com
Consolidated Manual
Table of Contents
Introduction of Manual............................................................................................................................................................5
Certificate of Warranty............................................................................................................................................................6
Symbols used in product.........................................................................................................................................................7
Warning before Use.................................................................................................................................................................8
Electrical Safety..........................................................................................................................................................................9
Radiation Safety......................................................................................................................................................................11
1. PAPAYA Plus Equipment Introduction........................................................................................................................12
1.1 Intended Use.........................................................................................................................................................12
1.2 Equipment’s Characteristics...............................................................................................................................12
1.3 System Basic Specification.................................................................................................................................13
1.4 Equipment Composition.....................................................................................................................................14
1.5 Accessories............................................................................................................................................................15
1.6 Patient Apply Part................................................................................................................................................16
1.7 On / Off Switch....................................................................................................................................................17
1.8 Hand Switch..........................................................................................................................................................18
1.9 Emergency Stop Switch......................................................................................................................................19
1.10 Equipment’s Control Button..............................................................................................................................20
1.11 Equipment’s 2nd Control Button.......................................................................................................................21
1.12 Laser Control Knob..............................................................................................................................................22
1.13 Changing Detector...............................................................................................................................................23
1.14 Device status LED.................................................................................................................................................24
1.15 External status LED..............................................................................................................................................25
1.16 Changing Fuse......................................................................................................................................................26
1.17 Detector status LED.............................................................................................................................................27
2. OP (Operation panel)......................................................................................................................................................28
2.1 OP Structure..........................................................................................................................................................28
2.2 X-ray Exposure Mode..........................................................................................................................................30
2.2.1 Panorama program...................................................................................................................................30
2.2.2 Cephalo Program......................................................................................................................................34
2.3 Choose Patient Size.............................................................................................................................................36
2.4 Exposure Speed and Change of Image Quality.............................................................................................36
2.5 Selection of Jaw shape.......................................................................................................................................36
Introduction of Manual
GENORAY’s PAPAYA Plus is digital panoramic X-ray system. This manual is designed for your safe use and
operation of PAPAYA Plus. When using PAPAYA Plus, use .
The potential risk always exists when using X-ray products and electric medical equipments. All users should
be aware of the methods of use, solutions in emergency conditions and other safety articles provided in this
manual.
The user must acknowledge the appropriate ways to protect oneself from the potential danger that could
occur at any time, and be aware of the condition of the product that can cause danger.
The manufacturer, assembler, and the importer hold responsibility upon the performance of safety, trust
and the article below.
- Guarantees the installation of the eligible professional, calibration, and repair.
GENORAY Co., Ltd is continuously developing the products. The manual or composition of the package
might change without notice
Certificate of Warranty
GENORAY guarantee this product without any expense for the mal-function or disorder occurred under
standard transportation and operation within 1 year from installation. Defect on its appearance is out of the
range of warranty.
When the malfunction or disorder appeared, stop operating and check again the related article of the “User’s
Manual.” Before your calling, please put off this equipment and check for the Model Name, serial No. and
purchasing date. GENORAY Co., Ltd. is not responsible for indirect harm. GENORAY Co., Ltd. cannot warrant
for defect or harm after warranty period.
AC (Exchange) Laser
This symbol is information about the special matters that require attention while using this software.
This symbol is warning the situation or environments which might bring danger to the patients or users.
This machine can be dangerous for all users if the safe exposure condition and appropriate using
method are not attended.
There is prohibition of installing different composite parts unless they are supplied by manufacturer.
Do not connect equipment’s power source with multi-adaptor or extension cord. This can cause defect or
become cause of mal-function.
Do not operate machine when the machine is defect or malfunctioning. Please contact engineer or
service center for help.
Do not use provided PC for unintended purpose. There is probability some problem could happen on
transferring images and reconstructing if the PC uses for unintended purpose.
Electrical Safety
The electric circuit of this product uses the electric current which can cause death due to severe
injury or electric shock. To avoid this kind of danger, do not remove any of the covering part.
To avoid failure and the danger that might cause serious injury or electric shock to users and patients, please
acknowledge the safety procedures listed below.
Before cleaning the product, always turn off the power and separate the power cord, and clean the
product with lightly wet cotton or sponge.
Never let anyone remove the cover except for the professional technician of GENORAY.
Do not place food on any part of the product. They might get into the electric circuit and get into
contact with electrical matters
The product is not waterproof to water, soap or liquid substances. If the liquid contacts the
product, it may cause fire or electric shock. If the liquid was in the product by accident, do not
touch the power cord connecting to the power, and wait until the liquid is completely dry.
Turn off the power and separate the power cable from the socket.
Stay far away from the place where the product has been installed.
Ask for help around you.
Use the fire extinguisher authorized only for the electric fire.
Abnormal use of fire extinguisher could be dangerous and cause fire or electric shock. To avoid
this kind of danger, check to see if the product can be used where the product has been
installed, and always use the authorized product.
Before connecting the power, check to see of the power supplying line is appropriate to this product.
Check the power and connect the power cable of this product to its power socket.
Check the categories mentioned above in case of power input of all accessory (the signal input/output
including accessory).
For reducing the risk of electrical shock, and when connecting to power, there needs protecting
ground connection.
Place this product away from other X-ray products, power generating equipment and
broadcasting stations. If the product is sharing the power with other electric devices, abnormal
image might occur.
The protection level of PAPAYA Plus is the 1st grade equipment or type-B equipment. This
product does not prevent the invasion of liquid substances.
Radiation Safety
This product needs X-ray room. Except the patient, no one can enter X-ray room unless they are patient.
While taking exposure, clear space between patient and equipment need to be maintained.
For this product’s safe operation, it needs to be installed at appropriate location with permission of
owner. Also, user needs to guarantee about whether equipment will be used at place where owner has
chosen.
This product’s owner needs to be guaranteed that all users have to wear X-ray protection devices while
using this equipment.
For this product, users need to wear TLD badge, and check dosage exposure regularly.
If the pregnant woman or young child needs to have dental X-ray taken, consult with specialist. It should
be considered to balance the benefit of the test and potential small risks of x-ray tests by specialist such
as doctors and the radiologist.
If you do not follow written rules in this manual, there can be risk for operator and patient.
Federal law restricts this device to sale by or on the order of physicians and dentists.
PAPAYA Plus
PC is used for acquiring and displaying image, and for image processing.
This machine has panoramic only model of PAPAYA and has added cephalo exposure that is
PAPAYA Plus. Each model’s composition or exterior, please refer to below pictures.
4 9
13
5 10
7
8 14
2 11
Hand Switch
5 X-ray source assembly 11
(Exposure button, Emergency stop switch)
12 Emergency stop switch
Detector
6 13 Device status LED
(Panorama, Cephalo using for both,
detachable structure)
14 Equipment’s 2nd control button
1.5 Accessories
For appropriate exposure mode, insert correct chin rest to designated hole.
Must use spare part that is provided by manufacturer. If you damage machine when using
different spare part, and then the manufacturer does not have responsibility of fault.
Cephalo part
Panorama part
Bite
Chin rest
Temple support
Nasal
Patient handle
positional
Ear rod
2nd control
button
OFF ON
Equipment’s on/off switch is like on picture, it is located at equipment’s below part. If turning on
equipment’s switch, and the equipment’s LED has light and displays equipment’s waiting condition.
Emergency stop switch is located at above picture. When emergency situation happens and push button and
equipment’s driving unit or X-ray source assembly’s power supply stops and machine will stop.
When pushing emergency stop switch, turning switch to clockwise to unwind and machine will
operate normally.
Turn on/off equipment’s laser. When pushing once, laser turns on and pushing
one more time and laser turns off. Laser will turn off after time. It will turn off
automatically.
Turn on/off equipment’s laser. When pushing once, laser turns on and pushing
one more time and laser turns off. Laser will turn off after time. It will turn off
automatically.
This equipment, as shown above picture, has laser control know. Before to exposure, when controlling
patient’s position, using laser control knob to adjust to Frankfort line.
A. Pushing detector handle area’s B. While pushing switch, pull C. By disassembling hook part
switch. detector downward. and pushing detector upward
and detector will fall out.
Following above change method and doing reverse way and detector will be attached.
A. Disassemble the B. Replace the broken fuse C. Assemble the fuse holder
fuse holder using with new fuse using driver
driver
2. OP (Operation panel)
2.1 OP Structure
1 9
2 11
4
12
10 6
5
7 8
10 Jaw shape selection button Choosing patient jaw shape and size
ease before take an exposure, always look equipment’s display and check whether equipment can
take an exposure. About equipment’s condition, please refer to 2.6.
Standard Panoramic
Standard panoramic has standard panorama’s image field of view and
Taking exposure.
Panorama mode
Orthogonal Panoramic
Orthogonal Panoramic is comparison to Standard panoramic
program that takes x-ray exposure with x-ray beam’s angle as 90° of
penetration to chin, and produce field of view. At standard
panoramic image, teeth is overlapping and effect is decreasing, but
exposure area’s opposite side’s shadow of teeth and chin and add an
effect.
Bitewing Panoramic
Bitewing Panoramic program works at panoramic image and take
exposure of bitewing area and take exposure of trajectory of patient.
In basic, Bitewing Panoramic is orthogonal panoramic and take
exposure with same angle.
TMJ lateral
TMJ lateral exposure has function of trajectory and side of
temporomandibular joint. Taking exposure of both sides of
temporomandibular joint and comparing, patient opens mouth and
the patient closes mouth to distinguish temporomandibular joint
exposure.
TMJ mode
TMJ PA
TMJ PA takes exposure of temporomandibular joint’s PA direction
and take exposure with trajectory direction. Both sides of
temporomandibular joint can take exposure and using as patient
opening mouth and closing mouth.
Double Lateral-PA
Lateral – PA program takes exposure of temporomandibular joints’
PA and Lateral direction alternately. Both sides of
temporomandibular joint can take exposure and using as patient
opening mouth and closing mouth.
Sinus Midsagittal
Sinus midsagittal exposure enables to take exposure of maxillary
sinus and use it. The image’s focus is on midsagittal surface.
Sinus mode
Sinus lateral
Sinus lateral exposure is like Sinus midsagittal, at maxillary sinus,
take a exposure at maxillary sinus’ trajectory. From midsagittal
surface’s standard, now focus surface can be toward left or right
side.
Sinus PA
Sinus PA takes an exposure at PA’s direction of maxillary sinus.
Cephalo lateral
Cephalo lateral exposure has at patient’s side, and X-ray beam’s
penetration of direction.
Cephalo mode
Cephalo AP
Cephalo AP exposure is that at toward AP’s direction, X-ray
penetrate and taking exposure of skull.
Cephalo PA
Cephalo PA exposure is that at toward PA’s direction, X-ray penetrate
and taking exposure of skull.
Water’s view
Water's view is when the patient is at with Frankfort line and wit
h detector surface, and taking exposure with 45 degree angle slo
pe and is exposuring area of head.
SMV(Submento vertex)
SMV takes exposure of according to patient’s Frankfort line and the
surface of detector is leveled with horizon and put patient in this
position to take exposure.
Carpus
Carpus takes exposure of wrist. In order to do Carpus exposure,
using carpus exposure wrist panel and install it and take exposure.
Among 4 different size, can choose one size, and according to choice of size of patient, each exposure
program has saved X-ray exposure condition with trajectory. At each exposure program, X-ray exposure
condition is set up differently.
Fast image has lower image quality and HD has longer image processing time but image.
Average
Large
You can choose different jaw shape and size of patients. There are 9 options by following 3 jaw shape (Sharp,
Normal, Square) and 3 jaw size (Small, Average, Large).
The exposure condition value is chosen automatically when choosing patient size & exposure
program. If you’d like to change the exposure condition by yourself, you can do like above.
2 3
1
Click the order list button, refer to 2.1, then the order list management window will be appeared. Functions of
each part are described in below table.
This below picture shows a window appears when you click ‘Register’ button, user can input patient information
and select exposure program with this window.
Condition Description
It shows the connection of device & PC. When the device is connected with
PC, it turns orange color.
2.9.1 General
Name Explanation
R-Mark Setting exposure images' position mark's printing out / not printing out
Save Raw images Setting Raw image's setting (If checking and taking up a lot of hard disk
2.9.2 DB Server
Name Explanation
2.9.3 DB Server
Name Explanation
System Log When pushing test button and printing out processing result
2.9.4 Version
Name Explanation
3. Image acquisition
Please refer the Triana manual provided with device for more details.
Please follow the process below after running Triana software to operate OP for the device operation.
Patient registration
A. OP operation
[ Panorama ] Choose the device among device selection tap once the patient
registration is done.
Papaya is a panoramic device so you can choose panorama. If the case
[ Cephalo ] of the device including cephalo function, you can choose one between
panorama & cephalo.
Please let the patient remove eye glasses, hearing aid, false/artificial teeth, hair pin, other
accessories and metal etc. before taking X-ray images. If not, the shadow or reflected images
can be seen. Please remove the cloths wearing around the neck such as scarf, muffler, necktie
for the safety.
3. Operate OP by clicking panorama exposure button after registering a patient according to 3.1
4. When you click “Confirm” button after setting panorama mode & exposure condition on operating
OP, “Confirm” button turns “Ready” button, and then patient position arrange window comes out.
You can’t change the X-ray condition after clicking ‘Confirm’ button. X-ray condition & exposure
change is possible on start screen by ‘Back’ button.
5. After being the patient positioned, set height on patient’s height by using 1.9 device set button.
6. Cover bite with clean cover, let the patient put their chin on
chin-rest and bite the bite.
If patient can’t raise their neck, set the height of device by using device set button.
8. Set the position of patient by pressing laser button ( ) of device set button while the patient
bites the bite correctly.
Laser is consists of 3, you should let the patient be positioned for all 3 lasers correctly to get a good quality of image
Frankfort plane is the straight line connecting upper part of an earhole & lower part of eyes.
Laser is turned off automatically after few minutes or if the exposure starts.
Please let the patient remove eye glasses, hearing aid, false/artificial teeth, hair pin, other
accessories and metal etc. before taking X-ray images. If not, the shadow or reflected images
can be seen. Please remove the cloths wearing around the neck such as scarf, muffler, neck tie
for the safety.
3. Operate OP by clicking panorama exposure button after registering patient according to 3.1
4.Set TMJ mode & exposure condition on operating OP and click “Confirm” on right-low side, then
“Confirm” button turns to “Ready” button, patient position control screen comes out.
You can’t change the X-ray condition after clicking ‘Confirm’ button. X-ray condition & exposure
change is possible on start screen by ‘Back’ button.
5. Set the height by using 1.9 device control button after being the patient positioned.
If patient can’t raise their neck, set the height of device by using device set button.
8.Set the position of patient by pressing laser button ( ) of device set button while the patient
stands on correctly.
Laser is consists of 3, you should let the patient be positioned for all 3 lasers correctly to get a good quality of image
Frankfort plane is the straight line connecting upper part of an earhole & lower part of eyes.
Laser is turned off automatically after few minutes or if the exposure starts.
Please let the patient remove eye glasses, hearing aid, false/artificial teeth, hair pin, other
accessories and metal etc. before taking X-ray images. If not, the shadow or reflected images
can be seen. Please remove the cloths wearing around the neck such as scarf, muffler, neck tie
for the safety.
3. Operate OP by clicking panorama exposure button after registering patient according to 3.1
4. Set Sinus mode & exposure condition on operating OP and click “Confirm” on right-low side,
then “Confirm” button turns to “Ready” button, patient position control screen comes out.
You can’t change the X-ray condition after clicking ‘Confirm’ button. X-ray condition & exposure
change is possible on start screen by ‘Back’ button.
5. Set the height by using 1.9 device control button after being the patient positioned.
If patient can’t raise their neck, set the height of device by using device set button.
8. Set the position of patient by pressing laser button ( ) of device set button while the patient stands
on correctly.
Laser is consists of 3, you should let the patient be positioned for all 3 lasers correctly to get a good quality of image
Frankfort plane is the straight line connecting upper part of an earhole & lower part of eyes.
Laser is turned off automatically after few minutes or if the exposure starts.
Please let the patient remove eye glasses, hearing aid, false/artificial teeth, hair pin, other
accessories and metal etc. before taking X-ray images. If not, the shadow or reflected images
can be seen. Please remove the cloths wearing around the neck such as scarf, muffler, and
necktie for the safety.
2. Operate OP by clicking cephalo exposure button after registering patient according to 3.1
3. Set mode & exposure condition on operating OP and click “Confirm” on right-low side, then
“Confirm” button turns to “Ready” button, patient position control screen comes out.
You can’t change the X-ray condition after clicking ‘Confirm’ button. X-ray condition & exposure
change is possible on start screen by ‘Back’ button.
4. Set the height by using 1.10 device control button after being the patient positioned.
5. Let the patient positioned according to exposure program after being Ear rod positioned to ear hole
of the patient.
6. Set the position of patient by pressing laser button ( ) of device set button while the patient
bites the bite correctly.
7. Laser enters horizontally. Control the angle of laser & patient’s Frankfort according to exposure program.
Frankfort plane is the straight line connecting upper part of an earhole & lower part of eyes.
Please keep pressing the switch during exposure. If the users take their hands off from shot
switch, the exposure is stopped and then, the error message comes out about the abnormal
shut down.
Focus on patient & condition of device during exposure. If the emergency situation is occurred,
take the hands off right away and then press emergency switch
When the exposure starts, OP shows X-ray warning message & the image taking, and final image comes out
again after the complete.
After image exposure finishes and on screen, image is displayed and before push "save" button and
can do image control
Double click mouse's left button Re-size image with size of window
Double click mouse's right button initialize image to before control brightness
1. Connect TWAIN session by choosing "PAPAYA" among TWIN data source whithin user's program
2. At user's program, the command of Acquire and send this command, and progress on OP's
full screen mode.
3. From contents of 3.2 to 3.6, depending upon exposure mode and choose exposure mode
and control patient position, and then try exposure of X-ray.
4. After exposure steps are finished normally and click "save" button and send user's program
called, "PAPAYA."
5. After sending is completed, and OP screen is closed automatically and ends with TWAIN session.
TWAIN data source can send image according to usage of Multi Focus that enables to send
more than 1 page. If on user's program and try to receive 1 page of image, and un-check the
check mark on content of 2.9.1's at general option tab and un-check All Save's checkbox.
At pop-up menu, by choosing send, and this function allows to pre-set to send image to DICOM
server. The DICOM server information is written on 2.9.3 and on DICOM option tab's lower left side
and set on Storage Server. By pushing "Insert Item" button, and input information window appears
and input AETitle, IP, Port, and Timeout and push "Insert" button. At Storage Server's Preset list, and
choose server in which will be sent and finish job by pushing "Save" button.
3.9.2 Burning CD
After taking exposured image and save them by using CD-ROM like media storage device or hard disk.
On pop-up menu, choose "Save to CD" and progress CD burning window.
For media storage device, choosing right side at device list and save image. Checking by Total Space
Used and Free Space and click "Burning" button. After burning is completed and finish and click
"Close" button.
When saving at hard disk or USB, by pushing button and choose save path and process burning.
With image, image viewer of CDSee is also saved.
Media device is connected, but device list doesn't appear and can't choose device, and push "Refresh"
button for updating the device list
On same path inside of CDSee folder, progress CDSee.exe and view saved image.
Additional order function is that on previous ordered list, and patient need additonal exposure and
doesn't input basic personal information. Then choose exposure program and do registration for order.
If you choose additional order, and have chosen patient's basic information is updated and progress
with registration ordered window. By choosing exposure program, and click "Register" button, and the
order is added on ordered list.
When choosing opening of folder, search window is opened by starting from where the images are in
saved folder.
4. Use condition
Please avoid these conditions for the operation/storing of this device.
Exposure to humidity
Exposure to
Exposure to dust
Exposure to high humidity
Exposure to condition where has problem with ventilation
Exposure to salty air.
Exposure to chemical / harmful gas.
Please avoid use the product the place where has strong vibration or over heat or humidity.
5. Cleaning
You can clean Bite piece, patient support table, chin rest & patient handle by high pressure sterilization up to
135℃ or cotton/gauze with alcohol.
Please clean softly OP panel or other parts with wet soft cloth.
6. Handling of waste
It is regular custom to follow the law of each country when handling of waste including devices. Please
follow the regulations of each country.
7. Message
6 No program at FPGA
No. Message 설명
8. Technical Data
8.1 Classification
You are recommended to use AC power to avoid crash of power requirements for other
devices.
74User manual
Technical Data
kV range 60~90 kV
mA range 4~12 mA
Superheated Temperature 50 ℃
Form Rectangle
8.6 Detector
중 량 110 kg ± 5%
PAPAYA Plus
중 량 140 kg ± 5%
8.8 Movement
Panorama : 480 mm
SID
Cephalo: 1650 mm
Panorama : 248 mm
SSD
Cephalo : 1359 mm
Type Stationary
Target Angle 5°
Thermal Characteristics
Anode Heat Storage Capacity 30 kJ
8.0 V
Filament characteristics (Max.)
4.0 A
Type Stationary
Target Angle 5°
Thermal Characteristics
Anode Heat Storage Capacity 35 kJ
4.2 V
Filament characteristics (Max.)
3.5 A
9. Labeling
9.1 PAPAYA
• Digital Sensor
• X-ray tube
• Beam limiting device
• Generator
• Rotating Arm
• Laser
• Hand S/W
• Manufacturer Label
• X-ray Support
• Digital Sensor
• X-ray tube
• Beam limiting device
• Generator
• Detector
• Rotating Arm
• Laser
• Hand S/W
• Manufacturer Label
• X-ray Support
[Manufacturer label]
- PAPAYA
- PAPAYA Plus
Appendix 1.
MAINTENANCE
The user needs to check that all safe related the equipment operation before using it.
Routine maintenance
- Machine and parts should be maintained by basic condition.
- Machine and parts should be checked by routine maintenance.
Regular maintenance
- Check list every 1 year
Accuracy of X-ray tube voltage
Accuracy of X-ray tube current
Exposure dose of patient
Test method
- Accuracy of X-ray tube voltage
Test instrument: X-ray multi meter
Test method: Measure the tube voltage value and compare setting voltage with measured voltage.
Test limit: The error of the values of the X-ray tube voltage shall be not greater than 10%
User manual87
Maintenance
If maintenance parameter is over the limit, some part is changed or repaired in manufacturer’s
judgment.
Please refer to the “PAPAYA Plus constancy test manual“ about detector calibration.
INSPECTION RESULT
INSPECTION ITEM INSPECTOR
PERIOD 1 2 3 4 5 Remarks
On / Off Switch
Patient Registration
WHEN THE
Tube Voltage Accuracy Test
MACHINE HAS
NOT BEING
USED FOR 6 Tube Current Accuracy Test
MONTH
Exposure Dose to the patient
Dose information
Big 72 11 12 67.05
Male 68 9 12 48.38
Female 65 7 12 33.60
Child 62 5 12 17.52
Big 72 11 12 67.05
Male 68 9 12 48.38
Female 65 7 12 33.60
Child 62 5 12 17.52
Big 72 11 8 44.70
Male 68 9 8 32.25
Female 65 7 8 22.40
Child 62 5 8 11.68
D. PA TMJ mode
Big 75 11 8 48.40
Male 72 9 8 36.57
Female 68 7 8 25.08
Child 65 6 8 15.93
Big 72 11 5 27.94
Male 68 9 5 20.16
Female 65 7 5 14.00
Child 62 5 5 7.30
F. PA Sinus
Big 75 11 9 54.45
Male 72 9 9 41.14
Female 68 7 9 28.22
Child 65 6 9 17.92
Big 85 10 8 8.67
Male 83 10 8 8.23
Female 81 9 8 7.02
Child 80 9 8 5.66
H. Carpus mode
Big 70 7 8 3.96
Male 67 6 8 3.04
Female 63 5 8 2.18
Child 60 4 8 1.27
Obelis s.a.
Boulevard Général Wahis 53, 1030 Brussels,
BELGIUM Tel: +32-2-7325954 Fax: +32-2-
7326003
Installation Manual
Copyright© 2012
2013 / 05 / 28
Document Ver. 0.2
Table of Contents
This manual provides guidelines for the safe use and operation of the system.
The user should be trained in the use of equipment of this type and should ref
er to this manual in conjunction with that training.
Potential hazards exist in the use of medical electronic devices and X-ray syste
ms.
All users must understand the safety and emergency procedures, and the operat
ing instructions given in this manual.
The user must know how to recognize hazardous and potentially hazardous con
ditions, and know how to adequately protect themselves and others from possibl
e injury.
The manufacturer, assembler, and importer are responsible for the safety, reliability
and performance of the unit only if:
- Installation, calibration, modification and repairs are carried out by qualified
authorized personnel.
SYMBOLS Description
Attention!
consult accompanying documents
WEEE Mark
Laser
The sign indicates information that needs attention to use the device.
<PAPAYA>
Required Tools
PANORAMA BOX
CEPHALO BOX
Distinguish
packaging's up and
down
SENSOR BOX
ACCESSORY BOX
- Accessory Box
1
6
7
3 4 5
PAPAYA CD 1
TRIANA CD 1
2 CD
CALLIBRATION
1
DATA CD
- 10 - Installation Manual
CEPHALO M8x25L
ASSEMBLY WRENCH spring washer 4
BOLTS and flat washers
CHINREST HANDLE
M6x20L
ASSEMBLY WRENCH 4
spring washer
BOLTS
COVER ASSEMBLY
M4x8L 6
SEM’S BOLTS
SHOT SWITCH
M4x12L
HOLDER ASSEMBLY 2
Washer head
SELF DRILL SCREW
2 CHINREST HANDLE 1
RIGHT AND
2 TEMPLE SUPPORT 1SET
LEFT
2 BITE 2
2 BITE COVER 2
3 CHINREST SUPPORT 1
4 CHINCUP SUPPORT 1
TMJ-SINUS
5 1
SUPPORT
6 SHOT SWITCH 1
7 LAN CABLE 1
Installation Manual - 11 -
1) PANORAMA UNIT disassemble
①
②
①
②
Lying down number 2 styrofoam and take off number 1 styrofoam. (Outside Styrofoam number is displayed)
Pulling out sensor box, base, and anchor base (option) while lying the machine on floor.
Put machine on upright position and use handle jig to tilt. Then disassemble the Styrofoam.
- 12 - Installation Manual
2) Moving to installation place
Installation Manual - 13 -
3) Base Sub Plate assembly
4-M8x25L, SW
assembly
- 14 - Installation Manual
4) Handling Jig disassemble
Installation Manual - 15 -
5) Stand Column
M8x20L,
3-M8x 25L, SW
assembly
- 16 - Installation Manual
6) Column Bottom Cover A/B assembly
Installation Manual - 17 -
7) Hinge & Driving Top Cover disassembly
- 18 - Installation Manual
8) Safety Pin & Cephalo Align Bolt disassemble
M8 NUT disassemble
Installation Manual - 19 -
9) Cephalo Unit assembly
Assembling machine while placing cephalo unit without unpacking from packing fo
- 20 - Installation Manual
10) Cephalo Cable assembly
Coming from Cephalo and organize cable toward red arrow side
Assemble Main Harness BKT
Identifying cable's label and connect main B/D to connector (Included LAN cable)
Installation Manual - 21 -
11) Cephalo Box disassemble
- 22 - Installation Manual
12) Driving Top Cover assembly
Installation Manual - 23 -
13) Patient Handle & Accessory assembly
4-M6x20L, SW
assembly
- 24 - Installation Manual
2.2 Level Adjustment
(1) Make the rotating part and supporting part at a line(180°) to each other and lay a dig
ital level on the datum plane. And keep the part horizontal with adjusting 6EA of screws.
(2) Make the rotating part and supporting part at right angles(90°) to each other and lay
a digital level on the datum plane. And keep the part horizontal with adjusting 6EA of s
crews.
Required tools : Digital Level, 5mm T Wrench (Or 5mm L Wrench)
Datum Plane
Installation Manual - 25 -
5mm T-Wrench
- 26 - Installation Manual
2.3 Connecting PC to Device / HAND SWITCH
NO. Equipment PC
1 Equipment _ LAN PC_LAN
Installation Manual - 27 -
2.4 PAPAYA APP Set up.
- 28 - Installation Manual
3. Choosing installation option and click 'Next' button.
If MS-SQL DB is not needed for installation and do not choose option
Installation Manual - 29 -
5. The installation process will proceed as shown in the figure below
- 30 - Installation Manual
7. When the installation is complete, click 'FINISH' button to exit.
Installation Manual - 31 -
3. Test After Installation
Test operating condition of each part.
OFF ON
- 32 - Installation Manual
3.2 Check Rotating Unit
(1) Press button to test after CONTROL screen came out. Check if Rotating Unit is
moving right way.
(2) If Rotating Unit is not moving, turn it manually a little and press button again.
(3) Repeat above procedure 2-3 times to confirm Rotating Unit is fine.
When the Rotating Unit does not stop and crashes into main body, turn off power
(1) Press Laser Aimer button on OP Panel and three Laser Aimers show up.
(2) Press Laser Aimer button again, three Laser Aimer disappear.
Installation Manual - 33 -
(4) After entering PAPAYA adapter properties (Properties), and the following screen will appear..
- 34 - Installation Manual
- Setting flow control (Flow Control) to Disable mode (Disable).
- Setting IPv4 Checksum Offload (IPv4 Checksum Offload) to all use
- setting Jumbo Frame (Jumbo Packet) to 9014 Byte or 9KB MTU
- Setting large send offload to use
- Setting receiving buffer (Receive Buffers) to 2048
- Setting TCP check sum off load (TCP checksum Offload) to all use..
Installation Manual - 35 -
Locating PAPAYA's adapter as above picture
- 36 - Installation Manual
GENORAY Co., Ltd.
#512 Byucksan Technopia 434-6 Sangdaewon 1-Dong Jungwon-Gu
Seongnam-City Gyeonggi-Do Korea, 462-716
TEL : +82-31-737-8020 FAX : +82-31-737-8025
www.genoray.com
Obelis s.a.
Boulevard Général Wahis 53, 1030 Brussels,
BELGIUM Tel: +32-2-7325954 Fax: +32-2-7326003
Installation Manual - 37 -
PAPAYA Plus
(PAPAYA)
Technical Manual
Copyright© 2013
2013/ 07 /15
Document Ver. 0.1
2. Composition of PAPAYA................................................................................7
3. FUNCTION CHECK.......................................................................................................8
3.1 Rotating part shift and reversion TEST..............................................................................8
4. DIAGRAM.....................................................................................................................11
4.1 TOTAL BLOCK DIAGRAM...........................................................................................11
5. Replacement PARTS..........................................................................................24
5.1 Chinrest Assy....................................................................................................................24
6. Calibration method........................................................................................33
This manual is designed for your safe use and operation of PAPAYA & PAPAYA PLUS, use
this manual to guide you. It must be used by a trained professional.
The potential risk always exists when using X-ray products and electric medical
equipments. All users should be aware of the methods of use solutions in emergency
conditions and other safety articles provided in this manual.
The user must acknowledge the appropriate ways to protect oneself from the potential
danger that could occur at any time, and be aware of the condition of the product that
can cause danger.
The manufacturer, assembler, and the importer hold responsibility upon the performance of safety, trust and the
- Guarantees the installation of the eligible professional, calibration, and repair
GENORAY Co., Ltd is continuously developing the products. The manual or composition of the package might ch
1.2 Cleaning and environment for using
Bite piece, (patient support table, chin rest and temple support can be cleaned by alcohol
soaked cotton. The other surface including the control panel of the product must be cleaned
gently by lightly wet, soft cotton. For the operation and installation of this product, please
avoid the categories listed below.
Exposure to moisture
Exposure to direct ray of light
Exposure to the space with dust
Exposure in high dampness
The environment which can cause problem on ventilation of product
Exposure in atmosphere including sodium
Exposure to chemicals harmful gas
During the use of the product, if there is a strong vibration at normal condition or in the
places that exceed the range of dampness or temperature below do not put the product
nearby.
SYMBOLS Description
X-Ray Source
Assembly, X-ray
조사
Caution!
Refer to the user’s manual
WEEE Mark
Laser
This symbol is information about the special matters that require attention while using this software.
This symbol is warning the situation or environments which might bring danger to the patients or users.
2. Composition of PAPAYA
14
3
9 13
4
10
5
7
7
8
11
2
12
Likewise on picture below, the equipment’s downside switch is on, and the light is on at
LED and display waiting condition of the machine.
.
12
OFF ON
At machine, the power is inputted, and proceed TRIANA program at PC, and push exposure
button and connect to the machine. OP program is on and start checking the machine.
If Rotating Unit is not rotating, and by hands, turning Rotating Unit, and push laser button
again( ).
When reversing back to its starting point, if the Rotating Unit does not stop
and contacts the main body, the use should press on the off button
immediately and replace the original place using the technical manual.
4. DIAGRAM
4.1 TOTAL BLOCK DIAGRAM
4.2 CT PWR B/D
1) POWER ON: By receiving main power from NOISE FILTER, it gets power from TB1.
2) PSU B/D power supply:Main power is going through LF1 and is going through CN1,
and main power (AC220V) is out toward PSU B/D.
3) TANK DRV B/D DC Voltage Supply : From PSU B/D, and going through CN3, and
after receiving +24V, and main power (220V) changes to DC 310V and permit to give
to CN2.
1) By coming from CN406, and goes through +24V, and goes through
relay(K1), and goes through DC CONVERTER(PSD10-24-1212), and
produce +24V,+12V,-12V.
2) Through relay(K1), +24V is going through PANO_CN1,CEPH_CN1 and
permit to give voltage to each board.
3) By using MCU, connecting each communication and controls the lift
through CN405.
4.4 MAIN CONTROL B/D
2) IP PWM CONTROL: Between IP_REF and IP_FB’s voltage difference, the frequency’s
range need to be controlled, and send signal (output) to IP_GATE1, IP_GATE2.
3) LEVER OVER: According to DRIVER B/D’s defect and TANK’s defect, FEEDBACK value
that is higher than one area or set-up value and send LEVER OVER signal to MAIN
B/D.
0.5V 이하
5V
5-1-1
5-1-3
5-1-2
5-1-4
5-1-6
5-1-5
PARTS LIST
NO. ITEM SERIAL NO. QNT REMARKS
5-1-1 BITE P03RP412 1
5-1-2 CHINREST P03RP409 1
5-1-3 TEMPLE SUPPORT A P03RP403 1
5-1-4 TEMPLE SUPPORT B P03RP404 1
5-1-5 CHINREST HANDLE P03RP408 1
5-1-6 CHINREST OP BOARD P03JB018 1
5.2 Main Column Assy.
PARTS LIST
5-3-1
5-3-3
5-3-2
PARTS LIST
No. Item Serial no. Qnt Remarks
5-3-1 Main board P03JB011 1
5-3-2 Driving LED board P03JB019 1
5-3-3 Pulley assy P03DS001 1
5.4 Gantry Assy.
5-4-1
5-4-6
5-4-3
5-4-2
5-4-4
5-4-5
PARTS LIST
5-5-1
5-5-5
5-5-4
5-5-2
5-5-3
5-5-6
5-5-7
PARTS LIST
5-6-4
5-6-5
5-6-1
5-6-2
5-6-3
PARTS LIST
No. Item Serial no. Qnt Remarks
5-6-1 Terminal Block Dft-20a-3p 1
5-6-2 LAN Connector JACK-3101 1
5-6-3 Spiral Cable Grand PECG001 1
5-6-4 Hanger’s D-sub Connector P03JH121 1
5-6-5 Waring system’s D-sub Connector P03JH122 1
6. Calibration method
- Refer ‘Machine Alignment & Image calibration’ document.
TROUBLE SHOOTING, Error message Solutions & modification
POWER ON
N
1) SYSYEM CHECK
Device ON
N 2) SYSYEM CHECK
IMAGE Normal
3) Detector CHECK
Device Normal
1) SYSTEM CHECK
① Is power of device turned on normally?
-Is LED1 turned on POWER SEQ
B/D? OFF : Check FUSE1
ON : Check +5V between U3 No. 2 & 5.
- Check FUSE B/D
Check F3,F4 FUSE
2) IMAGE CHECK
- In case of No image on the screen
Solution => Check whether X-RAY generates normally with KV METER.
=> Check the beam of X-RAY TANK by using a X-ray intensifying screen
=> Replace TANK ASS’Y & DRIVER B/D
=> Replace Detector
Replace motor.
Permanent Replace motor harness.
Solution Replace encoder.
Replace encoder harness.
Replace sensor.
Obelis s.a.
Boulevard Général Wahis 53, 1030 Brussels,
BELGIUM Tel: +32-2-7325954 Fax: +32-2-7326003
PAPAYA plus (PAPAYA)
Alignment Manual
Copyright© 2013
2013 / 06 / 07
Document Ver. 0.0
1. INTRODUCTION.............................................................................................................................3
1.1. Regular Condition......................................................................................................................3
1.2. Preparations................................................................................................................................3
1.2.1. Required Tools...............................................................................................................3
1.2.2. Jig or phantom set..........................................................................................................4
2. Machine’s Alignment.........................................................................................................................6
2.1. When Machine Production........................................................................................................6
2.1.1. Execution of Alignment Program and Connection with Machine.................................6
2.1.2. Execution of Calibration Wizard (Panoramic mode).....................................................7
2.1.3. Panoramic Mode’s Machine Alignment and Calibration..............................................9
2.1.4. Execution of Calibration Wizard (Cephalometric mode)............................................27
2.1.5. Cephalometric Mode’s Machine Alignment and Calibration......................................29
1.2. Preparations
<Wrench set>
<SD-line jig>
3 4
<Preparing>
Before machine alignment and modification, this explains necessary tools.
- 10 - Installation Manual
<Preparing2>
Before alignment of machine, on patient’s support part, and need to remove parts, in
which are machine’s cover and handle, and explains how to remove them.
- 11 - Installation Manual
On red circle, remove indicated bolts and remove
cover on that parts.
- 12 - Installation Manual
<Step1>
After preparing all preparation, set up Y-Axis starting point. Machine’s OP and using
machine’s ‘FORWARD’, ‘BACKWARD’ button and move machine’s Y-Axis to front
and back to match of indicated driving part’s and aluminium block for Y-axis, refer to
<Figure 2.1.3.3>. In this condition, push ‘Set current position’ button and then current
Y-Axis location is displayed and saved.
- 13 - Installation Manual
<Step2>
According to <Figure 2.1.3.5>, fix ‘Theta jig block’ to driving side and fix with bolts.
Then, installing ‘Theta jig block’ if because Y-Axis is coming front and rotating side
and ‘Theta jig block’ conflicts, and then using OP’s ‘BACKWARD’ button and move
Y-Axis to backward and make installation. On rotating part and ‘Theta jig block’
meets and parallel each other, and then push ‘Set current position’ button and save
current rotating side’s location by pushing button.
- 14 - Installation Manual
<Figure 2.1.3.6> Using of ‘Ball phantom’ and ‘SD-line jig’ (Panoramic)
<Figure 2.1.3.8> Setting of ‘Ball phantom’ and ‘SD-line jig’s starting point
- 15 - Installation Manual
Adjusting as repeating below steps to align ‘Ball phantom’ and rotating axis’ location.
1. By pushing ‘0°’ button, making perpendicular between rotating part and driving
part.
2. ‘SD-line jig’s line and ‘Ball phantom’s reference line need to be matched by
turning ‘Ball phantom’ stage’s knob and move ‘Ball phantom’s location.
3. ‘180°’s button need to be pressed and rotating part move to 180°.
4. If, at 0° location and 180° location, ‘SD-line jig’ is not on same location, on 0°
adjusted ‘Ball phantom’s reference line and at 180° location. ‘Ball phantom’s
reference line and ‘SD-line jig’s line that needs to be located by turning each
knob and locating them to the middle of ‘SD-line jig’s locations of 0° and 180°.
5. By repeating 1-4’s to reduce the error.
After steps of 1~5, ‘SD-line jig’ is located on machine’s rotating axis. On this
condition, according to <Figure 2.1.3.9>, by using ‘0°’, ‘90°’, ‘180°’ button, at each
degree’s location, ‘SD-line jig’s line is at ‘Ball phantom’s matching with appropriate
reference line and adjust chin-rest’s location and tighten chin-rest’s bolt. Then, need to
be caution to not move ‘Ball phantom’s knob.
After chin-rest is tightly fixed, again at ‘0°’, ‘90°’, ‘180°’ location, and checking ‘Ball
phantom’ and ‘SD-line jig’ are matching. If location is not matching, again chin-rest’s
bolt need to be unscrew and adjusted.
- 16 - Installation Manual
<Step3>
On Step3, before Collimator’s accurate alignment, this is step where finding estimated
position. On Step2, ‘Ball phantom’ is installed on PAPAYA and click ‘Find Collimator
Offset’ button. After clicking button, the machine is automatically finding for
Collimator’s position by proceeding necessary steps and displays ‘Adult Panoramic
collimator position’s value. In this condition, click ‘Set Offset & Get Image’ button
then X-ray is exposed and display an image. Then distinguishing whether collimator’s
is good position by checking with left illustrated picture. If Collimator’s position is
well found, according to picture’s Good position, on image, by Collimator, X-ray’s
exposure’s limited area is almost same with left area and right area.
- 17 - Installation Manual
<Step4>
On Step4, this step where finding detector’s position so that the X-ray beam center and
the center of rotation are on a straight line. Repeat the following process.
1.Click ‘Rotate 0 ° & Get Image’ button to expose X-ray and to get an X-rayed
image.
2.Select line1 of ‘Select Line’ Combo box located on bottom of display window
and then click the center of the pin, imaged by X-ray exposure, then blue
dashed line is being on the center of the pin.
3.Click ‘Rotate 180 ° & Get Imagte’ button to expose X-ray and to get an X-rayed
image.
4.Select line2 of ‘Select Line’ Combo box located on bottom of display window
and then click the center of the pin, imaged by X-ray exposure, then yellow
dashed line is being on the center of the pin.
5.Red line is being on the middle of two lines between blue and yellow line. Check
whether the red line is located inside of two green line or not (Pass/Fail
information will be displayed). If fail, refer to <Figure 2.1.3.11>, then loosen
screws marked with red circle and move the detector position left or right and
- 18 - Installation Manual
repeat 1-5 until Pass is displayed.
In the condition Pass in on, refer to <Figure 2.1.3.12>, join the red circled screws and
put the detector cover on.
- 19 - Installation Manual
<Step5>
On Step5, this step is where to adjust rotation origin angle which is set before by using
‘Theta jig block’ exactly. Repeat following procedure until Pass is being displayed on
the window.
1.Click ‘Set angle and Get image’ to expose X-ray and to get an X-rayed image of
degree 90°.
2.Select blue line of ‘Select Line’ Combo box located on bottom of display
window and then click the center of the pin, imaged by X-ray exposure, then
blue dashed line is being on the center of the pin
3.Select yellow line of ‘Select Line’ Combo box located on bottom of display
window and then click the center of the metal ball, imaged by X-ray exposure,
then yellow dashed line is being on the center of the pin
4.Make a judgment of Pass/Fail. If failed, refer to <Figure 2.1.3.13> , then adjust
‘Gantry rotation angle offset of 90°’ value slightly little. Repeat 1-4 until Pass
is displayed
- 20 - Installation Manual
<Step6>
- 21 - Installation Manual
<Step7>
On Step7, this step is where to adjust collimator position exactly. Repeat following
procedure until Pass is being displayed on the window.
- 22 - Installation Manual
<Figure 2.1.3.16> Adjustment of Collimator
- 23 - Installation Manual
<Step8>
- 24 - Installation Manual
<Step9>
<Step10>
Step10 is a step to check whether the whole calibration and alignment process
performed rightly or not.
Install ‘Ball pahntom’ to PAPAYA and press ‘Confirm’ button and when the ‘Ready &
- 25 - Installation Manual
Shot’ button is been able then click the button to take panoramic of ‘Ball phatom’.
When it finish to reconstruct panoramic of ‘Ball phantom’, draw 4 lines by using
‘Select line’ combo box like as before steps, refer to the image on bottom side. If get
‘Pass’, then every calibration and alignment of panoramic is finished. Or if get ‘Fail’,
then refer to appendix to find solution and repeat step10 to check.
<Offset Overall>
<Figure 2.1.3.20>
Offset Overall – Live image and manual input of Offset parameter
(Panoramic)
Offset Overall, possible to input offset parameter manually and to take live X-ray
image.
- 26 - Installation Manual
2.1.4. Execution of Calibration Wizard (Cephalometric mode)
Connect to PAPAYA and choose Cephalometric tab, refer to <Figure 2.1.4.1>.
Click 'Calibration Wizard' button at the bottom right in the Cephalometric tab (<Figure
2.1.4.2>)
When you click the button, Calibration Wizard is run like <Figure 2.1.4.3>.
- 27 - Installation Manual
1
3 4
- 28 - Installation Manual
2.1.5. Cephalometric Mode’s Machine Alignment and Calibration
This section explains about what to do on each step for machine’s alignment and
calibration. User could change the steps by using step tab, item number 1 of<Figure
2.1.4.3>. When production of machine, progress with each steps to align and to
calibrate machine.
<Preparing>
Explanation of preparations and tools.
.
<Figure 2.1.5.1> Preparing Step(Cephalometric)
The image on bottom side explains how to remove the covers ahead to calibration and
alignment.
- 29 - Installation Manual
Remove the top cover.
The position is marked with a red circle and remove the
screws to remove the cover and lift it up.
- 30 - Installation Manual
<Step1>
Once all the preparations is ready, set the detector’s reference position. Use 5mm pin,
refer to <Figure 2.1.5.2>, to set detector position to the reference hole of cephalometric
base. When the detector is on the reference point, then remove the pin carefully. Click
‘Set current position’ button.
- 31 - Installation Manual
<Step2>
Step2 is where to align 1st collimator position. Click ‘Find Collimator Offset’ button
then procedure will proceed automatically.
When automatic procedure is finished, then click ‘Set Offset & Get image’ button to
check 1st collimator is on proper position.
- 32 - Installation Manual
<Step3>
Step3 is a step to check whether X-ray and ear posts are on a straight line or not and to
align their to be on a straight line. The laser alignment also is performed in this step.
First, adjust the height of two ear posts so that they have same length, refer to <Figure
2.1.5.5>.
- 33 - Installation Manual
<Figure 2.1.5.6> Adjustment of Cephalometric laser
Set ear posts and nasal positioner to Lateral direction and perform this below
procedure for alignment of base angle of cephalometric part.
- 34 - Installation Manual
<Figure 2.1.5.7> Adjustment of base angle and Pass/Fail determination
- 35 - Installation Manual
<Step4>
- 36 - Installation Manual
<Step5>
- 37 - Installation Manual
<Step6>
Step 6 performs FFC(Flat Field Correction) and Scan calibration to calibration 2nd
collimator position of scanning and to equalize brightness of cephalometric image.
First, set ear posts and nasal positioner to AP direction and click ‘Auto scan/FFC
calibration’ button, then procedure will proceed automatically.
- 38 - Installation Manual
<Step7>
Step 7 is a step to test that whole steps for cephalometric calibration is performed well
until now.
First, set ear posts and nasal positioned to Lateral direction and click ‘Confirm’ button.
When it is ready then click ‘Ready & Shot’ button to take cephalometric image.
Determine, when the reconstruction is finished, whether there is a problem or not, refer
to right image of <Figure 2.1.5.11>.
If you find there is a problem on the image. Then refer to Appendix to find solution for
each problem case. And calibrate and align PAPAYA again by following the instruction.
And then do step7 again.
- 39 - Installation Manual
<Offset Overall>
<Figure 2.1.5.12>
Offset Overall – Live image and manual input of Offset parameter
(Cephalometric)
Offset Overall, possible to input offset parameter manually and to take live X-ray
image.
- 40 - Installation Manual
GENORAY Co., Ltd.
#512 Byucksan Technopia 434-6 Sangdaewon 1-Dong Jungwon-Gu
Seongnam-City Gyeonggi-Do Korea, 462-716
TEL : +82-31-737-8020 FAX : +82-31-737-8025
www.genoray.com
Obelis s.a.
Boulevard Général Wahis 53, 1030 Brussels,
BELGIUM Tel: +32-2-7325954 Fax: +32-2-7326003
- 41 - Installation Manual
- !GENORAY