Professional Documents
Culture Documents
Albúmina
Verde bromocresol. Colorimétrico
CALIBRACIÓN
Utilizar un suero de calibración adecuado (CFAS Ref. 1002011)
CONDICIONES OPERATORIAS
Introducir en la memoria del ordenador los siguientes parámetros:
1.GENERAL:
REACTTION TYPE INCREASE INCREASE
SAMPLE TYPE: SER (SUERUM) SER (SUERUM)
UNITS: mg/dL mg/dL
# DE REPLICTES: 1 1
# DECIMALS: 1 1
ASSAY METHOD: END POINT END POINT
2.PROCEDURE:
BOCHROMATIC READING
1 ST WAVELENGTH: 635 635
2 ND WAVELENGTH:
3.VOLUMES:
SAMPLE VOL: 2 ul 2 ul
REAG1 VOL: 350 ul 350 ul
REAG2 VOL:
CLEAN: 1.2 1.2
3.TIME:
READING 1: 192 SEC / C9 195 SEC / C14
READING 2:
REAGENT 2:
4.CALIBRATION.
CALIBRATOR. SPECIFIC SPECIFIC
CAL NUMB: 1 1
CAL REPLICATES: 3 3
BLK REPLICATES: 3 3
MULTIPOINT CAL:
5.CONTROLS:
# CONTRLS: 1 1
REJECTION CRITERIA 3 DS 3 DS
CONTROL TYPE: MULTIPLE. MULTIPLE
CALCUL METHOD: MANUAL MANUAL
#BATCH:
BATCH:
NAME:
LIMITS:
6.OPTIONS
BLK ABS. LIMIT YES YES
VALUE:
Ed. 2018
SPINREACT, S.A.U Ctra. Santa Coloma, 7 E-17176 SANT ESTEVE DE BAS (GI) SPAIN
Tel. +34 972 69 08 00 Fax +34 972 69 00 99 e-mail: spinreact@spinreact.com
ALBUMIN
Albumin
Bromcresol Green. Colorimetric
CALIBRATION
Use a human serum calibrator (CFAS Ref. 1002011)
OPERATORY CONDITIONS
Introduce the following data in the computer:
1.GENERAL:
REACTTION TYPE INCREASE INCREASE
SAMPLE TYPE: SER (SUERUM) SER (SUERUM)
UNITS: mg/dL mg/dL
# DE REPLICTES: 1 1
# DECIMALS: 1 1
ASSAY METHOD: END POINT END POINT
2.PROCEDURE:
BOCHROMATIC READING
1 ST WAVELENGTH: 635 635
2 ND WAVELENGTH:
3.VOLUMES:
SAMPLE VOL: 2 ul 2 ul
REAG1 VOL: 350 ul 350 ul
REAG2 VOL:
CLEAN: 1.2 1.2
3.TIME:
READING 1: 192 SEC / C9 195 SEC / C14
READING 2:
REAGENT 2:
4.CALIBRATION.
CALIBRATOR. SPECIFIC SPECIFIC
CAL NUMB: 1 1
CAL REPLICATES: 3 3
BLK REPLICATES: 3 3
MULTIPOINT CAL:
5.CONTROLS:
# CONTRLS: 1 1
REJECTION CRITERIA 3 DS 3 DS
CONTROL TYPE: MULTIPLE. MULTIPLE
CALCUL METHOD: MANUAL MANUAL
#BATCH:
BATCH:
NAME:
LIMITS:
6.OPTIONS
BLK ABS. LIMIT YES YES
VALUE:
Ed. 2018