Stat Fax: 3300
Chemistry Analyzer
Flip-up display
screen
_ Welcome to a new
generation oa
Stat Fax’
AWARENESS TECHNOLOGY, INC.See the curve,
then calculate kinetic
results based on the
entire read time
OR
the segment
[reece you choose.
Large 240 x128 pixel graphic display is easy to view and use
Stat Fax® 3300 Stat Fax® 3300
stores calibration stores control values
and standard curves with many QC options to assure linear
kinetics and reliable results.
For each run, cone
choose a new 21.8
curve or use a 1.0
options to modify
multi-point curves
and revert to
previous curve.
stored curve with 2.5 +
Create Levey-Jennings plots for QC
Stat Fax® 3300 stores and reports patient data
Create a work list. Output data to:
- On-board 40-column
Run the tests. = Thermal printer
- External printer
Print Patient Reports. -PC and LIS
DEST U RUC Mutu MCR)
AWARENESS TECHNOLOGY, INC.
Werth aie enti rue mac ec
and other laboratory instruments.Bichromatic optical system with
6 filters and spaces for 2 optional
additional filters (340-700nm).
Adapts for use with 12mm round
tubes, 1cm square cuvettes, or
built-in flow cell.
Optional external keyboard for
| easy entry of patient names
and data.
Vacuum system speeds sampling
and reduces carryover.
Read cell temperature controlled
to 37°C.
Peltier cell cooling available as
an option.
Calibration traceable to NIST
standard.
Optional incubation block
available with 18 wells,
temperature controlled to 37°C.
Use Redi-Check® Photometer QA,
the easy way to monitor accuracy,
‘a linearity, and repeatability.
Se ee
eo as
http://www.awaretech.comStat Fax’ 3300 Technical Specifications
Photometer
Tube or cuvette. 0 to 2.5 absorbance units.
With flowcell...... 0 to 3.5 absorbance units.
Stability, ++ seeee Drift of no more than 0.005A in 8 hours/bichromatic.
Light source Tungsten halogen lamp, with lamp saver feature:
Standard wavelengths ...........340, 405, 505, 545, 580, 630nm plus 2 optional
Optional filters, tee eeee es. 840- 700nm,
Filter types ses seeeesee Ultra long life multiple cavity IAD hard coat.
Sampling
Sample Volume - Flowcell ........ Programmable 250L-750yL.
Fiowcell Body................ Surgical grade stainless steel
Windows. [boc Be Pyrex.
Illuminated Volume... « 2A.
Minimum Volume - Tube (12mm)... tml.
Minimum Volume - Cuvette ....... 400uL.
Electronic
Processor . . .ceseeeseessesee+++Z180,
Program Memory 1 megabit.
Non-volatile Memory vee eee T megabit
RAM... cosa ea 32K byte.
Intemal printer... . «40/80 column thermal
Display....... 0.5 +++. Graphical LCD 240x128,
Keyboard 20 key, domed membrane.
PC keyboard. .... oe Compatible with standard PS2 keyboard.
Output - R232... - 9600 8 data, 1 stop, no parity.
Output - Parallel Any IBM printer with simple ASCII compatibility.
Operating Modes
Single and multipoint standardization
Rate (factor/standard) . . . . live graphie display of reaction.
Clive tice ames. . linear, point to point, or regression.
Miscellaneous. blank and curve recall, sample blanking.
Special Features
Password Protection for test results and parameters.
Levey-Jennings QC tracking, patient reports, job creation.
User tests More than 100.
Stored patient results More than 500.
‘Stored control results More than 500.
Optional external heat block 18 wells, controlled temp @ 37°C.
Optional read cell cooling Peltier cooling system - cell only.
Cell Temperature selections....... 25°C, 30°C, or 37°C.
Enclosure... fire retardant ABS and steel.
Dimensions. 16" (40cm) x 14.5" (37 om) x 5.5" (14 cm)*
*12"- 30 om tall with screen up.
Weight (17 Ibs) 6.4 kg.
Power . 100-240 VAC (auto-sensing) 50/60 HZ input, 90 watts.
A aaa PO Drawer 1679, Palm City, FL 34991