No. MPDUS0085EAN
INTRODUCTION
This system achieves high sensitivity and high quality by
using state-of-the-art digital technology in the T/R section,
which is the core of a diagnostic ultrasound system.
This system features advanced algorithms that run on circuits
that incorporate the latest circuit technology, semiconductor
technology, and surface-mount technology.
This system is designed to support a full range of applications
and can be used as a general purpose system or a specialized
system, depending on the installed software.
CW kit UICW-AA000A Software and hardware which enables Steering CWD. Software license is required. OP. OP.
This kit is used to display reference signals (ECG waveforms etc.).
Reference Signal kit UJUR-AA000A OP. OP.
(UJUR-AI900A or UJUR-AI901A is required.)
For cardiovascular examinations (for regions other than the USA):
UJUR-AI900A ECG, Respiration, ECG gating, Heart rate. OP. OP.
(UJUR-AA000A is required.)
Reference Signal cable
For cardiovascular examinations (for the USA and Canada):
UJUR-AI901A ECG, Respiration, ECG gating, Heart rate. OP. OP.
(UJUR-AA000A is required.)
PCG and Pulse sensor.
Reference Signal Sensor unit UJUR-AI902A OP. OP.
(UJUR-AA000A is required.)
USSE-AI900A/7D, Adds the cardiac stress examination function to the system.
Stress Echo kit OP. OP.
USSE-AI900A/7L (UJUR-AA000A is required.)
USWT-AI900A, Adds the cardiac wall motion analysis function to the system.
2D Wall Motion Tracking kit OP. OP.
USWT-AI900A/EL (UJUR-AA000A is required.)
This kit enables cardiac wall motion analysis function for fetal heart using data from a convex
2D Wall Motion Tracking USWT-AI904A,
transducer. *1 OP.
Fetal kit USWT-AI904A/EL
(USWT-AI900A is required.)
USEF-AI600A, Software to enable Auto EF LV.
Auto EF Measurement kit OP. OP.
USEF-AI600A /EL (UJUR-AA000A is required.)
USWN-AA550A, This kit activates automatically and indicates the next procedure based on the ASE guideline
Workflow Navigator kit OP. OP.
USWN-AA550A/EL after completing the preceding step in the workflow examination in Cardiology.
Automatic measurements which supports the reduction in the variability of standard
measurements. Following functions are included in the Measurement
USQM-AA550A,
Measurement Assistant kit Assistant: Auto EF LA (Auto EF LA. requires USEF-AI600A.), Auto TR, Auto LVOT, Auto Ao. OP. OP.
USQM-AA550A/EL
Not available in the USA and Canada.
(UJUR-AA000A is required.)
11
Auto GLS kit
USWT-AI907A, Enables easy full LV assessment including bull's eye, Auto EF and GLS (Quick Strain).* OP. OP.
USWT-AI907A/EL (USWT-AI900A is required.)
USCT-AI900A, Enables the examiner to improve visibility of the cavities and thus the delineation of the
Contrast Enhance kit 11 OP. OP.
USCT-AI900A/EL myocardium (Clarity).*
*1: Advanced Software kit or Premium Software kit is required. *6: Premium Software kit is required.
*2: The Mounting kit for fusion sensor supporting the transducer *7: USMV-AI900A or Smart Sensor 3D kit is required.
used, Magnetic Generator kit UIFR-AA550A, and Sensor kit for *8: Option for WH model.
Fusion unit UIFR-A501A are required. *9: The black-and-white digital printer P95DW-DC cannot be installed
*3: Magnetic Generator kit UIFR-AA550A, Smart Navigation Sensor kit at the rear of the system if the Mounting kit for external HDD
UISN-A500A and Smart Fusion kit USFN-AA550A are required. UZHI-AI900A is installed.
*4: Some options are in addition to UIFR-AA550A. *10: Only available in the CE marking regions and Malaysia.
*5: Magnetic Generator kit UIFR-AA550A and Smart Fusion kit *11: Not available in the CE marking regions.
USFNAA550A are required.
2
MPDUS0085EAN
Main unit
Unit Model name Remarks CUS- WH
AA000 Model
This kit enables the automatic generation of standard fetal heart views from a
USFP-AI900A,
Smart Fetal Heart kit 4 chamber volume data set with mechanical 4D. Not available in the USA and Canada. *1 OP.
USFP-AI900A/EL
(USMV-AI900A is required)
USZS-AI900A,
Measurement Z score kit This kit enables Z-score analysis for the measurement results. OP. OP.
USZS-AI900A/EL
USSI-AA550A, This kit enables to detect fetal ultrasound screening section in real-time mainly based on ISUOG
Smart Area Indication OB kit OP. OP.
USSI-AA550A/EL guidelines.Not available in the USA and Canada.
Fetal Heart MPI USFH-AI600A, This kit enables Myocardial Performance Index (MPI).
*1 *1
Measurement kit USFH-AI600A/EL The MPI value can be calculated from the time change curve in TDI.
USHI-AA550A,
CHI kit Adds the Contrast Imaging function to the system. *1 *1
USHI-AA550A/EL
USCQ-AI900A, Adds the Time Curve Analysis (TCA) function to the system.
CHI-Q kit *6 *6
USCQ-AI900A/EL (USHI-AA550A and USPO-AA000A are required.)
USCQ-AI901A, Function for calculating characteristic value parameters by fitting curve.
Fitting Curve kit *6 *6
USCQ-AI901A/EL (USHI-AA550A, USCQ-AI900A, and USPO-AA000A are required.)
USEL-AA551A, This kit enables Elastography (with FLR measurement) with linear and convex transducers.
Elastography-FLR kit OP. OP.
USEL-AA551A/EL Not available in the USA.
USEL-AA550A,
Elastography kit The kit enables Elastography without FLR measurement. Only available in the USA. OP. OP.
USEL-AA550A/EL
USSW-AA550A, This kit allows tissue stiffness to be visualized by generating images that show shear wave
Shear Wave kit OP. OP.
USSW-AA550A/EL propagation.
Upper limit and range are expanded from 200 kPa to 700 kPa to provide measurements of stiffer
USSW-AA551A,
Shear Wave Hard kit targets. OP. OP.
USSW-AA551A/EL
(USSW-AA550A or USLP-AA550A is required.)
USFN-AA550A, CT/MRI/US volume data is loaded, and a CT/MRI/US planar image and an ultrasound image at the
Smart Fusion kit 2 *1 *1
USFN-AA550A/EL same position are displayed together.*
This kit allows display of a guideline in the image by simulating the pathway of the needle and
USSN-AI600A,
Smart Navigation kit the position of the needle tip based on positional information acquired using the magnetic *6 *6
USSN-AI600A/EL 3
sensor. Not available in the USA.*
This kit is used to generate the magnetic field for acquiring positional information for transducers
and needles in Smart Fusion, Smart Navigation, and Smart Sensor 3D modes. Sensor securing
Magnetic Generator kit UIFR-AA550A *1 *1
adaptors and magnetic sensors are provided
(one set for the PVT-375BT or PVT-375SC, and one set for the PVT-475BT).
This magnetic sensor is added for performing Smart Fusion, Smart Navigation, and Smart Sensor
Sensor kit for Fusion unit UIFR-A501A *1 *1
3D using multiple transducers.*4
USFN-AI901A, Used for the Smart Fusion function. Performs automatic position matching between the
Auto Registration kit 5 *1 *1
USFN-AI901A/EL ultrasound volume data acquired in a previous exam and the real-time 2D ultrasound image.*
*1: Advanced Software kit or Premium Software kit is required. *6: Premium Software kit is required.
*2: The Mounting kit for fusion sensor supporting the transducer *7: USMV-AI900A or Smart Sensor 3D kit is required.
used, Magnetic Generator kit UIFR-AA550A, and Sensor kit for *8: Option for WH model.
Fusion unit UIFR-A501A are required. *9: The black-and-white digital printer P95DW-DC cannot be installed
*3: Magnetic Generator kit UIFR-AA550A, Smart Navigation Sensor kit at the rear of the system if the Mounting kit for external HDD
UISN-A500A and Smart Fusion kit USFN-AA550A are required. UZHI-AI900A is installed.
*4: Some options are in addition to UIFR-AA550A. *10: Only available in the CE marking regions and Malaysia.
*5: Magnetic Generator kit UIFR-AA550A and Smart Fusion kit *11: Not available in the CE marking regions.
USFNAA550A are required.
3
Main unit
Unit Model name Remarks CUS- WH
AA000 Model
This pole cart allows the magnetic field generator included in the system main unit to be
Fusion Pole Cart UZWT-A500A *1 *1
positioned independently.
Both superficial and deep structures in a specific region can be observed simultaneously by
USSG-AI900A,
Shadow Glass kit superimposing them. Simultaneous display with a color 4D image showing internal blood flow *6, *7 OP.
USSG-AI900A/EL
(CDI and SMI) is also possible. Premium Software kit is not needed in WH model.
Auto Volume Measurement USOB-AI900A, Used for calculation of the volume by extracting the contours for regions with lower brightness in
*7 OP.
kit USOB-AI900A/EL the 3D Volume image acquired in 4D mode. Enables volume measurement, e.g., antral follicle count.
USPF-AI900A,
3D printer format export kit This kit enables output of volume data for Smart 3D, Mecha4D and Smart Sensor 3D to 3D printers. *7 OP.
USPF-AI900A/EL
Superb Micro Vascular USMI-AI600A,
Visualize low-velocity blood flow at a high frequency rate. Std. in WH model. OP. Std.
Imaging kit USMI-AI600A/EL
USSS-AI900A, This function is used to display high-precision Smart 3D images by detecting the position of the
Smart Sensor 3D kit *1 *1
USSS-AI900A/EL transducer based on positional information acquired using the magnetic sensor.*4
USMP-AI900A,
MicroPure kit This kit enables MicroPure, which supports visualization of small structures. OP. OP.
USMP-AI900A/E
USPV-AI900A, B/W images can be obtained with a wider field of view by moving the transducer in the lateral
Panoramic View kit OP. OP.
USPV-AI900A/EL direction.
USRC-AI900A, This function is used to visualize images with fewer artifacts by cancelling multiple reflections
Multi-reflection Canceller kit *1 *1
USRC-AI900A/EL from the body.
USAT-AI900A, This kit enables visualization of ultrasound frequency-dependent attenuation coefficient within
Attenuation Imaging kit OP. OP.
USAT-AI900A/EL tissue.
USLP-AA550A, Following optional kits are included.
Liver Package Basic kit OP. OP.
USLP-AA550A/EL Optional kits: USSW-AA550A, USAT-AI900A
USLD-AI900A,
Doppler Luminance kit This kit allows display of the pseudo color doppler in three dimensions. OP. OP.
USLD-AI900A/EL
USGI-AA550A,
General Imaging kit This kit allows usage of General Imaging. *8 OP.
USGI-AA550A/EL
*1: Advanced Software kit or Premium Software kit is required. *6: Premium Software kit is required.
*2: The Mounting kit for fusion sensor supporting the transducer *7: USMV-AI900A or Smart Sensor 3D kit is required.
used, Magnetic Generator kit UIFR-AA550A, and Sensor kit for *8: Option for WH model.
Fusion unit UIFR-A501A are required. *9: The black-and-white digital printer P95DW-DC cannot be installed
*3: Magnetic Generator kit UIFR-AA550A, Smart Navigation Sensor kit at the rear of the system if the Mounting kit for external HDD
UISN-A500A and Smart Fusion kit USFN-AA550A are required. UZHI-AI900A is installed.
*4: Some options are in addition to UIFR-AA550A. *10: Only available in the CE marking regions and Malaysia.
*5: Magnetic Generator kit UIFR-AA550A and Smart Fusion kit *11: Not available in the CE marking regions.
USFNAA550A are required.
4
MPDUS0085EAN
Main unit
Unit Model name Remarks CUS- WH
AA000 Model
USUH-AA550A, 11
Ultra High Freq kit This kit enables to use ultra high frequency transducer, PLI-2002BT.* OP. OP.
USUH-AA550A/EL
Rack for mounting a B/W printer.
UZRI-AI900A OP. OP.
(For except for UP-D711MD/ P95DW-DC)
Rack for mounting a DVD recorder and a color printer on two levels.
UZRI-AI902A OP. OP.
(Compatible with UIFR-AA550A)
Mounting kit for Peripheral
Rack for mounting a DC B/W printer in the front/back.
units
UZRI-AA550A (For UP-D711/AC or P95DW-DC) OP. OP.
9
When this kits is installed in the back, UZRI-AI902A can be installed same time.*
UZRI-AA551A One B/W printer and a color printer, or one B/W printer and DVD recorder can be installed. OP. OP.
UZRI-AA000A Rack for mounting a B/W printer in the front left side. OP. OP.
Foot switch UZFS-004A Switch used for freezing, printing, and some other operations by foot. OP. OP.
Gel warmer UZGW-008A This unit warms the ultrasound gel to a suitable temperature. OP. OP.
Transducer cable hanger kit UZMK-AI900A Long hanger on which the transducer cable is hooked. OP. OP.
A basket to store transducer connectors is added to the side of the system main unit (One basket
Transducer Holder kit UZBK-AI900A OP. OP.
is provided in the standard configuration.)
This kit used to establish connection to the DICOM network via wireless LAN.
UIWL-A500A Complied with the Radio Law of Japan and applicable laws and regulations of USA, OP. OP.
Wireless LAN kit Canada, EU member states, Iceland, Norway, Liechtenstein, and Switzerland.
This kit enables connection to the DICOM network via wireless LAN.
UIWL-AI900A OP. OP.
(For Singapore, Malaysia, Saudi Arabia, Russia, Australia, and Turkey.)
EV/ER Transducer Holder kit UZPH-AI900A This kit is used to add a dedicated holder for endocavitary transducers. OP. OP.
Transducer Holder kit (left) UZPH-AA000A This kit is used to add transducer holder on the left side. OP. OP.
Keyboard kit UIKB-AI900A This kit is used to add a hardware full keyboard for entering the patient ID and comments. OP. OP.
UZKF-AI900A (French)
UZKG-AI900A (German)
UZKI-AI900A (Italian)
This kit is intended for changing the key tops of the full keyboard (UIKB-AI900A) to support
UZKS-AI900A (Spanish)
Local Language Key-Top kit specific languages. OP. OP.
UZKD-AI900A (Danish)
(UIKB-AI900A is required.)
UZKN-AI900A (Norwegian)
UZKW-AI900A (Scandinavian)
UZKR-AI900A (Russian)
OLED Monitor unit UIOM-001A 21.6-inch wide OLED Monitor to replace 23/21.5-inch LCD Monitor with LED back light. OP. OP.
VIDEO unit UIVP-AA550A This unit enable DVI signal output. OP. OP.
DataBase for External USDB-AI900A, Function for setting up the patient database in the USB HDD in order to perform examinations.
OP. OP.
HDD kit USDB-AI900A/EL (UZHI-AI900A is required.)
Mounting kit for External Box with lock for installing the external HDD. HDD itself is not included.
UZHI-AI900A OP. OP.
HDD (USDB-AI900A is required.)
ECG Cable Hanger kit UZMK-AI901A Hook used to hang the ECG cables on the front of the operating panel. OP. OP.
Kit for adding a USB port to the operating panel.
Panel USB Port kit UIUB-AI900A OP. OP.
(Note that this kit cannot be installed if the UIST-AI900A is installed.)
A battery kit to provide approximately 30 minutes of operation without being connected to a
Battery unit UEBT-AA550A OP. OP.
power outlet.
Track Ball kit UZTB-AI900A Newly designed (heavier) trackball to improved response to user and therefore workflow. OP. OP.
Pointing device, dial on track ball allows easier Gain adjustment without reaching out for another
Palm Controller kit UZPT-001A OP. OP.
button.
*1: Advanced Software kit or Premium Software kit is required. *6: Premium Software kit is required.
*2: The Mounting kit for fusion sensor supporting the transducer *7: USMV-AI900A or Smart Sensor 3D kit is required.
used, Magnetic Generator kit UIFR-AA550A, and Sensor kit for *8: Option for WH model.
Fusion unit UIFR-A501A are required. *9: The black-and-white digital printer P95DW-DC cannot be installed
*3: Magnetic Generator kit UIFR-AA550A, Smart Navigation Sensor kit at the rear of the system if the Mounting kit for external HDD
UISN-A500A and Smart Fusion kit USFN-AA550A are required. UZHI-AI900A is installed.
*4: Some options are in addition to UIFR-AA550A. *10: Only available in the CE marking regions and Malaysia.
*5: Magnetic Generator kit UIFR-AA550A and Smart Fusion kit *11: Not available in the CE marking regions.
USFNAA550A are required.
5
Main unit
Unit Model name Remarks CUS- WH
AA000 Model
Tilt panel kit UITP-AA550A Kit for enabling tilting of the touch panel. OP. OP.
For PLT-1204BX.
Dynamic Micro Slice unit UIMS-AA550A *6 *6
(USMS-AA550A is required.)
USMS-AA550A, For PLT-1204BX.
Dynamic Micro Slice kit *6 *6
USMS-AA550A/EL (UIMS-AA550A is required.)
USHE-AI900A,
Online Help kit Kit for displaying the operation manual on the viewing monitor. OP. OP.
USHE-AI900A/EL
USPA-AI900A, A sequence of operations is registered, and each operation is executed by single switch operation.
Protocol Assistant kit OP. OP.
USPA-AI900A/EL (Not necessary when USSE-AI900 is installed.)
A demonstrating movie by Dr Inigo Iriate, Professor of Ultrasound of the Spanish Society of
USPA-AI901A, Rehabilitation & Physical Medicine which tells the practical scanning tips of the anatomy and
MSK Protocol Movie kit OP. OP.
USPA-AI901A/EL scanning of the shoulder provides examiners’ skill development.
(USPA-AI900A or USSE-AI900A is required.)
USDL-AI900A, This kit enables a combined report for the abdominal measurement applications.
Multi Parametric Report kit OP. OP.
USDL-AI900A/EL (USSW-AA550A or USLP-AA550A is required.)
Security Management kit USSM-AI900A This kit provides software for security management of the system. OP. OP.
USRA-AI900A,
RADS kit This kit enables Reporting and Data System. OP. OP.
USRA-AI900A/EL
USIO-AI900A, 10
IOTA kit This kit enables International Ovarian Tumor Analysis.* OP. OP.
USIO-AI900A/EL
USNA-AI900A, This kit enables large capacity RAW data management with Network Attached Storage (NAS)
Network Storage kit OP. OP.
USNA-AI900A/EL which is commercially available. (NAS itself is not included.)
USTR-AI900A,
Tricefy Access kit This kit enables to access Tricefy which is a cloud service for clinical images. OP. OP.
USTR-AI900A/EL
Video capture unit, HDMI to USB converter.
ApliGate kit UIAG-001A OP. OP.
(USAG-001A is required.)
USAG-001A, ApliGate software.
ApliGate Soft kit OP. OP.
USAG-001A/EL (UIAG-001A is required.)
USWC-AI900A, Video clip captured by the camera which is connected to the main unit can be displayed on the
ApliCam kit 11 OP. OP.
USWC-AI900A/EL screen as picture-in-picture.*
USRI-AI600A,
Reference Image kit Software to enable images from a previous examination to be displayed at the right of the screen. OP. OP.
USRI-AI600A/EL
This kit enables that the information in the DICOM data from the digital Mammography MLO/CC
USMB-AI900A,
Breast Scan Guide kit images to be used to create an overlay position on the Ultrasound body mark in reference mode. *6 *6
USMB-AI900A/EL
Reference Image kit (USRI-AI600A) is required.
It automatically traces and displays the anatomical position of transducer mark based on the
USSB-AI900A,
Smart Body Mark kit transducer position using magnetic sensor. OP. OP.
USSB-AI900A/EL
(UIFR-AA550A is required.)
Package of Fusion unit (UIFR-AA550A) and Fusion adapter(UAFS-004A).
Breast Package kit UIBP-AA550A *1 *1
(USBP-AA550A is required.)
Breast Package software. Package of USSB-AI900A, USFN-AA550A, USSW-AA550A,
USBP-AA550A,
Breast Package Soft kit and USEL-AA550A or USEL-AA551A. *1 *1
USBP-AA550A/EL
(UIBP-AA550A is required.)
USPS-AA000A,
Probe 4Port kit This kit activates 4th transducers port. OP. OP.
USPS-AA000A/EL
USAO-AA000A, This kit enables advanced software kits
Advanced Software kit OP. OP.
USAO-AA000A/EL (same level as Aplio a450).
USPO-AA000A, This kit enables premium software kits
Premium Software kit OP. OP.
USPO-AA000A/EL (same level as Aplio a550).
*1: Advanced Software kit or Premium Software kit is required. *6: Premium Software kit is required.
*2: The Mounting kit for fusion sensor supporting the transducer *7: USMV-AI900A or Smart Sensor 3D kit is required.
used, Magnetic Generator kit UIFR-AA550A, and Sensor kit for *8: Option for WH model.
Fusion unit UIFR-A501A are required. *9: The black-and-white digital printer P95DW-DC cannot be installed
*3: Magnetic Generator kit UIFR-AA550A, Smart Navigation Sensor kit at the rear of the system if the Mounting kit for external HDD
UISN-A500A and Smart Fusion kit USFN-AA550A are required. UZHI-AI900A is installed.
*4: Some options are in addition to UIFR-AA550A. *10: Only available in the CE marking regions and Malaysia.
*5: Magnetic Generator kit UIFR-AA550A and Smart Fusion kit *11: Not available in the CE marking regions.
USFNAA550A are required.
6
MPDUS0085EAN
Superb Micro Vascular Imaging kit Superb Micro Vascular Imaging kit
USMI-AI600A USMI-AI600A
4D kit 4D kit
* USAO-AA000A or
USMV-AI900A USPO-AA000A is required. USMV-AI900A
7
A
MicroPure kit
USMP-AI900A
CHI-Q kit
USCQ-AI900A
Fetal Heart MPI Measurement kit Fetal Heart MPI Measurement kit
USFH-AI600A USFH-AI600A
B C D
8
MPDUS0085EAN
B C D
Mounting kit for fusion sensor Mounting kit for fusion sensor
UAFS-001A, UAFS-002A, UAFS-003A, UAFS-004A, UAFS-005A, UAFS-001A, UAFS-002A, UAFS-003A, UAFS-004A, UAFS-005A,
UAFS-006A, UAFS-007A, UAFS-008A, UAFS-009A, UAFS-010A UAFS-006A, UAFS-007A, UAFS-008A, UAFS-009A, UAFS-010A
Sensor kit for Fusion unit Sensor kit for Fusion unit
UIFR-A501A UIFR-A501A
Linear Array Transducer Linear Array Transducer Linear Array Transducer Linear Array Transducer
PLI-2002BT PLI-2002BT PLI-2002BT PLI-2002BT
Ultra High Freq kit Ultra High Freq kit Ultra High Freq kit Ultra High Freq kit
USUH-AA550A USUH-AA550A USUH-AA550A USUH-AA550A
Linear ArrayTransducer Linear Array Transducer Linear ArrayTransducer Linear Array Transducer
PLT-1005BT PLT-1005BT PLT-1005BT PLT-1005BT
Breast Package kit Mounting kit for fusion sensor Breast Package kit Mounting kit for fusion sensor
UIBP-AA550A UAFS-004A UIBP-AA550A UAFS-004A
Breast Package Soft kit Breast Package Soft kit Breast Package Soft kit Breast Package Soft kit
USBP-AA550A USBP-AA550A USBP-AA550A USBP-AA550A
Smart Fusion kit Smart Sensor 3D kit Smart Fusion kit Smart Sensor 3D kit
Note: USFN-AI900A and USSS-AI900A Note: USFN-AI900A and USSS-AI900A Note: USFN-AI900A and USSS-AI900A Note: USFN-AI900A and USSS-AI900A
USFN-AA550A USSS-AI900A USFN-AA550A can be installed simultaneously.
USSS-AI900A can be installed simultaneously.
can be installed simultaneously. can be installed simultaneously.
Auto Registration kit Luminance kit Auto Registration kit Luminance kit
USFN-AI901A USLM-AI900A USFN-AI901A USLM-AI900A
E
9
E
CV kit
UACV-AA000A
CW kit
UICW-AA000A
10
MPDUS0085EAN
Mounting kit for Mounting kit for Mounting kit for Mounting kit for Mounting kit for * P95DW-DC
Peripheral units Peripheral units Peripheral units Peripheral units Peripheral units cannot be installed at the rear
of the system if the Mounting
UZRI-AA551A UZRI-AI902A UZRI-AI900A UZRI-AA000A UZRI-AA550A kit for external HDD UZHI-AI900A
is installed.
* When this kits is installed in the
back, UZRI-AI902A can be installed
Wireless LAN kit Wireless LAN kit same time.
UIWL-A500A UIWL-AI900A
G
Gel warmer
Tilt panel kit
UZGW-008A
UITP-AA550A
Keyboard kit
OLED Monitor unit
UIKB-AI900A
UIOM-001A
DataBase
for External HDD kit IOTA kit
USDB-AI900A USIO-AI900A
Mounting kit
Network Storage kit
for External HDD kit
USNA-AI900A
UZHI-AI900A
ApliGate kit
Online Help kit
UIAG-001A
USHE-AI900A
ApliCam kit
Transducer Holder kit
USWC-AI900A
UZBK-AI900A
11
TRANSDUCER OPTIONS/OPERATION MODES
Displayed range of
Freq. frequency Precision Ultra Micro
Model name Label (MHz) 2D ApliPure 1 BEAM M CDI Power
Imaging Wide View Pure*
(MHz)
PSI-70BT i10S4 7.0 3.5~9.0 - - - -
PST-25BT 5S1 2.5 1.7~4.7 - - - -
PST-28BT 6S1 3.0 1.5~6.0 - - - -
PST-30BT 5S2 3.0 1.7~5.2 - - - -
PST-50BT 6S3 5.0 3.0~8.2 - - - -
PST-65BT 12S4 7.0 3.5~12.0 - - - -
PVT-350BTP 6CP1 3.5 2.0~4.5 - - -
14
MPDUS0085EAN
SPECIFICATIONS • Camera
– Applicable OS: Windows 10
System
– Interface: USB2.0 type-A
• Scan methods
– Aspect ratio: 4:3
– Linear scan
– Resolution: 640:480 pixel or more
(Some transducers can perform oblique scanning.)
– Driver: Compatible with the driver
– Sector scan
installed in Windows 10.
– Convex scan
– Trapezoid scan 16
– Curved vector scan
2D mode (B mode)
• Viewing Depth
• Monitor
The viewing depth depends on the transducer used.
7-a – High-definition 21.5-inch Wide LCD monitor with LED
– Convex
backlight
· Minimum depth: 2 cm
· Resolution: 1920 × 1080 (Full HD)
1-b · Maximum depth: 50 cm
· Viewing angle: 178 degrees
– Linear
· Contrast ratio: typ. 1000: 1
· Minimum depth: 1 cm
· Response time (ms): typ. 14
· Maximum depth: 14 cm
· Luminance (cd/m2): typ. 300
– Sector
· Conformance Standard:DICOM Part 14
· Minimum depth: 1 cm
– High-definition 23-inch Wide LCD monitor with LED
· Maximum depth: 28 cm
backlight
• Line density
· Resolution: 1920 × 1080 (Full HD)
– The line density differs depending on the transducer used.
· Viewing angle: 178 degrees
– The line density can be changed.
· Contrast ratio: typ. 1000: 1
• Ultrasound Frame Rate
· Response time (ms): typ. 14
– The ultrasound frame rate can be adjusted by using the
· Luminance (cd/m2): typ. 300
following in combination.
· Conformance Standard: DICOM Part 14
– 4K 21.6-inch Wide OLED monitor · Line density
· Parallel signal processing
· Resolution: 3840 × 2160
· Viewing angle: 178 degrees *CR≧10 • Scan Angle and Scan Width
1-a – Adjustment of the field width (scan width, scan angle) is
· Contrast ratio: typ. 1000000: 1
· Response time (μs): typ. 40 (30 + 10) possible.
*Black White Black
– Adjustment of beam steering (scanning position) is
2
possible.
· Luminance (cd/m ): typ. 200 (7500 K) – Adjustment of linear beam steering is possible.
typ. 145 (13000 K)
• PAN/EXPAND
· Conformance Standard: DICOM Part 14 1-e Real-time PAN/EXPAND
• Presets
– Scale enlargement/reduction using the encoder is
– System preset: 1 type
possible.
Compatible Peripheral Devices – Movement to the desired section using the trackball is
• Black-and-white digital printer possible.
– UP-D711MD/WO: AC (100 V to 240 V, SONY) – The transmission focus is optimized in steps above.
– UP-D898MD: AC (100 V to 240 V, SONY) – The specified range of an image can be magnified.
– P95DW: AC (100 V to 240 V, MITSUBISHI) (Spot Zoom)
– P95DW-DC: DC (24 V, MITSUBISHI) • Transmission Focus
• Color digital printer – Transmission conditions: A maximum of 8 steps
– UP-D25MD: AC (100 V to 240 V, SONY) • Transmission Frequency
– CP30DW: AC (120 V, 220 V to 240 V, – Multi frequency: 3 frequencies can be selected
MITSUBISHI) from 13 types.
• DVD video recorder • GAIN
– HVO-550MD/FHD: AC (100 V to 240 V, NTSC/PAL, – The display brightness for 2D can be changed.
SONY) (Also available when the image is frozen.)
• USB flash drive – The display brightness for 2D and M can be changed
• External HDD simultaneously.
• Barcode reader
15
• STC 20 • Precision Imaging
– Software STC – Precision+
· Depth direction from · Structures in 2D-mode images can be displayed more
the body surface: 8-step slide controls clearly and the background can be displayed more
(common for 2D and M) smoothly.
· Lateral direction 19 · Saturation in high-intensity regions of tissue structures is
in the image: 6-step slide controls reduced, allowing the tissue structures to be displayed in
(common for 2D and M) a more natural manner.
– Hardware STC (UIST-AI900A is required.) • TSO (Tissue Specific Optimization)
· 8-step slide controls (common for 2D and M) – Reception focus compensation can be performed.
• Acoustic Output – Automatic reception focus compensation can be
– Adjustment is possible to 100%. performed (Auto TSO).
• Adjusting the 2D Image Quality
25-c • BEAM (Biopsy Enhancement Auto Mode)
– Dynamic range (Also available when the image is frozen.) – Display of the needle can be enhanced in the image.
– Time-smoothing (persistence) – The enhancement level can be adjusted.
– Gamma (Also available when the image is frozen.) • Doppler Luminance (USLD-AI900A is required.)
– Frame rate – Doppler Luminance is a function for displaying 3D images
– ApliPure created from 2D color images acquired in color modes
– Precision (CDI/Power/ADF/SMI).
– Reverb (USRC-AI900A is required.)
• 2D Map M mode
– The grayscale pattern can be changed and virtual color • M Transmission Frequency
setting for the 2D image is possible. – Multi frequency: 5 types (at maximum)
– Settings can also be changed when the image is frozen. • M Sweep Speed
2-b
14 • 2D Quick Scan – The Sweep Speed can be changed in M mode.
– The gain and STC can be adjusted automatically. • M Gain
5-d • THI (Tissue Harmonic Imaging) – M gain can be corrected for 2D gain.
– THI signal processing methods • M Image Processing Parameters
· Pulse subtraction method – M dynamic range (Can be changed even after the image
· Filtering method has been frozen.)
· Differential method 13 – M auto gain control
• Display Orientation – M gamma (Can be changed even after the image has
– Top/bottom reversal is possible. been frozen.)
– Left/right reversal is possible. 2-c • M Map
• Image Size – M image virtual color setting is possible. The setting can
– The displayed image size can be switched between small be changed even after the image has been frozen.
and large. • THI
• ApliPure – 2D mode and THI mode are linked, and M images can be
– ApliPure 17 displayed in THI mode.
This function reduces ultrasound wave interference within · Pulse subtraction method
tissues, which appear as speckle patterns or speckle noise · Filtering method
on 2D images. · Differential method
– ApliPure+ • M Mark
This function can display the boundaries between tissues – The M cursor can be displayed on 2D or C images.
more clearly and reduce speckle noise and acoustic – The M cursor displayed position can be adjusted.
shadows. • Flex-M
• MicroPure (USMP-AI900A is required.) – Any desired plane can be set on the 2D-mode image and
– Small structures can be extracted by performing filtering the M-mode image for the set plane can be reconstructed.
for 2D-mode images.
– Visualization of very small calcifications and other
extremely small lesions can be improved.
16
MPDUS0085EAN
Doppler (Spectrum Doppler) – Oblique scans are possible using a specific linear
• Doppler Mode transducer.
– PWD (pulsed-wave Doppler) – Auto Invert function
3-b – HPRF PWD (can be switched to HPRF mode) • Doppler Multifrequency
– CWD (continuous-wave Doppler) 4 – The PWD transmission frequency can be changed.
(UICW-AA000A is required.) • Doppler Sweep Speed
– Pencil CWD (pencil-type transducer) – The Sweep Speed can be changed in Doppler mode.
(UICW-AA000A and UIPC-AA550A are required.) • Doppler Display Dynamic Range
• Doppler Pulse Repetition Frequency (PRF) – The display dynamic range of the Doppler image can be
3-a – PWD: 0.3 kHz to 52.1 kHz changed.
4-b – CWD: 1.4 kHz to 52.1 kHz 4-a 4-e • Doppler Auto Trace (measurement performed after freezing
• Doppler Scan the image)
– 2D/D simultaneous scan – Measurement of peak velocity and mean velocity is
– D only scan possible by automatic velocity tracing.
• Doppler Sampling Volume – The following Doppler waveform trace is possible.
– The Doppler range gate width can be changed. – Trace style: Waveform Peak, Mean,
(Minimum 0.3 mm) Peak + Mean
• Doppler Sampling Shift – Trace area specified: Forward, Reverse, Full, Auto
– 0 cm to the maximum depth – Measurement item: Max, Min, Mean, PI, RI, etc.
• Doppler Cursor Mode
– Operation for the 2D live image is possible with the Color Doppler
Doppler sampling volume displayed in it. Color Doppler Mode
• Doppler Filter • Display mode
– The Doppler filter cutoff can be changed. – CDI mode
• Doppler Gain · Flow velocity
– The display brightness for Doppler can be changed. · Flow velocity/variance
4-d • Doppler Quick Scan 14 5-e · Power
– The Doppler scale and baseline shift can be adjusted – Power Angio mode
automatically. 12 – TDI mode
• Doppler Frequency Analysis and Image Processing 5-b – TwinView 5-f
– Method: FFT · Simultaneous dual-screen display with 2D mode is
– No. of data items: 255 (maximum) available.
• Indication of Doppler Spectrum Direction
21 – SMI (Superb Micro-vascular Imaging) mode
– Reverse display of the velocity spectrum is possible. (USMI-AI600A is required.) Pre-installed in WH model.
• Doppler Baseline Shift (Zero Shift) · Clutter suppressed
– The velocity baseline of Doppler images can be shifted. · Blood flow enhanced
– The baseline shift setting can also be adjusted when
images that were frozen are displayed.
15 ADF (Dynamic Flow) mode
–
· Direction display
4-b • Doppler Audio • C Map
– Stereo output (blood flow toward and away from the – C map can be selected for each color Doppler mode.
transducer) – Changes can also be made when the image is frozen.
4-c • Doppler Map • C Scale (Switching the Velocity Range)
– The brightness conversion table and the virtual color for – The velocity range can be changed.
Doppler images can be set. • C Time-Smoothing (Persistence)
• Display of Doppler Scale – The result of temporal correlation processing between
– 2 types (velocity, Doppler shift frequency) the previous image and current image can be displayed.
• Doppler Focus
• C Baseline (Zero Shift)
– Automatically follows the sample position.
– The velocity baseline of color Doppler images can be
• Doppler Angle Mark
shifted.
– This mark is displayed for measuring the angle between
– The baseline shift setting can also be adjusted when
the direction of the velocity and the direction of the
images that were frozen or images in the image memory
ultrasound beam.
are displayed.
5-a • Doppler Oblique Scan (PWD Steering)
17
• Reverse C Display • Black and White/Color Balance
– Coloring is reversed. – Color weighting for B/W images can be set by comparing
– Changes can also be made when the image is frozen. the M Color Doppler images and B/W images.
• Black and White/Color Balance – Changes can also be made when the image is frozen.
– By comparing the color Doppler images and B/W images, • Color Gain
color weighting to B/W can be set. – The display brightness of the M Color Doppler image can
– Changes can also be made when the image is frozen. be changed.
• C Gain • M Color Doppler Multi-Frequency
– The display brightness of color Doppler images can be – Doppler transmission frequency can be selected in M
changed. Color Doppler image acquisition.
• C Multifrequency • M Color Doppler Filter
– The transmission frequency for color Doppler image – M Color Doppler low-cut filter can be changed.
acquisition can be changed.
• C Line density Reference Signals
– The color Doppler image line density can be changed. • Type
• C ROI (region of interest) – Electrocardiogram (ECG)
– Position, size, and steering adjustment is possible for · Lead I is the standard connection.
color Doppler ROIs. · External input is possible.
· DC IN
• C Transmit Focus
The connected device must comply with IEC 60601-1.
– Automatically follows the color Doppler ROI position. · Top/bottom inversion is possible.
• C Filter · Lead switching
– Color Doppler low-cut filter can be changed. · Pacemaker
• Variance Curve · INST
– The display of the color variance component can be – Phonocardiogram (PCG)
adjusted. (UJUR-AA000A and UJUR-AI902A are required.)
• Color Quick Scan · PCG microphone: Acceleration type
– The following operations are possible when a linear · Filter: Switching is possible.
transducer is used. – Pulse (UJUR-AA000A and UJUR-AI902A are required.)
· The position of a color ROI and angle of color steering · Pulse transducer: Air conduction type
are adjusted automatically. – Respiration
· Impedance method using the Reference Signal cable
· When PWD sampling volume is displayed, the Doppler
• Heart Rate
gate position, Doppler steering angle, and Doppler – The heart mark blinks in synchronization with the heart
angle are adjusted automatically. beat detected by the ECG.
– The heart rate is displayed.
Color Doppler M mode (MDF Mode) • Reference Signal Sweep Speed
• Display mode – This changes the reference signal sweep speed.
– M-mode CDI
· Velocity display Other Diagnostic Function
· Velocity/variance display • CHI (Contract Harmonic Imaging) (USHI-AA550A is required.)
· Power display The second-harmonic wave signals from the microbubbles
– M-TDI mode in the contrast medium can be effectively visualized.
• M Color Doppler Map (CDI MAP) – The following image modes can be selected.
· PS (Pulse Subtraction) -Low, PS-Low2
– Color Doppler map can be selected for each mode.
· VRI (Vascular Recognition Imaging)
• M Color Doppler Velocity Range Selection (C Scale) · Fundamental
– The velocity range can be selected. · CHI ADF
• M Color Doppler Baseline (C Baseline) · SMI (Superb Micro-vascular Imaging)
– The zero-velocity line on the M Color Doppler image can (USMI-AI600 is required.) Pre-installed in WH model.
be shifted. – The following functions can be selected.
– The baseline shift setting can also be adjusted when · 2D TwinView
images that were frozen are displayed or when the · MFI (Micro Flow Imaging)
image in the image memory is played back. · Image Stabilizer
• Color Reverse Display · MI Constant function
– The colors can be reversed.
– Changes can also be made when the image is frozen.
18
MPDUS0085EAN
19
– The following functions can be used. – The following measurements can be performed.
· Blend display · MPR
· Segment display · Multi Auto Volume Measurement (USOB-AI900A is
· Multiplane display (Triple display) required.)
· Multivolume display (Quad display) • Smart 3D
· 3D body mark display (Quad display) 3D image can be generated from the 2D image and any
· Comparative display (Quad display) input volume shape.
· Auto Registration (USFN-AI901A is required.) – The following functions can be used.
· Auto Track (Auto Track kit is required.) · Volume Color
– The following measurements can be performed. · Multi View
· Basic measurement · Volume View
· Cardiac measurement (for ultrasound live image) · Magic Cut
• Protocol Assistant (USPA-AI900A is required.) · OmniView
A series of operations (a protocol) that has been created · CHI (Advanced Software kit or Premium Software kit
for the intended examination can be executed automatically. and USHI-AA550A are required.)
Protocols can be created and edited. · Shear Wave (USSW-AA550A is required.)
22 • Shear Wave (USSW-AA550A is required.) · STL export (Advanced Software kit or Premium
Images representing the speed of propagation of tissue Software kit and USPF-AI900A are required.)
displacement (Shear Wave speed) can be visualized (Shear • Reference (USRI-AI600A is required.)
Wave scan) by locally displacing tissues by transmitting a The ultrasound images of the current examination and the
burst wave with high acoustic pressure. previously acquired image of the patient can be displayed
– The following Shear Wave display modes are available. simultaneously.
· Speed: Shear Wave speed display (m/s) – The following video modes can be selected.
· Elasticity: elasticity display (kPa) · 2D mode (B mode)
· Propagation: propagation display · Color Doppler
– ECG Sync Acquisition function can be used. – The following image data can be referred to.
– Shear Wave measurement can be performed. · US image (still image)
– ATI (Attenuation Imaging) (USAT-AI900A is required.) · Raw data acquired using CUS-AA000
The ultrasound wave attenuation can be displayed in · CT
color parametric and measured. · MR
• Smart Navigation (USSN-AI600A etc. are required.) · MG
A needle navigation line can be superimposed on the · PET
ultrasound image based on the positional relationship – The following measurement can be performed.
between the magnetic sensor attached to the transducer Basic measurement
and the magnetic sensor attached to the puncture needle. • Smart Area Indication OB (USSI-AA550A is required.
• Smart Sensor 3D (Advanced Software kit or Premium Not available in the USA and Canada)
Software kit and USSS-AI900A, UIFR-AA550A, and The Smart Area Indication OB function is used in obstetrics
UIFR-AI501A are required.) to detect fetal measurement regions based on signals in
Generation of 3D images can be performed based on the images.
positional information obtained using the magnetic sensor – The following measurements are available.
attached to the transducer. · BPD
– The following functions can be used. · HC
· Volume Color · AC
· Multi View · FL
· Volume View – The following transducer supports Smart Area Indication
· Magic Cut (OB).
· CHI (USHI-AA550A is required.) · PVT-475BT
· Luminance (USLM-AI900A is required.)
· Shadow Glass (USSG-AI900A is required.)
· Shear Wave (USSW-AA550A is required.)
· OmniView
· ECG Sync Construction
· STL export (Advanced Software kit or Premium
Software kit and USPF-AI900A are required.)
20
MPDUS0085EAN
21
Measurement Functions · MV (mitral valve) measurements
Basic Measurement Functions · PV (pulmonary valve) measurements
• 2D-mode measurements · LV MASS measurements
– Distance · Auto EF measurements (USEF-AI600A is required.)
· Distance – M-mode measurements
· Trace Length · LV measurements
· Mean-IMT 6 · AV measurements
– Area · MV measurements
– Angle – Doppler measurements
· Angle · Trans-Aortic valve flow measurement
· Joint · Trans-Mitral valve flow measurement
– Volume · Trans-Pulmonary vein flow measurement
– Stenosis ratio · Trans-Tricuspid valve flow measurement
· %Stenosis (Distance) · Trans-Pulmonary valve flow measurement
· %Stenosis (Area) · Blood flow waveform auto measurements
• 4D-mode measurements (USMV-AI900A is required.) · Coronary measurements
– Distance · PISA measurements
· Distance – OB (obstetrics) measurements
· Trace Length · The data for determining fetal growth based on the
· Mean-IMT measured fetal size is displayed.
– Area · The list of measured data or a graph of the measured
– Angle value development (fetal growth conditions) is displayed.
· Angle · Week function (gestational age)
· Joint · Measurement data saving is possible.
– Volume 6 · Auto NT measurement
· Volume · Fetal heart MPI measurement (USFH-AI600A is required.)
· Auto Volume Measurement (Advanced Software kit or · Anatomy
Premium Software kit and USOB-AI900A are required.) · User chart registration
– Stenosis ratio – Vascular measurement
· %Stenosis (Distance) · CCA (Common Carotid Artery) measurement
· %Stenosis (Area) · ECA (External Carotid Artery) measurement
– Shear Wave measurement · ICA (Internal Carotid Artery) measurement
• M-mode measurements · Vert A (Vertebral Artery) measurement
– Slope · Subclav A (Subclavian Artery) measurement
– Distance · Auto-IMT measurement
– Time · IMT-C10 measurement
– Heart rate – User-registered measurements registration function.
• PW/CW Doppler measurements · Measurement items and calculation items based on
– Velocity the measured values
– Acceleration · Layout setting on the Worksheet screen
– Time · Switch layout setting of the touch panel
– Heart rate · Measurement package DICOM code registration
– PI
– RI Advanced Measurement Functions
– S/D • 2D wall motion tracking (USWT-AI900A are required.)
– Flow volume Wall motion can be analyzed by semi-automatically
– Doppler trace extracting the left ventricular myocardium from the image
data acquired by the system and displaying it for the
Application Measurement Functions evaluation of myocardial motion.
• Cardiac measurements – Wall motion tracking in 2D dynamic images
– 2D-mode measurements – Wall motion information display
· LV (left ventricular function) measurements – Polar map display
· LA (left atrial volume) measurements – Local/whole myocardial wall motion parameter curve
· AV (aortic valve) measurements display
22
MPDUS0085EAN
24
MPDUS0085EAN
• According to the degree of protection against electric shock • EU and other regions requiring compliance with European
– EQUIPMENT WITH TYPE-BF APPLIED PARTS Directive 93/42/EEC and subsequent amendments
(Transducer, ECG electrodes, PCG sensor, Pulse sensor) – General: EN 60601-1: 2006 + A1: 2013
• According to the degree of protection against harmful – Collateral: EN 60601-1-2: 2015
ingress of water – Particular: EN 60601-2-37: 2008 + A1: 2015
– IPX0 (enclosed EQUIPMENT without protection against • USA
ingress of water) – General: AAMI ES 60601-1: 2005 + C1: 2009 + A2:
– However, the footswitch is IPX8 and the transducers are 2010 + A1: 2012
IPX7 (excluding the connector part). – Collateral: IEC 60601-1-2: 2014
• According to the degree of safety of application in the IEC 60601-1-6: 2010 + A1: 2013
presence of a FLAMMABLE ANESTHETIC MIXTURE WITH AIR – Particular: IEC 60601-2-37: 2007 + A1: 2015
or WITH OXYGEN OR NITROUS OXIDE • Other regions requiring compliance with IEC 60601-1 Ed. 2
– EQUIPMENT not suitable for use in the presence of a – General: IEC 60601-1: 1988 + A1: 1991 + A2: 1995
FLAMMABLE ANESTHETIC MIXTURE WITH AIR or WITH – Collateral: IEC 60601-1-1: 2000
OXYGEN OR NITROUS OXIDE IEC 60601-1-2: 2001 + A1: 2004
• According to the mode of operation IEC 60601-1-4: 1996 + A1: 1999
– CONTINUOUS OPERATION – Particular: IEC 60601-2-37: 2001 + A1: 2004 + A2: 2005
• Sterilization method • Other regions requiring compliance with IEC 60601-1 Ed. 3
– System main unit – General: IEC 60601-1: 2005
· Not suitable for sterilization. – Collateral: IEC 60601-1-2: 2007
– Transducers – Particular: IEC 60601-2-37: 2007
· Sterilization methods are specified in the relevant • Other regions requiring compliance with IEC 60601-1 Ed. 3.1
operation manuals. – General: IEC 60601-1: 2005 + A1: 2012
– Collateral: IEC 60601-1-2: 2007
Conformance Standards – Particular: IEC 60601-2-37: 2007 + A1: 2015
• Canada Note: The above standards are applicable to the ultrasound system
– General: CAN/CSA-C22.2 No. 60601-1: 14 at the time of purchase. These standards continue to remain
– Collateral: IEC 60601-1-2: 2014 applicable even if the system configuration is changed as a
result of using options in combination. The standards of the
CAN/CSA-C22.2 NO. 60601-1-6A: 11 ultrasound system are applicable to transducers.
– Particular: IEC 60601-2-37: 2007 + A1: 2015
*1: 533 mm when the provided transducer connector holder is removed. (Main unit only; monitor width not included.)
*2: In Standby mode with internal battery UEBT-AA550A.
25
MASS
Model name Name of component Mass [kg] (lb) Model name Name of component Mass [kg] (lb)
System main unit UAFS-008A Mounting kit for fusion sensor 0.1 (0.2)
CUS-AA000 Aplio a 96 (211.6) UAFS-009A Mounting kit for fusion sensor 0.1 (0.2)
Options/Accessories for main unit UAFS-010A Mounting kit for fusion sensor 0.1 (0.2)
UICW-AA000A CW kit 1.1 (2.4) USMV-AI900A 4D kit 0.1 (0.2)
UJUR-AA000A Reference Signal unit 2.2 (4.8) USLM-AI900A Luminance kit 0.1 (0.2)
UJUR-AI900A Reference Signal Cable kit 0.3 (0.7) USSG-AI900A Shadow Glass kit 0.1 (0.2)
UJUR-AI901A Reference Signal Cable kit 0.1 (0.2) USOB-AI900A Auto Volume Measurement kit 0.1 (0.2)
UJUR-AI902A Reference Signal Sensor unit 0.4 (0.9) USPF-AI900A 3D printer format export kit 0.1 (0.2)
USSE-AI900A Stress Echo kit 0.1 (0.2) USMI-AI600A Superb Micro Vascular Imaging kit 0.1 (0.2)
USWT-AI900A 2D Wall Motion Tracking kit 0.1 (0.2) USSS-AI900A Smart Sensor 3D kit 0.1 (0.2)
USWT-AI904A 2D Wall Motion Tracking Fetal kit 0.1 (0.2) USMP-AI900A MicroPure kit 0.1 (0.2)
USEF-AI600A Auto EF Measurement kit 0.1 (0.2) USPV-AI900A Panoramic View kit 0.1 (0.2)
USWN-AA550A Workflow Navigator kit 0.1 (0.2) USRC-AI900A Multi-reflection Canceller kit 0.1 (0.2)
USQM-AA550A Measurement Assistant kit 0.1 (0.2) USAT-AI900A Attenuation Imaging kit 0.1 (0.2)
USWT-AI907A Auto GLS kit 0.1 (0.2) USLP-AA550A Liver Package Basic kit 0.1 (0.2)
USCT-AI900A Contrast Enhance kit 0.1 (0.2) USLD-AI900A Doppler Luminance kit 0.1 (0.2)
UIPC-AA550A Pencil Connector unit 0.3 (0.7) USGI-AA550A General Imaging kit 0.1 (0.2)
UAEH-AI900A M-TEE hanger kit 1.5 (3.3) USUH-AA550A Ultra High Freq kit 0.1 (0.2)
UAEH-AI901A TEE hanger kit 1.2 (2.6) UZRI-AI900A Mounting kit for peripheral unit 1.1 (2.4)
UIST-AI900A STC kit 0.1 (0.2) UZRI-AI902A Mounting kit for peripheral unit 3.1 (6.8)
UACV-AA000A CV kit 1.0 (2.2) UZRI-AA550A Mounting kit for peripheral unit 2.8 (6.2)
USFP-AI900A Smart Fetal Heart kit 0.1 (0.2) UZRI-AA551A Mounting kit for peripheral unit 4.5 (9.9)
USZS-AI900A Measurement Z score kit 0.1 (0.2) UZRI-AA000A Mounting kit for Peripheral units 2.4 (5.3)
USSI-AA550A Smart Area Indication OB kit 0.1 (0.2) UZFS-004A Foot switch 0.6 (1.3)
USFH-AI600A Fetal Heart MPI Measurement kit 0.1 (0.2) UZGW-008A Gel warmer 1.0 (2.2)
USHI-AA550A CHI kit 0.1 (0.2) UZMK-AI900A Transducer cable hanger kit 0.9 (2.0)
USCQ-AI900A CHI-Q kit 0.1 (0.2) UZBK-AI900A Transducer Holder kit 0.3 (0.7)
USCQ-AI901A Fitting Curve kit 0.1 (0.2) UIWL-A500A Wireless LAN kit 0.5 (1.1)
USEL-AA551A Elastography-FLR kit 0.1 (0.2) UIWL-AI900A Wireless LAN kit 0.5 (1.1)
USEL-AA550A Elastography kit 0.1 (0.2) UZPH-AI900A EV/ER Transducer Holder kit 1.6 (3.5)
USSW-AA550A Shear Wave kit 0.1 (0.2) UZPH-AA000A Transducer Holder kit (left) 0.8 (1.8)
USSW-AA551A Shear Wave Hard kit 0.1 (0.2) UIKB-AI900A Keyboard kit 1.1 (2.4)
USFN-AA550A Smart Fusion kit 0.1 (0.2) UZKF-AI900A Local Language Key-Top kit 0.5 (1.1)
USSN-AI600A Smart Navigation kit 0.1 (0.2) UZKG-AI900A Local Language Key-Top kit 0.5 (1.1)
UIFR-AA550A Magnetic Generator kit 15.0 (33.0) UZKI-AI900A Local Language Key-Top kit 0.5 (1.1)
UIFR-A501A Sensor kit for Fusion unit 0.1 (0.2) UZKS-AI900A Local Language Key-Top kit 0.5 (1.1)
USFN-AI901A Auto Registration kit 0.1 (0.2) UZKD-AI900A Local Language Key-Top kit 0.5 (1.1)
UZWT-A500A Fusion Pole Cart 26 (57.3) UZKN-AI900A Local Language Key-Top kit 0.5 (1.1)
UAFS-001A Mounting kit for fusion sensor 0.1 (0.2) UZKW-AI900A Local Language Key-Top kit 0.5 (1.1)
UAFS-002A Mounting kit for fusion sensor 0.1 (0.2) UZKR-AI900A Local Language Key-Top kit 0.5 (1.1)
UAFS-003A Mounting kit for fusion sensor 0.1 (0.2) UIOM-001A OLED Monitor unit 4.1 (9.0)
UAFS-004A Mounting kit for fusion sensor 0.1 (0.2) UIVP-AA550A VIDEO unit 1.23 (2.7)
UAFS-005A Mounting kit for fusion sensor 0.1 (0.2) USDB-AI900A DataBase for External HDD kit 0.1 (0.2)
UAFS-006A Mounting kit for fusion sensor 0.1 (0.2) UZHI-AI900A Mounting kit for External HDD kit 1.0 (2.2)
UAFS-007A Mounting kit for fusion sensor 0.1 (0.2) UZMK-AI901A ECG Cable Hanger kit 0.1 (0.2)
26
MPDUS0085EAN
Model name Name of component Mass [kg] (lb) Model name Name of component Mass [kg] (lb)
UIUB-AI900A Panel USB Port kit 0.1 (0.2) PVT-350BTP Convex array biopsy transducer 0.95 (2.1)
UEBT-AA550A Battery unit 13 (28.7) PVT-375BT Convex array transducer 0.95 (2.1)
UZTB-Al900A Track Ball kit 0.2 (0.4) PVT-375SC Convex array transducer 0.95 (2.1)
UZPT-001A Palm Controller kit 0.2 (0.4) PVT-382BT Convex array transducer 0.8 (1.8)
UITP-AA550A Tilt panel kit 1.5 (3.3) PVT-475BT Convex array transducer 0.8 (1.8)
UIMS-AA550A Dynamic Micro Slice unit 4.3 (9.5) PVT-482BT Convex array transducer 0.75 (1.7)
USMS-AA550A Dynamic Micro Slice kit 0.1 (0.2) PVT-574BT Convex array transducer 0.75 (1.7)
USHE-AI900A Online Help kit 0.1 (0.2) PVT-674BT Convex array transducer 0.9 (2.0)
USPA-AI900A Protocol Assistant kit 0.1 (0.2) PVT-675MVL Convex array transducer 1.1 (2.4)
USPA-AI901A MSK Protocol Movie kit 0.1 (0.2) PVT-675MVS Convex array transducer 1.05 (2.3)
USDL-AI900A Multi Parametric Report kit 0.1 (0.2) PVT-681MVL Endocavitary transducer 1.15 (2.5)
USSM-AI900A Security Management kit 0.1 (0.2) PVT-712BT Convex array transducer 0.8 (1.8)
USRA-AI900A RADS kit 0.1 (0.2) PVT-745BTF Convex array transducer 0.8 (1.8)
USIO-AI900A IOTA kit 0.1 (0.2) PVT-745BTH Convex array transducer 0.8 (1.8)
USNA-AI900A Network Storage kit 0.1 (0.2) PVT-745BTV Convex array transducer 0.78 (1.7)
USTR-AI900A Tricefy Access kit 0.1 (0.2) PVT-770RT Endocavitary transducer 2.0 (4.4)
UIAG-001A ApliGate kit 0.1 (0.2) PVT-781VT Endocavitary transducer 0.97 (2.1)
USAG-001A ApliGate Soft kit 0.1 (0.2) PVT-781VTE Endocavitary transducer 0.97 (2.1)
USWC-AI900A ApliCam kit 0.1 (0.2) PVL-715RST Endocavitary transducer 1.04 (2.3)
USRI-AI600A Reference Image kit 0.1 (0.2) PLT-308BTP Linear array biopsy transducer 0.8 (1.8)
USMB-AI900A Breast Scan Guide kit 1.0 (2.2) PLT-704SBT Linear array transducer 0.9 (2.0)
USSB-AI900A Smart Body Mark kit 0.1 (0.2) PLT-705BT Linear array transducer 0.85 (1.9)
UIBP-AA550A Breast Package kit 15.0 (33.1) PLT-705BTF Linear array transducer 0.78 (1.7)
USBP-AA550A Breast Package Soft kit 0.1 (0.2) PLT-705BTH Linear array transducer 0.78 (1.7)
USPS-AA000A Transducer connector 0.1 (0.2) PLT-1005BT Linear array transducer 0.85 (1.9)
USAO-AA000A Advanced Software kit 2.1 (4.6) PLT-1202BT Linear array transducer 0.75 (1.7)
USPO-AA000A Premium Software kit 2.1 (4.6) PLT-1204BT Linear array transducer 0.85 (1.9)
HVO-550MD/ PLT-1204BX Linear array transducer 0.92 (2.0)
DVD video recorder 3.2 (7.1)
FDH PLI-2002BT Linear array transducer 0.75 (1.7)
UP-D711MD/AC B/W printer 1.0 (2.2) Multi-plane transesophageal
PET-508MA 1.2 (2.6)
UP-D898MD B/W printer 2.6 (5.7) transducer
P95DW B/W printer 2.6 (5.7) Multi-plane transesophageal
PET-609MA 1.3 (2.87)
P95DW-DC B/W printer 1.6 (3.5) transducer
PST-50BT Phased array transducer 0.8 (1.8) PET-805LA Linear array transducer 1.17 (2.6)
PST-65BT Phased array transducer 0.73 (1.61) PC-20M CW Doppler pencil transducer 0.085 (0.2)
PC-50M CW Doppler pencil transducer 0.08 (0.2)
27
MPDUS0085EAN
https://global.medical.canon
©Canon Medical Systems Corporation 2019-2023. All rights reserved.
Design and specifications are subject to change without notice.
Model number: CUS-AA000 MPDUS0085EAN V5.0 2023-03 CMSC/Produced in Japan
Canon Medical Systems Corporation meets internationally recognized standards for Quality Management System ISO 9001, ISO 13485.
Canon Medical Systems Corporation meets the Environmental Management System standard ISO 14001.
Windows and AppLocker are registered trademarks of Microsoft Corporation in the United States and / or other countries.
omniTRAX and VirtuTRAX are trademarks of CIVCO Medical Solutions.
Tricefy is a trademark of Trice Imaging, Inc.
DICOM is the registered trademark of the National Electrical Manufacturers Association for its Standards publications relating to digital
communications of medical information.
This document may include trademarks or registered trademarks of their respective owners.