You are on page 1of 1

Document number: 841103_K_applic_CANNAB, Date: 02/04, s.w. 6.

0
Test definition CANNAB Calibr. parameters
Test type Photometric Test in use YES Calibration type NONLINEAR
Full name Cannabinoids Low High Units Bias corr. in use NO
Online name Test limit 0 100 µg/l Repeat time (d) 7
Result Unit µg/l low( Ö ), high (!) Abs. error (mA) 5
Point/Calibrator Single
Number of decimals 0 Initial absorbance 0 2 A Rel. error (%) 0
Acceptance MANUAL
Acceptance AUTOMATIC Response limit
(mA)
Type of calibrator SEPARATE Min. *
Dilution 1 + 0 Dilution limit (!) * * µg/l Calibrator Conc.(!) Dil. Ratio. 1+ Max. *
Sample type DoA Cal 0 0 0
Secondary dil. 1+ 0 0 DoA Cal C1 20 0
¨Serum ¨Plasma ýUrine DoA Cal C2 50 0
¨C.S.F ¨Other Correction factor 1 DoA Cal C3 100 0
More
Correction bias 0

Test flow Reagent definition


Blank NO Fixed cuvette
Antigen excess NO

Syringe BCR Id. Stable


speed on board
(days)
Name
Curve type
Incubation Incubation
Reagent Volume (µl) Time (sec.) Reagent Time (sec.) λ 1 (nm) NONLINEAR
CANNAB1 8 300 CANNAB2 120 340 Nonlinearity
Conc.(µg/l)

Volume (µl) Disp. with Volume (µl) λ 2 (nm) 2 CANNAB1 Slow 407 30
55 WATER 55 NONE %

Disp. with Volume (µl) Disp. with 10 CANNAB2 Slow 408 30


Extra with 10 Extra with
buffers buffers Time (sec)
90
Volume (µl) Dilution with Volume (µl)
8 WATER 8
Points & Inter.

Wash reagent Wash reagent Wash reagent Konelab 6/18


60/30/20XT
NONE [WATER] NONE

Konelab™ CANNABINOIDS - code: 981623 (2 x 30 ml) for cut off 50 µg/l

Reagent: Liquids, ready-to-use. It is recommended to store the reagents tightly sealed at 2...8 °C when not in use.
Calibrator: DoA Negative Cal., 981720, DoA Cal. C1, 981729, DoA Cal. C2, 981730, DoA Cal. C3, 981731. Liquids, ready to use.

Q.C. sample: DoA Control set C, 981733. Liquids, ready to use.