You are on page 1of 1

reREGD.

OFFICE & LAB:


Shop No. 3,4,6 & S,Pogoriyo - Nelge Ploze,
Pune - Noshik Highwoy,Bhosori,Pune - 4li 039.
Cell : 9860103446 I 9371151622 / 8459886810
E-moil : moruticolibrotions@gmoil,com / moruticolibrotions@yohoo,co.in

ffiffi
No. : MCU 2023-24164014 Certificate lssue Date : 1810312023
'l8l03l2o23 iNextw
ULR-CC251023000004849F i(As per Customer Requirement)
64014 Page No. : 1 of 1

18t03t2023
MSM PROCESS SOLUTIONS PW.LTD.
GAT NO.67,70,Sonwane Wasti,Jyotiba Nagar,
Talwade,Pune-412114.

Description of Unit Under Calibration (UUC)


of UUC : Pressure Gauge Range 0 to 20 kg/cm2
ldentification No. : M333.59-01391 L.C. 0.5 kg/cm'?
Make t1%fsd

Condition of UUC at Receipt: OK

Descriptron of Calibration Method Used


MCL Calibration Method No. : MCL-CP-41
Reference Standard Used : DKD-R-6-1 & lS 3624
Calibration Purpose : To check error in readings by comparison master digital pressure gauge within total range
& at regular intervals.
Uncertainty of Measurement (At 95.45o/o confidence level) : t2.S kg/cm2
Environmental Conditions : Temperature: 25"Ct4"C

Standards & Master Equipment Used & Their Traceability


Reference Standard / Equipment Used lD. No. Traceability (NABL
Certificate No.. Valid Upto
Lab No,l
Digital Pressure Gauge MCURS/39 cc-2294 Nv2207t076t001 22t07t2023
Traceability: Traceable to National / lnternational star,Jards through NABL Accredited calibration laboratory.

Calibrqtion Results: All readings are in kg/cm2


Set Value on Observed Value on Master kg/cm, Error in Repeatability=
T.r.'l:1".:':=
Uncertainty t in

UUC' kg/cm'z uP (M1) Down (M2) uP (M3) Mean Value


kg/cm'z kg/cm2 kg/cm2 kg/cm2
((M1+M3\12+M2Y21
0 0.00 0.00 0.00 0.00 0.00 0.00 0.00 2.50
5 5.10 5.09 5.11 5.10 -0.10 0.01 0.01 2.50
10 10.11 10.10 10.12 10.11 -0.11 0.01 0.01 2.50
15 15.13 15.12 15.14 15.13 -0.13 0.01 0.01 2.50
20 20.15 20,'14 20.16 20.15 -0.15 0.01 0.01 2.50
.UUC: Unit Under Calibration
)L-REC-45-04

Shivdas K. H.\)ty'
Engffi
oitiprki
Qualityfi/anager
Calibration

o
d,
The Calibration results reported in this certificate are the time of and under the conditions of measurement stated above.
o This Certificate should not be reproduced without our permission in writing.
o This Certificate refers to the above mentioned item only.