You are on page 1of 128

PB980 Diagnostic Codes

Page 1
I. Introduction

This chapter provides information to assist in the interpretation of diagnostics codes contained in the Puritan
Bennett™ 980 Ventilator System’s diagnostic logs. The ventilator’s diagnostic and status logs contain
information that may be useful when troubleshooting the ventilator.

The Puritan Bennett™ 980 contains two CPU PCBA’s (the BD CPU PCBA and the GUI CPU PCBA) which
generate six digit, alphanumeric diagnostic codes as part of their self-tests design.

The Self-tests include the following:

 Power On Self Test (POST) Diagnostics


 Ongoing or Background Diagnostics
 Extended Self Test (EST) Diagnostics
 Short Self Test (SST) Diagnostics

The Diagnostic Logs include the following:

System Diagnostic Log: Stores data on unexpected conditions (Faults and Failures) detected during POST
and background checks. The System Diagnostic Log stores up to 256 entries (first in, first out) organized by
date, time, type of event, the diagnostic code associated with the event, and notes.

System Communication Log: Stores data on ventilator events, including details on GUI/BD
communications. The System Communication Log stores up to 256 entries (first in, first out) organized by
date, time, type of event, the diagnostic code associated with the event, and notes.

EST/SST Diagnostic Log: Stores data on Alerts, Faults and Failures detected during SST and EST, along
with overall SST and EST test statuses. The EST/SST Log stores up to 256 entries (first in, first out)
organized by date, time, the diagnostic code associated with the event that occurred during EST or SST, and
notes.

Logs may be accessed via Normal Operation and in Service Mode.

Diagnostic Codes Interpretation:


st
To interpret a diagnostic code, via the 1 letter, decipher which CPU generated the code. Via the second
digit, determine during what diagnostic test the code was generated, and then read the message associated
with the code to obtain more information about the component, task or system affected.

1. The first letter of a diagnostic code identifies the CPU (BD or GUI) that generated the code. Refer to
the following table.

2. The second digit in a diagnostic code indicates where the fault occurred in software. See the type and
definition columns in the following table.

3. Test Identifier — the last four digits of a code represent the test identifier in the software.

Page 2
Refer to the table below to determine which CPU system generated the code and to identify during what
diagnostic test type the code was generated (SST, EST, Exception, Software test, Communication, POST or
Background Diagnostics). NOTE: the last four digits of a code signify the test identifier in the software. Read
the message associated with a code to obtain more information about the event for troubleshooting.

1st 2nd
Letter System Diagnostic Type Letter Test Type
BD or GUI A = Alert representing a Minor Failure S SST
A
Code in SST or EST (check 2nd Letter for type) E EST
BD T Exception
D (Check 2nd Letter for type)
Code
SYSTEM_EVENT_DIAGNOSTIC - E codes
represent System Events logged for
BD informational purposes only.
E
Code E codes are only generated by the BD
CPU. Review the System Diag & Comm
Log for related / more information.
BD or GUI F = Failure representing a Major Failure S SST
F
Code in SST or EST (check 2nd Letter for type) E EST
BD
H
Code NMI_DIAG - Nonmaskable Interrupt N
C Communication
BD
K Major Fault (check 2nd Letter for type) P POST
Code
B Background
C Communication
BD
L Minor Fault (check 2nd Letter for type) P POST
Code
B Background
SYSTEM_EVENT_DIAGNOSTIC -S codes
represent System Events logged for
GUI informational purposes only.
S
Code S codes are only generated by the GUI
CPU. Review the System Diag & Comm
Log for related / more information.
GUI T Exception
U (Check 2nd Letter for type)
Code
GUI
V
Code NMI_DIAG - Nonmaskable Interrupt N
C Communication
GUI
X Major Fault (check 2nd Letter for type) P POST
Code
B Background
C Communication
GUI
Z Minor Fault (check 2nd Letter for type) P POST
Code
B Background

Page 3
PB980 Diagnostic Codes

Non-technical Alarms

Non-technical alarms often result from a fault in the patient-ventilator interaction or external fault. Some non-
technical alarms relate to conditions the practitioner can correct, such as shifting to an alternate power source
or resolving gas supply issues. Others are patient data alarms associated with an abnormal condition of the
patient’s respiratory status. Reference the PB980 Operator’s Manual for a listing of non-technical alarms.

Technical Alarms

Technical alarms result from a violation of any of the ventilator’s self-monitoring conditions, such as failure of
POST or a background check. This includes faults detected by the ventilator’s background diagnostic system.
Technical Alarms cannot be reset by pressing the Alarm Reset key.

The ventilator reacts to the detection of a Technical alarm by entering one of the following states: Safety
Valve Open (SVO), Ventilator Inoperative (VENT INOP), Backup Ventilation (BUV), or generating High
Urgency and/or Medium Urgency Alarms.

i) SVO is a state which has the following conditions:


(a) The inspiratory PSOL is deactivated
(b) The safety valve is deactivated (de-energized)
(c) The exhalation valve is de-energized

ii) VENT INOP is a state which has the following conditions:


(a) The ventilator is placed in Safety Valve Open (SVO) state
(b) The gas mixing system PSOLs are disabled
(c) The accumulator of the gas mixing system is purged
(d) The VENT INOP indicator is annunciated
(e) The Primary alarm system annunciates an high urgency tone
(f) The secondary Alarm is activated
(g) The Alarm Urgency indicators displays a high urgency alarm condition
(h) The patient data displays are blanked and display the elapsed time without ventilator
support on the GUI

iii) BUV is a state which has 3 modes of operation:


 Mix BUV
 Inspiratory BUV
 Exhalation BUV

Page 4
BUV guidelines:

(a) Only one active mode of Mix BUV, Inspiratory BUV, or Exhalation BUV is allowed at a
time.
(b) During inspiratory or Exhalation Back Up Ventilation, the options gas supply solenoid is
disabled.
(c) During inspiratory or Exhalation Back up Ventilation, auto-zero of pressure sensors is not
allowed.
(d) Once the ventilator has entered Back Up Ventilation, normal ventilation is only entered by
cycling power and successfully completing Extended Self Test (EST).
(e) The Alarm Banner indicates Back Ventilation is active.

iv) Caution
(a) The 980 ventilator shall generate a high urgency alarm and add an entry into the System
Diagnostic Log.

v) Warning
(a) The 980 ventilator shall generate a medium urgency alarm and add an entry into the
System Diagnostic Log.

vi) Notification
(a) The 980 ventilator shall add an entry into the System Diagnostic Log.

Page 5
Upon detection of a specified technical alarm, the system displays an Alarm per the table below to inform the
user that a Background Fault has been detected :

Background Technical Alarm Type (BTAT) System Response

ALARM/BASE URGENCY/PRIORITY ANALYSIS REMEDY MESSAGE


MESSAGE MESSAGE
BTAT
#

1 DEVICE ALERT H/10 VENT INOP EST Required.

Go to System Diag. Log.

2 DEVICE ALERT H/11 Backup Provide alternate source of


Ventilation in ventilation. EST Required.
progress.

3 DEVICE ALERT H/11 Backup Provide alternate source of


Ventilation in ventilation. EST Required.
progress.

4 DEVICE ALERT H/13 Limited Mix Provide alternate source of


Capability, only ventilation. EST Required.
21 or 100% O2
supported.

6 DEVICE ALERT H/150 Background fault Ventilation continues as set.


detected
Go to System Diag. Log.

7 DEVICE ALERT M/330 Background fault Ventilation continues as set.


detected.
Go to System Diag. Log.

9 EXHALATION H/150 Background fault Check exhalation filter and


MODULE detected. exhalation flow sensor.
ALERT
Go to System Diag. Log.

Page 6
The displayed Alarm then provides a link to the System Diagnostic Log for BTAT conditions 1, 6, 7 and 9.
The system responds per the folllowing table for each Background Technical Alarm Type (BTAT).

BTAT # Name Urgency System Response Resettable*

VENT
1 INOP High Safe State No

2 Exh BUV High Backup Ventilation No

3 Insp BUV High Backup Ventilation No

4 Mix BUV High Backup Ventilation No

5 SVO High Safe State Yes

6 Caution High Alarm Condition -Ventilation continues Yes

7 Warning Med Alarm Condition -Ventilation continues Yes

8 Notification N/A Alarm Condition -Ventilation continues Yes

Exhalation
9 Module High Enter Idle Mode Yes

*Resettable conditions autoreset and allow normal operation once the condition that triggered the alarm goes
away. Conditions listed as nonresettable require at least EST to be completed and passed.

When a fault is detected in the primary audio alarm or a loss of communications with the GUI CPU, the
secondary alarm on the BD is activated producing a pulsating tone.

Once VENT INOP is triggered, a ventilator power cycle followed by a successful completion of Service Mode
Extended Self Test shall be required to enter Normal Mode Ventilation.

IMPORTANT NOTE: Whenever replacing a PCBA containing an FPGA, use the


PB980 ESS program to update the ventilator software package (which includes
FPGA’s).

Page 7
PB980 POST Code Table

POST Codes Vent


POST Tests Corrective Action
BD Id GUI Id Response

VENT INOP
POST ERROR CPU if BD CPU;
KP0002 XP0002 Replace applicable CPU PCBA
UNIT GUI INOP
if GUI CPU

VENT INOP
POST ERROR if BD CPU;
KP0006 XP0006 Replace applicable CPU PCBA
PHASE 2 GUI INOP
if GUI CPU

VENT INOP
if BD CPU; Run EST and SST. Power cycle unit. If
POST ERROR
KP0008 XP0008 ALARM problem persists, then replace applicable
NOVRAM Phase I
CONDITION CPU PCBA.
if GUI CPU

When POST has started 3 or more times


not via the power switch or during a power
POST ERROR fail without POST completely finishing, a
KP0009 VENT INOP POST Failure is generated.
ROLLING THUNDER

Run EST to troubleshoot. If persists,


replace BD CPU PCBA.

POST ERROR VENT INOP


WATCHDOG - 3 if BD CPU;
KP0013 XP0013 Replace applicable CPU PCBA
unexpected Boot GUI INOP
loader resets. if GUI CPU

VENT INOP
POST ERROR LOCAL if BD CPU;
KP0082 XP0082 Replace applicable CPU PCBA
BUS GUI INOP
if GUI CPU

VENT INOP
POST ERROR DDR if BD CPU;
KP0084 XP0084 Replace applicable CPU PCBA
MEMORY GUI INOP
if GUI CPU

VENT INOP Review the System Diagnostic Log to


POST ERROR
if BD CPU; determine causal codes. If no supporting
KP0087 XP0087 UNEXPECTED
GUI INOP codes are present, replace applicable CPU
RESET UMPIRE
if GUI CPU PCBA.

Page 8
POST Codes Vent
POST Tests Corrective Action
BD Id GUI Id Response

Run EST and SST. Power cycle unit. If


POST ERROR POST ALARM
KP0088 XP0088 problem persists, then replace applicable
NOVRAM Phase II CONDITION
CPU PCBA.

VENT INOP
POST ERROR TIMER if BD CPU;
KP0093 XP0093 Replace applicable CPU PCBA
TEST GUI INOP
if GUI CPU

POST ERROR SAAS 1. Run EST to determine root cause.


XP0097 ALARM
SELF TEST START 2. Replace the GUI CPU.
CONDITION
3. Replace the UI PCBA.

1. Run EST to determine root cause.


POST ERROR SAAS ALARM
XP0098 2. Replace the GUI CPU.
TEST END CONDITION
3. Replace the UI PCBA.

POST ERROR ALARM


XP0101 Replace the GUI CPU PCBA
CARMINE CONDITION

POST ERROR ALARM


XP0102 Replace the GUI CPU PCBA
CARMINE DRAM CONDITION

ALARM
XP0104 POST GUI CPU UART Replace the GUI CPU PCBA
CONDITION

ALARM 1. Replace the Line Interface 2 PCBA.


XP0105 POST GUI CPU HDMI
CONDITION 2. Replace the GUI CPU PCBA.

POST ERROR VENT


KP0112 VENT INOP Replace the BD CPU PCBA
INOP

Run a complete EST to troubleshoot and


KP0115 POST ERROR DAC VENT INOP identify the analog system or part at fault.

POST ERROR Run a complete EST to troubleshoot and


KP0116 VENT INOP
ANALOG DEVICES identify the analog system / part at fault.

Page 9
POST Codes Vent
POST Tests Corrective Action
BD Id GUI Id Response

1. Ensure all modules and major


assemblies are properly seated and
connected, then run EST to
troubleshoot.
POST ERROR SAFE
KP0120 VENT INOP 2. Replace the exhalation valve.
STATE SYSTEM
3. Replace the safety valve.
4. Replace the delivery PSOL.
5. Replace the delivery flow sensor.
6. Replace the Pneumatic Interface PCBA.

1. Do not hold the service switch in for


POST ERROR longer than 1 second when entering
KP0121 Alarm Condition
SERVICE SWITCH Service Mode.
2. Replace the Line Interface I PCBA.

1. Run EST to troubleshoot.


POST ERROR 2. Check / reseat the flat ribbon cable
POWER between the BD Power Controller
DISTRIBUTION PCBA and the BDU Backplane PCBA.
KP0132 VENT INOP
SUBSYSTEM - BD 3. Replace the BDU Power Controller
Power Controller ADC PCBA.
out of range. 4. Replace the DC to DC Converter
PCBA.

1. Run EST to troubleshoot.


2. Check the Compressor Circuit Breaker.
POST ERROR
ALARM 3. Check the Compressor Batteries.
KP0133 COMPRESSOR
CONDITION Replace Compressor Interface PCBA.
SUBSYSTEM
4. Replace Compressor Power Controller
PCBA.

POST ERROR MIX


1. Run EST to troubleshoot.
SUBSYSTEM - Mix
KP0134 VENT INOP 2. Check / reseat all Pneumatic cables.
FPGA ADC out of
3. Replace the Mix Module PCBA.
range.
POST ERROR
INSPIRATORY 1. Run EST to troubleshoot.
SUBSYSTEM - 2. Replace the Pneumatic Interface
KP0135 VENT INOP
pneumatic interface - PCBA.
inspiratory ADC out of 3. Replace the IFM PCBA.
range.
POST ERROR
EXHALATION
SUBSYSTEM - 1. Run EST to troubleshoot.
KP0136 VENT INOP
pneumatics interface - 2. Replace the Exhalation Sensor PCBA.
exhalation ADC out of
range.

Page 10
POST Codes Vent
POST Tests Corrective Action
BD Id GUI Id Response

1. Check primary battery. Restart vent.


PRIMARY BATTERY ALARM 2. If condition persists, replace battery with
LP0123
NOT INSTALLED CONDITION a known good battery.
3. Run EST.

1. Check primary battery. Restart vent.


2. If condition persists, replace battery
PRIMARY BATTERY ALARM
LP0125 with a known good battery.
NOT INSTALLED CONDITION
3. Run EST.
4. Replace the Battery Backplane PCBA.

1. Check exhalation valve flow sensor


(EVQ). Restart vent.
2. If condition persists with a properly
connected sensor, then run EST.
EXHALATION FLOW 3. Replace the EVQ.
ALARM
LP0127 SENSOR NOT 4. Check the tubing from F5 filter to the
CONDITION
INSTALLED Exhalation Sensor PCBA.
5. Perform a flow sensor calibration,
exhalation valve calibration, EST and
SST. Restart vent to see if problem
has been resolved.

Page 11
PB980 Background Diagnostics Code Table

Background
Diagnostic Codes Vent
Diagnostic Tests Corrective Action
Response
BD Id GUI Id

1. Run EST.
KB0001 2. Replace the Safety Valve.
Safety valve current ALARM
CONDITION 3. Replace the IFM PCBA.
out of range
4. Replace the Pneumatic
Interface PCBA.

1. Check exhalation valve flow


sensor (EVQ). Restart vent.
2. If condition persists with a
properly connected sensor, then
run EST.
Exhalation flow sensor reading ALARM
KB0002 CONDITION 3. Replace the EVQ. Perform a
out of range
flow sensor calibration,
exhalation valve calibration,
EST and SST. Restart vent to
see if problem has been
resolved.

1. Run EST.
2. Replace the Breath Delivery
PSOL.
KB0003 Inspiratory PSOL current out of 3. Replace the IFM PCBA.
IBUV
range 4. Replace the Breath Delivery
Flow Sensor.
5. Replace the Pneumatic
Interface PCBA.

1. Run EST.
2. Replace the Exhalation Valve
KB0004 Exhalation Motor current out Assembly and perform an
of range Exhalation Valve Calibration.
EBUV
3. Replace the Pneumatic
Interface PCBA.
4. Replace Exhalation Sensor
PCBA.

1. Run EST.
KB0006 2. Replace the Exhalation Valve
Exhalation Pressure reading VENT
Sensor PCBA.
out of range INOP
3. Replace the Pneumatic
Interface PCBA.

Page 12
Background
Diagnostic Codes Vent
Diagnostic Tests Corrective Action
Response
BD Id GUI Id

1. Run EST.
KB0007 Inspiratory Pressure reading VENT 2. Replace the IFM PCBA.
out of range INOP 3. Replace the Pneumatic
Interface PCBA.

1. Run a complete EST.


2. Replace the Delivery Flow
KB0008 Sensor and run the Flow
Inspiratory Flow reading out of VENT Sensor Calibration.
range high INOP 3. Replace the Delivery PSOL.
4. Replace the IFM PCBA.
5. Replace the Pneumatic
Interface PCBA.

1. Run a complete EST.


2. Replace the Delivery Flow
KB0009 Sensor and run the Flow
Inspiratory Flow reading out of VENT Sensor Calibration.
range low INOP 3. Replace the Delivery PSOL.
4. Replace the IFM PCBA.
5. Replace the Pneumatic
Interface PCBA.

1. Ensure the ventilator has been


warmed up at ambient
KB0010 temperature for at least 10
minutes. Run a complete EST.
Inspiratory Flow temperature out of VENT 2. Replace the Delivery Flow
range INOP Sensor and run the Flow
Sensor Calibration.
3. Replace the IFM PCBA.
4. Replace the Pneumatic
Interface PCBA.

1. Ensure the ventilator has been


warmed up at ambient
temperature for at least 10
KB0011 minutes. Run a complete EST.
Exhalation Flow temperature out of VENT
2. Replace the Exhalation Valve
range INOP
Flow Sensor and run the Flow
Sensor Calibration.
3. Replace the Pneumatic
Interface PCBA.

Page 13
Background
Diagnostic Codes Vent
Diagnostic Tests Corrective Action
Response
BD Id GUI Id

1. Run a complete EST.


KB0020 2. Replace the Delivery PSOL.
VENT
Inspiratory PSOL stuck open 3. Replace the IFM PCBA.
INOP
4. Replace the Pneumatic
Interface PCBA.

KB0021 1. Perform an Atmospheric


Atmospheric Pressure Sensor Device Pressure Transducer
reading out of range Alert Calibration.
2. Replace the IFM PCBA.

1. Replace the oxygen sensor and


KB0022 perform an oxygen sensor
VENT
O2 Sensor reading out of range calibration.
INOP
2. Run EST.
3. Replace the IFM PCBA.

KB0023 O2 sensor reset No action required.

1. Run EST.
KB0024 2. Replace the Safety Valve.
Safety Valve Open Current out of VENT
3. Replace the IFM PCBA.
range INOP
4. Replace the Pneumatic
Interface PCBA.

1. Run EST.
KB0025 2. Check or replace the IFM
Inspiratory Pressure Sensor reading VENT
PCBA.
out of range – stuck at ambient INOP
3. Replace the Pneumatic
Interface PCBA.

1. Run EST.
KB0026 2. Check or replace the Exhalation
Expiratory Pressure Sensor reading VENT
Sensor PCBA.
out of range – stuck at ambient INOP
3. Replace the Pneumatic
Interface PCBA.

1. Run EST.
KB0027 2. Check or replace the IFM
VENT
Inspiratory Autozero Failed PCBA.
INOP
3. Replace the Pneumatic
Interface PCBA.

Page 14
Background
Diagnostic Codes Vent
Diagnostic Tests Corrective Action
Response
BD Id GUI Id

1. Run EST.
KB0028 2. Check or replace the Exhalation
VENT
Expiratory Autozero Failed Sensor PCBA.
INOP
3. Replace the Pneumatic
Interface PCBA.

1. Run EST.
KB0029 2. Replace the Power Distribution
Power Fail Capacitor reading out of VENT
PCBA.
range INOP
3. Replace the Power Controller
PCBA.

KB0030 1. Run EST.


Alarm Cable Voltage reading out of VENT
2. Replace the BDU Alarm Cable.
range INOP
3. Replace the BDU Alarm.

1. Run EST.
2. Perform the GUI Touchscreen
Background Touch Screen check ALARM
XB0034 CONDITION Calibration.
failed
3. Replace the GUI Touchscreen
Assembly.

1. Remove any obstructions from


the GUI.
Device 2. Clean the GUI.
XB0035 Touch Screen Blocked
Alert 3. Run EST.
4. Replace the GUI Touchscreen
Assembly.

XB0036 Touch Screen Resumed No action required.

KB0038 1. Run EST. If condition resets


Alarm return to operation. If code
BD NOVRAM Checksum test failed
Condition reoccurs, replace CPU.
2. Replace the BD CPU PCBA.

VENT
KB0039 BD Time of Date Clock failed Replace the BD CPU PCBA
INOP

ALARM
XB0040 GUI Time of Date Clock failed CONDITION Replace the GUI CPU PCBA

ALARM
XB0041 GUI NOVRAM Checksum test failed CONDITION Replace the GUI CPU PCBA

Page 15
Background
Diagnostic Codes Vent
Diagnostic Tests Corrective Action
Response
BD Id GUI Id

1. Run EST.
KB0045 LVDT Temperature out of range 2. Verify that the BDU cooling fan
is connected and working.
3. Replace the LVDT PCBA.

1. Remove any obstructions from


the GUI.
ALARM 2. Clean the GUI.
XB0046 GUI Offscreen key stuck CONDITION 3. Run EST.
4. Replace the GUI Touchscreen
Assembly.

ALARM 1. Run EST.


GUI SAAS communications failed CONDITION
XB0049 2. Replace the GUI CPU PCBA.

1. Check the System Diagnostic


Log for other codes. If no
ALARM
KB0052 Loss of GUI communications CONDITION
other codes are present, no
action required.
2. Check the GUI cable.

1. Check the System Diagnostic


XB0053 ALARM Log for other codes. If no
Loss of BD communications CONDITION other codes are present, no
action required.

KB0054 Resumed GUI communications No action required.

XB0055 Resumed BD communications


No action required.

VENT
KB0056 Background check failed. EST Run EST.
INOP
required.

1. Run EST.
2. Replace the GUI CPU PCBA.
XB0057 GUI Audio Alarm failed ALARM 3. Replace the GUI Audio Alarm.
CONDITION 4. Replace the User Interface
PCBA.

Page 16
Background
Diagnostic Codes Vent
Diagnostic Tests Corrective Action
Response
BD Id GUI Id

1. Run EST.
2. Replace the Safety Valve.
VENT
KB0059 Safety valve current out of range 3. Replace the IFM PCBA.
INOP
4. Replace the Pneumatic
Interface PCBA.

1. Run EST.
2. Ensure the cooling fan is
connected and operational.
KB0060 3. Replace the Primary Battery.
Ventilator Primary Battery Charger VENT 4. Replace the Power Controller
Fault INOP PCBA.
5. Replace the Power Distribution
PCBA.
6. Replace the Battery Backplane
PCBA.

1. Run EST.
2. Replace the Primary Battery.
3. Replace the Power Controller
KB0061 Ventilator Primary Battery Fault – ALARM PCBA.
battery not charging CONDITION 4. Replace the Power Distribution
PCBA.
5. Replace the Battery Backplane
PCBA.

1. Run EST.
2. Ensure the cooling fan is
connected and operational.
KB0062 3. Replace the Extended Battery.
Ventilator Extended Battery ALARM 4. Replace the Power Controller
Charging Fault CONDITION PCBA.
5. Replace the Power Distribution
PCBA.
6. Replace the Battery Backplane
PCBA.

1. Run EST.
2. Replace the Extended Battery.
3. Replace the Power Controller
KB0063 Ventilator Extended Battery Fault – ALARM PCBA.
battery not charging CONDITION 4. Replace the Power Distribution
PCBA.
5. Replace the Battery Backplane
PCBA.

Page 17
Background
Diagnostic Codes Vent
Diagnostic Tests Corrective Action
Response
BD Id GUI Id

1. Run EST.
2. Ensure the cooling fan is
connected and operational.
KB0064 Compressor Primary Battery 3. Replace the Compressor
Charger Fault Primary Battery.
ALARM
CONDITION 4. Replace the Compressor
Power Controller PCBA.
5. Replace the Compressor
Power Distribution PCBA.
6. Replace the Compressor
Battery Backplane PCBA.

1. Run EST.
2. Ensure the compressor cooling
fan is connected and
operational.
KB0065 Compressor Primary Battery 3. Replace the Compressor
Charger Fault ALARM Primary Battery.
CONDITION 4. Replace the Compressor
Power Controller PCBA.
5. Replace the Compressor
Power Distribution PCBA.
6. Replace the Compressor
Battery Backplane PCBA.

1. Run EST.
2. Replace the Compressor
Primary Battery.
Compressor Primary Battery 3. Replace the Compressor
ALARM
Charger Fault - battery not charging CONDITION Power Controller PCBA.
KB0065 4. Replace the Compressor
Power Distribution PCBA.
5. Replace the Compressor
Battery Backplane PCBA.

1. Run EST.
2. Ensure the compressor cooling
fan is connected and
KB0066 operational.
3. Replace the Compressor
Compressor Extended Battery ALARM
CONDITION Extended Battery.
Charger Fault
4. Replace the Compressor
Power Controller PCBA.
5. Replace the Compressor
Power Distribution PCBA.
6. Replace the Compressor
Battery Backplane PCBA.

Page 18
Background
Diagnostic Codes Vent
Diagnostic Tests Corrective Action
Response
BD Id GUI Id

1. Run EST.
2. Replace the Compressor
Extended Battery.
KB0067 3. Replace the Compressor Power
Controller PCBA.
Compressor Extended Battery
4. Replace the Compressor Power
Charger Fault - battery not charging
Distribution PCBA.
ALARM
CONDITION 5. Replace the Compressor
Battery Backplane PCBA.

1. Run EST.
Alarm
KB0069 Datakey/USB Update Failed 2. Re-enter the Options Code.
Condition
3. Obtain a new data key.

BD runs POST.
BD Task Monitor Failed Reset
KB0070 Replace the BD CPU PCBA.

1. Check the logs for other codes.


Alarm
GUI logged dropped a waveform 2. If the code reoccurs, replace the
ZB0071 Condition
packet GUI CPU PCBA.

1. Check the logs for other codes.


Alarm
KB0072 BD dropped a waveform packet 2. If the code reoccurs, replace the
Condition
GUI CPU PCBA.

VENT 1. Run EST.


KB0073 Forced Ventilator Inoperative
INOP 2. Replace the BD CPU PCBA.

Reset –
KB0074 Multiple BD Background Main Cycle 1. Run EST.
Runs
Messages 2. Replace the BD CPU PCBA.
POST

KB0075 Multiple BD Background Second Reset –


1. Run EST.
Cycle Messages Runs
2. Replace the BD CPU PCBA.
POST

Alarm
Replace BD CPU PCBA.
KB0076 XB0076 Watchdog Failure. Condition

KB0077 Init resume GUI communications No action required.

XB0078 Init resume BD communications No action required.

Page 19
Background
Diagnostic Codes Vent
Diagnostic Tests Corrective Action
Response
BD Id GUI Id

1. Check the System Diagnostic


Log for other codes. If no other
Alarm
KB0079 Init loss of GUI communications codes are present, no action
Condition
required.
2. Check the GUI cable.
3. Replace the GUI CPU PCBA.

1. Check the System Diagnostic


Log for other codes. If no other
Settings
XB0080 Init loss of BD communications codes are present, no action
locked out
required.
2. Replace the BD CPU PCBA.

KB0081 Compressor Serial Number updated Notification No action required.

KB0082 Compressor update PM hours Notification No action required.

Alarm Clear Obstruction


XB0084 Touch Screen Blocked
Condition Run EST

XB0085 Touch Screen Resumed ? No action required.

1. Run SST with a known good


Prox Flow Sensor.
KB0088 Proximal Flow Sensor Error Alarm 2. Replace the Prox Module
Condition PCBA.
3. Replace the Options Host
PCBA.

1. Run EST.
2. Check for leaks and occlusions
KB0089 Mix Accumulator Pressure Out of in the Mix Subsystem.
Range MBUV 3. Replace affected Mix PSOL.
4. Replace the IFM PCBA.
5. Replace Solenoid 1.
6. Replace PS1.

1. Ensure high pressure air and


oxygen supplies are within 35 –
KB0090 Mix Inlet Pressure Out of Range 87 psig.
MBUV
2. Run EST.
3. Replace the Gas Supply
Sensors PCBA.

Page 20
Background
Diagnostic Codes Vent
Diagnostic Tests Corrective Action
Response
BD Id GUI Id

1. Run EST.
2. Verify high pressure source
gases are within ventilator
specifications and that there are
no restrictions or occlusions in
the gas supplies.
3. If the Mix Air PSOL is affected
and an external, wall-air water
trap is present, remove it from
the circuit. NOTE: do NOT
KB0091 remove it if water is present in
Mix PSOL Current Out of Range MBUV
the high pressure wall air
supply. Instead, check/replace
the wall-air water trap filter and
rerun EST.
4. Replace affected Mix PSOL.
5. Replace affected Mix Flow
Sensor.
6. Replace the IFM PCBA.
7. Replace the Gas Supply
Sensors PCBA.

1. Power Cycle the ventilator.


2. Check cable connections to the
Background
KB0094 BDU Status Display Error Notification BDU Status Display PCBA.
3. Replace the BDU Status
Display PCBA.

1. Power Cycle the ventilator.


2. Check cable connections to the
KB0095 BD Status Display Communications Background BDU Status Display PCBA.
Test Failed Notification 3. Replace the BDU Status
Display PCBA.
4. Replace the BD CPU PCBA.

VENT 1. Run EST.


KB0096 Multiple Backup Ventilation States INOP 2. Replace the BD CPU PCBA.
Invoked. 3.

VENT
Mix BUV failed to activate. Replace the BD CPU CPBA.
INOP
KB0097

Page 21
Background
Diagnostic Codes Vent
Diagnostic Tests Corrective Action
Response
BD Id GUI Id

1. Run EST.
2. Replace the BD CPU PCBA.
KB0098 3. Replace the affected Mix PSOL.
VENT 4. Replace the affected Mix Flow
Unintentional Mix BUV activation.
INOP Sensor.
5. Replace the Mix Module PCBA.

1. Run EST.
2. Replace the Mix Air Flow
KB0099 Mix Air Flow Sensor Reading Out of Sensor.
MBUV
Range 3. Replace the Mix Air PSOL.
4. Replace the Mix Module PCBA.
5. Replace the IFM PCBA.

1. Run EST.
2. Replace the Mix Oxygen Flow
KB0100 Mix Oxygen Flow Sensor Reading Sensor.
MBUV
Out of Range 3. Replace the Mix Oxygen PSOL.
4. Replace the Mix Module PCBA.
5. Replace the IFM PCBA.

1. Run EST.
2. Ensure no leaks or occlusions in
the Mix Subsystem.
3. Replace affected Mix PSOL.
KB0101 Mix Accumulator Pressure Reading VENT 4. Replace the IFM PCBA.
Out of Range INOP 5. Verify Solenoid 1 functionality
via EST. Replace Solenoid 1.
6. Verify PS1 functionality via EST.
Replace PS1.

1. Run EST.
VENT
KB0102 Mix Digital to Analog Converter 2. Replace the Mix Module PCBA.
INOP
Failure

1. Run EST.
2. Try a known good USB flash
KB0103 drive.
Alarm 3. Ensure flash drive is not full.
USB Failed.
Condition 4. Replace Line Interface STD2
PCBA.
5. Replace the Communications
Backplane PCBA.

Page 22
Background
Diagnostic Codes Vent
Diagnostic Tests Corrective Action
Response
BD Id GUI Id
Alarm
XB0104 GUI Video Failure 1. Replace the GUI CPU PCBA.
Condition

COMP Run EST.


KB0105 Compressor Inoperative.
INOP

1. Run EST.
2. Replace the IFM PCBA.
3. Replace the Pneumatic
KB0106 Delivery Subsystem Error – PI, VENT Interface PCBA.
Delivery PSOL, Delivery Flow INOP 4. Replace the BD CPU PCBA.
Sensor, Safety Valve error 5. Replace the Options Backplane
PCBA.
6. Replace the BDU Backplane
PCBA.

1. Run EST.
KB0107 Pneumatic Interface Analog to VENT
2. Replace the Pneumatic
Digital Converter Failure. INOP
Interface PCBA.

1. Run EST.
KB0108 2. Replace the Backup Ventilation
VENT
Unexpected Inspiratory BUV Error Solenoid 3.
INOP
3. Replace the Pneumatic
Interface PCBA.

1. Run EST.
2. Replace the Backup Ventilation
VENT
KB0109 Inspiratory BUV Transition Error Solenoid 3.
INOP
3. Replace the Pneumatic
Interface PCBA.

1. Run EST.
KB0110 Expiratory Pressure Sensor Analog 2. Replace the Pneumatic
VENT
to Digital Converter Failure Interface PCBA.
INOP
3. Replace the Exhalation Sensor
PCBA.

1. Run EST.
2. Replace the Exhalation Valve.
Unexpected Expiratory Backup
KB0111 VENT 3. Replace the Pneumatic
Ventilation Error
INOP Interface PCBA.
4. Replace the Exhalation Sensor
PCBA.

Page 23
Background
Diagnostic Codes Vent
Diagnostic Tests Corrective Action
Response
BD Id GUI Id

1. Run EST.
KB0112 Exhalation Backup Ventilation VENT 2. Replace the Pneumatic
Transition Error INOP Interface PCBA.
3. Replace the Exhalation Valve.

1. Run EST.
2. Replace the Exhalation Valve.
KB0113 3. Replace the Exhalation Sensor
Expiratory Poppet Control Reading VENT
PCBA.
Out of Range INOP
4. Replace the Pneumatic
Interface PCBA.

1. Run EST.
2. Replace the Exhalation Valve.
KB0114 3. Replace the Exhalation Sensor
Exhalation Valve velocity control VENT
PCBA.
reading out of range. INOP
4. Replace the Pneumatic
Interface PCBA.

1. Replace the CO2 Sensor.


2. Replace the Prox Flow
CO2 Sensor Fault Alarm
KB0115 Capnostat Module.
Condition
3. Replace the Options Host
PCBA 1.

CO2 Sensor zero offset initiated Notification No action required.


KB0116

1. No action required unless


accompanied by other codes.
2. Run EST.
3. Check/reseat cables in the GUI.
XB0117 User Interface Communications Notification
4. Reseat the GUI cable.
Error
5. Replace the GUI cable.
6. Replace the User Interface
PCBA.

XB0118 Video lock error Notification Replace the GUI CPU PCBA.

1. Replace the GUI CPU PCBA.


XB0119 User Interface Heartbeat Error Notification 2. Replace the User Interface
PCBA.

Page 24
Background
Diagnostic Codes Vent
Diagnostic Tests Corrective Action
Response
BD Id GUI Id

1. Run EST.
Compressor Inspiratory Flow Alarm 2. Replace the Compressor Motor.
KB0120 Temperature too High Condition 3. Replace the Compressor
Interface PCBA.

KB0121 1. Run EST


Inspiratory 3.3V reference voltage VENT 2. Replace the IFM PCBA.
out of range INOP 3. Replace the Pneumatic
Interface PCBA.

1. Run EST
Inspiratory 3.3V reference OPM
KB0122 VENT 2. Replace the IFM PCBA.
Scale out of range
INOP 3. Replace the Pneumatic
Interface PCBA.

1. Run EST
VENT 2. Replace the IFM PCBA.
Inspiratory Pressure 3.3V reference
KB0123 INOP 3. Replace the Pneumatic
voltage scale out of range
Interface PCBA.

1. Run EST
2. Replace the Pneumatic
Expiratory Pressure 3.3V reference VENT
KB0124 Interface PCBA.
voltage scale out of range INOP
3. Replace the Exhalation Sensor
PCBA.

1. Run EST
2. Replace the Pneumatic
Expiratory Pressure 3.3V reference VENT
KB0125 Interface PCBA.
voltage scale out of range INOP
3. Replace the Exhalation Sensor
PCBA.

1. Run EST
KB0126 2. Replace the Mix Module PCBA.
VENT 3. Replace the IFM PCBA.
Mix 5V Plus Supply out of range
INOP 4. Replace the DC to DC
Converter PCBA.

1. Run EST
2. Replace the Mix Module PCBA.
Mix 5V Minus Supply out of range VENT
KB0127 3. Replace the IFM PCBA.
INOP
4. Replace DC to DC Converter
PCBA.

Page 25
Background
Diagnostic Codes Vent
Diagnostic Tests Corrective Action
Response
BD Id GUI Id

1. Run EST
2. Replace the Mix Module PCBA.
VENT
KB0129 Mix 12V Minus Supply out of range 3. Replace the IFM PCBA.
INOP
4. Replace the DC to DC
Converter PCBA.

1. Run EST
2. Replace the Mix Module PCBA.
VENT
KB0130 Mix 24V Minus Supply out of range 3. Replace the IFM PCBA.
INOP
4. Replace the DC to DC
Converter PCBA.

1. Run EST
VENT 2. Replace the Mix Module PCBA.
Mix Air Flow ADC OOR
KB0131 INOP 3. Replace the Air Flow Sensor.
4. Replace the IFM PCBA.

1. Run EST
KB0132 2. Replace the Mix Module PCBA.
Mix Oxygen Flow ADC OOR VENT
3. Replace the Oxygen Flow
INOP
Sensor.
4. Replace the IFM PCBA.

1. Run EST.
2. Replace the exhalation valve.
KB0133 Exhalation valve current out of VENT 3. Replace the Pneumatic
range INOP Interface PCBA.
4. Replace the Exhalation Sensor
PCBA.

1. Check/test the Primary Battery.


Alarm 2. If fully depleted, charge the
KB0134 Ventilator Primary Battery Fault Condition Primary Battery.
3. Replace the Primary Battery.

1. Check/test the Extended


Battery.
Alarm
KB0135 Extended Battery Fault 2. If fully depleted, charge the
Condition
Extended Battery.
3. Replace the Extended Battery.

Page 26
Background
Diagnostic Codes Vent
Diagnostic Tests Corrective Action
Response
BD Id GUI Id

1. Ensure the cooling fan is


Ventilator Primary Battery Alarm connected.
KB0136 Temperature too high Condition 2. Replace the primary battery.
3. Run EST.

1. Ensure the cooling fan is


Ventilator Primary Battery connected.
Alarm
KB0137 Temperature too high 2. Replace the primary battery.
Condition
3. Run EST.

1. Ensure the cooling fan is


Ventilator Primary Battery
connected.
KB0138 Temperature out of range Alarm
2. Replace the primary battery.
Condition
3. Run EST.

1. Ensure the cooling fan is


KB0139 Ventilator Extended Battery
connected.
Temperature out of range Alarm
2. Replace the extended battery.
Condition
3. Run EST.

1. Run EST.
2. Replace the primary battery.
KB0140 Ventilator Primary Battery ADC
Alarm 3. Replace the Power Distribution
Voltage out of range
Condition PCBA.
4. Replace the Power Controller
PCBA.

1. Run EST.
KB0141 Ventilator Extended Battery ADC 2. Replace the primary battery.
Voltage out of range Alarm 3. Replace the Power Distribution
Condition PCBA.
4. Replace the Power Controller
PCBA.

Page 27
Background
Diagnostic Codes Vent
Diagnostic Tests Corrective Action
Response
BD Id GUI Id

Alarm
KB0142 Ventilator Primary Battery Cycles Replace Primary Battery.
Condition
Exceeded.

Alarm
KB0143 Ventilator Extended Battery Cycles Replace Extended Battery.
Condition
Exceeded.

Alarm
KB0144 Ventilator Battery Charge time fault Replace battery.
Condition

Alarm Ensure power cord is properly


KB0145 Ventilator Mains OOR.
Condition connected to AC Mains.

Run EST to isolate whether the


KB0146 Ventilator 3.3 voltage ref out of VENT problem is at PI or PE or
range. INOP compressor.

Run EST to isolate whether the


KB0147 Ventilator 3.3 voltage monitor out of VENT problem is at PI or PE or
range INOP compressor.

1. Run EST.
KB0148 2. Replace the DC to DC
Ventilator negative 5 volts out of VENT
Converter PCBA.
range INOP
3. Replace the Mix Module
PCBA.

1. Run EST.
KB0149 Ventilator VCC Standby out of 2. Replace the Power Distribution
VENT
range PCBA.
INOP
3. Replace the Power Controller
PCBA.

Alarm
KB0150 Compressor Primary Battery Fault Replace battery.
Condition

Page 28
Background
Diagnostic Codes Vent
Diagnostic Tests Corrective Action
Response
BD Id GUI Id

Compressor Extended Battery Alarm


KB0151 Fault. Condition Replace battery.

KB0152 Compressor Primary Battery Temp


Alarm Replace battery.
too high.
Condition

KB0153 Compressor Extended Battery


Alarm
Temp too high. Replace battery.
Condition

Compressor Primary Battery


Alarm
KB0154 Voltage out of range. Replace battery.
Condition

KB0155 Compressor Extended Battery Alarm


Replace battery.
Voltage out of range. Condition

Compressor Primary Battery ADC Alarm


Replace battery.
KB0156 Voltage out of range. Condition

Compressor Extended Battery ADC Alarm


Replace battery.
KB0157 Voltage out of range. Condition

KB0158 Compressor Primary Battery Cycles


Alarm
Exceeded. Replace battery.
Condition

Compressor Extended Battery


Alarm
KB0159 Cycles Exceeded. Replace battery.
Condition

KB0160 Compressor Battery Charge Time Alarm


Replace battery.
Fault. Condition

Page 29
Background
Diagnostic Codes Vent
Diagnostic Tests Corrective Action
Response
BD Id GUI Id

1. Run EST.
KB0161 2. Replace the Compressor
Power Supply.
COMP
Compressor 24 V Supply Out of 3. Replace the Compressor
INOP
Range Power Distribution PCBA.
4. Replace the Compressor
Power Controller PCBA.

Compressor 3.3 voltage monitor out


Notification No action required.
KB0162 of range.

KB0163 Compressor negative 5 volts out of


Notification No action required.
range.

Perform the specified calibration as


directed by the ventilator.

NOTE: With the exception of the


oxygen sensor calibration, all
KB0166 Calibration Expired. Notification calibrations are performed in
Service Mode. Never perform
Service Mode calibrations while
the ventilator is on a patient.

Perform the specified preventive


KB0167 Vent PM Expired. Notification
maintenance.

Perform the specified preventive


KB0168 Compressor PM Expired. Notification
maintenance.

Provide alternate care for the


KB0169 EST Expired. Notification patient. Remove the ventilator
from the patient and run EST.

KB0170 Ventilator Primary Battery Comm Alarm


Replace the battery.
Error. Condition

KB0171 Ventilator Extended Battery Comm Alarm


Replace the battery.
Error. Condition

Page 30
Background
Diagnostic Codes Vent
Diagnostic Tests Corrective Action
Response
BD Id GUI Id

KB0172 Compressor Primary Battery Comm Alarm


Replace the battery.
Error. Condition

Alarm
KB0173 Compressor Extended Battery Replace the battery.
Condition
Comm Error.

KB0174 BD Status Display Comm Resume Notification No action required.

1. Run EST.
2. Replace the Backup Ventilation
KB0175 BUV End Inspiratory Pressure is out VENT Solenoid (Solenoid 3).
of range INOP 3. Replace the IFM PCBA.
4. Replace the Pneumatic
Interface PCBA.

1. Run EST.
KB0176 2. Replace the Exhalation Sensor
BUV End Expiratory Pressure is out VENT PCBA.
of range INOP 3. Replace the Exhalation Valve.
4. Replace the Pneumatic
Interface PCBA.

1. Run EST.
2. Replace the Backup Ventilation
KB0177 BUV Breath Rate out of range. VENT Solenoid (Solenoid 3).
INOP 3. Replace the IFM PCBA.
4. Replace the Pneumatic
Interface PCBA.

KB0178 Offscreen Keys event. Notification No action required.

KB0179 System Logs Invalid Media Failure Notification. No action required.

System Logs Database Initialization


KB0180 Notification. No action required.
Failure

Page 31
Background
Diagnostic Codes Vent
Diagnostic Tests Corrective Action
Response
BD Id GUI Id

KB0181 System Logs Action Failure Notification. No action required.

Ensure High Pressure Air Inlet


KB0182 Air Pressure Too High. Notification. Pressure does not exceed 87
PSIG.

KB0189 System Logs Database Drop


Notification. No action required.
Failure.

KB0190 System Logs Database Drop Event. Notification. No action required.

1. Remove then reinsert USB.


KB0191 2. Try the other USB port.
Unable to save config files to USB
Notification. 3. Try another USB flash drive.
4. Replace the Line Interface
STD2 PCBA.

1. Remove then reinsert USB.


2. Try the other USB port.
Unable to load config files from
KB0192 Notification. 3. Try another USB flash drive.
USB.
4. Replace the Line Interface
STD2 PCBA.

1. Run EST.
2. Replace the IFM PCBA.
VENT
3. Replace Solenoid 1.
KB0193 Mix Pressure Switch Activated INOP
4. Replace PS1.

1. Run EST.
2. Replace the IFM PCBA.
KB0194 3. Replace the Pneumatic
Inspiratory 24 volts out of range Notification
Interface PCBA.
4. Replace the DC to DC
Converter.

Page 32
Background
Diagnostic Codes Vent
Diagnostic Tests Corrective Action
Response
BD Id GUI Id

KB0195 Ventilator Primary Battery No Fault. Notification No action required.

KB0196 Ventilator Extended Battery No


Notification No action required.
Fault.

KB0197 Ventilator Primary Battery Temp OK Notification No action required.

KB0198 Ventilator Extended Battery Temp Notification No action required.


OK

KB0199 Ventilator Primary Battery Voltage Notification No action required.


OK

KB0200 Ventilator Extended Battery Voltage


Notification No action required.
OK

KB0201 Ventilator Primary Battery ADC


Notification No action required.
Voltage OK

KB0202 Ventilator Extended Battery ADC Notification No action required.


Voltage OK

Ventilator Primary Battery No


Notification No action required.
KB0203 Comm Error

KB0204 Ventilator Extended Battery No


Notification No action required.
Comm Error

KB0205 Compressor Primary Battery No


Fault Notification No action required.

Page 33
Background
Diagnostic Codes Vent
Diagnostic Tests Corrective Action
Response
BD Id GUI Id

KB0206 Compressor Extended Battery No


Notification No action required.
Fault

KB0207 Compressor Primary Battery Temp Notification No action required.


OK

KB0208 Compressor Extended Battery Notification No action required.


Temp OK

KB0209 Ventilator Compressor Battery Notification No action required.


Voltage OK

KB0210 Ventilator Compressor Extended Notification No action required.


Battery Voltage OK

KB0211 Ventilator Compressor Primary Notification No action required.


Battery ADC Voltage OK

Ventilator Compressor Extended


Notification No action required.
KB0212 Battery ADC Voltage OK

Alarm
KB0213 Ventilator Compressor Primary Condition
Replace battery.
Battery No Comm Error

Ventilator Compressor Extended Alarm


Condition
Replace battery.
KB0214 Battery No Comm Error

KB0215 Ventilator Primary Battery Charge Alarm


Condition
Replace battery.
Error

KB0216 Ventilator Primary Battery No Notification No action required.


Charge Error

Page 34
Background
Diagnostic Codes Vent
Diagnostic Tests Corrective Action
Response
BD Id GUI Id

KB0217 Ventilator Extended Battery Charge Alarm


Replace the battery.
Error Condition

KB0218 Ventilator Extended Battery No Alarm


Replace the battery.
Charge Error Condition

KB0219 Compressor Primary Battery Alarm


Replace the battery.
Charge Error Condition

KB0220 Compressor Primary Battery No Alarm


Replace the battery.
Charge Error Condition

Compressor Extended Battery Alarm


Replace the battery.
KB0221 Charge Error Condition

KB0222 Compressor Extended Battery No Alarm


Replace the battery.
Charge Error Condition

Ventilator Primary Battery Charge Alarm


Replace the battery.
KB0223 Time too Long Condition

KB0224 Ventilator Primary Battery Charge


Notification No action required.
Time OK

KB0225 Ventilator Extended Battery Charge Alarm


Replace the battery.
Time too Long Condition

KB0226 Ventilator Extended Battery Charge


Notification No action required.
Time OK

KB0227 Compressor Primary Battery Alarm


Replace the battery.
Charge Time Too Long Condition

KB0228 Compressor Primary Battery


Notification No action required.
Charge Time OK

KB0229 Compressor Extended Battery


Alarm
Charge Time Too Long Replace the battery.
Condition

Page 35
Background
Diagnostic Codes Vent
Diagnostic Tests Corrective Action
Response
BD Id GUI Id

KB0230 Compressor Extended Battery


Notification No action required.
Charge Time OK

KB0231 Alarm Replace the USB flash drive in the


Datakey Failure
Condition Line Interface STD2 PCBA.

1. Replace the BD CPU PCBA.


BD NOVRAM operational time Alarm
KB0232 2. Replace the USB flash drive in
values out of range Condition
the Line Interface STD2 PCBA.

KB0233 GUI Volatile Operation Time values


Alarm
out of range Replace the GUI CPU PCBA.
Condition

KB0234 BD from GUI Operational Time


Alarm 1. Replace the BD CPU PCBA.
values out of range
Condition 2. Replace the GUI CPU PCBA.

KB0235 GUI from BD Operational Time 1. Replace the GUI CPU PCBA.
Alarm
values out of range 2. Replace the BD CPU PCBA.
Condition

KB0236 Alarm 1. Replace the BD CPU PCBA.


Operational Hours Failure
Condition 2. Replace the GUI CPU PCBA.

KB0237 1. Ensure the cooling fan is


Ventilator Primary Battery Charge Alarm connected.
inhibited Condition 2. Replace the primary battery.
3. Run EST.

1. Ensure the cooling fan is


KB0238 Ventilator Primary Battery No
Alarm connected.
Charge inhibited
Condition 2. Replace the primary battery.
3. Run EST.

1. Ensure the cooling fan is


KB0239 Alarm connected.
Ventilator Extended Battery Charge
Condition 2. Replace the extended battery.
inhibited
3. Run EST.

Page 36
Background
Diagnostic Codes Vent
Diagnostic Tests Corrective Action
Response
BD Id GUI Id

KB0240 Ventilator Extended Battery No 1. Ensure the cooling fan is


Charge inhibited Alarm connected.
Condition 2. Replace the extended battery.
3. Run EST.

1. Ensure the cooling fan is


KB0241 connected.
Compressor Primary Battery Alarm
2. Replace the compressor
Charge inhibited Condition
primary battery.
3. Run EST.

1. Ensure the cooling fan is


Compressor Primary Battery No
connected.
KB0242 Charge inhibited Alarm
2. Replace the compressor
Condition
primary battery.
3. Run EST.

1. Ensure the cooling fan is


Compressor Extended Battery
Alarm connected.
KB0243 Charge inhibited
Condition 2. Replace the extended battery.
3. Run EST.

1. Ensure the cooling fan is


Compressor Extended Battery No Alarm connected.
KB0244 Charge inhibited Condition 2. Replace the extended battery.
3. Run EST.

KB0245 No Event Reset – Start Reset Notification No action required.

KB0246 Inspiratory Flow Temperature out of


Notification No action required.
range reset

KB0247 Exhalation Flow Temperature out of


Notification No action required.
range reset

KB0248 Compressor Fault Reset Notification No action required.

KB0249 Background Max Event Notification No action required.

Page 37
PB980 Short Self-Test (SST)

Introduction
This chapter describes the Puritan Bennett™ 980 Ventilator System built-in self-tests, including how to run
them.

SST is a short sequence of tests designed to verify proper operation of breath delivery hardware (including
pressure and flow sensors), checks the patient circuit (including tubing, humidification device, and filters) for
leaks, measures the circuit compliance, resistance, and the resistance of the expiratory filter. SST lasts
approximately 5 minutes long. SST, in normal mode, can only be performed at start up, prior to initiation of
ventilation. The ventilator does not allow access to SST if it senses a patient is connected.

Covidien recommends running SST under these circumstances and intervals:

 Every 15 days
 Prior to ventilating a new patient
 Between patients
 When replacing the patient circuit and expiratory filter
 When connecting a different patient circuit to the ventilator
 When changing the patient circuit type
 When installing a new or sterilized expiratory filter
 When changing the humidification device type
 When adding or removing accessories to the breathing system such as a humidifier or water trap

Required Equipment

• Proposed patient circuit for patient ventilation


• Accessories (water traps, etc.)
• Expiratory filter and condensate vial
• Humidifier, if applicable
• #1 stopper to block the patient wye when prompted
• Two gas sources (air and oxygen) connected to the ventilator at a pressure between 35 psi and 87 psi
(241.3 kPa and 599.8 kPa)

Page 38
To run SST

1. Ensure a patient is NOT connected to the ventilator.

2. So that the ventilator does not detect a patient connection, ensure that the breathing circuit wye is not
attached to a test lung or covered in any way that would cause an increase in pressure at the wye.

3. Turn the ventilator on using the power switch located at the front of the BDU, below the status display. The
ventilator runs POST when the power switch is turned on. Ensure the ventilator is operating on full AC power.
Otherwise, SST test failures may result.

4. Wait at least 15 minutes to allow the ventilator to warm up and stabilize to ensure accurate results.

5. At the ventilator startup screen, touch SST or the Configure icon (wrench) displayed in the lower right area
of the GUI. The SST history log appears along with Patient Setup, Run Leak Test, and Run All SST buttons.

6. Connect the patient circuit, filters, condensate vial, and all accessories to be used in patient ventilation.
Ensure the patient wye is not blocked.

7. Touch Run All SST to perform all SST tests or touch Run Leak Test to perform the SST Leak test of the
ventilator breathing circuit.

8. Touch Accept to continue or Cancel to go back to the previous screen.

9. After accepting, touch the Circuit Type button corresponding to the patient circuit type used to perform SST
and to ventilate the patient (adult, pediatric, or neonatal.

10. Touch the Humidification Type button corresponding to the humidification type used for patient ventilation.
If no humidifier is used, touch HME. If a humidifier is used, touch Humidification Volume and turn the knob to
enter the volume.

11. Touch Accept to start SST.

12. Follow the prompts. Certain SST tests require operator intervention, and will pause indefinitely for a
response.

13. After each test, the ventilator displays the results. If a particular test fails, the test result appears on the
screen and a choice to repeat the test or perform the next test is given. When all of the SST tests are
complete, the SST status screen displays the individual test results.

14. To proceed to patient set up, (if SST did not detect an ALERT or FAILURE) touch EXIT SST, then touch
Accept or cycle the ventilator’s power.

Page 39
SST Outcomes

Final Outcome Meaning Response

PASS All SST tests passed. Touch Patient Setup to set up the patient
for ventilation:
Alert The ventilator detected one or more To override the alert, touch Override SST,
faults. Alerts occur when the ventilator then touch Accept. If touching Override
detects one or more non-critical faults. SST, the following warning may be
A non-critical fault does not jeopardize displayed: A single circuit leak test can be
the operation of the ventilator. If run, but the full suite of SST test must
detected, this type of error may be successfully pass before releasing the
acknowledged by the operator (a.k.a. ventilator for clinical use.
overridden) and ventilation will be
allowed. Choose to override the
ALERT status and authorize ventilation
only when absolutely certain this cannot
create a patient hazard or add to risks
arising from other hazards.
Fail One or more critical faults were Check the patient circuit to determine the
detected. The individual SST test did problem or restart SST with a different
not meet its requirements. Such patient circuit. Touch Repeat Test, Run All
failures may jeopardize safe operation SST, or Exit SST, then touch Accept.
of the ventilator. If detected, this type of
error must be corrected before
ventilation is allowed. The ventilator
enters the SVO state and cannot be
used for normal ventilation until SST
passes.

Page 40
PB980 SST Test Sequence

SST Test & Function Code and Corrective Actions


Message

SST Flow Sensor Cross Check

Purpose: This test verifies the following: a)


Verifies operation of the Mix O2, Mix Air,
Delivery, and Exhalation Flow Sensors; b)
Verifies Delivery PSOL liftoff command
current corresponding to establishing 0.3
lpm. At the start of this test, the safety valve
is closed, exhalation valve is open.
1. Flush Circuit With O2: Flush residual
gas from system by flowing oxygen (15 1. Ensure that the ventilator has been
liters for 18 seconds, if an adult circuit is warmed up for at least 15 minutes.
specified; 11 liters for 13.6 seconds, if a 2. Ensure that adequate source gases are
pediatric circuit is specified; 3.5 liters for present.
10 seconds, if neonatal circuit is 3. Ensure the exhalation bacteria filter
specified). latch is in the down position.
2. Establish 120, 60, 20, 5 and 1 slpm 4. Ensure that the proper patient circuit
oxygen flows. [Note: Mix Flow Sensor type and humidifier type selection
not tested at 5 and 1 slpm]. Verify Insp matches the actual equipment used.
(Delivery) flow sensor, Exp flow sensor, 5. Ensure that there are no leaks or
and Mix flow sensor are within calculated occlusions anywhere in the patient
limits. circuit (includes the inspiratory and
FAILURE (Unable to establish stable flow) FS0001 expiratory bacteria filters, humidifier,
FS0002 proximal flow sensor if present, water
FAILURE (Inspiratory flow sensor FS0003 traps, or any other accessories). If
OOR) FS0004 possible, run the SST Leak Test and
FAILURE (Exhalation flow sensor OOR) verify that it passes. If it does not,
FAILURE (Mix gas flow sensor OOR) troubleshoot the system for leaks or
3. Verify Delivery PSOL lift-off current is occlusions first.
within limit. 6. Ensure that the EVQ is properly
FS0005
FAILURE (Delivery PSOL current OOR) aligned.
4. Flush Circuit With Air: Flush residual gas 7. Exit out of SST, perform and pass a
from system by flowing oxygen (15 liters flow sensor calibration in Service
for 18 seconds, if an adult circuit is Mode. Then rerun SST from the
specified; 11 liters for 13.6 seconds, if a beginning.
pediatric circuit is specified; 3.5 liters for
10 seconds, if neonatal circuit is
specified).
5. Establish 120, 60, 20, 5 and 1 slpm
oxygen flows. [Note: Mix Flow Sensor
not tested at 5 and 1 slpm]. Verify Insp
(Delivery) flow sensor, Exp flow sensor,
and Mix flow sensor are within calculated
limits.
FS0001
FAILURE (Unable to establish stable flow)
FS0002
FAILURE (Inspiratory flow sensor FS0003
OOR) FS0004
FAILURE (Exhalation flow sensor OOR)
FAILURE (Mix “gas” flow sensor OOR)
FS0005
6. Verify Delivery PSOL lift-off current is
within limit.

Page 41
FAILURE (Delivery PSOL current OOR)

SST EV Performance

Purpose: This test verifies that the current 1. Verify no leaks or occlusions.
vs. pressure values stored in the exhalation 2. Exit out of SST and put the ventilator
valve (EV) calibration table agree with actual in Service Mode. Push the Manual
valve performance. At the start of this test,
and then Analog Data buttons. Verify
the safety valve is closed and PSOL Air is
used to pressurize the accumulator to 11 psi that the LVDT temperature displayed
th
with 21% O2. in the first column, 7 row, in screen A
1. Using the exhalation valve calibration is at least 10 degrees centigrade. If
table, verify that the target pressures the temperature is too low, allow the
established are within the specified ventilator to warm-up for a minimum
software range for Minimum and of 15 minutes then rerun SST.
Maximum pressure values.
3. Exit out of SST, perform and pass a
FAILURE (Bad Exhalation Valve Calibration FS0101
flow sensor calibration, exhalation
or Exhalation Valve Not Calibrated)
valve calibration and EST in Service
2. System reads the LVDT temperature to
Mode. Then rerun SST from the
verify it is within range.
beginning.
FAILURE (LVDT Temperature OOR) FS0105 4. Replace the EVQ. Rerun the flow
3. System generates 5 lpm for 2 seconds
and verifies stable flow. sensor calibration, exhalation valve
FS0102
FAILURE (Unable to establish stable flow) calibration, EST and SST.
4. The delivery PSOL is commanded open 5. Replace the LVDT PCBA. Rerun the
to generate the following pressures and flow sensor calibration, exhalation
the exhalation valve calibration table is valve calibration, EST and SST.
used to control the exhalation valve in 6. Replace the exhalation valve. Rerun
establishing the following pressure
the flow sensor calibration, exhalation
targets: 2 cmH2O, 5 cmH2O, 10 cmH2O,
20 cmH2O, 45 cmH2O and 95 cmH2O. valve calibration, EST and SST.
FS0103
FAILURE (Pressure OOR)

SST Circuit Pressure

Purpose: This test provides the following


functions: a) Checks inspiratory and
expiratory autozero solenoids and pressure
transducer zero offsets; and, b) Cross-
checks inspiratory and expiratory pressure
transducers at three pressure targets. At the
start of this test, the safety valve is closed
and PSOLAir is used to pressurize the
accumulator to 11 psi with 21% O2.

1. Activate the inspiratory autozero


solenoid and obtain a zero offset for PI.
Verify offset is within limits and store the 1. Verify no leaks, tubing disconnects or
reading in memory. occlusions.
FAILURE Zero offset OOR. FS0201 2. Run a complete EST: noting all
2. Activate the expiratory autozero solenoid
failures and looking for components
and obtain a zero offset for PE. Verify
offset is within limits and store the common to all the failures.
reading in memory. 3. Replace the applicable PCBA

Page 42
FAILURE if Zero offset OOR. FS0202 containing the components common
3. Using air from the accumulator, the to all the failures:
delivery PSOL is commanded open to a. IFM PCBA for Sol 1, PI, PSV
generate the following target pressures:
failures.
10 cmH2O, 50 cmH2O, 100 cmH2O.
FAILURE if pressure OOR or timeout occurs: FS0203 b. Exhalation Sensor PCBA for Sol
Target pressure not achieved. 2 or PE failures.
4. PI and PE readings are verified to be
within test limits.
FAILURE if inspiratory and expiratory FS0205
pressure sensors differ outside test limits.
5. The exhalation pressure transducer is
verified to be within test specifications of
the inspiratory pressure transducer.
6. The safety valve pressure transducer is
verified to be within test limits.
7. The safety valve pressure transducer is
verified to be within test specifications of
the inspiratory pressure transducer.
8. With the system pressurized to
100cmH2O, autozero solenoids are
activated and zero offsets are logged for
PI and PE.
FAILURE if inspiratory and Safety Valve FS0204
pressure sensors differ outside test limits.
FAILURE if auto zero solenoid is not FS0206 (Insp
operational or if readings are not within Autozero)
range.
FS0207 (Exh
Autozero)

SST Leak Test

Purpose: Tests ventilator breathing system


for leaks
1. Ensure that the inspiratory port (FD 1) is
1. User prompted to block the inspiratory securely blocked with a #1 or #2
filter. Stopper.
2. Mix Accumulator pressure set to 11 psig. 2. Verify no leaks, tubing disconnects or
3. System closes the safety valve and occlusions.
samples the inspiratory pressure 3. Use a known good patient circuit and
transducer to verify the average value for bacteria filter set.
the Delivery PSOL leak is within test
4. Exit out of SST and run a complete
limits.
a) >28 cmH2O = Delivery PSOL leak test EST: noting all failures and looking for
failure components common to all the failures.
b) >14 cmH2O = Delivery PSOL leak test
alert
FAILURE (Delivery PSOL leak exceeds FS0303
failure limit)
ALERT (Delivery PSOL leak exceeds alert AS0303
limit)
4. Mix Accumulator pressure set to 25 psig.
FAILURE (Unable to establish accumulator
pressure)
System closes the safety valve and samples
the inspiratory pressure transducer to verify

Page 43
the average value for the Delivery PSOL leak
is within test limits.
a) >100 cmH2O = Delivery PSOL leak
test = Failure
b) >83 cmH2O = Delivery PSOL leak test
= Alert
6. Mix Accumulator pressure readjusted to
11 psig.
7. System tests the safety valve cracking
pressure.
8. System begins the Leak Test.
9. PI and PE transducers are auto zeroed.
10. Safety and exhalation valves are
closed.
11. System establishes a target of
90cmH2O using 2 lpm air and measuring
inspiratory pressure.
FAILURE (Unable to establish pressure) FS0303
12. Verifies no more than a 5cmH2O
pressure drop within 10 seconds. FS0304
FAILURE (Excessive Leak)
13. Verifies that the delta pressure
between PI and PE are within limits. Open AS0305
exhalation valve.
ALERT (System Leak Alert)
14. Run the Check Valve Reverse Flow
test. Verify that the pressure drop is
within limits and that the reverse flow rate
is less than .1 slpm at a back pressure of
95 cmH2O.

SST Exhalation Filter Test

Purpose: This test checks for exhalation


filter and exhalation compartment occlusion.
1. User prompted to disconnect circuit at
expiratory filter.
2. System auto zeroes the inspiratory and
expiratory autozero solenoids.
3. With the safety valve closed, system
establishes 60 lpm if Adult/Ped circuit
selected and 30 lpm if Neonate circuit
selected.
4. Measure the pressure drop at the
inspiratory module and verify that the
circuit was disconnected by checking if
the Exhalation Flow Sensor is reading <1
lpm.
5. User prompted to reconnect circuit to
exhalation filter.
6. With the safety and exhalation valves
closed, system establishes 60 lpm if
Adult/Ped circuit selected and 30 lpm if
Neonate circuit selected.
7. Measure the pressure drop at the
expiratory module and verify that the
expiratory flow sensor is reading 80% of
Page 44
target.
a. Verify that exhalation compartment is
not occluded.
i. Test fails if resistance > 7 cmH 2O for 1. Verify that there are no occlusions in
Adult/Ped circuit or > 4 cmH2O for the exhalation compartment.
Neonate circuit. 2. Use a known good Expiratory Filter.
FAILURE (occluded expiratory compartment) FS0403 3. Verify that the exhalation valve
diaphragm is not soiled or damaged.
ii. Test generates an alert if resistance 4. Run EST.
> 6 cmH2O for Adult/Ped circuit or > 5. Run exhalation valve calibration.
3 cmH2O for Neonate circuit.
ALERT (occluded expiratory compartment) AS0408
b. Calculate the pressure drop across
the exhalation filter.
ii. Test fails if resistance > 3 cmH 2O, or
< 0.1 cmH2O
FAILURE (occluded expiratory filter). FS0404
ALERT (occluded expiratory filter). AS0405
FAILURE (low expiratory filter resistance) FS0410
iii. Test generates an Alert if resistance
> 2.4 cmH2O, ALERT (occluded
expiratory filter) or < 0.4 cmH 2O.
ALERT (low expiratory filter resistance) AS0411

SST Circuit Resistance

Purpose: Checks for inspiratory and


expiratory limb occlusions, and calculates
and stores the inspiratory and expiratory limb
resistance parameters.

1. Pressurizes system to 100 cmH2O at 60 Block wye when prompted.


L/min. Failure: Unable to pressurize
system within time-out period. Check for circuit leaks.

Test: Characterizes total circuit resistance


over a range of flows using either gas.

2. Failure: Peak flow < alert threshold but >


failure threshold:
i. Alert threshold (adult/pediatric) = AS0507
80 L/min Check for kinked or occluded patient
ii. Alert threshold (neonatal) = No tubing.
alert state for neonatal; pass or AS0511
fail only
iii. Failure threshold (adult/pediatric)
= 60 L/min; Failure threshold FS0504
(neonatal) = 20 L/min; Failure:
Peak flow < failure threshold.
iv. Failure threshold (adult/pediatric) FS0510
= 60 L/min; Failure threshold
(neonatal) = 20 L/min

User prompted to unblock wye.

3. Test: Measures pressure drop across


inspiratory limb at 60 L/min (adult), 30
Page 45
L/min (pediatric), or 10 L/min (neonatal).
Failure: Flow controller unable to
establish and control gas flow.

Test: Determines inspiratory limb pressure


drop, and checks for an occlusion.

i. Alert: Inspiratory limb pressure > 8.5


cmH2O (adult) or 5.5 cmH2O AS0503 Ensure a known good inspiratory filter is
(pediatric), but < 12.5 cmH2O (adult) connected.
or 7.5 cmH2O (pediatric), indicating
occlusion. No Alert state for
neonatal; pass or fail only

ii. Failure: Inspiratory limb pressure > FS0501 Check for occluded patient tubing.
12.5 cmH2O (adult), 7.5 cmH2O
(pediatric), or 3.5 cmH2O (neonatal),
indicating occlusion.

iii. Failure: Inspiratory limb pressure < FS0502


0.6 cmH2O (adult) or 0.5 cmH2O Replace patient circuit.
(pediatric), but 0.2 cmH2O. No Alert
state for neonatal; pass or fail only.

iv. Failure: Inspiratory limb pressure < FS0508


0.2 cmH2O.

SST Circuit Compliance

Purpose: Calculates the attached patient


circuit compliance.

User prompted to block wye.


Check patient circuit for leaks. Replace
Test: Measures circuit compliance using as necessary.
10 L/min flow, storing pressure and volume
measurements at 5 cmH2O intervals from 5
to
100 cmH2O. Uses air as default gas.

Failure: System cannot pressurize to one of FS0601 Repeat test, making sure circuit is
the test pressure points. undisturbed during test.

Test: Measures circuit compliance using a Repeat SST to retest pressure


high flow, storing pressure and volume transducers.
measurements at 5 cmH2O intervals from 5
to 100 cmH2O. Run EST and verify all pneumatics tests
pass.
Failure: System cannot pressurize to one of FS0602
the test pressure points.

Failure: Out-of-range pressure ratio FS0603


calculation. Make sure correct patient tubing type
was specified in SST. Replace patient
Test: Determines compliance and checks its circuit.
reasonableness.

Page 46
Alert: High compliance > 6 mL/cmH2O AS0605 Make sure correct patient tubing type
(adult) or 4.5 mL/cmH2O (pediatric), or 2.5 was specified in SST. Replace patient
mL/cmH2O (neonatal) but < 12 mL/cmH2O circuit.
(adult) or 9 mL/cmH2O (pediatric), or 3
mL/cmH2O (neonatal).
Make sure correct patient tubing type
was specified in SST.
Failure: High compliance > 12 mL/cmH2O FS0604
(adult), 9 mL/cmH2O (pediatric), or 3 Replace patient circuit with a known
mL/cmH2O good circuit and filter set.
(neonatal).
Run EST to check the operation of PI
and PE during the Circuit Pressure Test.
Alert: High compliance falls below AS0607 Replace applicable pressure transducer.
1.56 mL/cmH2O (adult), 1.34 mL/cmH2O
(pediatric), or 0.6 mL/cmH2O (neonatal) but Run Atmospheric Pressure Transducer
not lower than 1.05 mL/cmH2O (adult or calibration.
pediatric) or 0.25 mL/cmH2O (neonatal).
Make sure correct patient tubing type
Failure: High compliance falls below 1.05 FS0606 was specified in SST. Replace patient
mL/cmH2O (adult or pediatric) or 0.25 circuit.
mL/cmH2O (neonatal).

Alert: Low compliance > 6 mL/cmH2O AS0609


(adult), 4.5 mL/cmH2O (pediatric), or 2.5 Make sure correct patient tubing type
mL/cmH2O (neonatal), but < 12 mL/cmH2O was specified in SST.
(adult), 9 mL/cmH2O (pediatric), or 3
mL/cmH2O (neonatal). Replace patient circuit with a known
good circuit and filter set.
Failure: Low compliance > 12 mL/cmH2O FS0608
(adult), 9 mL/cmH2O (pediatric), or 3 Run EST to check the operation of PI
mL/cmH2O (neonatal). and PE
during the Circuit Pressure Test. Replace
Alert: Low compliance falls below applicable pressure transducer.
1.56 mL/cmH2O (adult), 1.34 mL/cmH2O AS0611
(pediatric), or 0.6 mL/cmH2O (neonatal) but Run Atmospheric Pressure Transducer
not below 1.05 mL/cmH2O (adult or calibration.
pediatric) or 0.25 mL/cmH2O (neonatal).
Replace patient circuit with a known
good circuit and filter set.
Failure: Low compliance falls below
1.05 mL/cmH2O (adult or pediatric) or 0.25 FS0610
(neonatal).

Page 47
Note: for any Alerts or Failures generated while performing any of the SST tests below, here are some
general troubleshooting guidelines:

1. Ensure that the ventilator has been warmed up for at least 15 minutes.
2. Ensure that adequate source gases are present.
3. Ensure that there are no leaks or occlusions anywhere in the patient circuit (includes the inspiratory
and expiratory bacteria filters, humidifier, proximal flow sensor if present, water traps, or any other
accessories).
4. Ensure the exhalation bacteria filter latch is in the down position.
5. Ensure that the proper patient circuit type and humidifier type selection matches the actual equipment
used.
6. Use a known good inspiratory filter, expiratory filter and patient circuit.
7. Ensure that the EVQ is properly aligned.
8. Run EST.
9. Run EST, Flow Sensor and/or Exhalation Valve Calibrations as needed to troubleshoot the ventilator.

Page 48
EST (Extended Self Test)

Introduction

EST is a user-initiated self test that verifies the integrity of the ventilator’s subsystems using operator
participation. EST checks the pneumatics, memory, safety system, indicators, digital and analog electronics,
power supplies, transducers and options. EST is intended to be run by trained service personnel.

The three primary purposes for running EST are:

1. To verify the integrity of PB980 ventilator pneumatics and electronics systems.


2. To troubleshoot the ventilator.
3. To reset “Pre-Existing Device Alerts” preventing the ventilator from operating in normal operation.

NOTE: When the PB980 generates a VENT INOP, GUI INOP or BUV condition, at a minimum, a complete
EST must be performed before normal operation is allowed. Within EST, there are 27 possible tests total.

EST requires a “gold standard” test circuit, available from Covidien. All required software support to perform
an EST is resident on the ventilator. EST testing, excluding tests of optional equipment, such as the
compressor, takes about 10 minutes to complete.

EST contains a feature that allows one or more EST tests to be run individually. This feature is called Single
Test EST. Single Test EST may be run to verify if a particular system or part may be causing an EST fault or
failure. The ventilator software requires that when any Single Test EST test is initiated, EST must be
completed in its entirety and all individual tests must pass before normal operation is allowed.

NOTE:
EST must always be performed with an ADULT exhalation filter installed. Attempting to run EST with a
NeoMode filter can cause some EST tests to fail.

WARNING:
Always disconnect the ventilator from the patient before running EST or Single Test EST. Running EST or
Single Test EST while the ventilator is connected to the patient can injure the patient. A fault identified in EST
indicates that the ventilator or an associated component is defective. A defective ventilator or associated
component should be repaired before the ventilator is returned to service, unless it can be determined with
certainty that the defect cannot create a hazard for the patient, or add to the risks
which may arise from other hazards.

NOTE:
If the ventilator has not reached operating temperature from recent usage, allow it to warm up for at least 10
minutes in service mode before running EST or single Test EST to ensure accurate testing.
Be aware that each time EST is performed, the patient alarm log is cleared.

When to run EST

• Prior to initial installation of a new ventilator


• Every six months
• Before and after any preventive maintenance
• Following ventilator service or repair

Page 49
Hardware requirements

Running EST requires the equipment listed in the table below. High pressure, compressed air and oxygen are
required for EST. Always use an adult exhalation bacteria filter and gold standard circuit to conduct EST. Use
of other hardware can cause some EST tests to fail.

Description Manufacturer
Gold Standard Test Circuit 4-018506-00

Stopper, wye (no.2) 4-009523-00

Compressed air and oxygen source both at 35 to


87 psi (241 to 599 kPa)
TM
Note: Due to excessive restriction of Air Liquide , Local Supplier
TM
Drager and SIS air/oxygen hose assemblies,
certain EST tests may fail when supply pressures
< 50 psi (345 kPa) are used in conjunction with
those hoses.

Adult/Pediatric expiratory filter and vial 10063033

or

Pediatric-Adult expiratory filtration system (carton 10043551


of 12)

Note:
If the ventilator has not reached normal operating temperature from recent usage, allow it to warm up for at
least 15 minutes in service mode prior to running EST to ensure accurate testing.

Page 50
How to run EST

1. Ensure the ventilator is not connected to the patient.


2. Before running EST, verify the following:
a. The ventilator is operating on AC power.
b. Ensure the primary ventilator battery is at least 90% charged. All green bars on the battery
charge indicator should be lit.
c. Ensure that the ventilator has been warmed up for a minimum of 15 minutes.
d. By viewing the status display in Service Mode, verify that the compressed air and oxygen
sources are both at 35 to 87 psi (241 to 599 kPa).

3. EST is run from Service Mode. If not already in Service Mode, enter Service Mode by turning the
ventilator power switch on, and then pressing and immediately releasing the service switch located in
the upper left corner of the BDU. Do not hold the service switch in.

Page 51
4. Install the gold standard test circuit from the To Patient and From Patient ports. NOTE: Do not install
an inspiratory filter.
5. Select the SELF TESTS tab from the horizontal banner at the top of the Service Mode screen.
6. Select the EST tab.
7. Press RUN ALL button to run all tests in sequence.
8. Press ACCEPT.
9. Follow prompts to complete EST. Some EST tests require operator input and will pause indefinitely
for a response.

EST Outcomes

Final Outcome Meaning Response

PASS All EST tests passed. EST successfully completed.

ALERT / The ventilator detected one or more To override the alert, touch Override EST,
OVERRIDDEN faults. Alerts occur when the ventilator then touch Accept.
detects one or more non-critical faults.
A non-critical fault does not jeopardize
the operation of the ventilator. If
detected, this type of error may be
acknowledged by the operator (a.k.a.
overridden) and ventilation will be
allowed. Choose to override the
ALERT status and authorize ventilation
only when absolutely certain this cannot
create a patient hazard or add to risks
arising from other hazards.
FAIL One or more critical faults were Troubleshoot and repair the problem, then
detected. In normal operation, the pass a complete EST.
ventilator enters the SVO state and
cannot be used for normal ventilation
until EST passes. Service is required.

Page 52
EST TESTS:

1. Zero Offset
2. Leak
3. Mix Leak
4. Mix PSOL
5. Mix Accumulator
6. Circuit Pressure
7. Flow Sensor Cross Check
8. Delivery PSOL
9. Exhalation Valve Loopback
10. Exhalation Valve Pressure Accuracy
11. Exhalation Valve Performance
12. Exhalation Valve Velocity Transducer
13. Safety System
14. Backup Ventilation
15. Communication
16. Internal Storage
17. LCB Backlight
18. Status Display
19. GUI Audio
20. BD Audio
21. Rotary Knob Test
22. Offscreen Key Test
23. Ventilator Battery
24. Compressor Battery
25. Compressor
26. Compressor Leak
27. Compressor Performance

Page 53
1. Test #1 Zero Offset Test:

Test Objectives: tests inspiratory and expiratory pressure transducers and flow sensors at ambient
pressure. Also exercises the PSOLs to “loosen” them up since the ventilator may have been in storage for an
extended period of time.

Page 54
1.1.1.1 Zero/Offset Pressure Sensors

Ref# Number Value

1 Inspiratory pressure <+/-10cmH2O

2 Exhalation pressure <+/-10cmH2O

3 Mix air inlet pressure <+/-1.5psig

4 Mix O2 inlet pressure <+/-1.5psig

5 Mix Accumulator pressure <+/-0.5psig

6 Insp & Exp. Manifold drain time 5 sec

7 Compressor Accum. drain time 45 sec

8 Stabilization time 5 sec

9 Open PSOL counts 2000

10 After Compressor drain settling time 2 sec

Table 3-1

 The offset of the inspiratory pressure transducer shall be verified to be within + 10 cmH2O at ambient
pressure.
 The offset of the exhalation pressure transducer shall be verified to be within + 10 cmH2O at ambient
pressure.
 The air and oxygen pressure transducers shall be verified to be within +/-1.5 psig when placed at
ambient pressure conditions.
 The accumulator pressure transducer shall be verified to be within +/-0.5 psig when placed at ambient
pressure conditions.

1.1.1.2 Flow Sensor Zero Offset

Ref# Number Value

1 QINSP <+/-0.125 lpm

2 QEXLH <+/-0.200 lpm

3 QAIR <+/-0.125 lpm

4 QO2 <+/-0.125 lpm

Table 3-2

 The offset of the inspiratory, exhalation, air and oxygen flow sensors shall be verified to be less than
0.05 slpm.
Method
2. Zero/Offset Pressure Sensors – Use Table 3-1

Page 55
a. Prompt user to disconnect all gasses
b. If compressor is installed, and activated
i. Deactivate the compressor (0 psig)
ii. Open the Safety Valve
iii. Open the Delivery PSOL to [9]
iv. Wait up to [7] for compressor to drain
v. Wait [10]
vi. Close the Safety Valve
c. Drain the system of pressure
i. Open Safety Valve
ii. Open Purge Valve
iii. Open Delivery Psol to [9]
iv. Open Exhalation Valve
v. Wait up to [6] for drain
d. Verify gasses are disconnected
i. If not, FAILURE: “Gas” still connected. NOTE: do not use air and oxygen high
pressure hoses with one-way check valves in line.
1. FE0414 – Failure if Air still connected. Disconnect Air when prompted.
2. FE0415 – Failure if O2 still connected. Disconnect Oxygen when
prompted.
e. Wait [8]
i. Collect 5 samples of the Air inlet pressure and verify zero/offset of the pressure
sensor is within limits.
1. FE0405 - inspiratory pressure sensor zero offset failed.
a. Ensure tubing is properly connected to Sol 4 the Inspiratory Autozero
Solenoid and to PI.
b. Replace the IFM PCBA.
2. FE0406 - exhalation pressure sensor zero offset failed.
a. Ensure tubing is properly connected to Sol 5 the Expiratory Autozero
Solenoid and to PE.
b. Replace the Exhalation Sensor PCBA.
3. FE0401 - air inlet pressure sensor zero offset failed.
a. Ensure the vent for R1 is unobstructed.
b. Replace the Gas Supply Sensors PCBA.
4. FE0402- oxygen inlet pressure sensor zero offset failed.
a. Replace the Gas Supply Sensors PCBA.
5. FE0404 - mix accumulator pressure sensor zero offset failed.
a. Ensure Solenoid 1 is properly connected.
b. Replace the IFM PCBA.
3. Flow Sensor Zero Offset – Use Table 3-2
a. Collect 5 samples of the Air inlet pressure and verify that the average value is within limits.
1. FE0412 - inspiratory flow sensor zero offset OOR
a. Reseat the delivery PSOL to ensure it is not leaking.
b. Check for leaks at the BUV Solenoid 3.
c. Replace the delivery flow sensor.
d. Replace the delivery PSOL.
2. FE0413 - exhalation flow sensor zero offset OOR
a. Reseat the EVQ.
b. Replace the EVQ.
3. FE0409 - air flow sensor zero offset OOR
a. Replace the air flow sensor.
4. FE0410 - O2 flow sensor zero offset OOR
a. Replace the oxygen flow sensor.
b. Enable Compressor and controls.

Page 56
2. Test #2 Leak Test:

Test Objective: using air, tests the system’s ability to hold pressure.

Page 57
1.1.1.3 Delivery PSOL Leak

Ref# Number Value

1 Open SV Delay 1s

2 Close SV Delay 3.125 s

< 28 cmH2O for Failure


3 Insp Pressure (11psig)
< 14 cmH20 for Alert

4 Pressure Stability Time 3 sec

< 100 cmH2O for Failure


5 Insp Pressure (25psig)
< 83 cmH20 for Alert

6 Flow Stability Time 2 sec

7 Safety Valve cracking timeout 5 sec

8 Minimum cracking pressure 112 - (112 * 0.0455)

9 Maximum cracking pressure 125 + (125 * 0.0455)

Table 3-3

 The leak of the delivery PSOL shall cause less than 14 cmH20 (Alert) and 28 cmH20 (Failure) of
pressure to build between the PSOL and a blocked “To-Patient” port at input pressure of 11 psig;
 The leak of the delivery PSOL shall cause less than 83 cmH20 (Alert) and 100 cmH20 (Failure) of
pressure to build between the PSOL and a blocked “To-Patient” port at input pressure of 25 psig ; (
25 psig is the max pressure used during Inspiratory BUV)

1.1.1.4 Leak Test

Ref# Number Value

1 Initial Pressure 90 cmH2O

2 Flow 2 lpm

3 Pressurization Timeout < 100 + (((compliance * Adjusted Pressure) / Adjusted Flow) * 60)

4 Pressure wave decay time 200 ms

5 Start test pressure 85 cmH2O

6 Wait time to reach start pressure 10 s

7 Wait time after start pressure 10 s

8 Pressure drop < 5.0 cmH2O

Page 58
9 Insp/Exp Delta Max (insp Pressure + exp Pressure) * 0.0455 + 2.0 * 0.169

Table 3-4

 The leak rate of the inspiratory and exhalation subsystems shall be verified to be less than 0.042
slpm at an initial pressure of 90 cmH 2O.
 “gold standard” test circuit compliance = 1.4 ml/H2O, pressure drop = 5cmH2O, test time 10 sec,
hence the average leak at pressure 80cmH2O is: (1.4 * 5 / 10) * 60 = 42 cc/min.

1.1.1.5 Check Valve Reverse Flow

Ref# Number Value

1 Initial Pressure 95 cmH2O

2 Flow 5 lpm

3 < 100 + (((compliance * Adjusted Pressure) / Adjusted Flow) *


Pressurization Timeout
60)

4 Pressure wave decay


300 ms
time

5 Start test pressure 85 cmH2O

6 Full Decay Pressure 5 cmH2O

7 Full Decay Timeout 600 ms

8 Too Fast time < 5 cmH2O before 150 ms

9 Insp/Exp Delta Max (insp Pressure + exp Pressure) * 0.0455 + 2.0 * 0.169

Table 3-5

 The Check Valve is expected to have a leak.


 That leak cannot be too high or too low.
 At 95 cmH20, the pressure shall drop to 85 cmH20 in less than 300 ms, the pressure shall drop to 5
cmH20 in less than 600 ms.
 The pressure shall be greater than 5 cmH2O after 150 ms.
 The flow rate of the check valves shall be verified to be less than (0.1 slpm) at a back pressure of 95
cmH2O.
Method
1. Delivery PSOL Leak – Use Table 3-3
a. Activate over pressure switch bypass
b. Prompt operator to block “To-Patient” port.
c. Open the safety valve.
d. Wait [1]
e. Close the safety valve.
f. Wait [2]
g. Take a 5 sample average of the inspiratory pressure sensor.
h. Verify that the average value is within the test limits [3]
i. FE0601 – Failure if the leak rate exceeds failure limit > 28 cmH 2O

Page 59
1. Advance thru the first six EST tests to isolate where the leak is. If the system
is failing the Mix leak test, troubleshoot leaks in the Mix Subsystem, the
Delivery PSOL, and the BUV Solenoid 3.
2. Keep the cork plugged in the ventilator outlet to determine if the leak is in the
Delivery Subsystem. If the test continues to fail the leak test, but pass the
Mix leak test, verify the following:
a. that the delivery check valve is properly seated [NOTE: look at the
check valve seating by removing the internal bacteria filter at the
ventilator outlet];
b. verify the oxygen sensor is not leaking;
c. verify all IFM PCBA tubing connections;
d. check for no leaks at Solenoid 1 and 2;
e. check/replace the Safety Valve;
3. If the test passes with the cork in the ventilator outlet but fails when the gold
standard circuit is connected to the exhalation subsystem, rerun the test
while blocking the expiratory filter. If the system builds and holds pressure
with the expiratory filter blocked, replace the expiratory filter.
4. If the test still fails with the expiratory filter blocked, ensure the EVQ seals are
in place and properly seated. If yes, rerun the test after firmly slamming the
exhalation module latch down.
5. Try a new EVQ.
6. Try a new exhalation valve.
ii. AE0602 - ALERT if greater than 14 cmH 2O (Delivery PSOL leak exceeds alert
limit)
1. Perform the steps above.
i. Set the mix to 25 psig
j. Wait for accumulator pressure to stabilize or timeout of [4]
i. If not, FE0603 – FAILURE: Unable to establish accumulator pressure
1. Advance thru the first six EST tests to isolate where the leak is. If the system
is failing the Mix leak test, troubleshoot leaks in the Mix Subsystem, the
Delivery PSOL, and the BUV Solenoid 3.
2. Verify PMX pressure transducer connection
3. Ensure the mix check valve is seated properly
4. Ensure the mix accumulator is seated properly
k. Open the safety valve.
l. Wait [1]
m. Close the safety valve.
n. Wait [2]
o. Take a 5 sample average of the inspiratory pressure sensor.
p. Verify that the average value is within the test limits [5]
ii. FE0604 – FAILURE: at 25 psig, Delivery PSOL leak exceeds limit if pressure
established is greater than 100 cmH2O.
1. Reseat the Delivery PSOL
2. Replace the Delivery PSOL
i. AE0605 – ALERT: at 25 psig, Delivery PSOL leak exceeds limit if pressure
established is greater than 83 cmH2O.
1. Reseat the Delivery PSOL
2. Replace the Delivery PSOL
q. Set the mix to 11 psig
r. Close safety valve.
s. Establish a delivery test flow of 1.0 lpm with air.
t. Wait [5]. FE0609 – Failure: unable to establish stable flow
i. Verify no upstream restrictions to the high pressure air supply
ii. Swap the air and oxygen flow sensors, and then rerun the test. If test passes, return
oxygen flow sensor to its original position and replace the air flow sensor.
iii. Swap the air and oxygen mix PSOL’s. If test passes, return oxygen PSOL to its
original position and replace the air mix PSOL.
u. Wait for the safety valve to open or timeout of [6].

Page 60
v. Record the cracking pressure at every check for SV open.
w. Stop the flow controller.
iii. FE0610 – FAILURE: Safety Valve failed to open
i. Verify system builds pressure by running the first 6 tests in EST. If the system fails to
build pressure, verify all PSOL’s can generate pressures. If the PSOL’s can build
pressure, troubleshoot the leak.
ii. Verify the oxygen sensor is not leaking;
iii. Rerun the test by keeping the cork in the ventilator outlet to isolate leaks in the
delivery subsystem. Verify proper delivery check valve seating.
iv. Replace the safety valve.
x. Send the cracking pressure to GUI.
i. If the cracking pressure is less than [7] or greater than [8], FE0611 – FAILURE:
Safety Valve cracking pressure OOR.
1. Replace the safety valve
y. Prompt operator to unblock “To-Patient” port and reconnect test circuit.
2. System Leak Test – Use Table 3-4
a. Autozero the pressure transducers
b. Close the exhalation valve (with EV controller)
c. Establish [1] with either gas using [2] flow and measuring inspiratory pressure.
d. Wait up to the calculated pressurization timeout [3] for pressure to build.
i. FE0603 – FAILURE: Unable to establish pressure
1. Repeat steps above to troubleshoot the system’s ability to build and hold
pressure.
e. Stop pressure control
f. Wait [4] for pressure wave to decay.
g. Read inspiratory pressure sensor
h. Verify the initial inspiratory pressure is within the test limits [5]
i. FE0603 – FAILURE: Unable to establish pressure
1. Repeat steps above to troubleshoot the system’s ability to build and hold
pressure.
i. Wait [6] or until inspiratory pressure drops below [5].
i. If timeout, then test passes. Skip the rest of the test.
j. When pressure reaches [5], wait additional [7]
k. Verify pressure drops less than [8]
i. FE0604 – FAILURE: Unable to establish pressure
1. Repeat steps above to troubleshoot the system’s ability to build and hold
pressure.
l. Verify the delta between the inspiratory and expiratory pressures is below [9].
i. AE0605 – ALERT: Test circuit not connected
1. Connect good test circuit.
2. Repeat steps above to isolate if the leak is in the inspiratory or the expiratory
subsystem.
m. Open Exhalation valve
3. Check Valve Reverse Flow – Use Table 3-5
a. Close Exhalation valve (with EV controller)
b. Establish a pressure of [1] with [2] delivery flow, using inspiratory pressure sensor
c. Wait up to the calculated pressurization timeout [3] for pressure to build.
i. FE0603 – FAILURE: Unable to establish pressure
1. Repeat steps above to troubleshoot the system’s ability to build and hold
pressure.
d. Stop the pressure controller
e. Open the Safety Valve
f. Wait for pressure to drop below [5] or [4]
i. If timeout occurs, FE0606 – FAILURE: Inspiratory Check Valve reverse flow too
low
1. Verify proper seating of the delivery (inspiratory) check valve.
g. Wait for pressure to drop below [6] or [7]

Page 61
i. If pressure reaches [8], FE0607 – FAILURE: Inspiratory Check Valve reverse flow
too high
1. Verify proper seating of the delivery (inspiratory) check valve.
2. Replace the delivery check valve
ii. If pressure doesn’t reach [6] in [7], FE0606 – FAILURE: Inspiratory Check Valve
reverse flow too low:
1. Verify proper seating of the delivery (inspiratory) check valve.
2. Replace the delivery check valve

h. Open the Exhalation Valve


i. Close the Safety Valve.

Page 62
3. Test #3: Mix Leak Test

Test Objectives: verifies the integrity of the Mix System and checks that the Mix
System leaks are within the specified allowable limits.

Page 63
1.1.1.6 Mix PSOL Leak

Ref# Number Value

1 PSOL open 2000 counts

2 Pressure decay time 2s

3 Mix Air flow < 0.1 slpm

4 Mix O2 flow < 0.1 slpm

5 Accumulator Pressure 11 psig

6 Max Target Pressure Error < 0.5 psig

7 Delivery flow < 0.1 slpm

8 Damping Gain 0.8

9 O2 Only Percentage 100

10 Air Only O2 Percentage 21

Table 3-1

 PSOL leak test will be performed with accumulator at ambient pressure, which is worst case
condition.
 The leak rate of the air PSOL shall be verified to be less than 0.1 slpm.
 The leak rate of the O2 PSOL shall be verified to be less than 0.1 slpm.

1.1.1.7 Mix Accumulator Leak

Ref# Number Value

1 Pressure reached timeout 5 sec

2 Pressure stabilization time 3 sec

3 Accumulator pressure error < +/- 0.5 psig

4 PSOL open value 1000 counts

5 Target Pressure 25 psig

6 Pressure Build Time 1 sec

Table 3-2

 The accumulator leak shall be less than .5 psig when charged to 25 psig. (This includes leaks
causing a pressure rise and leaks causing a pressure drop.)

Page 64
1.1.1.8 Option Gas Supply Solenoid Test

Ref# Number Value

1 Target Pressure 11 psig

2 Delivery PSOL open 1000 counts

3 Mix PSOL open 1000 counts

4 Pressure build time 5s

5 Pressure stability timeout 2000 ms

6 Pressure drop time 3s

7 Max leak 0.4 psig

Table 3-3

 The leak rate of the option gas supply solenoid shall be verified to be less than 0.4 psig when the mix
accumulator is pressurized to 11 psig.

1.1.1.9 Check Valve Reverse Flow

Ref# Number Value

1 Pressure reached timeout 5s

2 Mix inlet pressure < 1.5 psig

3 Mix PSOL open 3500 counts

4 Check Valve stabilization time 1s

5 Mix flow < 1 slpm

6 Delivery PSOL open 2000 counts

7 Compressor Drain Time 45 sec

8 Compressor settling time 2 sec

9 Min O2 inlet connected pressure 35 psig

10 Pressure Decay time 2 sec

11 Mix Accumulator Target Pressure 11 psig

12 System Settling time 3 sec

Table 3-4

Page 65
 The flow rate of the mix check valves shall be verified to be less than 1 slpm when the accumulator is
pressurized to 11 psig.
Logged Parameters
 Accumulator leak rate, leak rate limit and elapsed time.
 Check valve reverse flow.
Method
1. Mix PSOL Leak – Use Table 3-1
a. Activate over pressure switch bypass
b. Put the mix in normal mode at [5] pressure
c. Set exhalation valve damping gain to [8]
d. Set mix O2 percentage to [9]
e. Open Delivery PSOL to [1]
f. Wait [2]
g. Read accumulator pressure
h. Verify accumulator pressure is within limits [6]
i. If not, FAILURE (Unable to establish stable accumulator pressure) FE0011 - if
less than minimum. FE0012 - if greater than max
ii. Check for leaks in the Mix Accumulator Subsystem:
1. Verify the integrity of the PMX pressure sensor tubing connections at
the IFM PCBA and at the accumulator manifold.
2. Verify Solenoid 2 is properly connected and assembled with its gasket.
3. Verify the Mix Check valve assembly is installed properly and is not
leaking or damaged.
4. Verify the Wall Air and Compressor check valves are not misaligned.
5. Verify the Mix Accumulator Relief Valve is not leaking.
6. Verify the Mix Air PSOL and Mix Air Flow Sensor is seated properly.
i. Set mix O2 percentage to [10]
j. Read accumulator pressure
i. Verify accumulator pressure is within limits [6] if less than min FE0013 AND if
greater than max FE0014
k.
i. If not, FAILURE (Unable to establish stable accumulator pressure)
1. Verify the Mix O2 PSOL and Mix Air Flow Sensor is seated properly.

l.Close BD PSOL
m.Wait [2]
n.Read Delivery flow sensor
o.Verify Delivery flow is within limits [7]
i. If not, FAILURE (Inspiratory flow too high) FE0016
1. Exit EST and perform a Flow Sensor Calibration, then rerun EST.
2. Replace the Delivery Flow Sensor.
p. Put the mix in test mode
q. Close all mix psols
r. Wait [2]
s. Read accumulator pressure
t. Verify accumulator pressure drop is less than [6] from initial reading
i. If not, FAILURE (Accumulator has excessive leak) FE0005
u. Take a 5 sample average of the air and O2 flow sensors
v. Verify that the average value is within the test limits ([3] for air FE0008, [4] for O2 FE0009)
i. If not, FAILURE (“Gas type” PSOL leak rate exceeds test limits)
ii. Swap Mix PSOL’s to see if the failure code changes. If it does, replace the
appropriate Mix PSOL identified by the EST failure.
2. Mix Accumulator Leak – Use Table 3-2
a. Close the safety valve
b. Close the purge valve
c. Open the delivery PSOL to [4]

Page 66
d. Put the mix controller in normal mode
e. Set the mix accumulator pressure to [5]
f. Wait [6] for pressure to build up
g. Wait until the pressure reaches [3] from target or timeout of [1]
i. If not, FAILURE (Unable to reach “target pressure”) FE0004
ii. Look for leaks in the Mix Subsystem:
1. PMX tubing
2. Solenoid 1 tubing
3. Solenoid 2 tubing
4. Proper Mix Checkvalve assembly
5. Mix flow sensor and PSOL orings
6. Mix Pressure Relief Valve
7. Delivery PSOL
8. BUV Solenoid 3

h. Decrement the delivery PSOL counts by 2 every 10 milliseconds until flow drops below 1lpm
or until PSOL is completely shut
i. Close the delivery PSOL
j. Put the mix controller in test mode
k. Close all mix psols
l. Wait [2] for stabilization
m. Take a sample of the mix accumulator pressure
n. Wait [2]
o. Take a 5 sample average of the mix accumulator pressure
p. Verify that the average value is within the test limits [3] of initial pressure
i. If not, FAILURE (Mix accumulator leak exceeds limit) FE0005
1. Swap the Mix PSOL’s
2. Replace the leaky Mix PSOL
3. Swap Mix flow sensors

3. Option Gas Supply Solenoid Test – Use Table 3-


a. Check if option host card is installed
i. If not, send “option host card not installed” to GUI and exit test
b. Check if prox module is installed
i. If not, send “prox module not installed” to GUI and exit test
c. Close option gas supply solenoid valve
d. Close prox purge valve
e. Deactivate option gas supply solenoid
f. Set mix in normal mode
g. Command mix accumulator to [1]
h. Open BD PSOL to [2]
i. Open mix air PSOL to [3]
j. Wait [4] for pressure to build up
k. Check mix accumulator pressure stability
i. If not, generate FAILURE FE0017
1. Verify proper connection of the Options Gas Supply Solenoid 2
2. Replace the Options Gas Supply Solenoid 2

l. Set mix in test mode


m. Close BD PSOL
n. Close mix air PSOL
o. Measure mix accumulator pressure P0
p. Activate option gas supply solenoid
q. Wait [6]
r. Measure mix accumulator pressure P1
s. Verify if (P0 –P1) < [7] psig
i. If not, generate FAILURE FE0018

Page 67
1. Verify that the Options Gas Supply Solenoid 2 is properly connected and that
the tubing is not leaking.
2. Verify that the Options Gas Supply Solenoid 2 gasket is in place.
3. Replace the Options Gas Supply Solenoid 2.
t. Deactivate option gas supply solenoid
4. Check Valve Reverse Flow – Use Table 3-4
a. Put mix in test mode
b. Close the Mix PSOLs
c. Prompt user to disconnect air
d. If the compressor is installed
i. Disable compressor controller
ii. Open the safety valve
iii. Open the delivery PSOL to [6]
iv. Wait up to [7] for compressor accumulator pressure to drop below 1 psig
v. Wait [8]
vi. Close the Safety valve
e. Read the O2 inlet pressure sensor
f. Verify the pressure is above [9]
i. If not, FAILURE (O2 not connected) FE0015
ii. Check / replace the Mix Check Valve.
g. Open delivery PSOL to [6]
h. Open the air PSOL to [3]
i. Wait [10]
j. Read air inlet pressure
k. Verify the pressure is less than [2]
i. If not, FAILURE (Air still connected) FE0007
ii. Follow the ventilator prompts as directed.
iii. Look for occlusions to R1.
l. Set the mix accumulator to the target pressure [11]
m. Open the purge valve
n. Close the air PSOL
o. Set the mix percentage to 100% O2
p. Put the mix controller in normal mode
q. Wait [12]
r. Close the purge valve
s. Close the delivery PSOL
t. Wait for accumulator pressure to stabilize or timeout after [1]
i. If not, FAILURE (Unable to establish stable accumulator pressure) FE0006
ii. Swap the Mix PSOL’s.
iii. If problem transfers to the air side, reinstall the good PSOL back to its original
position and replace the Oxygen PSOL.
u. Put the mix into test mode
v. Prompt the user to disconnect O2
w. Wait up to [4] for pressure to drop below [2]. Continue testing as soon as pressure is below
[2]
i. If not, FAILURE (“Gas type” inlet pressure too high) FE0007 for air; FE0010 for
O2
ii. Follow test prompts as directed.
iii. Use high pressure, compressed gas sources between 35 – 87 psig.
iv. Replace the Gas Supply Sensors PCBA.
x. Open the mix PSOLs to [3]
y. Wait [4]
z. Take a 5 sample average value of the mix flow sensors.
aa. Verify that the average value is less than [5]
i. If not, FAILURE (“Gas type” check valve leak rate exceeds limit) FE0008 for air
and FE0009 for O2
ii. Check/replace the Mix Check Valve.
iii. Swap the Mix flow sensors.

Page 68
4. Test #4: Mix PSOL Test

Test Objectives: To calculate the Mix PSOL liftoff current and check it against the
expected range. The liftoff value is calculated for testing only. It is not used as a
parameter into the mix control algorithm like the liftoff for the delivery PSOL.

Page 69
1.1.1.10 Mix PSOL liftoff

Ref# Number Value

1 Mix PSOL starting current (Air, O2) 25 mA

2 PSOL increment 1 mA

3 Minimum Liftoff current (AIR,O2) 50 mA

4 Max Liftoff current (Air, O2) 250mA

5 Mix O2 Percentage Settling Time 1 sec

6 Mix System Settling Time 3 sec

7 Mix Pressure Settling Time 1 sec

8 PSOL liftoff flow 0.1 slpm

9 PSOL current settling time 100 msec

10 Mix PSOL loopback current (Air, O2) < +/- (9.1% + 13.73 mA) of the commanded current

11 Mix Stabilization Time 500 msec

12 Pressure Relief time 2 sec

Table 3-3

 Mix PSOL liftoff shall be when flow exceeds .100 slpm.


 The liftoff average shall be the average of 3 liftoffs.
 The liftoff average for air shall be within 50mA to 250mA.
 The liftoff average for oxygen shall be within 50mA to 250mA.
 The starting current for the liftoff shall be 25 mA.

1.1.1.11 Mix PSOL Loopback

Ref# Number Value

1 Mix PSOL loopback current (Air, O2) < +/- (9.1% + 13.73 mA) of the commanded current

2 Stabilization time 1 sec

3 PSOL Open Count 3500 counts

4 PSOL current settling time 100 msec

Table 3-4

 The loopback current read shall be within (9.1% + 13.73 mA) of the commanded current.

Method
Page 70
1. Mix PSOL Liftoff Test – Use Table 3-3
a. Set Mixer to normal mode
b. Set O2 percentage to 21 % (100% for O2 PSOL)
c. Wait [5]
d. Cycle the PSOLs
e. Wait [6]
f. Put the mix into test mode
g. Close all mix PSOLs
h. Wait [7]
i. Read air flow sensor
j. Verify initial flow is less than limit [8]
i. If not, FAILURE( “Gas Type” PSOL has excessive leak) if Air it’s FE0101 and
FE0102 for O2
ii. Swap the Mix Air and Oxygen PSOL’s. If the problem transfers, replace the
appropriate PSOL.

k. Command the air PSOL with starting current [1]


l. Wait [9]
m. Read air PSOL current
n. Verify current is within limits [10]
i. If not, FAILURE( “Gas Type” PSOL current OOR) if Air it’s FE0103 and FE0104
for O2
ii. Swap the Mix Air and Oxygen PSOL’s. If the problem transfers, replace the
appropriate PSOL.

o. Increment the PSOL by [2] every 20 milliseconds until flow exceeds [8] or current exceeds
[4]
i. If not, FAILURE( “Gas Type” PSOL maximum liftoff current exceeded) if Air it’s
FE0107 and FE0108 for O2
ii. Swap the Mix Air and Oxygen PSOL’s. If the problem transfers, replace the
appropriate PSOL.

p. Wait [11]
q. Get the air PSOL current
r. Check current is within limits [3] [4]
i. If not, FAILURE(“Gas Type” PSOL liftoff current OOR) if Air it’s FE0109 and
FE0110 for O2
ii. Swap the Mix Air and Oxygen PSOL’s. If the problem transfers, replace the
appropriate PSOL.

s. Close the air PSOL


t. Open the purge valve
u. Wait [12]
v. Repeat two more times.
w. Average the 3 values to calculate the liftoff average.
x. Repeat for the O2 PSOL.
2. PSOL Loopback – Use Table 3-4
a. Put the mix controller in test mode
b. Close all mix PSOLs
c. Wait [2]
d. Open delivery PSOL to [3]
e. Command the air PSOL driver to deliver 178 mA.
f. Wait [4]
g. Take a 5 sample average of the PSOL current
h. Verify current loopback reading is within test limits [1]
i. If not, FAILURE (Mix air PSOL loopback current OOR) FE0111
ii. Swap the Mix Air and Oxygen PSOL’s. If the problem transfers, replace the
appropriate PSOL.

Page 71
i. Repeat for currents of 374 mA, and 730 mA.
j. Repeat for oxygen PSOL

Page 72
5. Test #5: Mix Accumulator Test

Test Objectives: Using air, verify the integrity of the accumulator pressure sensor
and over pressure switch.

Page 73
1.1.1.12 Pressure Sensor

NOTE: The accumulator is charged to a target pressure, the pressure is measured, and then the volume is
discharged to ambient. Assuming the accumulator volume is 500 ml, it has a compliance of 0.5 ml/cm H2O.
There are 70 cm H2O per psi, so this is the same as 35 ml/psi. In other words, an error in volume discharge
measurement of 35 ml would create an error in pressure estimation of 1 psi. Using the same compliance
value of 35 ml/psi, the accumulator should contain 875 ml when charged to 25 psi (i.e., 25 psi*35 ml/psi).
Charge volumes for 11 psi and 2 psi can be calculated in the same way.

Ref# Number Value

1 Mix pressure stability timeout 5 sec

2 Delivery PSOL open amount 700 counts

3 Empty Accumulator Volume timeout 5 sec

4 Volume limits < +/- 75 mL of target volume

5 Pressure Drop time 1 sec

6 Target pressure 11 psig

7 Additional Stability time 4 sec

Table 3-5

 The accumulator pressure transducer shall be verified to read within an accuracy of +/- (5% + 0.3)
psig.
 The accumulator pressure transducer shall be verified at pressures of 11 psig, and 25 psig.

1.1.1.13 Over Pressure Switch

Ref# Number Value

1 Mix Air PSOL open amount 3500 counts

2 Max Over pressure switch activation pressure 20 < Activation Pressure < 28 psig

3 Min Over pressure switch deactivation pressure Deactivation Pressure > 6 psig

4 Delay Prior to checking deactivation 1 sec

5 Pressure drain time 3 sec

Table 3-6

 The over pressure switch shall be verified to activate at 24 psig +/- 4 psig.
 The over pressure switch shall be verified to deactivate at 11 psig +/- 5 psig.

Method
1. Pressure Sensor – Use Table 3-5

Page 74
a. Activate overpressure bypass switch
b. Put mix in Test mode
c. Open the delivery PSOL to [2]
d. Wait [5]
e. Close the delivery PSOL
f. Set Mix to [6]
g. Put mix in Normal mode
h. Wait for pressure stability or timeout of [1]
i. If not, FAILURE (Unable to establish accumulator pressure) FE0201
1. Verify no leaks in the Mix Accumulator Subsystem.
2. Check the Mix and Delivery PSOL and flow sensors functions via the other
EST tests.

i. Wait [7]
j. Put mix in Test mode
k. Close the mix PSOLs
l. Read accumulator pressure sensor
m. Open the delivery PSOL to [2] and monitor the flow
n. Wait [3]
o. Read accumulator pressure sensor
p. Get measured volume
q. Calculate the volume using the starting and end pressure
r. Calculate pressure based on measured volume and end pressure
s. Verify volume within limits [4]
i. If not, FAILURE (Pressure sensor OOR) FE0203
ii. Check / replace the IFM PCBA.
t. Close delivery PSOL
u. Repeat for 25 psig
2. Over Pressure switch – Use Table 3-6
a. Put mix in test mode
b. Close the mix psols
c. Open the delivery PSOL to [1]
d. Wait [4]
e. Wait for pressure to drain and switch to deactivate up to [5]
i. If not, FAILURE (Pressure too high/Over pressure switch failed to deactivate)
FE0205
i. Check / replace PS1.
ii. Replace the IFM PCBA.
f. Put mix in Normal mode
g. Open the purge Valve
h. Close the delivery PSOL
i. Fill accumulator gradually from 10 to 30 psig while checking over pressure switch to activate
[2]
i. If not, FAILURE (Over pressure switch failed to activate.) FE0204
ii. Check / replace PS1.
iii. Replace the IFM PCBA.
j. Drain accum. gradually down to 1 psig while checking over pressure switch to deactivate [3]
i. If not, FAILURE (Over pressure switch failed to deactivate) FE0205
ii. Check / replace PS1.
iii. Replace the IFM PCBA.

Page 75
6. Test #6: Circuit Pressure Test

Test Objectives: To test the inspiratory/exhalation offsets using the auto zero
solenoids, cross-check the inspiratory, exhalation pressure and safety valve
transducers at various test pressures and verify that the autozero solenoids are
working.

Page 76
1.1.1.14 Inspiratory / Exhalation Pressure Transducer Auto Zero

Ref# Number Value

1 Autozero response time 40 ms

2 inspiratory pressure 1465 to 1763 counts

3 exhalation pressure 1465 to 1763 counts

4 Pressure Decay Time 1 sec

Table 3-7

 The zero pressure offset error of the inspiratory pressure transducer shall be verified to be
≤+10cmH2O at ambient pressure - equivalent to an ADC reading between 1465 and 1763 counts.
 The zero pressure offset error of the exhalation pressure transducer shall be verified to be
≤+10cmH2O at ambient pressure - equivalent to an ADC reading between 1465 and 1763 counts.

1.1.1.15 Pressure Transducer Cross Check

Ref# Number Value

1 Flow 5L/min

2 Time for pressure build 2000 ms

3 Pressure stability time 200 ms

4 Abs(PINSP - PSV) <4.55%*(PINSP+PSV)+10.169cmH2O

5 Abs(PINSP - PEXLH) <4.55%*(PINSP+PEXLH)+2*0.169cmH2O

Table 3-8
 The safety valve pressure transducer shall be verified to be within +(4.55% + 10.169 cmH2O) of the
inspiratory pressure transducer.
 The exhalation pressure transducer shall be verified to be within + (4.55% + 2 * 0.169 cmH2O) of the
inspiratory pressure transducer.
 The pressure transducer cross check shall be verified at the following pressures: 10 cmH 2O, 50
cmH2O, 100 cmH2O.

1.1.1.16 Auto Zero Solenoids

Ref# Number Value

1 Autozero response time 40ms

2 | PINSP | <0.419 cm H2O

3 |PEXLH| <0.419cm H2O

Table 3-9

Page 77
 Activation of the inspiratory autozero solenoid shall be verified to take the pressure at the inlet of the
inspiratory transducer from 100 cmH 2O to within + 0.419cm H2O at ambient pressure.
 Activation of the exhalation autozero solenoid shall be verified to take the pressure at the inlet of the
exhalation transducer from 100 cmH 2O to within + 0.419cm H2O at ambient pressure.
Method
2. Inspiratory / Exhalation Pressure Transducer Auto Zero – Use Table 3-7
a. Open Safety Valve
b. Wait [4]
c. Activate the inspiratory autozero solenoid and wait [1] to respond. If fails, FE0306
i. Replace the IFM PCBA.
1
d. Take an 8-sample average of the inspiratory pressure transducer ADC count at 0 cmH 2O.
e. Deactivate the inspiratory autozero solenoid and wait [1] to respond.
f. Verify the average value read is within limits ([2] for inspiratory, [3] for exhalation)
i. FE0301 - FAILURE Insp Zero offset OOR.
1. Replace the IFM PCBA.

ii. FE0302 - FAILURE Exp Zero offset OOR.


1. Replace the Exhalation Sensors PCBA.

g. Store the inspiratory pressure 0 cmH 2O count.


h. Repeat steps a through e for exhalation pressure sensor.
3. Pressure Transducer Cross Check – Use Table 3-8
a. Close the safety valve.
b. Close the exhalation valve (with EV controller)
c. Establish 10 cmH2O pressure with [1] flow and reading inspiratory pressure.
d. Wait for pressure to build or timeout of [2]
i. If not, FAILURE (Target pressure not achieved FE0303.) and display the
inspiratory and exhalation pressures.
1. Replace the IFM PCBA for low inspiratory pressure.
2. Replace the Exhalation Sensors PCBA for low expiratory pressure.

e. Stop Pressure controller


f. Wait [3] for pressure to stabilize.
2
g. Take a reading of the inspiratory and exhalation pressures, and a 3-sample average of the
safety valve pressure.
h. Verify that the difference of inspiratory and exhalation are within the limits [5]
i. If not, FAILURE (Inspiratory and Safety Valve pressure sensor difference
outside test limits.) FE0305
1. Replace the IFM PCBA

i. Verify that the difference of inspiratory and safety valve are within the limits [4]
i. If not, FAILURE (Inspiratory and exhalation pressure sensor difference outside
test limits.) FE0304
1. Replace the IFM PCBA for low inspiratory pressure.
2. Replace the Exhalation Sensors PCBA for low expiratory pressure.

j. Repeat steps c through i at pressures of 50 and 100 cmH 2O.


4. Auto Zero Solenoids – Use Table 3-9
a. Activate the inspiratory autozero solenoid
b. wait [1]
c. Take a reading of the inspiratory pressure sensor.
d. Deactivate the inspiratory autozero solenoid
e. wait [1] to respond.
i. Verify that the average reading is within the limits ([2] for inspiratory, [3] for
exhalation) FE0306 for Insp and FE0307 for Exp
1. Replace the IFM PCBA for low inspiratory pressure.
2. Replace the Exhalation Sensors PCBA for low expiratory pressure.

Page 78
ii. If not, FAILURE (“Sensor Side” auto zero solenoid not operational.)
1. Replace the IFM PCBA for inspiratory autozero failure.
2. Replace the Exhalation Sensors PCBA for expiratory autozero failure.

f. Repeat Steps a through f for exhalation auto zero solenoid.


g. Open Exhalation Valve

Page 79
7. Test #7: Flow Sensor Cross Check

Delivery (Insp) Flow compared to Expiratory Flow and Oxygen Mix Flow

NOTE: graphic above to be updated with latest max target flow of 90 slpm instead of 120 slpm.

Page 80
Delivery (Insp) Flow compared to Expiratory Flow and Air Mix Flow

NOTE: graphic above to be updated with latest max target flow of 90 slpm instead of 120 slpm.

Page 81
Test Objectives: To verify all flow sensors: mix, inspiratory and exhalation. To verify
that PSOL command current is in range for each test flow, and perform Delivery
PSOL liftoff (i.e. determine and store the PSOL command corresponding to 0.3 lpm).

Note 1: The exhalation flow sensor has separate air and O 2 calibration tables. The appropriate table for the
test gas is used during the flow cross-checks.

Note 2: The SST flow sensor cross-check is identical to this except that it is able to cross-check a single gas
supply. SST also differs in the times used for flushing the circuit.

Ref# Number Value

1 Stable flow timeout 1 sec

2 Inspiratory flow +/- (2% + 0.05 slpm) of the target flow

3 Mix flow (Air, O2)


+/- (4.5% + 0.15 slpm) of the insp flow
and Exhalation flow
4 3 2
4 Delivery PSOL ≥ (-1.79E-07x + 7.56E-05x - 1.17E-02x + 2.62x + 110.01) * (1-
current min PsolAccuracy) mA where x = (insp flow * 0.98) – 0.05
4 3 2
5 Delivery PSOL ≤ -3.94E-08x + 2.49E-05x – 7.85E-03x + 2.95x + 180.7 * (1+ PsolAccuracy)
current max mA where x = (insp flow * 1.02) + 0.05

6 Compressor
18 sec
recharge time

7 Flow Settling time 2 sec

Test Volume (ml) Flow (slpm) Timeout (sec)


EST 1500 120 5
Circuit Flush
SST Pediatric 11000 120 13.6
SST Adult 15000 120 18.0
SST Neo 3500 30 10.0

Table 3-10

 The inspiratory flow sensor shall be within (0.1slpm + 1%) of the flow target.
 The mix, and exhalation flow sensors shall be verified to be within + (4.5% + 0.15 slpm) of the
inspiratory flow sensor.
 The delivery PSOL current required to deliver a flow rate shall be verified to be within:
Current_min ≤ Current ≤ Current_max. This doesn’t not apply to the mix.

The at each flow rate is listed as table below:

Page 82
flow rate PSOL drive PSOL PSOL
target signal heating Accuracy
accuracy effects
from from 3-55
70550-45
1 slpm

5 slpm

20 slpm

60 slpm

90 slpm

 The flow sensor cross check, flow stability test, and current vs. flow test shall be verified at the
following flow rates: 90, 60, 20, 5, and 1 slpm. Mix flow sensors shall be verified at 90, 60 and 20
slpm only.

1.1.1.17 Delivery PSOL Liftoff Calibration and Lookup Table Creation

Ref# Number Value

1 Delivery PSOL counts 448 counts

2 PSOL increment 1 count

3 PSOL increment time 10 ms

4 Liftoff flow 0.300 lpm

5 Max PSOL liftoff 909 counts

Table 3-11

Logged Parameters
 If a PSOL current check fails, log the expected current and actual current of the PSOL.
Method
1. Flow Sensor Cross Check – Use Table 3-10
a. Flush the system
b. If using compressor, wait [6] for compressor to recharge
c. Establish a target flow rate of 90 slpm using air
d. Wait [7]
e. Wait for stable flow using inspiratory flow sensor or timeout at [1]
i. If not, FAILURE (Unable to establish stable flow) FE0501
1. Verify source gas is unobstructed and able to source the ventilator pressure
and flow requirements.
2. Run a Flow Sensor Calibration.
3. Swap the Mix PSOL’s. Replace the appropriate PSOL if failure transfers to
the other side.
4. Swap the Mix flow sensors. Replace the appropriate sensor if failure
transfers to the other side.
5. Check / replace the Delivery PSOL.
Page 83
6. Check / replace the Delivery Flow Sensor.
f. Take a reading of the alpha-filtered air flow sensor, inspiratory flow sensor, unoffsetted
exhalation flow sensor
g. Calculate delivery PSOL commanded current
h. Stop flow controller
i. Verify inspiratory flow is within limits of target [2]
i. If not, FAILURE(Inspiratory flow sensor OOR) FE0502
1. Run a Flow Sensor Calibration
2. Replace the Delivery Flow Sensor.
j. Verify exhalation flow is within limits of inspiratory [3]
i. If not, FAILURE (Exhalation flow sensor OOR) FE0503
1. Run a Flow Sensor Calibration
2. Replace the EVQ.
k. Verify air flow is within limits of inspiratory [3]
i. If not, FAILURE (Mix “gas” flow sensor OOR) FE0504
1. Run a Flow Sensor Calibration
2. Swap the Mix flow sensors. Replace the appropriate sensor if failure
transfers to the other side.
l. Verify that the delivery PSOL current is within the test limits [4] [5]
i. If not, FAILURE (Delivery PSOL current OOR) FE0505
1. Replace the Delivery PSOL.
m. Repeat for flow rates: 60, 20, 5, 1 slpm
n. Repeat for oxygen and heliox (if installed)
2. Delivery PSOL Liftoff Calibration and Lookup Table Creation – Use Table 3-11
a. Command the delivery PSOL to [1]
b. Increment the delivery PSOL by [2] every [3].
c. Store the count value when flow equals 0.100, 0.150, 0.200, 0.250, and 0.300 lpm.
d. Stop incrementing when flow equals or exceeds [4] for two consecutive readings or until
count reaches [5].
e. Repeat two more times.
f. Average values in the 3 PSOL lookup tables generated and save final PSOL lookup table.
g. Display and store the liftoff current. The liftoff current is determine at [4]

Page 84
8. Test #8: Delivery PSOL Test

Page 85
Test Objective: This test checks the PSOL current sensor and commanded PSOL
current.
1.1.1.18 Delivery PSOL Loopback

Ref# Number Value

1 Delivery PSOL loopback <+/-(9.1%+13.73ma)

2 PSOL Stabilization time 100 msec

Table 3-12

 The loopback current read is with (9.1% + 13.73 mA) of the commanded current at 178, 374, and 730
mA.
Method
1. PSOL Loopback – Use Table 3-12
a. Command the delivery PSOL driver to deliver 178 mA.
b. Wait [2]
c. Read Delivery PSOL loopback current
d. Verify current loopback reading is within test limits [1]
i. If not, FAILURE (Delivery PSOL Current loopback reading OOR) FE1401
1. Replace the Delivery PSOL.
e. Repeat for currents of 374 mA, and 730 mA.
f. Close the delivery PSOL.

Page 86
9. Test #9: Exhalation Valve Loopback Test

Test Objective: To verify that the exhalation valve loopback current and the
exhalation valve command current are within a maximum range from each other.
Ref# Number Value

1 Damping Gain 0.0 Volts

2 Settling Time for EV 500 ms

3 EV Nominal Offset 6.5 mA

4 EV Loopback Gain Error 11 %

5 EV Loopback Offset Error 26.8 mA

6 Max Error -[3]+/-(TargetCommand * [4] + [5])

Table 3-13
Method
1. Exhalation Valve Current Loopback – Use Table 3-13
a. Command the exhalation valve damping gain to [1]
b. Set the exhalation valve target current to 700 mA
c. Wait [2]
d. Read the exhalation valve loopback current
e. Verify current is within [6]
i. If not, ALERT (Exhalation Valve Loopback current OOR). AE1601
1. Run an Exhalation Valve Calibration.
2. Replace the Exhalation Valve.
f. Repeat for all commanded current targets (300, 0, -30, -60, -98)
g. Open Exhalation valve (non-controller)

Page 87
10. Test #10: Exhalation Valve Pressure Accuracy Test

Test Objective: This test verifies that the exhalation valve (EV) seal and poppet
friction are acceptable.

Page 88
Ref# Number Value

1 Damping Gain 0.8 Volts

2 EV Pressure 45 cmH2O

3 Delivery Flow 5 lpm

4 Stable Flow Time 2s

5 Flow Settling Time 500 ms

6 PSOL Ramp Down 1 count per 50 ms

7 Flow Min 1 lpm

8 Pressure Settling Time 5s

9 Pressure Decay (Failure) < 4.5 cmH2O

10 Pressure Decay (Alert) < 3.5 cmH2O

11 Min LVDT Temperature 10 C

12 KT KT = 1 + 0.0015 * (LVDT temp – 21.1C)

13 EV Pressure Equivalent 43.3 cmH2O

Table 3-14

 Flow stability (reference stable flow) at a flow rate of 5 slpm shall be verified after commanding the
exhalation valve to a system pressure of 45 cmH 2O.
 Pressure accuracy shall be verified to be within 4.5 cmH2O (FAILURE) and 3.5 cmH2O (ALERT) at a
commanded pressure of 45 cmH 2O when flow is ramped from 5 slpm down to 1 slpm.

Method
1. EV Pressure Accuracy – Use Table 3-14
a. Verify EV Calibration table checksum is valid and last calibration performed was completed
i. If not, FAILURE (Bad Exhalation Valve Calibration) FE1701
1. Rerun the Exhalation Valve Calibration.
b. Autozero pressure transducers
1
c. Read the LVDT temperature and verify temperature is larger than [11] .
i. If not, FAILURE(LVDT Temperature OOR) FE1705
1. Replace the Exhalation Valve Assembly (contains the LVDT).
d. Calculate the temperature compensation factor [12]
e. Command the exhalation valve to [13] * [12], which corresponds to a desired system
pressure of [2]
f. Command the exhalation valve damping gain to [1]
g. Establish a delivery flow of [3] or timeout of [4]
i. If not, FAILURE (Unable to establish flow) FE1702
1. Perform a flow sensor calibration.
2. Check / replace the Delivery Flow Sensor.
3. Check / replace the EVQ.

Page 89
h. Wait [5]
i. Read alpha-filtered exhalation pressure (P1)
j. Stop delivery flow controller, but DO NOT reset the PSOL command in order to maintain flow.
k. Ramp PSOL down at [6] until exhalation flow reaches [7]
i. If not, FAILURE (Exhalation Flow too High) FE1703
1. Perform a flow sensor calibration.
2. Check / replace the EVQ.
3. Replace the Exhalation Sensors PCBA.
l. Wait [8]
m. Read alpha-filtered exhalation pressure (P2)
n. Verify pressure difference (P1 – P2) is within limits [9] [10]
i. If P1-P2 greater than FAILURE limit, FAILURE (Exhalation Valve Pressure OOR)
FE1704
1. Perform an Exhalation Valve Calibration.
2. Replace the Exhalation Valve.
3. Replace the Exhalation Sensors PCBA.
ii. If P1-P2 greater than ALERT limit, ALERT (Exhalation Valve Pressure OOR)
AE1704
1. Perform an Exhalation Valve Calibration.
2. Replace the Exhalation Valve.
3. Replace the Exhalation Sensors PCBA.

Page 90
11. Test #11 Exhalation Valve Performance Test:

Test Objective: This test verifies that the current vs. pressure values stored in the
exhalation valve (EV) calibration table agrees with actual valve performance.

Page 91
Ref# Number Value

1 Damping Gain 0.8 volts

2 EV Pressure 5 cmH2O

3 Delivery Flow 5 lpm

4 Stable flow timeout 2 sec

5 Flow Settling Time 300 msec

6 KT KT = 1 + 0.0015 * (LVDT temp – 21.1C)

7 Min LVDT Temperature 10 C

Table 3-15

 The pressure established using the exhalation valve calibration table pressure values shall be verified
to be within the range specified by the Min Pressure and Max Pressure values.
 The pressure generated using the calibration tables shall be verified at the following pressures: 2
cmH2O , 5 cmH2O , 10 cmH2O, 20 cmH2O, 45 cmH2O, 95 cmH2O.
Method
1. EV Performance – Use Table 3-15
a. Verify EV Calibration table checksum is valid and last calibration performed was completed
i. If not, FAILURE (Bad Exhalation Valve Calibration) FE1801
1. Perform an Exhalation Valve Calibration.
2. Replace the Exhalation Valve.
3. Replace the Exhalation Sensors PCBA.
1
b. Read the LVDT temperature and verify temperature is larger than [7] .
i. If not, FAILURE(LVDT Temperature OOR) FE1805
1. Perform an Exhalation Valve Calibration.
2. Replace the Exhalation Valve.
3. Replace the Exhalation Sensors PCBA.

c. Calculate the temperature compensation factor [6]


d. Autozero the pressure transducers
e. Establish a delivery flow of [3]
f. Command the exhalation valve to (EV command for test pressure of 2 cmH 2O) * [6] cmH2O,
which corresponds to a desired system pressure of 2 cmH 2O
g. Command the exhalation valve damping gain to [1]
h. Wait for stable flow for up to [4]
i. If not, FAILURE (Unable to establish stable flow) FE1802
1. Perform a Flow Sensor Calibration.
2. Perform an Exhalation Valve Calibration.
3. Replace the EVQ.

i. Wait for [5]


j. Read alpha-filtered exhalation pressure
k. Verify pressure is within test limits (In Display Parameters above)
i. In not, FAILURE (Pressure OOR) FE1803
1. Perform an Exhalation Valve Calibration.
2. Replace the Exhalation Sensors PCBA.

Page 92
l. Repeat steps f to k for 5, 10, 20 , 45 and 95 cmH 2O

Page 93
12. Test # 12 Exhalation Valve Velocity Transducer Test:

Test Objective: This test verifies that the exhalation valve (EV) velocity transducer is
sending a signal and that the control circuit is recognizing the signal. It does not
check for the quality of the signal. This test verifies the driver system's ability to
control negative current to lift the poppet open, to check the operation of the LVDT
position sensor and to make a gross check of the velocity feedback signal.

Page 94
1.1.1.19 Exhalation Valve Damping

Ref# Number Value

1 EV resting pressure 2.5 cmH2O

2 Valve response Time 200 msec

3 Delivery Flow 180 lpm

4 Stable flow timeout 500 msec

5 Initial damping gain 0.8 volt

6 Sum Period 100 ms @ 10 ms intervals

7 Final damping gain 0.05 volt

8 Sum Delta >20

9 Pressure Stability Time 200 msec

10 EV testing pressure 15 cmH2O

11 Compressor Recharge Time 3 sec

12 System Settling Time 3 sec

Table 3-16

 Damping circuitry shall be verified to generate a difference in pressure-time area greater than 20
cmH2O/time value by measuring the exhalation pressure step response from 0.3 cmH 2O to 15
cmH2O, with damping initially set to a gain of 1.0 V (high gain) then to a gain of 0.05V (low gain).
 Measurement will be taken from the time the pressure command is issued for 100 ms at 10 ms
requirements.

1.1.1.20 Exhalation Valve Mechanical Position

Ref# Number Value

1 Mechanical settling Time 50 msec

2 Stable position time (-0.080 in) 1 sec

3 Stable position time (-0.150 in) 3 sec

Table 3-17

 The exhalation valve shall poppet shall be verified to move between -0.080 inches to -0.150 inches
Method
1. Exhalation Valve Damping – Use Table 3-16
a. Autozero the inspiratory and exhalation pressure transducers.
b. Open Purge Valve to stabilize accumulator pressure.
c. Wait [12] for the System to settle
Page 95
d. Command the EV to [1]
e. Wait [2] for the valve to respond.
f. Establish a flow of [3]
g. Wait up to [4] for flow to stabilize.
i. If not, FAILURE (Unable to establish flow) FE1901
1. Perform a flow sensor calibration.
2. Replace the Delivery Flow Sensor.
h. Stop Flow controller, but keep PSOL at last command
i. Wait [9] for pressure stability
j. Command the EV to [10]
k. Command the EV damping gain to [5]
l. Read and Sum the exhalation pressure for [6] intervals (Psum1)
m. Close delivery PSOL
n. Open the EV
o. Wait [11]
p. Repeat steps b through l for EV damping gain of [7] (Psum2)
q. Verify Psum2 - Psum1 is greater than minimum test limit [8]
i. If not, FAILURE (Exhalation Velocity Transducer OOR) FE1902
1. Perform an Exhalation Valve Calibration.
2. Replace the Exhalation Valve Assembly.
r. Close Purge Valve.
2. Exhalation Valve Mechanical Position – Use Table 3-17
a. De-energize the EV
b. Close Delivery PSOL
c. Delay [1] to allow any mechanical transients to settle out.
d. Establish a position of -80 milli-inches
e. Wait for stable position up to [2]
ii. If not, FAILURE (Unable to stabilize position: Reseat the exhalation valve body)
FE1903
1. Perform an Exhalation Valve Calibration.
2. Replace the Exhalation Valve Assembly.

f. Repeat steps d through e for -150 milli-inches and wait for stability up to [3]
g. Stop Position controller

Page 96
13. Test #13: Safety System Test

Test Objective: To test safety valve operation.

Page 97
1.1.1.21 Safety Valve Occlusion

Ref# Number Value

1 Delivery Flow 60 lpm

2 Flow stabilization timeout 2 sec

3 Insp Pressure ≤1.5 cmH2O

4 Flow stability Delay 3 sec

Table 3-18

 The safety valve shall be verified to not be occluded by commanding the SV to open and establishing
a flow of 60 lpm followed by waiting 2 sec for the inspiratory manifold pressure pressure to stabilize.
Once it is stable read the residual pressure and compare it with an empirically derived threshold
pressure of 1.5 cmH2O.

1.1.1.22 Safety Valve Loopback Current

Ref# Number Value

1 Open SV current 10 mA

2 Closed SV current 100ms 360 ~ 840 mA

3 Closed SV current 500ms 240 ~ 560 mA

4 SV current settling time 100 msec

5 Extended SV settling time 400 msec

Table 3-19

 Safety valve loopback current test consists of measuring the loopback current of the SV under the
following varying conditions (i) open valve, (ii)closed valve during initial closing current pulse, and (iii)
steady state closed condition.
 During SV open condition the loopback current must be less than 10ma
 During Initial closed state – 100 ms after close command issued, the current must be between 360
ma and 840 ma
 During the steady state closed condition, about 500 ms after the close command is issued, the
loopback current should be between 240 ma and 560 ma.
Method
1. Safety Valve Occlusion – Use Table 3-18
a. Autozero the pressure sensors
b. Close the Exhalation Valve (with EV controller)
c. Open the Safety Valve
d. Establish a delivery flow of [1]
e. Wait [2] for flow stabilization.
i. If not, FAILURE (Unable to establish flow) FE1501
1. Perform a Flow Sensor Calibration.
2. Replace the Delivery Flow Sensor.
3. Replace the Delivery PSOL.

Page 98
f. Wait [4]
g. Read inspiratory pressure filtered value
h. Verify pressure less than or equal to [3]
i. If not, FAILURE (Safety Valve occluded) FE1502
1. Check the Safety Valve cable connection.
2. Replace the Safety Valve.
i. Stop the flow controller
2. Safety Valve Loopback Current (safety valve is still open from the previous test) – Use Table 3-19
a. Read the Safety Valve current
b. Verify current is within test limits [1]
i. If not, FAILURE(Safety Valve Loopback current OOR) FE1503
1. Check the Safety Valve cable connection.
2. Replace the Safety Valve.

c. Close the Safety Valve


d. Wait [4]
e. Read the Safety Valve current during the initial close current pulse
f. Verify current is within test limits [2]
i. If not, FAILURE (Safety Valve Loopback current OOR) FE1503
1. Check the Safety Valve cable connection.
2. Replace the Safety Valve.

g. Wait [5]
h. Read the Safety Valve steady state current
i. Verify current is within test limits [3]
i. If not, FAILURE (Safety Valve Loopback current OOR) FE1503
1. Check the Safety Valve cable connection.
2. Replace the Safety Valve.

Page 99
14. Test #14: Backup Ventilation Test

Test Objectives: To verify the backup ventilation systems: Mix, Inspiratory and
Exhalation.

1.1.1.23 Mix Backup Ventilation

Ref # Number Value

1 Open Delivery PSOL 2000 counts

2 Time for accumulator to empty 1 sec

3 Mix PSOL Desired gas Min flow 30 lpm

4 Mix PSOL non-desired gas Max flow 1 lpm

5 Wait for accumulator to fill 1 sec

6 Mix Accumulator Pressure limit < +/- 3 psig of target

7 BUV transition Time 50 msec

8 Stable Pressure Time 5 sec

Table 3-20

 The mix accumulator shall be pressurized to 11 psig +- 3 psig for 21% Oxygen.

Page 100
 The mix accumulator shall be pressurized to 11 psig +- 3 psig for 100% Oxygen.
 Time for accumulator to empty: the capacity of mix accumulator is 0.5 Liter, 2000 counts generates
about 115 lpm flow in delivery PSOL. So time is calculated as: 0.5*60/115 = 0.26 s.
 Wait for accumulator to fill: here 3 times of 0.5 Liter is used to calculate the fill time, i.e. 0.26*3 =
0.78s.

1.1.1.24 Inspiratory Backup Ventilation – Delivery PSOL shutdown

Ref# Number Value

1 Open Delivery PSOL 1000 counts

2 Wait time after transfer to BUV 50 msec

3 Delivery PSOL Closed Flow limit < 0.5 lpm

4 Delivery PSOL Open Flow limit > 10 lpm

5 Settling time 500 msec

Table 3-21

 The delivery PSOL shall set to zero when Inspiratory BUV is active. (Inspiratory BUV shuts down the
delivery PSOL.)
 BD PSOL flow less than 0.5 lpm indicates the BD PSOL is closed. Refer to PSOL spec 10042304.

1.1.1.25 Inspiratory Backup Ventilation – Flow Response

Ref# Number Value

1 Stable flow timeout 5000 msec

2 Flow Error percentage 15.5%

3 High Accumulator Pressure 25 psig

4 Additional stability time 2 sec

5 Normal Accumulator Pressure 11 psig

Table 3-22

 When mix accumulator is set to 25 psig, the inspiratory flow shall within +- 15.5% of the expected
flow, which includes 10% of insp BUV solenoid uncertainty, 3.2% of mix accumulator pressure
transducer uncertainty, and 2.3% of flow sensor uncertainty.

1.1.1.26 Exhalation Backup Ventilation – Exhalation Valve driver shutdown

Ref# Number Value

1 Initial EV BUV current 0 counts

Page 101
2 Initial EV current 300 mA

3 Initial EV BUV loopback current < 50 mA

4 EV loopback current < 50 mA

5 Final EV BUV current 648 counts

6 Final EV BUV loopback current < 300 mA

7 Min EV Pressure 20 cmH2O

8 Settling Time 200 msec

9 Open PSOL counts 1000 counts

10 Flow settling time 500 msec

11 BUV transition Time 50 msec

Table 3-23

 The exhalation valve driver shall set to zero +- 50 mA when Exhalation BUV is active.
 The exhalation valve shall be driven by the redundant driver and read by the current loopback.
 Loopback current threshold 50 mA is from EV BUV driver design spec.

1.1.1.27 Exhalation Backup Ventilation – Pressure Response

Ref# Number Value

1 Delivery Flow 5 lpm

2 Stable flow time < 1 sec

3 Delay after stability 500 msec

4 Sample pressure time 300 ms at 10 ms intervals

5 EV pressure < +/-12% of target

6 3 cmH2O PWM 134 counts

7 18 cmH2O PWM 648 counts

Table 3-24

 Using a 5 lpm flow, the exhalation pressure sensor shall read within 12% of the target pressure: 3
and 18 cmH2O.
 Error limit 12% is from EV BUV driver design spec.
Method
1. Mix Backup Ventilation – Use Table 3-20
a. Put the mix into test mode.
b. Close both mix PSOLs.

Page 102
c. Open the delivery PSOL to [1].
d. Open the purge valve
e. Wait [2] for the accumulator to empty.
f. Close the purge valve.
g. Go into Mix Backup Ventilation – Air.
h. Wait [7]
i. Check that the Mix is reporting that it is in BUV.
i. If not, FAILURE (Unable to enable mix backup ventilation) FE2301
1. Check / replace the Mix Air PSOL.
2. Check / replace the Mix Air Flow Sensor.
j. Wait for accumulator to reach target pressure for up to [8]
iii. If not, FAILURE (Mix backup ventilation failed to reach target pressure) FE2302
1. Check the PMX sensor pneumatic connections.
2. Replace the IFM PCBA.
k. Read the Air flow sensor
l. Verify that the mix is filling the accumulator within flow limits [3]
i. If not, FAILURE( No air delivery in Mix backup ventilation) FE2303
1. Check / replace the Mix Air PSOL.
2. Check / replace the Mix Air Flow Sensor.

ii. For oxygen, if not, FAILURE (No O2 delivery in Mix backup ventilation) FE2304
1. Check / replace the Mix O2 PSOL.
2. Check / replace the Mix O2 Flow Sensor.

m. Read the O2 flow sensor


n. Verify that the mix is not filling the accumulator within flow limits [4]
i. If not, FAILURE (O2 delivery when AIR expected) FE2305
1. Verify cable connections.
2. Swap the Mix PSOL’s.
3. Replace the IFM PCBA.
4. Replace the Mix Module PCBA.

ii. For oxygen, if not, FAILURE (Air delivery when O2 expected) FE2306
1. Verify cable connections.
2. Swap the Mix PSOL’s.
3. Replace the IFM PCBA.
4. Replace the Mix Module PCBA.

o. Disable Mix BUV


p. Repeat for oxygen.
q. Set Test State to initial
2. Inspiratory Backup Ventilation - Delivery PSOL shutdown – Use Table 3-21
a. Open the delivery PSOL to [1]
b. Wait [5]
c. Verify that delivery flow is within the limits [4]
i. If Not, FAILURE(Unable to establish stable flow) FE2307
1. Perform a Flow Sensor Calibration.
2. Replace the Delivery Flow Sensor.
3. Replace the Delivery PSOL.
d. Close the delivery PSOL
e. Wait [5]
f. Verify that the delivery flow is not more than [3]
i. If not, FAILURE (Delivery flow is higher than expected) FE2308
1. Perform a Flow Sensor Calibration.
2. Replace the Delivery Flow Sensor.
3. Replace the Delivery PSOL.

g. Go into Inspiratory Backup Ventilation

Page 103
h. Wait [2]
i. Check that the Inspiratory Module is reporting that it is in BUV.
i. If not, FAILURE (Unable to enable inspiratory backup ventilation) FE2309
1. Check / replace the BUV Solenoid 3.

3. Inspiratory Backup Ventilation – Flow Response – Use Table 3-22


a. Activate Over Pressure Switch bypass
b. Set mix pressure to [3]
c. Put the mix into normal mode.
d. Wait for stable pressure or timeout of [1]
i. If not, FAILURE (Unable to establish stable pressure) FE2310
1. Check / replace the IFM PCBA.

e. Wait [4]
f. Verify that delivery flow is within the limits [2]
i. If Not, FAILURE (Unable to establish stable flow) FE2311
1. Perform a Flow Sensor Calibration.
2. Replace the Delivery Flow Sensor.
3. Replace the Delivery PSOL.

g. Disable Inspiratory BUV


h. Set mix pressure to [5]
i. Set Test State to initial
4. Exhalation Backup Ventilation – Exhalation Valve driver shutdown – Use Table 3-23
a. Set Exhalation Valve BUV driver current to [1]
b. Set Exhalation Valve NON-BUV driver to [2]
c. Wait [8]
d. Open the delivery PSOL to 1000 [9]
e. Wait [10]
f. Read Exhalation filtered pressure
g. Verify that the exhalation pressure is more than [7]
i. If Not, FAILURE(Unable to reach minimum pressure) FE2316
1. Replace the Exhalation Valve.
2. Check / replace the Exhalation Sensors PCBA.

h. Go into Exhalation BUV


i. Wait [11] for BUV transition
j. Check that the Exhalation Module is reporting that it is in BUV.
i. If not, FAILURE (Unable to enable exhalation backup ventilation) FE2312
1. Replace the Exhalation Valve.
2. Check / replace the Exhalation Sensors PCBA.

k. Verify that Exhalation BUV loopback current is zero [3]


i. If not, FAILURE (Exhalation backup ventilation current OOR) FE2313
1. Replace the Exhalation Valve.
2. Check / replace the Exhalation Sensors PCBA.

l. Verify that the Exhalation Valve driver loopback current is zero [4]
i. If not, FAILURE (Exhalation valve current OOR)
1. Replace the Exhalation Valve.
2. Check / replace the Exhalation Sensors PCBA.

m. Set a current to the Exhalation BUV [5]


n. Wait [11] for BUV transition
o. Verify that the loopback current is within range [6]
i. If not, FAILURE (Exhalation backup ventilation current OOR) FE2314
1. Replace the Exhalation Valve.
2. Check / replace the Exhalation Sensors PCBA.

Page 104
5. Exhalation Backup Ventilation – Pressure Response – Use Table 3-24
a. Establish a delivery flow of [1]
b. Wait [3]
c. Command the exhalation valve to [6]
d. Wait for stable flow for up to [2]
i. If not, FAILURE (Unable to establish stable flow) FE2311
1. Perform a Flow Sensor Calibration.
2. Replace the Delivery Flow Sensor.
3. Check / replace the Exhalation Valve.

e. Wait [3]
1
f. Read alpha-filtered pressure for [4] and average the samples
g. Verify pressure is within test limits [5]
i. In not, FAILURE (Pressure out of range) FE2315
1. Replace the Exhalation Valve.
2. Check / replace the Exhalation Sensors PCBA.

h. Repeat step a to g for [7]


i. Disable Exhalation BUV
Notes
1. Note that the BD should read pressure continuously for 300 msec, and then average all samples.
The rationale for not simply taking the last filtered sample is that when the valve is resonating,
pressure can wander high or low for a brief period, so a total average may be more accurate.

Page 105
15. Test #15: Communication Test

Test Objectives: To verify the GUI CPU communication ports. For the serial ports,
this test verifies that the UART is functioning properly. It does not verify the circuitry
that connects the UART to the connectors. For the Ethernet, this test verifies that the
Ethernet is working between the BD/GUI boards, not that the connector is
functioning.
Method
1. Communication
a. Perform serial loopback test on UART for Line Interface 1 RS-232
i. If loopback fails, ALERT (Bad serial port) AE2001
1. Reseat communications PCBA Line Interface 1;
2. Replace communications PCBA Line Interface 1

b. Perform the following for Ethernet (ETH0_IF):


i. Ping the BD board 4 times.
ii. If ping fails three out of four, FAILURE (GUI to BD ping failed on Ether1) FE2002 –
1. Replace the GUI CPU PCBA.
2. Replace the BD CPU PCBA.

Page 106
16. Test #16: Internal Storage Test
Purpose
To verify that the internal storage device (USB Thumb Drive located on the Communications Line Interface 2
PCBA) is functioning.

Troubleshooting for all codes below:


1) Replace USB Thumb Drive;
2) Replace Communications Line Interface 2 PCBA;

Method
1. Internal Storage
a. Check the integrity of the file system (fx_media_check)
i. If check fails, FAILURE (No media driver) FE0701
b. Check the CRCs of the critical files stored on the internal storage device.
i. If check fails, FAILURE (file CRC failure) not checking CRC
c. Open a test file on the internal USB drive.
i. If open fails, FAILURE (file open error) FE0702
ii.
d. Write to test file on the internal USB drive.
i. If write fails, FAILURE (file write error) FE0703
ii.
e. Close the test file
f. Reopen it as read only.
i. If open fails, FAILURE (file open error) FE0702
ii.
g. Read the test file out of the internal USB drive.
h. Compare the read contents with the original contents
i. If comparison fails, FAILURE (file read error) FE0704
ii.
i. Cleanup the drive.
i. If cleanup fails, FAILURE (file cleanup error) FE0705

Page 107
Test #17: LCD Backlight

Test Objective: To verify that the GUI LCD Backlight intensity is functioning
properly.

Ref# Number Value

1 Minimum backlight sensor reading at max brightness 150 to 500

2 Minimum backlight sensor reading at min brightness 40 to 150

3 Number of seconds to transition 2s

Table 3-25

When the LCD Backlight is set to the maximum value the backlight sensor reading must be greater than 0.5*
BMIN-MAX. A temporary range of from 150 to 500 is established.

When the LCD Backlight is set to the minimum value the backlight sensor reading must be greater than
0.125* BMIN-MAX . A temporary range of from 40 to 150 is established.

Method
1. LCD Backlight – Use Table 3-25
a. Read user interface status from FPGA
b. Verify no light sensor I2C error
i. If not, ALERT (LCD backlight I2C error) AE1001
c. Turn on the LCD Backlight
d. Set the LCD Backlight to the minimum.
e. Wait for [3] seconds.
f. Check that the backlight sensor is within ([1] for min AE1003, [2] for max AE1004)
i. If not, ALERT (LCD backlight OOR)
ii.
g. Repeat steps d to f for the maximum setting.
h. AE1002 Ambient Light Sensor I2C

Troubleshooting:

1) Ensure ventilator has been warmed up for a minimum of 15 minutes prior to running EST.
2) Replace the backlight inverter.
3) Replace the touchscreen assembly.

Page 108
18. Test #18: Status Display Test

Test Objective: verifies that the Status Display:


 LCD is functioning properly,
 Communicates with the Breath Delivery CPU.
 Communicates with the Power Distribution module,
 Communicates with the Compressor module (if installed),
Method
1. Status Display
a. Verify that the status display is reporting that it is communicating with the BD.
i. If not, then FAILURE (Status display to BD comm failed) FE1101
b. Verify that BD is reporting that it is communicating with the status display system.
i. If not, then FAILURE (BD to status display comm failed) FE1102
i.
c. Verify that the status display is okay.
i. If not, then FAILURE (Status Display failed) FE1105

Troubleshooting:

1) Replace the BDU Status Display.


2) Replace the BDU Status Display LCD.

Page 109
19. Test #19 GUI Audio Test

Test Objectives:
To verify that the GUI 360 Omni-directional Alarm Indicators are fully functional. The 360 Omni-directional
Alarm Indicators consists of the following:

1. A single central active status indicator.


2. Two latched alarm status indicators adjacent to the central active status indicator.
3. The alternate gas supply indicator.
4. To verify that the audible alarm is working and can be controlled by the GUI CPU.
5. To verify that the backup system is functioning properly.

Display Parameters
A graphical representation of the 360 Omni-directional Alarm Indicator and visual cues to aid the user.

Method
1. 360 Omni-directional Alarm Indicator
a. Cycle through the central active status indicator colors in the following order: RED, YELLOW,
GREEN, OFF.
b. Prompt the user to confirm that the active status indicator.
i. If not, FAILURE (Active Alarm Indicator failure) FE1201
1. Answer the ventilator prompts appropriately.
2. Replace the GUI Alarm LED PCBA.

c. Set the central active status indicator to OFF.


d. Cycle through the latched alarm status indicator colors in the following order: RED, YELLOW,
OFF.
e. Prompt the user to confirm that the latched alarm status indicator.
i. If not, FAILURE (Latched Alarm Indicator failure) FE1202
1. Answer the ventilator prompts appropriately.
2. Replace the GUI Alarm LED PCBA.

f. Set the latched active status indicator to OFF.


g. Set the alternative gas indicator to ON.

Page 110
h. Prompt the user to confirm that alternative gas indicator is ON.
i. If not, FAILURE (Alternative gas Indicator failure) FE1203
1. Answer the ventilator prompts appropriately.
2. Replace the GUI Alarm LED PCBA.

2. Primary Audio
a. Play the low urgency alarm tone.
b. Prompt the user to confirm that they hear the low urgency alarm tone.
i. If not, FAILURE (Low Alarm sound failure) FE1204
1. Replace the GUI Audio Alarm.

c. Stop playing the low urgency alarm tone.


d. Repeat for medium, high urgency alarm and test tones.
i. If medium urgency alarm fails, FAILURE (Medium Alarm sound failure) FE1205
1. Answer the ventilator prompts appropriately.
2. Replace the GUI Alarm LED PCBA.

ii. If low urgency alarm fails, FAILURE (Medium Alarm sound failure) FE1206
1. Answer the ventilator prompts appropriately.
2. Replace the GUI Alarm LED PCBA.

Page 111
20. Test #20: BD Audio Test

Test Objectives:
To verify that the BD audible alarm is working and that the power fail capacitor has sufficient capacity to meet
the minimum 2 minute alarm time for loss of power.

The BD Application can activate the 360 Omni-directional alarm indicators to identify a high priority alarm by
means of the BD CPLD during occasion when the GUI SPI may have failed. During a high priority alarm the
360 Omni-directional alarm indicators will all flash red and the high priority alarm sound will be played.

Ref# Number Value

1 Number of seconds to charge the power fail capacitor 120 sec

2 Alarm cable voltage range 4.3 to 5.125 volts

3 Power fail capacitor voltage 4.3 to 5.125 volts

4 Max voltage drop (V0 - V1) over 2 s > 0.3 volts

5 Wait time before taking power-fail cap voltage measurements 1 sec

6 Discharge time 2 secs

Table 3-26
Method
1. Secondary Audio – Use Table 3-26
a. Turn off secondary alarm
1
b. Take an 8-sample average of the alarm cable voltage.
c. Verify alarm cable voltage is within limits [2] with the alarm turned off.
i. If not, FAILURE (Bad alarm cable) FE1301
1. Check the BD audio alarm cable.
2. Replace the BD audio alarm.

Page 112
1
d. Take an 8-sample average of the power-fail capacitor voltage.
e. If power-fail capacitor voltage below minimum limit, then allow up to [1] for it to charge.
f. Verify that the power-fail capacitor voltage is within limits [3]
i. If not, FAILURE (Bad power fail cap) FE1302
1. Ensure the ventilator had been plugged into adequate AC power and
warmed up for at least 15 minutes in Service Mode.
2. Replace the Power Distribution PCBA.

g. Stop charging the power-fail capacitor.


h. Activate the BD secondary alarm.
i. Delay [5] seconds
1
j. Take an 8-sample average of the power-fail capacitor voltage = V0
k. Delay [6] seconds.
1
l. Take another 8-sample average of the power-fail capacitor voltage = V1
m. Verify cap voltage decreased
i. If not, FAILURE (Power-Fail Cap not discharging) FE1303
1. Replace the Power Distribution PCBA.

n. Verify V0 - V1 is within limits [4]


i. If not, FAILURE (Bad BD audio). FE1304
1. Check the BD audio alarm cable.
2. Replace the BD audio alarm.

o. Prompt the user to confirm that they hear the alarm tone
i. If not, FAILURE (Piezo alarm failure). FE1308
1. Check the BD audio alarm cable.
2. Replace the BD audio alarm.

p. De-activate the BD secondary alarm.

Page 113
21. Test# 21: Rotary Knob Test

Test Objective:
To verify that the knob is functioning properly by detecting clockwise and counterclockwise knob rotations.

Method
1. Knob
a. Prompt operator to turn knob counterclockwise.
b. Wait up to 10 seconds for knob turn
i. If not, FAILURE (bad knob) FE0901 -
1. Ensure the knob is being rotated per test instructions.
2. Check and/or reseat the GUI rotary encoder cable.
3. Replace GUI rotary encoder.
c. Repeat a to b for clockwise knob turn

Page 114
22. Test # 22: Offscreen Key Test

Test Objective: To verify that each GUI bezel keys are working and to verify that the
GUI Bezel Button Backlight LEDs are operational and may be changed from minimal
to maximum backlight brightness levels.
Method
1. Keyboard
a. For each GUI bezel key, prompt the user to press each key in turn.
b. If a key is not pressed within 15 seconds, FAILURE (error listed below)
c. If the wrong key is pressed, inform the user but do not fail the test. Allow the user the
opportunity to press the right key until timeout.
i. screen brightness fails FE0801

ii. display lock fails FE0802

iii. alarm volume fails FE0805

iv. MAN INSP fails FE0803

v. Inspiratory pause failscc FE0810

vi. Expiratory pause fails FE0811

vii. alarm reset failscc FE0804

viii. alarm silence fails FE0806

d. Turn ON the Screen Lock LED.


e. Prompt the user to verify that the Screen Lock LED is ON.
i. If not, ALERT (Screen lock key LED failure ) AE0807
f. Turn OFF the Screen Lock LED
g. Turn ON the Alarm Silence LED
h. Prompt the user to verify that Alarm Silence LED is ON.
i. If not, ALERT (Alarm Silence key LED failure ) AE0808
i. Turn OFF the Alarm Silence LED

Page 115
j. Pulse the Bezel Backlight.
k. Prompt the user to verify that the Bezel Backlight brightness is changing.
i. If not, ALERT (Key-panel LED back-light failure ) AE0809
l. Restore the Bezel Backlight to its current setting.

Troubleshooting:

1. Ensure you are responding to the prompts as directed within 15 seconds.


2. Clean the GUI touchscreen and then repeat the test.
3. Replace the GUI touchscreen assembly.
4. If failing a brightness test, replace the GUI Backlight PCBA.

Page 116
23. Test # 23: Ventilator Battery Test

Test Objective: test the primary ventilator battery.


1.1.1.28 Installation and Status

Ref# Number Value

1 Max Fan RPM 3600

2 Min Fan RPM 2400

3 Max Battery Temperature 50 C

4 Min Battery Temperature 5C

5 Max board temperature 55 C

6 Min board temperature 10 C

7 Discharge Time 5 mins

8 Charger Mode Wait Time 5 sec

9 Charge Time 5 sec

10 Max Battery Voltage 29.5 V

11 Min Battery Voltage 27.3

Page 117
12 Delivery PSOL open count 1280

13 Exhalation Valve close count 546

14 Exhalation Valve damping 0.8 V

15 Max Battery Voltage Delta for V0 >= 28 V 0.4 V

16 Max Battery Voltage Delta for V0 < 28 V 0.25 V

Table 3-27

 Extended battery is optional so EST is not required.


 The fan shall be operating at 3000 RPM +- 20%.
 The battery temperature shall be within 5 to 50 degrees Celsius.
 The PCBA temperature shall be within 10 to 55 degrees Celsius
 The battery voltage shall be within 27.3 to 29.5 Volts

1.1.1.29 Discharge

 Test the ability to detect loss of AC. (Unplug the vent)


 When on DC test that the battery is discharging.

1.1.1.30 Charge

 When on AC test that the battery is charging.


Method
2. Installation and Status – Use Table 3-27
a. Check that the fan RPM is within the test limits [1] [2]
i. If not, FAILURE (Card cage fan malfunction) FE2101
1. Ensure the card cage fan is properly connected and operational.
2. Replace the fan.
3. Replace the Power Distribution PCBA.

b. Check that board temperature is within test limits [5] [6]


i. If not, FAILURE (Board temperature OOR) FE21010
1. Ensure the card cage fan is properly connected and operational.
2. Replace the fan.
3. Replace the Power Distribution PCBA.

c. Check the installation status of the primary battery.


i. If not, FAILURE (Ventilator primary battery not installed) FE2102
1. Install a primary battery (at least charged to 100 %).
2. Replace the primary battery.

d. Verify the operational status of the primary battery is normal.


i. If not, FAILURE (Ventilator primary battery malfunction) FE2103
1. Install a primary battery (at least charged to 25 %).
2. Replace the primary battery.

e. Verify the temperature of the primary battery is within the test limits [3] [4]
i. If not, FAILURE( Ventilator primary battery temperature OOR) FE2104
1. Ensure the card cage fan is properly connected and operational.
2. Replace the primary battery.

Page 118
f. Check that the primary battery is charged.
i. If not, ALERT (Ventilator primary battery not charged) AE2115
1. Install a primary battery (at least charged to 25 %).
2. Replace the primary battery.

g. Verify the voltage of the primary battery is within the test limits [10] [11]
i. If less than limit [11], ALERT (Ventilator primary battery not charged)
1. Install a primary battery (at least charged to 25 %).
2. Replace the primary battery.

ii. If greater than limit [10], FAILURE (Ventilator primary battery voltage OOR)
FE2116
1. Install a primary battery (at least charged to 25 %).
2. Replace the primary battery.

h. If the extended battery is installed, prompt user to remove extended battery


i. Check the installation status of the extended battery
i. If installed, FAILURE(Ventilator extended battery installed) FE2114
1. Remove extended battery as directed.

3. Discharge – Use Table 3-27


a. Prompt user to remove AC power
b. Apply system load using [12] [13] [14]
c. Verify vent is Not on AC power
i. If not, FAILURE (Still detecting AC power) FE2113
1. Disconnect AC power as directed.

d. Verify Primary battery is discharging


i. If not, FAILURE (Ventilator primary battery malfunction) FE2103
1. Replace the primary battery.

e. Verify operational status of the primary battery is normal


i. If not, FAILURE (Ventilator primary battery malfunction) FE2103
1. Replace the primary battery.

f. Wait [7]. Measure Voltages V0 (at T = 0), V180 (at T = 180 secs), V300 (at T = 300 secs)
g. Verify V0 is within the test limits [3] [4]
i. If not, FAILURE( Ventilator primary battery voltage OOR) FE2116
1. Replace the primary battery.

h. Verify VDelta = (V180 – V300) is within the test limits [15] [16]

4. Charge – Use Table 3-27


a. Prompt User to plug in AC
b. Verify vent is on AC power
c. Verify Primary battery is charging
i. If not, FAILURE (Ventilator primary battery malfunction) FE2103
1. Replace the primary battery.

d. Verify operational status of the primary battery is normal


i. If not, FAILURE (Ventilator primary battery malfunction) FE2116
1. Replace the primary battery.

Page 119
24. Test #24: Compressor Battery

See Ventilator Battery/Power Distribution test with the following exceptions:

 All messages reference the compressor batteries and compressor power system
 Compressor Battery test does not check fan RPM range
 Compressor Battery test does not check board temperature
 Apply compressor load using Table 3-28:

Ref# Number Value

1 Compressor Mode TEST

2 Compressor Unload Solenoid OPEN

3 Compressor Motor Speed 1100 RPM

Table 3-28

Troubleshooting:

1. Follow the prompts as directed.


2. Ensure the compressor cooling fan is working.
3. Use compressor primary battery that is at least 25% charged.
4. Replace the compressor primary battery.

Page 120
25. Test #25: Compressor Test

Purpose
To verify the following components’ operation: pressure transducer, fan, motor and pressure relief valve. The
pressure transducer is cross checked against the air inlet pressure transducer.

Test Objectives
1.1.1.31 Pressure Transducer Zero/Offset

Ref# Number Value

1 Open PSOL counts 2000

2 Pressure drop time out 10 sec

3 Accumulator pressure drop 5 psi

4 Additional pressure drop wait time 5 sec

5 Number of samples 5

6 Low pressure high limit 0.5 psi

Table 3-29
 The offset of the compressor accumulator pressure transducer shall be verified to be within 0.5 of
ambient pressure.

1.1.1.32 Fan

 Fan tachometer shall read at 150 +- 20% RPS.

1.1.1.33 Unload solenoid and motor speed

Ref# Number Value

1 Motor on timeout 5 sec

2 Additional Motor on wait 1 sec

3 Motor failure RPM 0

4 Number of samples 5

5 Motor target RPM 3800

6 Motor RPM limit Target RPM * 0.1

7 Pressure stabilization timeout 5 sec

8 Max pressure limit 15 psi

Table 3-30
 When the unload solenoid is open and the compressor is drained the accumulator pressure shall be
less than 15 psig.

Page 121
 Target motor speed is 3800 rpm + 10%

1.1.1.34 Pressure Transducer Cross Check

Ref# Number Value

1 Delivery PSOL open count 2000

2 Pressure drop timeout 10 sec

3 Compressor high pressure 15 psig

4 Compressor low pressure 12 psig

5 Motor on timeout 2 sec

6 Motor off timeout 30 sec

7 Pressure stabilization timeout 200 ms

8 Number of samples 5

9 Pressure failure limit 2 psig

Table 3-31

 The compressor accumulator pressure transducer shall be verified to be within + 2 psig of the air inlet
pressure transducer.
 The pressure transducer cross check shall be verified at the following pressures: 15 and 25 psig.

1.1.1.35 Pressure Relief Valve

1 Target motor speed 3800

2 Pressurization timeout 10 sec

3 Pressure stabilization timeout 30 sec

4 Max allowed pressure 40 psig

5 Number of samples 5

Table 3-32
 Max RPM for motor speed is 3800.
 The pressure relief valve shall limit the compressor pressure to 40 psig

1.1.1.36 Pressure Transducer Offset

 offset, min, max

1.1.1.37 Fan

 Tachometer value for the fan

Page 122
1.1.1.38 Unload solenoid and motor speed

 pressure, max
 Target motor speed, actual motor speed, min, max

1.1.1.39 Pressure Relief Valve

 Expected relief pressure, actual relief pressure

1.1.1.40 Temperature

 Actual temperature reading, min, max.


Method
1. Compressor Test Initialization
a. Prompt user to disconnect wall air, if wall air is connected
b. Check for wall air connection.
i. If connected, FAILURE(Wall air pressure detected) FE2402
1. Follow the prompts as directed.

2. Pressure Transducer Zero/Offset – Use Table 3-29


a. Deactivate the compressor
b. Open the safety valve
c. Open the mix purge valve
d. Open delivery PSOL to [1]
e. Wait up to [2] seconds for accumulator pressure to drop below [3] psig.
i. If not, FAILURE (Unable to drain accumulator) FE2403
1. Check for crimps or occlusions in the compressor/BDU interconnect tubing.
2. Replace the Compressor Interface PCBA.

f. Wait [4] seconds.


g. Take a [5] sample average of the compressor accumulator pressure sensor
h. Verify that the average value is within test limits [6].
i. If not, FAILURE (Pressure sensor zero offset out of range) FE2404
1. Replace the Compressor Interface PCBA.

i. Close the safety valve


j. Close the mix purge valve
k. Close the delivery PSOL ( 0 counts)
3. Verify the compressor unload solenoid and compressor motor – Use Table 3-30Error! Reference
source not found.
a. Set compressor into test mode
b. Open the unload solenoid
c. Set the motor speed to target rpm [5]
d. Turn compressor motor on
e. Wait up to [1] sec for the motor to turn on
f. Verify the motor speed using the tachometer.
i. If the motor speed is [3], then FAILURE (Motor speed at zero) FE2406
1. Replace the compressor motor.
g. Take a [4] sample average of the motor speed.
h. Verify that the difference between average value and target value is within the test limit [6]
i. If not ALERT (Motor speed out of range) AE2407
1. Replace the Compressor Motor.

i. Wait [7] seconds for pressure to stabilize


j. Take a [4] sample average of the accumulator pressure
k. Verify that the average value is within the test limits [8].
i. If not, FAILURE Pressure exceeded maximum value FE2408
Page 123
1. Replace the Compressor Interface PCBA.

l. Close unload solenoid


4. Cross check the compressor pressure sensor at 15 psig and 25 psig – Use Table 3-31
a. Disable compressor
b. Open mix safety valve
c. Open mix purge valve
d. Open delivery PSOL to [1] counts
e. Wait [2] seconds for the compressor pressure to drop
f. Close safety valve
g. Close purge valve
h. Close delivery PSOL
i. Put compressor into normal mode
j. Enable compressor
k. Set compressor high pressure target to [3] psig
l. Set compressor low pressure target to [4] psig
m. While the compressor motor is turning, test compressor fan (do this only when testing the first
pressure – 15psig)
n. Verify that the fan is within test limits
i. If not, a FAILURE (Fan speed out of range) FE2405
1. Replace the compressor cooling fan.

o. Wait up to [6] seconds for the compressor motor to turn off (pressure is built)
i. If not, a FAILURE (Compressor motor failed to stop) ) FE2409
1. Replace the Compressor Interface PCBA.
2. Replace the Compressor Motor.

p. Wait [7] milliseconds for pressure to stabilize.


q. Take a [8] sample average of the air inlet and compressor accumulator pressure readings.
r. Verify that the difference of the average readings are within the test limits [9].
i. If not, FAILURE (Pressure sensor out of range) ) FE2410
1. Check for leaks or occlusions.
2. Ensure the Compressor Pressure Sensor tubing is properly connected.
3. Replace the Compressor Interface PCBA.
4. Replace the Compressor Motor.

s. Repeat for 25 psig.


5. Verify the pressure relief valve – Use Table 3-32
a. Put the compressor into test mode
b. Set the motor speed to max rpm [1].
c. Wait [2] for pressure to build
d. Wait up to [3] seconds for pressure to stabilize.
i. If not, FAILURE (Unable to establish pressure) FE2411
1. Check for leaks or occlusions.
2. Ensure the Compressor Pressure Sensor tubing is properly connected.
3. Replace the Compressor Interface PCBA.

e. Take a [5] sample average of the accumulator pressure.


f. Verify that the average is less than the test limit [4].
i. If not, FAILURE (Pressure exceeded maximum value) FE2412
1. Check for leaks or occlusions.
2. Ensure the Compressor Pressure Sensor tubing is properly connected.
3. Replace the Compressor Interface PCBA.

Page 124
26. Test # 26: Compressor Leak Test

Test Objective: check the compressor system for leaks.


Ref# Number Value

1 Pressurization timeout 12 s

2 Maximum leak 0.43 psig

3 Time before taking first measurement 10 s

4 Time between first and second measurements 5s

5 Delivery PSOL open count 2000

6 Mix PSOLs open count 2000

7 Pressure drop timeout 5 sec

8 Compressor initial RPM 2000

9 Compressor initial high pressure 32 psig

10 Compressor initial low pressure 28 psig

11 Compressor motor on timeout 1 sec

12 Compressor motor off timeout 12 sec

13 Number of samples 5

Table 3-33

The leak shall be +- .25 psig after 5 seconds. The test shall fail if the pressure drops below 28 psig or rises
above 35 psig at any time during the test.

Method
1. Compressor Test Initialization
a. Prompt the user to disconnect wall air, if wall air is connected
b. Check for wall air connection.
i. If connected, FAILURE (Wall air pressure detected) FE2502
1. Follow prompts as directed.

2. Leak – Use Table 3-33


a. Prompt user to disconnect oxygen, if oxygen is connected.
b. Put the mix into test mode
c. Open the safety valve
d. Open the delivery PSOL to [5] counts
e. Open the air and oxygen PSOLs to [6] counts
f. Wait up to [7] seconds for pressure to drop below 25 psig.
i. If not, FAILURE (Unable to drain accumulator) FE2503
1. Check for crimps or occlusions in the compressor/BDU interconnect tubing.
2. Replace the Compressor Interface PCBA.

Page 125
g. Close the safety valve
h. Close the delivery PSOL
i. Prompt user to block the inlet port to prevent a delivery PSOL leak from skewing the results.
j. Put compressor into normal mode
k. Activate the compressor ( set to initial conditions [9]/[10] psig,[8] RPM, pressurization
timeout to [1] seconds )
l. Wait up to [11] second for the compressor motor to turn on.
i. If not FAILURE (Unable to establish pressure) FE2504
1. Check for leaks or occlusions.
2. Verify compressor motor turns on.
3. Replace the compressor Interface PCBA.

m. Wait up to [12] seconds for compressor motor to turn off.


n. Check that the compressor did not report a pressurization timeout fault.
o. Check that pressure is less than maximum.
p. Deactivate the compressor
q. Wait [3] seconds
r. Take a 5 sample average
s. Wait [4] seconds
t. Take a [13] sample average.
u. Take the difference of the two averages. Verify difference is less than [2] psig.
i. If not FAILURE (Excessive leak) FE2505
1. Check for leaks or occlusions.
2. Verify compressor motor turns on.
3. Replace the compressor Interface PCBA.

v. Check that pressure is greater than minimum.


w. Prompt user to unblock the inlet port

Page 126
27. Test # 27: Compressor Performance Test

Test Objective: test the compressor under dynamic conditions

Ref# Number Value

1 Delivery PSOL open count 2000

2 Compressor low pressure 23 psig

3 Compressor pressure drop timeout 5 sec

4 Compressor motor high RPM 3800

5 Compressor normal pressure 28 psig

6 Compressor normal pressure timeout 10 sec

7 Exhalation valve damping 0.1 V

8 Performance test flow rate 35 lpm

9 Flow stabilization timeout 2 sec

10 Pressure and flow monitoring time 20 sec

Table 2.18-34

 Flow rate shall be 35±2 (lpm BTPS).


 Compressor pressure shall be within 23 to 33 psig.
Method
1. Preconditions
a. Prompt user to disconnect wall air, if wall air is connected.
b. Check for wall air connection.
i. If connected, FAILURE (Wall air pressure detected) FE2602
1. Follow prompts as directed.

2. Performance- Use Table 2.18-34


a. Open the safety valve
b. Open the delivery PSOL to [1] counts
c. Wait up to [3] seconds for accumulator pressure to drop below [2] psig.
i. If not, FAILURE (Unable to drain accumulator) FE2603
1. Check for crimps or occlusions in the compressor/BDU interconnect tubing.
2. Replace the Compressor Interface PCBA.

d. Close the safety valve


e. Close the delivery PSOL
f. Put compressor into test mode
g. Set the high motor speed to [4] RPM.
h. Turn compressor motor on
i. Wait up to [6] seconds for compressor pressure to rise above [5] psig.
i. If not, FAILURE (Target pressure not reached) FE2604

Page 127
1. Check for leaks or occlusions.
2. Replace the Compressor Interface PCBA.
3. Replace the Compressor Motor.

j. Set exhalation valve damping to [7] V


k. Set a flow rate of [8] slpm using the delivery PSOL
l. Wait [9] seconds for flow to stabilize.
i. If not, FAILURE (Unable to establish flow) FE2605
1. Check for leaks or occlusions.
2. Replace the Compressor Motor.
3. Replace the Compressor Interface PCBA.

m. Monitor pressure and flow for [10] seconds.


n. If pressure is not within test limits, ALERT ( Unable to maintain pressure) AE2606
1. Check for leaks or occlusions.
2. Replace the Compressor Motor.
3. Replace the Compressor Interface PCBA.

o. If flow is not within test limits, ALERT ( Unable to maintain flow) AE2607
1. Check for leaks or occlusions.
2. Replace the Compressor Interface PCBA.
3. Replace the Compressor Motor.

Page 128

You might also like