Date 2021-12-07 Written by Jason Jung
Model HS40 Approved by Jaeyk Kim
HS40_XH40_v1.01.07b.3526 HS40_XH40_v1.01.07b.3531
OS UAI
_OS_0x72A249F0.iso _UAI_0xEA933F75.iso
1. Summary
HS40 v1.01.07b release
2. Details
1) SW information
As-Is To-Be S/W Compatibility
v1.01.07a v1.01.07 Compatible
* You must use only approved versions of SW in your country.
① Original S/W License key is not available as S/W Option License Key Generation Standard changed.
Ask a relevant ASM to issue new License Key.
2) Details
① H/W Change
Touch Screen has been changed (Will be applied in January 2022)
- Before : MI96-01413G → After : MI96-01413H
CONFIDENTIAL
THIS DOCUMENT CONTAINS CONFIDENTIAL INFORMATION WHICH IS PROPRIETARY TO SAMSUNG MEDISON CO., Ltd. THE DOCUMENT AND ITS CONTENTS
(INFORAMTION) CANNOT BE COPIED OR DISCLOSED WITHOUT THE EXPRESSED WRITTEN CONSENT OF SAMSUNG MEDISON CO., Ltd.
3) S/W & H/W upgrade guide
① How to S/W upgrade
SW Version in use Target version How to S/W upgrade
Both OS & UAI installation
v1.00.00 ~ v1.00.07 v1.01.07b
* Need to do ‘FPGA download’
v1.01.05 HS40_UAI_V1.01.05.3480.iso
HS40_OS_V1.01.05.3472.iso
v1.01.06 HS40_V1.01.06.3498.UAI.iso
v1.01.07 HS40_V1.01.07.3513_UAI_0x30187151.iso
CONFIDENTIAL
THIS DOCUMENT CONTAINS CONFIDENTIAL INFORMATION WHICH IS PROPRIETARY TO SAMSUNG MEDISON CO., Ltd. THE DOCUMENT AND ITS CONTENTS
(INFORAMTION) CANNOT BE COPIED OR DISCLOSED WITHOUT THE EXPRESSED WRITTEN CONSENT OF SAMSUNG MEDISON CO., Ltd.
Technical Note (v1.01.07a)
Date 2020-10-30 Written by Jason Jung
Model HS40 Approved by Donghyun Kim
HS40_V1.01.07a.3517_UAI_only for
OS HS40_V1.01.07.3502_OS_0x53DD619F.iso UAI
china_0xF192C040.iso
1. Summary
HS40 v1.01.07a release
2. Details
1) SW information
As-Is To-Be S/W Compatibility
v1.01.07 v1.01.07a Compatible
* You must use only approved versions of SW in your country.
2) Details
① MR (Modification Request)
Improve the issue that translation of Chinese is incorrect
3) Installation information
① UAI (Ultrasound Application Installer) software
SW Version in USE Target version For SW Upgrade
OS + UAI
v1.00.00 ~ v1.00.07 v1.01.07a
* FPGA update is required
v1.01.05 HS40_UAI_V1.01.05.3480.iso
HS40_OS_V1.01.05.3472.iso
v1.01.06 HS40_V1.01.06.3498.UAI.iso
v1.01.07 HS40_V1.01.07.3513_UAI_0x30187151.iso
CONFIDENTIAL
THIS DOCUMENT CONTAINS CONFIDENTIAL INFORMATION WHICH IS PROPRIETARY TO SAMSUNG MEDISON CO., Ltd. THE DOCUMENT AND ITS CONTENTS
(INFORAMTION) CANNOT BE COPIED OR DISCLOSED WITHOUT THE EXPRESSED WRITTEN CONSENT OF SAMSUNG MEDISON CO., Ltd.
Technical Note (v1.01.07)
Date 19-June-2020 Written by Heechul Kim
Model HS40 Approved by Donghyun Kim
1. Summary
HS40 v1.01.07 release
2. Details
1) SW information
As-Is To-Be S/W Compatibility
v1.01.06 v1.01.07 Compatible
* You must use only approved versions of SW in your country.
2) Details
① Improve ‘Image quality’ and ‘function’
LN5 12 PW wave measure, Auto trace, Limitied Trace detection is improved
- Sensitivity to detect peak/ end point and wave line is improved in all application
- Caution: New Touch (MI96-01413G) will generate "Error Code 106001" when used in v1.01.07 or lower
④ MR (Modification Request)
Symptom occured when saving 3D/4D preset of User Application/Preset which named with
multi language or character is fixed (Error code 102001)
Trend mode of OB measurement error is fixed
- Symptom : Measured values which were measured in the different study was not displayed on the
CONFIDENTIAL
THIS DOCUMENT CONTAINS CONFIDENTIAL INFORMATION WHICH IS PROPRIETARY TO SAMSUNG MEDISON CO., Ltd. THE DOCUMENT AND ITS CONTENTS
(INFORAMTION) CANNOT BE COPIED OR DISCLOSED WITHOUT THE EXPRESSED WRITTEN CONSENT OF SAMSUNG MEDISON CO., Ltd.
trend mode of OB Graph in same patient
Improved: Store Event during store didn’t transfer when ‘Export image save’ failed.
Improved: System was occasionally turend off during booting-up after the message appears.
Improved: System was occasionally truned off during scanning after the message
(‘Unexpected system error’) appears
Improved: System was occasionally locked up during operating the system
Improved : Probe temperature rise under certain conditions during HPRF operation
Auto HR issue is fixed
- Symptom : Auto HR is calculated in half using AutoCalc. of PW mode
3) Installation information
① UAI (Ultrasound Application Installer) software
SW Version in USE Target version For SW Upgrade
OS + UAI
v1.00.00 ~ v1.00.07 v1.01.07
* FPGA update is required
v1.00.01 HS40_UAI_V1.00.01.3321.iso
v1.00.03 HS40_UAI_V1.00.03.3375.iso
v1.00.04 HS40_UAI_V1.00.04.3420.iso
v1.00.05 HS40_UAI_V1.00.05.3421.iso
HS40_OS_V1.00.04.3420.iso
v1.00.06 HS40_UAI_V1.00.06.3422.iso
v1.00.07 HS40_UAI_V1.00.07.3423.iso
v1.01.00 HS40_UAI_V1.01.00.3419.iso
v1.01.02 HS40_UAI_V1.01.02.3428.iso
v1.01.03 HS40_UAI_V1.01.03.3470.iso
HS40_OS_V1.01.03.3442.iso
v1.01.04 HS40_UAI_V1.01.04.3471.iso
v1.01.05 HS40_UAI_V1.01.05.3480.iso
HS40_OS_V1.01.05.3472.iso
v1.01.06 HS40_V1.01.06.3498.UAI.iso
CONFIDENTIAL
THIS DOCUMENT CONTAINS CONFIDENTIAL INFORMATION WHICH IS PROPRIETARY TO SAMSUNG MEDISON CO., Ltd. THE DOCUMENT AND ITS CONTENTS
(INFORAMTION) CANNOT BE COPIED OR DISCLOSED WITHOUT THE EXPRESSED WRITTEN CONSENT OF SAMSUNG MEDISON CO., Ltd.
Technical Note (v1.01.06)
Date 04-Jul-2019 Written by Heechul Kim
Model HS40 Approved by Jason Jung
1. Summary
HS40 v1.01.06 release
2. Details
1) SW information
As-Is To-Be S/W Compatibility
v1.01.05 v1.01.06 Compatible
* You must use only approved versions of SW in your country.
2) Details
① H/W change information
Apply SSD
③ MR (Modification Request)
Improved the slowness when moving ROI box in Color mode
3) Installation information
① UAI (Ultrasound Application Installer) software
SW Version in USE Target version For SW Upgrade
OS + UAI
v1.00.00 ~ v1.00.07 v1.01.06
* FPGA update is required
CONFIDENTIAL
THIS DOCUMENT CONTAINS CONFIDENTIAL INFORMATION WHICH IS PROPRIETARY TO SAMSUNG MEDISON CO., Ltd. THE DOCUMENT AND ITS CONTENTS
(INFORAMTION) CANNOT BE COPIED OR DISCLOSED WITHOUT THE EXPRESSED WRITTEN CONSENT OF SAMSUNG MEDISON CO., Ltd.
② OS & UAI Table
SW Version OS UAI
v1.00.00 HS40_OS_V1.00.00.2204.iso HS40_UAI_V1.00.00.3300.iso
v1.00.01 HS40_UAI_V1.00.01.3321.iso
v1.00.03 HS40_UAI_V1.00.03.3375.iso
v1.00.04 HS40_UAI_V1.00.04.3420.iso
v1.00.05 HS40_UAI_V1.00.05.3421.iso
HS40_OS_V1.00.04.3420.iso
v1.00.06 HS40_UAI_V1.00.06.3422.iso
v1.00.07 HS40_UAI_V1.00.07.3423.iso
v1.01.00 HS40_UAI_V1.01.00.3419.iso
v1.01.02 HS40_UAI_V1.01.02.3428.iso
v1.01.03 HS40_UAI_V1.01.03.3470.iso
HS40_OS_V1.01.03.3442.iso
v1.01.04 HS40_UAI_V1.01.04.3471.iso
v1.01.05 HS40_UAI_V1.01.05.3480.iso
HS40_OS_V1.01.05.3472.iso
v1.01.06 HS40_V1.01.06.3498.UAI.iso
CONFIDENTIAL
THIS DOCUMENT CONTAINS CONFIDENTIAL INFORMATION WHICH IS PROPRIETARY TO SAMSUNG MEDISON CO., Ltd. THE DOCUMENT AND ITS CONTENTS
(INFORAMTION) CANNOT BE COPIED OR DISCLOSED WITHOUT THE EXPRESSED WRITTEN CONSENT OF SAMSUNG MEDISON CO., Ltd.
Technical Note (v1.01.05)
Date 24-Dec-2018 Written by Sean Park
Model HS40 Approved by Duke Choi
1. Summary
HS40 v1.01.05 release
2. Details
1) S/W information
AS–IS TO-BE S/W Compatibility
v1.01.04 v1.01.05 Compatibility
* You must use only approved versions of SW in your country.
2) Change list
① SW changes
Cyber Security (Anti-Virus SW)
- New UI supports for installing Windows Defender and AVAST.
② MR (Modification Request)
Fixed: Delay between ECG and Doppler signals
Fixed: to make System Serial Number change function fully functional under Service Core Login
3) Installation Information
① UAI (Ultrasound Application Installer) software
SW Version in USE Target Version For SW Upgrade
OS + UAI
v1.00.00 ~ v1.00.07 v1.01.05
* FPGA update is required
OS + UAI
v1.00.00 ~ v1.00.03 v1.00.07
* FPGA update is required
CONFIDENTIAL
THIS DOCUMENT CONTAINS CONFIDENTIAL INFORMATION WHICH IS PROPRIETARY TO SAMSUNG MEDISON CO., Ltd. THE DOCUMENT AND ITS CONTENTS
(INFORAMTION) CANNOT BE COPIED OR DISCLOSED WITHOUT THE EXPRESSED WRITTEN CONSENT OF SAMSUNG MEDISON CO., Ltd.
② OS & UAI Table
SW Version OS UAI
v1.00.00 HS40_OS_V1.00.00.2204.iso HS40_UAI_V1.00.00.3300.iso
v1.00.01 HS40_UAI_V1.00.01.3321.iso
v1.00.03 HS40_UAI_V1.00.03.3375.iso
v1.00.04 HS40_UAI_V1.00.04.3420.iso
v1.00.05 HS40_UAI_V1.00.05.3421.iso
HS40_OS_V1.00.04.3420.iso
v1.00.06 HS40_UAI_V1.00.06.3422.iso
v1.00.07 HS40_UAI_V1.00.07.3423.iso
v1.01.00 HS40_UAI_V1.01.00.3419.iso
v1.01.02 HS40_UAI_V1.01.02.3428.iso
v1.01.03 HS40_UAI_V1.01.03.3470.iso
HS40_OS_V1.01.03.3442.iso
v1.01.04 HS40_UAI_V1.01.04.3471.iso
CONFIDENTIAL
THIS DOCUMENT CONTAINS CONFIDENTIAL INFORMATION WHICH IS PROPRIETARY TO SAMSUNG MEDISON CO., Ltd. THE DOCUMENT AND ITS CONTENTS
(INFORAMTION) CANNOT BE COPIED OR DISCLOSED WITHOUT THE EXPRESSED WRITTEN CONSENT OF SAMSUNG MEDISON CO., Ltd.
Technical Note (v1.01.04, v1.00.07)
Date 2018-10-22 Written by Sean Park
Model HS40 Approved by Duke Choi
1. Summary
HS40 v1.01.04, v1.00.07 release
2. Details
1) S/W information
AS–IS TO-BE S/W Compatibility
v1.01.03 v1.01.04 Compatibility
v1.00.06 v1.00.07 Compatibility
* You must use only approved versions of SW in your country.
2) Change list
① MR (Modification Request)
Fixed: No biopsy needle displayed at 0 degree with V5-9 and EV3-10B
3) Installation Information
① UAI (Ultrasound Application Installer) software
SW Version in USE Target Version For SW Upgrade
OS + UAI
v1.00.00 ~ v1.00.06 v1.01.04
* FPGA update is required
OS + UAI
v1.00.00 ~ v1.00.03 v1.00.07
* FPGA update is required
CONFIDENTIAL
THIS DOCUMENT CONTAINS CONFIDENTIAL INFORMATION WHICH IS PROPRIETARY TO SAMSUNG MEDISON CO., Ltd. THE DOCUMENT AND ITS CONTENTS
(INFORAMTION) CANNOT BE COPIED OR DISCLOSED WITHOUT THE EXPRESSED WRITTEN CONSENT OF SAMSUNG MEDISON CO., Ltd.
v1.00.02 HS40_UAI_V1.00.02.3346.iso
v1.00.03 HS40_UAI_V1.00.03.3375.iso
v1.00.04 HS40_UAI_V1.00.04.3420.iso
v1.00.05 HS40_UAI_V1.00.05.3421.iso
HS40_OS_V1.00.04.3420.iso
v1.00.06 HS40_UAI_V1.00.06.3422.iso
v1.00.07 HS40_UAI_V1.00.07.3423.iso
v1.01.00 HS40_UAI_V1.01.00.3419.iso
v1.01.02 HS40_UAI_V1.01.02.3428.iso
v1.01.03 HS40_UAI_V1.01.03.3470.iso
HS40_OS_V1.01.03.3442.iso
v1.01.04 HS40_UAI_V1.01.04.3471.iso
CONFIDENTIAL
THIS DOCUMENT CONTAINS CONFIDENTIAL INFORMATION WHICH IS PROPRIETARY TO SAMSUNG MEDISON CO., Ltd. THE DOCUMENT AND ITS CONTENTS
(INFORAMTION) CANNOT BE COPIED OR DISCLOSED WITHOUT THE EXPRESSED WRITTEN CONSENT OF SAMSUNG MEDISON CO., Ltd.
Technical Note (v1.01.03)
Date 2018-07-16 Written by Ricky Jeong
Model HS40 Approved by Duke Choi
1. Summary
HS40 v1.01.03 release
2. Details
1) S/W information
AS–IS TO-BE S/W Compatibility
v1.01.02 v1.01.03 Compatibility
* You must use only approved versions of SW in your country.
2) Change list
① S/W function add or Chagnged
Applied to improvements item of lithotripter by Wolf-company.
- Improvement to Spec of size and shape of cross target in live or freeze mode properly
when instaling a UAI file if not correctly compatibility of OS file, It can not install the UAI
file.(Disable of install button)
② MR (Modification Request)
Improved to problem of turn off forced, during booting or during the turn off of the system
Logo display to prevent file corruption when the end of the system
Improved to problem that can not be changed to chinese with shift key after Chinese setting
Fixed the screen broken of multi-language when created an exam file on Multi-language
CONFIDENTIAL
THIS DOCUMENT CONTAINS CONFIDENTIAL INFORMATION WHICH IS PROPRIETARY TO SAMSUNG MEDISON CO., Ltd. THE DOCUMENT AND ITS CONTENTS
(INFORAMTION) CANNOT BE COPIED OR DISCLOSED WITHOUT THE EXPRESSED WRITTEN CONSENT OF SAMSUNG MEDISON CO., Ltd.
(Korean, Chiness, Rusian)
Improved Trace measurement type,
Improved problem of change the print out direction on the print out on printer(P95DE/P95DW)
Intermittently
Improved problem of intermittent long print output when printing(P95DW)
Improved problem that occurred to pop-up("Two service parts has been changed") without
replace the service parts
Improved problem that stopped in loading bar 9% during booting intermittently.
Improved problem that occurred to error message("Exception") on the start of worklisk exam
intermittently
Send images when using Merge PACS, which checks the role information in the DICOM
Association stage Improve the problem of transmission failure with PACS rejecting it
If multiple Avaliable Transfer Syntax is delivered during the Association course, the product does
not abort, select one Transfer Syntax as normal Improve to transmit
Improved to problem of dot in 2D image
Improved to problem that occurred to message ("An error has occurred")when the strain+
3) Installation Information
① UAI (Ultrasound Application Installer) software
SW Version in USE Target Version For SW Upgrade
OS + UAI
v1.00.00 ~ v1.00.03 v1.01.03
* FPGA update is required
v1.00.01 HS40_UAI_V1.00.01.3321.iso
v1.00.03 HS40_UAI_V1.00.03.3375.iso
v1.00.04 HS40_UAI_V1.00.04.3420.iso
v1.00.06 HS40_UAI_V1.00.06.3422.iso
v1.01.00 HS40_UAI_V1.01.00.3419.iso
v1.01.02 HS40_UAI_V1.01.02.3428.iso
CONFIDENTIAL
THIS DOCUMENT CONTAINS CONFIDENTIAL INFORMATION WHICH IS PROPRIETARY TO SAMSUNG MEDISON CO., Ltd. THE DOCUMENT AND ITS CONTENTS
(INFORAMTION) CANNOT BE COPIED OR DISCLOSED WITHOUT THE EXPRESSED WRITTEN CONSENT OF SAMSUNG MEDISON CO., Ltd.
Technical Note (v1.01.02 / v1.00.06)
Date 2018-06-04 Written by Sean Park
Model HS40 Approved by Duke Choi
HS40_OS_1.01.00.3414.iso HS40_UAI_V1.01.02.3428.iso
OS UAI
HS40_OS_1.00.04.3420.iso HS40_UAI_V1.00.06.3422.iso
1. Summary
HS40 v1.01.02 / v1.00.06 release
2. Details
1) SW Information
AS-IS TO-BE S/W compatibility
v1.01.01 v1.01.02 Compatible
v1.00.05 v1.00.06 Compatible
* You must use only approved versions of SW in your country.
2) Change List
① MR (Modification Request)
PW Doppler sensitivity improved for EVN4-9(USP-GN49N0A/xx) and ER4-9(USP-E049N0A/xx)
* Applied to OB application – 1st Trimester and FetalHeart presets
* For some cases, while heartbeat can be seen from 2D image, there is no PW spectrum.
3) Installation Information
① UAI (Ultrasound Application Installer) software
SW Version in USE Target Version For SW Upgrade
OS + UAI
OS + UAI
CONFIDENTIAL
THIS DOCUMENT CONTAINS CONFIDENTIAL INFORMATION WHICH IS PROPRIETARY TO SAMSUNG MEDISON CO., Ltd. THE DOCUMENT AND ITS CONTENTS
(INFORAMTION) CANNOT BE COPIED OR DISCLOSED WITHOUT THE EXPRESSED WRITTEN CONSENT OF SAMSUNG MEDISON CO., Ltd.
② OS & UAI Table
SW Version OS UAI
v1.00.00 HS40_OS_1.00.00.2204.iso HS40_UAI_V1.00.00.3300.iso
v1.00.01 HS40_UAI_V1.00.01.3321.iso
v1.00.03 HS40_UAI_V1.00.03.3375.iso
v1.00.04 HS40_UAI_V1.00.04.3420.iso
v1.00.06 HS40_UAI_V1.00.06.3422.iso
v1.01.00 HS40_UAI_V1.01.00.3419.iso
v1.01.02 HS40_UAI_V1.01.02.3428.iso
CONFIDENTIAL
THIS DOCUMENT CONTAINS CONFIDENTIAL INFORMATION WHICH IS PROPRIETARY TO SAMSUNG MEDISON CO., Ltd. THE DOCUMENT AND ITS CONTENTS
(INFORAMTION) CANNOT BE COPIED OR DISCLOSED WITHOUT THE EXPRESSED WRITTEN CONSENT OF SAMSUNG MEDISON CO., Ltd.
Technical Note (v1.01.01 / v1.00.05)
Date 2018-05-04 Written by Sean Park
Model HS40 Approved by Duke Choi
HS40_OS_1.01.00.3414.iso HS40_UAI_V1.01.01.3428.iso
OS UAI
HS40_OS_1.00.04.3420.iso HS40_UAI_V1.00.05.3421.iso
1. Summary
HS40 v1.01.01 / v1.00.05 release
2. Details
1) SW Information
AS-IS TO-BE S/W compatibility
v1.01.00 v1.01.01 Compatible
v1.00.04
(excluded six new types of probes v1.00.05 Compatible
from the v1.01.00)
2) Change List
① MR (Modification Request)
Fixed: System auto-shutdown during mode changes when LN5-12 (USP-L05CN4A/xx) probe is
used.
* Sequence: [Harmonic OFF] B mode → Color mode → B mode → Color mode
3) Installation Information
① UAI (Ultrasound Application Installer) software
SW Version in USE Target Version For SW Upgrade
OS + UAI
OS + UAI
CONFIDENTIAL
THIS DOCUMENT CONTAINS CONFIDENTIAL INFORMATION WHICH IS PROPRIETARY TO SAMSUNG MEDISON CO., Ltd. THE DOCUMENT AND ITS CONTENTS
(INFORAMTION) CANNOT BE COPIED OR DISCLOSED WITHOUT THE EXPRESSED WRITTEN CONSENT OF SAMSUNG MEDISON CO., Ltd.
SW Version OS UAI
v1.00.00 HS40_OS_1.00.00.2204.iso HS40_UAI_V1.00.00.3300.iso
v1.00.01 HS40_UAI_V1.00.01.3321.iso
v1.00.03 HS40_UAI_V1.00.03.3375.iso
v1.00.04 HS40_UAI_V1.00.04.3420.iso
HS40_OS_1.00.04.3420.iso
v1.00.05 HS40_UAI_V1.00.05.3421.iso
v1.01.00 HS40_UAI_V1.01.00.3419.iso
HS40_OS_1.01.00.3414.iso
v1.01.01 HS40_UAI_V1.01.01.3428.iso
CONFIDENTIAL
THIS DOCUMENT CONTAINS CONFIDENTIAL INFORMATION WHICH IS PROPRIETARY TO SAMSUNG MEDISON CO., Ltd. THE DOCUMENT AND ITS CONTENTS
(INFORAMTION) CANNOT BE COPIED OR DISCLOSED WITHOUT THE EXPRESSED WRITTEN CONSENT OF SAMSUNG MEDISON CO., Ltd.
Technical Note (v1.01.00 / v1.00.04)
Date 2018-01-25 Written by Sean Park
Model HS40 Approved by Duke Choi
HS40_OS_1.01.00.3414.iso HS40_UAI_V1.01.00.3419.iso
OS UAI
HS40_OS_1.00.04.3420.iso HS40_UAI_V1.00.04.3420.iso
1. Summary
HS40 v1.01.00 release
2. Details
1) SW Information
AS-IS TO-BE S/W compatibility
v1.00.03 v1.01.00 Compatible
v1.00.03 v1.00.04 Compatible
* Please check the SW version for your country before you use.
2) Change List
① New Probes(v1.01.00 Only)
C2-8 (Abdomen, OB, GYN)
C2-5 (Abdomen, OB, GYN)
LN5-12 (Abdomen, MSK, Small Parts)
L5-12/50 (Abdomen, MSK, Small Parts)
ER4-9 (OB, GYN, Urology)
CA2-6BM (Renal)
② H/W Changes
FPGA on BE board download required for Write Zoom + SCI function
* When Main SW version and FPGA version are mismatched, a guide message will show up.
HS40_BE board
FPGA Upgrade manual.pptx
③ S/W Changes
Enable to use Write Zoom + SCI
Enable to save TGC setting for each Probe + Preset
Default settings for Screen Saver and Auto Freeze changed from OFF to ON
* Screen Saver: OFF -> 10 min, Auto Freeze: OFF -> 20 min
CONFIDENTIAL
THIS DOCUMENT CONTAINS CONFIDENTIAL INFORMATION WHICH IS PROPRIETARY TO SAMSUNG MEDISON CO., Ltd. THE DOCUMENT AND ITS CONTENTS
(INFORAMTION) CANNOT BE COPIED OR DISCLOSED WITHOUT THE EXPRESSED WRITTEN CONSENT OF SAMSUNG MEDISON CO., Ltd.
* when going in and out of 4D, when retrieving 4D thumbnail from active exam
3) Installation Information
① UAI (Ultrasound Application Installer) software
SW Version in USE Target Version For SW Upgrade
OS + UAI
OS + UAI
v1.00.01 HS40_UAI_V1.00.01.3321.iso
v1.00.03 HS40_UAI_V1.00.03.3375.iso