Professional Documents
Culture Documents
Instrument Verification
2
Instrument Verification
3
Instrument Verification
4
Instrument Verification
5
Instrument Verification
6
Instrument Verification
7
Instrument Verification
8
Instrument Verification
Protocol Troubleshooting
Room temperature: 22 – 25°C Room temperature below 20
solutions stand at room temperature: 30 °C or above 25°C, standing
min times must be adjusted.
solutions stand in analyser: 30 min
Solutions used: Peltier / Thermistor cable
130 µl thermo labile with 240 µl Extra Heating resistance /
volume Thermistor cables
130 µl thermo stabile with 240 µl Extra Cooling module
volume
9
Instrument Verification
10
Instrument Verification
11
Instrument Verification
12
Instrument Verification
Protocol Troubleshooting
Solutions used: Lamp
0 – 2.5 A 340 nm filter
130 µl of reagent with 240 µl Extra Photocard
13
Instrument Verification
14
Instrument Verification
Protocol Troubleshooting
Solutions used: Needle tip
10 µl K2Cr2O 7 - 2.5A (L340_5) Bubbles in fluidic circuit
+ Ground cable
150 µl H2SO 4 - 0A (DISP_ACC)
15
Instrument Verification
16
Instrument Verification
Protocol Troubleshooting
Solutions used: Needle
Yellow food 24 A+ common salt Bubbles
0, 2, 4, 10 µl Ground cable
17
Instrument Verification
18
Instrument Verification
19
Instrument Verification
20
Instrument Verification
• Do not let the samples stand in the cups in the Konelab longer than
1 hour!
• Do not let the reagents stand open in the Konelab longer than 2
hours!
• From one set of bottles it is possible to perform the tests twice.
Second set is there because additional maintenances may take
longer than the half an hour that is left after the first test load is
analysed.
21
Instrument Verification
• Insert the empty sample segment no 1 into the Konelab and press
Start.
• Insert segment no 2 immediately after segment 1 is measured (start
again if necessary) and then segment 3 once the segment no 2 has
been measured.
• After all 3 segments are ready. Check the Accuracy Factors window
for right: lot number, factors and reference values.
22
Instrument Verification
23
Instrument Verification
24
Instrument Verification
25
Instrument Verification
26