You are on page 1of 2

Global Systems Support r

cobas c 111
Service Bulletin No. 018

Subject: Update to 3.2L Water and Waste Containers


Date: 04-Dec-2007 SB Version: 1

Information:

Author: Walter Oehen

Approver: Ralph Harvey

Manufacturer: ROCHE

Modification X Hardware Software X Spare Parts Documentation


Attachment(s):

References:

Urgency
At next opportunity.

Units Concerned
Optional update.
This update may concern cobas c 111 instruments having the following serial numbers:
< 50483 (with ISE)
< 1591 (without ISE)
All cobas c 111 instruments having a higher serial number are already factory equipped with the larger
3.2L containers and do not require this update.

Summary
Requests were received from different affiliates asking for a larger type of water- and waste container
for the cobas c 111.
Such larger containers are now available. A fluid update kit has been prepared which includes all parts
needed to perform this retrofit.

Action
Install the fluid update kit where required. Proceed as follows:
1. Disconnect the water suction tube from the instrument. Remove the water suction assy from the
water container and put it aside.
2. Remove and empty the currently installed water and waste containers and put them aside.
3. Fill and install the new water container. Install the new, associated suction assembly.
Connect the water suction tube at the instrument
4. Install the new waste container. The existing waste discharge assembly must be reused.
Store the spare waste container at a suitable place.
5. Remove the cleaner suction assy from the cleaner bottle. Unscrew the black nipple at the end of the
suction pipe and install the new, grey nipple instead. This grey nipple is part of the update kit.
Reinstall the cleaner suction pipe
6. Go to <Utilities> <Maintenance> and select <Prime Fluid System>. Perform the prime function
twice.

Page: 1/2
Content of the Fluid Update Kit Replacing the Nipple of the Cleaner Suction Pipe

Reasons of Change
The new, larger containers minimize the number of operator interventions regarding refilling /
emptying the external fluid containers.

Details
The new, larger containers fit into the existing fluid rack. Compared to the currently used 1.5 liter
containers the new types offer now a filling capacity of 3.2 liters. The belonging water suction assembly
includes also a modified type of holder water inlet filter which allows a convenient replacement of the
water filter.
The currently used black nipple installed at the end of the cleaner suction pipe should be replaced by
the grey nipple. The reason is that the grey nipple is more resistant against cleaner compared to the
black one.

Parts Required per Instrument

Qty. Part Number Description


1 05127696001 Update Kit Fluid Supply 3.2L

Beside the Update Kit Fluid Supply 3.2L the new type of containers, the new type Suction Assy Water
Container 3.2L and the new type Fitting 1/4“ Cleaner Suction Tube are also available as discrete parts.
Part numbers and descriptions are as follows:

Part Number Description


05127700001 Container 3.2L Water
05127718001 Container 3.2L Waste
05127726001 Suction Assy Water Cont. 3.2L
05136474001 Fitting 1/4“ Cleaner Suction Tube

Note: The smaller 1.5L water and waste container will be discontinued as spare parts.

Order Information