Test Parameter

Test Detail
Test IRON
Report Name IRON
Unit MICROGRAM/DL
Wavelenth-Primary 570
Assay Type 2 POINT
M1 Start 24
M2 Start 63
Sample Replicates 2
Control Replicate 1
Reaction Direction INCREASING
Prozone Limit % 0
Linearity Limit %
Technical Minimum 0.0
Y=aX+b A= 1
Reagent Abs Min 0.0

Test Volume
Test IRON
Sample Type Serum

Sample Volume

Normal 34 Dilution Ratio


Increase 34 Dilution Ratio
Decrease 34 Dilution Ratio

Standard Volume 34

Reagent Volumes and Stirrer Speed

RGT-1 Volume 200 R1 Stirrer Speed


RGT-2 Volume 20 R2 Stirrer Speed

Page 1
IRON

Reference Range

Test IRON
Sample Type Serum

Reference Range Default

Category Default

Reference Range – Decimal Places :2

Lower Limit Upper Limit

Normal

Panic 0 0

Calibration

Test IRON Curve Type

Calibration Expiry
Last Calibration Date Automatic Display Limit

Consumable: Reagent

Page 2
IRON

Consumable Type Reagent

Consumable IRON

Lot Detail

Manufacturer ERBA
Lot No as per current LOT
Expiry Date After 3 years Lot Status

Onboard Stability 0 Unit

Reagent Type R1 √
R2 √

Utility
Position Reagent Reagent TypeBottle Type
IRON R1 large
IRON R2 SMALL

Page 3
IRON

R2 time

m1s
Host Name IRON Auto Rerun √ M1E
Online Calibration M2S
Decimal Places 4 M2E
Secondary SELECT Total Reagent 2
Curve Type K-Factor Reagent R1 UIBCR1
M1 End 24 Reagent R2 UIBCR2
M2 End 63
Stadard Replicate 3
Control Interval 0
Reaction ABS limit 0
Prozone Check LOWER
Delta Abs/Min
Technical Maximum 2,000.0
b= 0
Reagent Abs Max 0.00

Sample Type

1 √ Serum
1 Urine
1 CSF
Plasma
Whole Blood
Other

ed

MEDIUM
MEDIUM

Page 4
IRON

Sample Type

√ Serum
Urine
CSF
Plasma
Whole Blood
Other

K Factor

0 Acceptable Limit % 0

Calibration Date: Automatic Display K Factor 1528

R1 Lot No

Page 5
IRON

Stock Over

Active (Automatic)

Hours

Lot No
as per current LOT
as per current LOT

Page 6
IRON

XL_1000

234
lag period
lag period after R2 Xl 1000 cycle
after R2 (in XL1000 (in N in
second) cycle No) instrument

Page 7
UIBC

Test Parameter

Test Detail
Test UIBC
Report Name UIBC
Unit MICROGRAM/DL
Wavelenth-Primary 570
Assay Type 2 POINT
M1 Start 24
M2 Start 63
Sample Replicates 2
Control Replicate 1
Reaction Direction INCREASING
Prozone Limit % 0
Linearity Limit %
Technical Minimum 0.0
Y=aX+b A= 1
Reagent Abs Min 0.0

Test Volume
Test UIBC
Sample Type Serum

Sample Volume

Normal 50 Dilution Ratio


Increase 50 Dilution Ratio
Decrease 50 Dilution Ratio

Standard Volume 50

Reagent Volumes and Stirrer Speed

RGT-1 Volume 250 R1 Stirrer Speed


RGT-2 Volume 10 R2 Stirrer Speed

Page 8
UIBC

Reference Range

Test UIBC
Sample Type Serum

Reference Range Default

Category Default

Reference Range – Decimal Places :2

Lower Limit Upper Limit

Normal 3.2 5

Panic 0 0

Calibration

Test UIBC Curve Type

Calibration Expiry
Last Calibration Date Automatic Display Limit

Consumable: Reagent

Page 9
UIBC

Consumable Type Reagent

Consumable UIBC

Lot Detail

Manufacturer ERBA
Lot No as per current LOT
Expiry Date After 3 years Lot Status

Onboard Stability 0 Unit

Reagent Type R1 √
R2 √

Utility
Position Reagent Reagent TypeBottle Type
49 UIBC R1 large
50 UIBC R2 SMALL

Page 10
UIBC

R2 time

m1s
Host Name URE Auto Rerun √ M1E
Online Calibration M2S
Decimal Places 4 M2E
Secondary SELECT Total Reagent 2
Curve Type K-Factor Reagent R1 UIBCR1
M1 End 24 Reagent R2 UIBCR2
M2 End 63
Stadard Replicate 3
Control Interval 0
Reaction ABS limit 0
Prozone Check LOWER
Delta Abs/Min
Technical Maximum 2,000.0
b= 0
Reagent Abs Max 0.00

Sample Type

1 √ Serum
1 Urine
1 CSF
Plasma
Whole Blood
Other

ed

MEDIUM
MEDIUM

Page 11
UIBC

Sample Type

√ Serum
Urine
CSF
Plasma
Whole Blood
Other

K Factor

0 Acceptable Limit % 0

Calibration Date: Automatic Display K Factor 1

R1 Lot No

Page 12
UIBC

Stock Over

Active (Automatic)

Hours

Lot No
as per current LOT
as per current LOT

Page 13
UIBC

XL_1000

234
lag period
lag period after R2 Xl 1000 cycle
after R2 (in XL1000 (in N in
second) cycle No) instrument

Page 14
HCL

Test Parameter
Test Detail
Test HCL
Report Name HCL
Unit mg/dl
Wavelenth-Primary 660
Assay Type 2 point
M1 Start 1
M2 Start 20
Sample Replicates 1
Control Replicate 1
Reaction Direction INCREASING
Prozone Limit % 0
Linearity Limit %
Technical Minimum 0.0
Y=aX+b A= 1
Reagent Abs Min 0.0

Test Volume
Test HCL
Sample Type Serum

Sample Volume

Normal 2 Dilution Ratio


Increase 2 Dilution Ratio
Decrease 2 Dilution Ratio

Standard Volume 2

Reagent Volumes and Stirrer Speed

RGT-1 Volume 300 R1 Stirrer Speed

Page 15
HCL

Reference Range

Test HCL
Sample Type Serum

Reference Range Default

Category Default

Reference Range – Decimal Places :2

Lower Limit Upper Limit

Normal 0 0

Panic 0 0

Calibration

Test HCL Curve Type

Calibration Expiry
Last Calibration Date Automatic Display Limit

Consumable: Reagent

Page 16
HCL

Consumable Type Reagent

Consumable HCL

Lot Detail

Manufacturer ERBA
Lot No as per current LOT
Expiry Date After 3 years Lot Status

Onboard Stability 0 Unit

Reagent Type R1 √

Utility
Position Reagent Reagent TypeBottle Type
HCL R1 SMALL

Page 17
HCL

m1s
Host Name HCLL Auto Rerun √ M1E
Online Calibration M2S
Decimal Places 1 M2E
Secondary SELECT Total Reagent 1
Curve Type K-Factor Reagent R1 HCL R1
M1 End 1
M2 End 20
Stadard Replicate 3
Control Interval 0
Reaction ABS limit 2
Prozone Check LOWER
Delta Abs/Min
Technical Maximum 0.0
b= 0
Reagent Abs Max 1.00

Sample Type

1 √ Serum
1 Urine
2 CSF
Plasma
Whole Blood
Other

ed

MEDIUM

Page 18
HCL

Sample Type

√ Serum
√ Urine
√ CSF
Plasma
Whole Blood
Other

K Factor

0 Acceptable Limit % 0

Calibration Date: Automatic Display K Factor 1

R1 Lot No

Page 19
HCL

Stock Over

Active (Automatic)

Hours

Lot No
as per current LOT

Page 20
HCL

Xl 1000 cycle
N in
instrument
1 1
1 1
180 10 10
180 10 10

Page 21
HYPO

Test Parameter
Test Detail
Test HYPO
Report Name HYPO
Unit MG/DL
Wavelenth-Primary 340
Assay Type 1 POINT
M1 Start 0
M2 Start 63
Sample Replicates 10
Control Replicate 1
Reaction Direction INCREASING
Prozone Limit % 0
Linearity Limit % 0
Technical Minimum 0.0
Y=aX+b A= 1
Reagent Abs Min 0.0

Test Volume
Test HYPPO
Sample Type Serum

Sample Volume

Normal 4 Dilution Ratio


Increase 4 Dilution Ratio
Decrease 4 Dilution Ratio

Standard Volume 4

Reagent Volumes and Stirrer Speed

RGT-1 Volume 200 R1 Stirrer Speed


RGT-2 Volume 20

Page 22
HYPO

Reference Range

Test HYPO
Sample Type Serum

Reference Range Default

Category Default

Reference Range – Decimal Places :2

Lower Limit Upper Limit

Normal 2 10

Panic 0 0

Calibration

Test HYPO Curve Type

Calibration Expiry
Last Calibration Date Automatic Display Limit

Consumable: Reagent

Page 23
HYPO

Consumable Type Reagent

Consumable HYPO

Lot Detail

Manufacturer ERBA
Lot No as per current LOT
Expiry Date After 3 years Lot Status

Onboard Stability 0 Unit

Reagent Type R1 √
R2 √

Utility
Position Reagent Reagent TypeBottle Type
54 HYPO R1 SMALL
56 HYPO R2 SMALL

Page 24
HYPO

Host Name HYPO Auto Rerun √


Online Calibration
Decimal Places 1
Secondary SELECT Total Reagent 2
Curve Type K-Factor Reagent R1 HYPO R1
M1 End 0 Reagent R1 HYPO R2
M2 End 63
Stadard Replicate 3
Control Interval 0
Reaction ABS limit 0
Prozone Check LOWER
Delta Abs/Min 0
Technical Maximum 0.0
b= 0
Reagent Abs Max 0.00

Sample Type

1 √ Serum
1 Urine
10 CSF
Plasma
Whole Blood
Other

ed

MEDIUM

Page 25
HYPO

Sample Type

√ Serum
Urine
CSF
Plasma
Whole Blood
Other

K Factor

0 Acceptable Limit % 0

Calibration Date: Automatic Display K Factor 1

R1 Lot No

Page 26
HYPO

Stock Over

Active (Automatic)

Hours

Lot No
as per current LOT
as per current LOT

Page 27
UREA

Test Parameter

Test Detail
Test URE
Report Name UREA
Unit MG/DL
Wavelenth-Primary 340
Assay Type 2 POIN T
M1 Start 28
M2 Start 35
Sample Replicates 1
Control Replicate 1
Reaction Direction DECREASING
Prozone Limit % 0
Linearity Limit %
Technical Minimum 0.0
Y=aX+b A= 1
Reagent Abs Min 0.0

Test Volume
Test URE
Sample Type Serum

Sample Volume

Normal 2 Dilution Ratio


Increase 2 Dilution Ratio
Decrease 2 Dilution Ratio

Standard Volume 2

Reagent Volumes and Stirrer Speed

RGT-1 Volume 160 R1 Stirrer Speed


RGT-2 Volume 40 R2 Stirrer Speed

Page 28
UREA

Reference Range

Test URE
Sample Type Serum

Reference Range Default

Category Default

Reference Range – Decimal Places :2

Lower Limit Upper Limit

Normal 13 43

Panic 0 0

Calibration

Test URE Curve Type

Calibration Expiry
Last Calibration Date Automatic Display Limit

Consumable: Reagent

Page 29
UREA

Consumable Type Reagent

Consumable URE

Lot Detail

Manufacturer ERBA
Lot No as per current LOT
Expiry Date After 3 years Lot Status

Onboard Stability 0 Unit

Reagent Type R1 √
R2 √

Utility
Position Reagent Reagent TypeBottle Type
11 URE R1 large
12 URE R2 SMALL

Page 30
UREA

R2 time

m1s
Host Name URE Auto Rerun √ M1E
Online Calibration M2S
Decimal Places 0 M2E
Secondary SELECT Total Reagent 2
Curve Type K-Factor Reagent R1 UREA R1
M1 End 28 Reagent R2 UREA R2
M2 End 35
Stadard Replicate 3
Control Interval 0
Reaction ABS limit -0.3
Prozone Check UPPER
Delta Abs/Min
Technical Maximum 0.0
b= 0
Reagent Abs Max 0.00

Sample Type

1 √ Serum
1 Urine
2 CSF
Plasma
Whole Blood
Other

ed

MEDIUM
MEDIUM

Page 31
UREA

Sample Type

√ Serum
Urine
CSF
Plasma
Whole Blood
Other

K Factor

0 Acceptable Limit % 0

Calibration Date: Automatic Display K Factor -589

R1 Lot No

Page 32
UREA

Stock Over

Active (Automatic)

Hours

Lot No
as per current LOT
as per current LOT

Page 33
UREA

234
lag period
lag period after R2 Xl 1000 cycle
after R2 (in XL1000 (in N in
second) cycle No) instrument
252 18 1 12
252
315 63 3.5 15
315

Page 34
TG

Carryover of CHOLESTEROL occur on Triglyceride.


so wash cycle should be kept before TG.

Test Parameter Kept Detergent wash 300 microliter 3 cycle

Test Detail
Test TG
Report Name TRIGLYCERIDES
Unit MG/DL
Wavelenth-Primary 505
Assay Type 2 POINT
M1 Start 1
M2 Start 63
Sample Replicates 1
Control Replicate 1
Reaction Direction INCREASING
Prozone Limit % 0
Linearity Limit %
Technical Minimum 0.0
Y=aX+b A= 1
Reagent Abs Min 0.0

Test Volume
Test TG
Sample Type Serum

Sample Volume

Normal 2 Dilution Ratio


Increase 2 Dilution Ratio
Decrease 2 Dilution Ratio

Standard Volume 2

Reagent Volumes and Stirrer Speed

RGT-1 Volume 200 R1 Stirrer Speed

Page 35
TG

Reference Range

Test TG
Sample Type Serum

Reference Range Default

Category Default

Reference Range – Decimal Places :2

Lower Limit Upper Limit

Normal 40 160

Panic 0 0

Calibration

Test TG Curve Type

Calibration Expiry
Last Calibration Date Automatic Display Limit

Consumable: Reagent

Page 36
TG

Consumable Type Reagent

Consumable TG

Lot Detail

Manufacturer ERBA
Lot No as per current LOT
Expiry Date After 3 years Lot Status

Onboard Stability 0 Unit

Reagent Type R1 √

Utility
Position Reagent Reagent TypeBottle Type
2 TG R1 SMALL

Page 37
TG

LESTEROL occur on Triglyceride. Carryover of CHOLESTEROL occur on Triglyceride.


ould be kept before TG. so wash cycle should be kept before TG.

sh 300 microliter 3 cycle Kept Detergent wash 300 microliter 3 cycle

m1s
Host Name TG Auto Rerun √ M1E
Online Calibration M2S
Decimal Places 0 M2E
Secondary SELECT Total Reagent 1
Curve Type K-Factor Reagent R1 TG R1
M1 End 1
M2 End 63
Stadard Replicate 3
Control Interval 0
Reaction ABS limit 2.2
Prozone Check LOWER
Delta Abs/Min
Technical Maximum 1,000.0
b= 0
Reagent Abs Max 0.10

Sample Type

1 √ Serum
1 Urine
2 CSF
Plasma
Whole Blood
Other

ed

MEDIUM

Page 38
TG

Sample Type

√ Serum
Urine
CSF
Plasma
Whole Blood
Other

K Factor

0 Acceptable Limit % 0

Calibration Date: Automatic Display K Factor 712.9

R1 Lot No

Page 39
TG

Stock Over

Active (Automatic)

Hours

Lot No
as per current LOT

Page 40
TG

cur on Triglyceride.

Xl 1000 cycle
N in
instrument
1 1
1 1
567 31.5 31
567 31.5 31

Page 41
CHE

Test Parameter

Test Detail
Test CHE
Report Name CHOLINESTERASE
Unit U/L
Wavelenth-Primary 405
Assay Type RATE A
M1 Start 0
M2 Start 5
Sample Replicates 1
Control Replicate 1
Reaction Direction INCREASING
Prozone Limit % 0
Linearity Limit % 0
Technical Minimum 500.0
Y=aX+b A= 1
Reagent Abs Min 0.0

Test Volume
Test CHE
Sample Type Serum

Sample Volume

Normal 2 Dilution Ratio


Increase 10 Dilution Ratio
Decrease 2 Dilution Ratio

Standard Volume 2

Reagent Volumes and Stirrer Speed

RGT-1 Volume 200 R1 Stirrer Speed

Page 42
CHE

Reference Range

Test CHE
Sample Type Serum

Reference Range Default

Category Default

Reference Range – Decimal Places :2

Lower Limit Upper Limit

Normal 3700 15000

Panic 0 0

Calibration

Test CHE Curve Type

Calibration Expiry
Last Calibration Date Automatic Display Limit

Consumable: Reagent

Page 43
CHE

Consumable Type Reagent

Consumable CHE

Lot Detail

Manufacturer ERBA
Lot No as per current LOT
Expiry Date After 3 years Lot Status

Onboard Stability 0 Unit

Reagent Type R1 √

Utility
Position Reagent Reagent TypeBottle Type
16 CHE R1 SMALL

Page 44
CHE

Host Name CHE Auto Rerun √


Online Calibration
Decimal Places 0
Secondary SELECT Total Reagent 1
Curve Type K-Factor Reagent R1 CHE R1
M1 End 0
M2 End 12
Stadard Replicate 3
Control Interval 0
Reaction ABS limit 2
Prozone Check LOWER
Delta Abs/Min 0
Technical Maximum 50,000.0
b= 0
Reagent Abs Max 1.00

Sample Type

1 √ Serum
1 Urine
1 CSF
Plasma
Whole Blood
Other

ed

MEDIUM

Page 45
CHE

Sample Type

√ Serum
Urine
CSF
Plasma
Whole Blood
Other

K Factor

0 Acceptable Limit % 0

Calibration Date: Automatic Display K Factor 32076.7

R1 Lot No

Page 46
CHE

Stock Over

Active (Automatic)

Hours

Lot No
as per current LOT

Page 47
CHE

for xl-1000

0
lag period
lag period after R1 Xl 1000 cycle
after R1 (in XL1000 (in N in
second) cycle No) instrument
m1s 0 0
M1E 0
M2S 3 3 0.166666667 2
M2E 7 7 0.388888889 5
technical minimum
technical maximum

use a and b rather than using factor.


as che has clinical importance at lower value
so by using “b” we can remove water colour and can give perfect result at lower value.
at room temprature, reaction will be occurred automatically which will give false results.

Page 48
HDL(CHOX)

Test Parameter
Test Detail
Test CHOX
Report Name HDL
Unit mg/dl
Wavelenth-Primary 505
Assay Type 2 point
M1 Start 1
M2 Start 63
Sample Replicates 1
Control Replicate 1
Reaction Direction INCREASING
Prozone Limit % 0
Linearity Limit %
Technical Minimum 0.0
Y=aX+b A= 1
Reagent Abs Min 0.0

Test Volume
Test CHOX
Sample Type Serum

Sample Volume

Normal 2 Dilution Ratio


Increase 2 Dilution Ratio
Decrease 2 Dilution Ratio

Standard Volume 2

Reagent Volumes and Stirrer Speed

RGT-1 Volume 200 R1 Stirrer Speed

Page 49
HDL(CHOX)

Reference Range

Test CHOX
Sample Type Serum

Reference Range Default

Category Default

Reference Range – Decimal Places :2

Lower Limit Upper Limit

Normal 120 250

Panic 0 0

Calibration

Test CHOX Curve Type

Calibration Expiry
Last Calibration Date Automatic Display Limit

Consumable: Reagent

Page 50
HDL(CHOX)

Consumable Type Reagent

Consumable CHO

Lot Detail

Manufacturer ERBA
Lot No as per current LOT
Expiry Date After 3 years Lot Status

Onboard Stability 0 Unit

Reagent Type R1 √

Utility
Position Reagent Reagent TypeBottle Type
10 CHO R1 SMALL

Page 51
HDL(CHOX)

m1s
Host Name CHOX Auto Rerun √ M1E
Online Calibration M2S
Decimal Places 0 M2E
Secondary SELECT Total Reagent 1
Curve Type K-Factor Reagent R1 CHO R1
M1 End 1
M2 End 63
Stadard Replicate 3
Control Interval 0
Reaction ABS limit 0
Prozone Check LOWER
Delta Abs/Min
Technical Maximum 750.0
b= 0
Reagent Abs Max 0.20

Sample Type

1 √ Serum
1 Urine
1 CSF
Plasma
Whole Blood
Other

ed

MEDIUM

Page 52
HDL(CHOX)

Sample Type

√ Serum
Urine
CSF
Plasma
Whole Blood
Other

K Factor

0 Acceptable Limit % 0

Calibration Date: Automatic Display K Factor 477.98

R1 Lot No

Page 53
HDL(CHOX)

Stock Over

Active (Automatic)

Hours

Lot No
as per current LOT

Page 54
HDL(CHOX)

Xl 1000 cycle
N in
instrument
1 1
1 1
567 31.5 31
567 31.5 31

Page 55
CK

Test Parameter

Test Detail
Test CK
Report Name CK
Unit mg/dl
Wavelenth-Primary 340
Assay Type RATE A
M1 Start 0
M2 Start 15
Sample Replicates 1
Control Replicate 1
Reaction Direction INCREASING
Prozone Limit % 0
Linearity Limit % 0
Technical Minimum 0.0
Y=aX+b A= 1
Reagent Abs Min 0.0

Test Volume
Test CK
Sample Type Serum

Sample Volume

Normal 4 Dilution Ratio


Increase 4 Dilution Ratio
Decrease 4 Dilution Ratio

Standard Volume 4

Reagent Volumes and Stirrer Speed

RGT-1 Volume 200 R1 Stirrer Speed

Page 56
CK

Reference Range

Test CK
Sample Type Serum

Reference Range Default

Category Default

Reference Range – Decimal Places :2

Lower Limit Upper Limit

Normal 0 200

Panic 0 0

Calibration

Test CK Curve Type

Calibration Expiry
Last Calibration Date Automatic Display Limit

Consumable: Reagent

Page 57
CK

Consumable Type Reagent

Consumable CK

Lot Detail

Manufacturer ERBA
Lot No as per current LOT
Expiry Date After 3 years Lot Status

Onboard Stability 0 Unit

Reagent Type R1 √

Utility
Position Reagent Reagent TypeBottle Type
38 CK R1 SMALL

Page 58
CK

XL_1000

m1s
Host Name CK Auto Rerun M1E
Online Calibration M2S
Decimal Places 0 M2E
Secondary SELECT Total Reagent 1
Curve Type K-Factor Reagent R1 CK R1 FILTER
M1 End 0 METHOD
M2 End 28
Stadard Replicate 3
Control Interval 0
Reaction ABS limit 2.4
Prozone Check LOWER
Delta Abs/Min 0
Technical Maximum 0.0
b= 0
Reagent Abs Max 0.50

Sample Type

1 √ Serum
1 Urine
2 CSF
Plasma
Whole Blood
Other

ed

MEDIUM

Page 59
CK

Sample Type

√ Serum
Urine
CSF
Plasma
Whole Blood
Other

K Factor

0 Acceptable Limit % 0

Calibration Date: Automatic Display K Factor 8947

R1 Lot No

Page 60
CK

Stock Over

Active (Automatic)

Hours

Lot No
as per current LOT

Page 61
CK

0
lag period
lag period after R1 Xl 1000 cycle
after R1 (in XL1000 (in N in
second) cycle No) instrument
0 0.444444444
0
8
13

340
RATE A

Page 62
CAL

Test Parameter
Test Detail
Test CAL
Report Name CALCIUM
Unit mg/dl
Wavelenth-Primary 660
Assay Type 2 point
M1 Start 1
M2 Start 20
Sample Replicates 1
Control Replicate 1
Reaction Direction INCREASING
Prozone Limit % 0
Linearity Limit %
Technical Minimum 0.0
Y=aX+b A= 1
Reagent Abs Min 0.0

Test Volume
Test CAL
Sample Type Serum

Sample Volume

Normal 2 Dilution Ratio


Increase 2 Dilution Ratio
Decrease 2 Dilution Ratio

Standard Volume 2

Reagent Volumes and Stirrer Speed

RGT-1 Volume 200 R1 Stirrer Speed

Page 63
CAL

Reference Range

Test CAL
Sample Type Serum

Reference Range Default

Category Default

Reference Range – Decimal Places :2

Lower Limit Upper Limit

Normal 8.5 10.3

Panic 0 0

Calibration

Test CAL Curve Type

Calibration Expiry
Last Calibration Date Automatic Display Limit

Consumable: Reagent

Page 64
CAL

Consumable Type Reagent

Consumable CAL

Lot Detail

Manufacturer ERBA
Lot No as per current LOT
Expiry Date After 3 years Lot Status

Onboard Stability 0 Unit

Reagent Type R1 √

Utility
Position Reagent Reagent TypeBottle Type
30 CAL R1 SMALL

Page 65
CAL

m1s
Host Name CAL Auto Rerun √ M1E
Online Calibration M2S
Decimal Places 1 M2E
Secondary SELECT Total Reagent 1
Curve Type K-Factor Reagent R1 CAL R1
M1 End 1
M2 End 20
Stadard Replicate 3
Control Interval 0
Reaction ABS limit 2
Prozone Check LOWER
Delta Abs/Min
Technical Maximum 0.0
b= 0
Reagent Abs Max 1.00

Sample Type

1 √ Serum
1 Urine
2 CSF
Plasma
Whole Blood
Other

ed

MEDIUM

Page 66
CAL

Sample Type

√ Serum
√ Urine
√ CSF
Plasma
Whole Blood
Other

K Factor

0 Acceptable Limit % 0

Calibration Date: Automatic Display K Factor 16.02

R1 Lot No

Page 67
CAL

Stock Over

Active (Automatic)

Hours

Lot No
as per current LOT

Page 68
CAL

Xl 1000 cycle
N in
instrument
1 1
1 1
180 10 10
180 10 10

Page 69
TP

Test Parameter
Test Detail
Test TP
Report Name TOTAL PROTEIN
Unit mg/dl
Wavelenth-Primary 546
Assay Type 2 point
M1 Start 1
M2 Start 63
Sample Replicates 1
Control Replicate 1
Reaction Direction INCREASING
Prozone Limit % 0
Linearity Limit %
Technical Minimum 0.0
Y=aX+b A= 1
Reagent Abs Min 0.0

Test Volume
Test TP
Sample Type Serum

Sample Volume

Normal 4 Dilution Ratio


Increase 4 Dilution Ratio
Decrease 4 Dilution Ratio

Standard Volume 4

Reagent Volumes and Stirrer Speed

RGT-1 Volume 200 R1 Stirrer Speed

Page 70
TP

Reference Range

Test TP
Sample Type Serum

Reference Range Default

Category Default

Reference Range – Decimal Places :2

Lower Limit Upper Limit

Normal 6.5 8.5

Panic 0 0

Calibration

Test TP Curve Type

Calibration Expiry
Last Calibration Date Automatic Display Limit

Consumable: Reagent

Page 71
TP

Consumable Type Reagent

Consumable TP

Lot Detail

Manufacturer ERBA
Lot No as per current LOT
Expiry Date After 3 years Lot Status

Onboard Stability 0 Unit

Reagent Type R1 √

Utility
Position Reagent Reagent TypeBottle Type
22 TP R1 SMALL

Page 72
TP

m1s
Host Name TP Auto Rerun √ M1E
Online Calibration M2S
Decimal Places 1 M2E
Secondary SELECT Total Reagent 1
Curve Type K-Factor Reagent R1 TP R1
M1 End 1
M2 End 63
Stadard Replicate 3
Control Interval 0
Reaction ABS limit 2.5
Prozone Check LOWER
Delta Abs/Min
Technical Maximum 15.0
b= 0
Reagent Abs Max 0.17

Test Volume
Test
Sample Type

Sample Type S

1 √ Serum Normal
1 √ Urine Increase
2 √ CSF Decrease
Plasma
Whole Blood Standard Vol
Other

ed Reagent Vo

MEDIUM RGT-1 Volume

Page 73
TP

e Reference R

Test TP
Sample Type urine

Reference Ra Default

Category Default

Sample Type Reference Range – Decimal Plac

√ Serum Lower Limit


√ Urine
√ CSF
Plasma Normal 0
Whole Blood
Other Panic 0

K Factor

0 Acceptable Limit % 0

Calibration Date: Automatic Display K Factor 13.33

R1 Lot No

Page 74
TP

Stock Over

Active (Automatic)

Hours

Lot No
as per current LOT

Page 75
TP

Xl 1000 cycle
N in
instrument
1 1
1 1
567 31.5 31
567 31.5 31

Test Volume
TP
urine

Sample Volume

4 Dilution Ratio 1
4 Dilution Ratio 1
4 Dilution Ratio 2

Reagent Volumes and Stirrer Speed

200 R1 Stirrer SpeMEDIUM

Page 76
TP

Reference Range Reference Range

Test TP
Sample Type csf

Reference Ra Default

Category Default

nce Range – Decimal Places :2 Reference Range – Decimal Places :2

Upper Limit Lower Limit

10 Normal 0

0 Panic 0

Page 77
TP

Test Volume
Test TP
Sample Type csf

Sample Volume

Normal 4 Dilution Ratio 1


Increase 4 Dilution Ratio 1
Decrease 4 Dilution Ratio 2

Standard Vol 4

Reagent Volumes and Stirrer Speed

RGT-1 Volume 200 R1 Stirrer SpeMEDIUM

Page 78
TP

Reference Range

nce Range – Decimal Places :2

Upper Limit

10

Page 79
TP

Page 80
CREAT

Test Parameter

Test Detail
Test CR
Report Name CREATINE
Unit mg/dl
Wavelenth-Primary 505
Assay Type 2 point
M1 Start 28 14
M2 Start 42 21
Sample Replicates 2
Control Replicate 1
Reaction Direction INCREASING
Prozone Limit % 0
Linearity Limit % 0
Technical Minimum 0.0
Y=aX+b A= 29.1
Reagent Abs Min 0.0

Test Volume
Test DBIL
Sample Type Serum

Sample Volume

Normal 16 Dilution Ratio


Increase 16 Dilution Ratio
Decrease 16 Dilution Ratio

Standard Volume 16

Reagent Volumes and Stirrer Speed

RGT-1 Volume 100 R1 Stirrer Speed


RGT-2 Volume 100 R2 Stirrer Speed

Page 81
CREAT

Reference Range

Test CR
Sample Type Serum

Reference Range Default

Category Default

Reference Range – Decimal Places :2

Lower Limit Upper Limit

Normal 0.7 1.3

Panic 0 0

Calibration

Test CR Curve Type

Calibration Expiry
Last Calibration Date Automatic Display Limit

Consumable: Reagent

Page 82
CREAT

Consumable Type Reagent

Consumable CR

Lot Detail

Manufacturer ERBA
Lot No as per current LOT
Expiry Date After 3 years Lot Status

Onboard Stability 0 Unit

Reagent Type R1 √
Reagent Type R2 √

Utility
Position Reagent Reagent TypeBottle Type
25 CR R1 large
27 CR R2 SMALL

Page 83
CREAT

14 7 R2 time

m1s
Host Name CR Auto Rerun √ M1E
Online Calibration M2S
Decimal Places 1 M2E
Secondary SELECT Total Reagent 2
Curve Type K-Factor Reagent R1 CR R1 r2-m2S
M1 End 28 14 Reagent R2 CR R2
M2 End 42 21
Stadard Replicate 3
Control Interval 0
Reaction ABS limit 0.7
Prozone Check LOWER
Delta Abs/Min 0.0
Technical Maximum 0
b= -0.48
Reagent Abs Max 0.0

Test Volume
Test
Sample Type

Sample Type

1 √ Serum Normal
1 √ Urine Increase
2 CSF Decrease
Plasma
Whole Blood Standard Vol
Other

ed Reagen

MEDIUM RGT-1 Volume


MEDIUM RGT-2 Volume

Page 84
CREAT

Sample Type

√ Serum
Urine
CSF
Plasma
Whole Blood
Other

K Factor

0 Acceptable Limit % 0

Calibration Date: Automatic Display K Factor 1

R1 Lot No

Page 85
CREAT

Stock Over

Active (Automatic)

Hours

Lot No
as per current LOT
as per current LOT

Page 86
CREAT

234
lag period
after R2 Xl 1000 cycle
lag period after R2 XL1000 (in N in
(in second) cycle No) instrument
252 18 1 12
252
378 126 7 19
378

144

Test Volume
DBIL
urine

Sample Volume

16 Dilution Ratio 10
16 Dilution Ratio 10
16 Dilution Ratio 10

16

Reagent Volumes and Stirrer Speed

100 R1 Stirrer SpeMEDIUM


100 R2 Stirrer SpeMEDIUM

Page 87
DBIL

Test Parameter

Test Detail
Test DBIL
Report Name DIRECT BILIRUBIN
Unit MG/DL
Wavelenth-Primary 546
Assay Type 2 point
M1 Start 5
M2 Start 63
Sample Replicates 1
Control Replicate 1
Reaction Direction INCREASING
Prozone Limit % 0
Linearity Limit %
Technical Minimum 0.0
Y=aX+b A= 1
Reagent Abs Min 0.0

Test Volume
Test DBIL
Sample Type Serum

Sample Volume

Normal 20 Dilution Ratio


Increase 20 Dilution Ratio
Decrease 20 Dilution Ratio

Standard Volume 20

Reagent Volumes and Stirrer Speed

RGT-1 Volume 160 R1 Stirrer Speed


RGT-2 Volume 40 R2 Stirrer Speed

Page 88
DBIL

Reference Range

Test ALT
Sample Type Serum

Reference Range Default

Category Default

Reference Range – Decimal Places :2

Lower Limit Upper Limit

Normal 1.1 2.3

Panic 0 0

Calibration

Test ALT Curve Type

Calibration Expiry
Last Calibration Date Automatic Display Limit

Consumable: Reagent

Page 89
DBIL

Consumable Type Reagent

Consumable DBIL

Lot Detail

Manufacturer ERBA
Lot No as per current LOT
Expiry Date After 3 years Lot Status

Onboard Stability 0 Unit

Reagent Type R1 √
Reagent Type R2 √

Utility
Position Reagent Reagent TypeBottle Type
5 DBIL R1 large
4 DBIL R2 SMALL

Page 90
DBIL

R2 time

m1s
Host Name DBIL Auto Rerun √ M1E
Online Calibration M2S
Decimal Places 1 M2E
Secondary SELECT Total Reagent 2
Curve Type K-Factor Reagent R1 DBIL R1 r2-m2S
M1 End 5 Reagent R2 TBIL R2
M2 End 63
Stadard Replicate 3
Control Interval 0
Reaction ABS limit 2
Prozone Check LOWER
Delta Abs/Min
Technical Maximum 25
b= 0
Reagent Abs Max 0.0

Sample Type

1 √ Serum
1 Urine
5 CSF
Plasma
Whole Blood
Other

ed

MEDIUM
MEDIUM

Page 91
DBIL

Sample Type

√ Serum
Urine
CSF
Plasma
Whole Blood
Other

K Factor

0 Acceptable Limit % 0

Calibration Date: Automatic Display K Factor 12.53

R1 Lot No

Page 92
DBIL

Stock Over

Active (Automatic)

Hours

Lot No
as per current LOT
as per current LOT

Page 93
DBIL

234
lag period lag period
lag period lag period after R1 after R2 Xl 1000 cycle
after R1 (in after R2 (in XL1000 (in XL1000 (in N in
second) second) cycle No) cycle No) instrument
45 45 2.5 3 3
45
567 333 18.5 29.5 31
567

333

Page 94
TBIL

Test Parameter

Test Detail
Test TBIL
Report Name TOTAL BILIRUBIN
Unit MG/DL
Wavelenth-Primary 546
Assay Type 2 point
M1 Start 5
M2 Start 63
Sample Replicates 1
Control Replicate 1
Reaction Direction INCREASING
Prozone Limit % 0
Linearity Limit %
Technical Minimum 0.0
Y=aX+b A= 1
Reagent Abs Min 0.0

Test Volume
Test TBIL
Sample Type Serum

Sample Volume

Normal 20 Dilution Ratio


Increase 20 Dilution Ratio
Decrease 20 Dilution Ratio

Standard Volume 20

Reagent Volumes and Stirrer Speed

RGT-1 Volume 160 R1 Stirrer Speed


RGT-2 Volume 40 R2 Stirrer Speed

Page 95
TBIL

Reference Range

Test ALT
Sample Type Serum

Reference Range Default

Category Default

Reference Range – Decimal Places :2

Lower Limit Upper Limit

Normal 1.1 2.3

Panic 0 0

Calibration

Test ALT Curve Type

Calibration Expiry
Last Calibration Date Automatic Display Limit

Consumable: Reagent

Page 96
TBIL

Consumable Type Reagent

Consumable TBIL

Lot Detail

Manufacturer ERBA
Lot No as per current LOT
Expiry Date After 3 years Lot Status

Onboard Stability 0 Unit

Reagent Type R1 √
Reagent Type R2 √

Utility
Position Reagent Reagent TypeBottle Type
3 TBIL R1 large
4 TBIL R2 SMALL

Page 97
TBIL

R2 time

m1s
Host Name TBIL Auto Rerun √ M1E
Online Calibration M2S
Decimal Places 1 M2E
Secondary SELECT Total Reagent 2
Curve Type K-Factor Reagent R1 TBIL R1 r2-m2S
M1 End 5 Reagent R2 TBIL R2
M2 End 63
Stadard Replicate 3
Control Interval 0
Reaction ABS limit 2
Prozone Check LOWER
Delta Abs/Min
Technical Maximum 0
b= 0
Reagent Abs Max 0.0

Sample Type

1 √ Serum
1 Urine
5 CSF
Plasma
Whole Blood
Other

ed

MEDIUM
MEDIUM

Page 98
TBIL

Sample Type

√ Serum
Urine
CSF
Plasma
Whole Blood
Other

K Factor

0 Acceptable Limit % 0

Calibration Date: Automatic Display K Factor 12.53

R1 Lot No

Page 99
TBIL

Stock Over

Active (Automatic)

Hours

Lot No
as per current LOT
as per current LOT

Page 100
TBIL

234
lag period lag period
lag period lag period after R1 after R2 Xl 1000 cycle
after R1 (in after R2 (in XL1000 (in XL1000 (in N in
second) second) cycle No) cycle No) instrument
45 45 2.5 3 3
45
567 333 18.5 29.5 31
567

333

Page 101
ALT

Test Parameter

Test Detail
Test ALT
Report Name ALANINE TRANSFERASE
Unit U/L
Wavelenth-Primary 340
Assay Type rate A
M1 Start 1
M2 Start 32
Sample Replicates 1
Control Replicate 1

Reaction Direction Decreasing


Prozone Limit % 0
Linearity Limit % 0
Technical Minimum 0.0
Y=aX+b A= 1
Reagent Abs Min 0.0

Test Volume
Test ALT
Sample Type Serum

Sample Volume

Normal 20 Dilution Ratio


Increase 20 Dilution Ratio
Decrease 20 Dilution Ratio

Standard Volume 20

Reagent Volumes and Stirrer Speed

Page 102
ALT

RGT-1 Volume 160 R1 Stirrer Speed


RGT-2 Volume 40 R2 Stirrer Speed

Reference Range

Test ALT
Sample Type Serum

Reference Range Default

Category Default

Reference Range – Decimal Places :2

Lower Limit Upper Limit

Normal 0 45

Panic 0 0

Calibration

Test ALT Curve Type

Calibration Expiry
Last Calibration Date Automatic Display Limit

Page 103
ALT

Consumable: Reagent

Consumable Type Reagent

Consumable ALT

Lot Detail

Manufacturer ERBA
Lot No as per current LOT
Expiry Date After 3 years Lot Status

Onboard Stability 0 Unit

Reagent Type R1 √
Reagent Type R2 √

Utility
Position Reagent Reagent TypeBottle Type
41 ALT R1 large
42 ALT R2 SMALL

Page 104
ALT

R2 time

m1s
Host Name ALT Auto Rerun √ M1E
Online Calibration M2S
Decimal Places 0 M2E
Secondary SELECT Total Reagent 2
Curve Type K-Factor Reagent R1 ALT R1 r2-m2S
M1 End 1 Reagent R2 ALT R2
M2 End 52
Stadard Replicate 3
Control Interval 0
when reaction
absorbance is
below 0.9, at M2E
sample will be
undergo dilution.
Reaction ABS limit 0.9 NOTE:- (auto rerun)
Prozone Check upper
Delta Abs/Min 0.0
Technical Maximum 0.0
b= 0
Reagent Abs Max 0.0

Sample Type

1 √ Serum
1 Urine
10 CSF
Plasma
Whole Blood
Other

ed

Page 105
ALT

MEDIUM
MEDIUM

Sample Type

√ Serum
Urine
CSF
Plasma
Whole Blood
Other

K Factor

0 Acceptable Limit % 0

Calibration Date: Automatic Display K Factor -1909

R1 Lot No

Page 106
ALT

Stock Over

Active (Automatic)

Hours

Lot No
as per current LOT
as per current LOT

Page 107
ALT

234
lag period
lag period after R2 Xl 1000 cycle
after R2 (in XL1000 (in N in
second) cycle No) instrument
1 0 0
1 0
288 54 3 14
468 234 13 27

54

Page 108
GLC

Test Parameter
Test Detail
Test GLC
Report Name GLUCOSE
Unit Mg/dl
Wavelenth-Primary 505
Assay Type 2 point
M1 Start 1
M2 Start 63
Sample Replicates 1
Control Replicate 1
Reaction Direction increasing
Prozone Limit % 0
Linearity Limit %
Technical Minimum 0.0
Y=aX+b A= 1
Reagent Abs Min 0.0

Test Volume
Test GLC
Sample Type Serum

Sample Volume

Normal 2 Dilution Ratio


Increase 2 Dilution Ratio
Decrease 2 Dilution Ratio

Standard Volume 2

Reagent Volumes and Stirrer Speed

RGT-1 Volume 200 R1 Stirrer Speed

Page 109
GLC

Reference Range

Test GLC
Sample Type Serum

Reference Range Default

Category Default

Reference Range – Decimal Places :2

Lower Limit Upper Limit

Normal 60 120

Panic 50 500

Calibration

Test GLC Curve Type

Calibration Expiry
Last Calibration Date Automatic Display Limit

Consumable: Reagent

Page 110
GLC

Consumable Type Reagent

Consumable GLC

Lot Detail

Manufacturer ERBA
Lot No as per current LOT
Expiry Date After 3 years Lot Status

Onboard Stability 0 Unit

Reagent Type R1 √

Utility
Position Reagent Reagent TypeBottle Type
1 GLC R1 large

Page 111
GLC

m1s
Host Name GLC Auto Rerun √ M1E
Online Calibration M2S
Decimal Places 0 M2E
Secondary SELECT Total Reagent 1
Curve Type K-Factor Reagent R1 GLC R1
M1 End 1
M2 End 63
Stadard Replicate 3
Control Interval 0
Reaction ABS limit 1.9
Prozone Check lower
Delta Abs/Min
Technical Maximum 0
b= 0
Reagent Abs Max 0.2

Sample Type

1 √ Serum
1 Urine
2 CSF
Plasma
Whole Blood
Other

ed

MEDIUM

Page 112
GLC

Sample Type

√ Serum
Urine
CSF
Plasma
Whole Blood
Other

K Factor

0 Acceptable Limit % 0

Calibration Date: Automatic Display K Factor 261.7

R1 Lot No

Page 113
GLC

Stock Over

Active (Automatic)

Hours

Lot No
as per current LOT

Page 114
GLC

Xl 1000 cycle
N in
instrument
1 1
1 1
567 31.5 31
567 31.5 31

Page 115
MPR

Test Parameter
Test Detail
Test MPR
Report Name Micro-protein
Unit Mg/dl
Wavelenth-Primary 600
Assay Type 2 point
M1 Start 1
M2 Start 63
Sample Replicates 2
Control Replicate 1
Reaction Direction increasing
Prozone Limit % 0
Linearity Limit %
Technical Minimum 0.0
Y=aX+b A= 1
Reagent Abs Min 0.0

Test Volume
Test MPR
Sample Type Serum

Sample Volume

Normal 4 Dilution Ratio


Increase 4 Dilution Ratio
Decrease 4 Dilution Ratio

Standard Volume 4

Reagent Volumes and Stirrer Speed

RGT-1 Volume 250 R1 Stirrer Speed

Page 116
MPR

Reference Range

Test MPR
Sample Type Serum

Reference Range Default

Category Default

Reference Range – Decimal Places :2

Lower Limit Upper Limit

Normal 0 30

Panic 0 0

Calibration

Test MPR Curve Type

Calibration Expiry
Last Calibration Date Automatic Display Limit

Consumable: Reagent

Page 117
MPR

Consumable Type Reagent

Consumable MPR

Lot Detail

Manufacturer ERBA
Lot No as per current LOT
Expiry Date After 3 years Lot Status

Onboard Stability 0 Unit

Reagent Type R1 √

Utility
Position Reagent Reagent TypeBottle Type
32 MPR R1 SMALL

Page 118
MPR

m1s
Host Name MPR Auto Rerun M1E
Online Calibration M2S
Decimal Places 1 M2E
Secondary SELECT Total Reagent 1
Curve Type K-Factor Reagent R1 MPR R1
M1 End 1
M2 End 63
Stadard Replicate 3
Control Interval 0
Reaction ABS limit 0
Prozone Check lower
Delta Abs/Min
Technical Maximum 200
b= 0
Reagent Abs Max 0.2

Test
Sample Type

Sample Type S

1 √ Serum Normal
1 √ Urine Increase
5 √ CSF Decrease
Plasma
Whole Blood Standard Vol
Other

ed Reagent Vo

MEDIUM RGT-1 Volume

Page 119
MPR

e Reference R

Test MPR
Sample Type urine

Reference Ra Default

Category Default

Sample Type Reference Range – Decimal Plac

√ Serum Lower Limit


√ Urine
√ CSF
Plasma Normal 21.3
Whole Blood
Other Panic 0

K Factor

0 Acceptable Limit % 0

Calibration Date: Automatic Display K Factor 447

R1 Lot No

Page 120
MPR

Stock Over

Active (Automatic)

Hours

Lot No
as per current LOT

Page 121
MPR

Xl 1000 cycle
N in
instrument
1 1
1 1
567 31.5 31
567 31.5 31

MPR
urine

Sample Volume Sample Type

4 Dilution Ratio 1 √ Serum


4 Dilution Ratio 1 √ Urine
4 Dilution Ratio 5 √ CSF
Plasma
4 Whole Blood
Other

Reagent Volumes and Stirrer Speed

250 R1 Stirrer SpeMEDIUM

Page 122
MPR

Reference Range Reference

Test
Sample Type

Reference Ra

Category

nce Range – Decimal Places :2 Reference Range – Decimal

Upper Limit

119.6 Normal

0 Panic

Page 123
MPR

Test MPR
Sample Type csf

Sample Volume

Normal 4 Dilution Ratio 1


Increase 4 Dilution Ratio 1
Decrease 4 Dilution Ratio 5

Standard Vol 4

Reagent Volumes and Stirrer Speed

RGT-1 Volume 250 R1 Stirrer SpeMEDIUM

Page 124
MPR

Reference Range

MPR
Serum

Default

Default

Reference Range – Decimal Places :2

Lower Limit Upper Limit

60 120

0 0

Page 125
MPR

Sample Type

√ Serum
√ Urine
√ CSF
Plasma
Whole Blood
Other

Page 126
CGLC

Test Parameter
Test Detail
Test CGLC
Report Name GLUCOSE
Unit Mg/dl
Wavelenth-Primary 505
Assay Type 2 point
M1 Start 1
M2 Start 63
Sample Replicates 1
Control Replicate 1
Reaction Direction increasing
Prozone Limit % 0
Linearity Limit %
Technical Minimum 0.0
Y=aX+b A= 1
Reagent Abs Min 0.0

Test Volume
Test CGLC
Sample Type Serum

Sample Volume

Normal 2 Dilution Ratio


Increase 2 Dilution Ratio
Decrease 2 Dilution Ratio

Standard Volume 2

Reagent Volumes and Stirrer Speed

RGT-1 Volume 200 R1 Stirrer Speed

Page 127
CGLC

Reference Range

Test CGLC
Sample Type Serum

Reference Range Default

Category Default

Reference Range – Decimal Places :2

Lower Limit Upper Limit

Normal 60 120

Panic 0 0

Calibration

Test CGLC Curve Type

Calibration Expiry
Last Calibration Date Automatic Display Limit

Consumable: Reagent

Page 128
CGLC

Consumable Type Reagent

Consumable CGLC

Lot Detail

Manufacturer ERBA
Lot No as per current LOT
Expiry Date After 3 years Lot Status

Onboard Stability 0 Unit

Reagent Type R1 √

Utility
Position Reagent Reagent TypeBottle Type
1 CGLC R1 large

Page 129
CGLC

Host Name CGLC Auto Rerun √


Online Calibration
Decimal Places 0
Secondary SELECT Total Reagent 1
Curve Type K-Factor Reagent R1 GLC R1
M1 End 1
M2 End 63
Stadard Replicate 3
Control Interval 0
Reaction ABS limit 2
Prozone Check lower
Delta Abs/Min
Technical Maximum 0
b= 0
Reagent Abs Max 0.2

Sample Type

1 √ Serum
1 Urine
2 CSF
Plasma
Whole Blood
Other

ed

MEDIUM

Page 130
CGLC

Sample Type

√ Serum
Urine
CSF
Plasma
Whole Blood
Other

K Factor

0 Acceptable Limit % 0

Calibration Date: Automatic Display K Factor 261.7

R1 Lot No

Page 131
CGLC

Stock Over

Active (Automatic)

Hours

Lot No
as per current LOT

Page 132
ALB

Test Parameter
Test Detail
Test ALB
Report Name ALBUMIN
Unit g/dl
Wavelenth-Primary 600
Assay Type 2 point
M1 Start 1
M2 Start 5
Sample Replicates 1
Control Replicate 1
Reaction Direction increasing
Prozone Limit % 0
Linearity Limit %
Technical Minimum 0.0
Y=aX+b A= 1
Reagent Abs Min 0.0

Test Volume
Test ALB
Sample Type Serum

Sample Volume

Normal 2 Dilution Ratio


Increase 2 Dilution Ratio
Decrease 2 Dilution Ratio

Standard Volume 2

Reagent Volumes and Stirrer Speed

RGT-1 Volume 300 R1 Stirrer Speed

Page 133
ALB

Reference Range

Test ALB
Sample Type Serum

Reference Range Default

Category Default

Reference Range – Decimal Places :2

Lower Limit Upper Limit

Normal 3.2 5

Panic 0 0

Calibration

Test ALB Curve Type

Calibration Expiry
Last Calibration Date Automatic Display Limit

Consumable: Reagent

Page 134
ALB

Consumable Type Reagent

Consumable ALB

Lot Detail

Manufacturer ERBA
Lot No as per current LOT
Expiry Date After 3 years Lot Status

Onboard Stability 0 Unit

Reagent Type R1 √

Utility
Position Reagent Reagent TypeBottle Type
31 ALB R1 large

Page 135
ALB

m1s
Host Name ALB Auto Rerun √ M1E
Online Calibration M2S
Decimal Places 1 M2E
Secondary SELECT Total Reagent 1
Curve Type K-Factor Reagent R1 ALB R1
M1 End 1
M2 End 5
Stadard Replicate 3
Control Interval 0
Reaction ABS limit 0
Prozone Check lower
Delta Abs/Min
Technical Maximum 7
b= 0
Reagent Abs Max 0.4

Sample Type

1 √ Serum
1 Urine
5 CSF
Plasma
Whole Blood
Other

ed

LOW

Page 136
ALB

Sample Type

√ Serum
Urine
CSF
Plasma
Whole Blood
Other

K Factor

0 Acceptable Limit % 0

Calibration Date: Automatic Display K Factor 5.36

R1 Lot No

Page 137
ALB

Stock Over

Active (Automatic)

Hours

Lot No
as per current LOT

Page 138
ALB

Xl 1000 cycle
N in
instrument
1 1
1 1
45 2.5 3
45 2.5 3

Page 139
CHO

Carryover of CHOLESTEROL occur on Triglyceride.


so wash cycle should be kept before TG.

Test Parameter Kept Detergent wash 300 microliter 3 cycle

Test Detail
Test CHO
Report Name Total cholesterol
Unit mg/dl
Wavelenth-Primary 505
Assay Type 2 point
M1 Start 1
M2 Start 63
Sample Replicates 1
Control Replicate 1
Reaction Direction increasing
Prozone Limit % 0
Linearity Limit %
Technical Minimum 0.0
Y=aX+b A= 1
Reagent Abs Min 0.0

Test Volume
Test CHO
Sample Type Serum

Sample Volume

Normal 2 Dilution Ratio


Increase 2 Dilution Ratio
Decrease 2 Dilution Ratio

Standard Volume 2

Reagent Volumes and Stirrer Speed

RGT-1 Volume 200 R1 Stirrer Speed

Page 140
CHO

Reference Range

Test CHO
Sample Type Serum

Reference Range Default

Category Default

Reference Range – Decimal Places :2

Lower Limit Upper Limit

Normal 120 250

Panic 0 0

Calibration

Test CHO Curve Type

Calibration Expiry
Last Calibration Date Automatic Display Limit

Consumable: Reagent

Page 141
CHO

Consumable Type Reagent

Consumable CHO

Lot Detail

Manufacturer ERBA
Lot No as per current LOT
Expiry Date After 3 years Lot Status

Onboard Stability 0 Unit

Reagent Type R1 √

Utility
Position Reagent Reagent TypeBottle Type
10 CHO R1 SMALL

Page 142
CHO

LESTEROL occur on Triglyceride. Carryover of CHOLESTEROL occur on Triglyceride.


ould be kept before TG. so wash cycle should be kept before TG.

sh 300 microliter 3 cycle Kept Detergent wash 300 microliter 3 cycle

m1s
Host Name CHO Auto Rerun √ M1E
Online Calibration M2S
Decimal Places 0 M2E
Secondary SELECT Total Reagent 1
Curve Type K-Factor Reagent R1 CHO R1
M1 End 1
M2 End 63
Stadard Replicate 3
Control Interval 0
Reaction ABS limit 0
Prozone Check lower
Delta Abs/Min
Technical Maximum 750
b= 0
Reagent Abs Max 0.2

Sample Type

1 √ Serum
1 Urine
5 CSF
Plasma
Whole Blood
Other

ed

HIGH

Page 143
CHO

Sample Type

√ Serum
Urine
CSF
Plasma
Whole Blood
Other

K Factor

0 Acceptable Limit % 0

Calibration Date: Automatic Display K Factor 520.24

R1 Lot No

Page 144
CHO

Stock Over

Active (Automatic)

Hours

Lot No
as per current LOT

Page 145
CHO

cur on Triglyceride.

Xl 1000 cycle
N in
instrument
1 1
1 1
567 31.5 31
567 31.5 31

Page 146
UA

Test Parameter
Test Detail
Test UA
Report Name Uric acid
Unit mg/dl
Wavelenth-Primary 505
Assay Type 2 point
M1 Start 1
M2 Start 63
Sample Replicates 1
Control Replicate 1
Reaction Direction increasing
Prozone Limit % 0
Linearity Limit %
Technical Minimum 0.0
Y=aX+b A= 1
Reagent Abs Min 0.0

Test Volume
Test UA
Sample Type Serum

Sample Volume

Normal 5 Dilution Ratio


Increase 5 Dilution Ratio
Decrease 5 Dilution Ratio

Standard Volume 5

Reagent Volumes and Stirrer Speed

RGT-1 Volume 200 R1 Stirrer Speed

Page 147
UA

Reference Range

Test UA
Sample Type Serum

Reference Range Default

Category Default

Reference Range – Decimal Places :2

Lower Limit Upper Limit

Normal 3.4 7.4

Panic 0 0

Calibration

Test UA Curve Type

Calibration Expiry
Last Calibration Date Automatic Display Limit

Consumable: Reagent

Page 148
UA

Consumable Type Reagent

Consumable UA

Lot Detail

Manufacturer ERBA
Lot No as per current LOT
Expiry Date After 3 years Lot Status

Onboard Stability 0 Unit

Reagent Type R1 √

Utility
Position Reagent Reagent TypeBottle Type
26 UA R1 SMALL

Page 149
UA

m1s
Host Name UA Auto Rerun √ M1E
Online Calibration M2S
Decimal Places 1 M2E
Secondary SELECT Total Reagent 1
Curve Type K-Factor Reagent R1 UA R1
M1 End 1
M2 End 63
Stadard Replicate 3
Control Interval 0
Reaction ABS limit 0
Prozone Check lower
Delta Abs/Min
Technical Maximum 20
b= 0
Reagent Abs Max 0.2

Sample Type

1 √ Serum
1 Urine
10 CSF
Plasma
Whole Blood
Other

ed

MEDIUM

Page 150
UA

Sample Type

√ Serum
Urine
CSF
Plasma
Whole Blood
Other

K Factor

0 Acceptable Limit % 0

Calibration Date: Automatic Display K Factor 26

R1 Lot No

Page 151
UA

Stock Over

Active (Automatic)

Hours

Lot No
as per current LOT

Page 152
UA

Xl 1000 cycle
N in
instrument
1 1
1 1
567 31.5 31
567 31.5 31

Page 153
PHO

Carry over of PHOSPHORUS reagent occur on many test


EXAMPLE glucose over phosphorus

Test Parameter
Test Detail
Test PHO
Report Name phosphorus
Unit mg/dl
Wavelenth-Primary 340
Assay Type 2 point
M1 Start 1
M2 Start 3
Sample Replicates 2
Control Replicate 1
Reaction Direction increasing
Prozone Limit % 0
Linearity Limit %
Technical Minimum 0.0
Y=aX+b A= 1
Reagent Abs Min 0.0

Test Volume
Test PHO
Sample Type Serum

Sample Volume

Normal 2 Dilution Ratio


Increase 2 Dilution Ratio
Decrease 2 Dilution Ratio

Standard Volume 2

Reagent Volumes and Stirrer Speed

RGT-1 Volume 180 R1 Stirrer Speed

Page 154
PHO

Reference Range

Test PHO
Sample Type Serum

Reference Range Default

Category Default

Reference Range – Decimal Places :2

Lower Limit Upper Limit

Normal 2.5 4.5

Panic 0 0

Calibration

Test PHO Curve Type

Calibration Expiry
Last Calibration Date Automatic Display Limit

Consumable: Reagent

Page 155
PHO

Consumable Type Reagent

Consumable PHO

Lot Detail

Manufacturer ERBA
Lot No as per current LOT
Expiry Date After 3 years Lot Status

Onboard Stability 0 Unit

Reagent Type R1 √

Utility
Position Reagent Reagent TypeBottle Type
13 PHO R1 large

Page 156
PHO

cur on many test

m1s
Host Name PHO Auto Rerun √ M1E
Online Calibration M2S
Decimal Places 1 M2E
Secondary SELECT Total Reagent 1
Curve Type K-Factor Reagent R1 PHO R1
M1 End 1
M2 End 3
Stadard Replicate 3
Control Interval 0
Reaction ABS limit 0
Prozone Check lower
Delta Abs/Min
Technical Maximum 15
b= 0
Reagent Abs Max 0.0

Sample Type

1 √ Serum
1 Urine
2 CSF
Plasma
Whole Blood
Other

ed

HIGH

Page 157
PHO

Sample Type

√ Serum
Urine
CSF
Plasma
Whole Blood
Other

K Factor

0 Acceptable Limit % 0

Calibration Date: Automatic Display K Factor 15.09

R1 Lot No

Page 158
PHO

Stock Over

Active (Automatic)

Hours

Lot No
as per current LOT

Page 159
PHO

Xl 1000 cycle
N in
instrument
1 1
1 1
27 1.5 2
27 1.5 2

Page 160
CKMB

Test Parameter

Test Detail
Test CKMB
Report Name CK-MB
Unit U/L
Wavelenth-Primary 340
Assay Type RATE-A
M1 Start 0
M2 Start 49
Sample Replicates 1
Control Replicate 1
Reaction Direction increasing
Prozone Limit % 0
Linearity Limit % 0
Technical Minimum 0.0
Y=aX+b A= 1
Reagent Abs Min 0.0

Test Volume
Test CKMB
Sample Type Serum

Sample Volume

Normal 20 Dilution Ratio


Increase 20 Dilution Ratio
Decrease 20 Dilution Ratio

Standard Volume 20

Reagent Volumes and Stirrer Speed

RGT-1 Volume 200 R1 Stirrer Speed

Page 161
CKMB

Reference Range

Test CKMB
Sample Type Serum

Reference Range Default

Category Default

Reference Range – Decimal Places :2

Lower Limit Upper Limit

Normal 0 25

Panic 0 0

Calibration

Test CKMB Curve Type

Calibration Expiry
Last Calibration Date Automatic Display Limit

Consumable: Reagent

Page 162
CKMB

Consumable Type Reagent

Consumable CKMB

Lot Detail

Manufacturer ERBA
Lot No as per current LOT
Expiry Date After 3 years Lot Status

Onboard Stability 0 Unit

Reagent Type R1 √

Utility
Position Reagent Reagent TypeBottle Type
36 CKMB R1 SMALL

Page 163
CKMB

m1s
Host Name CKMB Auto Rerun √ M1E
Online Calibration M2S
Decimal Places 0 M2E
Secondary SELECT Total Reagent 1
Curve Type K-Factor Reagent R1 CKMB R1
M1 End 0
M2 End 63
Stadard Replicate 3
Control Interval 0
Reaction ABS limit 2.4
Prozone Check lower
Delta Abs/Min 0.0
Technical Maximum 0
b= 0
Reagent Abs Max 0.5

Sample Type

1 √ Serum
1 Urine
2 CSF
Plasma
Whole Blood
Other

ed

HIGH

Page 164
CKMB

Sample Type

√ Serum
Urine
CSF
Plasma
Whole Blood
Other

K Factor

0 Acceptable Limit % 0

Calibration Date: Automatic Display K Factor 3537

R1 Lot No

Page 165
CKMB

Stock Over

Active (Automatic)

Hours

Lot No
as per current LOT

Page 166
CKMB

0
lag period
lag period after R1 Xl 1000 cycle
after R1 (in XL1000 (in N in
second) cycle No) instrument
1 0
1
441 441 24.5 24
567 567 31.5 31

Page 167
AMY

Test Parameter

Test Detail
Test AMY
Report Name Amylase
Unit U/L
Wavelenth-Primary 415
Assay Type RATE-A
M1 Start 0
M2 Start 7
Sample Replicates 1
Control Replicate 1
Reaction Direction increasing
Prozone Limit % 0
Linearity Limit % 0
Technical Minimum 0.0
Y=aX+b A= 1
Reagent Abs Min 0.0

Test Volume
Test AMY
Sample Type Serum

Sample Volume

Normal 4 Dilution Ratio


Increase 4 Dilution Ratio
Decrease 4 Dilution Ratio

Standard Volume 4

Reagent Volumes and Stirrer Speed

RGT-1 Volume 200 R1 Stirrer Speed

Page 168
AMY

Reference Range

Test AMY
Sample Type Serum

Reference Range Default

Category Default

Reference Range – Decimal Places :2

Lower Limit Upper Limit

Normal 0 80

Panic 0 0

Calibration

Test AMY Curve Type

Calibration Expiry
Last Calibration Date Automatic Display Limit

Consumable: Reagent

Page 169
AMY

Consumable Type Reagent

Consumable AMY

Lot Detail

Manufacturer ERBA
Lot No as per current LOT
Expiry Date After 3 years Lot Status

Onboard Stability 0 Unit

Reagent Type R1 √

Utility
Position Reagent Reagent TypeBottle Type
18 AMY R1 SMALL

Page 170
AMY

m1s
Host Name AMY Auto Rerun √ M1E
Online Calibration M2S
Decimal Places 0 M2E
Secondary SELECT Total Reagent 1
Curve Type K-Factor Reagent R1 AMY R1
M1 End 0
M2 End 14
Stadard Replicate 3
Control Interval 0
Reaction ABS limit 1.7
Prozone Check lower
Delta Abs/Min 0.0
Technical Maximum 0
b= 0
Reagent Abs Max 0.4

Sample Type

1 √ Serum
1 Urine
10 CSF
Plasma
Whole Blood
Other

ed

HIGH

Page 171
AMY

Sample Type

√ Serum
Urine
CSF
Plasma
Whole Blood
Other

K Factor

0 Acceptable Limit % 0

Calibration Date: Automatic Display K Factor 3625

R1 Lot No

Page 172
AMY

Stock Over

Active (Automatic)

Hours

Lot No
as per current LOT

Page 173
AMY

0
lag period
lag period after R1 Xl 1000 cycle
after R1 (in XL1000 (in N in
second) cycle No) instrument
1 0
1
63 63 3.5 3
126 126 7 6

63 54

Page 174
ALP

Test Parameter

Test Detail
Test ALP
Report Name Alkaline phophatase
Unit IU/L
Wavelenth-Primary 415
Assay Type RATE-A
M1 Start 0
M2 Start 30
Sample Replicates 1
Control Replicate 1
Reaction Direction increasing
Prozone Limit % 0
Linearity Limit % 20
Technical Minimum 0.0
Y=aX+b A= 1
Reagent Abs Min 0.0

Test Volume
Test ALP
Sample Type Serum

Sample Volume

Normal 4 Dilution Ratio


Increase 4 Dilution Ratio
Decrease 4 Dilution Ratio

Standard Volume 4

Reagent Volumes and Stirrer Speed

RGT-1 Volume 160 R1 Stirrer Speed


RGT-2 Volume 40 R2 Stirrer Speed

Page 175
ALP

Reference Range

Test ALP
Sample Type Serum

Reference Range Default

Category Default

Reference Range – Decimal Places :2

Lower Limit Upper Limit

Normal 38 94

Panic 0 0

Calibration

Test ALP Curve Type

Calibration Expiry
Last Calibration Date Automatic Display Limit

Consumable: Reagent

Page 176
ALP

Consumable Type Reagent

Consumable ALP

Lot Detail

Manufacturer ERBA
Lot No as per current LOT
Expiry Date After 3 years Lot Status

Onboard Stability 0 Unit

Reagent Type R1 √
R1 √

Utility
Position Reagent Reagent TypeBottle Type
6 ALP R1 SMALL
8 ALP R2 SMALL

Page 177
ALP

m1s
Host Name ALP Auto Rerun √ M1E
Online Calibration M2S
Decimal Places 0 M2E
Secondary SELECT Total Reagent 2
Curve Type K-Factor Reagent R1 ALP R1
M1 End 0 Reagent R2 ALP R2
M2 End 37
Stadard Replicate 3
Control Interval 0
Reaction ABS limit 1.5
Prozone Check lower
Delta Abs/Min 0.0
Technical Maximum 0
b= 0
Reagent Abs Max 0.0

Sample Type

1 √ Serum
1 Urine
10 CSF
Plasma
Whole Blood
Other

ed

Medium
HIGH

Page 178
ALP

Sample Type

√ Serum
Urine
CSF
Plasma
Whole Blood
Other

K Factor

0 Acceptable Limit % 0

Calibration Date: Automatic Display K Factor 2883

R1 Lot No

Page 179
ALP

Stock Over

Active (Automatic)

Hours

Lot No
as per current LOT
as per current LOT

Page 180
ALP

0
lag period
lag period after R1 Xl 1000 cycle
after R1 (in XL1000 (in N in
second) cycle No) instrument
1 0
1
270 270 15 15
333 333 18.5 18

63 54

Page 181
LDH

Test Parameter

Test Detail
XL 1000
Test LDH LDHH
Report Name Lactate Dehydrogenase
Unit IU/L
Wavelenth-Primary 340
Assay Type RATE-A
M1 Start 0
M2 Start 7 4
Sample Replicates 1
Control Replicate 1
Reaction Direction increasing
Prozone Limit % 0
Linearity Limit % 0
Technical Minimum 0.0
Y=aX+b A= 1
Reagent Abs Min 0.0

Test Volume
Test LDH
Sample Type Serum

Sample Volume

Normal 10 Dilution Ratio


Increase 10 Dilution Ratio
Decrease 10 Dilution Ratio

Standard Volume 10

Reagent Volumes and Stirrer Speed

RGT-1 Volume 200 R1 Stirrer Speed

Page 182
LDH

Reference Range

Test LDH
Sample Type Serum

Reference Range Default

Category Default

Reference Range – Decimal Places :2

Lower Limit Upper Limit

Normal 225 450

Panic 0 0

Calibration

Test LDH Curve Type

Calibration Expiry
Last Calibration Date Automatic Display Limit

Consumable: Reagent

Page 183
LDH

Consumable Type Reagent

Consumable LDH

Lot Detail

Manufacturer ERBA
Lot No as per current LOT
Expiry Date After 3 years Lot Status

Onboard Stability 0 Unit

Reagent Type R1 √

Utility
Position Reagent Reagent TypeBottle Type
20 LDH R1 Small

Page 184
LDH

XL 1000 m1s
Host Name LDH LDHH Auto Rerun √ M1E
Online Calibration M2S
Decimal Places 0 M2E
Secondary SELECT Total Reagent 1
Curve Type K-Factor Reagent R1 LDH R1
M1 End 0
M2 End 14 7
Stadard Replicate 3
Control Interval 0
Reaction ABS limit 1.8
Prozone Check lower
Delta Abs/Min 0.0
Technical Maximum 1800
b= 0
Reagent Abs Max 0.0

Sample Type

1 √ Serum
1 Urine
10 CSF
Plasma
Whole Blood
Other

ed

Medium

Page 185
LDH

Sample Type

√ Serum
Urine
CSF
Plasma
Whole Blood
Other

K Factor

0 Acceptable Limit % 0

Calibration Date: Automatic Display K Factor 4774

R1 Lot No

Page 186
LDH

Stock Over

Active (Automatic)

Hours

Lot No
as per current LOT

Page 187
LDH

0
lag period
lag period after R1 Xl 1000 cycle
after R1 (in XL1000 (in N in
second) cycle No) instrument
1 0
1
63 63 3.5 3
126 126 7 6

63 54

Page 188
K

Test Parameter
Test Detail
Test K
Report Name Potassium
Unit mmol/L
Wavelenth-Primary -
Assay Type -
M1 Start 0
M2 Start 0
Sample Replicates 1
Control Replicate 1
Reaction Direction -
Prozone Limit %
Linearity Limit %
Technical Minimum
Y=aX+b A= 1.041
Reagent Abs Min

Test Volume
Test K do change in urine also
Sample Type Serum

Sample Volume

Normal 70 Dilution Ratio


Increase 70 Dilution Ratio
Decrease 70 Dilution Ratio

Standard Volume

Reagent Volumes and Stirrer Speed

RGT-1 Volume R1 Stirrer Speed


RGT-2 Volume R2 Stirrer Speed

Page 189
K

Reference Range

Test K
Sample Type Serum

Reference Range Default

Category Default

Reference Range – Decimal Places :2

Lower Limit Upper Limit

Normal 3.5 5

Panic 0 0

Calibration

Test Curve Type

Last Calibration Date Calibration Expiry Limit

Consumable: Reagent

Consumable Type

Page 190
K

Consumable

Lot Detail

Manufacturer
Lot No
Expiry Date Lot Status

Onboard Stability Unit

Reagent Type

Utility
Position Reagent Reagent TypeBottle Type

Page 191
K

Host Name K Auto Rerun


Online Calibration
Decimal Places 2
Secondary - Total Reagent -
Curve Type -
M1 End 0
M2 End 0
Stadard Replicate
Control Interval 0
Reaction ABS limit
Prozone Check
Delta Abs/Min
Technical Maximum
b= 0
Reagent Abs Max

Test Volume
Test
Sample Type

Sample Type Sa

1 √ Serum Normal
1 √ Urine Increase
1 CSF Decrease
Plasma
Whole Blood Standard Volume
Other

ed Reagent Volu

Low RGT-1 Volume


High RGT-2 Volume

Page 192
K

e Refere

Test
Sample Type

Reference Ra

Category

Sample Type Reference Range – D

√ Serum
√ Urine
CSF
Plasma Normal
Whole Blood
Other Panic

ration Expiry Limit Acceptable Limit %

Calibration Date: K Factor

R1 Lot No
R2 Lot No

Page 193
K

Stock Over

Lot No

Page 194
K

Test Volume
K
urine

Sample Volume Sample Type

140 Dilution Ratio 10 √ Serum


140 Dilution Ratio 3 √ Urine
140 Dilution Ratio 15 CSF
Plasma
Standard Volume Whole Blood
Other

Reagent Volumes and Stirrer Speed

RGT-1 Volume R1 Stirrer SpeLow


RGT-2 Volume R2 Stirrer SpeHigh

Page 195
K

Reference Range

K
urine

Default

Default

Reference Range – Decimal Places :2 Sample Type

Lower Limit Upper Limit √ Serum


√ Urine
CSF
25 120 Plasma
Whole Blood
0 0 Other

Page 196
Na

Test Parameter
Test Detail
Test Na
Report Name Sodium
Unit mmol/L
Wavelenth-Primary -
Assay Type -
M1 Start 0
M2 Start 0
Sample Replicates 1
Control Replicate 1
Reaction Direction -
Prozone Limit %
Linearity Limit %
Technical Minimum
Y=aX+b A= 0.995
Reagent Abs Min

Test Volume do change in urine also

Test Na
Sample Type Serum

Sample Volume

Normal 70 Dilution Ratio


Increase 70 Dilution Ratio
Decrease 70 Dilution Ratio

Standard Volume

Reagent Volumes and Stirrer Speed

RGT-1 Volume R1 Stirrer Speed


RGT-2 Volume R2 Stirrer Speed

Page 197
Na

Reference Range

Test Na do change in urine also


Sample Type Serum

Reference Range Default

Category Default

Reference Range – Decimal Places :2

Lower Limit Upper Limit

Normal 136 7

Panic 0 0

Calibration

Test Curve Type

Last Calibration Date Calibration Expiry Limit

Consumable: Reagent

Consumable Type

Page 198
Na

Consumable

Lot Detail

Manufacturer
Lot No
Expiry Date Lot Status

Onboard Stability Unit

Reagent Type

Utility
Position Reagent Reagent TypeBottle Type

Page 199
Na

Host Name Na Auto Rerun


Online Calibration
Decimal Places 2
Secondary - Total Reagent -
Curve Type -
M1 End 0
M2 End 0
Stadard Replicate
Control Interval 0
Reaction ABS limit
Prozone Check
Delta Abs/Min
Technical Maximum
b= 0
Reagent Abs Max

Test Volume
Test
Sample Type

Sample Type Sa

1 √ Serum Normal
1 √ Urine Increase
1 CSF Decrease
Plasma
Whole Blood Standard Volume
Other

ed Reagent Volu

Low RGT-1 Volume


High RGT-2 Volume

Page 200
Na

e Refere

Test
Sample Type

Reference Ra

Category

Sample Type Reference Range – D

√ Serum
√ Urine
CSF
Plasma Normal
Whole Blood
Other Panic

ration Expiry Limit Acceptable Limit %

Calibration Date: K Factor

R1 Lot No
R2 Lot No

Page 201
Na

Stock Over

Lot No

Page 202
Na

Test Volume
Na
urine

Sample Volume Sample Type

140 Dilution Ratio 10 √ Serum


140 Dilution Ratio 3 √ Urine
140 Dilution Ratio 15 CSF
Plasma
Standard Volume Whole Blood
Other

Reagent Volumes and Stirrer Speed

RGT-1 Volume R1 Stirrer SpeLow


RGT-2 Volume R2 Stirrer SpeHigh

Page 203
Na

Reference Range

Na
urine

Default

Default

Reference Range – Decimal Places :2 Sample Type

Lower Limit Upper Limit √ Serum


√ Urine
CSF
40 220 Plasma
Whole Blood
0 0 Other

Page 204
Sheet1_2

Test Parameter
Test Detail
Test
Report Name
Unit
Wavelenth-Primary
Assay Type
M1 Start
M2 Start
Sample Replicates
Control Replicate
Reaction Direction
Prozone Limit %
Linearity Limit %
Technical Minimum
Y=aX+b a=
Reagent Abs Min

Test Volume
Test
Sample Type

Sample Volume

Normal Dilution Ratio


Increase Dilution Ratio
Decrease Dilution Ratio

Standard Volume

Reagent Volumes and Stirrer Speed

RGT-1 Volume R1 Stirrer Speed


RGT-2 Volume R2 Stirrer Speed

Page 205
Sheet1_2

Reference Range

Test
Sample Type

Reference Range

Category

Reference Range – Decimal Places :2

Lower Limit Upper Limit

Normal

Panic

Calibration

Test Curve Type

Last Calibration Date Calibration Expiry Limit

Consumable: Reagent

Consumable Type

Page 206
Sheet1_2

Consumable

Lot Detail

Manufacturer
Lot No
Expiry Date Lot Status

Onboard Stability Unit

Reagent Type

Utility
Position Reagent Reagent TypeBottle Type

Page 207
Sheet1_2

Host Name Auto Rerun


Online Calibration
Decimal Places
Secondary Total Reagent
Curve Type Reagent R1
M1 End Reagent R2
M2 End
Stadard Replicate
Control Interval
Reaction ABS limit
Prozone Check
Delta Abs/Min
Technical Maximum
b=
Reagent Abs Max

Sample Type

Serum
Urine
CSF
Plasma
Whole Blood
Other

ed

Page 208
Sheet1_2

Sample Type

Serum
Urine
CSF
Plasma
Whole Blood
Other

ration Expiry Limit Acceptable Limit %

Calibration Date: K Factor

R1 Lot No
R2 Lot No

Page 209
Sheet1_2

Stock Over

Lot No

Page 210
Sheet1

Test Parameter

Test Host Name


Report Name
Unit Decimal Places
Wavelenth-Primary Secondary
Assay Type Curve Type
M1 Start M1 End
M2 Start M2 End
Sample Replicates Stadard Replicate
Control Replicate Control Interval
Reaction Direction Reaction ABS limit
Prozone Limit % Prozone Check
Linearity Limit % Delta Abs/Min
Technical Minimum Technical Maximum
Y=aX+b a= b=
Reagent Abs Min Reagent Abs Max

Test Volume

Test
Sample Type

Sample Volume

Normal Dilution Ratio


Increase Dilution Ratio
Decrease Dilution Ratio

Standard Volume

Reagent Volumes and Stirrer Speed

RGT-1 Volume R1 Stirrer Speed


RGT-2 Volume R2 Stirrer Speed

Reference Range
Page 211
Sheet1

Test
Sample Type

Reference Range

Category

Reference Range – Decimal Places :2

Lower Limit Upper Limit

Normal

Panic

Calibration

Test Curve Type

Last Calibration Date Calibration Expiry Limit

Consumable: Reagent

Consumable Type

Consumable

Page 212
Sheet1

Lot Detail

Manufacturer Stock Over


Lot No
Expiry Date Lot Status

Onboard Stability Unit

Reagent Type

Utility

Position Reagent Reagent TypeBottle Type Lot No

Page 213
Sheet1

Auto Rerun
Online Calibration

Total Reagent
Reagent R1
Reagent R2

Sample Type

Serum
Urine
CSF
Plasma
Whole Blood
Other

Page 214
Sheet1

Sample Type

Serum
Urine
CSF
Plasma
Whole Blood
Other

Acceptable Limit %

Calibration Date: K Factor

R1 Lot No

Page 215
Explanation

Test Parameter
Test Detail
Test LDH
Report Name Lactate Dehydrogenase
Unit IU/L
Wavelenth-Primary 340
Assay Type RATE-A
M1 Start 0
M2 Start 7
Sample Replicates 1
Control Replicate 1
Reaction Direction increasing
Prozone Limit % 0
Linearity Limit % 0
Technical Minimum 0.0
Y=aX+b A= 1
Reagent Abs Min 0.0

Test Volume
Test LDH
Sample Type Serum

Sample Volume

Normal 10 Dilution Ratio


Increase 10 Dilution Ratio
Decrease 10 Dilution Ratio

Standard Volume 10

Reagent Volumes and Stirrer Speed

RGT-1 Volume 200 R1 Stirrer Speed

Page 216
Explanation

Reference Range

Test LDH
Sample Type Serum

Reference Range Default

Category Default

Reference Range – Decimal Places :2

Lower Limit Upper Limit

Normal 225 450

Panic 0 0

Calibration

Test LDH Curve Type

Calibration Expiry
Last Calibration Date Automatic Display Limit

Consumable: Reagent

Page 217
Explanation

Consumable Type Reagent

Consumable LDH

Lot Detail

Manufacturer ERBA
Lot No as per current LOT
Expiry Date After 3 years Lot Status

Onboard Stability 0 Unit

Reagent Type R1 √

Utility
Position Reagent Reagent TypeBottle Type
20 LDH R1 Small

Page 218
Explanation

Host Name LDH Auto Rerun √


Online Calibration
Decimal Places 0
Secondary SELECT Total Reagent 1
Curve Type K-Factor Reagent R1 LDH R1
M1 End 0
M2 End 14
Stadard Replicate 3
Control Interval 0
Reaction ABS limit 1.8 32 1.224
Prozone Check lower -68.724 52 1.116
Delta Abs/Min 0.0 63 1.0992
Technical Maximum 1800
b= 0
Reagent Abs Max 0.0
20

Sample Type

1 √ Serum
1 Urine
10 CSF
Plasma
Whole Blood
Other

ed

Medium

Page 219
Explanation

Sample Type

√ Serum
Urine
CSF
Plasma
Whole Blood
Other

K Factor

0 Acceptable Limit % 0

Calibration Date: Automatic Display K Factor 4774

R1 Lot No

Page 220
Explanation

Stock Over

Active (Automatic)

Hours

Lot No
as per current LOT

Page 221
Explanation

0.9

-0.108 -0.036

1282
1058

5294
5362

Page 222