Professional Documents
Culture Documents
-Amilasa
CNP-G3. Cinético. Líquido
CALIBRACIÓN
Utilizar un suero de calibración adecuado (CFAS Ref. 1002011).
CONDICIONES OPERATORIAS
Introducir en la memoria del ordenador los siguientes parámetros:
1.GENERAL:
REACTTION TYPE INCREASE INCREASE
SAMPLE TYPE: SER (SUERUM) SER (SUERUM)
UNITS: mg/dL mg/dL
# DE REPLICTES: 1 1
# DECIMALS: 1 1
ASSAY METHOD: KINETIC KINETIC
2.PROCEDURE:
BOCHROMATIC READING
1 ST WAVELENGTH: 405 405
2 ND WAVELENGTH:
3.VOLUMES:
SAMPLE VOL: 6 ul 6 ul
REAG1 VOL: 300 ul 300 ul
REAG2 VOL:
CLEAN: 1.2 1.2
3.TIME:
READING 1: 72 SEC / C4 75 SEC / C6
READING 2: 216 SEC / C10 210 SEC / C15
REAGENT 2:
4.CALIBRATION.
CALIBRATOR. SPECIFIC SPECIFIC
CAL NUMB: 1 1
CAL REPLICATES: 3 3
BLK REPLICATES: 3 3
MULTIPOINT CAL:
5.CONTROLS:
# CONTRLS: 1 1
REJECTION CRITERIA 3 DS 3 DS
CONTROL TYPE: MULTIPLE. MULTIPLE
CALCUL METHOD: MANUAL MANUAL
#BATCH:
BATCH:
NAME:
LIMITS:
6.OPTIONS
BLK ABS. LIMIT YES YES
VALUE:
Ed. 2018
SPINREACT, S.A.U. Ctra. Santa Coloma, 7 E-17176 SANT ESTEVE DE BAS (GI) SPAIN
Tel. +34 972 69 08 00 Fax +34 972 69 00 99 e-mail: spinreact@spinreact.com
AMYLASE
-Amylase
CNP-G3. Kinetic. Liquid
CALIBRATION
Use a human serum calibrator (CFAS Ref. 1002011).
OPERATORY CONDITIONS
Introduce the following data in the computer:
1.GENERAL:
REACTTION TYPE INCREASE INCREASE
SAMPLE TYPE: SER (SUERUM) SER (SUERUM)
UNITS: mg/dL mg/dL
# DE REPLICTES: 1 1
# DECIMALS: 1 1
ASSAY METHOD: KINETIC KINETIC
2.PROCEDURE:
BOCHROMATIC READING
1 ST WAVELENGTH: 405 405
2 ND WAVELENGTH:
3.VOLUMES:
SAMPLE VOL: 6 ul 6 ul
REAG1 VOL: 300 ul 300 ul
REAG2 VOL:
CLEAN: 1.2 1.2
3.TIME:
READING 1: 72 SEC / C4 75 SEC / C6
READING 2: 216 SEC / C10 210 SEC / C15
REAGENT 2:
4.CALIBRATION.
CALIBRATOR. SPECIFIC SPECIFIC
CAL NUMB: 1 1
CAL REPLICATES: 3 3
BLK REPLICATES: 3 3
MULTIPOINT CAL:
5.CONTROLS:
# CONTRLS: 1 1
REJECTION CRITERIA 3 DS 3 DS
CONTROL TYPE: MULTIPLE. MULTIPLE
CALCUL METHOD: MANUAL MANUAL
#BATCH:
BATCH:
NAME:
LIMITS:
6.OPTIONS
BLK ABS. LIMIT YES YES
VALUE:
Ed. 2018