AMYLASE

-Amilasa
CNP-G3. Cinético. Líquido

PREPARACIÓN Y ESTABILIDAD A_15 / A_25


Ref. 41201 y 41202. Los reactivos se encuentran listos para su uso.
Son estables hasta la fecha de caducidad indicada en el envase conservado a 2-8ºC.

CALIBRACIÓN
Utilizar un suero de calibración adecuado (CFAS Ref. 1002011).

CONDICIONES OPERATORIAS
Introducir en la memoria del ordenador los siguientes parámetros:

AMYLASE A_15 A_25

1.GENERAL:
REACTTION TYPE INCREASE INCREASE
SAMPLE TYPE: SER (SUERUM) SER (SUERUM)
UNITS: mg/dL mg/dL
# DE REPLICTES: 1 1
# DECIMALS: 1 1
ASSAY METHOD: KINETIC KINETIC

2.PROCEDURE:
BOCHROMATIC READING
1 ST WAVELENGTH: 405 405
2 ND WAVELENGTH:

3.VOLUMES:
SAMPLE VOL: 6 ul 6 ul
REAG1 VOL: 300 ul 300 ul
REAG2 VOL:
CLEAN: 1.2 1.2

3.TIME:
READING 1: 72 SEC / C4 75 SEC / C6
READING 2: 216 SEC / C10 210 SEC / C15
REAGENT 2:

4.CALIBRATION.
CALIBRATOR. SPECIFIC SPECIFIC
CAL NUMB: 1 1
CAL REPLICATES: 3 3
BLK REPLICATES: 3 3
MULTIPOINT CAL:

5.CONTROLS:
# CONTRLS: 1 1
REJECTION CRITERIA 3 DS 3 DS
CONTROL TYPE: MULTIPLE. MULTIPLE
CALCUL METHOD: MANUAL MANUAL
#BATCH:
BATCH:
NAME:
LIMITS:

6.OPTIONS
BLK ABS. LIMIT YES YES
VALUE:

Ed. 2018

SPINREACT, S.A.U. Ctra. Santa Coloma, 7 E-17176 SANT ESTEVE DE BAS (GI) SPAIN
Tel. +34 972 69 08 00 Fax +34 972 69 00 99 e-mail: spinreact@spinreact.com
AMYLASE

-Amylase
CNP-G3. Kinetic. Liquid

PREPARATION AND STABILITY A_15 / A_25


Ref. 41201 and 41202. Reagent ready to use.
Unopened reagent is stable until the expiration date printed on the label when stored at 2-8ºC.

CALIBRATION
Use a human serum calibrator (CFAS Ref. 1002011).

OPERATORY CONDITIONS
Introduce the following data in the computer:

AMYLASE A_15 A_25

1.GENERAL:
REACTTION TYPE INCREASE INCREASE
SAMPLE TYPE: SER (SUERUM) SER (SUERUM)
UNITS: mg/dL mg/dL
# DE REPLICTES: 1 1
# DECIMALS: 1 1
ASSAY METHOD: KINETIC KINETIC

2.PROCEDURE:
BOCHROMATIC READING
1 ST WAVELENGTH: 405 405
2 ND WAVELENGTH:

3.VOLUMES:
SAMPLE VOL: 6 ul 6 ul
REAG1 VOL: 300 ul 300 ul
REAG2 VOL:
CLEAN: 1.2 1.2

3.TIME:
READING 1: 72 SEC / C4 75 SEC / C6
READING 2: 216 SEC / C10 210 SEC / C15
REAGENT 2:

4.CALIBRATION.
CALIBRATOR. SPECIFIC SPECIFIC
CAL NUMB: 1 1
CAL REPLICATES: 3 3
BLK REPLICATES: 3 3
MULTIPOINT CAL:

5.CONTROLS:
# CONTRLS: 1 1
REJECTION CRITERIA 3 DS 3 DS
CONTROL TYPE: MULTIPLE. MULTIPLE
CALCUL METHOD: MANUAL MANUAL
#BATCH:
BATCH:
NAME:
LIMITS:

6.OPTIONS
BLK ABS. LIMIT YES YES
VALUE:

Ed. 2018

SPINREACT,S.A.U. Ctra.Santa Coloma, 7 E-17176 SANT ESTEVE DE BAS (GI) SPAIN


Tel. +34 972 69 08 00 Fax +34 972 69 00 99 e-mail: spinreact@spinreact.com