Test definition PHENOB Calibr. parameters
Test type Photometric Test in use YES Calibration type LINEAR
Full name Phenobarbital Low High Units Bias corr. in use NO
Online name Test limit 4.1 345 µmol/l Repeat time (d) 0
Result Unit µmol/l low( Ö ), high (!) Abs. error (mA) 3
Point/Calibrator Duplicate
Number of decimals 1 Initial absorbance 0 2 A Rel. error (%) 1.5
Acceptance MANUAL
Acceptance AUTOMATIC Response limit
(mA)
Type of calibrator SEPARATE Min. *
Dilution 1 + 0 Dilution limit (!) * * µmol/l Calibrator Conc.(!) Dil. Ratio. 1+ Max. *
Sample type TDM Cal B0 0 0
Secondary dil. 1+ 0 0 TDM Cal B1 345 0
ýSerum ýPlasma ¨Urine
¨C.S.F ¨Other Correction factor 1
More
Correction bias 0
Volume (µl) Disp. with Volume (µl) λ 2 (nm) 9 PHENOB1 Slow 459 30
97 EXTRA 97 NONE %
Reagent: Dissolve the Reagent A lyoph. with Reagent A buffer and Reagent B lyoph. with Reagent B buffer.
It is recommended to store the reagents tightly sealed at 2...8 °C when not in use.
Calibrator: TDM Calibration set B, 981648. Liquids, ready to use.
! Values are lot dependent, refer to Calibration set value sheets.
Ö determination limit.