You are on page 1of 2

ALBUMIN

Albúmina
Verde bromocresol. Colorimétrico

PREPARACIÓN Y ESTABILIDAD BX_400 / TYB_40 / RX IMMOLA


Ref. 1001020, 1001022 y 1001023. El reactivo está listo
para su uso. Es estable hasta la fecha de caducidad indicada en el envase. Proteger de la luz.

CALIBRACIÓN
Utilizar un suero de calibración adecuado (CFAS Ref. 1002011).

CONDICIONES OPERATORIAS
Introducir en la memoria del ordenador los siguientes parámetros:

CHEMISTRY PARAMETER

Method * Name ALB Reagent Name Reagent (µL) Water (µL)


Print Name ALBUMIN R1 ALBUMIN 350
Sample SERUM R2 Enable
Unit g/dL
Assay Type END POINT Diluent Enable

Mesuring Points Decimal Points 1


Start End
1 79 80 Normal Range
Enable 2 10 11 Nº Range Name Min. Max.
1 * * *
Wave Lenght 2 * * *
Prim. 605 Sec x Disable 3 * * *
4 * * *
Normal
Sampling Sample (µL) Diluent (µL) LOW HIGH
Dilution 2 Technical (Conc.) * *
Rerun (High Prozone) Range (mAbs/10) -999999 999999
Dilution
Rerun (Low) SPT Wash Enable Reag. Name
Dilution Stirring Speed R1-Middle R2-Middle

Limit Checks Blank measurement


Duplicate Limit mAbs/10
Sensitivity Limit mAbs/10 Blank measurement Enable Reagent Bk for S1 (linear)
Linearity Limit % (mAbs/10)/min Meas. of reagent BK during run Next Run
Prozone Limit Reag. BK. meas. at calibration Reagent BK (system water)
SL1-S SL1-F Nº of measurement Duplicate
SL2-S SL2-F
SENS mAbs/10 Reagent Blank Limit Checks:
Duplicate Limit mAbs/10
x Absorbance limit
Reaction Increase Instrument Factor
Limit * mAbs/10 A: 1.00 B: 0.00

Calibration: 2 linear point

REV.01 Ed. 2023/07

SPINREACT, S.A.U Ctra. Santa Coloma, 7 E-17176 SANT ESTEVE DE BAS (GI) SPAIN
Tel. +34 972 69 08 00 Fax +34 972 69 00 99 e-mail: spinreact@spinreact.com
ALBUMIN

Albumin
Bromocresol Green. Colorimetric

PREPARATION AND STABILITY BX_400 / TYB_40 / RX IMMOLA


Ref. 1001020, 1001022 and 1001023. Reagent is
ready to use. Stable up to the expiration date stated on the label.

CALIBRATION
Use a human serum calibrator (CFAS Ref. 1002011).

OPERATORY CONDITIONS
Introduce the following data in the computer:

CHEMISTRY PARAMETER

Method * Name ALB Reagent Name Reagent (µL) Water (µL)


Print Name ALBUMIN R1 ALBUMIN 350
Sample SERUM R2 Enable
Unit g/dL
Assay Type END POINT Diluent Enable

Mesuring Points Decimal Points 1


Start End
1 79 80 Normal Range
Enable 2 10 11 Nº Range Name Min. Max.
1 * * *
Wave Lenght 2 * * *
Prim. 605 Sec x Disable 3 * * *
4 * * *
Normal
Sampling Sample (µL) Diluent (µL) LOW HIGH
Dilution 2 Technical (Conc.) * *
Rerun (High Prozone) Range (mAbs/10) -999999 999999
Dilution
Rerun (Low) SPT Wash Enable Reag. Name
Dilution Stirring Speed R1-Middle R2-Middle

Limit Checks Blank measurement


Duplicate Limit mAbs/10
Sensitivity Limit mAbs/10 Blank measurement Enable Reagent Bk for S1 (linear)
Linearity Limit % (mAbs/10)/min Meas. of reagent BK during run Next Run
Prozone Limit Reag. BK. meas. at calibration Reagent BK (system water)
SL1-S SL1-F Nº of measurement Duplicate
SL2-S SL2-F
SENS mAbs/10 Reagent Blank Limit Checks:
Duplicate Limit mAbs/10
x Absorbance limit
Reaction Increase Instrument Factor
Limit * mAbs/10 A: 1.00 B: 0.00

Calibration: 2 linear point

REV.01 Ed. 2023/07

SPINREACT, S.A.U. Ctra.Santa Coloma, 7 E-17176 SANT ESTEVE DE BAS (GI) SPAIN
Tel. +34 972 69 08 00 Fax +34 972 69 00 99 e-mail: spinreact@spinreact.com