CHOLESTEROL LDL

DIRECT

The application parameters comprised here constitute a guide to facilitate the validation of our reagents by
the instrument. It is advisable to validate the use when there is any change in software or reagent versions.

INSTRUMENT: BIOBASE BK200 MINI

REAGENT PREPARATION
Reagents are provided ready to use.

ITEM PARAMETERS

LDL
Item Nº * Name CHOLESTEROL
Item Name LDL Sample volum (uL) 3
Min Absorbancy 0 R1 volume (uL) 300
Max Absorbancy 3 R2 Volume (uL) 100
Normal Low value <100 Sample type SER
Normal High Value >100 Units mg/dL
Lower Value of Test 0.4
Highest Value of
tests 990 Prozone check *
Sample Blank yes
Pri Wavelenght 546
Sub Wavelenght Pri-Point
Start Point 34
Decimals Places 1 End Point 35

Analysis Method Two point end Sub-Point


Start Point 12
End point 13

Linearity
K 1
b 0

* Use this factors to use as a compensated method. If not b=0.


CHOLESTEROL LDL
DIRECT

CALIBRATION

Item Name LDL Repeat Counts 3


Two point
Calibration Type linear Sample Type SER
Standard Counts 2 Factor

Standard Absorbance Concentration Cup nº


0 * 0.0 *
1 * * *

* User defined

Version 03/21