You are on page 1of 8

See discussions, stats, and author profiles for this publication at: https://www.researchgate.

net/publication/12473240

Dose calculations for asymmetric fields defined by independent collimators


using symmetric field data

Article in The British journal of radiology · May 2000


DOI: 10.1259/bjr.73.868.10844866 · Source: PubMed

CITATIONS READS

7 630

4 authors, including:

Ioannis A Tsalafoutas Stylianos Nikoletopoulos


Saint Savvas Hospital IASSO Hospital Athens
90 PUBLICATIONS 1,073 CITATIONS 8 PUBLICATIONS 107 CITATIONS

SEE PROFILE SEE PROFILE

All content following this page was uploaded by Stylianos Nikoletopoulos on 21 May 2014.

The user has requested enhancement of the downloaded file.


The British Journal of Radiology, 73 (2000), 403±409 E 2000 The British Institute of Radiology

Dose calculations for asymmetric ®elds de®ned by


independent collimators using symmetric ®eld data
1
I A TSALAFOUTAS, MSc, 2S XENOFOS, PhD, 2A PAPALEXOPOULOS, MSc and
2
S NIKOLETOPOULOS, MSc
1
Medical Physics Unit, Konstantopoulio ± Agia Olga Hospital, 3-5 Agias Olgas, Nea Ionia, 142 33, Athens
and 2Medical Physics Department, Iaso Hospital, 37-39 Ki®ssias, Maroussi, 151 23, Athens, Greece

Abstract. Several methods have been developed for the dosimetry of asymmetric radiation ®elds
formed by independently moving collimator jaws. Three of these methods, based on different
principles and modi®ed to comply with our set of available data, are utilized for the calculation of
asymmetric ®eld dose pro®les. All three methods use output factors and per cent depth doses or
tissue maximum ratios of symmetric ®elds. In the ®rst method, calculation of the off-centre ratio
(OCR) of the asymmetric ®eld is based on the symmetric ®eld from which the asymmetric is
originated, by setting the one jaw in an asymmetrical position. In the second method the OCR of
the symmetric ®eld is used for the OCR calculation of the asymmetric ®eld of the same size;
whereas the third method does not allow for the asymmetric OCR calculation. The results
obtained using data for the 6 MV photon beam of a Philips SL-20 linear accelerator indicate that
both the ®rst and second method can accurately reproduce asymmetric ®eld pro®les from
symmetric ®eld data; the third method does not allow for penumbra reproduction, but it is
accurate at the central part of the asymmetric ®eld. The problems encountered in the application
of the three methods are reported and their accuracy is compared.

Linear accelerators equipped with independent to reproduce asymmetric ®elds produced by a


collimators are capable of producing a wide Philips SL-20 linear accelerator (Crawley, UK)
variety of off-axis radiation ®elds, formerly located in the Iaso Radiotherapy Department.
done with the use of shielding blocks. This has The ®rst method, proposed by Kwa and collea-
found many applications in optimizing radiation gues [1], applies to the situation where only one of
treatment schemes [1] but has increased the the independent jaws is closed down to form an
volume of the basic dosimetric data required asymmetric ®eld of smaller width or length than
and thereby necessitated modi®cations of treat- the original symmetric ®eld. This method uses the
ment planning systems. original symmetric ®eld pro®le corrected point-
The implementation of symmetric ®elds into a by-point by a correction factor, and is based on
treatment planning system requires quite a large the principle of the Day's equivalent ®eld
set of data for various ®eld sizes: per cent depth calculation [10]. The second method, based on
doses (PDDs) or tissue maximum ratios (TMRs), work by several authors [3±7], generates asym-
output factors (Sc , Sp), cross-beam dose pro®les
metric ®eld pro®les by computing the off-centre
at various depths, etc. Several treatment planning
ratio (OCR) (often referred to as the off-axis
systems calculate asymmetric ®elds on the
ratio, OAR) of the asymmetric ®eld while using
assumption that these are equivalent to blocked
output factors and PDDs of the equivalent
®elds [2]. However, the parameters required for
symmetric ®eld. The third method, proposed by
treatment planning implementation depend on the
algorithm used, and the method of their deriva- Khan et al [3], is the simplest, but is limited to the
tion is often not explicit [2]. calculation of dose to a point on the central ray of
Since asymmetric ®elds are the simplest case of asymmetric ®elds. In this study it has also been
irregular ®elds, others [1, 3±9] have approached applied beyond the central ray of the asymmetric
this problem using appropriate algorithms that ®eld.
utilize the data available for symmetric ®elds. The ®rst two methods require software for their
Three of these later methods have been chosen application, with the ®rst being more demanding,
since interpolations and extrapolations are needed
for point-by-point corrections. We studied the
Received 1 March 1999 and in revised form 5 August problems arising from their application and
1999, accepted 25 August 1999. compared their accuracy.

The British Journal of Radiology, April 2000 403


I A Tsalafoutas, S Xenofos, A Papalexopoulos and S Nikoletopoulos

Materials and methods where 100 OCRs (WL, d, x) is the relative dose
pro®le and
The study was performed on an SL-20 linear
accelerator for the 6 MV X-ray beam. A comput- CF Wa L; WL; d; x†~
erized water tank system was used for dose s
measurements (Wellhofer water phantom with WP Sc;p ‰ 2Wa {W {2x†L; dŠTMRs ‰ 2Wa {W {2x†L; dŠ
700S, v3.20 Nucletron Software; Schwazembuck, Sc;p ‰ W {2x†LŠTMRs ‰ W {2x†LŠ
Germany). The measurements were carried out at (3)
source-to-surface distance (SSD) set to 100 cm,
using a 0.125 cm3 ionization chamber. where Wa and W are the asymmetric and the
Data for a full set of measurements for corresponding symmetric ®eld width, L is the ®eld
symmetric ®elds and a number of asymmetric length de®ned at the source-to-axis distance
®elds, including cross-beam pro®les at various (SAD), while TMR and Sc,p (5ScSp) refer to
depths, PDDs and output factors, were obtained symmetric ®elds. The symmetric ®eld correspond-
during dosimetric calibration of the accelerator. ing to a given asymmetric ®eld is that with width
These data were made available from the WP twice that of the longest opening of the asymmet-
700S software, in text ®le format for manipulation rically set jaws and equal length. To illustrate the
(in 2 mm steps). The asymmetric pro®les were formation of an asymmetric ®eld from the
®elds 666 cm2, 10610 cm2 and 15615 cm2 with corresponding symmetric ®eld, a schematic repre-
offsets 3 cm, 5 cm and 7.5 cm, respectively, and sentation of the beam's eye view of the radiation
666 cm2 with offset 8.7 cm at depths 1.5 cm, ®elds is given in Figure 1.
5 cm and 10 cm. We must note here that Kwa and colleagues
The symmetric pro®les were all taken as relative derived Equation (3) considering that the right
cross-beam pro®les with the 100% corresponding to collimator was the one that was moving to the
the centre of the ®eld (x50). To convert these to centre forming the asymmetric ®eld. If the left
absolute dose pro®les, the central axis dose at depth collimator is forming the asymmetric ®eld (which
1.5 cm for the 10610 cm2 symmetric ®eld was used was the case in our study) then 22x should be
as reference dose (unit dose or 100%). Thus, the substituted by +2x in Equation (3).
absolute dose Ds for any symmetric ®eld at any For the standard SSD method, Equations (2)
point and depth is given by the following equation: and (3) become:
Ds WL; d; x†~Sc WL†Sp WL†PDDs WL; d†OCRs WL; d; x†
Da Wa L; d; x†~‰ SSDzd max††= SSDzd†Š2 |
(1)
S c WL†Sp Wd Ld †|
where W and L are the width and length of the TMRs Wd Ld ; d†|
®eld (de®ned at SSD5100 cm), respectively, x is 100OCRs WL; d; x†|
the distance from the central axis, d is the depth, CF Wa L; WL; d; x†
the subscript ``s'' represents symmetric ®elds and
Sc and Sp are the collimator scatter factor and the (4)
phantom scatter factor, respectively (normalized where
to 10610 ®eld).
s
Sc r1 †Sp r1;d †TMR r1;d ; d†
CF Wa L; W L; d; x†~
Sc r2 †Sp r2;d †TMR r2;d ; d†
Method 1
(5a)
The ®rst method (Kwa et al) has been described
for the case of isocentric technique [1]. To meet r15 equivalent square of rectangular ®eld
the requirements of the standard SSD technique
‰ 2Wa {W z2x†LŠ 5b†
used in this study for cross-beam pro®le acquisi-
tion, modi®cations have to be made to the
corresponding equations. and
The original equation for the description of the r25 equivalent square of rectangular ®eld
asymmetric ®eld dose pro®le (Da) can be written
‰ W z2x†LŠ 5c†
using the same notation as before, as:

Symmetric and asymmetric ®eld dimensions are


Da Wa L; d; x†~Sc;p WL† TMRs WL; d†|
de®ned at the surface and, for the ®eld size at
100 OCRs WL; d; x†| depth the subscript ``d'' is used (i.e.
CF Wa L; WL; d; x† 2† Wd5W(SSD+d)/SSD).

404 The British Journal of Radiology, April 2000


Dose calculations for asymmetric ®elds

W 30630 cm2 are about 1±2% at the shoulder and


up to 3.5% at the tail. Thus the relative pro®les
(as OCRs) of the corresponding square symmetric
®elds can be used instead. This approximation
reduces the number of pro®les required for the
application of this method, but the accuracy is
slightly reduced.
L Wa The problem in the application of this method,
as mentioned by Kwa et al [1], is the penumbra of
the asymmetrically set jaw, which in fact deter-
mines the left ®eld edge. To deal with this they
have suggested that a correction factor should be
set semi-empirically, using as a basis the penum-
bra of the symmetric ®eld. The exact methodol-
ogy, however, was not made explicit. Assuming
that the penumbra of the asymmetric ®eld is
similar to that of the symmetric ®eld with equal
dimensions [1, 8], we used another procedure for
penumbra correction. The pro®le of the sym-
Figure 1. Schematic representation of the beam's eye metric ®eld with the same dimensions as the
view of an asymmetric ®eld and its corresponding asymmetric pro®le was offset so as to bring its
symmetric ®eld. centre to the centre of the asymmetric pro®le for
the depth in question. This pro®le was then
renormalized so as to match the dose value 1.5 cm
Equations (4) and (5) modify the original from the left ®eld edge (towards the central ray)
Equations (2) and (3), to ®t the case of standard with the dose value, at the same point, of the
SSD technique. TMRs were obtained from PDDs asymmetric pro®les produced. Finally, the left
using the following equation [11]: penumbra of the renormalized symmetric ®eld is
attached to the asymmetric pro®le to form the
" #2 new (penumbra-corrected) asymmetric pro®le.
PDD r; d; SSD† Sp rd max† † SSDzd
TMR rd ; d†~ The choice of 1.5 cm is based on the fact that
100 Sp rd † SSDzd max†
TMR and Sc,p of ®elds smaller than 363 cm2
(6) (Equation 5a) are not available and cannot be
accurately extrapolated [1].
where d(max) is the depth of maximum dose
(d(max)51.5 cm) and r, rd(max) and rd are the
equivalent square ®eld size at the surface, at depth Method 2
d(max) and at depth d, respectively. Symmetric The second method that will be used in this
®eld data (PDDs, output factors, pro®les etc.) study for the calculation of asymmetric ®elds was
were introduced in a computer program written developed by Loshek and Keller [5], based on
for the application of this method, which utilizes work by Chui and Mohan [4]. In its ®nal
Equations (4) to (6). Interpolation and extrapola- formulation, using the normalization proposed
tion algorithms and penumbral corrections (to be by Thomas and Thomas [6], this method predicts
described later) were also included, so that any the absolute asymmetric ®eld pro®le (Da) by
asymmetric ®eld pro®le could be produced from means of the following equation:
the corresponding symmetric ®eld data.
The method described above requires the Da WL; d; x†~Sc WL†S p WL†|
availability of the OCR of the symmetric ®eld PDDs WL; d†OCRa WL;x; d† 7a†
with the same length as the asymmetric ®eld and a
width twice that of the longest opening of the where
asymmetrically set jaws (Equation 4). That is, for
POCR x†
a 15615 cm2 asymmetric ®eld with offset 7.5 cm, OCRa WL; x; d†~OCRs WL; x{x0 ; d†
the corresponding symmetric ®eld OCR required POCR x{x0 †
is 30615 cm2. However, data for rectangular (7b)
®elds are not always available, whereas data for a
square ®eld with length twice that of the Sc , Sp and PDDs are the same as in Equation
asymmetric, i.e. 30630 cm2, are more likely to (1) (i.e. for the corresponding symmetric ®elds),
be available. As shown in Figure 2, the differences and OCRa and OCRs are the off-centre ratios of
in the relative dose pro®les of a 30615 cm2 and a the asymmetric ®eld (with centre at x5x0) and the

The British Journal of Radiology, April 2000 405


I A Tsalafoutas, S Xenofos, A Papalexopoulos and S Nikoletopoulos

Figure 2. Relative cross-beam dose


pro®les of the 30630 cm2 and
30615 cm2 symmetric ®elds. d,
depth.

symmetric ®eld (centre at x50), respectively. The both the 30630 cm2 and 30615 cm2 symmetric
POCR is the primary off-centre ratio, de®ned as ®eld OCRs, are shown in Figure 3. The pro®les
the ratio of dose at a point off the central ray to calculated using the OCR of the symmetric
the dose at the corresponding point at the central 30615 cm2 ®eld present errors only at the left
axis, for an in®nitely large ®eld [6]. ®eld edge. Those calculated using the OCR of the
symmetric 30630 cm2 ®eld present additional
errors at the right ®eld edge: the errors in the
Method 3 shoulder are about 1%, while the errors in the tail
The third method to be applied is the one are more prominent.
proposed by Khan et al [3] for predicting the dose Corrections in the right ®eld edge, when using
along the central ray of an asymmetric ®eld. It is the corresponding square symmetric ®eld pro®les,
described by Equation (7a) if OCRa (WL,x,d) is can be obtained in a similar way to the
substituted by POCR(x). As a result, this method corrections of the left ®eld edge, described in
does not allow for penumbra reproduction. the Materials and methods section. The right
The major dif®culty in the application of the geometrical ®eld edge at the corresponding depth
second and third methods is in the de®nition and was chosen as the normalization point, that is the
measurement of the POCR [4±6]. In this study, to point where the penumbra of the symmetric ®eld
simplify the method and avoid additional meas- with equal dimension is attached. This arbitrary
urements, POCR was approximated from the choice was based on the fact that this point is
40640 cm2 cross-beam pro®les at depth acquired known a priori for all ®elds, keeping the rationale
during the standard dosimetric calibration of the of the correction algorithm uniform.
accelerator. This approximation is expected to The pro®les for the 666 cm2, 10610 cm2 and
introduce small errors only: ®rst, the actual ®eld 15615 cm2 asymmetric ®elds with offsets 3 cm,
sizes are usually smaller than 40640 cm2 so that 5 cm and 7.5 cm, respectively, as predicted by the
the effects of the collimators in the POCR are not ®rst method, using the corresponding square
important [4]; and second, the errors introduced symmetric ®eld pro®les and the proposed penum-
by scatter when using the cross-beam pro®les at bral corrections (both left and right edge), as well
depth instead of the corresponding POCR are to as by the second and third method, are shown in
some extent cancelled out in the ratio of Equation Figures 4, 5 and 6.
(7b) as they affect both POCR(x) and It can be seen that both the ®rst method,
POCR(x2x0). corrected as proposed, and the second method
can accurately reproduce asymmetric pro®les.
Some offsets in the penumbra are due to the
Results and discussion fact that experimental points have been taken at
The pro®les for a 15615 cm2 asymmetric ®eld 0.2 cm intervals, which may introduce a maxi-
at depths 1.5 cm and 10 cm, calculated using the mum positional error of ¡1 mm. It may also be
®rst method, before penumbral corrections, using observed that the third method, even though

406 The British Journal of Radiology, April 2000


Dose calculations for asymmetric ®elds

Figure 3. Asymmetric cross-beam


dose pro®les of the 15615 cm2
(offset57.5 cm) ®eld, at depths
1.5 cm and 10 cm, calculated with
the Kwa et al method before the
penumbral corrections. The pro®les
have been calculated using both the
30630 cm2 and 30615 cm2 sym-
metric ®eld off-centre ratio (OCR).
d, depth.

Table 1. Dose values expressed relative to the dose at depth51.5 cm of a symmetric 10610 cm2 ®eld, which is
considered equal to 100

Field size (cm2) Depth Data point Kwa et al's Loshek and Khan et al's
(perpendicular ray (cm) method [1] Keller method [5] method [3]
off-axis distance (cm))

666 1.5 99 98.6 98.9 98.9


(3) 5 84.6 84.3 83.9 84.2
10 63.7 63.7 63 63.7
10610 1.5 103.5 103.9 104 104.2
(5) 5 89.3 89.7 89.7 89.4
10 69.5 69.3 69.8 69.6
15615 1.5 108.5 109 109 109
(7.6) 5 94.3 94.6 94.8 94.5
10 74 74 74 73.8

Figure 4. Asymmetric cross-beam


dose pro®les of 666 cm2 (off-
set53 cm) ®eld, calculated by three
different methods. d, depth.

The British Journal of Radiology, April 2000 407


I A Tsalafoutas, S Xenofos, A Papalexopoulos and S Nikoletopoulos

Figure 5. Asymmetric cross-beam


dose pro®les of a 10610 cm2 (off-
set55 cm) ®eld, calculated by three
different methods. d, depth.

originally proposed for the central ray of the different offsets. For example, the asymmetric
asymmetric ®eld, can also be applied with good 666 cm2 ®eld with offset 8.7 cm requires a
accuracy for the central part of the ®eld (about 23.466 cm2 pro®le, which is unlikely to be
60% of the ®eld width). The calculated dose available and has to be interpolated from the
values (by all three methods), for various depths treatment planning system. So the accuracy of the
along a perpendicular ray at the centre of the reproduced asymmetric pro®le depends on the
asymmetric ®eld, are essentially the same as accuracy of the interpolated symmetric pro®le.
shown in Table 1. The second method, however, does not present
It should be mentioned that the ®rst method is this problem, as it uses the symmetric ®eld with
quite demanding, as far as the number of pro®les the same dimensions as the asymmetric, irrespec-
required for its application is concerned. While tive of the offset. In Figure 7, the 666 cm2 ®eld
the data requirements can be greatly reduced by with offset 8.7 cm has been calculated using the
using square symmetric ®elds and the proposed second and the third method. Again, the asym-
right ®eld edge correction procedure, it is obvious metric pro®les calculated with Method 2 agree
that a different symmetric pro®le is required for well with those measured, while the third method
asymmetric ®elds of the same dimensions but with is accurate at the central part of the ®eld.

Figure 6. Asymmetric cross-beam


dose pro®les of a 15615 cm2 (off-
set57.5 cm) ®eld, calculated by
three different methods. d, depth.

408 The British Journal of Radiology, April 2000


Dose calculations for asymmetric ®elds

Figure 7. Asymmetric cross-beam


dose pro®les of a 666 cm2 (off-
set58.7 cm) ®eld, calculated by two
methods. d, depth.

From Figures 4±7, it can be seen that the References


approximation of POCR with the 40640 cm2 1. Kwa W, Kornelsen RO, Harrison RW, El-Khatib
cross-beam pro®les at depth does not evidently E. Dosimetry for asymmetric x-ray ®elds. Med Phys
affect the pro®les calculated by the second 1994;21:1599±604.
method. The differences observed between meas- 2. International Commission on Radiation Units and
ured and calculated pro®les at the edges of the Measurements. Representation and modi®cation of
photon beams. In: ICRU Report 42. Use of
®eld (Figure 6) could in principle be attributed to
computers in external beam radiotherapy proce-
the POCR approximation. If this were true, dures with high-energy photons and electrons.
however, the errors should increase with depth, Bethesda, MD: ICRU, 1987:8±11.
an effect that is not observed. So, although the 3. Khan FM, Gerbi BJ, Deibel FC. Dosimetry of
POCR approximation introduces a small error, asymmetric x-ray collimators. Med Phys 1986;13:
most of the differences should be attributable to 936±41.
4. Chui C, Mohan R. Off-center ratios for three
experimental and positional rounding errors. dimensional dose calculations. Med Phys 1986;13:
409±12.
5. Loshek DD, Keller KA. Beam pro®le generator for
Conclusions asymmetric ®elds. Med Phys 1988;15:604±10.
The method of Kwa and colleagues, modi®ed 6. Thomas SJ, Thomas RL. A beam generation
for standard SSD technique and with the addi- algorithm for linear accelerators with independent
collimators. Phys Med Biol 1990;35:325±32.
tional penumbral corrections proposed, works 7. Chui C, Mohan R, Fontenla D. Dose computations
well and exhibits very good accuracy. for asymmetric ®elds de®ned by independent jaws.
However, Loshek and Keller's method, which is Med Phys 1988;15:93±5.
also accurate, has the advantage of using a single 8. Palta JR, Ayyangar KM, Suntharalingam N.
equation for the whole pro®le and directly Dosimetric characteristics of a 6 MV beam from
a linear accelerator with asymmetric jaws. Int
producing the penumbra region. It is also less
J Radiat Oncol Biol Phys 1988;14:383±7.
demanding in the number of pro®le data required. 9. Gibbons JP, Khan FM. Calculation of dose in
The approximation of POCR with the asymmetric photon beams. Med Phys 1995;22:1451±
40640 cm2 cross-beam pro®les at depth results 7.
in calculated asymmetric pro®les that compare 10. Day MJ. A note on the calculation of dose in X-ray
well with measured data, and greatly simpli®es the ®elds. Br J Radiol 1950;23:368.
11. Khan FM. A system of dosimetric calculations. In:
application of this method. The physics of radiation therapy. Baltimore, MD:
For point dose calculations at the central part Wilkins & Wilkins, 1984:187.
of the asymmetric ®eld, the prominent simplicity
and good accuracy of the method of Khan et al
makes it the method of choice for straightforward
monitor unit calculations.

The British Journal of Radiology, April 2000 409

View publication stats

You might also like