Machine: S/No
Initial Date
Scaling
Setup blade calibration plate on couch top @ 100cm SSD
Level gantry to 90degree so KV detector is at floor
Place KV source at 100,0,0 & detector at 50,0,0
In OBI maintenance take an image using single pulse full resolution
Measure distance between 10X10 pair of lines in X & Y direction
Technique
KVP 60 MA 16.0 Ms 8.0
Imager Alignment
Using image from above zoom out until the centre circle on the plate is visible
Select area histogram icon and select show details
Change Hor Pos to 1024 & Vert Pos to 768
Drag upper left hand corner of the ROI box until it intersects the centre of the circle
Record the Hor Pos & Vert Pos of the centre pixel position
Center pixel position Pass/fail
Expected Measured Diff. (<10pixels)
Horizontal post 1024
Vertical post 768