HS40 v1.04.01 Technical Note Eng
HS40 v1.04.01 Technical Note Eng
01)
Date 2021-05-26 Written by Jason Jung
Model HS40 Approved by Donghyun Kim
HS40_V1.04.01.3136_OS_
OS UAI HS40_V1.04.01.3140_UAI_0x8D25FE74.iso
ADVENTECH_0x83F109E9.iso
1. Summary
HS40 v1.04.01 release
2. Details
1) Software Information
AS-IS TO-BE S/W Compatibility
v1.04.00b v1.04.01 Compatible
* You must use only approved versions of SW in your country.
2) Change List
① Add and Modify related to S/W functions
HQ Vision: Added LS6-15 probe (MSK → General preset)
- It was applied on v1.04.00. but, actual use is now available.
Display Diagnostic Version information in the bottom left when running BIST (Built In Self Test)
Added Tech Mode (Only for Russia)
Fixed multi language error in Chinese and Portuguese
Added: Add functionality and probe for US Region only
- Veterinary preset, NerveTrack functionaltity
- CA4-10M Probe
② MR (Modification Request)
The issue of PW doppler sound amplifying during scanning has been corrected.
- Modified PW sound parameter and windows equilizer for all probes/presets
CONFIDENTIAL
THIS DOCUMENT CONTAINS CONFIDENTIAL INFORMATION WHICH IS PROPRIETARY TO SAMSUNG MEDISON CO., Ltd. THE DOCUMENT AND ITS CONTENTS
(INFORAMTION) CANNOT BE COPIED OR DISCLOSED WITHOUT THE EXPRESSED WRITTEN CONSENT OF SAMSUNG MEDISON CO., Ltd.
v1.04.00a ~ v1.04.00b v1.04.01 Both OS & UAI installation
HS40_V1.04.01.3136_OS_
v1.04.01 HS40_V1.04.01.3140_UAI_0x8D25FE74.iso
ADVENTECH_0x83F109E9.iso
③ Upgrade Table
Target
SW Version in use HDD (SSD) in use Upgrade Type
SW Version
USO-US4W1P2/KR (BE board + PC Assy
HDD
v1.00.00 ~ v1.01.xx + SSD + Windows 10) + S/W upgrade
v1.04.00a ~
SSD USO-US4W1P1/KR (BE board + PC Assy
CONFIDENTIAL
THIS DOCUMENT CONTAINS CONFIDENTIAL INFORMATION WHICH IS PROPRIETARY TO SAMSUNG MEDISON CO., Ltd. THE DOCUMENT AND ITS CONTENTS
(INFORAMTION) CANNOT BE COPIED OR DISCLOSED WITHOUT THE EXPRESSED WRITTEN CONSENT OF SAMSUNG MEDISON CO., Ltd.
Technical Note (v1.04.00b)
Date 2021-03-19 Written by Jason Jung
Model HS40 Approved by Donghyun Kim
HS40_V1.04.00.3108_OS_
OS UAI HS40_XH40_V1.04.00b.3114_UAI_0xE7813F4F.iso
ADVENTECH_0xE03AA843.iso
1. Summary
HS40 v1.04.00b release
2. Details
1) Software Information
AS-IS TO-BE S/W Compatibility
v1.04.00a v1.04.00b Compatible
* You must use only approved versions of SW in your country.
2) Change List
① MR (Modification Request)
Improved : Add "Error code 106105" error generate function when XDMA Device is not detected
during boot
- Solution: Try rebooting first and replace with BE board, DDM Power priority if error persists.
3) Installation Information
① UAI (Ultrasound Application Installer) software
SW Version in USE Target Version For SW Upgrade
OS + UAI
v1.00.00 ~ v1.03.01 v1.04.00b
(‘Upgrade pkg’ purchase required)
CONFIDENTIAL
THIS DOCUMENT CONTAINS CONFIDENTIAL INFORMATION WHICH IS PROPRIETARY TO SAMSUNG MEDISON CO., Ltd. THE DOCUMENT AND ITS CONTENTS
(INFORAMTION) CANNOT BE COPIED OR DISCLOSED WITHOUT THE EXPRESSED WRITTEN CONSENT OF SAMSUNG MEDISON CO., Ltd.
Technical Note (v1.04.00a)
Date 2021-02-09 Written by Jason Jung
Model HS40 Approved by Donghyun Kim
HS40_V1.04.00.3108_OS_
OS UAI HS40_XH40_V1.04.00a.3110_UAI_0x0F0DCC4E.iso
ADVENTECH_0xE03AA843.iso
1. Summary
HS40 v1.04.00a release
2. Details
1) Software Information
AS-IS TO-BE S/W Compatibility
v1.03.01 v1.04.00a After purchasing ‘Upgrade pkg’
* You must use only approved versions of SW in your country.
① Original S/W License key is not available as S/W Option License Key Generation Standard changed.
Ask a relevant ASM to issue new License Key.
② Items required for v1.04.00a upgrade according (see below)
S/W version in USE HDD (SSD) in USE Target S/W Version Upgrade Type
※ When replacing only BE board and new PC and installing software with v1.04.00a instead of replacing SSD on
system where HDD is in use, a pop-up message appears asking to replace HDD with SSD.
2) Change List
① H/W Change
BE board change
- Prior to the change: MI82-02470A
CONFIDENTIAL
THIS DOCUMENT CONTAINS CONFIDENTIAL INFORMATION WHICH IS PROPRIETARY TO SAMSUNG MEDISON CO., Ltd. THE DOCUMENT AND ITS CONTENTS
(INFORAMTION) CANNOT BE COPIED OR DISCLOSED WITHOUT THE EXPRESSED WRITTEN CONSENT OF SAMSUNG MEDISON CO., Ltd.
Assy PC change (AMD RX-425BB → Intel i3-8100H)
- Prior to the change: MI96-01414D
Add "Keep Image and Database" option to keep IMAGE, DataBase folders on E drive when OS reinstallation
3) Installation Information
① UAI (Ultrasound Application Installer) software
SW Version in USE Target Version For SW Upgrade
OS + UAI
v1.00.00 ~ v1.03.01 v1.04.00a
(‘Upgrade pkg’ purchase required)
HS40_V1.04.00.3108_OS_
v1.04.00a HS40_XH40_V1.04.00a.3110_UAI_0x0F0DCC4E.iso
ADVENTECH_0xE03AA843.iso
CONFIDENTIAL
THIS DOCUMENT CONTAINS CONFIDENTIAL INFORMATION WHICH IS PROPRIETARY TO SAMSUNG MEDISON CO., Ltd. THE DOCUMENT AND ITS CONTENTS
(INFORAMTION) CANNOT BE COPIED OR DISCLOSED WITHOUT THE EXPRESSED WRITTEN CONSENT OF SAMSUNG MEDISON CO., Ltd.
Technical Note (v1.03.01)
Date 2020-07-06 Written by Jason Jung
Model HS40 Approved by Donghyun Kim
1. Summary
HS40 v1.03.01 release
2. Details
1) Software Information
AS-IS TO-BE S/W Compatibility
v1.03.00 v1.03.01 Compatible
v1.00.00 ~ v1.01.xx v1.03.01 CrystalLive upgrade pkg(Option) is necessary
* You must use only approved versions of SW in your country.
※ Caution
1. Crystal Live upgrade Pkg should be purchased to upgrade S/W products before v1.02
to above v1.03
- Procedure: Purchase ‘CrystalLive upgrade pkg’ → Install SSD after remove the existing HDD
→ Software Upgrade → FPGA Downlode → EC Update (Refer to the separate technical guide)
2) Change List
① H/W Change
SSD change
- Prior to the change: MI59-01242A , After the change : MI59-01283A
③ MR (Modification Request)
Improved : Probe temperature rise under certain conditions during HPRF operation
Improved : Intermittently system locked up after item change and save in "Setup" menu
CONFIDENTIAL
THIS DOCUMENT CONTAINS CONFIDENTIAL INFORMATION WHICH IS PROPRIETARY TO SAMSUNG MEDISON CO., Ltd. THE DOCUMENT AND ITS CONTENTS
(INFORAMTION) CANNOT BE COPIED OR DISCLOSED WITHOUT THE EXPRESSED WRITTEN CONSENT OF SAMSUNG MEDISON CO., Ltd.
Bug fix : Error code 100003
3) Installation Information
① UAI (Ultrasound Application Installer) software
SW Version in USE Target Version For SW Upgrade
v1.02.00 ~ v1.02.03
(CrystalLive upgrade pkg Option is necessary) OS + UAI
v1.00.00 ~ v1.01.05
v1.03.01 (FPGA download & EC Update)
(CrystalLive upgrade pkg Option is necessary)
v1.01.02 HS40_UAI_V1.01.02.3428.iso
v1.01.03 HS40_UAI_V1.01.03.3470.iso
HS40_OS_V1.01.03.3442.iso
v1.01.04 HS40_UAI_V1.01.04.3471.iso
v1.02.00 HS40_V1.02.00.3582.UAI.iso
v1.02.01 HS40_XH40_V1.02.01.3607_UAI.iso
HS40_V1.02.00.3554.OS.iso
v1.02.02 HS40_V1.02.02.3620_UAI.iso
v1.02.03 HS40_V1.02.03.3624_UAI.iso
v1.03.00 HS40_V1.03.00.3192_UAI.iso
HS40_V1.03.00.3188.OS.iso
v1.03.01 HS40_V1.03.01.3206_UAI_0xD8FF8846.iso
CONFIDENTIAL
THIS DOCUMENT CONTAINS CONFIDENTIAL INFORMATION WHICH IS PROPRIETARY TO SAMSUNG MEDISON CO., Ltd. THE DOCUMENT AND ITS CONTENTS
(INFORAMTION) CANNOT BE COPIED OR DISCLOSED WITHOUT THE EXPRESSED WRITTEN CONSENT OF SAMSUNG MEDISON CO., Ltd.
Technical Note (v1.03.00)
Date 2020-03-30 Written by Heechul Kim
Model HS40 Approved by Donghyun Kim
1. Summary
HS40 v1.03.00 release
2. Details
1) Software Information
AS-IS TO-BE S/W Compatibility
v1.02.03 v1.03.00 Compatible
v1.00.00 ~ v1.01.xx v1.03.00 CrystalLive upgrade pkg(Option) is necessary
* You must use only approved versions of SW in your country.
1. Crystal Live upgrade Pkg should be purchased to upgrade S/W products before v1.02
to above v1.03
- Procedure: Purchase ‘CrystalLive upgrade pkg’ → Install SSD after remove the existing HDD
→ Software Upgrade → FPGA Downlode → EC Update (Refer to the separate technical guide)
3. To register S-Detect for Breast or CrystalVue option, ‘Ace’ Option should be registered
first to activate the options for S-Detect for Breast or CrystalVue
- Ace License is issued at the same time when purchasing S-Detect for Breast or CrystalVue Option
CONFIDENTIAL
THIS DOCUMENT CONTAINS CONFIDENTIAL INFORMATION WHICH IS PROPRIETARY TO SAMSUNG MEDISON CO., Ltd. THE DOCUMENT AND ITS CONTENTS
(INFORAMTION) CANNOT BE COPIED OR DISCLOSED WITHOUT THE EXPRESSED WRITTEN CONSENT OF SAMSUNG MEDISON CO., Ltd.
2) Change List
① Add new Probes
EV2-10A Probe
Panoramic+ improved
3D/4D operation speed is improved
3D Oblique operation speed of angle control is improved
PW Auto Trace performance of Carotid/Arteiral is improved
CONFIDENTIAL
THIS DOCUMENT CONTAINS CONFIDENTIAL INFORMATION WHICH IS PROPRIETARY TO SAMSUNG MEDISON CO., Ltd. THE DOCUMENT AND ITS CONTENTS
(INFORAMTION) CANNOT BE COPIED OR DISCLOSED WITHOUT THE EXPRESSED WRITTEN CONSENT OF SAMSUNG MEDISON CO., Ltd.
According to FDA guideline, Probe Element Check Function is added (Only for USA)
Add New 5 Monitor Test Patterns (TG18-CT, TG18-MP, TG18-OIQ, TG18-PQC, TG18-QC)
Measurement Value is saved when turing off the system
System Activation function for displaying demo system is added
- Definition : System Activation Option for managing to demo product’s period.
- Explanation : Expired Date is set when demo product is shipped. Pop-up message appears when boot-up
and indicates It’s demo product. Pop-up message disappears after click ‘Ok’ button.
⑤ MR (Modification Request)
Symptom that Value of Depth is not displayed when accessing M mode during
Wide Dual mode is fixed
Improved: 2D Cine was occasionally lagging
Error Code(106160) appears when 3D Probe is not recognized.
When Error code(106001) appears caused by not recognizing Touch Panel or Trackball, The system
used to reset three times for re-recognizing but It resets five times.
Improved: Multi Exam Export was not working when exporing as DicomDIR after checking
‘Create patient Folder’
Trend mode of OB measurement error is fixed
- Symptom : Measured values which were measured in the different study was not displayed on the trend
Improved: 4D Cine Dicom file was saved as Single Frame(not Multi Frame) when exporting
Freeze issue in Dual R is fixed
- Symptom : Intermittently, Freeze action is delayed in Dual R mode
CONFIDENTIAL
THIS DOCUMENT CONTAINS CONFIDENTIAL INFORMATION WHICH IS PROPRIETARY TO SAMSUNG MEDISON CO., Ltd. THE DOCUMENT AND ITS CONTENTS
(INFORAMTION) CANNOT BE COPIED OR DISCLOSED WITHOUT THE EXPRESSED WRITTEN CONSENT OF SAMSUNG MEDISON CO., Ltd.
3) Installation Information
① UAI (Ultrasound Application Installer) software
SW Version in USE Target Version For SW Upgrade
v1.02.00 ~ v1.02.03
(CrystalLive upgrade pkg Option is necessary) OS + UAI
v1.00.00 ~ v1.01.05
v1.03.00 (FPGA download & EC Update)
(CrystalLive upgrade pkg Option is necessary)
v1.01.02 HS40_UAI_V1.01.02.3428.iso
v1.01.03 HS40_UAI_V1.01.03.3470.iso
HS40_OS_V1.01.03.3442.iso
v1.01.04 HS40_UAI_V1.01.04.3471.iso
v1.02.00 HS40_V1.02.00.3582.UAI.iso
v1.02.01 HS40_XH40_V1.02.01.3607_UAI.iso
HS40_V1.02.00.3554.OS.iso
v1.02.02 HS40_V1.02.02.3620_UAI.iso
v1.02.03 HS40_V1.02.03.3624_UAI.iso
CONFIDENTIAL
THIS DOCUMENT CONTAINS CONFIDENTIAL INFORMATION WHICH IS PROPRIETARY TO SAMSUNG MEDISON CO., Ltd. THE DOCUMENT AND ITS CONTENTS
(INFORAMTION) CANNOT BE COPIED OR DISCLOSED WITHOUT THE EXPRESSED WRITTEN CONSENT OF SAMSUNG MEDISON CO., Ltd.
Technical Note (v1.02.03)
Date 2020-02-11 Written by Heechul Kim
Model HS40 Approved by Donghyun Kim
1. Summary
HS40 v1.02.03 release
2. Details
1) Software Information
AS-IS TO-BE S/W Compatibility
v1.02.02 v1.02.03 Compatible
* You must use only approved versions of SW in your country.
2) Change List
① MR (Modification Request)
Changed color and shape of cross-hair based on requirement (Only for Germany)
Bug that cross-hair lines are not appeared in certain coordinate is fixed
3) Installation Information
① UAI (Ultrasound Application Installer) software
SW Version in USE Target Version For SW Upgrade
v1.02.00 ~ v1.02.03 OS + UAI
v1.00.00 ~ v1.01.05
(CrystalLive upgrade pkg Option is necessary) (FPGA download & EC Update)
v1.00.01 HS40_UAI_V1.00.01.3321.iso
v1.00.03 HS40_UAI_V1.00.03.3375.iso
v1.00.04 HS40_UAI_V1.00.04.3420.iso
HS40_OS_V1.00.04.3420.iso
v1.00.05 HS40_UAI_V1.00.05.3421.iso
CONFIDENTIAL
THIS DOCUMENT CONTAINS CONFIDENTIAL INFORMATION WHICH IS PROPRIETARY TO SAMSUNG MEDISON CO., Ltd. THE DOCUMENT AND ITS CONTENTS
(INFORAMTION) CANNOT BE COPIED OR DISCLOSED WITHOUT THE EXPRESSED WRITTEN CONSENT OF SAMSUNG MEDISON CO., Ltd.
v1.00.06 HS40_UAI_V1.00.06.3422.iso
v1.00.07 HS40_UAI_V1.00.07.3423.iso
v1.01.00 HS40_UAI_V1.01.00.3419.iso
v1.01.02 HS40_UAI_V1.01.02.3428.iso
v1.01.03 HS40_UAI_V1.01.03.3470.iso
HS40_OS_V1.01.03.3442.iso
v1.01.04 HS40_UAI_V1.01.04.3471.iso
v1.02.00 HS40_V1.02.00.3582.UAI.iso
v1.02.01 HS40_XH40_V1.02.01.3607_UAI.iso
HS40_V1.02.00.3554.OS.iso
v1.02.02 HS40_V1.02.02.3620_UAI.iso
v1.02.03 HS40_V1.02.03.3624_UAI.iso
CONFIDENTIAL
THIS DOCUMENT CONTAINS CONFIDENTIAL INFORMATION WHICH IS PROPRIETARY TO SAMSUNG MEDISON CO., Ltd. THE DOCUMENT AND ITS CONTENTS
(INFORAMTION) CANNOT BE COPIED OR DISCLOSED WITHOUT THE EXPRESSED WRITTEN CONSENT OF SAMSUNG MEDISON CO., Ltd.
Technical Note (v1.02.02)
Date 2019-10-15 Written by Heechul Kim
Model HS40 Approved by Donghyun Kim
1. Summary
HS40 v1.02.02 release
2. Details
1) Software Information
AS-IS TO-BE S/W Compatibility
v1.02.01 v1.02.02 Compatible
* You must use only approved versions of SW in your country.
2) Change List
① Add and Modify about S/W functions
Applied for RuScan60
Add the fuctnion of changing of the model name
→ Reason: The model name is changed :If you use the parts which are from HS30 with HS40
② MR (Modification Request)
Fixed: The calculated values about OB Measurement weren’t indicated in OB Trend.
Fixed: After using 4D, Brightness about 4D was abnormal after re-boot.
3) Installation Information
① UAI (Ultrasound Application Installer) software
SW Version in USE Target Version For SW Upgrade
v1.02.00 ~ v1.02.03 OS + UAI
v1.00.00 ~ v1.01.05
(CrystalLive upgrade pkg Option is necessary) (FPGA download & EC Update)
CONFIDENTIAL
THIS DOCUMENT CONTAINS CONFIDENTIAL INFORMATION WHICH IS PROPRIETARY TO SAMSUNG MEDISON CO., Ltd. THE DOCUMENT AND ITS CONTENTS
(INFORAMTION) CANNOT BE COPIED OR DISCLOSED WITHOUT THE EXPRESSED WRITTEN CONSENT OF SAMSUNG MEDISON CO., Ltd.
SW Version OS UAI
v1.00.00 HS40_OS_V1.00.00.2204.iso HS40_UAI_V1.00.00.3300.iso
v1.00.01 HS40_UAI_V1.00.01.3321.iso
v1.00.03 HS40_UAI_V1.00.03.3375.iso
v1.00.04 HS40_UAI_V1.00.04.3420.iso
v1.00.05 HS40_UAI_V1.00.05.3421.iso
HS40_OS_V1.00.04.3420.iso
v1.00.06 HS40_UAI_V1.00.06.3422.iso
v1.00.07 HS40_UAI_V1.00.07.3423.iso
v1.01.00 HS40_UAI_V1.01.00.3419.iso
v1.01.02 HS40_UAI_V1.01.02.3428.iso
v1.01.03 HS40_UAI_V1.01.03.3470.iso
HS40_OS_V1.01.03.3442.iso
v1.01.04 HS40_UAI_V1.01.04.3471.iso
v1.02.00 HS40_V1.02.00.3582.UAI.iso
v1.02.02 HS40_V1.02.02.3620_UAI.iso
CONFIDENTIAL
THIS DOCUMENT CONTAINS CONFIDENTIAL INFORMATION WHICH IS PROPRIETARY TO SAMSUNG MEDISON CO., Ltd. THE DOCUMENT AND ITS CONTENTS
(INFORAMTION) CANNOT BE COPIED OR DISCLOSED WITHOUT THE EXPRESSED WRITTEN CONSENT OF SAMSUNG MEDISON CO., Ltd.
Technical Note (v1.02.01)
Date 2019-08-23 Written by Heechul Kim
Model HS40 Approved by Donghyun Kim
1. Summary
HS40 v1.02.01 release
2. Details
1) Software Information
AS–IS TO-BE S/W Compatibility
v1.02.00 v1.02.01 Compatible
* You must use only approved versions of SW in your country.
2) Change List
① Add and Modify about S/W functions
Added: TCD Application
Modified: Only these probes(V5-9, LA3-16AD, EVN4-9) are possible using ElastoScan for China
system.
The unsupported preset would be deleted When you set the region as Canada
Added: Default Demo Image
Modified: Inactivate the function(5D Heart)
→ We do not support the function.
But, For previous verisions 5D Heart was activated automatically when you install the dongle.
② MR (Modification Request)
Fixed: It was inmtermittently failed to transmit by DICOM Printer
3) Installation Information
① UAI (Ultrasound Application Installer) software
SW Version in USE Target Version For SW Upgrade
v1.02.00 ~ v1.02.03 OS + UAI
v1.00.00 ~ v1.01.05
(CrystalLive upgrade pkg Option is necessary) (FPGA download & EC Update)
CONFIDENTIAL
THIS DOCUMENT CONTAINS CONFIDENTIAL INFORMATION WHICH IS PROPRIETARY TO SAMSUNG MEDISON CO., Ltd. THE DOCUMENT AND ITS CONTENTS
(INFORAMTION) CANNOT BE COPIED OR DISCLOSED WITHOUT THE EXPRESSED WRITTEN CONSENT OF SAMSUNG MEDISON CO., Ltd.
② OS & UAI Table
SW Version OS UAI
v1.00.00 HS40_OS_V1.00.00.2204.iso HS40_UAI_V1.00.00.3300.iso
v1.00.01 HS40_UAI_V1.00.01.3321.iso
v1.00.03 HS40_UAI_V1.00.03.3375.iso
v1.00.04 HS40_UAI_V1.00.04.3420.iso
v1.00.05 HS40_UAI_V1.00.05.3421.iso
HS40_OS_V1.00.04.3420.iso
v1.00.06 HS40_UAI_V1.00.06.3422.iso
v1.00.07 HS40_UAI_V1.00.07.3423.iso
v1.01.00 HS40_UAI_V1.01.00.3419.iso
v1.01.02 HS40_UAI_V1.01.02.3428.iso
v1.01.03 HS40_UAI_V1.01.03.3470.iso
HS40_OS_V1.01.03.3442.iso
v1.01.04 HS40_UAI_V1.01.04.3471.iso
v1.02.00 HS40_V1.02.00.3582.UAI.iso
HS40_V1.02.00.3554.OS.iso
v1.02.01 HS40_XH40_V1.02.01.3607_UAI.iso
CONFIDENTIAL
THIS DOCUMENT CONTAINS CONFIDENTIAL INFORMATION WHICH IS PROPRIETARY TO SAMSUNG MEDISON CO., Ltd. THE DOCUMENT AND ITS CONTENTS
(INFORAMTION) CANNOT BE COPIED OR DISCLOSED WITHOUT THE EXPRESSED WRITTEN CONSENT OF SAMSUNG MEDISON CO., Ltd.
Technical Note (v1.02.00)
Date 2019-04-19 Written by Heechul Kim
Model HS40 Approved by Duke Choi
1. Summary
HS40 v1.02.00 release
2. Details
1) SW Information
AS–IS TO-BE S/W Compatibility
v1.01.05 v1.02.00 CrystalLive upgrade pkg(Option) is necessary
* You must use only approved versions of SW in your country.
1. Crystal Live upgrade Pkg should be purchased to upgrade S/W products before v1.02
to above v1.03
- Procedure: Purchase ‘CrystalLive upgrade pkg’ → Install SSD after remove the existing HDD
→ Software Upgrade → FPGA Downlode → EC Update (Refer to the separate technical guide)
2) Details
① Add new Probes
SP3-8, L4-7, LS6-15, DP8B Probe
CONFIDENTIAL
THIS DOCUMENT CONTAINS CONFIDENTIAL INFORMATION WHICH IS PROPRIETARY TO SAMSUNG MEDISON CO., Ltd. THE DOCUMENT AND ITS CONTENTS
(INFORAMTION) CANNOT BE COPIED OR DISCLOSED WITHOUT THE EXPRESSED WRITTEN CONSENT OF SAMSUNG MEDISON CO., Ltd.
③ Add and Modify ‘S/W functions’
Add a function of ‘2D NT’
Add a function of ‘HQ Vision’
Add a function of ‘Trapezoidal MultiVision’
One of the option name changed (HelloMom → MobileExport)
Add a function of ‘Log Backup’, If an error would be happened during boot-up
- Menu Title
→ An error occurred when loading the system. Please contact the Customer Service Department.
→ Backup Log Button (USB Memory is needed)
→ Shut Down Button
back-up&Restore can be done in only Service Engineer authority
Add ‘cine save button’ on the touch screen
Add a button which enters Auto IMT+ at User key
- Pop up ‘Flash Message’ when User can not get into Auto IMT+
- Display all measurement values for Far and Near, Because a function of Near is added
- When User re-names the Preset, the name shows on the input field
- After done to diagnose, It is possible that User can choose whether reset the preset as initial state or not
- ‘Pointer’ would be disappeared when User applies ‘Single, Dual and Quad’
Add a fuctnion of ‘Color M Mode’ when User uses the Convex Probe for Fetal Heart
- Supported Probe : C2-5, C2-8, CA2-8AD, VN4-8 Probe
CONFIDENTIAL
THIS DOCUMENT CONTAINS CONFIDENTIAL INFORMATION WHICH IS PROPRIETARY TO SAMSUNG MEDISON CO., Ltd. THE DOCUMENT AND ITS CONTENTS
(INFORAMTION) CANNOT BE COPIED OR DISCLOSED WITHOUT THE EXPRESSED WRITTEN CONSENT OF SAMSUNG MEDISON CO., Ltd.
The symptoms which if Post Gain is changed, it is too much saturated has been improved.
The function which could search by 'AE Title' after add 'Another' to 'Scheduled Station AE Title' has
been added in search conditions of Worklist.
The message for Air Filter Cleaning has been changed.
- The image for replacing Air Filter has been changed.
Cyber Security
- HDD Encryption
- DICOM TLS
3) Installation Infomraiton
① UAI (Ultrasound Application Installer) software
SW Version in USE Target Version For SW Upgrade
v1.02.00 ~ v1.02.03 OS + UAI
v1.00.00 ~ v1.01.05
(CrystalLive upgrade pkg Option is necessary) (FPGA download & EC Update)
CONFIDENTIAL
THIS DOCUMENT CONTAINS CONFIDENTIAL INFORMATION WHICH IS PROPRIETARY TO SAMSUNG MEDISON CO., Ltd. THE DOCUMENT AND ITS CONTENTS
(INFORAMTION) CANNOT BE COPIED OR DISCLOSED WITHOUT THE EXPRESSED WRITTEN CONSENT OF SAMSUNG MEDISON CO., Ltd.
② OS & UAI Table
SW Version OS UAI
v1.00.00 HS40_OS_V1.00.00.2204.iso HS40_UAI_V1.00.00.3300.iso
v1.00.01 HS40_UAI_V1.00.01.3321.iso
v1.00.03 HS40_UAI_V1.00.03.3375.iso
v1.00.04 HS40_UAI_V1.00.04.3420.iso
v1.00.05 HS40_UAI_V1.00.05.3421.iso
HS40_OS_V1.00.04.3420.iso
v1.00.06 HS40_UAI_V1.00.06.3422.iso
v1.00.07 HS40_UAI_V1.00.07.3423.iso
v1.01.00 HS40_UAI_V1.01.00.3419.iso
v1.01.02 HS40_UAI_V1.01.02.3428.iso
v1.01.03 HS40_UAI_V1.01.03.3470.iso
HS40_OS_V1.01.03.3442.iso
v1.01.04 HS40_UAI_V1.01.04.3471.iso
CONFIDENTIAL
THIS DOCUMENT CONTAINS CONFIDENTIAL INFORMATION WHICH IS PROPRIETARY TO SAMSUNG MEDISON CO., Ltd. THE DOCUMENT AND ITS CONTENTS
(INFORAMTION) CANNOT BE COPIED OR DISCLOSED WITHOUT THE EXPRESSED WRITTEN CONSENT OF SAMSUNG MEDISON CO., Ltd.
Technical Note (v1.01.05)
Date 2018-12-24 Written by Sean Park
Model HS40 Approved by Duke Choi
1. Summary
HS40 v1.01.05 release
2. Details
1) S/W information
AS–IS TO-BE S/W Compatibility
v1.01.04 v1.01.05 Compatibility
* You must use only approved versions of SW in your country.
2) Change list
① SW changes
Cyber Security (Anti-Virus SW)
- New UI supports for installing Windows Defender and AVAST.
② MR (Modification Request)
Fixed: Delay between ECG and Doppler signals
Fixed: to make System Serial Number change function fully functional under Service Core Login
3) Installation Information
① UAI (Ultrasound Application Installer) software
SW Version in USE Target Version For SW Upgrade
OS + UAI
v1.00.00 ~ v1.00.07 v1.01.05
* FPGA update is required
OS + UAI
v1.00.00 ~ v1.00.03 v1.00.07
* FPGA update is required
CONFIDENTIAL
THIS DOCUMENT CONTAINS CONFIDENTIAL INFORMATION WHICH IS PROPRIETARY TO SAMSUNG MEDISON CO., Ltd. THE DOCUMENT AND ITS CONTENTS
(INFORAMTION) CANNOT BE COPIED OR DISCLOSED WITHOUT THE EXPRESSED WRITTEN CONSENT OF SAMSUNG MEDISON CO., Ltd.
v1.00.04 ~ v1.00.06 v1.00.07 UAI
v1.00.01 HS40_UAI_V1.00.01.3321.iso
v1.00.03 HS40_UAI_V1.00.03.3375.iso
v1.00.04 HS40_UAI_V1.00.04.3420.iso
v1.00.05 HS40_UAI_V1.00.05.3421.iso
HS40_OS_V1.00.04.3420.iso
v1.00.06 HS40_UAI_V1.00.06.3422.iso
v1.00.07 HS40_UAI_V1.00.07.3423.iso
v1.01.00 HS40_UAI_V1.01.00.3419.iso
v1.01.02 HS40_UAI_V1.01.02.3428.iso
v1.01.03 HS40_UAI_V1.01.03.3470.iso
HS40_OS_V1.01.03.3442.iso
v1.01.04 HS40_UAI_V1.01.04.3471.iso
CONFIDENTIAL
THIS DOCUMENT CONTAINS CONFIDENTIAL INFORMATION WHICH IS PROPRIETARY TO SAMSUNG MEDISON CO., Ltd. THE DOCUMENT AND ITS CONTENTS
(INFORAMTION) CANNOT BE COPIED OR DISCLOSED WITHOUT THE EXPRESSED WRITTEN CONSENT OF SAMSUNG MEDISON CO., Ltd.
Technical Note (v1.01.04, v1.00.07)
Date 2018-10-22 Written by Sean Park
Model HS40 Approved by Duke Choi
1. Summary
HS40 v1.01.04, v1.00.07 release
2. Details
1) S/W information
AS–IS TO-BE S/W Compatibility
v1.01.03 v1.01.04 Compatibility
v1.00.06 v1.00.07 Compatibility
* You must use only approved versions of SW in your country.
2) Change list
① MR (Modification Request)
Fixed: No biopsy needle displayed at 0 degree with V5-9 and EV3-10B
3) Installation Information
① UAI (Ultrasound Application Installer) software
SW Version in USE Target Version For SW Upgrade
OS + UAI
v1.00.00 ~ v1.00.06 v1.01.04
* FPGA update is required
OS + UAI
v1.00.00 ~ v1.00.03 v1.00.07
* FPGA update is required
CONFIDENTIAL
THIS DOCUMENT CONTAINS CONFIDENTIAL INFORMATION WHICH IS PROPRIETARY TO SAMSUNG MEDISON CO., Ltd. THE DOCUMENT AND ITS CONTENTS
(INFORAMTION) CANNOT BE COPIED OR DISCLOSED WITHOUT THE EXPRESSED WRITTEN CONSENT OF SAMSUNG MEDISON CO., Ltd.
v1.00.02 HS40_UAI_V1.00.02.3346.iso
v1.00.03 HS40_UAI_V1.00.03.3375.iso
v1.00.04 HS40_UAI_V1.00.04.3420.iso
v1.00.05 HS40_UAI_V1.00.05.3421.iso
HS40_OS_V1.00.04.3420.iso
v1.00.06 HS40_UAI_V1.00.06.3422.iso
v1.00.07 HS40_UAI_V1.00.07.3423.iso
v1.01.00 HS40_UAI_V1.01.00.3419.iso
v1.01.02 HS40_UAI_V1.01.02.3428.iso
v1.01.03 HS40_UAI_V1.01.03.3470.iso
HS40_OS_V1.01.03.3442.iso
v1.01.04 HS40_UAI_V1.01.04.3471.iso
CONFIDENTIAL
THIS DOCUMENT CONTAINS CONFIDENTIAL INFORMATION WHICH IS PROPRIETARY TO SAMSUNG MEDISON CO., Ltd. THE DOCUMENT AND ITS CONTENTS
(INFORAMTION) CANNOT BE COPIED OR DISCLOSED WITHOUT THE EXPRESSED WRITTEN CONSENT OF SAMSUNG MEDISON CO., Ltd.
Technical Note (v1.01.03)
Date 2018-07-16 Written by Ricky Jeong
Model HS40 Approved by Duke Choi
1. Summary
HS40 v1.01.03 release
2. Details
1) S/W information
AS–IS TO-BE S/W Compatibility
v1.01.02 v1.01.03 Compatibility
* You must use only approved versions of SW in your country.
2) Change list
① S/W function add or Chagnged
Applied to improvements item of lithotripter by Wolf-company.
- Improvement to Spec of size and shape of cross target in live or freeze mode properly
when instaling a UAI file if not correctly compatibility of OS file, It can not install the UAI
file.(Disable of install button)
② MR (Modification Request)
Improved to problem of turn off forced, during booting or during the turn off of the system
Logo display to prevent file corruption when the end of the system
Improved to problem that can not be changed to chinese with shift key after Chinese setting
CONFIDENTIAL
THIS DOCUMENT CONTAINS CONFIDENTIAL INFORMATION WHICH IS PROPRIETARY TO SAMSUNG MEDISON CO., Ltd. THE DOCUMENT AND ITS CONTENTS
(INFORAMTION) CANNOT BE COPIED OR DISCLOSED WITHOUT THE EXPRESSED WRITTEN CONSENT OF SAMSUNG MEDISON CO., Ltd.
Fixed the screen broken of multi-language when created an exam file on Multi-language
(Korean, Chiness, Rusian)
Improved Trace measurement type,
Improved problem of change the print out direction on the print out on printer(P95DE/P95DW)
Intermittently
Improved problem of intermittent long print output when printing(P95DW)
Improved problem that occurred to pop-up("Two service parts has been changed") without
replace the service parts
Improved problem that stopped in loading bar 9% during booting intermittently.
Improved problem that occurred to error message("Exception") on the start of worklisk exam
intermittently
Send images when using Merge PACS, which checks the role information in the DICOM
Association stage Improve the problem of transmission failure with PACS rejecting it
If multiple Avaliable Transfer Syntax is delivered during the Association course, the product does
not abort, select one Transfer Syntax as normal Improve to transmit
Improved to problem of dot in 2D image
Improved to problem that occurred to message ("An error has occurred")when the strain+
3) Installation Information
① UAI (Ultrasound Application Installer) software
SW Version in USE Target Version For SW Upgrade
OS + UAI
v1.00.00 ~ v1.00.03 v1.01.03
* FPGA update is required
v1.00.01 HS40_UAI_V1.00.01.3321.iso
v1.00.03 HS40_UAI_V1.00.03.3375.iso
v1.00.04 HS40_UAI_V1.00.04.3420.iso
v1.00.06 HS40_UAI_V1.00.06.3422.iso
v1.01.00 HS40_UAI_V1.01.00.3419.iso
v1.01.02 HS40_UAI_V1.01.02.3428.iso
CONFIDENTIAL
THIS DOCUMENT CONTAINS CONFIDENTIAL INFORMATION WHICH IS PROPRIETARY TO SAMSUNG MEDISON CO., Ltd. THE DOCUMENT AND ITS CONTENTS
(INFORAMTION) CANNOT BE COPIED OR DISCLOSED WITHOUT THE EXPRESSED WRITTEN CONSENT OF SAMSUNG MEDISON CO., Ltd.
Technical Note (v1.01.02 / v1.00.06)
Date 2018-06-04 Written by Sean Park
Model HS40 Approved by Duke Choi
HS40_OS_1.01.00.3414.iso HS40_UAI_V1.01.02.3428.iso
OS UAI
HS40_OS_1.00.04.3420.iso HS40_UAI_V1.00.06.3422.iso
1. Summary
HS40 v1.01.02 / v1.00.06 release
2. Details
1) SW Information
AS-IS TO-BE S/W compatibility
v1.01.01 v1.01.02 Compatible
v1.00.05 v1.00.06 Compatible
* You must use only approved versions of SW in your country.
2) Change List
① MR (Modification Request)
PW Doppler sensitivity improved for EVN4-9(USP-GN49N0A/xx) and ER4-9(USP-E049N0A/xx)
* Applied to OB application – 1st Trimester and FetalHeart presets
* For some cases, while heartbeat can be seen from 2D image, there is no PW spectrum.
3) Installation Information
① UAI (Ultrasound Application Installer) software
SW Version in USE Target Version For SW Upgrade
OS + UAI
OS + UAI
CONFIDENTIAL
THIS DOCUMENT CONTAINS CONFIDENTIAL INFORMATION WHICH IS PROPRIETARY TO SAMSUNG MEDISON CO., Ltd. THE DOCUMENT AND ITS CONTENTS
(INFORAMTION) CANNOT BE COPIED OR DISCLOSED WITHOUT THE EXPRESSED WRITTEN CONSENT OF SAMSUNG MEDISON CO., Ltd.
② OS & UAI Table
SW Version OS UAI
v1.00.00 HS40_OS_1.00.00.2204.iso HS40_UAI_V1.00.00.3300.iso
v1.00.01 HS40_UAI_V1.00.01.3321.iso
v1.00.03 HS40_UAI_V1.00.03.3375.iso
v1.00.04 HS40_UAI_V1.00.04.3420.iso
v1.00.06 HS40_UAI_V1.00.06.3422.iso
v1.01.00 HS40_UAI_V1.01.00.3419.iso
v1.01.02 HS40_UAI_V1.01.02.3428.iso
CONFIDENTIAL
THIS DOCUMENT CONTAINS CONFIDENTIAL INFORMATION WHICH IS PROPRIETARY TO SAMSUNG MEDISON CO., Ltd. THE DOCUMENT AND ITS CONTENTS
(INFORAMTION) CANNOT BE COPIED OR DISCLOSED WITHOUT THE EXPRESSED WRITTEN CONSENT OF SAMSUNG MEDISON CO., Ltd.
Technical Note (v1.01.01 / v1.00.05)
Date 2018-05-04 Written by Sean Park
Model HS40 Approved by Duke Choi
HS40_OS_1.01.00.3414.iso HS40_UAI_V1.01.01.3428.iso
OS UAI
HS40_OS_1.00.04.3420.iso HS40_UAI_V1.00.05.3421.iso
1. Summary
HS40 v1.01.01 / v1.00.05 release
2. Details
1) SW Information
AS-IS TO-BE S/W compatibility
v1.01.00 v1.01.01 Compatible
v1.00.04
(excluded six new types of probes v1.00.05 Compatible
from the v1.01.00)
2) Change List
① MR (Modification Request)
Fixed: System auto-shutdown during mode changes when LN5-12 (USP-L05CN4A/xx) probe is
used.
* Sequence: [Harmonic OFF] B mode → Color mode → B mode → Color mode
3) Installation Information
① UAI (Ultrasound Application Installer) software
SW Version in USE Target Version For SW Upgrade
OS + UAI
OS + UAI
CONFIDENTIAL
THIS DOCUMENT CONTAINS CONFIDENTIAL INFORMATION WHICH IS PROPRIETARY TO SAMSUNG MEDISON CO., Ltd. THE DOCUMENT AND ITS CONTENTS
(INFORAMTION) CANNOT BE COPIED OR DISCLOSED WITHOUT THE EXPRESSED WRITTEN CONSENT OF SAMSUNG MEDISON CO., Ltd.
② OS & UAI Table
SW Version OS UAI
v1.00.00 HS40_OS_1.00.00.2204.iso HS40_UAI_V1.00.00.3300.iso
v1.00.01 HS40_UAI_V1.00.01.3321.iso
v1.00.03 HS40_UAI_V1.00.03.3375.iso
v1.00.04 HS40_UAI_V1.00.04.3420.iso
HS40_OS_1.00.04.3420.iso
v1.00.05 HS40_UAI_V1.00.05.3421.iso
v1.01.00 HS40_UAI_V1.01.00.3419.iso
HS40_OS_1.01.00.3414.iso
v1.01.01 HS40_UAI_V1.01.01.3428.iso
CONFIDENTIAL
THIS DOCUMENT CONTAINS CONFIDENTIAL INFORMATION WHICH IS PROPRIETARY TO SAMSUNG MEDISON CO., Ltd. THE DOCUMENT AND ITS CONTENTS
(INFORAMTION) CANNOT BE COPIED OR DISCLOSED WITHOUT THE EXPRESSED WRITTEN CONSENT OF SAMSUNG MEDISON CO., Ltd.
Technical Note (v1.01.00 / v1.00.04)
Date 2018-01-25 Written by Sean Park
Model HS40 Approved by Duke Choi
HS40_OS_1.01.00.3414.iso HS40_UAI_V1.01.00.3419.iso
OS UAI
HS40_OS_1.00.04.3420.iso HS40_UAI_V1.00.04.3420.iso
1. Summary
HS40 v1.01.00 release
2. Details
1) SW Information
AS-IS TO-BE S/W compatibility
v1.00.03 v1.01.00 Compatible
v1.00.03 v1.00.04 Compatible
* Please check the SW version for your country before you use.
2) Change List
① New Probes(v1.01.00 Only)
C2-8 (Abdomen, OB, GYN)
C2-5 (Abdomen, OB, GYN)
LN5-12 (Abdomen, MSK, Small Parts)
L5-12/50 (Abdomen, MSK, Small Parts)
ER4-9 (OB, GYN, Urology)
CA2-6BM (Renal)
② H/W Changes
FPGA on BE board download required for Write Zoom + SCI function
* When Main SW version and FPGA version are mismatched, a guide message will show up.
HS40_BE board
FPGA Upgrade manual.pptx
③ S/W Changes
Enable to use Write Zoom + SCI
Enable to save TGC setting for each Probe + Preset
Default settings for Screen Saver and Auto Freeze changed from OFF to ON
* Screen Saver: OFF -> 10 min, Auto Freeze: OFF -> 20 min
CONFIDENTIAL
THIS DOCUMENT CONTAINS CONFIDENTIAL INFORMATION WHICH IS PROPRIETARY TO SAMSUNG MEDISON CO., Ltd. THE DOCUMENT AND ITS CONTENTS
(INFORAMTION) CANNOT BE COPIED OR DISCLOSED WITHOUT THE EXPRESSED WRITTEN CONSENT OF SAMSUNG MEDISON CO., Ltd.
4D transition time improved
* when going in and out of 4D, when retrieving 4D thumbnail from active exam
3) Installation Information
① UAI (Ultrasound Application Installer) software
SW Version in USE Target Version For SW Upgrade
OS + UAI
OS + UAI
v1.00.01 HS40_UAI_V1.00.01.3321.iso
v1.00.03 HS40_UAI_V1.00.03.3375.iso
CONFIDENTIAL
THIS DOCUMENT CONTAINS CONFIDENTIAL INFORMATION WHICH IS PROPRIETARY TO SAMSUNG MEDISON CO., Ltd. THE DOCUMENT AND ITS CONTENTS
(INFORAMTION) CANNOT BE COPIED OR DISCLOSED WITHOUT THE EXPRESSED WRITTEN CONSENT OF SAMSUNG MEDISON CO., Ltd.