Certificate No. : MCU 2023-24164012 Certificate lssue Date : 1810312023


Date Date :

Code ULR-CC251023000004847F (As per Customer Requirement)


Order No. 64012 Page No. : 1 of 1
Date of Receipt:
Customer Name MSM PROCESS SOLUTIONS PW.LTD.
Customer Address GAT N0.67,70,Sonwane Wasti,Jyotiba Nagar,
Talwade,Pune-412114.

Description of Unit Under Calibration (UUC)


Name of UUC Pressure Gauge lnange : 0 to 10 kg/cm2
Identification No. M364.59-02573 lu.c. : 0.2 kg/cm2
Make Baumer AccuracY : t1%fsd
Type 41alog Location :

Condition of UUC at OK

Description of Calibration Method Used


No.
tvtCl Calibration Method : MCL-CP-41
Used
Reference Standard : DKD-R-6-1 & lS 3624
Calibration purpose : To check error in readings by comparison master digital pressure gauge within total range
& at regular intervals.
Uncertainty of Measurement (At 95.45% confidence level) : t2.5 kg/cm2

Standards & Master Equipment Used & Their Traceability


Traceability (NABL Certificate No.. Valid Upto
Reference Standard / Equipment Used lD. No.
Lab No.)

Digital Pressure Gauge MCURS/39 cc-2294 NV2207t0761001 22t0712023

Traceability: Traceable to National / lnternational standards through NABL Accredited calibration laboratory.

Galibration Results: All readings are in kg/cm2


i in
RcprEtsbility= Hyst€resis= Uncertainty
Set Value on Observed Value on Ma';ter kg/cm'z Error in
lM3-Mll in lM2-M1lin
Mean Value kg/cm'a kg/cm'?
UUC' kg/cm'z uP (Ml) Down (M2) uP (M3) ({(M I +M3)/2+M2}/21
kg/cm'z kg/cm'z

0 0.00 0.00 0.00 0.00 0.00 0.00 0.00 2.50


2 2.05 2.04 2.06 2.05 -0.05 0.01 0.01 2.50
4 4.07 4.06 4.08 4.07 -0.07 0.01 0.01 2.50

6 6.12 6.11 6.13 6.12 -o.12 0.01 0.01 2.50


10 10.14 10.13 10.15 10.14 -0.14 0.01 0.01 2.50
*UUC: Unit Under Calibration MCL-REC-45-04

oitipx[kaai
Quality ffinager
"",i'?,IflT:^iW
v
D The Calibration results reported in this certificate are the time of and under the conditions of measurement stated above.
D This Certificate should not be reproduced without our permission in writing.
T This Certificate refers to the above mentioned item only.