7.00 pH
Aqua Calibration and sensor offset adjustment from the keypad, additional
magnetic key for programming
Large dual line display
SMARTPro® Automatic temperature compensation (ATC)
Manual temperature compensation setting without ATC probe, with
Transmitters independent setting for calibration and process temperature
One galvanically isolated current output 4…20mA with HART
communication protocol
Protection against electromagnetic interference
Three galvanic isolated open collector output for over-range, under-range
and error detection in non-Ex version.
Dimension Drawing
102
4.01
315
12.40
120
4.72
120 35 20 147
4.72 1.37 0.78 5.78
285
Millimeter 11.22
Inch
All dimensions ±5 mm
General Specifications
Display
Local display Two line dot matrix with 10 characters per line LCD display
Keypad
Keypad Three button keypad and magnetic pin programming with user friendly
menu interface.
Output
Current output 4-20 mA DC, galvanically isolated output with accuracy better than
0.3% of full scale
Open collector outputs Three open collector outputs with max load of 100mA @ 24VDC,
(for safe area version only) galvanically isolated from the sensor and the output current for:
Errordetection
Over range detection
Under range detection
Note: External relay option available for potential free contact requirement
Power Requirements
Mechanical Specifications
Housing material Die-cast aluminium (LM6) with Polyurethane (PU) paint (non-corrosive).
Protection class IP 67
Environmental Conditions
pH
Range 0 – 14 pH
Resolution 0.01 pH
Accuracy +/- 0.01 pH
Span Minimum 2pH to maximum 14pH (programable)
mV (ORP)
Range +/- 2000 mV
Resolution 1 mV
Accuracy +/- 1 mV
Span Minimum 200mV to maximum 4000 mV (programmable)
Temperature
Range -20º C to +200º C
Resolution 0.01º C
Accuracy ± 0.5 % of FS
Sensor PT 100 (default) / PT 1000
Compensation Automatic or manual
Calibration
pH Sensor slope 70% to 110%
pH sensor zero ± 2pH
Method Single / dual point
Diagnostics
Measurement mode Temperature sensor open
Temperature sensor short
Calibration mode Slope error
Calibration error
Output
Current output – pH/ORP 4-20 mA DC, galvanically isolated output from the sensor proportional to
the measured pH/ORP value with accuracy better than 0.3% of full scale
HART Superimposed on 4-20mA output
Conductivity
Range (with auto range facility) Measuring range Cell constant K Resolution
0 to 10 µS/cm 0.01 0.01 μS/cm
0 to 100 µS/cm 0.01 0.1 μS/cm
0 to 1000 µS/cm 0.1 0.1 μS/cm
0 to 10 mS/cm 1 0.01 mS/cm
0 to 100 mS/cm 10 0.01 mS/cm
Accuracy ± 0.5% of measuring range
Span Minimum 5% to maximum 100% of measurement range (programmable)
Calibration
Conductivity Single point
Resistivity
Range (with auto range facility) Measuring range Cell constant K Resolution
100 MΩ.cm to 10 KΩ.cm 0.01 0.01 MΩ.cm
10 MΩ.cm to 1 KΩ.cm 0.1 0.1 MΩ.cm
1 MΩ.cm to 0.1 KΩ.cm 1 0.01 KΩ.cm
0.1 MΩ.cm to 0.1KΩ.cm 10 0.1 KΩ.cm
Accuracy ± 0.5% of measuring range
Temperature
Resolution 0.01º C
Accuracy ± 0.5 % of FS
Diagnostics
Output
Current output – conductivity/ resistivity 4-20 mA DC, galvanically isolated output from the sensor proportional
to the measured Conductivity value with accuracy better than 0.3% of
full scale
DO
Range (with auto range facility) Measuring range Unit Resolution
0 to 500 % 0.1
50 to 9999 µg/l or PPB 1
0.05 to 20 mg/l or PPM 0.01
0.05 to 50 mg/l or PPM 0.1
Accuracy ± 0.5% of measuring range
Temperature
Range -10.0 to +150.0 Deg C
Resolution 0.01º C
Accuracy ± 0.5 % of FS
Sensor NTC 22 K Ohm
Compensation Automatic or Manual
Process parameter
Pressure 0 to 5 bar (Programmable)
Salinity 0 to 50 PPT (Programmable)
Humidity 0 to 100% (Programmable)
Calibration
DO Single/dual point calibration
Diagnostics
Measurement mode Temperature sensor open
Temperature sensor short
DO2 sensor short
Output
Current output - DO 4-20 mA DC, galvanically isolated output from the sensor proportional
to the measured DO value with accuracy better than 0.3% of full scale
HART Superimposed on 4-20mA output
FOR TDS – AquaTDS SMARTPro ® 8969
TDS
Range (with auto range facility) Measuring range* Cell constant K Resolution
0 to 7 ppm 0.01 0.01 ppm
0 to 70 ppm 0.01 0.01 ppm
0 to 700 ppm 0.1 0.1 ppm
0 to 7 ppt 1 0.01 ppt
0 to 70 ppt 10 0.1 ppt
*TDS ranges calculated considering TDS factor 0.7
Ordering Information
Opp 106th Milestone B-85, Phase II, Chakan Indl Area A-34/35, MIDC H Block
Bombay Poona Road Sawardari, Chakan, Tal. Khed Pimpri, Pune - 411 018. INDIA.
Kasarwadi, Pune - 411 034. INDIA Dist. Pune - 410 501. INDIA Tel : 91(0)20-27442020, 39851199
Tel : 91(0)20-27145595, 39858555 Tel : 91(0)2135-393400 Fax : 91(0)20-27442040
Forbes Marshall Fax : 91(0)20-27147413
Krohne Marshall CIN No.: U28996PN1985PTC037806
Forbes Marshall Arca Email : pasales@forbesmarshall.com, ccmidc@forbesmarshall.com www.forbesmarshall.com
Codel International © All rights reserved. Any reproduction or distribution in part or as a whole without written permission of Forbes Marshall Pvt Ltd, its associate
Forbes Solar companies or its subsidiaries (“FM Group”) is prohibited.
Information, designs or specifications in this document are subject to change without notice. Responsibility for suitability, selection, installation, use,
Forbes Vyncke operation or maintenance of the product(s) rests solely with the purchaser and/or user. The contents of this document are presented for informational
Forbes Marshall Steam Systems purposes only. FM Group disclaims liabilities or losses that may be incurred as a consequence of the use of this information.
DOC#FMPA/0617/1504/V3.R1