MML194x 3
INTRODUCTION
The PHILIPS Medical Grade MML194x is a monochrome active matrix, liquid crystal display exclusively designed for medical
imaging applications.
The monitor can be used for X-ray and PACS and other medical systems requiring a very high level of image quality for
examination and control purpose and can accept an analogue signal input as well as a digital input.
These monitors, certified/registered by the safety agencies/regulatory authorities as model number 19LCD-XR, follow the
provisions of Medical Directive:
93/42/EEC, Amended by 2007/47/EC (Medical device Class I)
The monitors have been designed to meet the following medical safety requirements, including;
EN 60601-1:1990 + A1:1993 + A2:1995 (2nd Ed.)
EN 60601-1:2006 (3rd Ed.)
IEC 60601-1:1988 + A1:1991 + A2:1995 (2nd Ed.)
IEC 60601-1:2005 + Corr.1: 2006 + Corr.2: 2007 (3rd Ed.)
UL60601-1 1st Ed, 2006-04-26
ANSI/AAMI ES 60601-1: 2005
CAN/CSA-C22.2 No. 601.1-M90, 2005
CAN/CSA-C22.2 No. 60601-1: 2008
• The monitor is intended for indoor use
• Class I Equipment, according to the type of protection against electric shock (*)
• Detachable supply connection
• Continuous Operation, according to the mode of operation,
• Pollution degree classification 2
• Degree of protection against ingress of water: IP20 / IP21
(IP21 with protective screen: monitor tested with a tilt of +5° towards the operator / -15° towards the back side)
• Stationary, in case of “slave” monitor placed in Control Room, not fixed on wall; Fixed, when mounted in MCS.
• Do not use in areas with flammable anesthetics
• Do not use in oxygen rich environments
• The monitor is not intended to be sterilized
• The monitor has been designed to be used in landscape position with a tilt of -15° (backward) and +5° (forward)
(*) To avoid the risk of electric shock, this equipment must only be connected to a supply mains with protective earth
(**) This equipment does not have a gas-sealed electronics enclosure and could ignite any flammable or explosive gases present
in its environment.
4 MML194x
Copyright ©
This manual is covered by Copyright with all rights reserved. Under the copy rights law, this manual may not be copied, in whole
or part, without written consent of Philips. Under the law, copying includes translating into another language or format.
FCC Notice
This equipment has been tested and found to comply with the limits for a class B digital device, pursuant to Part 15 of FCC Rules.
These limits are designed to provide reasonable protection against harmful interference in a residential installation.
This equipment generates, uses and can radiate radio frequency energy and, if not installed and used in accordance with the
instructions, may cause harmful interference to radio communications.
However, there is no guarantee that interference will not occur in a particular installation.
If this equipment does cause harmful interference to radio or television reception, while can be determined by turning the
equipment off and on, the User is encouraged to try to correct the interference by one or more of the following measures:
- Re orient or relocate the receiving antenna.
- Increase the separation between the equipment and receiver.
- Connect the equipment into an outlet on a circuit different from that to which the receiver is connected.
- Consult the dealer or an experienced radio TV technician for help.
Modification
The FCC requires the User to be notified that any changes or modifications made to this device that are not expressly approved by
the manufacturer may void the User’s authority to operate the equipment.
Cables
Connections to this device must be made with shielded signals cables with metallic RFI / EMI connector hoods to maintain
compliance with FCC Rules and Regulations.
The lighting flash with arrowhead symbol is intended to alert the User about the presence of not insulated “dangerous
voltages” within the product’s enclosure that may be sufficient magnitude to constitute a risk of electrical shock to
people.
The exclamation mark is intended to alert the User about the presence of important operating and maintenance
(servicing) instructions in literature accompanying the appliance.
CAUTION
RISK OF ELECTRICAL SHOCK
DO NOT OPEN
ATTENTION
RISQUE DE CHOC ELECTRIQUE
NE PAS OUVRIR
WARNING:
TO PREVENT DAMAGE WHICH MAY RESULT IN FIRE OR SHOCK HAZARD,
DO NOT EXPOSE THIS APPLIANCE TO RAIN OR EXCESSIVE MOISTURE.
MML194x 5
PRECAUTIONS
Installation
Power cord: due to the different power cable lengths necessary at system level, the power cable is not provided with this LCD
monitor.
in Europe, use a power cord H05VV-F or H05VVH2-F rated 0.75mm2 250V, with a Plug rated 10A 250V. At least one
European safety mark is required.
in US and Canada, use a power cord with plug “hospital grade”, provided with instructions to indicate that grounding
reliability can be achieved only when the equipment is connected to an equivalent receptacle marked hospital only or
hospital grade. Rating, min 18 AWG SJT 60°C 300V, Plug NEMA 5-15 P, Connector IEC 320. The cURus safety mark is
required.
Contact the local dealer if you have doubt on the right cord set to be used.
Protective earth: to avoid the risk of electric shock, this equipment must only be connected to a supply mains with protective
earth. (Class I Equipment)
Cables: when the monitor is assembled in the medical system, take care of the anchorage of all cables, to avoid unwanted
detachment
Connections: any external connection with other peripherals must follow the requirements of clause 16 of IEC60601-1 3rd ed. or
Table BBB.201 of IEC 60601-1-1 for the medical electrical systems.
6 MML194x
Serviceability
There are no User serviceable parts within the 19” LCD monitor. Attempting to open or disassemble the monitor may expose the
User to hazardous voltages, risk of excessive temperatures, fire or explosion from enclosed energy sources, or impairment of
function. This product is intended to be installed in a medical system only by trained system integrator.
Biological Hazard and Returns
The structure and the specifications of this device as well as the materials used for manufacturing make it easy to wipe and clean
and therefore suitable to be used for various applications in hospitals and other medical environments, where procedures for
frequent cleaning are specified. However, normal use shall exclude biological contaminated environments, to prevent spreading of
infections. Therefore use of this device in such environments is at the exclusive risk of Customer. In case this device is used where
potential biological contamination cannot be excluded, Customer shall implement the decontamination process as defined in the
latest edition of the ANSI/AAMI ST35 standard on each single failed Product that is returned for servicing, repair, reworking or
failure investigation to Seller (or to the Authorized Service Provider). At least one adhesive yellow label shall be attached on the
top site of the package of returned Product and accompanied by a declaration statement proving that the Product has been
successfully decontaminated. Returned Products that are not provided with such external decontamination label, and/or whenever
such declaration is missing, can be rejected by Seller (or by the Authorized Service Provider) and shipped back at Customer
expenses.
MML194x 7
For the design of MML194x LCD monitors, special attention has been placed to the environmental impact of the product:
packaging, product’s weight and energy consumption have been optimized vs. its performances. Then, please consider the
following notes in the disposal of the product.
The monitor’s packaging is re-usable, please retain the packaging for future use.
Disposal of the LCD monitor
Do not dispose of this monitor with general household waste.
The fluorescent tubes used in this LCD monitor contain mercury. Disposal of this monitor must be carried out in accordance with
the regulations of your local sanitation authority.
WARNING
The fluorescent lamps in the liquid crystal
display (LCD) contain mercury.
Do not put it in trash that is disposed of in
landfills.
Dispose of it as required by local ordinances
or regulations.
8 MML194x
Product disposal information
Directive 2002/95/EC on the restriction of certain hazardous substances in electrical and electronic equipment (RoHS)
According to what declared by our components supplier, this product is RoHS compliant.
Power management
If the power management function of the computer connected to the MML194x display is enabled, the monitor turns to and from
Stand-by automatically.
The User can control power management features from the computer.
MODE LED POWER CONSUMPTION (typical)
Power ON Green 55W (*)
Stand-by Orange 8.3W
Switch OFF Not illuminated 0W
(*) Factory settings
MML194x 9
TECHNICAL INFORMATION
Display Panel 19” (482 mm) SXGA active matrix monochrome TFT/LCD
Video interface
Analog Input: • BNC connection for Composite Video Signal
• DSUB-15 connection for Separate Sync Video Signal or SOG Video Signal.
• Analog on DVI (DVI-I connector)
Digital Input: • DVI-D connection for TMDS signal link standard version.
Analog Output: • BNC connection for Loop Through Composite Video Signal.
10 MML194x
DIMENSIONS
MML1942-PER
MML194x 11
MML1941-PCR
12 MML194x
INSTALLATION
Before unpacking the monitor, prepare a suitable workspace: a stable and level surface near a grounded wall outlet in
an area that is relatively free of glare from sunlight or other sources of bright light. Do not block the cooling vents.
NOTE: before installing this monitor, please refer to the User’s Guide of the attached computer and/or video adapter
to check if these devices require any change of setting.
Positioning
WARNING:
WHEN POSITIONING THIS EQUIPMENT, ENSURE THAT THE MAINS PLUG AND SOCKET ARE EASILY
ACCESSIBLE
While unpacking the monitor, inspect all package contents for any eventual shipping damage that could cause a fire or
shock hazard.
Immediately report any shipping damage to the carrier or transportation company and contact customer service for
assistance. Keep all packaging material in case you need to ship, store or return the monitor.
After unpacking the monitor, make sure the following items are included:
MML1942-PER
• Monitor
• This User’s Manual
• Warning Note DMR147472
• User Manual MML194x Series CCC version 4519 206 1101.x
• 4 VESA Screws M4x25mm 4519 200 40451
• 4 VESA Screws M4x25mm 4519 200 40181
• 4 VESA Screws M4x30mm 4519 200 40441 (mounted in VESA holes)
• 4 Lock washer 4519 200 40191 (mounted in VESA holes)
• 4 Plain washer 4519 200 40201 (mounted in VESA holes)
MML1941-PCR
• Monitor
• This User’s manual
• RS232 to RJ11 (RJ10) 4p4c (4 pins, 4 connectors) interface cable
• Warning Note DMR147472
MML194x 13
MOUNTING INSTRUCTION
Pedestal (MML1941-PCR)
With the built-in pedestal the monitor can be tilted for a most comfortable viewing angle as shown in figure here
below.
The arm with VESA (100mm x 100mm) interface must be fixed to the monitor with the 4 screws already present on
the back cover.
Note: If not available, use M4 x 25 screws, spring washers and plain washers (see figure above).
14 MML194x
CONNECTING INSTRUCTION
MML194x 15
Connecting AC Power
Plug the receptacle end of the cord set into the AC power adapter, then plug the power connector of the adapter into
the power port on the monitor.
This power port is located on the back of the monitor (see figure below).
The power cord set can be fixed to cabinet with proper cable clamp and screw (see figure below).
16 MML194x
Connecting analog video (SXGA)
Connect the supplied video cable from the monitor to the computer’s video port (see figure below) only when the
computer and the monitor are turned off.
Make sure the video cable connector is securely connected to the video port on the computer.
MML194x 17
Connecting composite video - BNC
Connect a BNC video cable from the monitor to the computer’s video port (see figure below) only when the computer
and the monitor are turned off.
For example: connection between MML1942-PER and Image source via BNC cable.
18 MML194x
INPUT INSTRUCTION
Mains switch
12A - 125/250V~
MML194x 19
CONTROL PANEL DESCRIPTION
On the front bottom part of the unit the following controls are placed:
KEY NAME SYMBOL DESCRIPTION
- BRIGHTNESS ADJ+ ☼+ Increase brightness
- BRIGHTNESS ADJ- ☼– Decrease brightness
- MENU SELECT - ▼ Move cursor in OSD
- MENU SELECT + ▲ Recall brightness Preset/Move cursor in OSD
- ENTER I (Store/Select)
- LOGO PHILIPS Turn keyboard backlight ON (touch PHILIPS logo for 2.5s)
20 MML194x
BRIGHTNESS KEY FUNCTIONS
KEY NAME SYMBOL DESCRIPTION
- BRIGHTNESS ADJ- ☼– It temporarily decrements the current brightness value, the LED turns
RED if the current brightness value is different from the User
brightness preset. Current brightness can be stored entering the
Brightness preset item in the OSD Service menu.
At power-on the stored brightness User preset value will be reloaded
- BRIGHTNESS ADJ+ ☼+ It temporarily increments the current brightness value, the LED turns
RED if the current brightness value is different from the User
brightness preset. Current brightness can be stored entering the
Brightness preset item in the OSD Service menu.
- MENU SELECT - ▼ It moves cursor in OSD menu from left to right and from up to down.
If the freeze feature is enabled, it toggles the image from normal to
frozen and vice versa. When the image is frozen, the LED is orange
and blinking.
- MENU SELECT + ▲ It moves cursor in OSD menu from right to left and from down to up.
When used to restore the User brightness preset, the LED turns green
and the brightness pop-up window turns up to show the User default
settings (see figure below, on the right)
When no video signal is available for the current video source, after
5s the MML194x enters the stand-by mode (the panel is black and the
front LED is orange). Touch ENTER for 1s to switch back to Normal
mode
- LOGO PHILIPS When touched for at least 3s, it turns the backlight illumination of the
keypad ON and OFF (toggle mode).
MML194x 21
Keyboard Lock/Unlock
To prevent unintentional access to the OSD menu, the keyboard can be locked.
If the Keyboard Status is locked, any access either to OSD or to the brightness control keys are inhibited.
- By touching any OSD key, the dialogue here below appears
Access to the OSD menu is granted only after touching the following key sequence:
• BRIGHTNESS ADJ-
• BRIGHTNESS ADJ+
• BRIGHTNESS ADJ+
• MENU SELECT +
The Keylock dialogue box disappears any time a wrong key in the sequence is touched.
For every right touched key, the corresponding empty circle in the row becomes embossed.
To enable/disable the keyboard lock see the corresponding page in the OSD menu description.
22 MML194x
OSD - ON SCREEN DISPLAY
MENU STRUCTURE
│ MENU │ SUBMENU │ ITEM │ NOTE │
MML194x 23
OSD KEY FUNCTIONS
Function Key Description
Enter It activates the OSD MAIN MENU if this is not active, otherwise selects and
executes a desired OSD Special Function.
Adjust When main menu or sub-menu is active, it allows to move or scroll to the
next (down) item or adjust the parameter.
Adjust When main menu or sub-menu is active, it allows to move or scroll to the
previous (upper) item or adjust the parameter.
Touching “Enter” button the OSD MENU appears, touching again “Enter” button the current menu is confirmed and the first
Sub-menu is highlighted
Touching “Adjust +” or “Adjust -” button, the highlighted selection of menu item moves downward/ upward.
Touching “Enter” button, the current menu window disappears and is replaced by the new sub-menu windows.
2) Touch the Adjust + or Adjust - button until the desired OSD Menu is highlighted.
3) When the desired OSD menu is highlighted, touch the Enter and the first Sub-menu is highlighted
4) Touch Adjust + or Adjust - button to adjust the value or to select the desired sub-menu and touch again Enter to
confirm.
5) Touch again Adjust + or Adjust - button to exit or to select another Sub-menu.
MAIN It shows the general features menu (this menu is always enabled)
IMAGE ADJUST It shows the image features menu (it is disabled if no or bad video
signal is recognized)
ADVANCED It shows the advanced features menu (it is disabled if no or bad video
signal is recognized)
SERVICE It shows the service feature menu (it is disabled if no or bad video signal is recognized)
24 MML194x
Main menu
After the activation of OSD Main menu the following window appears
Information It shows the hardware and firmware status (operating hours and release
number)
MML194x 25
Information
1) Touch Enter button when the Information is highlighted on Main menu.
The Information window appears:
MML19 Model
D1.15H - 04 - LP03 Firmware Release
h00002 Operating Hours
xxxx Warning 1
xxxx Warning 2
3) Touch Enter button to return to Main Menu.
4) Touch the Adjust + or Adjust - button to exit or to select
another item
Note:
Warning 1 can display:
Backlight NC = Backlight Not Calibrated
DVI 5V Over Current = DVI 5 Volt disabled for Over Current
Warning 2 can display:
BaudRate = XXXXX
EOL Backlight ON = End of Life of the Backlight is ON
Language
1) Touch Enter button when the Language is highlighted on Main Menu.
The Language window appears:
The ON SCREEN DISPLAY shows the available languages.
The default is English, but you can select Français, Deutsch, Español or
Italiano.
2) Touch the Adjust + or Adjust - button until the desired item is
highlighted.
3) Touch the Enter button to confirm your selection.
4) Touch the Adjust + or Adjust - button to exit or to select another item
Video Source
1) Touch Enter button when the Video Source is highlighted on Main
Menu.
The Video Source window appears:
26 MML194x
Key Lock
1) Touch Enter button when the Key Lock is highlighted on Main Menu.
The Key Lock window appears:
The ON SCREEN DISPLAY shows the following available choices:
- Disabled
- Enabled
2) Touch the Adjust + or Adjust - button until the desired item is
highlighted.
3) Touch the Enter button to confirm your selection.
4) Touch the Adjust + or Adjust - button to exit or to select another item
OSD Time-out
1) Touch Enter button when the OSD Time-out is highlighted on Main
Menu.
The OSD Time-out window appears:
The ON SCREEN DISPLAY shows the following available choices:
- 10 sec - 60 sec
- 20 sec - 90 sec
- 30 sec - 120 sec
- 45 sec - Disabled (Max 5 min. - to avoid image sticking)
2) Touch the Adjust + or Adjust - button to choose the time of activation
of the OSD-Menu.
3) Touch the Enter button to confirm your selection.
4) Touch the Adjust + or Adjust - button to exit or to select another item
MML194x 27
OSD Vert. Pos.
1) Touch Enter button when the OSD Vert. Pos. is highlighted on Main
Menu.
The OSD Vert. Pos. window appears:
2) Touch the Adjust + or Adjust - button until the desired value appears.
3) Touch the Enter button to confirm your selection.
4) Touch the Adjust + or Adjust - button to exit or to select another item
28 MML194x
Image Adjust
After the activation of Image Adjust menu the following window appears:
Hor. Position Manual adjust of horizontal position (only for analog inputs)
Vert. Position Manual adjust of vertical position (only for analog inputs)
Image Size Image format selection
LUT Linear, Dicom, CAL. TG3-PMS
Black Value Native, User
Color Emulation Activate/deactivate the emulation of Color monitors on
Monochrome
Border Color Grey level selection for the frame around the image (when available)
Freeze If enabled in the menu and activated with the Freeze ON/OFF
button, it freezes the image
MML194x 29
Hor. Position
1) Touch Enter button when the Hor. Position is highlighted on Image
Adjust Menu.
The Hor. Position window appears:
2) Touch the Adjust + or Adjust - button until the desired value appears.
3) Touch the Enter button to confirm your selection.
4) Touch the Adjust + or Adjust - button to exit or to select another item
Vert. Position
1) Touch Enter button when the Vert. Position is highlighted on Image
Adjust Menu.
The Vert. Position window appears:
2) Touch the Adjust + or Adjust - button until the desired value appears.
3) Touch the Enter button to confirm your selection.
4) Touch the Adjust + or Adjust - button to exit or to select another item
Image Size
1) Touch Enter button when the Image Size is highlighted on Image
Adjust Menu.
The Image Size window appears:
- Full
- Aspect/Ratio
- Center
2) Touch the Adjust + or Adjust - button until the desired value appears.
3) Touch the Enter button to confirm your selection.
4) Touch the Adjust + or Adjust - button to exit or to select another item
30 MML194x
LUT
1) Touch Enter button when the LUT is highlighted on Image Adjust
Menu.
The LUT window appears:
The ON SCREEN DISPLAY shows the following available choices, you
can select:
- Linear The Transfer Function is transparent
- Dicom The output curve follows the DICOM curve
- CAL. TG3 – PMS The output curve is the same as PMS
standard monitor. (Calibrated curve)
2) Touch the Adjust + or Adjust - button until the desired value appears.
3) Touch the Enter button to confirm your selection.
4) Touch the Adjust + or Adjust - button to exit or to select another item
Note: The selection of the color LUT works for all modes except for SOG
video formats.
Black Value
1) Touch Enter button when the Black Value is highlighted on Image
Adjust Menu.
The Black Value window appears:
The ON SCREEN DISPLAY shows the following available choices, you
can select:
- Native The first entry of the LUT is empty
- User The value of the first entry of the LUT is the one
passed using the proper RS232 command
2) Touch the Adjust + or Adjust - button until the desired value appears.
3) Touch the Enter button to confirm your selection.
4) Touch the Adjust + or Adjust - button to exit or to select another item
Color Emulation
1) Touch Enter button when the Color Emulation is highlighted on
Image Adjust Menu.
The Color Emulation window appears:
The ON SCREEN DISPLAY shows the following available choices:
- Disabled
- Enabled
2) Touch the Adjust + or Adjust - button until the desired value appears.
3) Touch the Enter button to confirm your selection.
4) Touch the Adjust + or Adjust - button to exit or to select another item
MML194x 31
Border Color
1) Touch Enter button when the Border Color is highlighted on Image
Adjust Menu.
The Border Color window appears:
2) Touch the Adjust + or Adjust - button until the desired value appears.
3) Touch the Enter button to confirm your selection.
4) Touch the Adjust + or Adjust - button to exit or to select another item
Freeze
1) Touch Enter button when the Freeze is highlighted on Image Adjust
Menu.
The Freeze window appears:
The ON SCREEN DISPLAY shows the following available choices:
- Disabled
- Enabled
After setting Freeze to Enabled it freezes the image if activated with the
Freeze ON/OFF (MENU SELECT -) button (see page 21)
2) Touch the Adjust + or Adjust - button until the desired value appears.
3) Touch the Enter button to confirm your selection.
4) Touch the Adjust + or Adjust - button to exit or to select another item
32 MML194x
Advanced
1) Touch Enter to enter in Sub-menu of Advanced menu
2) Touch the Adjust + or Adjust - button until the desired item is
highlighted.
3) When the desired Sub-menu is highlighted, touch the Enter and the
Item is highlighted
4) Touch Adjust + or Adjust - button to adjust the value or to select the
desired sub-menu and touch again Enter to confirm.
5) Touch again Adjust + or Adjust - button to exit or to select another
Sub-menu.
Phase Adjust Manual Adjust the Phase (only for analog inputs)
Clock Adjust Manual Adjust the Clock (only for analog inputs)
Auto Clock Auto Adjust of Clock & Position (only for analog inputs)
Auto Phase Auto Adjust of Phase (only for analog inputs)
Pixel ON-OFF pattern recommended
Video Calibration Auto Adjust of Gain & Offset (only for analog inputs)
Gain: White (254) Page Recommended
Offset: Black (001) Page Recommended
Restore Factory Reload factory parameters (Gain / Offset)
Baud Rate Set the Baud Rate of the Serial Communication Link
5 Volt - DVI Activate/Disable 5V on pin 16 of DVI
MML194x 33
Phase Adjust
1) Touch Enter button when the Phase Adjust is highlighted on Advanced
Menu.
The Phase Adjust window appears:
2) Touch the Adjust + or Adjust - button until the desired value appears.
3) Touch the Enter button to confirm your selection.
4) Touch the Adjust + or Adjust - button to exit or to select another item
Clock Adjust
1) Touch Enter button when the Clock Adjust is highlighted on Advanced
Menu.
The Clock Adjust window appears:
2) Touch the Adjust + or Adjust - button until the desired value appears.
3) Touch the Enter button to confirm your selection.
4) Touch the Adjust + or Adjust - button to exit or to select another item
Auto Clock
1) Touch Enter button when the Auto Clock is highlighted on Advanced Menu.
The Auto Clock window appears:
2) Touch the Enter button to confirm your selection and run the automatic
adjustment procedure.
3) Touch the Adjust + or Adjust - button to exit or to select another item
34 MML194x
Auto Phase
1) Touch Enter button when the Auto Phase is highlighted on Advanced
Menu.
The Auto Phase window appears:
2) Touch the Enter button to confirm your selection and run the automatic
adjustment procedure.
3) Touch the Adjust + or Adjust - button to exit or to select another item
NOTE: It is not allowed to run the procedure more than one
time: to execute it again, Exit the item first.
Video Calibration
1) Touch Enter button when the Video Calibration is highlighted on
Advanced Menu.
The Video Calibration window appears:
The ON SCREEN DISPLAY shows the following available choices:
- Offset it is necessary a black pattern (level 1)
- Gain it is necessary a white pattern (level 254)
- Offset +Gain it is necessary a half black and half white pattern
- Cancel to exit without execute none calibration
2) Touch the Adjust + or Adjust - button until the desired value appears.
3) Touch the Enter button to confirm your selection.
4) Touch the Adjust + or Adjust - button to exit or to select another item
Restore Factory
1) Touch Enter button when the Restore Factory is highlighted on Advanced Menu.
The Restore Factory window appears:
2) Touch the Enter button to confirm your selection.
3) Touch the Adjust + or Adjust - button to exit or to select another item
MML194x 35
AIC (Ambient Light Compensation 000 – 300 lux)
1) Touch Enter button when the AIC is highlighted on Advanced Menu.
The AIC window appears:
2) Touch the Adjust + or Adjust - button until the desired value appears.
3) Touch the Enter button to confirm your selection.
4) Touch the Adjust + or Adjust - button to exit or to select another item
Note: Phase Adjust, Clock Adjust, Auto Clock, Auto Phase, Video
Calibration, Restore Factory are valid only in analog video source. (DSub
and DVI-A)
Baud Rate
1) Touch Enter button when the Baud Rate is highlighted on Advanced
Menu.
The Baud Rate window appears:
2) Touch the Adjust + or Adjust - button until the desired value appears.
3) Touch the Enter button to confirm your selection.
4) Touch the Adjust + or Adjust - button to exit or to select another item
5 Volt - DVI
1) Touch Enter button when the 5 Volt - DVI is highlighted on Advanced
Menu.
The 5 Volt - DVI window appears:
The ON SCREEN DISPLAY shows the following available choices:
- On 5 Volt is supplied on DVI conn. (pin 16)
- Off No power available on DVI conn. (pin 16)
2) Touch the Adjust + or Adjust - button until the desired value appears.
3) Touch the Enter button to confirm your selection.
4) Touch the Adjust + or Adjust - button to exit or to select another item
36 MML194x
Service
Offset Red Manual adjust of Red offset (only for analog inputs)
Offset Green Manual adjust of Green offset (only for analog inputs)
Offset Blue Manual adjust of Blue offset (only for analog inputs)
Gain Red Manual adjust of Red gain (only for analog inputs)
Gain Green Manual adjust of Green gain (only for analog inputs)
Gain Blue Manual adjust of Blue gain (only for analog inputs)
Daisy Address FIMI proprietary protocol address (only for mono versions)
Analog Filter Select analog bandwidth (only for analog inputs)
Actual Brightness Store brightness preset*
*each video timing (Max 14) can use a different brightness preset
MML194x 37
Offset Red
1) Touch Enter button when the Offset Red is highlighted on Service
Menu.
The Offset Red window appears:
2) Touch the Adjust + or Adjust - button until the desired value appears.
3) Touch the Enter button to confirm your selection.
4) Touch the Adjust + or Adjust - button to exit or to select another item
Offset Green
1) Touch Enter button when the Offset Green is highlighted on Service
Menu.
The Offset Green window appears:
2) Touch the Adjust + or Adjust - button until the desired value appears.
3) Touch the Enter button to confirm your selection.
4) Touch the Adjust + or Adjust - button to exit or to select another item
Offset Blue
1) Touch Enter button when the Offset Blue is highlighted on Service
Menu.
The Offset Blue window appears:
2) Touch the Adjust + or Adjust - button until the desired value appears.
3) Touch the Enter button to confirm your selection.
4) Touch the Adjust + or Adjust - button to exit or to select another item
38 MML194x
Gain Red
1) Touch Enter button when the Gain Red is highlighted on Service
Menu.
The Gain Red window appears:
2) Touch the Adjust + or Adjust - button until the desired value appears.
3) Touch the Enter button to confirm your selection.
4) Touch the Adjust + or Adjust - button to exit or to select another item
Gain Green
1) Touch Enter button when the Gain Green is highlighted on Service
Menu.
The Gain Green window appears:
2) Touch the Adjust + or Adjust - button until the desired value appears.
3) Touch the Enter button to confirm your selection.
4) Touch the Adjust + or Adjust - button to exit or to select another item
Gain Blue
1) Touch Enter button when the Gain Blue is highlighted on Service
Menu.
The Gain Blue window appears:
2) Touch the Adjust + or Adjust - button until the desired value appears.
3) Touch the Enter button to confirm your selection.
4) Touch the Adjust + or Adjust - button to exit or to select another item
MML194x 39
Daisy Address
1) Touch Enter button when the Daisy Address is highlighted on Service
Menu.
The Daisy Address window appears:
The ON SCREEN DISPLAY shows the following available choices:
- 1 to -9
- A to -F
2) Touch the Adjust + or Adjust - button until the desired value appears.
3) Touch the Enter button to confirm your selection.
4) Touch the Adjust + or Adjust - button to exit or to select another item
Analog Filter
1) Touch Enter button when the Analog Filter is highlighted on Service
Menu.
The Analog Filter window appears:
The ON SCREEN DISPLAY shows the following available choices:
- 75 MHz
- 150 MHz
- 300 MHz
- Disabled (500 MHz)
2) Touch the Adjust + or Adjust - button until the desired value appears.
3) Touch the Enter button to confirm your selection.
4) Touch the Adjust + or Adjust - button to exit or to select another item
Actual Brightness
1) Touch Enter button when the Actual Brightness is highlighted on
Service Menu.
The Actual Brightness window appears:
The ACTUAL BRIGHTNESS can be adjusted via keyboard (brightness +
and brightness - keys - see 8.1.0)
and saved “User preset” via this OSD item.
40 MML194x
Exit
After the activation of Exit menu the following window appears
MML194x 41
WARNING MESSAGES
Check Input Signal
If the Video Signal is not available for the selected Video Source a warning message pops-up on the screen for about
5s.
When the monitor is properly calibrated (No warning messages in the information item) the displayed message is the
following:
Input Available
After a video source change, if in the selected video source no video signal is available, the MML194x switches back
to the previous video source and shows the dialogue box here below:
Unsupported Mode
If the Video Timing is not available a warning message will appear on the screen for about 5s.
42 MML194x
VIDEO MODES
MML194x 43
IDENTIFICATION LABEL
44 MML194x
SYMBOLS
The following symbols may appear on the product or its accessory:
MML194x 45
GENERAL DEFINITIONS
Abbreviations
CE European Conformity
DEMKO Danish Conformity Mark
CCFL Cold Cathode Fluorescent Lamp
CSA Canadian Standardization Authority
CVBS Composite Video Blanking and Sync
DVI Digital Visual Interface
EEDID Enhanced Extended Display Identification Data
EMC Electromagnetic Compatibility
ESD Electrostatic Discharge
FCC Federal Communications Commission
FRU Field Replaceable Unit
HDP Hot Plug Detect (DVI)
IEC International Electrotechnical Commission (Standards)
60601-1 Medical electrical equipment requirements
LCD Liquid Crystal Display
LUT Look-Up Table
LVDS Low voltage differential signal
MCS Monitor Ceiling Suspension
MTBF Mean Time Between Failure
MTTR Mean Time To Replace
OSD On Screen Display
PCA Printed Circuit Assembled
RFI Radio Frequency Interference
SDI Serial Digital Interface
SMPTE Society of Motion Picture and Television Engineers
SOG Sync On Green (monochrome video)
S-VIDEO Separated Video Y/C ; Y = luminance
C = color
TMDS Transition Minimized Differential Signalling
UL Underwriters Laboratories inc.
VESA Video Electronics Standards Association
VGA Video Graphics Adapter
WEEE Waste Electrical and Electronic Equipment
ROHS Restriction of Hazardous Substances
REACH Registration Evaluation Authorization and Restriction of Chemical Substances
Units of measurement
Length m meter
Time s second
Voltage V Volt
Current A Ampere
Power W Watt
Resistance Ω Ohm
Weight kg Kilogram
Luminance Cd/m² Candela per square meter
Illuminance lx lux
Temperature °C degree Celsius
Frequency Hz Hertz (Cycles per second)
Noise level dB(A) dBspl A weighted
Magnetic field T Tesla (= 10000 Gauss)
46 MML194x
INTENTIONALLY LEFT BLANK
MML194x 47
FIMI BARCO R&D .PI0612 R2012D
FIMI S.r.l. Saronno - Italy
All rights are reserved. Reproduction in whole or in part is prohibited without the written consent of the copyright owner
Product specification subject to change without notice.
Printed in Italy (Rev. 1.0)