HumaStar 200

anti-Streptolysin-O (Aso)
[REF] 11251P Material required but not provided:
[REF] 11351 ASO Standard
[REF] 13010 Turbidos
For detailed information about reagents
please refer to our corresponding leaflet
Test Name Anti-Streptolysin-O 9 - Pathological ranges
Code ASO #
Type Endpoint self blank
Group Turbi 10 - Result units
Units 1: IU/ml
1 - Pipetting 1st 2nd Units 2:
Sample vol: 3 µl Conversion: 0
Reagent 1 vol: 240 µl Decimal digits: 0
Reagent 2 vol: 60 µl
Diluent vol: 0 µl 11 - Calibrator factor limits
Min factor: 300
2 - Time Max factor: 1200
Incubation 1: 180 sec
Incubation 2: 576 sec
Reading: 36 sec Calibration

3 - Wavelengths Number of standards: 1 *


Wavelength 1: 546 nm repetitions: 3
Wavelength 2: Factor: cr
Validity (hours): 120
4 - Washing Interpolation type: Average
Needle: 3
Cuvette: 6 Calibrator: ASO Standard **
alternative Calibrator:
5 - Incompatibility Dilution rate: 1
1
2 Blank correction: yes
3 Validity (hours): 120
4
Reagents:
6 - Limits
Blank OD min: -0.500 Abs R1: [DIL]
Blank OD max: 0.800 Abs R2: [RGT]
Reaction slope positive C-Clean: sample position 60
OD Range min: -0.500 Abs * employed
OD Range max: 2.0 Abs ** Refer to the target value sheet
Min conc: 10 IU/ml
Max conc: 800 IU/ml cr calculated by run
# User defined

Blank OD ∆ max: NOTE:

The settings in the software of your HUMAN analyzer should represent the
7 - Autodilution latest version. If there are differences between the settings in this
Rate: 0 application sheet and those in your analyzer software, kindly contact the
Max OD: 0 Abs Customer Service of HUMAN: customer-support@human.de before
changing the analyzer settings according to this application sheet. The
laboratory has to validate the results of this method in its own
8 - Dilutions responsibility.
Serum 1:1

Date: Version:
Human Gesellschaft für Biochemica und Diagnostica mbH
SU-ASO-200 2014-08 001
Max-Planck-Ring 21 - D-65205 Wiesbaden - Germany
Tel.: +496122 99880 Fax: +496122 9988100 - e-mail: human@human.de