HumaStar 200

Acid Phosphatase total


[REF] 10660 Material required but not provided:
[REF] 13160 Autocal
For detailed information about reagents [REF] 13511/13512 HumaTrol N, HumaTrol P
please refer to our corresponding leaflet [REF] 13951/13151 Serodos, Serodos plus
Test Name Total ACP 9 - Pathological ranges
Code AcPT #
Type Kinetic
Group Clin 1 10 - Result units
Units 1: U/l
1 - Pipetting 1st 2nd Units 2:
Sample vol: 15 µl Conversion: 0
Reagent 1 vol: 205 µl Decimal digits: 1
Reagent 2 vol:
Diluent vol: 0 µl 11 - Calibrator factor limits
Min factor: 260
2 - Time Max factor: 480
Incubation 1: 324 sec
Incubation 2:
Reading: 180 sec Calibration

3 - Wavelengths Number of standards: 1 *


Wavelength 1: 405 nm repetitions: 3
Wavelength 2: 700 nm Factor: cr
Validity (hours): 120
4 - Washing Interpolation type: Average
Needle: 3
Cuvette: 3 Calibrator: [AUTOCAL] **
Dilution rate: 1
5 - Incompatibility
1 Blank correction: yes
2 Validity: 12
3
4 Reagent R1 (working reagent):
Dissolve the content of one bottle [SUB]
6 - Limits with exactly 15 ml of [BUF]
Blank OD min: -0.1 Abs Sample Preperation:
Blank OD max: 0.5 Abs Stabilise samples, AUTOCAL and controls
Reaction slope: positive by addition of one drop of [STAB]
OD Range min: -0.1 Abs immediately after seperation / reconstitution.
OD Range max: 1.0 Abs
Min conc: 1 U/l * employed
Max conc: 70 U/l ** Refer to the target value sheet
Linearity: 20 cr calculated by run
Substr. Depletion: # User defined
Blank OD ∆ max:
NOTE:
7 - Autodilution
The settings in the software of your HUMAN analyzer should
Rate: 0 represent the latest version. If there are differences between the
Max OD: 0 Abs/min settings in this application sheet and those in your analyzer software,
kindly contact the Customer Service of HUMAN: customer-
8 - Dilutions support@human.de before changing the analyzer settings according
to this application sheet. The laboratory has to validate the results of
Serum 1:1
this method in its own responsibility.
Urine 1:1

Date: Version:
Human Gesellschaft für Biochemica und Diagnostica mbH
EN-ACP-200 2013-04 001
Max-Planck-Ring 21 - D-65205 Wiesbaden - Germany