2 x 40 mL, 2 x 60 mL
LDL-CHOLESTEROL (D) WITH CALIBRATOR 11415003, 11415004
This assay method uses a surfactant for selectively solubilizing LDL alone in Ascorbic acid up to : 50 mg/dL
the cholesterol assay system that employs cholesterol esterase and cholesterol Haemoglobin up to : 500 mg/dL
oxidase. It passes the ester cholesterol and free cholesterol contained in the Triglyceride up to : 3000 mg/dL
LDL to the cholesterol reaction system to determine LDL cholesterol. The en-
zyme reactions to other, non-LDL l ipoproteins (HDL, VLDL, chilomicrons) are *(when triglyceride in a sample exceeds 1000 mg/dL, dilute the sample 1+9 with
inhibited by the syrfactant and by the sugar compounds. These l ipoproteins saline, repeat the assay and multiply result by 10)
are therefore not passed to the cholesterol reaction system and consequently
remain in the reaction l iquid as lipoproteins. Based on this principle it is thus Materials provided
possible to d irectly determine LDL-cholesterol on its own. LDL Direct Reagent R1,R2 & Calibrator
IN VITRO DIAGNOSTIC USE SEE PACKAGE INSERT FOR PROCEDURE LOT NUMBER MANUFACTURER’S ADDRESS MANUFACTURING DATE EXPIRY DATE TEMPERATURE LIMIT
2 x 40 mL, 2 x 60 mL
LDL-CHOLESTEROL (D) WITH CALIBRATOR 11415003, 11415004
Procedure Notes
Laboratory Procedure for Ful ly Auto Analyzer Performance
1. Linearity
Blank Calibrator Sample/control
This reagent is linear upto 700 mg/dL
Reagent 1 270 µL 270 µL 270 µL
If the concentration is greater than linearity (700 mg/dL), dilute the sample with
Calibrator - 3 µL - normal saline and repeat the assay. Multiply the result with dilution factor
Sample / control - 3 µL 2. Comparison
0
Mix & incubate for 5 min at 37 C. Measure the absorbance (OD1) at 600 nm/ A comparison study has been performed between Agappe Reagent and another
700 nm. international ly available reagent yielded a correlation coefficient of r2 =0.9922 and
a regression equation of y =0.9592x.
Reagent 2 90 µL 90 µL 90 µL
0
Mix and incubate for 5 min at 37 C.Measure the absorbance (OD2) at 600 nm/ 3. Precision
700 nm.
Intra Run
Calculation
Control Level 1 Control Level 2
(OD2-OD1) Sample n 20 20
LDL-C Concentration (mg/dL) = ———————-—— x Calibrator Concentration
Mean (mg/dL) 108.7 50.5
(OD2-OD1) Calibrator
SD 2.11 1.19
Laboratory Procedure for Semi Auto Analyzer CV(%) 1.94 2.36