Professional Documents
Culture Documents
0
Test definition COCAINE Calibr. parameters
Test type Photometric Test in use YES Calibration type NONLINEAR
Full name Cocaine Low High Units Bias corr. in use NO
metabolite
Online name Test limit 0 1000 µg/l Repeat time (d) 1
Result Unit µg/l low( Ö ), high (!) Abs. error (mA) 5
Point/Calibrator Duplicate
Number of decimals 0 Initial absorbance 0 2 A Rel. error (%) 0
Acceptance MANUAL
Acceptance AUTOMATIC Response limit
(mA)
Type of calibrator SEPARATE Min. *
Dilution 1 + 0 Dilution limit (!) * * µg/l Calibrator Conc.(!) Dil. Ratio. 1+ Max. *
Sample type DoA Cal 0 0 0
Secondary dil. 1+ 0 0 DoA Cal B2 300 0
¨Serum ¨Plasma ýUrine DoA Cal B4 1000 0
¨C.S.F ¨Other Correction factor 1
More
Correction bias 0
Volume (µl) Disp. with Volume (µl) λ 2 (nm) 50 COCA1 Slow 409 30
55 WATER 55 NONE %
Konelab™ COCAINE METABOLITE - code: 981624 (2 x 30 ml) for cut off 300 µg/l
Reagent: Liquids, ready-to-use. It is recommended to store the reagents tightly sealed at 2...8 °C when not in use.
Calibrator: DoA Negative Cal., 981720, (DoA Calibrator B1, 981724), DoA Calibrator B2, 981725, (DoA Calibrator B3, 981726), DoA Calibrator B4, 981727.