
Plug the tubing connector of the reference housing
assembly into the receptacle below the left side of
the measuring chamber. Close the front cover of the
analyzer.
See illustration B-13.

Preparing The Analyzer For Operation

Record the date of installation of the ISE Fluid Pack on


the pack label. Remove the red protective strip from the
Fluid Pack and slide the pack into position on the left
side of the analyzer.
See illustrations B-14 and B-15.
B-13. Reference connector
Note: Once the protective strip is removed, be sure
to keep the pack upright to avoid spillage. Save the
protective strip to use to close the ports on the pack
prior to disposal.

B-14. Removing protective strip

B-15. Installing fluid pack

B-5
VetLyte Operator’s Manual – Installation
B-16. Connecting RS232 cable
Connecting The VetLyte To The VetTest
Note: Do not plug in the VetTest before reading
instructions regarding set-up of the VetTest. It is critical that
the instructions found in the VetTest operator’s manual be
followed in detail during this procedure.

Warning: Do not unplug or power OFF the VetLyte


after initial set-up. Failure to leave the VetLyte plugged
in with a fluid pack in place could cause damage to the
system.

Connect only the RS232 cable from the VetLyte to the


VetTest as shown.
B-17. Power cord installation See illustration B-16

Locate the power switch on the back of the VetLyte and


make sure that it is in the OFF (O) position. Next, plug the
power cord into the power receptacle module on the back
of the VetLyte, then plug the cord into a grounded electrical
outlet.
See illustration B-17.

Push the power switch to the ON (I) position.


The VetLyte will automatically begin to operate.
See illustration B-18.

Now that the VetLyte Electrolyte Analyzer is functioning,


you will begin using the keypad interface to communicate
B-18. Turning power on with the instrument. Use the NO key to make changes, the
YES key to accept the displayed values or information.

B-6
Installation – VetLyte Operator’s Manual
The analyzer will display a default date and time and will allow you to input the correct
date and time.

Date: 01-JAN -80


Time: 00:00

You enter the correct date as follows:

1. Press the NO key until the correct day is displayed. Press YES. The cursor will move to
the month.
2. Press NO until the correct month is displayed. Press YES.
3. Press NO until the correct year is displayed. Press YES. The correct date should now be
displayed.
4. Follow the same procedure to enter the correct time.
5. After entering the time, the analyzer will prompt: OK? Press YES, if the date and time
you entered is correct, or press NO to make a change.
6. After entering YES, the system will enter a brief electronic warm-up period.

Note:: By keeping the NO key depressed, the analyzer will automatically scroll through the
numbers, first slowly, then fast.

Congratulations!
Your VetLyte Electrolyte Analyzer is now prepared for initial electrode maintenance.

Connecting the VetLyte to the VetLab


(See VetTest Appendix Q)

B-7
VetLyte Operator’s Manual – Installation
Initial Electrode Maintenance
Prior to performing your first calibration or running your
first sample, the VetLyte Electrolyte Analyzer needs to
undergo a simple cleaning and conditioning procedure
that helps ensure the unit will perform properly.

The process involves cleaning and conditioning the


sample path and electrodes, which prepares the VetLyte
analyzer for calibration. You should have ready the
bottles containing Cleaning Solution A and Electrolyte
Conditioning Solution, along with a package of lint-free
tissues that will be used to dry the probe.
B-19. Introducing cleaning solution
Note: In some cases, when the unit prompts you for
an action and you do not respond within a set period of
time, an alarm will sound and the unit will discontinue
its current operation.

To perform cleaning, you will communicate through


the keypad interface. The prompt Perform Weekly
Cleaning? will be displayed. Press YES to accept.

Pour a small amount of VetLyte Cleaning Solution A into


a clean sample cup. The prompt Open Sample
Door/Introduce Sample will be displayed.

Lift the sample door. Wait for the prompt Introduce


Sample to be displayed. The pump will begin to
B-20. Cleaning probe aspirate. Introduce the cleaning solution to the probe.
See illustration B-19.

Hold the solution under the probe until the prompt


Wipe Probe/Close Sample Door is displayed. Use
a lint-free tissue to wipe the probe, then close the door.
See illustration B-20.

The analyzer will now display Thank You! and a


brief countdown will begin, indicated by the clock in
the lower right of the display. While the countdown
is running, open the bottle of VetLyte Electrode
Conditioning Solution and pour a small amount into a
clean sample cup.

B-8
Installation – VetLyte Operator’s Manual
After the countdown is completed, the system will
prompt Perform Daily Conditioning? Press YES. The
prompt Open Sample Door/Introduce Sample will
be displayed. Lift the sample door. Wait for the prompt
Introduce Sample to be displayed. The pump will begin to
aspirate. Now, introduce the conditioning solution to the
probe.
See illustration B-21.

Hold the solution under the probe until the prompt Wipe
Probe/Close Sample Door is displayed. Use a lint-free
tissue to wipe the probe and close the door.
See illustration B-22.
B-19. Introducing cleaning solution
The analyzer will now display Thank You! and a
brief countdown will begin. Upon completion of the
countdown, the prompt Remain in Maintenance? will
be displayed. Press NO. The unit will prompt Are You
Sure? Press YES.

The unit will display Calibration in Process while the


analyzer prepares for automatic calibration. A countdown
will be displayed as soon as the actual calibration cycle
starts.

Note: Calibration is an automatic process. During this


time, the analyzer is conducting measurement operations
to ensure the accuracy of the instrument.

B-20. Cleaning probe Warning: It is very important that the main door is
closed during calibration, since it provides shielding from
sources of electromagnetic interference.

When calibration is completed, the analyzer will display


ANALYZE SAMPLE?, indicating that the instrument is
now prepared for operation.

The unit is factory set to default to Standby after 3


hours in ANALYZE SAMPLE? mode.

B-9
VetLyte Operator’s Manual – Installation
Section C - Operation

The VetLyte Electrolyte Analyzer provides fast, convenient analysis of serum, plasma or
anticoagulated whole blood.

Sample Preparation

Serum Sample
Draw patient blood into a syringe or vacuum collection tube. Do not heparinize syringe. If
using syringe, remove needle and quickly transfer to a red top OR serum separator tube. Let
tube sit for a minimum of 20 minutes. Make sure sample is fully clotted.
(Clotting time may vary by sample.)
Centrifuge for 10 minutes in a standard centrifuge (8,000 rpm) OR for 120 seconds in a high
speed (12,000 - 16,000 rpm) centrifuge.
Process immediately, or see sample storage guidelines.

Plasma Sample
Draw patient blood into a heparinized syringe or vacuum collection tube. If using syringe,
remove needle and quickly transfer to a lithium heparin tube. Fill 1/2 to 3/4 full. Gently invert
for 30 seconds to mix.

Centrifuge for 5 minutes in a standard centrifuge (8,000 rpm) OR for 90 seconds on a


high speed (12,000 - 16,000 rpm) centrifuge. Process immediately, or see sample storage
guidelines.

Note: Do not use anticoagulants such as EDTA, citrate, oxalate, etc.

Whole Blood Sample


Collect the blood in lithium heparin tube or heparinized syringe. Invert tube several times to
mix, and mix well just before analyzing sample. Process immediately, or see sample storage
guidelines.

Note: Do not use anticoagulants such as EDTA, citrate, oxalate, etc.

Sample Storage Guidelines

While we recommend that you process and analyze blood samples immediately after
collection for best results, following are sample storage and testing guidelines.

For storage, the serum or plasma must be separated and removed from the blood cells. Do
not attempt to pour off sample. Using a transfer pipette, carefully transfer to an untreated
collection tube, taking care not to draw up any white or red blood cells. Cap tube tightly to
avoid contamination and evaporation.

Storing serum/plasma
If you cannot perform analysis within four (4) hours of drawing and processing the sample,
refrigerate at 2 to 8°C (36 to 46°F). If you cannot perform analysis for more than 48 hours,
you should freeze serum/plasma at -18°C (0°F).

C-1
VetLyte Operator’s Manual – Operation
Analysis of Stored Samples
For samples stored between 2 and 8°C (36° and 46°F) and at - 18°C (0°F):
Allow the sample to come to room temperature, between 19 and 27°C (66 and 81°F).
Mix gently but thoroughly by inversion. Do not shake!
Centrifuge sample to remove any fibrin particles that may have formed during storage.
Analyze sample immediately after centrifugation.

Storing whole blood


Whole blood samples should be analyzed as soon as possible within one hour after collecting
the sample. If a brief storage is required, do not cool the sample as the erythrocytes could
burst and release the intracellular potassium, creating an inaccurate value of potassium in the
sample.

Operational Characteristics

With VetTest
The test results of the VetLyte will show in its display and will also printout on the VetTest
or an external printer. Electrolyte reference ranges will show on the printout, along with
the results. If results are beyond the reference ranges for any analyte, Hi or Lo indicators
will be displayed and printed on the VetTest printout. Results above the linearity range of
the analyzer will not print out on the VetTest, and will be displayed as dashes (--.--) on an
external printer.

Sample measurement may be initiated on either the VetTest or the VetLyte. It is recommended
that the sample be first presented to the VetTest, as the sample cycle time is longer than that
of the VetLyte. Under these conditions, chemistry testing on the VetTest may be in progress
while you begin sample aspiration on the VetLyte.

In this case, follow the prompts appearing on the displays of each analyzer. When each
analyzer has completed its test run, the data for both the VetLyte and VetTest will be printed
on the VetTest print-out under the patient number which was entered on the VetTest.

If the sample is entered on the VetLyte first, the VetLyte will prompt you to enter the patient
species. It is important to note that if patient species and patient ID data is not entered on the
VetTest, electrolyte data from the VetLyte will appear only in the VetLyte display and will not
appear on the VetTest print-out or be combined with the VetTest chemistry results.

With VetLab
(See VetTest Appendix R)

C-2
Operation – VetLyte Operator’s Manual
Operating the VetLyte Individually
The operator can also run the VetLyte analyzer when the VetTest analyzer is turned off. In this
case, the VetLyte test results will show only on the VetLyte analyzer display and will not be
available from the VetTest.

Running A Sample With The VetLyte


The VetLyte provides fast, easy operation. Whenever
ANALYZE SAMPLE? appears on the display, the unit is
prepared to conduct sampling measurements.

The analyzer prompts you to open the sample door and


introduce the sample. Lift the sample door. The prompt
Introduce Sample will be displayed and the pump will
begin to aspirate. Introduce the sample to the probe.
See illustration C-1.

Note: It is very important that the main door is closed


during sampling, since it provides shielding from sources
of electromagnetic interference.

C-1. Introducing a sample Hold the sample under the probe until Wipe Probe and
Close Sample Door is displayed. Use a lint-free tissue
to clean the probe, then close the sample door when
prompted.
See illustration C-2.

Note: It is very important that the sample probe is


carefully wiped clean each time following a sample
introduction.
The analyzer will display Thank You! and a brief
countdown will begin. Upon completion of the analysis,
the test results will be displayed on the VetLyte. If you
entered the sample type and the patient number on the
VetTest, the results will be displayed and printed on the
VetTest.

Note: Values that are higher or lower than the


programmed normal range will be indicated by an arrow
C-2. Cleaning probe pointing up or down.

View Last Patient Results


The last patient’s results will be displayed for one minute
after analysis, or until the NO or YES key is selected.

If the last patient result is required, press NO from


Analyze Sample? and press YES to View Last
Patient Results? The results will be displayed for one
minute.

C-3
VetLyte Operator’s Manual – Operation
Section D - Maintenance

Scheduled Maintenance:
Performing Electrode Maintenance
To ensure proper performance of the analyzer, a daily
conditioning and weekly cleaning is required.

Weekly Maintenance Requirements


To begin weekly maintenance, press NO until the prompt
WEEKLY MAINTENANCE? is displayed. Press YES to
accept.

Perform Weekly Cleaning? will be displayed. Press


YES to start the cleaning procedure.
D-1. Introducing Cleaning Solution A
Pour a small amount of VetLyte Cleaning Solution A
into a clean sample cup.
The prompt Open Sample Door Introduce Sample
will appear.

Lift the sample door. Wait for the prompt Introduce


Sample. The pump will begin to aspirate. Introduce the
cleaning solution to the sample probe.
See illustration D-1.

Continue holding the solution under the sample probe


until the prompt Wipe Probe/Close Sample Door is
displayed. Use a lint free tissue to wipe the probe, then
close the door.
See illustration D-2.
D-2. Cleaning Probe
The analyzer will now display Thank You!, and a brief
countdown will begin.

After the countdown is completed, the system will


prompt, Perform Daily Conditioning? To continue
with the procedure, press YES.

Daily Maintenance Requirements


To begin daily maintenance, press NO until the prompt
Perform Daily Conditioning? is displayed. Press
YES to accept.

Open the bottle of VetLyte Electrode Conditioning


Solution and pour a small amount into a clean sample
cup.

D-1
VetLyte Operator’s Manual – Maintenance
The prompt Open Sample Door Introduce Sample
will be displayed. Lift the sample door. Wait for the
prompt Introduce Sample. The pump will begin to
aspirate. Introduce the conditioning solution to the sample
probe.
See illustration D-3.

Continue holding the solution under the sample probe


until the prompt Wipe Probe/Close Sample Door is
displayed. After wiping the probe and closing the sample
door, the analyzer will display Thank You!
See illustration D-4.

D-3. Introduce conditioning solution Upon completion of the countdown, the prompt Remain
in Maintenance? will be displayed. Press NO. The
unit will prompt Are You Sure? press YES. The
VetLyte Electrolyte Analyzer will automatically initiate a
calibration cycle.

Note: It is very important that the main door is closed


during calibration, since it provides shielding from
sources of electromagnetic interference.

When calibration is completed, the analyzer will display


ANALYZE SAMPLE?, indicating that the instrument is
prepared to conduct electrolyte sampling.

D-4. Cleaning probe

D-2
Maintenance – VetLyte Operator’s Manual
Performing Monthly Maintenance
Monthly maintenance involves cleaning the reference
electrode housing, and should be performed prior to
daily maintenance. To complete this procedure, you will
need a small amount of household bleach.

Begin monthly maintenance by pressing NO until


ELECTRODE MAINTENANCE? is displayed. Press YES
and Perform Weekly Cleaning? will appear. Press
NO and Perform Daily Conditioning? will appear.
Do not press any keys.

D-5. Removing reference connector Open the front cover of the analyzer. Unplug the tubing
of the reference electrode from the receptacle below the
left side of the measuring chamber.
See illustration D-5.

Slide the measuring chamber forward until it locks in the


front position.
See illustration D-6.

Unclamp the left electrode holder by moving the lever


toward you.
See illustration D-7.

Remove the reference electrode assembly from the


analyzer. Unscrew the reference electrode from the
reference housing.
D-6. Sliding measuring chamber to
front locked position Warning: It is important the reference electrode
always be stored in reference solution and never allowed
to become dry. DO NOT bleach reference electrode.

Pour some bleach into a small container, and submerge


the reference housing into the bleach, insuring that no air
bubbles remain in the housing.

Note: The reference connector and tubing do not need


to be submerged.

Pour the reference solution from the reference electrode


into the red transport housing or a sample cup. Store the
reference electrode in the reference solution.

D-7. Unclamping left electrode holder After 15 minutes, remove the reference housing from the
bleach, thoroughly rinse the housing with tap water and
dry.

D-3
VetLyte Operator’s Manual – Maintenance
Remove the reference electrode from the red transport
housing or sample cup. Check that the o-ring on the
electrode is properly seated.

Carefully screw the reference electrode into the


reference electrode housing and place the assembly into
the left side of the measuring chamber. Note that the
reference electrode has a lip on the bottom that rests on
the flat edge of the measuring chamber.
See illustration D-8.

Close the left electrode holder by moving the lever away


D-8. Installing reference electrode from you until it locks in the back position. Ensure the
electrodes are seated properly.
See illustration D-9.

Plug the tubing of the reference electrode into the


receptacle below the left side of the measuring chamber.
See illustration D-10.

Slide the measuring chamber back until it snaps into


position, and close the front cover of the analyzer.
See illustration D-11.

At this time, you perform daily conditioning by pressing


YES. The prompt Remain In Maintenance? will
appear. Press NO. The unit will prompt Are You
Sure?. Press YES. A complete calibration cycle will be
D-9. Closing electrode holder clamp performed.

D-10. Installing reference electrode D-11. Installing measuring chamber


connector

D-4
Maintenance – VetLyte Operator’s Manual
Performing 6-Month Maintenance
Every six months, the peristaltic pump tubing needs to
be replaced.

To make sure that the pump does not turn on


during this procedure, press NO until ELECTRODE
MAINTENANCE? is displayed. Press YES and
Perform Weekly Cleaning? will appear. Press NO.
Perform Daily Conditioning? will appear. Do not
press any keys.

To change the tubing, open the analyzer front cover and


D-12. Removing pump windings slip the pump tubing from the analyzer pump rollers. See
illustration D-12.

Next, disconnect one tube at a time from the old pump


winding plate and reconnect to the same place on the
new pump tubing plate.
See illustration D-13.

Install the new pump tubing plate and slip the new pump
tubing over the analyzer pump rollers, being careful not
to twist or cross them.

Press NO to Perform Daily Conditioning? Press NO


to Remain In Maintenance? The unit will prompt
Are You Sure? press YES. A calibration will be
initiated.
D-13. Pump winding plate

D-5
VetLyte Operator’s Manual – Maintenance
Unscheduled Maintenance:
On a weekly basis, or whenever necessary, you should
clean the sample fill port and sample probe, as well as
the exterior analyzer surfaces.

To clean the sample fill port and probe, open the sample
door and clean the fill port, probe and surrounding area
with a damp cotton swab. When finished, close the
sample door. The exterior surfaces should be wiped
clean with a soft, damp cloth.

Note: Never use strong or abrasive cleaners on the


D-14. Slide chamber forward VetLyte Electrolyte Analyzer. Use a slightly damp cloth
to avoid getting fluid inside the analyzer.

Changing Electrodes
Slide the measuring chamber forward until it locks in the
front position.
See illustration D-14.

Unclamp the left electrode holder by moving the lever


toward you.
See illustration D-15.

Remove the used electrode from the measuring chamber.


See illustration D-16.

Remove the new electrode from its protective box and


check for the presence of an o-ring in the left side of the
D-15. Unclamp electrodes electrode.
See illustration D-17.

D-16. Remove used electrode D-17. Electrode o-ring

D-6
Maintenance – VetLyte Operator’s Manual
Install electrode in its labeled position in the measuring
chamber. Note that the electrode has a lip on the bottom
that rests on the flat edge of the measuring chamber to
aid in proper positioning.
See illustration D-18.

Close the left electrode holder by moving the lever away


from you until it locks in the back position. Ensure that
the electrodes are seated properly.
See illustration D-19.

Slide the measuring chamber back until it snaps into


position.
D-18. Installing new electrode
See illustration D-20.

After installing a new electrode, the VetLyte Electrolyte


Analyzer needs to undergo Electrode Maintenance and
calibration to verify the performance of the electrode.

D-19. Closing electrode clamp

D-20. Slide measuring chamber backward

D-7
VetLyte Operator’s Manual – Maintenance
Section E - Operator Functions

Checking Reagent Fluid Level And


Changing The ISE Fluid Pack
The VetLyte monitors the level of solutions in the ISE
Fluid Pack and displays the amount remaining. At 5%
fluid remaining, the unit will display the message Fluid
Pack Low at Analyze Sample? and Standby!
prompts. This message will go away only after the fluid
counter has been reset to 100%.

To check the status of fluid remaining in the pack:


1. Press NO until the prompt, Operator Functions?
is displayed. Press YES.
E-1. Removal of used ISE Fluid Pack 2. The prompt Check/Change Reagent Pack? will
be displayed.
3. Press YES. The analyzer will display the amount
of fluid remaining then the prompt New Reagent
Pack Installed? will appear.

Note: If ample fluid remains in the pack and it is not


necessary to change it, press NO to return to Analyze
Sample?.

4. To change the pack, just grasp the extended


portion of the pack and slide it out. If removal is
difficult, press on the end of the fluid pack guide pin
(protruding through the connector located to the left
of the measuring chamber inside the front door).
See illustration E-1.
5. Prepare the new ISE Fluid Pack by carefully
E-2. Installing ISE Fluid Pack removing the red protective strip.

Note: Once the protective strip is removed, be sure to


keep the pack upright to avoid spillage. Save the
protective strip to use to close the ports on the pack
prior to disposal.

6. Slide the new pack into position on the left side of


the analyzer.
See illustration E-2.
7. Press YES to indicate that a new fluid pack is
installed. The analyzer will prompt Are You Sure?.
8. Press YES, and the VetLyte Electrolyte Analyzer will
automatically reset the fluid pack counter to 100%
and commence system calibration.

E-1
VetLyte Operator’s Manual – Operator Functions
Standby Mode
Your analyzer has been programmed to automatically enter Standby mode three hours after
calibration. If no further sampling is required, you may place the analyzer in Standby mode.

To access this mode, follow the steps below:


1. Press NO until the prompt, Operator Functions? is displayed. Press YES.
2. Press NO until the prompt Go to Standby Mode? is displayed and press YES.
3. The analyzer will display Standby! to confirm that it has entered Standby mode.

To leave Standby and resume normal operation:


1. Press NO. The prompt Leave Standby Mode? will be displayed.
2. Press YES.

Setting Date And Time


The date and time of the VetLyte can be easily changed as necessary. To change the date
and time, enter the Operator Functions? menu and press NO until Set Time / Date? is
displayed.

1. Press YES to enter a change.


2. Press NO until the correct day is displayed. Press YES. The cursor will move to the
month.
3. Press NO until the correct month is displayed. Press YES.
4. Press NO until the correct year is displayed.
Press YES. The correct date should now be displayed.
5. Follow the same procedure to enter the correct time. The analyzer will prompt: OK?
Press YES if the date and time you entered is correct, or press NO to continue making
changes.

View Calibration Data


The most recent calibration data for each electrode can be viewed under this menu. Go to
Operator Functions. Press NO until View Calibration Data? is displayed. Press YES
and the 3 parameters will be displayed with the respective slope in mV. The number in
parentheses shows the number of STD A aspirations. This number can be between 3 and 6.
See Section F, page 1 to view voltage levels for each electrode.

Change Baud Rate


This menu option is for service use only.

Take Out Of Operation


Warning: If YES to Take Out of Operation? is selected accidentally, you must
remove fluid pack and turn OFF analyzer for 30 seconds. Replace fluid pack and turn
analyzer ON.

E-2
Operator Functions – VetLyte Operator’s Manual
The take-out-of-operation routine assists you in
performing a complete shutdown of the analyzer. A
complete shutdown may be indicated to prepare the
analyzer for shipping or in case the analyzer is not being
used for an extended period of time. For this procedure,
you will need a special shutdown kit (not supplied with
the analyzer).

Note: Never attempt to turn the power off for an


extended period of time without performing a complete
shutdown of the analyzer.

Note: In case the analyzer is not being used for several


E-3. Shutdown plug and containers
days only, it is not recommended to perform a complete
installed
shutdown, but to put the analyzer in Standby Mode.

To perform the shutdown, you will need the following


items:
• Two containers, one filled with at least 100 mL of
water, the other one empty.
• Four solenoid relief clamps (supplied with analyzer)
• Two reference electrode transport housings (one
supplied with the analyzer, the other supplied with
the shutdown kit)
• One shutdown plug (supplied with the shutdown kit)
• Protective strip for fluid pack

To get started, go to Operator Functions. Press YES,


E-4. Installing the second transport then NO until Take Out of Operation? appears.
housing Press YES.

Next the analyzer prompts Shutdown Plug and


Water Installed?. First remove the ISE Fluid Pack and
push the protective strip firmly onto the fluid pack ports.
Insert the shutdown plug carefully into the fluid pack
receptacle. Next, place the blue marked line into the
disposable container filled with water. The line with the
red mark is inserted into the empty container. Press YES.
See illustrations E-3.

The analyzer will prompt Transport Ref Housing


Installed? Pull the electrode holder forward and
remove the reference electrode assembly.

E-5. Reclamp electrode assembly

E-3
VetLyte Operator’s Manual – Operator Functions
Carefully unscrew the reference electrode from the
reference electrode housing. Temporarily place the
reference electrode on a clean, soft cloth. Next, pour the
remaining reference solution in the reference housing
into a transport housing. Then carefully screw the
reference electrode into the filled transport housing.

Now the second transport housing is screwed into the


reference housing. Place the reference housing back
into the electrode holder and move the lever backwards,
making sure all electrodes are seated properly. The
electrode holder remains in the forward position. Press
YES.
E-6. Blue line removed for purge cycle See illustrations E-4 and E-5.

The analyzer will take approximately one minute to


flush all lines with water. Upon completion, you will
receive the prompt Remove Water and Press Yes.
Remove the line with the blue mark from the water and
place it on a cloth, making sure it is not obstructed. The
line with the red mark remains in its container. Press
YES. During this cycle, all lines are purged of water.
Upon completion, the prompt All Electrodes & Plug
Removed? will appear.
See illustration E-6.

Unplug the reference connector below the left side of


the electrode holder. Then move the lever on the left side
forward and remove all electrodes, placing them on a
soft cloth. Next, move the lever back and push the empty
E-7. Shutdown plugs and electrodes
electrode holder into its back position.
removed
Unscrew the transport housing from the reference
housing. Empty the reference housing and place all
electrodes in their protective boxes. Remove the line
with the red mark from its container. Now the shutdown
plug can be removed. Press YES.
See illustration E-7.

The display will prompt All 4 Relief Clamps


Inserted? Install the red relief clamps making sure that
they are snapped securely into position. Press YES.
See illustration E-8.

Warning: Never insert the solenoid relief clamps


with the ISE Fluid Pack in place.

E-8. Installing red relief clamps

E-4
Operator Functions – VetLyte Operator’s Manual
The prompt Pump Windings Relieved? appears. First, grasp the front winding close to
the pump roller and gently pull it off the roller. Repeat the same procedure for the rear pump
winding. Press YES. The analyzer will ask Do You Wish to DELETE All Data?
Select NO.

Warning: If you press YES, conversion factors, date/time and ISE Fluid Pack volume
are reset to default.

Then the analyzer will prompt Shutdown Complete Turn Power Off. After the
power has been turned off, unplug the power cord and the cable from the receptacles. Close
the main door. Clean all external surfaces of the analyzer as well as the areas accessible
through the main door and the sample door.

For putting the analyzer back into operation, see Section B.

E-5
VetLyte Operator’s Manual – Operator Functions
Section F - Service Functions

Service Functions
To aid in testing and troubleshooting your VetLyte, there are various built-in functions
that you can access to evaluate the performance of the instrument. From the ANALYZE
SAMPLE? display, press NO until the prompt SERVICE FUNCTIONS? appears. Press YES.

Test Electrodes
You can test the voltage levels of the electrodes by using standard solutions or an external
sample. Press NO until the prompt Test Electrodes? is displayed. Press YES and the prompt
Test Standard A? will be displayed. Press YES and the voltages for each electrolyte will
appear. Record these values. Press NO to exit.

The prompt Test Standard B? will now be displayed. Press YES and the voltages for each
electrolyte will appear. Record these values and compare in this manner:

Electrode Standard Standard Allowable Difference


A B A-B
Na+ -600 to +2400 -1600 to +2200 +720 to +2000
K+ -700 to +1000 -2500 to +500 +520 to +1800
Cl¯ -3100 to -100 -1100 to +1800 -520 to -1200

After testing the electrodes against the standard solutions, the VetLyte Electrolyte Analyzer
will prompt Test External Sample? press NO to exit or to perform other service functions.

If you press YES, the instrument will prompt Open Sample Door Introduce Sample.
Press YES

Introduce the sample to the probe until the prompt Remove Sample/Wipe Probe is
received. After closing the sample door, the voltages will automatically appear. Check
for stable readings. Press NO to exit and after a short Please Wait message, the prompt
Remain in Test Electrodes? will appear. Press NO to exit or to perform other service
functions.

Test Sample Sensor


The sample sensor must provide a reading of 80 - 120 when air is detected, and should
indicate at least 40 higher when clear liquid is passed through the sensor.

For testing purposes, distilled water, Electrode Cleaning Solution A or Conditioning Solution
may be used.

To check the sample sensor, from the SERVICE FUNCTIONS? display, press YES and then
NO until Test Sample Sensor? appears. Press YES. The analyzer will display the normal
range and the measured value. The measured value is ideally around 100.

Pour a small amount of clear liquid into a clean sample cup. Press YES, open the sample door
and promptly introduce liquid into probe.

F-1
VetLyte Operator’s Manual – Service Functions
Wait a few seconds for liquid to be aspirated. Press NO to stop the pump. Wipe the probe
and close the sample door. The analyzer will display the normal range of 80-120; clear liquid
should read at least 40 units higher than the initial measured value. Press NO again to exit or
to perform other service functions.

Test Sample Door


To check the positioning of the sample door from the SERVICE FUNCTIONS? display, press
YES and then press NO until Test Sample Door? appears. Press YES.

Raise the sample door and the display should indicate that the door is open with an “O” in
the upper right corner of the display. Close the door and the “O” should change to a “C”
indicating the door is closed.

Test Peristaltic Pump


The peristaltic pump has four speeds of operation that can be automatically checked. From
the SERVICE FUNCTIONS? display, press YES and then NO until Test Pump? appears.
Open the front door and press YES.

The pump will progressively display Very Slow Speed, Slow Speed, Medium Speed
and Fast Speed, and an audible change should be heard in the pump’s speed of operation.
Close the front door and press NO to exit or perform other service functions.

Note: The pump speed is electronically controlled and should not require any adjustment.

Test Pinch Valves


The VetLyte Electrolyte Analyzer has four valves that function automatically during the
operation of the instrument.
See illustration F-1.

Open the analyzer front door. From the SERVICE FUNCTIONS? display, press YES and then
NO until Test Pinch Valves? appears. Press YES.

Test Valve A: YES/NO = ON/OFF will appear. When


you press the YES button, you should see the valve
move into the open position. The valve remains open for
approximately 30 seconds and then automatically closes.
R Pressing NO will close the valve. Pressing NO again will
V
move to the next successive valve. After testing all valves,
close the front door and press NO to exit or perform other
service functions.
B

F-1. Fluid control valves

F-2
Service Functions – VetLyte Operator’s Manual
Test Amplifier
This menu option is for service use only.

Enter Service Codes


The VetLyte can be programmed to increase the resolution of the measurement values by one
additional decimal place. It can also be programmed to report results in flame-equivalence
units.

To access either of these options, press NO until you reach the SERVICE FUNCTIONS?
display. Press YES. Press NO until Enter Service Code? appears. Press YES. Enter Code
will be displayed.

To increase resolution enter code DEC by selecting NO until desired letter is displayed.
Select YES to accept.

Note: VetLyte results printed through your VetTest will automatically use the additional
decimal place.

After entering the code, the analyzer will prompt Code Accepted, then it will display Enter
Another Service Code? press YES to do so, NO to exit.

To change reporting units from ISE to flame-equivalence, enter code ESI by selecting NO
until desired letter is displayed. Select YES to accept. After entering the code, the analyzer
will prompt Code Accepted, then it will display Enter Another Service Code? Press
YES to do so, NO to exit.

Note: If an incorrect code is entered during the process, the analyzer will prompt Code
Error Retry? Press YES to correct the code or NO to exit.

To remove a service code, enter the reversed sequence of the code letters (example: DEC
becomes CED and ISE become ESI).

F-3
VetLyte Operator’s Manual – Service Functions
Section G - Self Help Guide

Your VetLyte is designed to provide long, trouble-free service. However, any scientific
measuring device may occasionally malfunction, requiring the operator to identify the cause
of the problem.

The following information lists displayed messages and recommends steps that should
help you return your VetLyte to operation. If your VetLyte does not perform correctly after
conducting the basic steps outlined in this chapter, you should contact IDEXX for technical
support.

Self Help
STATUS: NOT CALIBRATED
This message will be displayed when calibration has been interrupted.

Possible remedies:
• Perform system calibration to return the instrument to the ANALYZE SAMPLE? mode.

STANDARD A NOT DETECTED


As the analyzer aspirates a sample of Standard A, the solution is detected by the sample
sensor, which must detect its presence without encountering any air bubbles. If the sample
sensor is unable to properly detect the presence of Standard A solution in a programmed time
period, the above error message will be displayed.

Possible remedies:
• Confirm the presence of the o-rings in the left side of all electrodes and in the right side
electrode holder. See illustrations B-4 and B-9.
• Replace the ISE Fluid Pack.
• Clean the sample sensor by performing Weekly Maintenance. Perform the sample
sensor test (see Section F, page 1). Ensure the sample sensor is securely plugged in. See
illustration B-9.
• Replace the peristaltic pump tube set to ensure correct aspiration of Standard A.
• Check for clots or crystals that may have formed in the Standard A tubing or the electrode
chamber. If clots or crystals are seen, contact technical support for assistance.

STANDARD B NOT DETECTED


The system for aspirating Standard B is like that of Standard A. If Standard A is aspirated
properly, but Standard B is not detected, check Standard B tubing for crystallization. If
Standard A is not aspirated either, check for leaks and blockages first.

Note: At the beginning of calibration, the analyzer checks for the presence of Standard
B first. Therefore, a missing o-ring or a blockage in the sample path will result in a
STANDARD B NOT DETECTED message.

G-1
VetLyte Operator’s Manual – Self Help Guide
CHECK SAMPLE SENSOR
The sample sensor is calibrated with air during each calibration. For proper functioning, the
sensor must provide a reading of 80 - 120 units when air is detected. For transparent fluids
(e.g. water) the reading should increase by at least 40 units. To check sample sensor response,
perform Test Sample Sensor (see Section F, page 1). Then perform a calibration to remove the
error message.

Possible remedies:
• Clean sample sensor by using the Weekly Maintenance procedure. Ensure the sample
sensor is properly plugged in. See illustration B-9.
• Replace the peristaltic pump tube set.

CHECK REFERENCE HOUSING


When the analyzer fails to detect a flow of Reference Solution into the measuring chamber,
CHECK REFERENCE HOUSING will be displayed. This test is performed at the beginning of
each calibration cycle.

Possible remedies:
• Check the double reference tubing for large air bubbles. If seen, repeat calibration.
• Ensure that the double reference tubing is securely connected to the receptacle.
• Clean the reference housing (See Performing Monthly Maintenance in Section D).
• Replace the ISE Fluid Pack.
• Replace the peristatic pump tube set.

PLEASE CLOSE SAMPLE DOOR


This message occurs when the sample door is not closed within 20 seconds after the sample is
in place, or when the door has been opened and no sample is detected.

Possible remedy:
• Close the sample door within time allowed.

NO SAMPLE
Under certain conditions, the analyzer may not detect the presence of a sample and an error
message will occur. This could be caused by the sample door remaining open more than 20
seconds after sample is introduced or prompts are not followed. Also check for air bubbles in
the sample, a highly lipemic or viscous sample, a sample volume too small to analyze, or no
sample being aspirated through the sampling mechanism.

Possible remedies:
• First, repeat the sample to see if it is detected properly on a second trial.
• Close the sample door within time allowed.
• When you open the door, wait for the prompt Introduce Sample?

G-2
Self Help Guide – VetLyte Operator’s Manual
• Check sample aspiration and look for the presence of clots in the sampling system.
• Check for the presence of o-rings and for proper sealing of the electrodes.
• Ensure that the sample sensor is plugged in and perform TEST SAMPLE SENSOR to
verify that it is operating correctly (see Section F, page 1).
• Replace pump windings.

CLEANING FLUID NOT DETECTED


Under certain conditions, the analyzer may not detect the presence of cleaning solution.
This could be caused by prompts not being followed, air bubbles in the cleaning solution,
too small a volume of cleaning solution, or the cleaning solution not being aspirated
properly through the sampling mechanism.

Possible remedies:
• When you open door, wait for the prompt Introduce Sample.
• Check for the presence of o-rings and for proper sealing of the electrodes.
• Check for aspiration of cleaning solution and look for presence of clots in the sampling
system.
• Check for the proper sealing of the pump windings.
• Ensure that the sample sensor is plugged in and perform Test Sample Sensor to verify
that it is operating correctly (see Section F, page 1).

CONDITIONING FLUID NOT DETECTED


Under certain conditions, the analyzer may not detect the presence of conditioning solution.
This could be caused by prompts not being followed, air bubbles in the conditioning solution,
too small a volume of conditioning solution, or the conditioning solution not being aspirated
properly through the sampling mechanism.

Possible remedies:
• When you open door, wait for the prompt Introduce Sample.
• Check for the presence of o-rings and for proper sealing of the electrodes.
• Check for aspiration of conditioning solution and look for presence of clots in the
sampling system.
• Check for the proper sealing of the pump windings.
• Ensure that the sample sensor is plugged in and perform Test Sample Sensor to verify that
it is operating correctly (see Section F, page 1).

CHECK ELECTRODES
If none of the electrodes passes the calibration, the Check Electrodes message will be
displayed.

Possible remedies:
• Ensure that electrodes are properly plugged in. See illustrations B-7 through B-11.
• Perform Monthly Maintenance.
• Replace the ISE fluid pack.

G-3
VetLyte Operator’s Manual – Self Help Guide
CLOG IN SAMPLE PATH
If the unit is unable to clear the sample path, the message CLOG IN SAMPLE PATH will be
displayed.

Possible remedies:
• Make sure that the electrode o-rings are present and seated properly.
See illustration B-4.
• Perform weekly cleaning.
• Check the ISE Fluid Pack volume; if less than 5% is remaining, replace pack.

↑↑↑↑ ↓↓↓↓
In case the unit displays arrows up or arrows down instead of the sample results, the
concentration of the sample is outside the measurement range (see specifications).

Possible remedies:
• Check for proper sample preparation (see Section C, page 1).
• Rerun sample. For high results, dilute sample 1:1 with sterile water and repeat.

Warning: Do not use saline for dilutions on the VetLyte.

ERR.
If the unit displays ERR. instead of the sample results, no valid voltage reading could be
obtained from the electrode.

Possible remedies:
• Check for proper sample preparation (see Section C, page 1).
• Ensure electrodes are properly plugged in.
See illustrations B-7 through B-11.
• Check for air bubbles in the sample.

G-4
Self Help Guide – VetLyte Operator’s Manual
Section H - Principles of Operation

The Measurement Principle


Your VetLyte Electrolyte Analyzer is a sophisticated instrument that uses the Ion Sensitive
Electrode (ISE) measurement principle to precisely determine electrolyte values. Although
the technology itself is quite complicated, understanding how the instrument performs
sampling analysis is relatively simple. Basically, the analyzer compares an unknown value
against a known value to compute the sample’s electrolyte level.

An ion-sensitive membrane undergoes a specific reaction with the type of electrolyte


contained in the sample. The membrane is an ion exchanger, reacting to the electrical change
of the ion causing a change in the membrane potential, or measuring voltage, which is built
up in the film between the sample and the membrane.

A galvanic measuring chain within the electrode determines the difference between the
two potential values on either side of the membrane. The galvanic chain is closed through
the sample on one side by the reference electrode, the reference electrolyte and the “open
terminal.” The membrane, inner electrolyte and inner electrode close the other side.
See illustration H-1.

The different ion concentrations between the inner electrolyte and the sample causes an
electrochemical potential to form on the membrane of the active electrode. The potential
is conducted by the inner electrode to the input of an amplifier. The reference electrode is
connected to ground as well as the second input of the amplifier.

Since the reference electrode is on ground, amplification of the electrode potential allows
further signal processing.

The ion concentration in the sample is then determined and displayed by using a calibration
curve determined by two measured points of standard solutions with precisely known ion
concentrations (two-point calibration), and by using the measured voltage of the sample and
the Standard A (one-point calibration).

Physical Principle
An ion-selective electrode is connected with a reference electrode to form a measuring
system. When immersed in a solution that contains the relative ion, the Nernst equation
applies:
R · T
1. E = E’± _______ · In ai
n · F
or
R · T
2. E = E’± _______ · In (fi · ci) (+) for cations,
n · F (-) for anions

H-1
VetLyte Operator’s Manual – Principles of Operation
The equation can also be written:

3. E=E’ ± S · log (fi · ci)

E the measured electric potential


E’ the e.m.f. of the system in a standard
solution
ai activity of the ion measured
R the general gas constant (8.31 J/Kmol)
T temperature
n valence of the measured ion
F Faraday constant 96.496 A.s/g
equivalent
fi the activity coefficient
ci the concentration of the measured ion
S the slope of the electrode
See illustration H-2.

If the ion concentration of one measuring solution is known, the ion concentration of the
sample can be determined on the basis of the difference of two measured potentials.

4. Esample = E’ + S · log (fi · ci sample)

5. Estandard = E’ + S · log (fi · ci standard)

ci sample
6. E = E sample - Estandard = S · log
ci standard

E the difference between the measured


potentials of the sample and the standard
S the potential difference of the electrode,
determined from the potential
difference of two measured standard
solutions
ci sample concentration of the measured ions in
the sample
ci standard concentration of the measured ions in
the standard solutions
The unknown ion concentration in the sample can now be determined by:

7. ci sample = ci standard · 10 ( E/S)

As demonstrated by these equations, the ion selective electrodes do not measure the ion
concentration but the activity of the ions concerned. This activity is a criterion of the ion’s
ability to interact with other ions, in which each ion binds a proportion of its energy.

H-2
Principles of Operation – VetLyte Operator’s Manual
The ion concentration is calculated on the basis of the ion activity. The correlation is affected
by the total number of ions in the solution. Because sodium in whole blood and serum is the
predominant ion, the known value of the sodium concentration makes it possible to ascertain
and adjust for total ion effect and strength.

Voltmeter

Inner
Reference Electrode
Solution

Ion-Selective Electrode
Current Terminal
Ion-Selective
Membrane

Measured solution

H-1. Measurement principle

Amplifier Display
Ground
Na+
Reference
Reference
electrode housing K+
electrolyte (KCl) Na+ electrode K+ electrode Cl-
Reference
electrode Cl- electrode
Inner electrode
Glass capillary
Inner electrolyte

Sample
Sodium sensitive Potassium sensitive Chloride sensitive
membrane membrane membrane

H-2. Measurement system

H-3
VetLyte Operator’s Manual – Principles of Operation
O-ring seal
Sodium Electrode Specifications
electrolyte chamber
The Sodium Electrode is a glass capillary electrode
used for in-vitro diagnostic measurement of sodium
ions present in fluid samples. See illustration H-3. It is
inner electrode designated with a Na+ marking on the top surface of the
housing.

Construction
glass
capillary Electrode housing: transparent acrylic plastic
serial number connecting pin Measuring capillary: sodium sensitive glass
Electrolyte chamber: airtight, filled with electrolyte
H-3. Sodium electrode construction solution for Na+ electrodes
Pin connector: silver, silver chloride
(Ag/AgCl)

Use and Care


VetLyte Sodium Electrodes are manufactured for use in
VetLyte Electrolyte Analyzer.

Proper care should be used in handling and storage of


the electrode. Never use strong or abrasive cleaners such
as alcohol or amphyl on the electrode. These will attack
the plastic housing.

Store the electrode in a clean, dry place only after the


electrode has been cleaned and rinsed with distilled
water and dried with a lint-free cloth. The o-ring seal
should be installed in the electrode during storage.

H-4
Principles of Operation – VetLyte Operator’s Manual
ion sensative Potassium Electrode Specifications
membrane
The Potassium Electrode is a membrane electrode used
for measurement of potassium ions present in fluid
O-ring seal

samples. See illustration H-4. It is designated with a K+


marking on the top surface of the housing.
electrode
chamber

Construction

glass
Electrode housing: transparent acrylic plastic
capillary Measuring membrane: potassium-ion sensitive
serial number inner
connecting pin
Electrolyte chamber: airtight, filled with electrolyte
electrode
solution for K+ electrodes
H-4. Potassium electrode construction Pin connector: silver, silver chloride
(Ag / AgCl)

Use and Care


VetLyte Potassium Electrodes are manufactured for use
in VetLyte Electrolyte Analyzer.

Proper care should be used in handling and storage of


the electrode. Never use strong or abrasive cleaners such
as alcohol or amphyl on the electrode. These will attack
the plastic housing.

Store the electrode in a clean, dry place only after the


electrode has been cleaned and rinsed with distilled
water and dried with a lint-free cloth. The o-ring seal
should be installed in the electrode during storage.

H-5
VetLyte Operator’s Manual – Principles of Operation
ion sensative Chloride Electrode Specifications
membrane
The Chloride Electrode is a membrane electrode used for
measurement of chloride ions present in fluid samples.
O-ring seal

See illustration H-5. It is designated with a Cl- marking


on the top surface of the housing.
electrode
chamber

Construction
Electrode housing: transparent acrylic plastic
measuring
capillary Measuring membrane: chloride-ion sensitive
serial number inner Electrolyte chamber: airtight, filled with electrolyte
solution for Cl- electrodes
connecting pin
electrode

H-5. Chloride electrode construction Pin connector: silver, silver chloride


(Ag / AgCl)

Use and Care


VetLyte Chloride Electrodes are manufactured for use in
VetLyte Series Electrolyte Analyzer.

Proper care should be used in handling and storage of


the electrode. Never use strong or abrasive cleaners such
as alcohol or amphyl on the electrode. These will attack
the plastic housing.

Store the electrode in a clean, dry place only after the


electrode has been cleaned and rinsed with distilled
water and dried with a lint-free cloth. The o-ring seal
should be installed in the electrode during storage.

H-6
Principles of Operation – VetLyte Operator’s Manual
entrance of ref. exit of ref. Reference Electrode Assembly Specifications
The Reference Electrode Assembly is a device used as
electrolyte electrolyte
glass
capillary an electrical junction between the sample and electrical
ground.

Construction
The Reference Electrode Assembly consists of two parts:
O-ring seal (c) the Reference Electrode Housing and the Reference
Electrode.
O-ring seal (b)
O-ring seal (a) serial number
Reference Electrode Housing
In the reference electrode housing, reference electrolyte
H-6. Reference electrode housing solution establishes the electrical contact between the
reference electrode and the sample. At the beginning of
each measurement, reference electrolyte is pumped into
the housing. At the same time a glass capillary allows
a small amount of reference electrolyte to pass into the
measuring capillary, thus establishing electrical contact
between the sample and the Reference Electrode.
See illustration H-6.

Use and Care


VetLyte Reference Electrode Housings are manufactured
for use in VetLyte Electrolyte Analyzer.

Proper care should be used in handling and storage of the


electrode housing. Never use strong or abrasive cleaners
such as alcohol or amphyl on the housing.

Store the Reference Electrode Housing in a clean,


dry place only after the Reference Electrode has been
removed and the housing cleaned and rinsed with
distilled water and dried with a lint-free cloth.

The o-ring seals should be installed in the housing


during storage.

H-7
VetLyte Operator’s Manual – Principles of Operation
Reference Electrode
The Reference Electrode completes the electrical circuit
platiunum wire cable housing between the reference electrolyte and electrical ground.
glass tube
This is accomplished by a cotton wool (saturated with
reference electrolyte)-calomel (Hg2Cl2)- mercury (Hg)-
platinum wire-connecting pin junction.
See illustration H-7.
cotton wool
calomel Use and Care
(Hg 2Cl2)
mercury (Hg) VetLyte Reference Electrodes are manufactured for use
in VetLyte Electrolyte Analyzer.

H-7. Reference electrode Proper care should be used in handling and storage of
the Reference Electrode. Never use strong or abrasive
cleaners such as alcohol or amphyl on the electrode.

Store the Reference Electrode in the transport housing


provided with electrode at the time of purchase. Make
sure that the transport housing is filled with reference
electrolyte solution (you may use the reference
electrolyte solution remaining in the Reference Electrode
Housing at the time of disassembly).

Warning: Never rinse the Reference Electrode with


distilled water. Never allow the reference electrode to
become dry.

H-8
Principles of Operation – VetLyte Operator’s Manual
Technical Specifications

Electrodes
Sodium (Na+) Sensor ion selective, flow-through, glass capillary electrode
Potassium (K ) Sensor
+
ion selective, flow-through, liquid membrane electrode
Chloride (Cl¯) Sensor ion selective, flow-through, liquid membrane electrode
Reference System open liquid junction, flow-through, electrode

Measuring Ranges (mmol/L) Range


Whole blood, serum, plasma Na resolution 0.1 mmol/L
+
40 - 205
K+ resolution 0.01 mmol/L 1.5 - 15
Cl¯ resolution 0.1 mmol/L 50 - 200

Reproducibility (within run) Na+ CV 1.0% @ 140 - 160 mmol/L


K+ CV 1.5% @ 4 - 6 mmol/L
Cl¯ CV 1.0% @ 90 - 130 mmol/L

Operating Parameters
Sample size 95 μL
Sample type serum, plasma, anticoagulated whole blood
Sample application sample cup and collection tube
Analysis time (approx.) 30 sec.
Calibration fully automatic 1- and 2-point calibrations
Diagnostic programs user-controlled diagnostics, plain language display
Electronics microprocessor-controlled
Display dot-matrix, 2 lines, 16 characters per line
Computer interface RS232C serial port
Power requirements 90 - 250 VAC 50/60 Hz
(self-adjusting) 1.4 amp max.
375 watts max.
Temperature room temperature, 15˚ - 32˚ C ; 60˚ - 90˚ F
Humidity < 85% relative humidity, non-condensing
Dimensions (HxWxD) 13.2 x 12.4 x 12.0”; 335 x 315 x 295 mm
Weight (approx.) 13 lbs.; 6 kg

Data subject to change without notice. Technical information is supplied for general informational
purposes only.

I-1
VetLyte Operator’s Manual – Appendix: Technical Specifications
Appendix B

Program flow chart

NO NO NO NO
NO
ANALYZE View Last ELECTRODE OPERATOR SERVICE CALIBRATION?
STANDBY NO FUNCTIONS?
SAMPLE? 12:00 Sample Results? MAINTENANCE? FUNCTIONS?
NO YES YES YES YES YES

LEAVE STANDBY Perform Weekly Check/Change Test CALIBRATE ...


MODE? Cleaning? Reagent Pack? Electrodes?
Canine?
NO NO NO NO
YES
NO
Perform Daily Go to Test Sample
STANDBY Conditioning? Standby Mode? Sensor?
Feline? NO
NO NO

NO
Remain in Set Test
YES Maintenance? Time/Date? Sample Door?
Equine?
NO NO NO

NO
NO Are View Calibration Test
You Data? Pump?
Other? Sure?
NO NO
YES
NO
CALIBRATE ... Change Baud Test Pinch
Rate? Valves?
Next Sample?
NO NO NO
YES
Take Out of
Operation? Test Interface?
Jump Pins 2 & 3
NO NO

Remain in
YES Operator Func? Test
Amplifier?
NO
NO

Enter/Verify
Service Code?
NO

Remain in
YES Service Func?
NO

J-1
VetLyte Operator’s Manual – Appendix: Program Flow Chart
Appendix C
Maintenance log master

Note: You may use this page as a master for duplicating.

VETLYTE MAINTENANCE SCHEDULE

MONTH OF __________________

1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31
DAILY
Perform Daily Conditioning

WEEKLY
Perform Weekly Cleaning
Clean Sample Probe and Fill Port
Clean Analyzer Sur faces

MONTHLY
Clean Reference Electrode Housing Date __________ Initial __________
(check)

EVERY 6 MONTHS
Replace Peristaltic Pump T ubing Date __________ Initial __________
(check)

UNSCHEDULED MAINTENANCE
Fluid Pack Date __________ Initial __________
(check)

Electrode Replacement Type_________ Date __________ Initial __________


(check)

Other ____________________

VetLyte

K-1
VetLyte Operator’s Manual – Appendix: Maintenance Log Master
Warranty

IDEXX LABORATORIES, INC.


Limited Warranty for VetTest, VetLyte, VetCom and
QBC® VetAutoread™ System

IDEXX Laboratories, Inc. (“IDEXX”) warrants this instrument to be free from defects in
materials and workmanship, when stored under appropriate conditions and given normal, proper
and intended usage, for one year from the date of delivery of the product to the original end user
purchaser or, in the case of computer software, licensee (“Buyer”). IDEXX agrees during the
applicable warranty period to repair or replace all defective products within 30 days after the
date of receipt by IDEXX of such defective products and without cost to Buyer. IDEXX shall
not have any obligation under this Limited Warranty to make repairs or replacements which
result, in whole or in part, from catastrophe, fault or negligence of the Buyer, or anyone claiming
through or on behalf of the Buyer, or from improper use of the products, or use of the products in
a manner for which they were not designed, or by causes external to the products. In addition, this
Limited Warranty shall not apply to expendable components, such as, but not limited to, fuses and
bulbs, which are incorporated or sold for use in instruments.

Buyer shall notify IDEXX of any products which it believes to be defective during the
warranty period. At IDEXX’s option, such products shall be returned by Buyer, transportation
and insurance prepaid, to IDEXX’s designated facility for examination and testing. IDEXX
shall repair or replace, within 30 days of receipt by IDEXX, any such product found to be so
defective and promptly return such repaired or replacement product to Buyer, transportation and
insurance prepaid. Should IDEXX’s examination and testing not disclose any defect covered
by the foregoing warranty, IDEXX shall so advise Buyer and dispose of or return the product in
accordance with Buyer’s instructions and at Buyer’s sole expense.

THE PROVISIONS OF THE FOREGOING LIMITED WARRANTY ARE EXCLUSIVE AND


ARE EXPRESSLY IN LIEU OF ANY OTHER WARRANTY, WHETHER EXPRESS OR
IMPLIED, WRITTEN OR ORAL (INCLUDING ANY WARRANTY OF MERCHANTABILITY
OR FITNESS FOR A PARTICULAR PURPOSE). IDEXX NEITHER ASSUMES NOR
AUTHORIZES ANY EMPLOYEE, AGENT, DISTRIBUTOR OR OTHER PERSON OR
ENTITY TO ASSUME FOR IT ANY OTHER LIABILITY IN CONNECTION WITH THE
MANUFACTURE, SALE, SUPPLYING OR USE OF THE PRODUCTS. IDEXX’S LIABILITY
ARISING OUT THE MANUFACTURE, SALE OR SUPPLYING OF THE PRODUCTS OR
THEIR USE OR DISPOSITION, WHETHER BASED UPON WARRANTY, CONTRACT,
TORT OR OTHERWISE, SHALL NOT EXCEED THE ACTUAL PURCHASE PRICE
PAID BY BUYER FOR THE PRODUCTS. IN NO EVENT SHALL IDEXX BE LIABLE
TO BUYER OR ANY OTHER PERSON OR ENTITY FOR SPECIAL, INCIDENTAL OR
CONSEQUENTIAL DAMAGES (INCLUDING, BUT NOT LIMITED TO, LOSS OF PROFITS,
DATA OR USE DAMAGES) ARISING OUT OF THE MANUFACTURE, SALE, SUPPLYING
OR USE OF THE PRODUCTS. THE FOREGOING WARRANTIES EXTEND TO BUYER
ONLY AND SHALL NOT BE APPLICABLE TO ANY OTHER PERSON OR ENTITY
INCLUDING, WITHOUT LIMITATION, CUSTOMERS OF BUYER.

L-1
VetTest Operator’s Manual – Appendix: Warranty

You might also like