You are on page 1of 4
Stat Fax: 3300 Chemistry Analyzer Flip-up display screen _ Welcome to a new generation oa Stat Fax’ AWARENESS TECHNOLOGY, INC. See the curve, then calculate kinetic results based on the entire read time OR the segment [reece you choose. Large 240 x128 pixel graphic display is easy to view and use Stat Fax® 3300 Stat Fax® 3300 stores calibration stores control values and standard curves with many QC options to assure linear kinetics and reliable results. For each run, cone choose a new 21.8 curve or use a 1.0 options to modify multi-point curves and revert to previous curve. stored curve with 2.5 + Create Levey-Jennings plots for QC Stat Fax® 3300 stores and reports patient data Create a work list. Output data to: - On-board 40-column Run the tests. = Thermal printer - External printer Print Patient Reports. -PC and LIS DEST U RUC Mutu MCR) AWARENESS TECHNOLOGY, INC. Werth aie enti rue mac ec and other laboratory instruments. Bichromatic optical system with 6 filters and spaces for 2 optional additional filters (340-700nm). Adapts for use with 12mm round tubes, 1cm square cuvettes, or built-in flow cell. Optional external keyboard for | easy entry of patient names and data. Vacuum system speeds sampling and reduces carryover. Read cell temperature controlled to 37°C. Peltier cell cooling available as an option. Calibration traceable to NIST standard. Optional incubation block available with 18 wells, temperature controlled to 37°C. Use Redi-Check® Photometer QA, the easy way to monitor accuracy, ‘a linearity, and repeatability. Se ee eo as http://www.awaretech.com Stat Fax’ 3300 Technical Specifications Photometer Tube or cuvette. 0 to 2.5 absorbance units. With flowcell...... 0 to 3.5 absorbance units. Stability, ++ seeee Drift of no more than 0.005A in 8 hours/bichromatic. Light source Tungsten halogen lamp, with lamp saver feature: Standard wavelengths ...........340, 405, 505, 545, 580, 630nm plus 2 optional Optional filters, tee eeee es. 840- 700nm, Filter types ses seeeesee Ultra long life multiple cavity IAD hard coat. Sampling Sample Volume - Flowcell ........ Programmable 250L-750yL. Fiowcell Body................ Surgical grade stainless steel Windows. [boc Be Pyrex. Illuminated Volume... « 2A. Minimum Volume - Tube (12mm)... tml. Minimum Volume - Cuvette ....... 400uL. Electronic Processor . . .ceseeeseessesee+++Z180, Program Memory 1 megabit. Non-volatile Memory vee eee T megabit RAM... cosa ea 32K byte. Intemal printer... . «40/80 column thermal Display....... 0.5 +++. Graphical LCD 240x128, Keyboard 20 key, domed membrane. PC keyboard. .... oe Compatible with standard PS2 keyboard. Output - R232... - 9600 8 data, 1 stop, no parity. Output - Parallel Any IBM printer with simple ASCII compatibility. Operating Modes Single and multipoint standardization Rate (factor/standard) . . . . live graphie display of reaction. Clive tice ames. . linear, point to point, or regression. Miscellaneous. blank and curve recall, sample blanking. Special Features Password Protection for test results and parameters. Levey-Jennings QC tracking, patient reports, job creation. User tests More than 100. Stored patient results More than 500. ‘Stored control results More than 500. Optional external heat block 18 wells, controlled temp @ 37°C. Optional read cell cooling Peltier cooling system - cell only. Cell Temperature selections....... 25°C, 30°C, or 37°C. Enclosure... fire retardant ABS and steel. Dimensions. 16" (40cm) x 14.5" (37 om) x 5.5" (14 cm)* *12"- 30 om tall with screen up. Weight (17 Ibs) 6.4 kg. Power . 100-240 VAC (auto-sensing) 50/60 HZ input, 90 watts. A aaa PO Drawer 1679, Palm City, FL 34991 Eos pemieuies veneer)

You might also like