A P opt.

liquicolor
Enzymatic color test for the
quantitative determination of alkaline phosphatase

AU 400
NOTE:
RECOMMENDATION TO APPLY THE REAGENTS ON THE ANALYZER.
THE LABORATORY HAS TO VALIDATE THE SETTINGS IN ITS OWN RESPONSIBILITY.
HUMAN WILL ACCEPT NO RESPONSIBILITY FOR THE CORRECTNESS OF THIS APPLICATION AS WE HAVE
NO CONTROL ON SOFT- AND HARDWARE MODIFICATIONS OF THE RESPECTIVE ANALYZER.

Package Size
[REF] 12217 16 x 5 ml ALKALINE PHOSPHATASE opt. liquicolor Reagent kit
12017 10 x 10 ml ALKALINE PHOSPHATASE opt. liquicolor Reagent kit
12027 8 x 50 ml ALKALINE PHOSPHATASE opt. liquicolor Reagent kit
12037 4 x 250 ml ALKALINE PHOSPHATASE opt. liquicolor Reagent kit
13160 4 x 5 ml AUTOCAL
Recommended control serum
13511 6 x 5 ml HumaTrol N
13512 6 x 5 ml HumaTrol P
13951 6 x 5 ml SERODOS
13151 6 x 5 ml SERODOSplus
[IVD]
R1 = [BUF], ready for use
R2 = [SUB], ready for use
¼ For detailed information kindly refer to the most recent package insert.

Calibration
Calibration frequency is user defined. Re-calibration is recommended:
• following quality control procedures
• after reagent bottle change
• major preventive maintenance was performed on the analyser or a critical part was replaced.
Control frequency
It is recommended that 2 levels of control be assayed daily.
Setting Sheet Footnotes
# User defined
† Calibrator concentrations: see Calibration
¤ Analyser default values

EN-ap-li Human Gesellschaft für Biochemica und Diagnostica mbH


05-2018-1 Max-Planck-Ring 21 - D-65205 Wiesbaden - Germany
Telefon: +49 6122 9988 0 - Telefax: +49 6122 9988 100
Page 1 of 2 eMail: human@human.de – www.human.de
A P opt. liquicolor
Enzymatic color test for the
quantitative determination of alkaline phosphatase

AU 400
Specific Test Parameters
General LIH ISE Range

Test Name: ALP-2 ∇ < > Type: Serum ∇ Operation: Yes ∇

Sample: Volume 3 µL Dilution 0 µL Pre-Dilution Rate: 1


Reagents R1 Volume 200 µL Dilution 0 µL Min OD Max OD
R2 Volume 50 µL Dilution 0 µL L -0.1 H 2.5
Reagent OD Limit:
Wavelength: Pri. 410 ∇ Sec. 480 ∇ First L -0.1 First H 1.0
Method: RATE ∇ Last L -0.1 Last H 1.0
Reaction slope: + ∇ Dynamic Range:
Measuring Point 1: First 13 Last 19 L # H #
Measuring Point 2: First Last Correlation Factor:
Linearity : 15 % A 1 B 0
No Lag Time: NO ∇ On-board stability period: #

Specific Test Parameters


General LIH ISE Range

Test Name: ALP-2 ∇ < > Type: Serum ∇

Value/Flag: # ∇ Level L: # Level H: #


Normal Ranges: Age L Age H
Sex Year Month Year Month L H
❏ 1. # ∇ # # # # # #
❏ 2. # ∇ # # # # # #
❏ 3. # ∇ # # # # # #
❏ 4. # ∇ # # # # # #
❏ 5. # ∇ # # # # # #
❏ 6. # ∇ # # # # # #
7. Normal Ranges # #
8. Out of Range 1-6 L H # #
Panic Value: # # Unit: U/l Decimal places: 0

Calibration Specific
General ISE

Test Name: ALP-2 ∇ < > Type: Serum ∇

Calibration Type: AB ∇ Formula: Y = AX + B ∇ Counts: # Process: CONC ∇

Cal. No. OD CONC Factor/OD-L Factor/OD-H


Point 1: # † ¤ ¤
Point 2:
Point 3:
Point 4:
Point 5:
Point 6:
Point 7:
1-Point Cal. Point: ❏ With CONC-0 Extended Calibration: ∇
MB Type Factor: Calibration Stability Period: #

EN-ap-li Human Gesellschaft für Biochemica und Diagnostica mbH


05-2018-1 Max-Planck-Ring 21 - D-65205 Wiesbaden - Germany
Telefon: +49 6122 9988 0 - Telefax: +49 6122 9988 100
Page 2 of 2 eMail: human@human.de – www.human.de