Professional Documents
Culture Documents
Caution:
When Sarstedt test tube is not used, make sure that the Container type is "Normal.”
Make sure again that the Container type is normal after adjusting. When adjusting
Sarstedt test tube, adjust the position by referring to "4.1.6 Other Sample Tube
Aspiration Position Adjustment." If the Container type is Sarstedt when using test
tube type other than Sarstedt, aspiration error may occur.
Piercer
Narrow Portion
Rinse Cup
Piercer Tip
Caution:
This adjustment is for using Sarstedt. When using test tupe type other than Sarstedt,
this adjustment is not necessary.
2. Using “Up” or “Down” button, adjust “Up & Down pulse” to be same as recorded value in step
4.1.2.
3. After adjustment is completed, click “Save”
2013109
4.3.1 Preparation
Note: Tube pickup Left/Right position will be adjusted in 4.4.3 sampler profile adjustment.
1. This adjustment starts from main unit to sampler. Turn off sampler until main unit adjustment is
completed. If sampler rack needs to be moved, move sampler rack manually.
2. Open main unit cover and de-activate cover sensor using small piece of paper as shown below.
10. After adjusting hand clipper front/back tilt adjustment, adjust the followings.
3. After hand clipper front/back adjustment is adjusted, the following steps must be performed.
4.4.1 Preparation
1. Turn on sampler power and change mode to sampler mode.
2. Confirm sample rack is interfered with left and right profiles (positioning pin on the rack shifting
belt that push sample rack shifting) in both feed-in and feed-out positions.
*If sample rack is interfered with left and right profiles on both front side belt and back side belt or
spaces between profile and rack is right and left positions are much different, record pulse for
4.4.2 Rack Feed-in Position and 4.4.3 Rack Feed-out Position.
4. Click “Tube pickup position.” When system is XN-2000, select “Left” or “Right” on Analyzer to
select instrument to be adjusted.
5. Adjust sampler rack Left/Right position using “left” and “Right” button so that center of hand
clipper is aligned with center of first sample position on sampler rack.
6. After adjustment is completed, click “Save”
7. Select “Rack feed-out position” to remove sampler rack
8. Select “Close” to close “Sampler Pos. Adjustnment.”
9. Select “Sampler Pos. Adjustnment” again to initialize the movment.
10. Click “Back side” and adjust profile position in the same manner.
11. After both “Front side” and “Backside” adjustments are completed, click “Close” to close
“Sampler Pos. Adjustment” window.
12. Measure some samples in normal sampler mode to confirm both profiles, hand clipper and
sampler rack position are adjusted appropriately.
Note: In case of XN-2000, need to adjust both analyzers after switching “Left/Right” on Analyzer.
2014105
1. Click mode switch button (Gray color) to move sample set adaptor toward you.
2. Place JIG_ASSY NO.203 on front side adaptor, as shown below.
3. Place JIG_ASSY NO.205 on first position of sampler rack. When sample tube holder is attached
on sample rack, remove sample tube holder.
4. Place a sampler rack installed JIG_ASSY NO.204 on sample rack set position.
6. Select “Tube rotor position” to move the sampler rack to the barcode reading position.
7. Barcode reading starts by clicking “Start” button on “Tube Rotation”
8. Loosen fixing screws to adjust barcode reader position so that barcode reader can read barcode
properly. After adjustment is completed, tighten fixing screws.
9. Confirm barcode reader can read barcode properly after tightening fixing screws. When barcode
reader reads barcode properly, barcode LED light green and small beep sounds.
The following adjustment is mandatory after laser diode replacement. When laser diode is
replaced, optical adjustment is necessary as well as laser power adjustment.
201386
Phillips Screwdriver
Oscilloscope
201386
3. Loosen two screws and remove FLOWCELL_ASSY NO.2. If screws are covered by base loosen
Screw F and adjust base position using adjustment screw 3.
Laser Diode (Optical axis direction (front / FSC collector unit (Optical axis direction (front/
1 6
back)) back))
2. Select sensor from maintenance menu on IPU, and confirm FCM detector heater temperature is
40°C ± 1°C.
Note: Wait for heating temperature.
3. Confirm two screws are tighten with approx. 4cN/m (0.4kgf/cm) torque on the following figure.
4. Confirm there are no gaps between laser diode base and FCM block base.
7. Adjust left/right and up/down position on PHOTO DETECTOR_ASSY NO.1 is at the center with
use of Adjustment screw 4 and 5.
5. Press start switch without sample to fill CELLPACK inside the flowcell by performing optical axis
adjustment measurement.
6. Close “Optical Axis Adj.(Fine)” and run one blank measurement. (without sample)
7. Select “Laser Power” on “SERVICE MENU” to turn laser on.
8. Adjust flowcell angle so that laser beam is aligned with PHOTO DETECTOR_ASSY NO.1
pinhole, and fix FLOWCELL_ASSY NO.2 by two hex socket bolts.
* When flowcell sample path lines are unclear, take the following steps.
a. Remove LENS_ASSY NO.2 to observe shadow.
b. Put frosted tape between laser diode and flowcell to view shadow clearly.
3. Install LENS_ASSY NO.2 and center beam stopper shadow on pinhole by adjusting adjustment
screw 10.
* When flowcell sample path line is unclear, take the following steps.
a. Put frosted tape between laser diode and flowcell to view shadow clearly.
2. Adjust adjustment screw 3 and 6 so that edge of LENS_ASSY NO.3 becomes one line with edge
of LENS_ASSY NO.2.
3. Adjust adjustment screw 1 so that chamfered corner of laser diode (cylinder) aligns with the
block surface.
• Since undiluted latex is easy to coagulate, shake it enough and add to CELLPACK DCL.
• Use latex material within 3 hours after making.
• Do not add raw latex to diluted latex.
SSC: 20mV/DIV
Time 0.5μs/DIV
4. Mix diluted FSC latex well and place it on sample set position.
5. Press start switch to measure.
6. Switch to CH1.
7. Perform the following steps so that FSC signal on oscilloscope is higher than 0.5V. If no signal
displays, use adjustment screw 10 so that beam will be split on FSC detector.
8. Switch to CH2.
9. Mix diluted SFL latex well and place it on sample set position and press start switch to measure.
10. Perform the following steps so that SFL signal on oscilloscope is higher than 0.5V.
a. Loosen screws E and D.
d. As wave height increase, change oscilloscope gain CH2: SFL VOLTS/DIV from 50V to 0.5V
stepwise.
e. Using SFL latex, adjust steps above until waveform height is higher than 0.5V and
maximized.
12. Mix diluted SFL latex well and place it on sample set position, and press start switch to
measure.
13. Perform the following steps so that SFL signal on oscilloscope is higher than 1.3V:
a. Adjust laser diode position in laser beam direction (Adjustment screw 1) to maximize
waveform height.
b. Adjust the following order to make waveform height higher than 1.3V:
SSC collector unit laser beam direction (Adjustment screw 3) > APD up/down direction
(Adjustment screw 7) > APD left/right direction (Adjustment screw 8)
* When wave is uniformed, SFL (w) on IPU display will become lower than approx.0.150.
16. Analyze SFL latex and perform the following steps so that SSC signal is higher than 40mV.
a. Loosen two screws (Adjustment screw 9), and adjust left/right position of PHOTO
DETECTOR_ASSY NO.8 (SSC detector) to make the waveform height higher than 40mV.
Target of SSC (w) on IPU screen: lower than 0.250
Note: Cover detector with a palm or cover from light during
adjustment.
c. Put back cover and analyze SFL latex to confirm SSC signal is within the following ranges.
20. Mix diluted FSC latex well and place it on sample set position, and press start switch to
measure.
21. Analyze FSC latex and perform the following steps so that FSC signal is higher than 0.7V:
FSC collector unit laser direction (Adjustment screw 6) > FSC detector up/down direction
(Adjustment screw 5) > FSC detector left/right direction (Adjustment screw 4) to make the
waveform height is higher than 0.7V.
23. Change oscilloscope CH1 from AC to DC. Set oscilloscope to 0.5V range.
24. Analyze FSC latex and perform the following steps so that FSC signal height maintains and
base line is lowest.
a. Adjust beam stopper position (Adjustment screw 10) to one direction.
b. When signal disappears, rotate counter direction.
25. Change oscilloscope CH1 from DC to AC. Set oscilloscope to 0.2V range.
27. Change oscilloscope CH1 from AC to DC. Set oscilloscope to 0.5V range.
28. Analyze FSC latex and perform the following steps so that FSC signal height maintains and
base line is lowest.
29. Change oscilloscope CH1 from DC to AC. Set oscilloscope to 0.2V range.
Caution:
- The last half of the sampling data will be decreased as shown below. This
is a normal behaivier for the optical alignment fine.
SSC signal: higher than 30mV SSC (w): lower than 0.250
FSC signal curve does not have to be a single peak. (Bimodal peaks will not
affect the results.)
201435
FSC signal: higher than 0.7V FSC(w): lower than 0.150
*If signal is out of range, adjust FSC collector position in laser direction (Adjustment screw 6)
FSC signal curve does not have to be a single peak. (Bimodal peaks will not
affect the results.)
Lower limit: 1216 (=76 x 16) Upper limit: 2256 (=141 x 16)
3. Mix diluted FSC latex well and place it on sample set position and press start switch to measure.
While checking FSC signal of oscilloscope, FSC (x) and FSC (w) on IPU, tighten the following
screws.
4. If waveform becomes bad after fixing screws, loosen screws A, B, C and readjust by adjustment
screws 4, 5, 6.
5.Mix diluted SFL latex well and place it on sample set position, and press start switch to measure.
While checking SFL signal of oscilloscope, SFL (x), and SFL(W) on IPU, fixing the following
screws.
a. APD left/right direction (D) and up/down direction (E) 4 screws equally.
b. SSC collector unit (F) 2 screws are torqued to approx. 30cN/m (3kgf/cm)
6. If waveform becomes bad after fixing screws, loosen screws D, E, F and readjust by Adjustment
screw 3, 7, 8.
* If SFL signal is getting lower by loosing screw F, perform steps in the following order:
a. Loosen D, E, F screws.
7. Mix diluted FSC latex well and place it on sample set position and press start switch to measure.
While checking FSC signal of oscilloscope, FSC (x) and FSC (W) on IPU, fixing the following
screws.
a. Tighten laser diode (G) 4 screws with approx.10cN/m (1kgf/cm) torque.
b. Tighten laser diode left/right direction (H) 2 screws with approx. 30cN/m (3kgf/cm) torque.
c. Tighten laser diode (G) 4 screws with approx. 30cN/m (3kgf/cm) torque.
d. Tighten beam stopper (I).
8. If waveform becomes bad after fixing screws, check the situation in the following order.
a. Loosen G, H, I screws and confirm waveform become good.
b. When waveform becomes good, identify which screws cause bad waveform while
analyzing FSC latex.
9. If waveform becomes bad after fixing screw H, check the situation in the following order.
a. Tighten screw H.
b. Analyze SFL latex and adjust adjustment screw 2 so that SFL signal is uniformed.
c. Confirm waveform is within specified range with FSC latex.
d. Fix screw H.
Note: If waveform becomes bad by above steps, try tightening while pushing heater down.
Recommend to wear gloves.
Upper
<130 <0.150 <110 <0.150
Limit
Lower
>66 - >60 -
Limit
Gain 70 x 8 70 x 8
2014124
Upper
- <0.200 - <0.250
Limit
Lower
>46 - >48 -
Limit
Gain 30 x 4 230 x 16
Gain -
FSC signal curve does not have to be a single peak. (Bimodal peaks will not
affect the results.)
201435
FSC signal curve does not have to be a single peak. (Bimodal peaks will not
affect the results.)
201435
<Procedure>
Enter the following value and then click “Send”
HB Blank Gain: 0 - 255
Switching: x1, x2, x4
3. Unmark use box of “Blood Aspiration Sensor” on “Analyzer Setting” and measure blank. (Press
start switch without a sample.) Wait 10 minutes to stabilize LED light.
4. Select “Aspiration Sensor” in “SERVICE MENU”. Enter “125” on Span and click “Send.”
4. Click Execute.
5. Automatically the rack is shifted to blood sensor position. After adjustment the rack is come out
to the left table.
6. Click close.
Note:
a. Adjust sensitivity after registering reagent information.
b. Aspiration sensor and HGB blank adjustment must be completed before
sensitivity adjustment.
6. Click “Close”
7. Place XN-CAL and press start switch.
8. After measurement is completed, click “Sensitivity” to confirm all parameters are within Assay
sheet targets.
2. If values are out of ranges, adjust the parameters which are out of ranges with the following
adjustment procedure. (Enter target and click send)
3. After adjustment is completed, confirm that the data is within ranges, scattergram is normal and
gains are within the following ranges.
Gain
WNR-X 82 - 210
<WDF>
1. Confirm values are within the following ranges. When they are within ranges, close the screen.
WDF-X: XN-CAL Target Value ± 2
2. If values are out of ranges, adjust the parameters which are out of ranges with the following
adjustment procedure. (Enter target and click “Send”)
Gain
WDF-Y: 74 - 194
Adjust sensitivity (4.12 Sensitivity Adjustment) and check that WDF-Y ch SFL Gain is within the
following range.
Lower limit: 1216 (=76 x 16) Upper limit: 2256 (=141 x 16)
2. If values are out of ranges, adjust the parameters which are out of ranges with the following
adjustment procedure. (Enter target and click “Send”)
3. After adjustment is completed, confirm that the data is within ranges, scattergram is normal and
gain is within the following ranges.
Gain
RBC-X: 98 - 244
2. If values are out of ranges, adjust the parameters which are out of ranges with the following
adjustment procedure. (Enter target and click “Send”)
3. After adjustment is completed, confirm that the data is within ranges, scattergram is normal and
gain is within the following ranges.
WPC-Y: 98 - 244
<PLT-F>
Note: Use XN-CAL for the following adjustment
1. Confirm that values are within the following ranges. When they are within the ranges, close the
screen.
PLT-F-RBC-X: XN-CAL Target Value ± 2
2. If values are out of ranges, adjust the parameters which are out of ranges with the following
adjustment procedure. (Enter target and click “Send”)
3. After adjustment is completed, confirm that the data is within ranges, scattergram is normal and
gain is within the following ranges.
Gain
PLT-F-RBC-X: 87 - 220
2. If values are out of ranges, adjust the parameters which are out of ranges with the following
adjustment procedure. (Enter target and click “Send”)
3. After adjustment is completed, confirm that the data is within ranges, scattergram is normal and
gain is within the following ranges.
4. Confirm data on “Main” from “Browser”
2. If values are out of ranges, adjust the parameters which are out of ranges with the following
adjustment procedure. (Enter target and click “Send”)
XN-CAL
4.13.1 Measurement
1. Change mode to manual and whole blood mode on analyzer.
2. Select “Calibration” > “Calibrator Calibration” in Control Menu.
3. Mix XN-CAL, and place it in the sample tube holder. Press the start switch.
4. Analyzer performs 11 times measurement automatically.
313I003
(for RBC-He)
2. Verify whether the deference between the mean for RBC-He and the target value in the assay
sheet is within ±0.5pg.
3. If the mean for RBC-He exceed the range above, perform the calibration by “NVRAM RBC-He”.
4. After performing the calibration, run XN CAL again (repeat step 1 - 2).
5. Verify that the data are within the above range. If it fails, repeat the steps until it falls within the
range.
(for RET-He)
6. Verify whether the deference between the Mean for RET-He and the target value on the assay
sheet is within ±1.0pg.
7. If the Mean for RET-He exceeds the range, perform the calibration by “NVRAM RET-He”.
8. After performing the calibration, run XN CAL again (repeat step 1, 6).
9. Verify that the data are within the above range. If it fails, repeat the steps until it falls within the
range.
201341
Calibration
Acceptable Limit (%) Calibration parameter Acceptable Limit (%)
parameter
WBC 2.87 PLT-I 4.16
WBC-D 2.27 RET% 15.25
RBC 1.25 PLT-O 5.16
HGB 0.78 RBC-O 5.89
HCT 2.64 WBC-P 3.02
201345 201395
XN-CAL (PF)
Note: When analyzer does not have PLT-F parameter, the following calibration is not required.
4.14.1 Measurement
1. Change mode to manual and whole blood mode on analyzer.
2. Select “Calibration” > “Calibrator Calibration (PLT-F)” in control menu.
3. Mix XN-CAL (PF), and place it in the sample tube holder. Press the start switch.
4. Analyzer performs 11 times measurement automatically.
2. By checking calibrator parameter check box, select calibration parameters which need to be
modified. When calibration parameter does not meet the following condition, the check box for
modify cannot be checked.
3. If auto calculated calibration value needs to be changed, check the check box and enter new
calibration value manually.
4. Click “OK”
Note: If new calibration value is significantly changed, analyzer might be something wrong. Therefore, verify the
analyzer condition again before calibration.
201345
XN-CAL
5. Dilute XN-CAL 500µL by CELLPACK DCL 3ml(1:6 dilution), and measure diluted XN-CAL 10
times. When measuring it, remove XN-CAL cap.
201395 2014124
Calibration parameter Acceptable limit (%) Calibration parameter Acceptable limit (%)
WBC 3.76 PLT-I 7.00
WBC-D 2.97 RET% 12.45
RBC 2.10 PLT-O 7.66
HGB 1.31 RBC-O 4.48
201345 201395
XN-CAL (PF)
7. Dilute XN-CAL 500µL by CELLPACK DCL 3ml(1:7 dilution), and measure diluted XN-CAL 10
times. When measuring it, remove XN-CAL cap.
201395
Note: Mix sample each measurement
201345 201395
XN-CAL
Calibration Formula:
Calibration Value(%) = “Assay Value” / “5 times measurement average” x 100 x “Current Calibration Value”
RBC_BF_CAL 1000
Calibration parameter Acceptable limit (%) Calibration parameter Acceptable limit (%)
WBC-BF 2.27 RBC-BF 1.25
4. Click Execute.
5. The rack is shifted to blood sensor position automatically. The rack is come to the left table after
the adjustment is completed.
6. Click close.
201435
2)
3)
the center.
4)
5)
- Raise Bottom Tube cannot be allowed to set as standard sample tube. If selecting "Raised
Bottom Tube", Sample Aspiration error will occur due to the difference of piercer aspiration
position.
6) or
7) 8)
*When CV-60 connects to VIIT or G8, the settings are the same as SP-10.