You are on page 1of 2

33.2.

35 room temperature; use these each time fp value determinations


AOAC Official Method 961.07 are made. If it is suspected that salt secondary standards are in
Water (Added) in Milk error, check them against freshly pre pared su crose pri mary
Thermistor Method standards. It is responsibility of the analyst to be certain that fp
First Action 1961 values of salt secondary standards are same as fp values of freshly
Final Action 1961 prepared sucrose primary standards.

IDF–ISO–AOAC Method C. Calibration of Cryoscope


(See introduction to 922.07 [see 33.2.34].)
A. Cryoscope
Using calibration controls and fp values of standards, calibrate
Consists of cooling bath, sample agitator, seeding rod, thermistor
cryoscope to obtain correct “span” (0.600 - 0.408 = 0.192°C) and
probe (electric resistance thermometer), and Wheatstone bridge and
galvanometer, or taut band meter measuring circuit. Bath may be reference values (-0.600°C and -0.408°C). Follow directions in
cooled by mechanical or electrical means or by insulated ice–salt man u fac turer’s op er at ing man ual. Cal i bra tion con trols and
mixture. Test sample may be immersed in cooling bath or bath may procedures vary with make and model of cryoscope but with all
be “brought up” to test sample mechanically. Observed fp value is instruments 2 calibration controls (A and B or I and II) are adjusted,
read from measuring dial, calibrated in millidegrees C (0.001°C; individually or in combination, so that 7 and 10% sucrose primary
more correctly, degrees Hortvet or “H”), when galvanometer or standards and/or sucrose equivalent salt secondary standards yield
meter is nulled by rotation of dial. fp values of -0.408° and -0.600°C, respectively, with 0.192°C span.
Fill cooling bath at room temperature to proper level each time D. Determination (Final Action 1974)
instrument is used. Proper coolant level is determined by observing [If titratable acidity, 947.05 (see 33.2.06), is >0.18%, results may
coolant issuing from bath overflow or by visually checking coolant underestimate actual amount of added H2O in test portion.]
level, depending on the make and model of cryoscope. Cooling bath
Apply following technique in exactly same manner for both
temperature should be -7° or -8° ± 0.5°C, depending on make of standards and test sample to obtain valid milk fp value.
cryoscope, and temperature is measured by placing thermometer in
Check cooling bath level, cooling bath temperature, stirring
empty test sample well.
efficiency, and probe position in test portion tube as in A.
Locate thermistor probe, both horizontally and vertically, at
Check reference fp values and “span,” using salt secondary
midpoint of test sample. Check location visually, using 2.5 mL H2O
standards or sucrose primary standards. If “span” is other than
in test sample tube.
0.192°C, recalibrate cryoscope as in C. If “span” is correct but
Amplitude of stirring wire should be great enough to assure
reference fp values differ from known values of standards, it is not
uniform temperature in test sample throughout determination and
necessary to recalibrate cryoscope; simple arithmetic correction
may be checked visually, using 2.5 mL H2O in test sample tube to
will give correct observed fp value of test portion.
which small amount of dust, powder, or dye has been added.
Using clean, dry syringe or pipet, measure 2–3 mL test portion
(Apparatus and supplies available as Advanced Cryoscope
and transfer to clean, dry test portion tube supplied by manufacturer.
[current Model 4D3] from Advanced Instruments, Inc., Two
Set measuring dial to expected fp value. Place test portion tube in
Te c h n o l o g y Wa y, N o r w o o d , M A 0 2 0 6 2 , U S A ; Te l :
cooling bath test portion well or in operating head and lower
+1-800-225-4034; www.aicompanies.com; as Precision Cryoscope
operating head to position test portion in cryoscope cooling bath.
16 Tech Cir, Natick, MA 01760, USA; and as Fiske Cryoscope from
Advanced Instruments, Inc., Fiske Division.) Cool test portion if not already being cooled above. Proper
cooling is indicated by rapid and uniform (steady) movement of
B. Preparation of Standards light spot or needle from right to left over scale of galvanometer or
(Use distilled H2O recently boiled and cooled to 20°C for meter.
preparation of standards.) If cryoscope raises cooling bath to test portion level, begin slow
Prepare following sucrose primary standards or salt secondary cooling at -1.5° to -2.0°C, depending on extent of supercooling
standards. Determine fp values of salt standards as in (b). desired. If cryoscope immerses test portion in cooling bath, do not
(a) -0.408°C (-0.422°H) Standard.—(1) Weigh 7.0000 g NIST isolate test portion in air above cooling bath.
standard sucrose, SRM 17, into 100 mL volumetric flask and dilute Seed test portion at -2.0° or -3.0°C, depending on make of
to volume with H2O; or (2) weigh 100 g H2O into 100 mL volumetric cryoscope. Galvanometer spot or meter needle will jump to right as
flask and add 0.6889 g reagent grade NaCl (dried to constant weight temperature of supercooled test portion rises toward fp. Extent of
just before weighing). supercooling (seeding point) must be same for sucrose primary
(b) -0.600°C (-0.621°H) Standard.—(1) Weigh 10.0000 g NIST standards or salt secondary standards as for milk test portions.
sucrose into 100 mL volumetric flask and dilute to volume with Adjust galvanometer spot or meter needle to 0 if necessary.
H2O; or (2) weigh 100 g H2O into 100 mL volumetric flask and add Switch galvanometer or meter to high sensitivity position. With
1.0207 g reagent grade NaCl. temperature dial, keep galvanometer or meter nulled (reading 0).
(Secondary salt standards with fp values equivalent to 7 and 10% Galvanometer spot or meter needle will cease to move to right,
sucrose primary standards may be purchased from cryoscope remain steady at 0, and finally begin to move to left. Read fp value
manufacturers.) from measuring dial to nearest millidegree, while spot or needle is
Mi cro or gan isms at tack su crose af ter lim ited stor age at steady just before movement to left begins. Do not read fp value
re frig er a tor tem per a tures, chang ing fp value of pri mary from measuring dial at some predetermined time after seeding test
standards. Salt (sucrose equivalent) secondary standards, stored portion; always wait for movement of galvanometer spot or meter
in poly ethylene bottles with screw caps, have long shelf life at needle to left before recording fp value. Spot or needle will become
2006 AOAC INTERNATIONAL

ã 2005 AOAC INTERNATIONAL


steady and begin to move to left sooner with standards than with sample and laboratory sample to be confirmed. If fps differ by
milk test portions. Check 0 point of galvanometer or meter. £0.010°C, laboratory sample is confirmed as H2O-free.
To confirm pooled herd milk as “added H2O” or “H2O-free,”
E. Interpretation
determine fp of authentic laboratory samples of all herd milks
If fp is -0.508°C (-0.525°H) or below, milk may be presumed to composing pooled herd milk, calculate weighted average fp for the
be H2O-free or may be confirmed as H2O-free by tests specified authentic laboratory samples of herd milk, and compare values as
be low. If fp is above -0.508°C, milk will be des ig nated above.
“presumptive added H2O” and will be confirmed as “added H2O” or To confirm processed milk as “added H2O” or “H2O-free,”
“H2O-free” by tests specified below. Evaluate extreme daily determine fp of test samples of all pooled herd milk received by
fluctuations in fp of herd, pooled herd, or processed milk for processing plant, calculate weighted average fp for milk received,
presence of added H2O. and compare with fp value to be confirmed. If fps differ by
To confirm herd milk as “added H2O” or “H2O-free,” determine £0.010°C, processed test sample is confirmed as H2O-free during
fp of authentic laboratory sample of herd milk obtained £ 72 h after processing. If ³1 test sample of pooled herd milk is “presumptive
laboratory sample to be “confirmed.” Authentic laboratory sample added H2O,” proceed as above for pooled herd milk. Fp for
is sample of milk from 1 complete, supervised herd milking (either pasteurized-homogenized milk should be same as that of pooled
AM or PM but beginning not <11 or >13 h after beginning of herd milk unless processing includes vacuum pasteurization, which
previous milking) obtained from bulk tank after entire herd has been raises the freezing point approximately 0.005°C.
milked through approved, properly sanitized, and thoroughly References: JAOAC 44, 438(1961); 51, 816(1968);
drained milking system into empty bulk tank but before rinsing or 53, 539(1970); 55, 410, 504(1972).
washing of system has begun. Compare fp of authentic laboratory ISO 5764:2002.

2006 AOAC INTERNATIONAL

ã 2005 AOAC INTERNATIONAL

You might also like