You are on page 1of 2

To,

Indian Army
Kolkata

Ref: GeM Bid Number: GEM/2023/B/3728891, Date: 26-08-2023


Sub: Clarification Letter

Dear Sir,

This is with reference demo held on 31.10.2023 for referred GeM bid, please find below line by line reply against the
query asked by you.

S.No Bid Specification


3 Each machine should be supplied with latest generation sevoflurane, We confirm that we will provide
g(iii) isoflurane vaporizers. They should be temperature, pressure and flow latest generation sevoflurane,
compensated and maintenance free isoflurane vaporizers. They shall be
temperature, pressure and flow
compensated and maintenance free
h(i) Pipeline input range: 300kPa to 600kPa as per installed hospital We confirm that Quoted model
system. Pipeline input range: 300kPa to
600kPa as per installed hospital
system.
h(ii) (ii) Pipeline connection: As per installed Hospital Pipeline system We confirm that Quoted Model have
for oxygen, nitrous oxide, and air. Pipeline connection: As per installed
Hospital Pipeline system for oxygen,
nitrous oxide, and air
H(iii) (iii) All fittings available for oxygen, Nitrous Oxide and Air which We confirm that All fittings available
contains pipeline filter and check valve to be certified by the for oxygen, Nitrous Oxide and Air
manufacture. which contains pipeline filter and
check valve to be certified by the
manufacture
4(f) (f) All materials in contact with anaesthesia gases & vapouriser We confirm that All materials in
should be free of natural rubber latex contact with anaesthesia gases &
vapouriser should be free of natural
rubber latex
15(e) Monitor should be capable of being upgraded at least to interface We confirm that Monitor shall be
third party devices, hospital network-PACS, DICOM imaging etc capable of being upgraded at least
to interface third party devices,
hospital network-PACS, DICOM
imaging etc
17 Monitor should have basic and advanced arrhythmia detection and ST We confirm that Monitor shall have
analysis with trends. It should have a filter for protection against basic and advanced arrhythmia
defibrillator and electro surgery potentials. detection and ST analysis with
trends. It shall have a filter for
protection against defibrillator and
electro surgery potentials
21(a) Electrical safety IEC-6060/IS-13450 Certification to be provided by the IEC Certificates Enclosed
OEM/Vendor
21(b) Power cord length should be at least 5 meters We confirm tha Power cord length
ii shall be up to 5 meters
21(c) (c) All materials in contact with the patient breathing gases We confirm that All materials in
should be free of natural rubber latex. contact with the patient breathing
gases should be free of natural
rubber latex
System Operational
Temperature: 10-30 degrees Celsius/50-95 degrees Fahrenheit (AC Temperature Operation:10 ~ 40˚C
Workstation) Storage: -20~55˚C
Humidity: 15 to 95 % relative humidity non-condensing as per IEC 68- Relative HumidityOperation:80% ,
2-3 non-condensing
Storage: ≤93%,non-condensing
The Machine Monitor and Ventilator should be European CE with These Specifications we not
proof of marketability in Europe and US FDA Approved mentioned in Bid.

Quoted model Anesthesia machine


is BIS Certified.
Monitor EU CE Approved by 4 digit
notified body number 0297.
Certificate enclosed
23(m) m) The firm must provide one high quality double stage oxygen We confirm that we will provide
regulator for connection to jumbo cylinders and connectors to the high quality double stage oxygen
piped oxygen supply where central medical gas pipeline exists, as per regulator for connection to jumbo
user requirements. cylinders and connectors to the
piped oxygen supply where central
medical gas pipeline exists, as per
user requirements.

For Allied Medical Limited

Sarla Verma
Asst. Manager Tendering