Dental X-ray Imaging System PAPAYA

Consolidated Manual

This Product is Medical Equipment.

The specification above can be changed to improve performance.


Copyright© 2012
2013 / 07 / 15
Document Ver. 0.3
Table of Contents

Table of Contents

Introduction of Manual............................................................................................................................................................5
Certificate of Warranty............................................................................................................................................................6
Symbols used in product.........................................................................................................................................................7
Warning before Use.................................................................................................................................................................8
Electrical Safety..........................................................................................................................................................................9
Radiation Safety......................................................................................................................................................................11
1. PAPAYA Plus Equipment Introduction........................................................................................................................12
1.1 Intended Use.........................................................................................................................................................12
1.2 Equipment’s Characteristics...............................................................................................................................12
1.3 System Basic Specification.................................................................................................................................13
1.4 Equipment Composition.....................................................................................................................................14
1.5 Accessories............................................................................................................................................................15
1.6 Patient Apply Part................................................................................................................................................16
1.7 On / Off Switch....................................................................................................................................................17
1.8 Hand Switch..........................................................................................................................................................18
1.9 Emergency Stop Switch......................................................................................................................................19
1.10 Equipment’s Control Button..............................................................................................................................20
1.11 Equipment’s 2nd Control Button.......................................................................................................................21
1.12 Laser Control Knob..............................................................................................................................................22
1.13 Changing Detector...............................................................................................................................................23
1.14 Device status LED.................................................................................................................................................24
1.15 External status LED..............................................................................................................................................25
1.16 Changing Fuse......................................................................................................................................................26
1.17 Detector status LED.............................................................................................................................................27
2. OP (Operation panel)......................................................................................................................................................28
2.1 OP Structure..........................................................................................................................................................28
2.2 X-ray Exposure Mode..........................................................................................................................................30
2.2.1 Panorama program...................................................................................................................................30
2.2.2 Cephalo Program......................................................................................................................................34
2.3 Choose Patient Size.............................................................................................................................................36
2.4 Exposure Speed and Change of Image Quality.............................................................................................36
2.5 Selection of Jaw shape.......................................................................................................................................36

2 PAPAYA PLUS User manual


Table of contents

2.6 X-ray exposure condition change....................................................................................................................37


2.7 Order list................................................................................................................................................................37
2.8 Device condition display screen.......................................................................................................................39
2.9 Option setting.......................................................................................................................................................40
2.9.1 General........................................................................................................................................................41
2.9.2 DB Server....................................................................................................................................................42
2.9.3 DB Server....................................................................................................................................................43
2.9.4 Version.........................................................................................................................................................44
3. Image acquisition............................................................................................................................................................45
3.1 Registration of patients and orders.................................................................................................................45
3.1.1 Registration with Triana software....................................................................................................................45
3.1.2 Registration with OP software..........................................................................................................................46
3.2 Panorama mode Set position of patient........................................................................................................47
3.3 TMJ mode patient position set.........................................................................................................................51
3.4 Sinus mode patient position set.......................................................................................................................56
3.5 Cephalo mode patient position set.................................................................................................................61
3.6 Image taking.........................................................................................................................................................63
3.7 Image Control.......................................................................................................................................................65
3.8 TWAIN INTERFACE...............................................................................................................................................66
3.9 OP Additional Function......................................................................................................................................66
3.9.1 Sending to DICOM Server.......................................................................................................................67
3.9.2 Burning CD.................................................................................................................................................67
3.9.3 Register addition order............................................................................................................................68
3.9.4 Opening of patient's folder.....................................................................................................................68
4. Use condition...................................................................................................................................................................69
5. Cleaning.............................................................................................................................................................................70
6. Handling of waste...........................................................................................................................................................70
7. Message.............................................................................................................................................................................71
7.1 Error message.......................................................................................................................................................71
7.2 Condition message..............................................................................................................................................73
8. Technical Data..................................................................................................................................................................74
8.1 Classification.........................................................................................................................................................74
8.2 Power Requirements...........................................................................................................................................74
8.3 Complied Standards............................................................................................................................................74

PAPAYA PLUS User manual 3


Table of Contents

8.4 X-ray source assembly........................................................................................................................................75


8.5 Beam Limiting Device.........................................................................................................................................75
8.6 Detector.................................................................................................................................................................76
8.7 Dimensions and Weights...................................................................................................................................77
8.8 Movement..............................................................................................................................................................78
8.9 Other Information................................................................................................................................................78
8.10 X-ray Tube.............................................................................................................................................................79
7.10.1 OPX – 105 (CEI)............................................................................................................................................79
7.10.2 D-054SB (Toshiba).......................................................................................................................................81
9. Labeling.............................................................................................................................................................................83
9.1 PAPAYA...................................................................................................................................................................83
9.2 PAPAYA Plus..........................................................................................................................................................84
Appendix 1...............................................................................................................................................................................87

4 PAPAYA PLUS User manual


Introduction of Manual

Introduction of Manual

GENORAY’s PAPAYA Plus is digital panoramic X-ray system. This manual is designed for your safe use and
operation of PAPAYA Plus. When using PAPAYA Plus, use .

The potential risk always exists when using X-ray products and electric medical equipments. All users should
be aware of the methods of use, solutions in emergency conditions and other safety articles provided in this
manual.

The user must acknowledge the appropriate ways to protect oneself from the potential danger that could
occur at any time, and be aware of the condition of the product that can cause danger.

The manufacturer, assembler, and the importer hold responsibility upon the performance of safety, trust
and the article below.
- Guarantees the installation of the eligible professional, calibration, and repair.

GENORAY Co., Ltd is continuously developing the products. The manual or composition of the package
might change without notice

Tel : +82-31-737-8020 FAX : +82-31-737-801 8


E- mail : genoray@ genoray. com
Address : #512 Byucksan Technopia 434-6, Sangdaewon 1 -Dong, Jungwon- Gu,
Seongnam-City, Gyeonggi- Do, Korea

5 PAPAYA PLUS User manual


Table of Contents

Certificate of Warranty

Range and terms for warranty

GENORAY guarantee this product without any expense for the mal-function or disorder occurred under
standard transportation and operation within 1 year from installation. Defect on its appearance is out of the
range of warranty.

Requirements on After service call

When the malfunction or disorder appeared, stop operating and check again the related article of the “User’s
Manual.” Before your calling, please put off this equipment and check for the Model Name, serial No. and
purchasing date. GENORAY Co., Ltd. is not responsible for indirect harm. GENORAY Co., Ltd. cannot warrant
for defect or harm after warranty period.

6 PAPAYA PLUS User manual


Warning before Use

Symbols used in product

Symbol description Symbol description

X-Ray Source Assembly,


Manufacturer, Address
X-ray radiation

Caution Manufacturing Date

Refer to User’s Manual Product serial number

Type b applied part WEEE Mark

AC (Exchange) Laser

This symbol is information about the special matters that require attention while using this software.

This symbol is warning the situation or environments which might bring danger to the patients or users.

PAPAYA PLUS User manual 7


Electrical Safety

Warning before Use

 This machine can be dangerous for all users if the safe exposure condition and appropriate using
method are not attended.

 This machine can only operated by authorized user.

 This machine need to operate after studying of manual.

 Random modification on machine is prohibited.

 There is prohibition of installing different composite parts unless they are supplied by manufacturer.

 Do not connect equipment’s power source with multi-adaptor or extension cord. This can cause defect or
become cause of mal-function.

 Do not operate machine when the machine is defect or malfunctioning. Please contact engineer or
service center for help.

 Do not use provided PC for unintended purpose. There is probability some problem could happen on
transferring images and reconstructing if the PC uses for unintended purpose.

8 PAPAYA PLUS User manual


Warning before Use

Electrical Safety

The electric circuit of this product uses the electric current which can cause death due to severe
injury or electric shock. To avoid this kind of danger, do not remove any of the covering part.

To avoid failure and the danger that might cause serious injury or electric shock to users and patients, please
acknowledge the safety procedures listed below.

 Before cleaning the product, always turn off the power and separate the power cord, and clean the
product with lightly wet cotton or sponge.
 Never let anyone remove the cover except for the professional technician of GENORAY.
 Do not place food on any part of the product. They might get into the electric circuit and get into
contact with electrical matters

The product is not waterproof to water, soap or liquid substances. If the liquid contacts the
product, it may cause fire or electric shock. If the liquid was in the product by accident, do not
touch the power cord connecting to the power, and wait until the liquid is completely dry.

The user should follow the safety guideline provided below.

 Turn off the power and separate the power cable from the socket.
 Stay far away from the place where the product has been installed.
 Ask for help around you.
 Use the fire extinguisher authorized only for the electric fire.

Abnormal use of fire extinguisher could be dangerous and cause fire or electric shock. To avoid
this kind of danger, check to see if the product can be used where the product has been
installed, and always use the authorized product.

PAPAYA PLUS User manual 9


PAPAYA Equipment Introduction

Before use this product, please check below.

 Before connecting the power, check to see of the power supplying line is appropriate to this product.
Check the power and connect the power cable of this product to its power socket.
 Check the categories mentioned above in case of power input of all accessory (the signal input/output
including accessory).

For reducing the risk of electrical shock, and when connecting to power, there needs protecting
ground connection.

Place this product away from other X-ray products, power generating equipment and
broadcasting stations. If the product is sharing the power with other electric devices, abnormal
image might occur.

The protection level of PAPAYA Plus is the 1st grade equipment or type-B equipment. This
product does not prevent the invasion of liquid substances.

10 PAPAYA PLUS User manual


PAPAYA Equipment Introduction

Radiation Safety

 This product has to follow designated country’s radiation law.

 This product needs X-ray room. Except the patient, no one can enter X-ray room unless they are patient.
While taking exposure, clear space between patient and equipment need to be maintained.

 For this product’s safe operation, it needs to be installed at appropriate location with permission of
owner. Also, user needs to guarantee about whether equipment will be used at place where owner has
chosen.

 This product’s owner needs to be guaranteed that all users have to wear X-ray protection devices while
using this equipment.

 For this product, users need to wear TLD badge, and check dosage exposure regularly.

 If the pregnant woman or young child needs to have dental X-ray taken, consult with specialist. It should
be considered to balance the benefit of the test and potential small risks of x-ray tests by specialist such
as doctors and the radiologist.

If you do not follow written rules in this manual, there can be risk for operator and patient.

PAPAYA PLUS User manual 11


PAPAYA Equipment Introduction

1. PAPAYA Plus Equipment Introduction

1.1 Intended Use


PAPAYA Plus is a digital extraoral source X-ray system intended to produce panoramic and cephalometric
images of the oral and craniofacial anatomy for a precise treatment planning in adult and pediatric care. The
system is used for dental & skull radiographic examination and diagnosis of teeth, jaw, oral structure, and
skull by exposing an X-ray image receptor to ionizing radiation, with a digital imaging capability for taking
both panoramic and cephalometric images.

Federal law restricts this device to sale by or on the order of physicians and dentists.

1.2 Equipment’s Characteristics


This equipment is used for diagnosing of teeth, chin, and head. This equipment is producing and is
controlling X-ray, and composed of X-ray generator, X-ray control part, and X-ray support part. The basic
concept of acquiring image is to produce high voltage between X-ray tube’s both positive and negative, and
exposure X-ray to subject, and transmitted X-ray to exposure at sensor and acquire the image.

High Frequency, Fixed Type Tube,


X-ray Generator 60 ~ 90kV, 4 ~ 12mA

Focal Spot Size 0.5mm


Image Detector CdTe flat panel

Panorama: 4.8mm x 150mm


FOV Ceph: 4.8mm x 240mm

Image Acquisition / Angle


(Standard panoramic) 210°

Scan Time / Exposure Time Panorama: 17sec(max), Cephalo: 17sec(max)

Patient Position Standing

Reconstruction Type Fan beam

Reconstruction Time Less than 30 seconds

Weight 110 kg ± 5kg (Basic), 140 kg ± 5kg (Cephalo)

Main body 1010 x 1055 x 2312 mm (Basic)


1010 x 1775.4 x 2312 mm (Cephalo)

12 PAPAYA PLUS User manual


PAPAYA Equipment Introduction

1.3 System Basic Specification

PAPAYA Plus

[ System Connection Image ]

PC is used for acquiring and displaying image, and for image processing.

PAPAYA PLUS User manual 13


PAPAYA Equipment Introduction

1.4 Equipment Composition

This machine has panoramic only model of PAPAYA and has added cephalo exposure that is
PAPAYA Plus. Each model’s composition or exterior, please refer to below pictures.

4 9

13

5 10
7
8 14

2 11

[ PAPAYA ] [ PAPAYA Plus ]

No. description 1 No. 12 description


1 BASE 7 Patient supporting area

2 Fixed column 8 Equipment’s control button

3 Driving Unit 9 Cephalo Arm

4 Rotating Unit 10 On/off switch

Hand Switch
5 X-ray source assembly 11
(Exposure button, Emergency stop switch)
12 Emergency stop switch
Detector
6 13 Device status LED
(Panorama, Cephalo using for both,
detachable structure)
14 Equipment’s 2nd control button

14 PAPAYA PLUS User manual


PAPAYA Equipment Introduction

1.5 Accessories

Patient’s condition or according to exposure mode, chin rest can be changeable.

[Standard Exposure] [For No Teeth Patient] [SINUS & TMJ mode]

 Chin rest change method

For appropriate exposure mode, insert correct chin rest to designated hole.

Must use spare part that is provided by manufacturer. If you damage machine when using
different spare part, and then the manufacturer does not have responsibility of fault.

PAPAYA PLUS User manual 15


PAPAYA Equipment Introduction

1.6 Patient Apply Part

Cephalo part

Panorama part

Bite

Chin rest

Temple support

Nasal
Patient handle
positional
Ear rod
2nd control
button

After exposure, sanitize patient apply part.

16 PAPAYA PLUS User manual


PAPAYA Equipment Introduction

1.7 On / Off Switch

OFF ON

Equipment’s on/off switch is like on picture, it is located at equipment’s below part. If turning on
equipment’s switch, and the equipment’s LED has light and displays equipment’s waiting condition.

PAPAYA PLUS User manual 17


PAPAYA Equipment Introduction

1.8 Hand Switch

Hand switch is made with magnet and can be hung


on wall. At hand switch, as seen on picture right,
there is X-ray exposure switch and emergency stop
switch, they are two functions together. While X-ray exposure switch
emergency switch is used when equipment is
malfunctioning or it can harm patient. At moment,
equipment will stop.

Emergency stop switch


Exposure switch need to be pushed continuously while is to exposure patient’s diagnostic image.
If during exposure, switch is not pushed and machine’s movement is stopped instantly and at
OP, advisor word will be displayed.

18 PAPAYA PLUS User manual


PAPAYA Equipment Introduction

1.9 Emergency Stop Switch

Emergency stop switch is located at above picture. When emergency situation happens and push button and
equipment’s driving unit or X-ray source assembly’s power supply stops and machine will stop.

When pushing emergency stop switch, turning switch to clockwise to unwind and machine will
operate normally.

PAPAYA PLUS User manual 19


PAPAYA Equipment Introduction

1.10 Equipment’s Control Button


Equipment is controlled by equipment’s control button and controls equipment’s movement or controls laser.
Each button’s function is as below.

Controlling equipment’s height. ‘LIFT DOWN’ button moves equipment


downward and ‘LIFT UP’ button moves equipment upward.

Using when user wants to change the detector.

Turn on/off equipment’s laser. When pushing once, laser turns on and pushing
one more time and laser turns off. Laser will turn off after time. It will turn off
automatically.

Using when equipment’s gentry is moved forward and backward. By using


laser, use to decide patient’s exposure location. ‘FORWARD’ button used for
moving gentry to user and ‘BACKWARD’ button used for moving genty,
opposite side to user.

20 PAPAYA PLUS User manual


PAPAYA Equipment Introduction

1.11 Equipment’s 2nd Control Button


Equipment is controlled by equipment’s 2nd control button and controls equipment’s movement or controls
laser. Each button’s function is as below.

Controlling equipment’s height. ‘LIFT DOWN’ button moves equipment


downward and ‘LIFT UP’ button moves equipment upward.

Turn on/off equipment’s laser. When pushing once, laser turns on and pushing
one more time and laser turns off. Laser will turn off after time. It will turn off
automatically.

PAPAYA PLUS User manual 21


PAPAYA Equipment Introduction

1.12 Laser Control Knob

Laser Control Knob

This equipment, as shown above picture, has laser control know. Before to exposure, when controlling
patient’s position, using laser control knob to adjust to Frankfort line.

22 PAPAYA PLUS User manual


PAPAYA Equipment Introduction

1.13 Changing Detector


This equipment is according to exposure mode, detector needs to move.

When to exposure Panorama


Detector’s Location

When to exposure Cephalo


Detector’s Location

 Detector Detachable Method

A. Pushing detector handle area’s B. While pushing switch, pull C. By disassembling hook part
switch. detector downward. and pushing detector upward
and detector will fall out.

Following above change method and doing reverse way and detector will be attached.

PAPAYA PLUS User manual 23


OP (Operation Panel)

1.14 Device status LED

LED Color description

Blue Rotating side is at rotation

Yellow while X-ray exposure

Green When X-ray exposure, ready

Red While taking exposure, error occurs or pushing


emergency switch

24 PAPAYA PLUS User manual


OP (Operation Panel)

1.15 External status LED

LED Color description

Blue Rotating side is at rotation

Yellow while X-ray exposure

Green When X-ray exposure, ready

Red While taking exposure, error occurs or pushing


emergency switch

PAPAYA PLUS User manual 25


OP (Operation Panel)

1.16 Changing Fuse


If equipment’s fuse is break, user can replace the fuse.

Patient or uneducated person must not operate equipment. Please be aware.

 Replace Fuse Method

A. Disassemble the B. Replace the broken fuse C. Assemble the fuse holder
fuse holder using with new fuse using driver
driver

26 PAPAYA PLUS User manual


OP (Operation Panel)

1.17 Detector status LED

LED Color description

Blue while detector is active

Blue while detector is active

Green while power is on

PAPAYA PLUS User manual 27


OP (Operation Panel)

2. OP (Operation panel)

2.1 OP Structure

Manual’s used picture can differ from real program’s version.

Patient or uneducated person must not operate OP. Please be aware.

1 9

2 11

4
12

10 6
5

7 8

[ Operation Panel Program / Main Screen ]

28 PAPAYA PLUS User manual


OP (Operation Panel)

No. Subject description

Exposure Time Display


1 Displaying current time.
Screen
Patient’s Information
2 Displaying patient’s information who wait for current exposure.
Display Screen
Exposure Mode Choose
3 Choosing equipment’s exposure mode.
Button
Exposure Program Choose
4 Choosing equipment’s exposure mode by exposure program.
Button
Patient’s Size Choose
5 Choosing according to patient’s size by pre-set value.
Button
Image Quality Choose
6 Deciding exposure image’s quality.
Button
Exposure Condition Display
7 Controlling equipment’s exposure condition manually.
and Change Part
Equipment Movement Deciding exposure condition and preparation completion and
8
Button Part opinion.

9 Equipment display Display equipment’s condition

10 Jaw shape selection button Choosing patient jaw shape and size

11 Order list button Managing patient information and order list

12 Exposure Program Display Displaying selected exposure program

ease before take an exposure, always look equipment’s display and check whether equipment can
take an exposure. About equipment’s condition, please refer to 2.6.

PAPAYA PLUS User manual 29


OP (Operation Panel)

2.2 X-ray Exposure Mode


Papaya is supporting various types of X-ray exposure mode. At 2.2.1 of Operation Panel Program, at main
screen, by pushing wanted x-ray exposure mode and push x-ray button and be able to choose program.

2.2.1 Panorama program

Exposure Mode Mode Type Program

Standard Panoramic
Standard panoramic has standard panorama’s image field of view and
Taking exposure.

Panorama mode

Orthogonal Panoramic
Orthogonal Panoramic is comparison to Standard panoramic
program that takes x-ray exposure with x-ray beam’s angle as 90° of
penetration to chin, and produce field of view. At standard
panoramic image, teeth is overlapping and effect is decreasing, but
exposure area’s opposite side’s shadow of teeth and chin and add an
effect.

30 PAPAYA PLUS User manual


OP (Operation Panel)

Exposure Mode Mode Type Program

 Bitewing Panoramic
Bitewing Panoramic program works at panoramic image and take
exposure of bitewing area and take exposure of trajectory of patient.
In basic, Bitewing Panoramic is orthogonal panoramic and take
exposure with same angle.

 TMJ lateral
TMJ lateral exposure has function of trajectory and side of
temporomandibular joint. Taking exposure of both sides of
temporomandibular joint and comparing, patient opens mouth and
the patient closes mouth to distinguish temporomandibular joint
exposure.

TMJ mode

 TMJ PA
TMJ PA takes exposure of temporomandibular joint’s PA direction
and take exposure with trajectory direction. Both sides of
temporomandibular joint can take exposure and using as patient
opening mouth and closing mouth.

PAPAYA PLUS User manual 31


OP (Operation Panel)

Exposure Mode Mode Type Program

 Double Lateral-PA
Lateral – PA program takes exposure of temporomandibular joints’
PA and Lateral direction alternately. Both sides of
temporomandibular joint can take exposure and using as patient
opening mouth and closing mouth.

 Sinus Midsagittal
Sinus midsagittal exposure enables to take exposure of maxillary
sinus and use it. The image’s focus is on midsagittal surface.

Sinus mode

 Sinus lateral
Sinus lateral exposure is like Sinus midsagittal, at maxillary sinus,
take a exposure at maxillary sinus’ trajectory. From midsagittal
surface’s standard, now focus surface can be toward left or right
side.

32 PAPAYA PLUS User manual


OP (Operation Panel)

Exposure Mode Mode Type Program

 Sinus PA
Sinus PA takes an exposure at PA’s direction of maxillary sinus.

PAPAYA PLUS User manual 33


OP (Operation Panel)

2.2.2 Cephalo Program

Exposure Mode Mode Type Program

Cephalo lateral
Cephalo lateral exposure has at patient’s side, and X-ray beam’s
penetration of direction.

Cephalo mode
Cephalo AP
Cephalo AP exposure is that at toward AP’s direction, X-ray
penetrate and taking exposure of skull.

Cephalo PA
Cephalo PA exposure is that at toward PA’s direction, X-ray penetrate
and taking exposure of skull.

34 PAPAYA PLUS User manual


OP (Operation Panel)

Exposure Mode Mode Type Program

 Water’s view
Water's view is when the patient is at with Frankfort line and wit
h detector surface, and taking exposure with 45 degree angle slo
pe and is exposuring area of head.

 SMV(Submento vertex)
SMV takes exposure of according to patient’s Frankfort line and the
surface of detector is leveled with horizon and put patient in this
position to take exposure.

 Carpus
Carpus takes exposure of wrist. In order to do Carpus exposure,
using carpus exposure wrist panel and install it and take exposure.

PAPAYA PLUS User manual 35


OP (Operation Panel)

2.3 Choose Patient Size


At all exposure program, according to patient’s size and can choose appropriate patient’s size.

Among 4 different size, can choose one size, and according to choice of size of patient, each exposure
program has saved X-ray exposure condition with trajectory. At each exposure program, X-ray exposure
condition is set up differently.

2.4 Exposure Speed and Change of Image Quality


At each exposure program, you can choose different image quality. The image quality can choose between
Normal, Fast, HD(High Definition). They are total of 3 types and can choose one.

Normal Fast HD ( High Definition)

Fast image has lower image quality and HD has longer image processing time but image.

2.5 Selection of Jaw shape


Sharp Normal Square
Small

Average

Large

You can choose different jaw shape and size of patients. There are 9 options by following 3 jaw shape (Sharp,
Normal, Square) and 3 jaw size (Small, Average, Large).

36 PAPAYA PLUS User manual


OP (Operation Panel)

2.6 X-ray exposure condition change


X-ray condition can be changed by clicking the left or right arrow button of kVp & mA condition value on
the x-ray condition display screen like the below image. kVp can be changed 1kVp per 1 click and mA can be
changed 0.5mA per 1 click. The exposure time can’t be changed.

The exposure condition value is chosen automatically when choosing patient size & exposure
program. If you’d like to change the exposure condition by yourself, you can do like above.

2.7 Order list

2 3
1

PAPAYA PLUS User manual 37


OP (Operation Panel)

Click the order list button, refer to 2.1, then the order list management window will be appeared. Functions of
each part are described in below table.

No. Subject description


Displaying patients’ information of registered order by following
selected filtration option.
1 Order status filter Ready: Displaying order information be in ready and wait status.
All: Displaying all order information.
Done: Displaying order information exposure is finished.

2 Worklist Loading patients’ information from worklist server.

3 Search Searching order information.

Displaying registered order information. Time, status, ID, name and


4 Order information window
exposure program information are displayed.
Managing order.
Register: Executing order registration window.
5 Order management button Select: Selecting registered order of order list for taking image.
Modify: Modifying registered order information.
Delete: Deleting registered order information.
Close: Closing order list and returning to main screen of OP.

This below picture shows a window appears when you click ‘Register’ button, user can input patient information
and select exposure program with this window.

38 PAPAYA PLUS User manual


OP (Operation Panel)

2.8 Device condition display screen


Device condition display screen shows the current condition of device.

Condition Description

It shows the connection of device & PC. When the device is connected with
PC, it turns orange color.

It shows the temperature of X-ray generating part of device. If the X-ray


generating part is overheated, it turns red color. In this case, please don’t
exposures X-ray until the temperature becomes normal.
Green: normal, Orange: caution, Red: warning.

It shows the connection condition of the device sensor to the panorama or


cephalo position. “P” turns violet when the sensor is connected to panorama
position. And “C” turns green when the sensor is connected to panorama
position.

PAPAYA PLUS User manual 39


Image acqusition

2.9 Option setting

PAPAYA OP program needs to set

No. Name Explanation

1 General Display composition or user's convenience need to be set

2 DB Server Material composition or data management need to be set.

3 DICOM Outside compatibility or DICOM function need to be set

4 Version Machine or program version need to be displayed

40 PAPAYA PLUS User manual


Image acquisition

2.9.1 General

Name Explanation

Language Setting language

Font Setting Font

DAP Setting DAP output measurement

Data Format Setting data format.

Clinic Name Save by inputting Clinic Name

Volume control Controlling machine's sound

Setting Multi Focus using or not using

Setting multi focus' pages


Multi Focus Number (3,5 page setting is abled)

Step Setting space between images

All save Saving images that is processed with Multi focus

Automatically setting processing / no processing


Auto Process
After taking exposure, automatically send with DICOM
Send DICOM
format
Test Operation Setting for exposure test.

R-Mark Setting exposure images' position mark's printing out / not printing out

Use Cephalo Adding cephalo mode function (*Option)

Save Raw images Setting Raw image's setting (If checking and taking up a lot of hard disk

PAPAYA PLUS User manual 41


Image acqusition

2.9.2 DB Server

Name Explanation

Server Name Setting Order DB's usage path


SQL Server Information
Triana Server Name Setting DB path at Triana

Temp Image At PAPAYA OP, setting saved image path

Data folder Information At Triana, by managing image, making im


Triana Image
age saving path

Parameter Setting parameter path

Trouble shoot When problem happens, pushing button to solve problem.

42 PAPAYA PLUS User manual


Image acquisition

2.9.3 DB Server

Name Explanation

Linkage Setting about interface about compatibility

Equipment Information Inputting machine's information for DICOM usage

Storage Server Setting DICOM Storage server

MWL Server Setting Work list server

DICOM Printer Setting DICOM Printer

System Log When pushing test button and printing out processing result

PAPAYA PLUS User manual 43


Image acqusition

2.9.4 Version

Name Explanation

PAPAYA OP Displaying program version at PAPAYA OP

PAPAYA Bridge Displaying program version of PAPAYA Bridge

PAPAYA REC Displaying version of PAPAYA REC program

M/B Displaying Papaya's main board version

sound Displaying PAPAYA sound controller

LIFT Displaying PAPAYA LIFT version

Sensor Serial num Displaying sensor serial number

44 PAPAYA PLUS User manual


Image acquisition

3. Image acquisition

3.1 Registration of patients and orders.


3.1.1 Registration with Triana software.

Please refer the Triana manual provided with device for more details.

Please follow the process below after running Triana software to operate OP for the device operation.
Patient registration

Click the patient registration button after


running Triana.

Put patient information in


patient registration window.
Click the patient registration
after done.

A. OP operation

[ Panorama ] Choose the device among device selection tap once the patient
registration is done.
Papaya is a panoramic device so you can choose panorama. If the case
[ Cephalo ] of the device including cephalo function, you can choose one between
panorama & cephalo.

Order information will be registered when you click the acquisition


button.

PAPAYA PLUS User manual 45


Image acqusition

3.1.2 Registration with OP software.

Execute order registration input window; Refer


to 2.7, Press register button positioned at
bottom-left of order list.
Input patient information and select exposure
program to the window. Order information will
be registered when you click the register button.

Then you can see the registered order


information in the order list.

46 PAPAYA PLUS User manual


Image acquisition

3.2 Panorama mode Set position of patient


Panorama takes panorama images of upper/lower jaw.

1. Set detector on the exposure position.

Please refer 1.12 about how to install the detector.

2. [ For regular teeth ] Set chin-rest & bite on


the device.

The patient who doesn’t have teeth or hard to bite


the bite should install & use the chin-rest for
[ For regular teeth ] [For patient without teeth]
patient without teeth.

In case of using a chin-rest [For regular teeth],


cover it with clean cover. Remove clean cover used
and don’t use again.

Refer 1.5 about how to install the chin-


rest.

PAPAYA PLUS User manual 47


Image acqusition

Please let the patient remove eye glasses, hearing aid, false/artificial teeth, hair pin, other
accessories and metal etc. before taking X-ray images. If not, the shadow or reflected images
can be seen. Please remove the cloths wearing around the neck such as scarf, muffler, necktie
for the safety.

3. Operate OP by clicking panorama exposure button after registering a patient according to 3.1

4. When you click “Confirm” button after setting panorama mode & exposure condition on operating
OP, “Confirm” button turns “Ready” button, and then patient position arrange window comes out.

You can’t change the X-ray condition after clicking ‘Confirm’ button. X-ray condition & exposure
change is possible on start screen by ‘Back’ button.

48 PAPAYA PLUS User manual


Image acquisition

5. After being the patient positioned, set height on patient’s height by using 1.9 device set button.

6. Cover bite with clean cover, let the patient put their chin on
chin-rest and bite the bite.

If patient can’t raise their neck, set the height of device by using device set button.

7. When the position of patient is correct, lead patient


to grab a handle.

8. Set the position of patient by pressing laser button ( ) of device set button while the patient
bites the bite correctly.

Laser is consists of 3, you should let the patient be positioned for all 3 lasers correctly to get a good quality of image

9. When the laser is turned on, set midsagittal laser to midsagittal


of patient by moving patient’s head.

PAPAYA PLUS User manual 49


Image acqusition

10. Set Frankfort laser to Frankfort of patient by controlling angle of


patient’s head.

If the position of laser is not correct, set the height of laser by


moving Frankfort laser control knob.

Frankfort plane is the straight line connecting upper part of an earhole & lower part of eyes.

11. Finally move the position of laser by using ‘FORWARD’ &


‘BACKWARD’ button on patient control button to be canine
laser positioned on dogtooth of patient.

Laser is turned off automatically after few minutes or if the exposure starts.

12. Check whether the position of 3 lasers is correct after


positioning the patient by using Temple support. If it is
not correct, be the patient positioned correctly.

50 PAPAYA PLUS User manual


Image acquisition

3.3 TMJ mode patient position set


TMJ mode takes the image of lower teeth on TMJ part.

1. Set detector on panorama position.

Refer 1.12 on how to install detector

2. the chin-rest [for TMJ mode]

[ For TMJ mode ]

Please let the patient remove eye glasses, hearing aid, false/artificial teeth, hair pin, other
accessories and metal etc. before taking X-ray images. If not, the shadow or reflected images
can be seen. Please remove the cloths wearing around the neck such as scarf, muffler, neck tie
for the safety.

PAPAYA PLUS User manual 51


Image acqusition

3. Operate OP by clicking panorama exposure button after registering patient according to 3.1

4.Set TMJ mode & exposure condition on operating OP and click “Confirm” on right-low side, then
“Confirm” button turns to “Ready” button, patient position control screen comes out.

You can’t change the X-ray condition after clicking ‘Confirm’ button. X-ray condition & exposure
change is possible on start screen by ‘Back’ button.

52 PAPAYA PLUS User manual


Image acquisition

5. Set the height by using 1.9 device control button after being the patient positioned.

6.Let the patient lean on the chin-rest like the


image left.

Let the patient be positioned according to open/close the mouth.

If patient can’t raise their neck, set the height of device by using device set button.

7.When the position of patient is correct, lead the


patient grab the handle.

8.Set the position of patient by pressing laser button ( ) of device set button while the patient
stands on correctly.

Laser is consists of 3, you should let the patient be positioned for all 3 lasers correctly to get a good quality of image

9.When the laser is turned on, set midsagittal laser to midsagittal


of patient by moving patient’s head.

PAPAYA PLUS User manual 53


Image acqusition

10. Set Frankfort laser to Frankfort of patient by controlling angle of


patient’s head.

If the position of laser is not correct, set the height of laser by


moving Frankfort laser control knob.

Frankfort plane is the straight line connecting upper part of an earhole & lower part of eyes.

In the case of the exposure program is set to PA mode, move the


Frankfort laser to the TMJ PA laser marked position and adjust patient’s
head angle to position the TMJ is on to the laser line.

11. Finally move the position of laser by using ‘FORWARD’ &


‘BACKWARD’ button on patient control button to be canine
laser positioned on TMJ of patient.

‘FORWARD’ button moves to user’s direction but ‘BACKWARD’


button moves to opposite direction of user’s.

Laser is turned off automatically after few minutes or if the exposure starts.

54 PAPAYA PLUS User manual


Image acquisition

12. Check whether the position of 3 lasers is correct after


positioning the patient by using Temple support. If it is
not correct, be the patient positioned correctly.

PAPAYA PLUS User manual 55


Image acqusition

3.4 Sinus mode patient position set


TMJ mode takes the image of lower teeth on TMJ part.

1. Set detector on panorama position.

Refer 1.12 on how to install detector

2. Install the chin-rest for [Sinus mode] on device.

[ for Sinus mode ]

Please let the patient remove eye glasses, hearing aid, false/artificial teeth, hair pin, other
accessories and metal etc. before taking X-ray images. If not, the shadow or reflected images
can be seen. Please remove the cloths wearing around the neck such as scarf, muffler, neck tie
for the safety.

56 PAPAYA PLUS User manual


Image acquisition

3. Operate OP by clicking panorama exposure button after registering patient according to 3.1

4. Set Sinus mode & exposure condition on operating OP and click “Confirm” on right-low side,
then “Confirm” button turns to “Ready” button, patient position control screen comes out.

You can’t change the X-ray condition after clicking ‘Confirm’ button. X-ray condition & exposure
change is possible on start screen by ‘Back’ button.

PAPAYA PLUS User manual 57


Image acqusition

5. Set the height by using 1.9 device control button after being the patient positioned.

6. Let the patient lean on the chin-rest like


the image left.

Let the patient be positioned according to open/close the mouth.

If patient can’t raise their neck, set the height of device by using device set button.

7. When the position of patient is correct, lead


the patient grab the handle.

8. Set the position of patient by pressing laser button ( ) of device set button while the patient stands
on correctly.

Laser is consists of 3, you should let the patient be positioned for all 3 lasers correctly to get a good quality of image

58 PAPAYA PLUS User manual


Image acquisition

9. When the laser is turned on, set midsagittal laser to midsagittal


of patient by moving patient’s head.

10. Set Frankfort laser to Frankfort of patient by controlling angle of


patient’s head.

If the position of laser is not correct, set the height of laser by


moving Frankfort laser control knob.

Frankfort plane is the straight line connecting upper part of an earhole & lower part of eyes.

11. Finally move the position of laser by using ‘FORWARD’ &


‘BACKWARD’ button on patient control button to be canine
laser positioned on Sinus of patient.

‘FORWARD’ button moves to user’s direction but ‘BACKWARD’


button moves to opposite direction of user’s.

Laser is turned off automatically after few minutes or if the exposure starts.

PAPAYA PLUS User manual 59


Image acqusition

1. Check whether the position of 3 lasers is correct after


positioning the patient by using Temple support. If it is
not correct, be the patient positioned correctly.

60 PAPAYA PLUS User manual


Image acquisition

3.5 Cephalo mode patient position set

1. Set detector on cephalo position.

Refer 1.12 on how to install detector

Please let the patient remove eye glasses, hearing aid, false/artificial teeth, hair pin, other
accessories and metal etc. before taking X-ray images. If not, the shadow or reflected images
can be seen. Please remove the cloths wearing around the neck such as scarf, muffler, and
necktie for the safety.

2. Operate OP by clicking cephalo exposure button after registering patient according to 3.1

3. Set mode & exposure condition on operating OP and click “Confirm” on right-low side, then
“Confirm” button turns to “Ready” button, patient position control screen comes out.

PAPAYA PLUS User manual 61


Image acqusition

You can’t change the X-ray condition after clicking ‘Confirm’ button. X-ray condition & exposure
change is possible on start screen by ‘Back’ button.

4. Set the height by using 1.10 device control button after being the patient positioned.

5. Let the patient positioned according to exposure program after being Ear rod positioned to ear hole
of the patient.

6. Set the position of patient by pressing laser button ( ) of device set button while the patient
bites the bite correctly.

7. Laser enters horizontally. Control the angle of laser & patient’s Frankfort according to exposure program.

Frankfort plane is the straight line connecting upper part of an earhole & lower part of eyes.

62 PAPAYA PLUS User manual


Image acquisition

3.6 Image taking

After finishing the fixing the position of patient, press ‘Ready’


button of right-low part of OP after informing the patient not to
move. After the message for taking a X-ray image from OP, the
exposure starts when pressing X-ray shot switch.

Users should exposure from the position safe from X-ray.

Please keep pressing the switch during exposure. If the users take their hands off from shot
switch, the exposure is stopped and then, the error message comes out about the abnormal
shut down.

Focus on patient & condition of device during exposure. If the emergency situation is occurred,
take the hands off right away and then press emergency switch

PAPAYA PLUS User manual 63


Image acqusition

When the exposure starts, OP shows X-ray warning message & the image taking, and final image comes out
again after the complete.

64 PAPAYA PLUS User manual


Image acquisition

3.7 Image Control

After image exposure finishes and on screen, image is displayed and before push "save" button and
can do image control

Drag mouse's left button Move image within window (Up/down/left/right)

Drag mouse's right button Control Brightness

Drag mouse's both left and right View enlarge or minimize

Double click mouse's left button Re-size image with size of window

Double click mouse's right button initialize image to before control brightness

View if images are many and previous or next i


Move mouse wheel to up or down mage

PAPAYA PLUS User manual 65


Image acqusition

3.8 TWAIN INTERFACE


If user's program supports Twain interface, and by using TWAIN data source in which is named,
"PAPAYA" (TWAIN Version 2.1) and can send image to user's program. Sending methods are
supported with Native Mode Transfer and Buffered Memoray Mode Transfer.
The image acquisition by using PAPAYA data source is following.

1. Connect TWAIN session by choosing "PAPAYA" among TWIN data source whithin user's program
2. At user's program, the command of Acquire and send this command, and progress on OP's
full screen mode.
3. From contents of 3.2 to 3.6, depending upon exposure mode and choose exposure mode
and control patient position, and then try exposure of X-ray.
4. After exposure steps are finished normally and click "save" button and send user's program
called, "PAPAYA."
5. After sending is completed, and OP screen is closed automatically and ends with TWAIN session.

TWAIN data source can send image according to usage of Multi Focus that enables to send
more than 1 page. If on user's program and try to receive 1 page of image, and un-check the
check mark on content of 2.9.1's at general option tab and un-check All Save's checkbox.

3.9 OP Additional Function

OP enables burning CD, sending to DICOM server, registering


additional command, opening of patient date folder. Burning CD
and sending of DICOM server function is able for patient who
completed taking exposure. On order list window, clicking
mouse's right button and by choosing patient and pop-up menu
appears for choosing additional function.

66 PAPAYA PLUS User manual


Image acquisition

3.9.1 Sending to DICOM Server

At pop-up menu, by choosing send, and this function allows to pre-set to send image to DICOM
server. The DICOM server information is written on 2.9.3 and on DICOM option tab's lower left side
and set on Storage Server. By pushing "Insert Item" button, and input information window appears
and input AETitle, IP, Port, and Timeout and push "Insert" button. At Storage Server's Preset list, and
choose server in which will be sent and finish job by pushing "Save" button.

3.9.2 Burning CD
After taking exposured image and save them by using CD-ROM like media storage device or hard disk.
On pop-up menu, choose "Save to CD" and progress CD burning window.

For media storage device, choosing right side at device list and save image. Checking by Total Space
Used and Free Space and click "Burning" button. After burning is completed and finish and click
"Close" button.
When saving at hard disk or USB, by pushing button and choose save path and process burning.
With image, image viewer of CDSee is also saved.
Media device is connected, but device list doesn't appear and can't choose device, and push "Refresh"
button for updating the device list
On same path inside of CDSee folder, progress CDSee.exe and view saved image.

PAPAYA PLUS User manual 67


Image acqusition

3.9.3 Register addition order

Additional order function is that on previous ordered list, and patient need additonal exposure and
doesn't input basic personal information. Then choose exposure program and do registration for order.
If you choose additional order, and have chosen patient's basic information is updated and progress
with registration ordered window. By choosing exposure program, and click "Register" button, and the
order is added on ordered list.

3.9.4 Opening of patient's folder

When choosing opening of folder, search window is opened by starting from where the images are in
saved folder.

68 PAPAYA PLUS User manual


Image acquisition

4. Use condition
Please avoid these conditions for the operation/storing of this device.

 Exposure to humidity
 Exposure to
 Exposure to dust
 Exposure to high humidity
 Exposure to condition where has problem with ventilation
 Exposure to salty air.
 Exposure to chemical / harmful gas.

Please avoid use the product the place where has strong vibration or over heat or humidity.

Condition of movement & Storing


 Temperature : -20°C ~ 60°C
 Relative humidity: 5% ~ 95% RH (But, No dew condensation)
 Air pressure : 500 ~ 1100 hPa

Condition of movement when using


 Temperature: 15°C ~ 30°C
 Relative humidity : 5% ~ 75% RH (But, No dew condensation.)

PAPAYA PLUS User manual 69


Message

5. Cleaning

You can clean Bite piece, patient support table, chin rest & patient handle by high pressure sterilization up to
135℃ or cotton/gauze with alcohol.

You can patient handle with cotton/gauze with alcohol.

Please clean softly OP panel or other parts with wet soft cloth.

Clean the device after turning off the power.

Don’t spray liquid type of cleaning product directly to the device.

6. Handling of waste

It is regular custom to follow the law of each country when handling of waste including devices. Please
follow the regulations of each country.

70 PAPAYA PLUS User manual


Message

7. Message

7.1 Error message


Please refer the solutions below when the messages like below come out on OP. If, below methods do not solve problems a

Error message kV Low Occurred


1
Temporary Please turn off machine and turn on.
Solution
Error message kV High Occurred
2
Temporary Please turn off machine and turn on.
Solution
Error message mA Low Occurred
3
Temporary Please turn off machine and turn on.
Solution
Error message mA High Occurred
4
Temporary Please turn off machine and turn on.
Solution
Error message OCP
5
Temporary Please turn off machine and turn on.
Solution
Error message EPIP
6
Temporary Please turn off machine and turn on.
Solution
Error message Tank Overheat (SW)
7
Temporary Turning off machine for one hour and use again
Solution
Error message Tank Overheat (sensor)
8
Temporary Turning off machine for one hour and use again
Solution
Error message Exposure button was released
9
Temporary If user press off button by mistake, and after 10 seconds, error message disappear and
Solution
processing again
Error message Interrupt by something to rotate gantry
10
Temporary If user touched rotating part while rotation, after 10 seconds, and error message
Solution
disappear and use again
Error message Please, recheck laser alignment
11
Temporary If Canine Laser is located at back of ear, and after 10 seconds and error message
Solution
disappeared and use again.

PAPAYA PLUS User manual 71


Message

Please contact to manufacturer when the error messages like below.

No. Error message

1 Gantry rotating overload

2 Y-axis moving overload

3 Ceph detector moving overload

4 Ceph collimator moving overload

5 Collimator moving overload

6 No program at FPGA

7 Exposure program was wrong.

72 PAPAYA PLUS User manual


Message

7.2 Condition message

No. Message 설명

1 Scanning.. Taking a image during rotating

Making a quality of image better.


2 Processing
Reorganizing the image for diagnosis.

3 Testing Testing the rotating part.

4 Make by Ready Heating a X-ray tube.

5 Press the switch Exposure X-ray pressing the switch.

6 Returning... Returning the rotating part to starting point.

7 Emergency Pressing emergency button.

8 Technical Mode Setting OP system from PC.

PAPAYA PLUS User manual 73


Technical Data

8. Technical Data

8.1 Classification

Electrical classification Class I, Type B

8.2 Power Requirements

Input Power Supply 120 V~

Power Supply Frequency 60 Hz

Capacity of Power Supply 2.2 kVA

You are recommended to use AC power to avoid crash of power requirements for other
devices.

8.3 Complied Standards


- IEC/EN 60601-1 - IEC/EN 60601-1-2
- IEC/EN 60601-1-3 - IEC/EN 60601-2-7
- IEC/EN 60601-2-28 - IEC/EN 60601-2-32

74User manual
Technical Data

8.4 X-ray source assembly

kV range 60~90 kV

mA range 4~12 mA

PANORAMA: max 17sec


Exposure time range
CEPH: max 17sec

Inherent Filtration : 0.5 mmAl (OPX-


105) Inherent Filtration : 0.8 mmAl (D-
Total filtration 054SB) Added Filtration : 2mmAl
Total filtration : 2.5mmAl (OPX-105)
2.8mmAl (D-054SB)

Superheated Temperature 50 ℃

8.5 Beam Limiting Device

Material Aluminum + 3.0 mm Pb

Form Rectangle

PAPAYA PLUS User manual 75


Technical Data

8.6 Detector

 XID-C15DP (Only for Panorama)

Part Items Specifications


Sensor Material CdTe, 750 um
Pixel Size (um) 100
Sensor Chip Active Area (mm) 150 mm x 4.8 mm
Dynamic Range (dB) ≥ 72dB
DQE(70kV,0lp,mm) ≥ 0.8
MTF(2lp,mm) ≥ 0.8
Power Supply Requirement Digital +5 V/1 A, Analog +5 V/1 A
Command I/O 3 TTL trigger
Electronics
Data I/O GigE
Frame Rate (fps) 200 fps
Data Bits Per Pixel 12 bit
Mechanical Size (mm) 73W*168H*21.05D
Dimension
Weight (g) 340

 XID-C24DC (For Panorama and Cephalo)

Part Items Specifications


Sensor Material CdTe, 750 um
Pixel Size (um) 100
Sensor Chip Active Area (mm) 240 mm x 4.8 mm
Dynamic Range (dB) ≥ 72dB
DQE(70kV,0lp,mm) ≥ 0.8
MTF(2lp,mm) ≥ 0.8
Power Supply Requirement Digital +5 V/1 A, Analog +5 V/1 A
Command I/O 3 TTL trigger
Electronics
Data I/O GigE
Frame Rate (fps) Pano: 200 fps, Ceph: 125 fps
Data Bits Per Pixel 12 bit
Mechanical Size (mm) 73W*255H*21.7D
Dimension
Weight (g) 505 g

76 PAPAYA PLUS User manual


Technical Data

8.7 Dimensions and Weights


 PAPAYA

치 수 1040 x 1155 x 2317 mm

중 량 110 kg ± 5%

 PAPAYA Plus

치 수 1040 x 1775 x 2317 mm

중 량 140 kg ± 5%

PAPAYA PLUS User manual 77


Technical Data

8.8 Movement

Movement (Distance) : 700 mm


Up/Down Movement
Speed : 16.7 mm/s

Movement (Distance) : 140 mm


Right/left Movement
Speed : 6.8 mm/s

Rotating Angle 220 °

8.9 Other Information

Maximum No. of Exposure (Per an hour) 10 times

X-ray Exposure method Continuous

Panorama : 480 mm
SID
Cephalo: 1650 mm

Panorama : 248 mm
SSD
Cephalo : 1359 mm

Maximum tube current at maximum tube 12 mA


voltage
Maximum tube voltage at maximum tube 90 kV
current
Tube current and voltage at maximum 90 kV, 12 mA
power output
Minimum mAs 36 mAs

Over current Release Fuse 15A

Ground connection B type Ground

78 PAPAYA PLUS User manual


Technical Data

8.10 X-ray Tube


7.10.1 OPX – 105 (CEI)

Type Stationary

Focal Spot 0.5 mm

Target Angle 5°

Target Material Tungsten

Inherent Filtration 0.5 mmAl

Cooling Method Oil cooling

Nominal Anode Input Energy at 0.1S(DC) 2000 W

Nominal Voltage 110 kV

Thermal Characteristics
Anode Heat Storage Capacity 30 kJ

Maximum Anode Cooling rate 200 W

8.0 V
Filament characteristics (Max.)
4.0 A

 X-Ray tube characteristic curve

(1) Maximum Rating Charts (Absolute maximum rating charts)

PAPAYA PLUS User manual 79


Technical Data

(2) Emission Characteristics

(3) Anode Thermal Characteristics

(4) OPX/105 size (Unit:mm)

80 PAPAYA PLUS User manual


Technical Data

7.10.2 D-054SB (Toshiba)

Type Stationary

Focal Spot 0.5 mm

Target Angle 5°

Target Material Tungsten

Inherent Filtration 0.8 mmAl

Cooling Method Oil cooling

Nominal Anode Input Energy at 0.1S(DC) 1750 W

Nominal Voltage 110 kV

Thermal Characteristics
Anode Heat Storage Capacity 35 kJ

Maximum Anode Cooling rate 250 W

4.2 V
Filament characteristics (Max.)
3.5 A

 X-Ray tube characteristic curve

(1) Maximum Rating Charts (Absolute maximum rating charts)

PAPAYA PLUS User manual 81


Technical Data

(2) Emission Characteristics (3) Anode Thermal Characteristics

(4) D-054SB size (Unit: mm)

82 PAPAYA PLUS User manual


Labeling

9. Labeling
9.1 PAPAYA

• Digital Sensor

• X-ray tube
• Beam limiting device
• Generator
• Rotating Arm
• Laser
• Hand S/W

• Manufacturer Label
• X-ray Support

PAPAYA PLUS User manual 83


Labeling

9.2 PAPAYA Plus

• Digital Sensor

• X-ray tube
• Beam limiting device
• Generator
• Detector
• Rotating Arm
• Laser

• Hand S/W

• Manufacturer Label
• X-ray Support

84 PAPAYA PLUS User manual


Labeling

[Manufacturer label]

- PAPAYA

- PAPAYA Plus

PAPAYA PLUS User manual 85


Labeling

86 PAPAYA PLUS User manual


MAINTENANCE

Appendix 1.

MAINTENANCE

The user needs to check that all safe related the equipment operation before using it.

Routine maintenance
- Machine and parts should be maintained by basic condition.
- Machine and parts should be checked by routine maintenance.

Regular maintenance
- Check list every 1 year
Accuracy of X-ray tube voltage
Accuracy of X-ray tube current
Exposure dose of patient

Maintenance for long break


If the machine is not used for 6 months, the machine should be inspected by manufacturer.
- Check list
Accuracy of X-ray tube voltage
Accuracy of X-ray tube current
Exposure dose of patient

It must be used by a trained professional. This machine is maintained by GENORAY America.

Test method
- Accuracy of X-ray tube voltage
Test instrument: X-ray multi meter
Test method: Measure the tube voltage value and compare setting voltage with measured voltage.
Test limit: The error of the values of the X-ray tube voltage shall be not greater than 10%

- Accuracy of X-ray tube current


Test instrument: Oscilloscope
Test method: Measure the tube current value and compare setting current with measured current.
Test limit: The error of the values of the X-ray tube current shall be not greater than 20%

User manual87
Maintenance

- Exposure dose of patient


Test instrument: X-Ray multi-meter
Test method: Measured dosimetry.
Test limit: less than 7 mGy (Panorama), 3 mGy (skull AP/PA), 1.5 mGy (skull lateral)

If maintenance parameter is over the limit, some part is changed or repaired in manufacturer’s
judgment.

Please refer to the “PAPAYA Plus constancy test manual“ about detector calibration.

INSPECTION RESULT
INSPECTION ITEM INSPECTOR
PERIOD 1 2 3 4 5 Remarks

On / Off Switch

Up/Down, Right/Left buttons


EVERY DAY
Exposure Hand Switch
User

Patient Registration

Image Save and Load

Temperature & Humidity


EVERY WEEK
Emergency Button

Tube Voltage Accuracy Test


EVERY 1YEARS
Tube Current Accuracy Test

Exposure Dose to the patient Manufacturer

WHEN THE
Tube Voltage Accuracy Test
MACHINE HAS
NOT BEING
USED FOR 6 Tube Current Accuracy Test
MONTH
Exposure Dose to the patient

88 PAPAYA PLUS User manual


Appendix 2.

Dose information

A. Standard Panoramic mode

Recommended X-ray exposure DAP


*Patient type
kVp mA Exposure Time(sec) (mGy*cm²)

Big 72 11 12 67.05
Male 68 9 12 48.38
Female 65 7 12 33.60
Child 62 5 12 17.52

B. Orthogonal Panoramic mode

Recommended X-ray exposure DAP


*Patient type
kVp mA Exposure Time(sec) (mGy*cm²)

Big 72 11 12 67.05
Male 68 9 12 48.38
Female 65 7 12 33.60
Child 62 5 12 17.52

C. Bitewing Panoramic mode

Recommended X-ray exposure DAP


*Patient type
kVp mA Exposure Time(sec) (mGy*cm²)

Big 72 11 8.5 47.70


Male 68 9 8.5 34.42
Female 65 7 8.5 23.90
Child 62 5 8.5 17.52

*Patient type: See the last page in this section.


Lateral TMJ

Recommended X-ray exposure DAP


*Patient type
kVp mA Exposure Time(sec) (mGy*cm²)

Big 72 11 8 44.70
Male 68 9 8 32.25
Female 65 7 8 22.40
Child 62 5 8 11.68

D. PA TMJ mode

Recommended X-ray exposure DAP


*Patient type
kVp mA Exposure Time(sec) (mGy*cm²)

Big 75 11 8 48.40
Male 72 9 8 36.57
Female 68 7 8 25.08
Child 65 6 8 15.93

E. Lateral Sinus Midsagittal, Lateral Sinus mode

Recommended X-ray exposure DAP


*Patient type
kVp mA Exposure Time(sec) (mGy*cm²)

Big 72 11 5 27.94
Male 68 9 5 20.16
Female 65 7 5 14.00
Child 62 5 5 7.30

F. PA Sinus

Recommended X-ray exposure DAP


*Patient type
kVp mA Exposure Time(sec) (mGy*cm²)

Big 75 11 9 54.45
Male 72 9 9 41.14
Female 68 7 9 28.22
Child 65 6 9 17.92

*Patient type: See the last page in this section.

90 PAPAYA PLUS User manual


G. Lateral, PA, AP Cephalometric / Water’s View / SMV (Submento vertex) mode

Recommended X-ray exposure DAP


*Patient type
kVp mA Exposure Time(sec) (mGy*cm²)

Big 85 10 8 8.67
Male 83 10 8 8.23
Female 81 9 8 7.02
Child 80 9 8 5.66

H. Carpus mode

Recommended X-ray exposure DAP


*Patient type
kVp mA Exposure Time(sec) (mGy*cm²)

Big 70 7 8 3.96
Male 67 6 8 3.04
Female 63 5 8 2.18
Child 60 4 8 1.27

* See the table below for the Patient type.

Age Group PAPAYA Plus’s Standard


Child ≤ 12 -
Male -
Adult ≥ 13
Female -
Big ≥ 600mm, head
circumference

- A child is defined as a person who is younger than 12 years old.


- If the patient’s head is bigger than 600mm circumference, you can select the Big setting for exposure.
GENORAY Co., Ltd.
#512 Byucksan Technopia 434-6 Sangdaewon 1-Dong Jungwon-Gu Seongnam-City Gyeonggi-
Do Korea, 462-716
TEL : +82-31-737-8020 FAX : +82-31-737-8018
www.genoray.com

GENORAY America Inc.


3002 Dow Avenue, Suite 420, Tustin, CA 92780 USA
TEL : +1-714-289-8020 , FAX: +1-714-786-8919

Obelis s.a.
Boulevard Général Wahis 53, 1030 Brussels,
BELGIUM Tel: +32-2-7325954 Fax: +32-2-
7326003

The original version of PAPAYA is being written in Korean

92 PAPAYA PLUS User manual


PAPAYA plus (PAPAYA)

Installation Manual

Copyright© 2012

2013 / 05 / 28
Document Ver. 0.2

GENORAY Co., Ltd.


#812 Byucksan Technopia 434-6 Sangdaewon 1-Dong
Jungwon-Gu, Seongnam-City, Gyeonggi-Do, 462-716 KOREA
Tel: +82-31-737-8020 Fax: +82-31-737-8025
www.genoray.com

-1- Installation Manual


Table of Contents

Table of Contents

1. INTRODUCTION AND SAFETY............................................................................................... 3


1.1 General Introduction........................................................................................................ 3
1.2 Unit Symbols................................................................................................................... 4
1.3 LAYOUT..................................................................................................................... 5
1.4 Reference For Installation............................................................................................... 7
* These tools above are not components of product.2. INSTALLATION.......................................7
2. INSTALLATION......................................................................................................................... 8
2.1 Unpacking.................................................................................................................. 8
2.2 Level Adjustment........................................................................................................... 25
2.3 Connecting PC to Device / HAND SWITCH..................................................................27
2.4 PAPAYA APP Set up.................................................................................................................. 28
3. Test After Installation.............................................................................................................. 32
3.1 Input power................................................................................................................. 32
3.2 Check Rotating Unit....................................................................................................... 33
3.3 Check Patient Positioning Laser Aimer.........................................................................33
3.4 LAN card setting............................................................................................................ 33

-2- Installation Manual


1. INTRODUCTION AND SAFETY

1.1 General Introduction

This manual provides guidelines for the safe use and operation of the system.
The user should be trained in the use of equipment of this type and should ref
er to this manual in conjunction with that training.

Potential hazards exist in the use of medical electronic devices and X-ray syste
ms.
All users must understand the safety and emergency procedures, and the operat
ing instructions given in this manual.

The user must know how to recognize hazardous and potentially hazardous con
ditions, and know how to adequately protect themselves and others from possibl
e injury.

The manufacturer, assembler, and importer are responsible for the safety, reliability
and performance of the unit only if:
- Installation, calibration, modification and repairs are carried out by qualified
authorized personnel.

GENORAY pursues a policy of continual product development. Although every


effort is made to produce up-to-date product documentation this manual should
not be regarded as an infallible guide to current specifications. We reserve the
right to make changes without prior notice.

-3- Installation Manual


1.2 Unit Symbols

SYMBOLS Description

X-Ray Source Assembly,


X-ray Exposure

Attention!
consult accompanying documents

Type B applied part

Electrical Shock Hazard

WEEE Mark

Laser

The sign indicates information that needs attention to use the device.

The sign of warning gives information about the situations or


environments which can damage the product, or cause injury to users
or patients.

-4- Installation Manual


1.3 LAYOUT

<PAPAYA>

-5- Installation Manual


<PAPAYA plus>

-6- Installation Manual


1.4 Reference For Installation

Required Tools

+ SCREW DRIVER - SCREW DRIVER

L-WRENCH SET (Meter) 13mm Spanner

* These tools above are not components of product.

Installation Manual -7-


2. INSTALLATION
2.1 Unpacking

PANORAMA BOX

CEPHALO BOX

<PAPAYA plus (PAPAYA)> Initial Delivered Goods

-8- Installation Manual


- Styrofoam Information

Display Image Content


Disassemble
Styrofoam
Priority Number 1
- Disassemble
Styrofoam
Priority Number 2

Display where heavy


weight is put

Distinguish
packaging's up and
down

- Styrofoam information (Inside)

SENSOR BOX

ACCESSORY BOX

COLUMN BOTTOM COVER A/B

BASE SUB PLATE

ANCHOR JIG PLATE (OPTION)

Installation Manual -9-


- Sensor Box

- Accessory Box

1
6

7
3 4 5

- Accessory Box Part list


No Items Specification Figure QTY Confirmed

PAPAYA CD 1

TRIANA CD 1
2 CD
CALLIBRATION
1
DATA CD

1 SILICON CAP WHITE 7

1 BASE CAP WHITE 4

BASE ASSEMBLY M8x25L


4
WRENCH BOLTS spring washer
1 1-M8X20L
COLUMN ASSEMBLY
3-M8x25L 4
WRENCH BOLTS
spring washer

- 10 - Installation Manual
CEPHALO M8x25L
ASSEMBLY WRENCH spring washer 4
BOLTS and flat washers
CHINREST HANDLE
M6x20L
ASSEMBLY WRENCH 4
spring washer
BOLTS
COVER ASSEMBLY
M4x8L 6
SEM’S BOLTS
SHOT SWITCH
M4x12L
HOLDER ASSEMBLY 2
Washer head
SELF DRILL SCREW

2 CHINREST HANDLE 1

RIGHT AND
2 TEMPLE SUPPORT 1SET
LEFT

2 BITE 2

2 BITE COVER 2

3 CHINREST SUPPORT 1

4 CHINCUP SUPPORT 1

TMJ-SINUS
5 1
SUPPORT

6 SHOT SWITCH 1

7 LAN CABLE 1

Installation Manual - 11 -
1) PANORAMA UNIT disassemble




Lying down number 2 styrofoam and take off number 1 styrofoam. (Outside Styrofoam number is displayed)
Pulling out sensor box, base, and anchor base (option) while lying the machine on floor.
Put machine on upright position and use handle jig to tilt. Then disassemble the Styrofoam.

- 12 - Installation Manual
2) Moving to installation place

Using moving cart move with cart

Completion to move to X-ray room

Installation Manual - 13 -
3) Base Sub Plate assembly

4-M8x25L, SW
assembly

4-Rubber Cap assembly

- 14 - Installation Manual
4) Handling Jig disassemble

8-M8 BOLT disassembly

Installation Manual - 15 -
5) Stand Column

Displayed harness should not get caught on power sw bracket.

M8x20L,
3-M8x 25L, SW
assembly

- 16 - Installation Manual
6) Column Bottom Cover A/B assembly

6-M4x8L BOLT assembly

Installation Manual - 17 -
7) Hinge & Driving Top Cover disassembly

4-M8 BOLT disassembly

4-M4 BOLT disassembly

- 18 - Installation Manual
8) Safety Pin & Cephalo Align Bolt disassemble

2-Safety Pin disassemble

Cephalo Assembling Surface

M8 NUT disassemble

Installation Manual - 19 -
9) Cephalo Unit assembly

Assembling machine while placing cephalo unit without unpacking from packing fo

Making an even height with machine and push unit.

Guide pin is inserted and 4-M8x25L, SW, FW pre- installed


By using M8 siz-sided bolt and using 13mm Spanner, and tighten to basic surface
Tightening pre-installed bolt

- 20 - Installation Manual
10) Cephalo Cable assembly

Main Harness BKT disassemble

Coming from Cephalo and organize cable toward red arrow side
Assemble Main Harness BKT
Identifying cable's label and connect main B/D to connector (Included LAN cable)

Installation Manual - 21 -
11) Cephalo Box disassemble

- 22 - Installation Manual
12) Driving Top Cover assembly

4-M4 BOLT assembly

Installation Manual - 23 -
13) Patient Handle & Accessory assembly

4-M6x20L, SW
assembly

Chinrest, Temple support, Bite assembly

- 24 - Installation Manual
2.2 Level Adjustment
(1) Make the rotating part and supporting part at a line(180°) to each other and lay a dig
ital level on the datum plane. And keep the part horizontal with adjusting 6EA of screws.
(2) Make the rotating part and supporting part at right angles(90°) to each other and lay
a digital level on the datum plane. And keep the part horizontal with adjusting 6EA of s
crews.
Required tools : Digital Level, 5mm T Wrench (Or 5mm L Wrench)

Datum Plane

Installation Manual - 25 -
5mm T-Wrench

4-Level Adjustor Screw

- 26 - Installation Manual
2.3 Connecting PC to Device / HAND SWITCH

NO. Equipment PC
1 Equipment _ LAN PC_LAN

LAN port at the body


LAN port at the PC
- Location of board in the PC can be changed.

2.3.1 HAND S/W Assembly


Connect HAND S/W to LAN CONNECTOR in the picture below.

Installation Manual - 27 -
2.4 PAPAYA APP Set up.

1. Processing 'setup.exe' file

2. After processing 'Setup.exe' file, and displays like below picture


For beginning installation, click 'Next' button.

- 28 - Installation Manual
3. Choosing installation option and click 'Next' button.
If MS-SQL DB is not needed for installation and do not choose option

4. Click on the 'Install' button to install PAPAYA OP.

Installation Manual - 29 -
5. The installation process will proceed as shown in the figure below

6. After checking 'Install MS-SQL DB' and install MS-SQL DB

- 30 - Installation Manual
7. When the installation is complete, click 'FINISH' button to exit.

Installation Manual - 31 -
3. Test After Installation
Test operating condition of each part.

3.1 Input power


Turn switch ON. – On/Off switch is at rear of device.

OFF ON

After turning on, run TRIANA and choose any mode.


Run OP program.

- 32 - Installation Manual
3.2 Check Rotating Unit

(1) Press button to test after CONTROL screen came out. Check if Rotating Unit is
moving right way.

(2) If Rotating Unit is not moving, turn it manually a little and press button again.

(3) Repeat above procedure 2-3 times to confirm Rotating Unit is fine.

When the Rotating Unit does not stop and crashes into main body, turn off power

switch ( ) and adjust starting point according to Technical Manual.

3.3 Check Patient Positioning Laser Aimer

(1) Press Laser Aimer button on OP Panel and three Laser Aimers show up.

(2) Press Laser Aimer button again, three Laser Aimer disappear.

Laser Aimer can be different depend on purpose.


If intersecting point of lasers is not the vertical, adjust Laser Aimer according to Technical
Manual.

3.4 LAN card setting.


(1) At control panel (control), entering Network and Sharing Center (Network and Sharing
Center).
(2) Look on the right side of the menu, click "Change adapter setting" icon (Change adapter
setting).
(3) The following screen appears.
Changing Intel (R) PRO/1000 GT or Inter (R) Gigabit CT adapter's name to PAPAYA.

Installation Manual - 33 -
(4) After entering PAPAYA adapter properties (Properties), and the following screen will appear..

(5) Settings by pressing the button configuration (configure. ..).

- 34 - Installation Manual
- Setting flow control (Flow Control) to Disable mode (Disable).
- Setting IPv4 Checksum Offload (IPv4 Checksum Offload) to all use
- setting Jumbo Frame (Jumbo Packet) to 9014 Byte or 9KB MTU
- Setting large send offload to use
- Setting receiving buffer (Receive Buffers) to 2048
- Setting TCP check sum off load (TCP checksum Offload) to all use..

(6) Binding setting.

Pushing F10 key if the menu is not seen as above picture.


Entering in advance -> advance setting

Installation Manual - 35 -
Locating PAPAYA's adapter as above picture

- 36 - Installation Manual
GENORAY Co., Ltd.
#512 Byucksan Technopia 434-6 Sangdaewon 1-Dong Jungwon-Gu
Seongnam-City Gyeonggi-Do Korea, 462-716
TEL : +82-31-737-8020 FAX : +82-31-737-8025
www.genoray.com

GENORAY America Inc.


3002 Dow Avenue, Suite 420, Tustin, CA 92780 USA
TEL : +1-714-289-8020 , FAX: +1-714-786-8919

Obelis s.a.
Boulevard Général Wahis 53, 1030 Brussels,
BELGIUM Tel: +32-2-7325954 Fax: +32-2-7326003

Installation Manual - 37 -
PAPAYA Plus
(PAPAYA)

Technical Manual

Copyright© 2013

2013/ 07 /15
Document Ver. 0.1

GENORAY Co., Ltd.


#812 Byucksan Technopia 434-6 Sangdaewon 1-Dong
Jungwon-Gu, Seongnam-City, Gyeonggi-Do, 462-716 KOREA
Tel: +82-31-737-8020 Fax: +82-31-737-8025
www.genoray.com
Table of Content
1. Introduction and safety related article........................................................4
1.1 Introduction........................................................................................................................4

1.2 Cleaning and environment for using...................................................................................5

1.3 Symbols used in product.....................................................................................................6

2. Composition of PAPAYA................................................................................7

3. FUNCTION CHECK.......................................................................................................8
3.1 Rotating part shift and reversion TEST..............................................................................8

4. DIAGRAM.....................................................................................................................11
4.1 TOTAL BLOCK DIAGRAM...........................................................................................11

4.2 CT PWR B/D....................................................................................................................12

4.3 LIFT_CONTROL B/D.....................................................................................................13

4.4 MAIN CONTROL B/D................................................................................................14

4.5 COLIMATOR CONTROL B/D....................................................................................15

4.6 CEPH CONTROL B/D................................................................................................16

4.7 SENSOR connecting B/D.............................................................................................17

4.7 DRIVER B/D....................................................................................................................18

4.8 DRIVER SUB B/D...........................................................................................................20

4.9 CHINREST B/D...........................................................................................................22

4.10 SECOND COLIMATOR B/D.........................................................................................22

4.11 LED B/D.........................................................................................................................23

5. Replacement PARTS..........................................................................................24
5.1 Chinrest Assy....................................................................................................................24

5.2 Main Column Assy...........................................................................................................25

5.3 Driving Assy.....................................................................................................................27

5.4 Gantry Assy......................................................................................................................28


5.5 Cephalo Assy....................................................................................................................30

5.6 Rear Cover........................................................................................................................32

6. Calibration method........................................................................................33

7. TROUBLE SHOOTING, Error message Solutions & modification.............34


7.1 Trouble Shooting Flow.....................................................................................................34

7.2 How to replace FUSE B/D................................................................................................35

7.3 Error message...................................................................................................................36

7.4 Condition/Process message...............................................................................................41

7.5 X-RAY feedback data check.............................................................................................42


1. Introduction and safety related article
1.1 Introduction

This manual is designed for your safe use and operation of PAPAYA & PAPAYA PLUS, use
this manual to guide you. It must be used by a trained professional.

The potential risk always exists when using X-ray products and electric medical
equipments. All users should be aware of the methods of use solutions in emergency
conditions and other safety articles provided in this manual.

The user must acknowledge the appropriate ways to protect oneself from the potential
danger that could occur at any time, and be aware of the condition of the product that
can cause danger.

The manufacturer, assembler, and the importer hold responsibility upon the performance of safety, trust and the
- Guarantees the installation of the eligible professional, calibration, and repair

GENORAY Co., Ltd is continuously developing the products. The manual or composition of the package might ch
1.2 Cleaning and environment for using

Bite piece, (patient support table, chin rest and temple support can be cleaned by alcohol
soaked cotton. The other surface including the control panel of the product must be cleaned
gently by lightly wet, soft cotton. For the operation and installation of this product, please
avoid the categories listed below.

 Exposure to moisture
 Exposure to direct ray of light
 Exposure to the space with dust
 Exposure in high dampness
 The environment which can cause problem on ventilation of product
 Exposure in atmosphere including sodium
 Exposure to chemicals harmful gas

During the use of the product, if there is a strong vibration at normal condition or in the
places that exceed the range of dampness or temperature below do not put the product
nearby.

- Environment for movement or


Storage Temperature : -20°C ~
60°C
Relative humidity : 5% ~ 95% RH (water drop due to humidity should not
occur) Air pressure : 500 ~ 1100 hPa

- Environment for using


Temperature : 15°C ~
30°C
Relative humidity : 5% ~ 75% RH (water drop due to humidity should not occur)
1.3 Symbols used in product

SYMBOLS Description

X-Ray Source
Assembly, X-ray
조사

Caution!
Refer to the user’s manual

Type B applied part

Electric shock danger

WEEE Mark

Laser

This symbol is information about the special matters that require attention while using this software.

This symbol is warning the situation or environments which might bring danger to the patients or users.
2. Composition of PAPAYA

14
3
9 13
4

10
5
7
7
8
11
2

12

[ PAPAYA ] [ PAPAYA Plus ]

No. description No. description

1 BASE 7 Patient supporting area

2 Fixed column 8 Equipment’s control button

3 Driving Unit 9 Cephalo Arm

4 Rotating Unit 10 On/off switch

Hand Switch (Exposure button, Emergency stop


5 X-ray source assembly 11
switch)

12 Emergency stop switch


Detector (Panorama, Cephalo using
6 13 Device status LED
for both, detachable structure)

14 Equipment’s 2nd control button


3. FUNCTION CHECK
3.1 Rotating part shift and reversion TEST

Likewise on picture below, the equipment’s downside switch is on, and the light is on at
LED and display waiting condition of the machine.
.

12

OFF ON
At machine, the power is inputted, and proceed TRIANA program at PC, and push exposure
button and connect to the machine. OP program is on and start checking the machine.

Check of the machine is completed, and below screen appears on OP.

< OP Screen >


OP screen appears and by checking of machine’s normal operation, push the machine’s

laser button ( ). To check the machine’s Rotating Unit is working normally.

If Rotating Unit is not rotating, and by hands, turning Rotating Unit, and push laser button

again( ).

Repeat the procedure above more than 2 times for check-ups.

When reversing back to its starting point, if the Rotating Unit does not stop
and contacts the main body, the use should press on the off button
immediately and replace the original place using the technical manual.
4. DIAGRAM
4.1 TOTAL BLOCK DIAGRAM
4.2 CT PWR B/D
1) POWER ON: By receiving main power from NOISE FILTER, it gets power from TB1.
2) PSU B/D power supply:Main power is going through LF1 and is going through CN1,
and main power (AC220V) is out toward PSU B/D.
3) TANK DRV B/D DC Voltage Supply : From PSU B/D, and going through CN3, and
after receiving +24V, and main power (220V) changes to DC 310V and permit to give
to CN2.

4.3 LIFT_CONTROL B/D

1) By coming from CN406, and goes through +24V, and goes through
relay(K1), and goes through DC CONVERTER(PSD10-24-1212), and
produce +24V,+12V,-12V.
2) Through relay(K1), +24V is going through PANO_CN1,CEPH_CN1 and
permit to give voltage to each board.
3) By using MCU, connecting each communication and controls the lift
through CN405.
4.4 MAIN CONTROL B/D

Through FPGA, each motor and sensors permit to give signal.


1) Through MCU, RS232 and CAN communicate.
2) Through FPGA, LED SIGNAL is given a permission to give signal.
4.5 COLIMATOR CONTROL B/D

1) It is entered Signal through CN100 connector, and shows


LASER Signal received from Main B/D.(CN102,CN103) ,
2) Colimator motor’s Sensor signal is detected.(CN101)
3) Through MCU of Main B/D, Colimator motor control.(CN104)
4.6 CEPH CONTROL B/D

1) Ap Sensor, Pa Sensor, Ear Sensor, Carpus SW, Nasal SW’s


SIGNAL is detected.
2) Through CN207, Cep Sen motor & Cep Col motor control.
4.7 SENSOR connecting B/D

1) SESNOR SPRING B/D: Through receiving SIGNAL and by detachable SPRING


transmit signal to MASKING.
2) SENSOR MASKING B/D: Among received SIGNAL, +5V,X-RAY TRIGER,X-RAY LED,
through CN505, giving signal to SENSOR CONTROL B/D, and about ETHENET,
directly give a permission to sensor part.
3) SENSOR CONTROL B/D: By receiving signal from SENSOR MASKING B/D, and give
permission to give signal to I3 sensor part.
4.7 DRIVER B/D
1) OCP BOARD: In DRIVER B/D, consume voltage is that convert both end of R5’s
voltage path. Then, you compare to standard OCP_REF voltage and is higher than preset
voltage, and at OCP LED1’s LED is on. With VR1, it is able to set up the moment
maximum voltage.

2) They are composed with DRIVER way and FULL-BRIDGE way.

3) DRV_SUB BOARFD: With below timing chart, DIRVER B/D is operating.


4.8 DRIVER SUB B/D
1) EP PWM CONTROL: Between EP_REF and EP_FB’s voltage difference, the
frequency’s range need to be controlled, and send signal (output) to EP_GATE1,
EP_GATE2.

2) IP PWM CONTROL: Between IP_REF and IP_FB’s voltage difference, the frequency’s
range need to be controlled, and send signal (output) to IP_GATE1, IP_GATE2.

3) LEVER OVER: According to DRIVER B/D’s defect and TANK’s defect, FEEDBACK value
that is higher than one area or set-up value and send LEVER OVER signal to MAIN
B/D.

1) VR2: Control PRE-HEAT COURVE

0.5V 이하
5V

85KV 7mA 0.3SEC

PEAK DECREASE PEAK INCREASE


VR2
DRV_SUB B/D
FRONT VIEW
4.9 CHINREST B/D

1) It is entered SIGNAL of each SW through CN1 connector, and


shows LED SIGNAL received from MAIN B/D.

4.10 SECOND COLIMATOR B/D

1) CN1 connector conveys each SW SIGNAL to CEP


CONTROL B/D
2) CN1 connector shows Laser LED Signal which can show
its condition depending on SW SIGNAL conveyed.
4.11 LED B/D

1) Conveyed voltage & SIGNAL show LED display by


using photocoupler(SFH690ABT) and TR(MMBT2222A).
5. Replacement PARTS
5.1 Chinrest Assy.

5-1-1
5-1-3
5-1-2

5-1-4

5-1-6

5-1-5

PARTS LIST
NO. ITEM SERIAL NO. QNT REMARKS
5-1-1 BITE P03RP412 1
5-1-2 CHINREST P03RP409 1
5-1-3 TEMPLE SUPPORT A P03RP403 1
5-1-4 TEMPLE SUPPORT B P03RP404 1
5-1-5 CHINREST HANDLE P03RP408 1
5-1-6 CHINREST OP BOARD P03JB018 1
5.2 Main Column Assy.

5-2-1 5-2-2 5-2-3 5-2-4


5-2-6

PARTS LIST

No. Item Serial no. Qnt Remarks


5-2-1 SMPS SNP-B309 1
5-2-2 Lift Control Board P03JB023 1
5-2-3 Power board P03JB001 1
5-2-4 Noise filter MC1220 1
5-2-5 Lift TLG10-23700000 1
5.3 Driving Assy.

5-3-1

5-3-3

5-3-2

PARTS LIST
No. Item Serial no. Qnt Remarks
5-3-1 Main board P03JB011 1
5-3-2 Driving LED board P03JB019 1
5-3-3 Pulley assy P03DS001 1
5.4 Gantry Assy.

5-4-1

5-4-6

5-4-3
5-4-2

5-4-4

5-4-5
PARTS LIST

No. Item Serial no. Qnt Remarks


5-4-1 Tank Drv board P03JB002 1
5-4-2 Colimator board P03JB012 1
5-4-3 Tank Assy P03TS001 1
5-4-4 Collimator Assy P03MS001 1
5-4-5 Frankfort Laser Assy P03GS001 1
5-4-6 Sensor Assy P03SS001 1
5.5 Cephalo Assy.

5-5-1

5-5-5

5-5-4

5-5-2
5-5-3
5-5-6

5-5-7

PARTS LIST

No. Item Serial no. QTY Remarks


5-5-1 Sensor Assy P03SS001 1
5-5-2 Ear Post Assy P03ES002 2
5-5-3 Nasal Assy P03ES001 1
5-5-4 2nd Collimator Assy P03ES003 1
5-5-5 Cephalo Arm Assy P03ES004 1
5-5-6 Ceph Control Board P03JB013 1
5-5-6 Cephalo LED Board P03JB019 1
5.6 Rear Cover

5-6-4
5-6-5
5-6-1

5-6-2

5-6-3

PARTS LIST
No. Item Serial no. Qnt Remarks
5-6-1 Terminal Block Dft-20a-3p 1
5-6-2 LAN Connector JACK-3101 1
5-6-3 Spiral Cable Grand PECG001 1
5-6-4 Hanger’s D-sub Connector P03JH121 1
5-6-5 Waring system’s D-sub Connector P03JH122 1
6. Calibration method
- Refer ‘Machine Alignment & Image calibration’ document.
TROUBLE SHOOTING, Error message Solutions & modification

7. TROUBLE SHOOTING, Error message Solutions & modification


7.1 Trouble Shooting Flow

POWER ON

N
1) SYSYEM CHECK
Device ON

N 2) SYSYEM CHECK
IMAGE Normal

3) Detector CHECK

Device Normal

1) SYSTEM CHECK
① Is power of device turned on normally?
-Is LED1 turned on POWER SEQ
B/D? OFF : Check FUSE1
ON : Check +5V between U3 No. 2 & 5.
- Check FUSE B/D
Check F3,F4 FUSE
2) IMAGE CHECK
- In case of No image on the screen
Solution => Check whether X-RAY generates normally with KV METER.
=> Check the beam of X-RAY TANK by using a X-ray intensifying screen
=> Replace TANK ASS’Y & DRIVER B/D
=> Replace Detector

- In case of Dark image on the screen


Solution => Check whether X-RAY generates normally with KV METER.
=> Replace TANK ASS’Y & DRIVER B/D
=> Replace Detector

7.2 How to replace FUSE B/D

F1,F2 FUSE rated


Model name : 6.35X31
FUSE Model No. : 65TL
250V 15A MAKER :
ORISEL
7.3 Error message

Below table is the list of error messages.


These messages are showed on main screen of operating panel when the user uses the
product inappropriately or the functions are not operated normally with unknown reasons.
Error message kV Low Occurred!
• When it is beamed lower then kV value set.
Cause
• When the temperature of X-Ray tube low.
1
• When connector related X-Ray tube connects badly.
Temporary 1. Check whether cable sheath is peeled or cable is squashed
Solution 2. Preheat enough by using “Ready” function more than twice.
Permanent
1. Replace DRIVER B/D, MONO TANK & Harness.
Solution
Error message kV High Occurred!
• When it is beamed higher then kV value set.
Cause
2 • When connector related X-Ray tube connects badly.
Temporary 1. Check whether cable sheath is peeled or cable is squashed
Solution 2. Preheat enough by using “Ready” function more than twice.
Permanent
1. Replace DRIVER B/D, MONO TANK & Harness.
Solution
Error message mA Low Occurred!
• When it is beamed lower then mA value set.
Cause
• When the temperature of X-Ray tube low.
3
• When B/D inside connects badly.
Temporary 1. Check whether cable sheath is peeled or cable is squashed
Solution 2. Preheat enough by using “Ready” function more than twice.
Permanent
1. Replace DRIVER B/D, MONO TANK & Harness.
Solution
Error message mA High Occurred!
• When it is beamed higher then mA value set.
Cause
4 • When B/D inside connects badly.
Temporary 1. Check whether cable sheath is peeled or cable is squashed
Solution 2. Preheat enough by using “Ready” function more than
twice.
Permanent
1. Replace DRIVER B/D, MONO TANK & Harness.
Solution
Error message OCP (Over Current) Occurred!
• When the voltage currents over than the voltage set.
• When the temperature of X-Ray tube low.
Cause
• When B/D inside connects badly.
5
• When the shooting function is used too much.
• When it has malfunction cause of exterior static electricity or
NOISE.
Temporary 1. Check whether cable sheath is peeled or cable is squashed
Solution 2. Turn OFF the power and ON after 10 minutes later.
Permanent 1. Check whether cable sheath is peeled or cable is squashed
Solution 1. Replace DRIVER B/D, MONO TANK & Harness.
Error message EPIP Over Occurred!
• When kV & mA value current over than the value set.
• When the temperature of X-Ray tube low.
Cause
• When B/D inside connects badly.
6
• When the shooting function is used too much.
• When it has malfunction cause of exterior static electricity or
NOISE.
Temporary 1. Check whether cable sheath is peeled or cable is squashed
Solution 2. Turn OFF the power and ON after 10 minutes later.
Permanent
1. Replace DRIVER B/D, MONO TANK & Harness.
Solution
Error message Tank Outside overheat
• When the temperature switch is dropped cause of overheated
Cause
7 tank outside by shooting too much.
Temporary 1. Check whether cable sheath is peeled or cable is squashed
Solution 2. Turn OFF the power and ON after 10 minutes later.
Permanent
Replace harness / Replace sensor.
Solution
Error message Tank Inside overheat
• When the temperature switch is dropped cause of overheated
Cause
8
tank inside by shooting too much.
Temporary 1. Check whether cable sheath is peeled or cable is squashed
Solution 2. Turn OFF the power and ON after 10 minutes later.
Permanent 1. Replace DRIVER B/D, MONO TANK & Harness.
Solution
Error message Exposure button was released.
Cause • When exposure switch is released while it is exposure
9
Temporary
• Automatically finished after 10 seconds.
Solution
Permanent
Solution
Error message Gantry Rotating overload
• When there are human or objects while gantry is rotating.
Cause
Malfunction of sensor
• Away human or objects from the rotating space.
Temporary Check whether motor moves to the opposite direction.
10
Solution Check whether the sensor operates.
Check whether the sensor is put correctly.

Replace motor.
Permanent Replace motor harness.
Solution Replace encoder.
Replace encoder harness.
Replace sensor.

Error message Y-axis Moving Overload


• When there are interrupting objects while Y-axis motor is
Cause
moving.
Malfunction of sensor
11 Remove the elements which interrupt the moving from moving
Temporary space.
Solution Check whether the motor moves to the opposite side.
Check whether the sensor operates.
Check whether the sensor is put correctly.

Permanent Replace motor.


Solution Replace motor harness.
Replace sensor.
Error message Cephalo Sensor Moving Overload
12 • When there are interrupting objects while Detector is moving left
Cause & right
Malfunction of sensor
Remove the elements which interrupt the moving from moving
space.
Temporary
Check whether the motor moves to the opposite side.
Solution
Check whether the sensor operates.
Check whether the sensor is put correctly.

Permanent Replace motor.


Solution Replace motor harness.
Replace sensor.
Error message Cephalo Collimator Moving Overload
• When there are interrupting objects while Cephalo 2nd collimator
Cause
is moving.
Remove the elements which interrupt the moving from moving
13
Temporary space.
Solution Check whether the motor moves to the opposite side.
Check whether the sensor operates.
Check whether the sensor is put correctly.

Permanent Replace motor.


Solution Replace motor harness.
Replace sensor.
Error message Collimator Moving Overload
Cause • When there are interrupting objects while collimator is moving.
Remove the elements which interrupt the moving from moving
Temporary space.
14
Solution Check whether the motor moves to the opposite side.
Check whether the sensor operates.
Check whether the sensor is put correctly.

Permanent Replace motor.


Solution Replace motor harness.
Replace sensor.
Error message No program at FPGA
Cause • When there is no program at FPGA
15 Temporary
Update FPGA program
Solution
Permanent
Update FPGA program
Solution
Error message Pressed Exposure switch.
Cause • When the Exposure switch is pressed
16 Temporary
Release the pressure of exposure switch.
Solution
Permanent
Replace the exposure switch.
Solution
Error message Gantry is interrupted by something
Cause • When human or objects interrupt the rotation
22 Temporary
Press & Take off the emergency button
Solution
Permanent
Replace the encoder
Solution
Error message Rotate Exposure program was wrong.
• When pressing the ready button, tracing value over the moving
Cause
23 distance.
Temporary
Press & Take off the emergency button
Solution
Permanent
Set the starting point of rotating axis again.
Solution
Error message Please, recheck laser aligment
• When pressing the ready button, tracing value over the moving
Cause
distance.
24
Temporary Press & Take off the emergency button
Solution Check the position of Y-axis by turning on laser.
(It can be problem if the position of Y-axis is too much behind)
Permanent
Require A/S to Research team / CS team
Solution
Error message Cephalo Sensor Exposure program was wrong.
• When pressing the ready button, tracing value over the moving
Cause
25 distance.
Temporary
Press & Take off the emergency button
Solution
Permanent
Set the starting point of Cephalo sensor axis again.
Solution
26 Error message Cephalo Collimator Exposure program was wrong. .
• When pressing the ready button, tracing value over the moving
Cause
distance.
Temporary
Press & Take off the emergency button
Solution
Permanent
Set the starting point of Cephalo collimator axis again.
Solution

7.4 Condition/Process message


The below table is the list of Condition/Process message

No Main message Description of condition


1 • Scanning.. • Scanning the images by rotating.

2 • Processing • Improving the quality of the image

3 •Converting DICOM (3/3)... • Converting DICOM image

4 • Testing • Testing the rotating part.

5 • Ready... • Heating a X-ray tube

6 • Press exposure button • X-ray generates if pressing the exposure button.

7 • Returning • Rotate the rotating part to the starting point.

8 •Emergency • Emergency button has pressed.

9 •Technical mode • Setting OP system in PC.


7.5 X-RAY feedback data check

7.5.1 Peak tube potential (kVp) measurement


(1) Connect the probe of digital multimeter (or oscilloscope) measurement point on
the main board (positive probe – TP15, negative probe – TP16).
(2) Select the appropriate DC voltage measurement range for 1 to 4.5 V signal level.
(3) Take an exposure with desired kV setting.
(4) Reading the voltage value and record it.
(5) The actual tube voltage relates to the measured feedback signal an follows :
Actual tube voltage = 20kV * measurement feedback voltage ( in volts)
The resulting tube voltage should be within ±10% of the voltage indicated in the user
interface.

7.5.2 Tube current (mA) measurement


(1) Connect the probe of digital multimeter (or oscilloscope) measurement point on
the main board (positive probe – TP17, negative probe – TP16).
(2) Select the appropriate DC voltage measurement range for 1 to 6 V signal level.
(3) Take an exposure with desired mA setting.
(4) Reading the voltage value and record it.
(5) The actual tube voltage relates to the measured feedback signal an follows :
Actual tube current = 2mA * measurement feedback voltage ( in volts)
The resulting tube current should be within ±10% of the voltage indicated in the user
interface.
GENORAY Co., Ltd.
#512 Byucksan Technopia 434-6 Sangdaewon 1-Dong Jungwon-Gu
Seongnam-City Gyeonggi-Do Korea, 462-716
TEL : +82-31-737-8020 FAX : +82-31-737-8025
www.genoray.com

GENORAY America Inc.


3002 Dow Avenue, Suite 420, Tustin, CA 92780 USA
TEL : +1-714-289-8020 , FAX: +1-714-786-8919

Obelis s.a.
Boulevard Général Wahis 53, 1030 Brussels,
BELGIUM Tel: +32-2-7325954 Fax: +32-2-7326003
PAPAYA plus (PAPAYA)

Alignment Manual

Copyright© 2013

2013 / 06 / 07
Document Ver. 0.0

GENORAY Co., Ltd.


#812 Byucksan Technopia 434-6 Sangdaewon 1-Dong
Jungwon-Gu, Seongnam-City, Gyeonggi-Do, 462-716 KOREA
Tel: +82-31-737-8020 Fax: +82-31-737-8025
www.genoray.com

-1- Installation Manual


Table of Contents

1. INTRODUCTION.............................................................................................................................3
1.1. Regular Condition......................................................................................................................3
1.2. Preparations................................................................................................................................3
1.2.1. Required Tools...............................................................................................................3
1.2.2. Jig or phantom set..........................................................................................................4
2. Machine’s Alignment.........................................................................................................................6
2.1. When Machine Production........................................................................................................6
2.1.1. Execution of Alignment Program and Connection with Machine.................................6
2.1.2. Execution of Calibration Wizard (Panoramic mode).....................................................7
2.1.3. Panoramic Mode’s Machine Alignment and Calibration..............................................9
2.1.4. Execution of Calibration Wizard (Cephalometric mode)............................................27
2.1.5. Cephalometric Mode’s Machine Alignment and Calibration......................................29

-2- Installation Manual


1. INTRODUCTION
1.1. Regular Condition

This installation manual includes necessary contents of when production and


installation of machine.

1.2. Preparations

1.2.1. Required Tools


<Driver>

<Figure 1.2.1.1> + or – Driver

<Wrench set>

<Figure 1.2.1.2> Wrench set

-3- Installation Manual


1.2.2. Jig or phantom set
<Theta jig>

<Figure 1.2.2.1> Thata jig

<SD-line jig>

<Figure 1.2.2.2> SD line jig

-4- Installation Manual


<Ball phantom>

<Figure 1.2.2.3> Ball phantom

<Laser alignment jig>

<Figure 1.2.2.4> Laser alignment jig

-5- Installation Manual


2. Machine’s Alignment

2.1. When Machine Production

2.1.1. Execution of Alignment Program and Connection with Machine


“C:/GENORAY/PAPAYA”. Inside of the folder, find PAPAYA_REC.exe and execute
the file.

<Figure 2.1.1.1> Machine Alignment Program

According to <Figure 2.1.1.1>, after execution of the machine alignment program,


push ‘Connect’ button where it is located at middle upper side and connecting
program with machine (<Figure 2.1.1.2>). After connection with machine ‘Connect’
button changes to ‘Disconnect’. If it is not connected with machine, refer to appendix

-6- Installation Manual


A.

<Figure 2.1.1.2> Connect Button’s Location

2.1.2. Execution of Calibration Wizard (Panoramic mode)


After connecting with machine, refer to <Figure 2.1.2.1> and choose Panoramic tab.

<Figure 2.1.2.1> Panoramic tab’s Location

At Panoramic tab, on right down corner, click ‘Calibration Wizard’ button(<Figure


2.1.2.2>).

<Figure 2.1.2.2> Location of Calibration Wizard’s Button

After clicking button <Figure 2.1.2.3>; According to <Figure 2.1.2.3>, Calibration


Wizard is being executed.

-7- Installation Manual


1

3 4

<Figure 2.1.2.3> Calibration Wizard (Panoramic) Calibration

Wizard’s each part’s name and functions are as below.

Number Name Function


Using for changing steps for machine alignment and
1 Step tab
calibration.
Displaying necessary commands and functions for machine
2 Step window
alignment and calibration.
Machine PAPAYA machine’s current status is displayed.
3
Status Window
Displaying progress status of necessary functions of
4 Progress Bar
machine’s alignment and calibration.
<Table 2.1.2.1> Calibration Wizard’s Each Parts Name

-8- Installation Manual


2.1.3. Panoramic Mode’s Machine Alignment and Calibration
This section explains about what to do on each step for machine’s alignment and
calibration. User could change the steps by using step tab, item number 1 of <Figure
2.1.2.3>. When production of machine, progress with each steps to align and to
calibrate machine.

<Preparing>
Before machine alignment and modification, this explains necessary tools.

<Figure 2.1.3.1> Preparing Step(Panoramic)

Number Name Function


Using for machine’s alignment of rotating degree to
1 Theta jig block
starting point.
Using for machine’s center of rotation alignment and for
2 SD-line jig
alignment of ball phantom’s position.
3 Ball phantom Using with SD-line jig and use for chin-rest’s alignment.
Laser alignment Using as put above Ball phantom and use for laser
4
jig alignment.
<Table 2.1.3.1> Preparation Tools

-9- Installation Manual


On window’s below side, before doing alignment, explaining what needs to be
removed from machine.

Remove upper cover of driving.


On circled with red, remove bolts and remove
cover by pulling up.

Remove cover, following the steps. (The removal


method is as below contents)

On red circle, remove bolts on red circles and


remove cover on back side where X-ray tank.

After removing red circle and their bolts, and


remove front cover of where X-ray tank.

After removing of red circle and remove bolts and


on step by step removing cover.

- 10 - Installation Manual
<Preparing2>

<Figure 2.1.3.2> Preparing2 Step(Panoramic)

Before alignment of machine, on patient’s support part, and need to remove parts, in
which are machine’s cover and handle, and explains how to remove them.

Patient support’s cover and handle need to be


removed. (Removal method is as below)

After removing bolts on red circles and remove the


handle.

- 11 - Installation Manual
On red circle, remove indicated bolts and remove
cover on that parts.

On red circle, remove indicated bolts and remove


cover on that parts.

On red circle, remove indicated bolts and remove


below cover.

On red circle, remove indicated bolts and remove


upper cover.

- 12 - Installation Manual
<Step1>

<Figure 2.1.3.3> Step1 – Y-Axis adjusting of staring point(Panoramic)

After preparing all preparation, set up Y-Axis starting point. Machine’s OP and using
machine’s ‘FORWARD’, ‘BACKWARD’ button and move machine’s Y-Axis to front
and back to match of indicated driving part’s and aluminium block for Y-axis, refer to
<Figure 2.1.3.3>. In this condition, push ‘Set current position’ button and then current
Y-Axis location is displayed and saved.

- 13 - Installation Manual
<Step2>

<Figure 2.1.3.4> Step2 – Adjustment of center of rotation (Panoramic)

<Figure 2.1.3.5> Using ‘Theta jig block’ (Panoramic)

According to <Figure 2.1.3.5>, fix ‘Theta jig block’ to driving side and fix with bolts.
Then, installing ‘Theta jig block’ if because Y-Axis is coming front and rotating side
and ‘Theta jig block’ conflicts, and then using OP’s ‘BACKWARD’ button and move
Y-Axis to backward and make installation. On rotating part and ‘Theta jig block’
meets and parallel each other, and then push ‘Set current position’ button and save
current rotating side’s location by pushing button.

- 14 - Installation Manual
<Figure 2.1.3.6> Using of ‘Ball phantom’ and ‘SD-line jig’ (Panoramic)

According to <Figure 2.1.3.6>, ‘Ball phantom’ and ‘SD-line jig’ is installed on


machine. ‘SD-line jig’ is refer as below <Figure 2.1.3.7>, and install on X-ray tank
side by using bolts. When installing, dip side and SD-line jig’s fixed side need to be
meet perfectly and need to be installed with caution.

<Figure 2.1.3.7> ‘SD-line jig’s Installation Location

After completion of installation of ‘Ball phantom’ and ‘SD-line jig’ according to


<Figure 2.1.3.6>, unscrew 2 bolts in which are fixing chin-rest.

<Figure 2.1.3.8> Setting of ‘Ball phantom’ and ‘SD-line jig’s starting point

- 15 - Installation Manual
Adjusting as repeating below steps to align ‘Ball phantom’ and rotating axis’ location.

1. By pushing ‘0°’ button, making perpendicular between rotating part and driving
part.
2. ‘SD-line jig’s line and ‘Ball phantom’s reference line need to be matched by
turning ‘Ball phantom’ stage’s knob and move ‘Ball phantom’s location.
3. ‘180°’s button need to be pressed and rotating part move to 180°.
4. If, at 0° location and 180° location, ‘SD-line jig’ is not on same location, on 0°
adjusted ‘Ball phantom’s reference line and at 180° location. ‘Ball phantom’s
reference line and ‘SD-line jig’s line that needs to be located by turning each
knob and locating them to the middle of ‘SD-line jig’s locations of 0° and 180°.
5. By repeating 1-4’s to reduce the error.

After steps of 1~5, ‘SD-line jig’ is located on machine’s rotating axis. On this
condition, according to <Figure 2.1.3.9>, by using ‘0°’, ‘90°’, ‘180°’ button, at each
degree’s location, ‘SD-line jig’s line is at ‘Ball phantom’s matching with appropriate
reference line and adjust chin-rest’s location and tighten chin-rest’s bolt. Then, need to
be caution to not move ‘Ball phantom’s knob.

<Figure 2.1.3.9> Fixing ‘Ball phantom’s location

After chin-rest is tightly fixed, again at ‘0°’, ‘90°’, ‘180°’ location, and checking ‘Ball
phantom’ and ‘SD-line jig’ are matching. If location is not matching, again chin-rest’s
bolt need to be unscrew and adjusted.

- 16 - Installation Manual
<Step3>

<Figure 2.1.3.10> Step3 – Alignment of Collimator’s Position (Panoramic)

On Step3, before Collimator’s accurate alignment, this is step where finding estimated
position. On Step2, ‘Ball phantom’ is installed on PAPAYA and click ‘Find Collimator
Offset’ button. After clicking button, the machine is automatically finding for
Collimator’s position by proceeding necessary steps and displays ‘Adult Panoramic
collimator position’s value. In this condition, click ‘Set Offset & Get Image’ button
then X-ray is exposed and display an image. Then distinguishing whether collimator’s
is good position by checking with left illustrated picture. If Collimator’s position is
well found, according to picture’s Good position, on image, by Collimator, X-ray’s
exposure’s limited area is almost same with left area and right area.

- 17 - Installation Manual
<Step4>

<Figure 2.1.3.11> Step4 – Adjustment of detector position(Panoramic)

On Step4, this step where finding detector’s position so that the X-ray beam center and
the center of rotation are on a straight line. Repeat the following process.

1.Click ‘Rotate 0 ° & Get Image’ button to expose X-ray and to get an X-rayed
image.
2.Select line1 of ‘Select Line’ Combo box located on bottom of display window
and then click the center of the pin, imaged by X-ray exposure, then blue
dashed line is being on the center of the pin.
3.Click ‘Rotate 180 ° & Get Imagte’ button to expose X-ray and to get an X-rayed
image.
4.Select line2 of ‘Select Line’ Combo box located on bottom of display window
and then click the center of the pin, imaged by X-ray exposure, then yellow
dashed line is being on the center of the pin.
5.Red line is being on the middle of two lines between blue and yellow line. Check
whether the red line is located inside of two green line or not (Pass/Fail
information will be displayed). If fail, refer to <Figure 2.1.3.11>, then loosen
screws marked with red circle and move the detector position left or right and

- 18 - Installation Manual
repeat 1-5 until Pass is displayed.

In the condition Pass in on, refer to <Figure 2.1.3.12>, join the red circled screws and
put the detector cover on.

<Figure 2.1.3.12> Adjustment of detector position

- 19 - Installation Manual
<Step5>

<Figure 2.1.3.13> Step5 – Detailed Offset Angle adjustment(Panoramic)

On Step5, this step is where to adjust rotation origin angle which is set before by using
‘Theta jig block’ exactly. Repeat following procedure until Pass is being displayed on
the window.

1.Click ‘Set angle and Get image’ to expose X-ray and to get an X-rayed image of
degree 90°.
2.Select blue line of ‘Select Line’ Combo box located on bottom of display
window and then click the center of the pin, imaged by X-ray exposure, then
blue dashed line is being on the center of the pin
3.Select yellow line of ‘Select Line’ Combo box located on bottom of display
window and then click the center of the metal ball, imaged by X-ray exposure,
then yellow dashed line is being on the center of the pin
4.Make a judgment of Pass/Fail. If failed, refer to <Figure 2.1.3.13> , then adjust
‘Gantry rotation angle offset of 90°’ value slightly little. Repeat 1-4 until Pass
is displayed

- 20 - Installation Manual
<Step6>

<Figure 2.1.3.14> Step6 – Laser alignment(Panoramic)

Step 6 adjust the position of the laser of the equipment.


Install 'Laser Alignment jig' on the ‘Ball phantom’ and adjust laser position, refer to
<Figure 2.1.3.14>.

- 21 - Installation Manual
<Step7>

<Figure 2.1.3.15> Step7 – Collimator alignment(Panoramic)

On Step7, this step is where to adjust collimator position exactly. Repeat following
procedure until Pass is being displayed on the window.

1.Click ‘Find Collimator Offset’ button to find collimator position


2.Click ‘Set Offset & get Image’ button to expose X-ray and to get an X-rayed
image.
3.Make a decision whether collimator position and shape is right or not, refer to
the image on the left side of window. If the collimator is rotated or sided to one
side then refer to bottom-left image and adjust collimator position correctly,
refer to <Figure 2.1.3.15>.
4.Judge whether it is Pass or Fail. If failed, repeat 1-3 until Pass is displayed

- 22 - Installation Manual
<Figure 2.1.3.16> Adjustment of Collimator

- 23 - Installation Manual
<Step8>

<Figure 2.1.3.17> Step8 – Detector Calibration(Panoramic)

Step8, performing the detector calibration. 'Auto' button to perform calibration


automatically. Make sure that the Dark Ref. Level value is 0, Bright1 is 200, Bright2 is
800 and press 'Auto' button.

- 24 - Installation Manual
<Step9>

<Figure 2.1.3.18> Step9 – Collimator shading calibration(Panoramic)

Step9 is where to calibrate attenuation of X-ray due to collimator. 'Auto shading


calibration' button will proceed automatically.

<Step10>

<Figure 2.1.3.19> Step10 – Final test(Panoramic)

Step10 is a step to check whether the whole calibration and alignment process
performed rightly or not.
Install ‘Ball pahntom’ to PAPAYA and press ‘Confirm’ button and when the ‘Ready &

- 25 - Installation Manual
Shot’ button is been able then click the button to take panoramic of ‘Ball phatom’.
When it finish to reconstruct panoramic of ‘Ball phantom’, draw 4 lines by using
‘Select line’ combo box like as before steps, refer to the image on bottom side. If get
‘Pass’, then every calibration and alignment of panoramic is finished. Or if get ‘Fail’,
then refer to appendix to find solution and repeat step10 to check.

<Offset Overall>

<Figure 2.1.3.20>
Offset Overall – Live image and manual input of Offset parameter
(Panoramic)

Offset Overall, possible to input offset parameter manually and to take live X-ray
image.

- 26 - Installation Manual
2.1.4. Execution of Calibration Wizard (Cephalometric mode)
Connect to PAPAYA and choose Cephalometric tab, refer to <Figure 2.1.4.1>.

<Figure 2.1.4.1> Location of Cephalometric tab

Click 'Calibration Wizard' button at the bottom right in the Cephalometric tab (<Figure
2.1.4.2>)

<Figure 2.1.4.2> Location of calibration Wizard button

When you click the button, Calibration Wizard is run like <Figure 2.1.4.3>.

- 27 - Installation Manual
1

3 4

<Figure 2.1.4.3> Calibration Wizard (Cephalometric) Calibration

Wizard’s each part’s name and functions are as below.

No. Name Function


Using for changing steps for machine alignment and
1 Step tab
calibration.
Displaying necessary commands and functions for machine
2 Step window
alignment and calibration.
Machine Status PAPAYA machine’s current status is displayed.
3
Window
Displaying progress status of necessary functions of
4 Progress Bar
machine’s alignment and calibration.
<Table 2.1.4.1> Calibration Wizard’s Each Parts Name

- 28 - Installation Manual
2.1.5. Cephalometric Mode’s Machine Alignment and Calibration

This section explains about what to do on each step for machine’s alignment and
calibration. User could change the steps by using step tab, item number 1 of<Figure
2.1.4.3>. When production of machine, progress with each steps to align and to
calibrate machine.

<Preparing>
Explanation of preparations and tools.

.
<Figure 2.1.5.1> Preparing Step(Cephalometric)

No. Name Function


This is used to set cephalometric base angled and
1 5mm pin
reference position of 2nd collimator.
<Table 2.1.5.1> Preparations

The image on bottom side explains how to remove the covers ahead to calibration and
alignment.

- 29 - Installation Manual
Remove the top cover.
The position is marked with a red circle and remove the
screws to remove the cover and lift it up.

Remove the cover of the 2nd collimator. Remove the


screws on the red circle location Peel off the cover.

- 30 - Installation Manual
<Step1>

<Figure 2.1.5.2> Step1 – Setting of reference position of detector (Cephalometric)

Once all the preparations is ready, set the detector’s reference position. Use 5mm pin,
refer to <Figure 2.1.5.2>, to set detector position to the reference hole of cephalometric
base. When the detector is on the reference point, then remove the pin carefully. Click
‘Set current position’ button.

- 31 - Installation Manual
<Step2>

<Figure 2.1.5.3> Step2 – Alignment of 1st collimator’s position (Cephalometric)

Step2 is where to align 1st collimator position. Click ‘Find Collimator Offset’ button
then procedure will proceed automatically.
When automatic procedure is finished, then click ‘Set Offset & Get image’ button to
check 1st collimator is on proper position.

- 32 - Installation Manual
<Step3>

<Figure 2.1.5.4> Step3 – Alignment of Ear post and base(Cephalometric)

Step3 is a step to check whether X-ray and ear posts are on a straight line or not and to
align their to be on a straight line. The laser alignment also is performed in this step.
First, adjust the height of two ear posts so that they have same length, refer to <Figure
2.1.5.5>.

<Figure 2.1.5.5> Adjustment of Ear post’ height

- 33 - Installation Manual
<Figure 2.1.5.6> Adjustment of Cephalometric laser

Adjust laser position, refer to <Figure 2.1.5.6>.

Set ear posts and nasal positioner to Lateral direction and perform this below
procedure for alignment of base angle of cephalometric part.

1.Click ‘Confirm’ button


2.Click ‘Ready & Shot’ button when it is ready.
3.Select blue cross of ‘Select Cross’ combo box which is located under the image
window and click center of one of two ear posts.
4.Select yellow cross of ‘Select Cross’ combo box which is located under the
image window and click center of another one.
5.Determine it is Pass or Fail. If it failed , adjust base angle refer to <Figure
2.1.5.7> and repeat 1-5 until get Pass

- 34 - Installation Manual
<Figure 2.1.5.7> Adjustment of base angle and Pass/Fail determination

- 35 - Installation Manual
<Step4>

<Figure 2.1.5.8> Step4 – 2nd Collimator alignment(Cephalometric)

Step 4 is for aligning 2nd collimator position exaclty.


Repeat this below procedure until get ‘Pass’

1.Click ‘Find Collimator Offset’ button to find 2nd collimator position


automatically.
2.Click ‘Set Offset & Get Image’ button to get an X-ray exposed image.
3.Figure out that the collimator has a right shape and is on a proper position. If it
has not a right shape or be sided. Then adjust 2nd collimator position to right
position, refer to the bottom-left image of <Figure 2.1.5.8>
4.Determine Pass/Fail. If it failed, repeat 1 -3.

- 36 - Installation Manual
<Step5>

<Figure 2.1.5.9> Step5 – Detector Calibration(Cephalometric)

Step5, performing the detector calibration. 'Auto' button to perform calibration


automatically. Make sure that the Dark Ref. Level value is 0, Bright1 is 200, Bright2 is
800 and press 'Auto' button.

- 37 - Installation Manual
<Step6>

<Figure 2.1.5.10> Step6 – Scan calibration/FFC Calibration (Cephalometric)

Step 6 performs FFC(Flat Field Correction) and Scan calibration to calibration 2nd
collimator position of scanning and to equalize brightness of cephalometric image.
First, set ear posts and nasal positioner to AP direction and click ‘Auto scan/FFC
calibration’ button, then procedure will proceed automatically.

- 38 - Installation Manual
<Step7>

<Figure 2.1.5.11> Step7 – Final image test(Cephalometric)

Step 7 is a step to test that whole steps for cephalometric calibration is performed well
until now.
First, set ear posts and nasal positioned to Lateral direction and click ‘Confirm’ button.
When it is ready then click ‘Ready & Shot’ button to take cephalometric image.
Determine, when the reconstruction is finished, whether there is a problem or not, refer
to right image of <Figure 2.1.5.11>.
If you find there is a problem on the image. Then refer to Appendix to find solution for
each problem case. And calibrate and align PAPAYA again by following the instruction.
And then do step7 again.

- 39 - Installation Manual
<Offset Overall>

<Figure 2.1.5.12>
Offset Overall – Live image and manual input of Offset parameter
(Cephalometric)

Offset Overall, possible to input offset parameter manually and to take live X-ray
image.

- 40 - Installation Manual
GENORAY Co., Ltd.
#512 Byucksan Technopia 434-6 Sangdaewon 1-Dong Jungwon-Gu
Seongnam-City Gyeonggi-Do Korea, 462-716
TEL : +82-31-737-8020 FAX : +82-31-737-8025
www.genoray.com

GENORAY America Inc.


3002 Dow Avenue, Suite 420, Tustin, CA 92780 USA
TEL : +1-714-289-8020 , FAX: +1-714-786-8919

Obelis s.a.
Boulevard Général Wahis 53, 1030 Brussels,
BELGIUM Tel: +32-2-7325954 Fax: +32-2-7326003

- 41 - Installation Manual
- !GENORAY
w ww.g e no ra y .c o m
Ont he baseof Experience and Technology, Genoray Co.,