You are on page 1of 4

JACC: CARDIOVASCULAR INTERVENTIONS VOL. -, NO.

-, 2016
ª 2016 BY THE AMERICAN COLLEGE OF CARDIOLOGY FOUNDATION ISSN 1936-8798/$36.00

PUBLISHED BY ELSEVIER http://dx.doi.org/10.1016/j.jcin.2016.09.012

EDITORIAL COMMENT

Comparing Hemodynamics of
Contemporary Mechanical
Circulatory Support
Moving from In Silico to In Vivo Results*

Morton J. Kern, MD, MSCAI, Arnold H. Seto, MD, MPA

T he degree of myocardial recovery determines


the function and ultimately the survival of a
patient in cardiogenic shock. Therapeutic
efforts to improve myocardial function include inten-
involving LV unloading and other performance
characteristics. To review, the intra-aortic balloon
pump inflates in diastole, generating an augmented
pressure pulse and deflates abruptly in systole
sive pharmacotherapy (inotropes, vasopressors), and reducing LV afterload. The net hemodynamic effects
use of mechanical circulatory support (MCS) devices, of intra-aortic balloon pump produce (to variable
including the intra-aortic balloon pump, the left ven- degrees) an increase in mean arterial pressure, coro-
tricular (LV)-to-aorta axial pump catheter (Impella CP nary perfusion, and mild decrease in myocardial
[IMP]) and the left atrial (LA)-to-femoral artery pump work. The magnitude of these effects is smaller than
(TandemHeart [TH]). All act through different mecha- those of either the IMP or TH and requires a contracting
nisms, and to different degrees, to unload the heart, heart to function effectively. The IMP works through a
increase blood pressure, cardiac output and systemic catheter-mounted axial pump, drawing blood from the
perfusion, while reducing myocardial work and oxy- LV and expelling it into the ascending aorta. The TH
gen consumption. pulls blood from the left atrium through a trans-septal
The hemodynamics of MCS devices are known to inserted large (22-F) inflow cannula to an external
us predominately from computer simulations (“in centrifugal pump, returning the blood to the arterial
silico” models) (1,2), in vivo animal studies (3,4), and circulation through the femoral artery. Both IMP and
series of cardiogenic shock patients (5,6). In practice, TH substantially reduce LV stroke volume, LV loading,
a specific device is selected based both on its pre- and myocardial work, and increase cardiac output and
sumed hemodynamic power as well as the patient’s systemic pressure. Although the hemodynamics of
clinical features (e.g., status of the aortic valve, pe- IMP and TH have been explored extensively by com-
ripheral vascular disease, etc.) and technical chal- puter simulations (1,2), it remains largely unknown to
lenges of device implantation (e.g., cannula size, ease what degree the MCS devices perform under in vivo
and speed of insertion, and availability of a trained conditions, in the same subject with the same flow
team on call). rates. Moreover, hemodynamic differences from
Because of the different mechanisms of action, we computer simulations would be anticipated since any
can expect unique hemodynamic profiles, especially in silico model cannot not completely replicate the
more complex behavior of the intact, in vivo animal
cardiovascular system (7).
*Editorials published in JACC: Cardiovascular Interventions reflect the Moving from in silico to in vivo, Weil et al. (8)
views of the authors and do not necessarily represent the views of JACC: compared the hemodynamics of the IMP and TH
Cardiovascular Interventions or the American College of Cardiology.
from a closed chest animal model of myocardial
From the Veterans Administration Long Beach Health Care System, infarction. Using a LV pressure–volume (PV)
University of California, Long Beach, California. Dr. Kern is a speaker for
catheter, PV loops were obtained before and after a
Abiomed Inc., St. Jude Medical, Volcano Therapeutics, Acist Medical,
Opsens Inc., and Heartflow Inc. Dr. Seto has reported that he has no moderate-sized myocardial infarction was induced
relationships relevant to the contents of this article to disclose. with a 2-h circumflex coronary artery occlusion.
2 Kern and Seto JACC: CARDIOVASCULAR INTERVENTIONS VOL. -, NO. -, 2016
Hemodynamics of Contemporary Mechanical Circulatory Support - 2016:-–-

more effective decreases in myocardial oxygen de-


T A B L E 1 Comparative Hemodynamic Effects of Percutaneous Ventricular
Assist Devices
mand would lead to greater myocardial recovery after
infarction, or more support during cardiogenic shock.
IMP TH
This well-performed study, comparing 2 commonly
Heart rate (beats/min) ¼ ¼
used MCS devices head to head for the first time in
Mean aortic pressure (mm Hg) ¼ ¼
this way, suggests that, at comparable device flow
Aortic pulse pressure (mm Hg), 34  2 to 27  4 34  2 to 14  1*†
control to device activation rates, the degree of LV unloading depends on
LV end-diastolic pressure (mm Hg) 15  1 to 11  1* 15  2 to 7  4*† whether blood is withdrawn directly from the LA or
Pulmonary capillary wedge pressure 15  2 to 13  2 15  2 to 9  1* the LV. The IMP draws blood from the LV and
(mm Hg)
decreased LV end-diastolic volume and LV end-
End-systolic elastance (mm Hg/ml) ¼ ¼
LV stiffness constant (b) þ þ diastolic pressure and maintained arterial pressure,
LV dP/dtmax reduction - þþ but in this study did not affect PV loop-derived
LV stroke work reduction þ þþþ indices of myocardial work significantly. In contrast,
LV pressure volume area reduction þ þþþ TH withdrawal of blood from the LA also reduced LV
LV preload recruitable stroke work - þ end-diastolic volume and LV end-diastolic pressure,
reduction
and maintained arterial pressure while also reducing
Values are mean  SEM. *p < 0.05 vs. post-MI; †p < 0.05 vs. Impella CP (IMP). Adapted from native LV stroke volume, stroke work, dP/dtmax, PV
Weil et al. (8).
area, and preload-recruitable stroke work consistent
þ ¼ improved compared with control MI state; ¼ ¼ equivalent change between devices; - ¼ no
improvement compared to control; LV ¼ left ventricular; TH ¼ TandemHeart. with enhanced LV unloading.
Why should the findings of Weil et al. (8) differ
from prior studies? Although neither a computer
Although frank cardiogenic shock was not produced, model nor the induced animal infarction model
myocardial infarction shifted the PV loop to the right exactly duplicates the human cardiogenic shock
due to an increased LV end-diastolic pressure state, the in vivo conditions incorporate a number of
accompanied by reduced systolic pressure with influences on myocardial function, arterial imped-
mildly reduced stroke volume. Both MCS devices ance and afterload, and peripheral resistance that can
were then sequentially placed, with the order ran- only be estimated in a more rudimentary fashion in
domized and the goal flow rates matched. computer simulations. In contrast, the present study
Both MCS devices maintained the aortic pressure did not truly induce a state of cardiogenic shock.
but, compared to IMP, the TH had a greater reduction Because unloading with IMP occurs most effectively
in LV end-diastolic pressure (but not volume), native when cardiac output is low and LV end-diastolic
LV stroke volume, dP/dt max , stroke work, PV area, pressure and volume are increased (and poorly
and preload stroke work slope (Table 1). In short and when these states are not present), the present model
at odds with some prior reports (2–5), TH unloaded may have unfairly tilted the scale against IMP.
and rested the heart to a greater degree than IMP. Of Why aspiration of similar flow rates from the left
particular interest and unique to this study was that atrium compared with the LV would have such
the hemodynamics and PV loops of the 2 MCS devices different hemodynamic effects remains unclear.
were not only different from each other but also diff- Potentially, this result reflects the greater capacitance
erent from the computer model predictions (Figure 1). of the LA (compared with the LV) to release volume to
The investigators deserve our compliments for their the MCS device, particularly in euvolemic, nonshock
rigorous experimental methods. The greater decrease states, but this would presumably have been evident
in native stroke work by TH suggests a more powerful from the flow rates measured. Alternatively, the au-
volume unloading of the heart occurs despite the thors suggest that, by decreasing the LV end-diastolic
relative increase in native end-systolic volume pressure to a greater extent, LA withdrawal may
compared with IMP as demonstrated by the different improve coronary blood flow. However, coronary
shapes of the in vivo PV loops (Figure 1). The IMP blood flow was not measured in this study and would
maintained end-systolic pressure, reduced stroke not explain how LV end-diastolic pressure is reduced
volume and shifted the PV loop leftward, whereas the more effectively.
TH increased end-systolic pressure with a pronounced The IMP and TH are powerful MCS devices that are
reduction in stroke volume and preload. As a result, frequently used for high-risk percutaneous coronary
the TH produced a greater decrease in PV area (a well- interventions and cardiogenic shock. In clinical prac-
validated index of myocardial oxygen consumption) tice, the selection of a percutaneous MCS device de-
than the IMP in the animal model, a contrasting pends not only on the hemodynamic power of the
observation from the computer models. Conceivably, device, but also and more important on the specific
JACC: CARDIOVASCULAR INTERVENTIONS VOL. -, NO. -, 2016 Kern and Seto 3
- 2016:-–- Hemodynamics of Contemporary Mechanical Circulatory Support

F I G U R E 1 Ventricular Effects of LA-to-Arterial MCS and LV-to-Aorta MCS

(Top) Computer simulation of ventricular effects of left atrial (LA)-to-arterial mechanical circulatory support (MCS; TH; left) and left
ventricular (LV)-to-aorta MCS (IMP; right) showing flow-dependent changes in pressure–volume loops, reductions in end-diastolic pressures,
increased end-systolic volume, and decreased LV stroke volume (from Burkhoff et al. [9]). (Bottom) PV loops for TH and IMP from the in vivo
animal model of Weil BR et al. (8) (IMP in blue and TH in green) compared with the corresponding unassisted post-myocardial infarction
baseline time-point (red). Both devices shifted the PV loop to the left and decreased LV end-diastolic pressure, but TH produced a greater
reduction in native LV stroke volume. Ao ¼ aorta; CGS ¼ cardiogenic shock; LVD ¼ left ventricular assist device.

clinical scenario. The TH takes an average of 45 to 60 patient in need of LV mechanical support, the operator
min to insert, compared with <10 min for the IMP, must choose the device that is most clinically and he-
making the TH less useful in patients with unstable modynamically appropriate.
cardiogenic shock (9). The risk of major bleeding with
the TH is also higher than IMP (25% vs. 6%) (4). The REPRINT REQUESTS AND CORRESPONDENCE: Dr.
carefully measured hemodynamic in vivo differences Morton J. Kern, Department of Medicine, Veterans
of LV performance by Weil et al. (8) shift conventional Administration Long Beach Health Care System,
wisdom about the MCS device power order and should Building 1, 5901 East 7th Street, Long Beach,
bring new attention to the TH. Nonetheless, for each California 90822. E-mail: Morton.kern@va.gov.

REFERENCES

1. Rihal CS, Naidu SS, Givertz MM, et al. 2015 SCAI/ 2. Burkhoff D, Cohen H, Brunckhorst C, et al., therapy with intraaortic balloon pumping for
ACC/HFSA/STS clinical expert consensus state- TandemHeart Investigators Group. A randomized treatment of cardiogenic shock. Am Heart J 2006;
ment on the use of percutaneous mechanical cir- multicenter clinical study to evaluate the safety 152:469.e1–8.
culatory support devices in cardiovascular care. and efficacy of the TandemHeart percutaneous 3. Burkhoff D, Naidu SS. The science
J Am Coll Cardiol 2015;65:e7–26. ventricular assist device versus conventional behind percutaneous hemodynamic support:
4 Kern and Seto JACC: CARDIOVASCULAR INTERVENTIONS VOL. -, NO. -, 2016
Hemodynamics of Contemporary Mechanical Circulatory Support - 2016:-–-

a review and comparison of support strategies. cannulation for acute circulatory support in a circulatory support devices in a porcine model
Catheter Cardiovasc Interv 2012;80:816–29. bovine model of left heart injury. ASAIO J 2015;61: of acute myocardial infarction. J Am Coll Car-
301–6. diol Intv 2016;xx:xxx–xxx.
4. Naidu SS. Novel percutaneous cardiac assist
devices: the science of and indications for hemo- 7. Atkinson TM, Ohman EM, O’Neill WW, Rab T, 9. Kar B, Adkins LE, Civitello AB, et al. Clinical
dynamic support. Circulation 2011;123:533–43. Cigarroa JE. A practical approach to mechanical experience with the TandemHeart percutaneous
5. Burkhoff D. Hemodynamic support: science and circulatory support in patients undergoing percu- ventricular assist device. Tex Heart Inst J 2006;33:
evaluation of the assisted circulation with percu- taneous coronary intervention: an interventional 111–5.
taneous assist devices. Interventional Cardiology perspective. J Am Coll Cardiol Intv 2016;9:871–83.
Clinics 2013;2:407–16. 8. Weil BR, Konecny F, Suzuki G, Iyer V,
6. Kapur NK, Paruchuri V, Pham DT, et al. Hemo- Canty JM. Comparative hemodynamic effects KEY WORDS cardiac output, hemodynamics,
dynamic effects of left atrial or left ventricular of contemporary percutaneous mechanical ventricular assist device

You might also like