Reagent definition ASO

Reagent ASO BU TRIS ASO LATEX


Stable on board (days) 1** 1**
Alarm limit (ml)
Vial volume 10 10
Barcode id
Syringe speed Normal Slow

Alarm limit settings: 1 ml for 10 ml vial, 2 ml for 20 ml vial, 5 ml for 60 ml vial. Check the correct vial size.
** When stored tightly sealed at 2...8 °C, when not in use, the reagent can be used for 6 months after opening/reconstitution.

Test definition ASO


Test type Photometric Test in use YES
Full name Antistreptolysin O Low High Unit
MINDRAY
Test limit, low (⌦) 20 10000 IU/ml
Online name ASO Initial absorbance 0 2 A
Dilution limit 20 1000 IU/ml
Result unit IU/ml Secondary dil. 1 + 0 9
Number of decim. 0
Acceptance AUTOMATIC
Dilution 1 + 0

Sample type
Serum Plasma Urine Correction factor 1
CSF Other Correction bias 0 IU/ml

Konelab ™ / T SeriesANTISTREPTOLYSIN O - code: ASO0202/3/4


Check the reagent insert for the specific information of the reagent(s), calibrator(s) and control(s).
CE-marked application for Konelab / T Series by Thermo
⌦ Determination limit (measuring range low)
Fisher Scientific Oy, applicable for use outside the US.
Value is lot dependent, refer to the calibrators value sheet.
T Series is distributed by Siemens Medical Solutions
Diagnostics in selected countries.

Page 1/2
Document number: D06429_03_applic_ASO, Date: 10/09, s.w. 7.2

Calibration parameters ASO


Calibration type LINEAR Factor Bias
Repeat time (d) 0 Abs. error (mA) 10 Bias correction in use NO
Points/Calibrator Duplicate Rel. error (%) 10 Bias corr. repeat time (dd:hh)
Acceptance MANUAL Response limit (mA) Bias corr. limit (mA)
Type of calibrators SEPARATE Min * Total
Curve direction ASCENDING Max * Incremental
Calibrator id Calibrator Conc. ( ) Dil. ratio 1 + Bias cal. id
Concentration NaCl 0.9%l 0 0
Dil. ratio 1 + ASO MindrayCal 400 0

Test flow ASO


Blank YES
Antigen excess NO Normal cuvette

Reagent Sample and dilution Incubation Blank Reagent Incubation End point

Reagent Volume (µl) Time (sec.) Resp. min Reagent Time (sec.) Wavelength (nm)
ASO BU TRIS 2 120 * ASO LATEX 300 510
Volume (µl) Disp. with Resp. max Volume (µl) Side wavel. (nm)
100 EXTRA * 100 700
Disp. with Volume (µl) Disp. with
EXTRA 10 EXTRA
Volume (µl) Dilution with Volume (µl)
10 water. 10
Wash reagent Wash reagent Wash reagent Meas. type
NONE NONE NONE FIXED TIME

Konelab ™ / T Series ASLO II - code:0303


Check the reagent insert for the specific information of the reagent(s), calibrator(s) and control(s).

⌦ Determination limit (measuring range low)


Value is lot dependent, refer to the calibrators value sheet.