Professional Documents
Culture Documents
User Guide
Applies for Main Firmware version 1.10
Dear user,
We have made every effort to ensure the accuracy of the contents of this manual.
Should any errors be detected, we would greatly appreciate to receive sugges-
tions to improve the quality of the contents of this manual.
The above not withstanding, we can assume no responsibility for any errors in this
manual or their eventual consequences.
We reserve rights to make modifications to this manual without any further notice.
For more detailed technical data about the MC3 Portable Calibrator, please con-
tact the manufacturer.
Trademarks
QCAL® is a registered trademark owned by Oy Beamex Ab.
Other trademarks are property of their respective owners.
Contents
Contents
A General
Introduction 2 MC3 Firmware 14
About This Manual ................................. 2 General Description ............................. 14
Typographical Conventions .............. 3 Startup Procedure .......................... 14
Unpacking and Inspection ..................... 3 Basic Mode ..................................... 15
Maintenance ................................... 15
MC3 Hardware 4 Calibration Mode ............................ 15
Operational Sections and Connections . 4 The User Interface ............................... 16
The Upper Panel .............................. 5 The Status Bar ................................ 16
The Connector on The Function Key Bar .................... 17
the Left Side of MC3 ......................... 5 Menus ............................................. 17
The Front Panel ................................ 6 The Display Area ............................ 18
Memory .................................................. 9 MC3s Firmware Option ....................... 21
Batteries ............................................... 10
About the Charger and Safety 22
the Charging Procedure ................. 11 Certifications and Compliances
Removing/Replacing (EC Declaration of Conformity) ............ 22
the Battery Pack ............................. 12 Safety Precautions and Warnings ....... 23
Support for Table Top Use ................... 13 General Warnings .......................... 23
The Wrist Strap and Warnings Concerning the use
the Neck Support Strap ........................ 13 of E and ET Sections ...................... 24
The Optional Carrying Case ................ 13 General Warnings Concerning
Pressure Measurement .................. 24
Warnings Concerning
High Pressure ................................ 25
Service 26
Recalibrating MC3 ............................... 26
Cleaning the Contacts of
the Internal Reference Junction ........... 26
The Battery Charger ............................ 26
Contents
D Calibration
Introduction
Thank you for choosing a Beamex calibrator.
MC3 comes in three different models:
· MC3PE is meant for calibrating pressure instruments.
· MC3TE is intended for calibrating temperature instruments.
· MC3MF is a fully equipped multifunction calibrator including the
capabilities of both MC3PE and MC3TE.
This manual applies to MC3MF. Every time a feature that is not avail-
able in all MC3 models is presented, the models the feature applies
to are listed in the beginning of the presentation.
As a member of Beamexs QCAL Quality Calibration family MC3 is,
if equipped with the optional feature, able to communicate with the
available QCAL calibration software.
Thanks to the logical user interface MC3 is very easy to use. The
large backlit graphical display guides the user in different languages
and it displays results both numerically and graphically.
2
Introduction
Typographical Conventions
All examples of user interface texts are printed using 8 pt Arial
Black, e.g.
Selected port: ET: TCi(mea)
All front panel texts (fixed texts on MC3s cover) are printed using
8 pt Eurostile, e.g.
Connectors marked T/C, Low V
Function and Menu keys are often referred to using both the key
name in 8 pt Eurostile and the corresponding text (function)
displayed on the screen in 8 pt Arial Black, e.g.
Function key D/Menu
3
General
MC3 Hardware
General features:
· IP65 water/dust proof case (EN60529)
Battery pack IP30.
· Integrated impact protectors
· Both a wrist strap and a neck support strap
· A support for using the calibrator on the table
· Operating temperature: -10 +50 °C (14 122 °F).
+10 +40 °C (50 104 °F) when charging the batteries.
· Storage temperature: -20 +60 °C (-4 140 °F).
Note: The stickers and the batteries may be affected when
storing longer periods in extreme conditions.
· Humidity: 0 80 %RH
More comprehensive specifications are available in Appendix 1.
E le c tr ic a l
T h e S e c tio n
P re s s u re (E S e c tio n )
U p p e r In p u t
P a n e l
T h e
T e m p e ra tu re
S e c tio n
F ro n t
(E T S e c tio n ) P a n e l
T h e R e fe re n c e
C o m p u te r J u n c tio n
L e ft In te r fa c e B lo c k
S id e
4
MC3 Hardware
Warning!
Use only cables provided by Beamex when connecting MC3 to
a PC.
5
General
T / C IN T . R J 2 -w x m tr
T / C , L o w V R , R T D
4 -w m e a s
V , 1, + 2 4 V 1 m e a s / s in k
T / C O R E X T
W IR E S O N L Y
3 & 4 -w m e a s V , ,
M E A S U R E & S IM U L A T E O U T P U T M E A S U R E
E T E
T / C IN T . R J 2 -w x m tr
T / C , L o w V R , R T D
4 -w m e a s
V , 1, + 2 4 V 1 m e a s / s in k
T / C O R E X T
W IR E S O N L Y
3 & 4 -w m e a s V , ,
M E A S U R E & S IM U L A T E O U T P U T M E A S U R E
E T E
6
MC3 Hardware
7
General
Keyboard
The keys on the MC3s keyboard are grouped according to their
function as follows:
8
MC3 Hardware
Pressing the +/- key toggles the sign of the entered numeric value.
Note. The +/- key is applicable only in numeric fields.
The Decimal key adds the decimal point to the numeric value that
is currently edited.
Memory
MC3 has a dynamic memory allocation system. This means that
there is not a specific area of memory reserved for, e.g. instrument
data. All free memory may be used for anything that requires more
memory. Thus there is no exact limit for the number of instruments
that MC3 can maintain in its memory. It all depends on how much
memory is allocated by other data.
9
General
Batteries
MC3 uses rechargeable batteries. The charger for rechargeable
batteries operates in the following environments:
- Voltage: 100 240 VAC,
- Frequency: 50/60 Hz
The charging electronics is in the Battery Pack. Therefore the bat-
teries may be charged although the Battery Pack is disconnected
from MC3s Base Unit. If you have two sets of rechargeable batter-
ies you may charge the disconnected Battery Pack while at the
same time use MC3 with the connected Battery Pack.
The maximum operating time without recharging varies depending
on the usage and brightness setting of the display light. Also the
generated output current and the usage of the 24V transmitter sup-
ply affect the maximum operating time. Even with constant maxi-
mum load, the standard rechargeable batteries should last for 6
hours. A good average operating time is 10 hours.
Full batteries:
The upper left corner of MC3s display shows a picture of a battery.
The whiter the picture is, the more acute is the need for recharging.
Empty batteries:
Notes.
MC3s memory and the internal clock/calendar uses a small amount
of power although the calibrator is switched off. Remember to check
the capacity of the batteries from time to time although MC3 is not
in use.
Do not leave MC3 without a Battery Pack for a long time. MC3 may
loose its settings if it is left without a support voltage for an ex-
tended period.
10
MC3 Hardware
C h a rg e r c o n n e c to r
C h a r g e s ta tu s lig h t
Warnings!
USE ONLY THE CHARGER PROVIDED WITH THE CALIBRATOR.
The charger accepts input voltages from 100 to 240 VAC.
The charger should only be used indoors and the temperature
should not exceed 40 °C (104 °F).
11
General
1. Turn MC3 upside down (the display facing the table top) and
lift the support.
2. Pull the lever that is hidden under the support. The Battery
Pack pops out allowing you to pull it out.
To replace the Battery Pack, simply slide it on its place. When you
hear a click, the Battery Pack is secured in its place.
Note.
Although the Base Unit is IP65 protected, the Battery Pack is not.
The Battery Pack has holes in order to vent generated gas and
heat. Avoid exposing the Battery Pack to liquids.
12
MC3 Hardware
13
General
MC3 Firmware
General Description
The following picture shortly describes the functions of the firm-
ware. All main functions are marked with a black border. Each main
function has several tasks which are displayed as shaded boxes
without a black border.
S T A R T U P
P R O C E D U R E
S te p p in g a n d R a m p in g
M E N U S T R U C T U R E B A S IC
M O D E M e a s u r e m e n t/G e n e r a tio n /S im u la tio n
C A L IB R A T IO N M A IN T E -
M O D E N A N C E
In s tr u m e n t A d ju s tm e n t
V ie w in g th e R e s u lts S e ttin g T im e a n d D a te
In s tr u m e n t D a ta b a s e M a in te n a n c e A d ju s tin g th e C a lib r a to r
Startup Procedure
Every time MC3 is started the Startup Procedure checks the func-
tionality of the device by performing a self test.
If the self-test is passed successfully, some basic calibrator data is
displayed.
After that MC3 automatically proceeds to Basic Mode. A more com-
prehensive description of the Startup Procedure is in the beginning
of Part B of this manual.
14
MC3 Firmware
Basic Mode
In Basic Mode you can measure and generate/simulate signals.
There are two separately configurable windows available. Basic
Mode is often used for testing connections before starting the ac-
tual calibration procedure of an instrument.
Stepping and Ramping tools enable generating/simulating signals
that vary with time.
All main functions of Basic Mode are described in part B of this
manual.
Maintenance
This main function handles calibrator configuration settings.
Additionally there is the possibility to recalibrate MC3 (requires a
password).
Maintenance level subjects are handled in Part C of this Manual.
Calibration Mode
MC3s main duty is calibrating instruments. Therefore very special
attention was directed on this matter when creating the calibrator.
MC3 may be used as a stand-alone calibrator i.e. all instrument
data and calibration data is saved in MC3s own memory. Option-
ally MC3 also communicates with calibration software.
MC3 supports the use of instruction texts. They help the technician
to perform the calibration as fluently as possible. You may enter
three kinds of instruction texts: Starting Guide, Adjusting Guide
and Finishing Guide. Additionally, calibration notes can be entered
after the calibration procedure.
More calibration related information is available in Part D of this
manual.
Calibration Results
The graphical representation as well as numeric data of the calibra-
tion results may be viewed in MC3. Transferring the results to QCAL®
calibration software makes it possible to view the results in PC en-
vironment.
More information concerning calibration results is presented in Part
D of this manual.
15
General
b a r
P o p -u p lis t
D is p la y a r e a d iv id e d k P a U n it
P a P re s e n t
in to W in d o w 1 a n d 2 2 C u rre n t
b a r
[b a r]
s e le c tio n
E : C u r r e n t M e a s u p r se im e n t
0 .1 1 2m m H 2O
m m H g
P re s s u re
T y p e
[g a u g e ]
Z e ro
P re s s u re
M o d u le
W in d o w 1 W in d o w 2 O th e rs C lo s e
S e tu p S e tu p M o d e M E N U
F u n c tio n k e y b a r
All possible elements are not included in the previous picture, but
the important ones are discussed in the following chapters.
16
MC3 Firmware
W in d o w 1 W in d o w 2 C lo s e
O th e rs M E N U
S e tu p S e tu p M o d e
Menus
The Function Key for opening the 2 9 .0 1 .2 0 0 1 1 3 :2 0
menu is always D/Menu. The same 1 P re s s u re A Q l u a a r mn t i t< y
P 1 : P R 2 C / - 1 . 0 0 0 0 0 0 . . . 2 b [ P a rr e s
key is used when closing the menu. s u re ]
17
General
Key.
5 0 .0 0 ° C (IT S 9 0 )
Calibration: O u tp u t C u r r e n t [E : M e a s .]
Configuration Window: R A M P IN G
C u rre n t
There are plenty of configuration Q u a n tity
P o rt E : I(g e n )
windows in MC3. The picture be-
side is the configuration window for W a it in 0 % 2 s
Ramping settings. R is e T im e 1 s
W a it in 1 0 0 % 2 s
The common thing for all configu- F a ll T im e 2 s
figuration fields.
8 .0 0 0 0
Use the cursor keys to move be- R a n g e 0 %
1 0 0 % 1 6 .0 0 0 0 m A
tween fields. C a n c e l S ta rt
18
MC3 Firmware
Tables: 2 9 .0 1 .2 0 0 1 1 3 :2 7
1 2 .0 2 .2 0 0 0 1 0 :2 9 - A s F o u n d - P a s s e d
Tables are used, e.g. when view- In p u t O u tp u t E rro r
ing calibration results in numeric [V ] [V ] [% ]
format. Tables reserve the whole -0 .0 0 0 0 2 -0 .0 0 0 0 5 0 .0 0 3
display area. 0 .9 9 9 9 6
1 .9 9 9 9 8
0 .9 9 9 9 6
2 .0 0 0 0
0 .0
0 .0
0 0
0 2
3 .0 0 0 0 3 .0 0 0 1 0 .0 0 1
The tables are often larger than the 4 .0 0 0 0 4 .0 0 0 1 0 .0 0 1
display. In that case there are small 5 .0 0 0 0 5 .0 0 0 2 0 .0 0 2
6 .0 0 0 0 6 .0 0 0 1 0 .0 0 1
arrows added to the table borders. 7 .0 0 0 0 7 .0 0 0 1 0 .0 0 1
They indicate that more information 8 .0 0 0 0
9 .0 0 0 0
8 .0 0 0 1
9 .0 0 0 2
0 .0
0 .0
0 1
0 2
may be seen by using the arrow 9 .9 9 9 9 1 0 .0 0 0 1 0 .0 0 2
keys. 9 .0 0 0 0
8 .0 0 0 0
9 .0 0 0 2
8 .0 0 0 2
0 .0
0 .0
0 2
0 2
1 2
The and keys scroll the list one N e x t
B a c k M E N U
P a g e
line at a time. The and keys
scroll the list one page at a time (if 2 5 .0 4 .2 0 0 1 1 2 :1 3
applicable).
M F
Info window:
Pressing the Info key opens up M C 3
S e r ia l N u m b e r: 3 0 3 2 8 2 5 2
the info window. It can be called at M a in V e r s io n : 1 .1 0
E m o d u le v e r s io n : 1 .3 0
any time. The info window displays E T m o d u le v e r s io n : 1 .5 0
Numeric Fields
There are two ways to start editing a R a n g e 0 % 4 .0 0 0 0
1 6 .0 0 0 0
numeric field:
1 0 0 % m A
19
General
Notes.
You cannot add more digits if the length of the number is at its
maximum limit. Use the C/çDelete Function Key to remove un-
wanted digits first and then enter the new digits.
The dual function of the keys: , and is not available in a
numeric field. The keys only represent numbers.
Accept the new text with the D/Accept Function Key. To discard
(cancel) the edited text, use the A/Cancel Function Key.
20
MC3 Firmware
Selection Lists
Selection lists are used when you 2 9 .0 1 .2 0 0 1 1 3 :1 5
have to choose one of several op-
P O S IT IO N /D E V IC E ID
tions. Selection lists are often large, 1 0 1 -X L -0 0 1 .1
thus almost reserving the whole 1 1
1 1
2 -T
2 -T
T -0 0 3 .1
T -0 0 7 .1
window. Selection lists can be E S w
P T 1 0 6 .1
longer than the visible part. When P T 1 1 2 .1 2
P T 1 1 2 .1 5 -1
the cursor (the row with the inverted P T 1 1 2 .1 5 -2
text) is on the bottom and you press P T
T I
1 1 2 .1 6
21
General
Safety
MC3s case is water/dust proof (IP65). The battery pack does how-
ever have holes to enable proper ventilation and heat transfer. So
be careful when working in wet conditions.
The materials of MC3s case withstand normal industrial conditions.
MC3 endures shocks with the help of the built in impact protectors
22
Safety
General Warnings
Use only cables provided by Beamex when connecting MC3 to
a PC.
Use the MC3 battery charger in a non-hazardous indoor loca-
tion only and only with Beamex calibrators.
MC3 uses rechargeable batteries. They are considered as haz-
ardous waste. Dispose used batteries properly according to
local regulations.
Avoid short circuiting the batteries. The short circuit current
may cause burns to you, damage to the device or even fire.
Notice, that also new replacement batteries are shipped in
charged state.
Rechargeable batteries may vent small amounts of gas during
recharge. The vented gas mixture may be highly explosive, but
normally it diffuses rapidly into the atmosphere. To avoid dan-
ger, use only the original charger and never recharge in a gas-
tight container.
The charger should only be used indoors and the temperature
should not exceed 40 °C (104 °F).
23
General
24
Safety
25
General
Service
Only qualified service personnel may perform higher level mainte-
nance for MC3. Never open the case unless have explicit in-
structions from Beamex or a local representative.
There are, however a few things that anyone using MC3 may do.
Recalibrating MC3
Only laboratories approved by Beamex may recalibrate MC3. Con-
tact Beamex or your local representative for information concern-
ing the recalibration of MC3. Contact information is on the first pages
of this User Guide.
26
Startup and Basic
Operation
Starting MC3
Startup Procedure
When MC3 is started, a startup picture appears. After a self test,
some basic information of the calibrator at hand appears in the lower
part of the screen. The graphic in the upper part includes informa-
tion on the model of the MC3 at hand (MC3MF, MC4PE or MC3TE). If
you want to view the calibrator information for a longer period, press
1 8 .0 6 .2 0 0 1 1 2 :2 9
M F
S e r ia l n u m b e r 3 0 3 2 8 2 5 2
M a in v e r s io n 1 .1 0
C a lib r a tio n d u e d a te 1 0 .1 2 .2 0 0 1
W a it
28
Starting MC3
B a s ic M o d e :
M e a s u re m e n t
G e n e r a tio n ...
W h e re
s h o u ld I g o
to d a y ...
29
Startup and Basic Operation
2 C u rre n t
E : C u rre n t M e a s u re m e n t
1 2 .4 7 3 1 m A
C a lib r a tio n
M E N U
M o d e
Next
Measuring on page 31
Generating/Simulating on page 45
Special Generations on page 56.
30
Measuring
Measuring
All measurements in Basic Mode require that you first select the
Window to be used (Commands: Start with D/Menu and continue
either with A/Window 1 Setup or B/Window 2 Setup). Each mea-
surement has its own unique 1/Quantity and 2/Function/Port set-
tings in their windows menu. The other window menu settings, e.g.
measuring unit, refine the measurement characteristics.
Because all measurements are not available in all MC3 models,
the presentation of each measurement include a description of MC3
models supporting this measurement. The required section for the
measurement is also mentioned. If the section is not included in
your MC3, the 1/Quantity and 2/Function/Port settings needed
for the measurement are not available as choices in the pop-up
lists.
Each measurement also has at least one picture with a circle around
some of MC3s terminals, like the one below.
T / C IN T . R J 2 -w x m tr
T / C , L o w V R , R T D
4 -w m e a s
V , 1, + 2 4 V 1 m e a s / s in k
T / C O R E X T
W IR E S O N L Y
3 & 4 -w m e a s V , ,
M E A S U R E & S IM U L A T E O U T P U T M E A S U R E
E T E
T / C IN T . R J 2 -w x m t r
T / C , L o w V R , R T D
4 -w m e a s
V , 1, + 2 4 V 1 m e a s / s in k
T / C O R E X T
W IR E S O N L Y
3 & 4 -w m e a s V , ,
M E A S U R E & S IM U L A T E O U T P U T M E A S U R E
E T E
Warning!
Do not apply voltage higher than 50 V (max 2 A) between any
terminals.
31
Startup and Basic Operation
Pressure Measurement
Pressure measurement is possible with models MC3MF and MC4PE.
See chapter Things to Consider when Measuring Pressure on page 66 for more infor-
mation on pressure measurement and pressure inputs.
Required settings Options/description
Quantity Pressure
Port P1: PRxxxx (or P3:PRB, if applicable)
Pressure Type g gauge pressure or
abs absolute pressure.
3 & 4 -w m e a s V , ,
M E A S U R E & S IM U L A T E O U T P U T M E A S U R E
NOTE!
Zeroing a pressure input is especially important when the op-
erating position of MC3 is changed or the location of MC3 is
changed in the vertical direction. Both of the above mentioned
factors affect notably on the pressure measurement input.
Measuring pressure below 100 mbar (approx. 40 iwc) should
be done with a firmly mounted MC3 (e.g. placed on a table top).
Next
Calibration, see Part D.
32
Measuring
Current Measurement
Current measurement is possible with all MC3 models. The current measurement termi-
nals are located in the E section.
Required settings Options/description
Quantity Current
Funct/Port E: I(meas)
T / C IN T . R J 2 -w x m tr T / C IN T . R J 2 -w x m tr
T / C , L o w V R , R T D
4 -w m e a s
V , 1, + 2 4 V 1 m e a s / s in k T / C , L o w V R , R T D
4 -w m e a s
V , 1, + 2 4 V 1 m e a s / s in k
T / C O R E X T T / C O R E X T
W IR E S O N L Y W IR E S O N L Y
3 & 4 -w m e a s V , , 3 & 4 -w m e a s V , ,
M E A S U R E & S IM U L A T E O U T P U T M E A S U R E M E A S U R E & S IM U L A T E O U T P U T M E A S U R E
E T E E T E
Next
Current Generation on page 47
Calibration, see Part D.
33
Startup and Basic Operation
Voltage Measurement
Voltage measurement is possible with all MC3 models. Low Voltage measurement is
however only possible with MC3MF and MC3TE.
The E section has terminals for voltage measurement within ±50 V range. The ET section
has low voltage measurement terminals with a range of ±500 mV. The ET section termi-
nals are also used when measuring/simulating thermocouples using an external Refer-
ence Junction.
Required settings Options/description
Quantity Voltage
Funct/Port ET: LowV(mea), or E: V(meas)
3 & 4 -w m e a s V , ,
M E A S U R E & S IM U L A T E O U T P U T M E A S U R E
E T E
Hint!
Low voltage measurement can be used for non-standard thermo-
couple measurement. You will see the measured temperature in
millivolts and need a table to convert the measured millivolt value to
corresponding temperature values. In this case, use copper exten-
sion cords to connect the non-standard thermocouple to MC3s ter-
minals.
34
Measuring
T / C IN T . R J 2 -w x m tr
T / C , L o w V R , R T D
4 -w m e a s
V , 1, + 2 4 V 1 m e a s / s in k
T / C O R E X T
W IR E S O N L Y
3 & 4 -w m e a s V , ,
M E A S U R E & S IM U L A T E O U T P U T M E A S U R E
E T E
Warning!
Do not apply voltages higher than 50 V between any of MC3s
terminals.
Next
Voltage Generation on page 49
Thermocouple Measurement (Temperature) on page 43
Calibration, see Part D.
35
Startup and Basic Operation
Resistance Measurement
Resistance measurement is possible with models MC3MF and MC3TE. Resistance mea-
surement terminals are located in the ET section.
Required settings Options/description
Quantity Resistance
Funct/Port ET: R(meas)
The following picture displays the active The two leftmost terminals are used in 2-
terminals: wire systems. MC3 automatically checks
the connection and displays the found wir-
T / C IN T . R J 2 -w x m t r ing system (2-wire, 3-wire or 4-wire) in the
T / C , L o w V R , R T D
4 -w m e a s
V , 1, + 2 4 V 1 m e a s / s in k
measuring window. For more information
W
T / C O R E X T
IR E S O N L Y
concerning wiring options, see Resistance
M E A S U R E & S IM U L A T E
3 & 4 -w m e a s
O U T P U T
V , ,
M E A S U R E
and RTD Measurement, Connections on
E T E
page 72.
Note.
If the measured resistance value is infinite or very high (> 4000 ohm),
the text +OVER is displayed in the measuring window. This means
that the circuit is broken or the connection is wrong. Wrong connec-
tion may also cause erroneous reading, typically too low. If neces-
sary, use the 2-wire ohm measurement to check the wiring before
final connection.
Next
RTD and Resistance Simulation on page 53
RTD Measurement (Temperature) on page 42
Calibration, see Part D.
36
Measuring
Also check the Sound setting. The option shows voltages above approx. +1.5 V as
Change means that MC3 beeps every time open contact and voltages below approx.
the switch changes its state. When using +1.5 V as closed contact.
options Open or Closed the sound is con-
tinuously on when the switch is open re-
spectively closed. T / C IN T . R J 2 -w x m tr
4 -w m e a s
V , 1, + 2 4 V 1
T / C O R E X T
3 & 4 -w m e a s V , ,
M E A S U R E & S IM U L A T E O U T P U T M E A S U R E
Hint!
Switch state detection may also be used for binary signal detection.
Next
Limit Switch Test on page 38
Calibration, see Part D.
37
Startup and Basic Operation
Notes.
This limit switch test is a restricted version compared to the limit
switch test available in Calibration Mode. This test gives you the
approximates of the latest opening and closing points and also the
current switch status. The limit switch test in Calibration Mode of-
fers more accurate results and additionally, statistical data of re-
peated switch tests. More of the limit switch test in Calibration Mode
in Part D of this manual.
MC3 does not support limit switch testing when simulating resis-
tance or an RTD sensor.
38
Measuring
Note.
The accuracy of the Opened @ and Closed @ values depend
greatly on the change rate of the switchs input signal.
Hint.
If you use MC3 to generate/simulate the switchs input signal, you
can use ramping to create changing input signals for the switch
under test.
39
Startup and Basic Operation
Frequency Measurement
Frequency measurement is possible with all MC3 models.
The frequency measurement terminals are located in the E section.
Required settings Options/description
Quantity Freq.
Funct/Port E: f(meas)
Next
Switch State Sensing on page 37
Pulse Counting on page 41
Frequency Generation on page 51
Calibration, see Part D.
40
Measuring
Pulse Counting
Pulse counting is possible with all MC3 models. The pulse counter terminals are located
in the E section.
Required settings Options/description
Quantity Pulses
Funct/Port E: Pls(count)
counting pulses.
3 & 4 -w m e a s V , ,
M E A S U R E & S IM U L A T E O U T P U T M E A S U R E
E T E
Next
Frequency Measurement on page 40
Switch State Sensing on page 37
Pulse Generation on page 52
Calibration, see Part D.
41
Startup and Basic Operation
The following picture displays the active The two leftmost terminals are used in 2-
terminals: wire systems. MC3 automatically checks
the connection and displays the found wir-
T / C IN T . R J 2 -w x m t r
ing system (2-wire, 3-wire or 4-wire) in the
4 -w m e a s + 2 4 V 1
measuring window. For more information
V , 1,
T / C O R E X T
3 & 4 -w m e a s V , ,
page 72.
M E A S U R E & S IM U L A T E O U T P U T M E A S U R E
E T E
Note.
If the measured resistance value is infinite or very high (> 4000
ohm), the text +OVER is displayed in the measuring window. This
means that the circuit is broken or the connection is wrong. Wrong
connection may also cause erroneous reading, typically too low. If
necessary, use the 2-wire ohm measurement to check the wiring
before final connection.
Next
RTD and Resistance Simulation on page 53
Thermocouple Measurement (Temperature) on page 43
Resistance Measurement on page 36
Calibration, see Part D.
42
Measuring
T / C IN T . R J 2 -w x m t r T / C IN T . R J 2 -w x m tr
4 -w m e a s + 2 4 V 1 4 -w m e a s
V , 1, + 2 4 V 1
T / C , L o w V R , R T D V , 1, m e a s / s in k T / C , L o w V R , R T D m e a s / s in k
T / C O R E X T T / C O R E X T
W IR E S O N L Y W IR E S O N L Y
3 & 4 -w m e a s 3 & 4 -w m e a s V , ,
V , ,
M E A S U R E & S IM U L A T E O U T P U T M E A S U R E M E A S U R E & S IM U L A T E O U T P U T M E A S U R E
E T E E T E
43
Startup and Basic Operation
Warning!
If you connect an RTD sensor to the ET sections R, RTD con-
nectors, there is no galvanic isolation between the thermo-
couple and the RTD sensor.
Next
Thermocouple Simulation on page 54
RTD Measurement (Temperature) on page 42
Voltage Measurement on page 34
Calibration, see Part D.
44
Generating/Simulating
Generating/Simulating
General
MC3 is capable to perform the following generation/simulation func-
tions:
· Voltage generation
· Current generation
· Frequency and pulse generation
· Thermocouple simulation
· RTD and resistance simulation
Generation/simulation in Basic Mode require that you first select
the Window to be used (Commands: Start with D/Menu and con-
tinue either with A/Window 1 Setup or B/Window 2 Setup). Each
generation/simulation has its own unique 1/Quantity and 2/Func-
tion/Port settings in their windows menu. The other window menu
settings (generation/simulation unit etc.) refine the generation/simu-
lation characteristics.
Because all generations/simulations are not available in all MC3
models, the presentation of each generation/simulation include a
description of MC3 models supporting this generation/simulation.
The required section for the generation/simulation is also mentioned.
If the section is not included in your MC3, the 1/Quantity and
2/Function/Port settings needed for the generation/simulation are
not available as choices in the pop-up lists.
Each generation/simulation also has at least one picture with a circle
around some of MC3s terminals, like the one below.
T / C IN T . R J 2 -w x m t r
T / C , L o w V R , R T D
4 -w m e a s
V , 1, + 2 4 V 1 m e a s / s in k
T / C O R E X T
W IR E S O N L Y
3 & 4 -w m e a s V , ,
M E A S U R E & S IM U L A T E O U T P U T M E A S U R E
E T E
45
Startup and Basic Operation
Warnings!
When selecting a generation/simulation function, MC3 always
starts with zero output (not when simulating a resistance). This
is done to prevent damages in the connected instrument.
Be careful when increasing the generated/simulated signal. If
an instrument that cannot withstand the generated signal is
connected to MC3, the instrument may get seriously damaged.
Note.
The dual function of the keys: , and is not available in a
generation field. The keys only represent numbers.
46
Generating/Simulating
Current Generation
Current generation is possible with all MC3 models. MC3 can generate current both with
the E and the ET section. Additionally the E section can be used both in source and sink
mode.
Required settings Options/description
Quantity Current
Funct/Port ET: I(gen) or E: I(control)
3 & 4 -w m e a s V , ,
M E A S U R E & S IM U L A T E O U T P U T M E A S U R E
E T E
Notes.
If you are simultaneously using the ET sections output terminals
and some of the ET sections measuring terminals, there is no gal-
vanic isolation between the measuring circuit and the current gen-
eration.
The voltage between the output connectors of the ET section is
limited to ±15 Volts. There is no sinking possibility in the ET section.
T / C O R E X T
W IR E S O N L Y
3 & 4 -w m e a s V , ,
M E A S U R E & S IM U L A T E O U T P U T M E A S U R E
E T E
T / C IN T . R J 2 -w x m tr
T / C , L o w V R , R T D
4 -w m e a s
V , 1, + 2 4 V 1 m e a s / s in k
T / C O R E X T
W IR E S O N L Y
3 & 4 -w m e a s V , ,
M E A S U R E & S IM U L A T E O U T P U T M E A S U R E
E T E
Based on the terminals in use, MC3 automatically selects either source or sink mode.
47
Startup and Basic Operation
Notes applying to both the ET: I(gen) and the E: I(control) ports:
Check the polarity of your connections. The arrows in the previous
pictures describe the correct flow of current.
If the loop resistance of the external circuit is too high or infinite, the
message O.LOAD is displayed until the load is reasonable.
Warnings!
Make sure that you dont exceed the maximum current allowed
by the instrument under test.
If you open the mA generation loop, MC3 tries to maintain the
current by increasing the output voltage. If you then close the
loop again, the current is first too high, but returns quickly to
the correct level. If this current peak could damage the compo-
nents of the loop, make sure that the loop will not open or pro-
tect it against overcurrent. For the same reason, always enter
0 mA output before connecting the loop.
Next
Current Measurement on page 33
Special Generations on page 56
Calibration, see Part D.
48
Generating/Simulating
Voltage Generation
Voltage generation is possible with models MC3MF and MC3TE. The ET section has termi-
nals for voltage generation within ±12 V range and terminals for low voltage sensor simu-
lation with the range ±500 mV. The low voltage terminals are also used when measuring/
simulating thermocouples using an external Reference Junction.
Required settings Options/description
Quantity Voltage
Funct/Port ET: V(gen) or ET: LowV(sim)
T / C IN T . R J 2 -w x m t r T / C IN T . R J 2 -w x m tr
4 -w m e a s + 2 4 V 1 4 -w m e a s
V , 1, + 2 4 V 1
T / C , L o w V R , R T D V , 1, m e a s / s in k T / C , L o w V R , R T D m e a s / s in k
T / C O R E X T T / C O R E X T
W IR E S O N L Y W IR E S O N L Y
3 & 4 -w m e a s 3 & 4 -w m e a s V , ,
V , ,
M E A S U R E & S IM U L A T E O U T P U T M E A S U R E M E A S U R E & S IM U L A T E O U T P U T M E A S U R E
E T E E T E
internally measured value for the generated MC3 displays both the set value and the
voltage in the selected window. Max. out- internally measured value for the generated
put current depends on the voltage level voltage in the selected window.
as follows:
10 mA if the Voltage is within ±10 V or
3 mA if the Voltage is outside of ±10 V.
49
Startup and Basic Operation
Note.
If the resistance of the external circuit in voltage generation is very
low (obvious short circuit), the message O.LOAD is displayed until
the load is reasonable.
Hint!
Low voltage generation can be used when calibrating instruments
with millivolt input. You can use it also to simulate special thermo-
couples not defined in MC3. If you simulate a thermocouple in mil-
livolt mode, you must convert from temperature to millivolts manu-
ally.
Warning:
If you short circuit the voltage output, MC3 tries to maintain the
voltage by increasing the output current. If you then remove
the short circuit, the voltage is first too high, but returns quickly
to the correct level. If this voltage peak could damage the com-
ponents of the circuit, make sure that the circuit can not be
shorted or protect the circuit against overvoltage. For the same
reason, always enter 0 V output before connecting the circuit.
Next
Voltage Measurement on page 34
Thermocouple Simulation on page 54
Special Generations on page 56
Calibration, see Part D.
50
Generating/Simulating
Frequency Generation
Frequency generation is possible with models MC3MF and MC3TE. The frequency genera-
tion terminals are located in the ET section.
Required settings Options/description
Quantity Freq.
Funct/Port ET: f(gen)
3 & 4 -w m e a s V , ,
Next
Pulse Generation on page 52
Frequency Measurement on page 40
Special Generations on page 56
Calibration, see Part D.
51
Startup and Basic Operation
Pulse Generation
Pulse generation is possible with models MC3MF and MC3TE. The pulse generation termi-
nals are located in the ET section. Pulse generation is similar to square wave frequency
output, but it generates a predefined count of all positive or symmetric pulses.
Required settings Options/description
Quantity Pulses
Funct/Port ET: Pls(gen)
3 & 4 -w m e a s V , ,
to be used. M E A S U R E & S IM U L A T E O U T P U T
E T
M E A S U R E
E
Note.
If you change the frequency after the pulse generation has started,
the frequency is adjusted during the current cycle, but the frequency
is not correct until the next cycle.
Next
Frequency Generation on page 51
Pulse Counting on page 41
Special Generations on page 56.
52
Generating/Simulating
3 & 4 -w m e a s V , ,
Next
Thermocouple Measurement (Temperature) on page 43
RTD Measurement (Temperature) on page 42
Resistance Measurement on page 36
Special Generations on page 56
Calibration, see Part D.
53
Startup and Basic Operation
Thermocouple Simulation
MC3 is able to mimic a thermocouple at given temperatures. This property is called ther-
mocouple simulation. It enables checking and calibration of temperature indicators, re-
corders, transmitters and other equipment related to temperature measurement with a
thermocouple probe.
In thermocouple simulation the original thermocouple is disconnected from the instru-
ment under test and replaced with MC3. To the instrument under test MC3 appears as a
thermocouple at the given temperature.
Voltage measurement is possible with models MC3MF and MC3TE. Thermocouple simula-
tion terminals are located in the ET section.
Required settings Options/description
Quantity T/C-Temp
Funct/Port ET: TCi(sim) (for the internal Reference Junction), or
ET: TCx(sim) (for other RJ compensation methods)
Sensor Type Available thermocouple types
Reference Junction Depends on the Function/Port setting.
See subsequent chapters.
T / C IN T . R J 2 -w x m t r T / C IN T . R J 2 -w x m tr
4 -w m e a s + 2 4 V 1 4 -w m e a s
V , 1, + 2 4 V 1
T / C , L o w V R , R T D V , 1, m e a s / s in k T / C , L o w V R , R T D m e a s / s in k
T / C O R E X T T / C O R E X T
W IR E S O N L Y W IR E S O N L Y
3 & 4 -w m e a s 3 & 4 -w m e a s V , ,
V , ,
M E A S U R E & S IM U L A T E O U T P U T M E A S U R E M E A S U R E & S IM U L A T E O U T P U T M E A S U R E
E T E E T E
54
Generating/Simulating
Notes.
If the resistance of the external circuit in thermocouple simulation is
very low (obvious short circuit), the message O.LOAD is displayed
until the load is reasonable.
Check that the instrument under test is set for the same thermo-
couple type as selected in MC3. If the type is not one of the avail-
able types, you should simulate in millivolt mode and convert from
temperature to millivolts manually. More information concerning
millivolt generation is in chapter Voltage Generation on page 49.
Wire-wound thermocouple temperature indicators and record-
ers are calibrated with known loop resistance, commonly ten or
twenty ohms. The required resistance of the conductors is usually
marked on the device. When calibrating such a device, tune the
loop resistance to the nominal value with an additional resistor.
Use MC3s 2-wire resistance measurement to tune the loop resis-
tance of a wire wound device. Short circuit the measuring device
for the time of resistance measurement. The resistance measure-
ment may damage the device, if you do not install the short circuit.
For more information on resistance measurement, see chapter
Resistance Measurement on page 36.
Warning!
If you short circuit the thermocouple simulation output, MC3
tries to maintain the voltage by increasing the output current.
If you then open the output again, the voltage is first too high,
but returns quickly to the correct level. If this voltage peak could
damage the components of the circuit, make sure that the cir-
cuit can not be shorted or protect the circuit against overvolt-
age. For the same reason, always enter 0°C output before con-
necting the circuit.
If you connect an RTD sensor to the ET sections R, RTD con-
nectors, there is no galvanic isolation between the thermo-
couple and the RTD sensor.
Next
Thermocouple Measurement (Temperature) on page 43
RTD and Resistance Simulation on page 53
Special Generations on page 56
Calibration, see Part D.
55
Startup and Basic Operation
Special Generations
Manual change of generated signals is sometimes a restriction.
Therefore MC3 includes Stepping and Ramping as standard utili-
ties for creating automatically alternating output signals (special
generations).
To be able to start the definition of a special generation, a suitable
generation/simulation field has to be active in the Basic Mode Win-
dow. If a generation/simulation is not started, select D/Menu and
either A/Window 1 Setup or B/Window 2 Setup to configure a
generation/simulation. For more information on how to configure a
generation/simulation, see chapter Generating/Simulating on page
45 and its subchapters.
Notes.
Only one of the special generations may be active at a given time.
Special generations are available only in Basic Mode.
56
Special Generations
Stepping
Open the window setup menu for the gen- Configure the Stepping settings according
eration/simulation quantity that you want to to your needs. The following pictures are
configure a Stepping function for (D/Menu examples of stepping configurations. The
and either A/Window 1 Setup or B/Win- graph below the configuration window dis-
dow 2 Setup ). Choose 3/Step, Ramp and plays the result of the settings.
1/Step in the next menu.
3 0 .0 1 .2 0 0 1 1 3 :3 5 3 0 .0 1 .2 0 0 1 1 3 :3 6
S T E P P IN G S T E P P IN G
S te p p in g M o d e U p S te p p in g M o d e U p /D o w n
S te p T im e 1 .0 0 s S te p T im e 0 .5 0 s
R e p e a ts R e p e a ts
0 = c o n tin u o u s 3 0 = c o n tin u o u s 0
S ta r tin g P o in t F ro m p re s e n t S ta r tin g P o in t F ro m 0 %
S te p S iz e 1 .0 0 m A S te p S iz e 1 .5 0 m A
N u m b e r o f S te p s 4 N u m b e r o f S te p s -------
D iv is io n o f S te p s L in e a r D iv is io n o f S te p s L in e a r
R a n g e 0 % 8 .0 0 0 0 R a n g e 0 % 8 .0 0 0 0
1 0 0 % 1 2 .0 0 0 0 m A 1 0 0 % 1 2 .0 0 0 0 m A
C a n c e l F ie ld S E t d o i pt S ta rt C a n c e l F ie ld S E t d o i pt S ta rt
R a m p in g R a m p in g
C u rre n t, m A C u rre n t, m A
1 2 1 2
1 1 1 1
1 0 1 0
9 9
8 8
-1 0 1 5 1 0 1 4 T im e , s -1 0 1 5 1 0 1 4 T im e , s
S ta rt S ta rt
Note that because of the Starting point ger value for the Number of Steps field,
setting, the first repeat of the leftmost ex- MC3 displays a dashed line instead of a
ample is limited to a small starting step and value for the Number of Steps field. This
only two full steps. Generally: if the step- indicates that the size of the step that
ping settings limit the first repeat to only a reaches the 100 % range limit is smaller
fraction of a full repeat it is still calculated than the defined step size. See the
as one repeat. rightmost example.
You may enter either the Step Size or the To stop stepping press the function key C/
Number of Steps. Because these values Stop Stepping. This applies for both a
depend on each other, the value of the other continuous stepping (Repeats field set to
field is calculated whenever you make any zero) and a stepping with a predetermined
changes. Also, you can only enter an inte- amount of repeats (max. value 65535).
ger as the Number of Steps value. If you
enter a step size that results in a non-inte-
57
Startup and Basic Operation
Warning!
Do not configure the range settings so that they exceed the
allowed input range of the connected instrument. MC3 deter-
mines the limits of the range settings based on the selected
quantity and port, not the connected instrument.
Ramping
Open the window setup menu for the gen- Configure the ramping settings according
eration/simulation quantity that you want to to your needs. The following pictures are
configure a Ramping function for (D/Menu examples of ramping configurations. The
and either A/Window 1 Setup or B/Win- graphs presented after the configuration
dow 2 Setup ). Choose 3/Step, Ramp and windows display the result of the settings.
2/Ramp in the next menu.
3 0 .0 1 .2 0 0 1 1 3 :3 8 3 0 .0 1 .2 0 0 1 1 3 :3 9
R A M P IN G R A M P IN G
W a it in 0 % 2 s W a it in 0 % 0 s
R is e T im e 1 s R is e T im e 5 s
W a it in 1 0 0 % 1 s W a it in 1 0 0 % 0 s
F a ll T im e 2 s F a ll T im e 0 s
R e p e a ts R e p e a ts
0 = c o n tin u o u s 2 0 = c o n tin u o u s 0
R a n g e 0 % 8 .0 0 0 0 R a n g e 0 % 8 .0 0 0 0
1 0 0 % 1 6 .0 0 0 0 m A 1 0 0 % 1 6 .0 0 0 0 m A
C a n c e l F ie ld S E t d o i pt S ta rt C a n c e l F ie ld S E t d o i pt S ta rt
R a m p in g R a m p in g
C u rre n t, m A C u rre n t, m A
1 6 1 6
1 4 1 4
1 2 1 2
1 0 1 0
8 8
-1 0 1 5 1 0 1 4 T im e , s -1 0 1 5 1 0 1 4 T im e , s
S ta rt S ta rt
To stop ramping press the function key zero) and a ramping with a predetermined
C/Stop Ramping. This applies for both a amount of repeats (max. value 65535).
continuous ramping (Repeats field set to
58
Special Generations
Note.
MC3 actually makes the ramp in small steps. The steps are as small
as possible, slower ramps use smaller steps.
Warning!
Do not configure the range settings so that they exceed the
allowed input range of the connected instrument. MC3 deter-
mines the limits of the range settings based on the selected
quantity and port, not the connected instrument.
59
Startup and Basic Operation
(Empty)
60
Configurations and
Additional Information
Settings
The settings window has two pages. To open the settings window
starting from Basic Mode, press D/Menu, C/Others and 1/Set-
tings. The leftmost picture is the settings page that opens first.
2 9 .0 1 .2 0 0 1 1 0 :2 9 2 9 .0 1 .2 0 0 1 1 0 :3 0
S E T T IN G S S E T T IN G S
D a te F o rm a t d d .m m .y y y y N e t F re q u e n c y 5 0 H z
T im e F o r m a t h :m m 2 4 h B a c k lig h t P o w e r N o rm a l
L a n g u a g e E n g lis h
T e m p e r a tu r e U n it °C
A u to - o ff D e la y s ( 0 = n e v e r ) [m in ] T e m p e r a tu r e S c a le IT S 9 0
C a lib r a to r 0
B a c k lig h t 5
S o u n d V o lu m e s
K e y C lic k L o w
N e x t N e x t
C lo s e E d it C lo s e E d it
P a g e P a g e
You can scroll between the pages using the B/Next Page Function
Key.
Note.
All changes in settings become valid immediately.
62
Configuring the Calibrator
63
Advanced Operation and Configuration
2 9 .0 1 .2 0 0 1 1 0 :3 2
T IM E /D A T E S E T T IN G S
D a te (d d .m m .y y y y ) 2 9 0 1 2 0 0 1
T im e ( h h :m m :s s ) 1 0 3 2 2 2
C a n c e l E d it A c c e p t
Notes.
The date must always be given in dd.mm.yyyy format no matter
what the configured date format is.
The time must always be given in the 24 hour format no matter
what the configured time format is.
The date and time will be updated when you press the D/Accept
Function Key.
64
Additional Information
Additional Information
Most of the measurements, generations and simulations presented
in this manual are straight forward: Just make the required window
settings and connect the instrument under test in the active termi-
nals and thats it!
Some cases require additional settings and things to check before
you can be certain that the measurement, generation or simulation
works as expected. A typical example of this is temperature mea-
surement using a thermocouple. It is not enough to select the cor-
rect quantity and port in MC3. The Sensor type and the Reference
Junction Mode have to be correct too. Wrong settings give errone-
ous results.
Whenever this additional information may be of use in a measure-
ment, generation or simulation, text describing the function refers
to one of the following chapters. An experienced user need not jump
to read this supplemental text, but for a beginner we highly recom-
mend it.
65
Advanced Operation and Configuration
General
MC3 measures pressure with its internal gauge pressure input, if
present. Additionally, if the barometric option is present, the
measurement results of the gauge pressure input can also be shown
as absolute pressure.
Pressure Type
Every pressure measurement is in fact a pressure difference mea-
surement: a given pressure value is compared against a reference
pressure. In certain cases the reference pressure has a special
meaning (like the atmospheric pressure) and the pressure mea-
surements compared against these points are given a special name:
66
Additional Information
Square Rooting
Pressure transmitters or converters used for flow measurement of-
ten have square rooting character. To enable correct percentage
error calculation of square rooting transmitters, select square root-
ing as the Transfer Function.
Square rooting is required when entering instrument data for a pres-
sure instrument (see Part D, Calibration). Standard pressure mea-
surement in Basic Mode does not require any Transfer Function
information.
67
Advanced Operation and Configuration
o r a T / C O R E X T
W IR E S O N L Y
3 & 4 -w m e a s
T /C s ig n a l r e c e iv e r M E A S U R E & S IM U L A T E
Note.
The range of the Internal Reference Junctions temperature com-
pensation is -10 +50°C (14 122 °F).
See also
External Reference Junction on page 69
68
Additional Information
T /C s e n s o r T / C IN T . R J Entered:
o r a T / C , L o w V R , R T D
4 -w m e a s
To be used when:
T /C s ig n a l r e c e iv e r W
T / C O R E X T
IR E S O N L Y
3 & 4 -w m e a s
· The Reference Junction temperature is
M E A S U R E & S IM U L A T E
fixed to a temperature that differs from
0°C.
· The Reference Junction temperature is
controlled and the controllers set point
is not 0°C. Note: this method is of use
Warning! only if the accuracy of the controller is
If you connect an RTD sensor to the ET better than MC3s own Reference Junc-
inputs R, RTD connectors, there is no tion.
galvanic isolation between the thermo- · A compensation box is used and the
couple and the RTD sensor. Reference Junction temperature setting
is other than 0°C.
· There is no Reference Junction compen-
sation in the measuring/simulation cir-
cuit, but you want to manually enter an-
other Reference Junction temperature
than 0°C.
69
Advanced Operation and Configuration
T e m p e ra tu re
T re f T /C m a te r ia ls
C o n tr o lle r
t
C o p p e r
C o p p e r
T /C m a te r ia ls
T /C s e n s o r T / C IN T . R J
T /C s e n s o r T / C IN T . R J
o r a T / C , L o w V R , R T D
4 -w m e a s
o r a T / C , L o w V R , R T D
4 -w m e a s
T /C s ig n a l r e c e iv e r W
T / C O R E X T
IR E S O N L Y T /C s ig n a l r e c e iv e r W
T / C O R E X T
IR E S O N L Y
3 & 4 -w m e a s 3 & 4 -w m e a s
M E A S U R E & S IM U L A T E M E A S U R E & S IM U L A T E
T /C
m a te r ia ls C o p p e r
T re f
T /C s e n s o r T / C IN T . R J
o r a T / C , L o w V R , R T D
4 -w m e a s
T /C s ig n a l r e c e iv e r W
T / C O R E X T
IR E S O N L Y
3 & 4 -w m e a s
M E A S U R E & S IM U L A T E
See also
Internal Reference Junction on page 68
70
Additional Information
Error situations
The easiest way to avoid errors in thermocouple measurement and
simulation is to check carefully the used wiring and the Reference
Junction mode. The following table describes the typical error situ-
ations and possible causes/corrections when working with thermo-
couples:
PROBLEM CAUSE
MC3 (or the instrument • The thermocouple type selected in
under test when MC3 does not correspond with the
simulating used thermocouple.
thermovoltages) • The selected Reference Junction
measures the mode is not in accordance with the
temperature/millivolt used wiring.
signal, but the displayed
• Extension or compensation cable type
temperature reading is all
or connections are incorrect.
wrong.
(The error may vary from • The polarity of the cables are
0 to about ±50°C incorrect.
depending on the type of
the error)
MC3 (or the instrument • Incorrect connections.
under test) displays • The wiring is broken.
random readings during
• Interference from a mobile phone or a
thermocouple
radio transmitter affects the
measurement.
measurement.
MC3 displays unstable • The instrument under calibration uses
readings during voltage pulses to detect open sensor.
thermocouple simulation. When MC3 detects these pulses, it
tries to compensate for them, which
causes the unstable output. Prevent
these pulses for the time of
calibration. Refer to the service
manual of the instrument under
calibration on how to prevent these
open sensor detection pulses.
71
Advanced Operation and Configuration
T / C IN T . R J 2 -w x m tr T / C IN T . R J 2 -w x m tr
T / C , L o w V R , R T D
4 -w m e a s
V , 1, + 2 4 V 1 m e a s / s in k T / C , L o w V R , R T D
4 -w m e a s
V , 1, + 2 4 V 1 m e a s / s in k
T / C O R E X T T / C O R E X T
W IR E S O N L Y W IR E S O N L Y
3 & 4 -w m e a s V , , 3 & 4 -w m e a s V , ,
M E A S U R E & S IM U L A T E O U T P U T M E A S U R E M E A S U R E & S IM U L A T E O U T P U T M E A S U R E
E T E E T E
72
Additional Information
T / C IN T . R J 2 -w x m tr T / C IN T . R J 2 -w x m tr
4 -w m e a s + 2 4 V 1 4 -w m e a s
V , 1, + 2 4 V 1
T / C , L o w V R , R T D V , 1, m e a s / s in k T / C , L o w V R , R T D m e a s / s in k
T / C O R E X T T / C O R E X T
W IR E S O N L Y W IR E S O N L Y
3 & 4 -w m e a s V , , 3 & 4 -w m e a s V , ,
M E A S U R E & S IM U L A T E O U T P U T M E A S U R E M E A S U R E & S IM U L A T E O U T P U T M E A S U R E
E T E E T E
T / C O R E X T
W IR E S O N L Y
3 & 4 -w m e a s V , ,
M E A S U R E & S IM U L A T E O U T P U T M E A S U R E
E T E
Note.
In higher temperatures: The leakage of the diode in the instrument
may affect the accuracy.
73
Advanced Operation and Configuration
T / C O R E X T
E T s e c tio n
3 & 4 -w m e a s V , ,
M E A S U R E & S IM U L A T E O U T P U T M E A S U R E
Warning!
There is no galvanic isolation between the ET sections mea-
suring terminals and generating terminals.
74
Calibration
General
MC3 is a stand-alone calibrator as well as a calibrator communicat-
ing with calibration software.
MC3 supports stand-alone calibration and off-line calibration.
The following list briefly describes these methods:
· In stand-alone calibration, all instrument and calibration result
data is stored in MC3s memory. No external calibration data-
base is used.
· In off-line calibration (sometimes also called batch calibration),
the instrument data is downloaded from a calibration software.
The calibration is done just as in stand-alone calibration, but
calibration procedures are downloaded from a software and
the saved results are uploaded to the calibration software.
Note that off-line calibration requires the QCAL option. It also re-
quires the computer communication cable to connect MC3 to a PCs
serial port.
Note.
MC3 stores one As Found calibration and one As Left calibration
per instrument. Additional As Found or As Left calibrations for the
same instrument overwrite the older results. Use a calibration soft-
ware to transfer existing calibration results to the calibration
softwares database before performing more calibrations.
76
General
S ta rt
R e a d S ta r tin g
G u id e te x t
A s F o u n d
c a lib r a tio n
S a v e A s
F o u n d r e s u lts
n o A d ju s tm e n t y e s
r e q u ir e d ?
R e a d A d ju s tin g
G u id e te x t
A d ju s tm e n t
n o * W ith in y e s
lim its ?
* If it is n o t p o s s ib le
to a d ju s t th e
m a x im u m e r r o r
b e lo w th e " R e je c t if " A s L e ft
lim it, c o n s id e r c a lib r a tio n
r e p la c in g th e
in s tr u m e n t w ith S a v e A s L e ft
a m o re a c c u ra te o n e . r e s u lts
R e a d F in is h in g
G u id e te x t
E n d
77
Calibration
As Found Calibration
The As Found calibration documents the state of the instrument
before carrying out any adjustments. The calibration states the
amount of drift in the instrument during the calibration period.
E rro r
0 2 5 5 0 7 5 1 0 0 %
Adjustment
MC3 calculates (among other figures) the maximum error found
during the calibration. Depending on the value of the maximum er-
ror, you decide whether you adjust the instrument or not. MC3 sup-
ports four different error limit settings:
Reject if > acceptance limit for found maximum error,
Adjust if > the instrument need to be adjusted if this
limit is exceeded,
Do not adjust if < adjusting the instrument is not
necessary / of use and
Adjust to < after adjusting, the maximum error should
not exceed this limit.
MC3 has a special adjustment utility tool that can be used when the
input signal of the instrument is generated/simulated with MC3. See
chapter MC3s Support for Instrument Adjustment on page 96.
We recommend you use the utility whenever possible. In all other
situations, use the error graph and its error value to have a view on
how the adjustment affects the accuracy of the instrument and how
close the output signal is from the target value.
Use the instruments ZERO adjustment at lower end of the range.
Adjust the higher end of the range with the instruments SPAN ad-
justment. The two adjustment points do not necessarily need to be
the end points of the instrument range. You may want to use other
78
General
E rro r
S P A N
0
Z E R O
0 2 5 5 0 7 5 1 0 0 %
As Left Calibration
The As Left calibration documents the state of the instrument after
the possible adjustment.
MC3 does not require you to save an As Left calibration. If the As
Found calibration proves to be good and no adjustment is needed,
you may omit the As Left calibration.
E rro r E rro r
0 0
0 2 5 5 0 7 5 1 0 0 % 0 2 5 5 0 7 5 1 0 0 %
79
Calibration
Table for choosing the input section Table for choosing the output section
QUANTITY METHOD REQ. SECTION QUANTITY METHOD REQ. SECTION
Voltage Measured E Voltage Measured E
Generated ET Keyed none
Keyed none Low Voltage ET Measured ET
(
Low Voltage ET measured ET Keyed none *
Generated ET Current Measured E
(
Keyed none * Keyed none
Current Measured E Resistance Measured ET
(
E generated E Keyed none *
ET generated ET Pressure Measured Press.
(
Keyed none Keyed none **
Resistance Measured ET Temp. Keyed none
Simulated ET TC temperature Measured ET
( (
Keyed none * Keyed none *
Pressure Measured Press. RTD temperature Measured ET
( (
Keyed none ** Keyed none *
Temperature Keyed none Frequency Measured E
TC temperature Measured ET Keyed none
Simulated ET Value Keyed none
RTD temperature Measured ET Switch Measured E
Simulated ET
Frequency Measured E
Generated ET
Value Keyed none
80
General
The general principle is: The same section may not be used for
both the input and the output signal. There are however some ex-
ceptions to that rule:
If when checking the required sections both the input and the out-
put section is ET, but the cells are grayed, check the table below. It
presents all supported combinations where both the input signal
and the output signal is connected to the ET section.
81
Calibration
Calibrating an Instrument
Note.
If there are a lot instrument data in MC3s memory, displaying the
list may take a few seconds.
82
Calibrating an Instrument
83
Calibration
O u tp u t C u rre n t [E : M e a s ]
O u tp u t
S ig n a l 3 .9 7 6 2 m A
W in d o w
E rro r
1 .0 0 -0 .1 5 % o f s p a n
E rro r +
G ra p h
0
-
0 % 5 0 % 1 0 0 %
B a c k S ta rt A d ju s t M E N U
Note.
84
Calibrating an Instrument
Keys B to D.
3 0 .0 1 .2 0 0 1 1 3 :5 8
Notes. E N V IR O N M E N T A L D A T A
D e v ic e T e m p e ra tu r e 2 1 .0 °C
Some of the temperature data may E n v ir o n m e n t T e m p . 2 1 .0 °C
3 1 .6
be pre-entered by MC3, provided
In p u t M o d u le T e m p . °C
O u tp u t M o d u le T e m p . 2 9 .6 °C
bration). N o te s :
J u s t T e s tin g
See also:
MC3s Support for Instrument Adjustment on page 96.
Viewing Calibration Results on page 102.
85
Calibration
86
Calibrating an Instrument
87
Calibration
Pressure Transmitters
This procedure suits all instruments with a pressure input and any
kind of electrical output signal. Both the input and output signal
range should be measurable with MC3. Pressure transmitter cali-
bration can be done with models MC3MF and MC3PE.
Required sections
· A pressure input (PRxxx).
· The E section for measuring the electrical output signal.
Preparations
1. Connect the input signal of the instru- P
ment to the pressure input.
E
2. Connect the pressure source both to the P re s s u re
instruments input and to the pressure S o u rc e
input.
3. Connect the instruments output signal
to the chosen electrical input port in
MC3.
4. Test the connections in Basic Mode if
needed. To quickly configure the Basic T / C IN T . R J 2 -w x m t r
T / C O R E X T
W IR E S O N L Y
3 & 4 -w m e a s V , ,
0 . 4 9 8 2
1 1 : 4 9
[ P 1 : IN
G
T 6 C ]
a u g e
b a r
8 . 0 3 1 0 m A
Mode.
E r r o r
1 . 0 0 0 .2 8 % o f s p a n
+
-
0 % 5 0 % 1 0 0 %
A c c e p t
P a u s e
R e a d i n g s
M U L T I F U N C T I O N C A L I B R A T O R
88
Calibrating an Instrument
Calibration
1. Move to Calibration Mode, select the 4. Save or reject the results as is de-
instrument to be calibrated and start the scribed in chapter A Calibration Pro-
calibration as presented in chapter A cedure Using MC3 on page 84.
Calibration Procedure Using MC3 on 5. Do the possible adjustment and the As
page 84. Found calibration or return to the Instru-
2. Set the pressure using, e.g. a pressure ment window.
pump.
3. Accept the points using the B/Accept
Readings Function Key.
Warning!
Working with pressure instruments can be hazardous. Only
qualified personnel may use pressure instruments and pres-
sure sources.
Next
MC3s Support for Instrument Adjustment on page 96
89
Calibration
Temperature Sensors
This procedure suits temperature sensors no matter if they are RTDs
or thermocouples. The sensors output signal is measured with MC3
or it is keyed into MC3. The reference temperature is either mea-
sured with MC3 or entered to MC3. Temperature sensor calibration
can be done with models MC3MF and MC3TE.
If you measure the reference temperature using MC3, remember
the following conditions:
· If the sensor to be calibrated is a thermocouple, the reference
thermometer needs to be an RTD sensor.
· If the sensor to be calibrated is an RTD sensor, the reference
thermometer needs to be a thermocouple.
Required section
· The ET section for measuring the sensors output signal and
possibly also for measuring the reference temperature.
Preparations
1. Place the sensor of the instrument (and 2. Test the connections in Basic Mode if
the sensor of the reference thermom- needed. To quickly configure the Basic
eter) into the calibration bath/dry block. Modes windows, go to Calibration
Connect it/them to suitable terminals in Mode, select the instrument to be cali-
MC3 (RTD or T/C). If the sensor is a brated and the Function Key C/Cali-
thermocouple, select one of the avail- brate but immediately return to Basic
able reference junction methods: Mode.
· The internal Reference Junction R e fe re n c e
is in use. th e rm o m e te r
· The reference junction is fixed to S e n s o r
0°C.
· The reference junction tempera-
ture is entered.
· The reference junction tempera- T / C IN T . R J 2 -w x m t r
T / C O R E X T
W IR E S O N L Y
O U T P U T
E T
M
V ,
E
,
E A S U R E
minals.
See Part C for additional information In p u t
0 6 .1 0 .2 0 0 0
T C
1 1 :0 6
T e m p e r a tu r e [E T :C o n tr o l]
ods.
O u tp u t
2 4 .9 4
P t1 0 0 a 3 8 5
° C (IT S 9 0 )
E r r o r
1 .0 0 -0 .0 6 % o f s p a n
+
-
0 % 5 0 % 1 0 0 %
A c c e p t
P a u s e
R e a d in g s
M U L T I F U N C T I O N C A L I B R A T O R
90
Calibrating an Instrument
Calibration
1. Move to Calibration Mode, select the 4. If the reading of the reference tempera-
instrument to be calibrated and start the ture sensor is entered manually, enter
calibration as presented in chapter A the value and press B/Accept Read-
Calibration Procedure Using MC3 on ings Function Key. If the reference tem-
page 84. perature sensor is connected to MC3,
2. Start the calibration as presented in accept the readings using the B/Accept
Readings Function Key.
chapter A Calibration Procedure Us-
ing MC3 on page 84. 5. Save or reject the results as is de-
3. Set the temperature of the bath/dry scribed in chapter A Calibration Pro-
block to the required calibration point. cedure Using MC3 on page 84.
Wait until the temperature stabilizes. 6. Do the possible adjustment and the As
Found calibration or return to the Instru-
ment window.
Hint!
With small changes, this procedure should also suit all tempera-
ture instruments with a fixed sensor. In that case, use a reference
sensor that can be connected to MC3 to measure the input signal
and either measure or enter the output signal of the instrument.
Use one of the following examples for the output signal:
· Temperature Transmitter with sensor. See the example
Pressure Transmitters on page 88.
· Temperature Indicators/Recorders with sensor. See the
example Temperature Indicators and Recorders on page
92.
· Temperature switches (thermostats) with sensor. See the
example Electrical Limit Switches on page 94.
Next
MC3s Support for Instrument Adjustment on page 96
91
Calibration
Preparations
1. Connect suitable terminals in MC3 to 4. Test the connections in Basic Mode if
the input connectors of the indicator/ needed. To quickly configure the Basic
recorder. Modes windows, go to Calibration
Mode, select the instrument to be cali-
2. If the sensor to be simulated is a ther-
brated and the Function Key C/Cali-
mocouple, select one of the following
brate but immediately return to Basic
reference junction methods:
Mode.
· The internal Reference Junction
is in use.
· The reference junction is fixed to
0°C.
· The reference junction tempera-
ture is entered.
· The reference junction tempera-
ture is measured externally using T / C IN T . R J 2 -w x m tr
T / C O R E X T
minals.
W IR E S O N L Y
3 & 4 -w m e a s V , ,
M E A S U R E & S IM U L A T E O U T P U T M E A S U R E
E T E
ods.
P t1 0 0 a 3 8 5
7 5 .0 0 ° C (IT S 9 0 )
O u tp u t V a lu e [K e y e d ]
7 4 .8 0 0 °C
nals in MC3.
M U L T I F U N C T I O N C A L I B R A T O R
92
Calibrating an Instrument
Calibration
1. Move to Calibration Mode, select the This method is useful when cali-
instrument to be calibrated and start the brating analog indicators/record-
calibration as presented in chapter A ers and the input signal supports
Calibration Procedure Using MC3 on fine adjustment.
page 84. · Select the field displaying the de-
2. MC3 simulates the output signal of the fault value and enter the actual
temperature sensor for each calibration reading. This method is useful
point. The middlemost window displays when calibrating indicators with
the default value for the reading. You digital display or when the input
have two possibilities on how to con- signal cannot be fine adjusted.
tinue:
3. Accept the points using the B/Accept
· Adjust the simulated signal until Readings Function Key.
the reading of the indicator/re- 4. Save or reject the results as is de-
corder is exactly the same as the scribed in chapter A Calibration Pro-
default value displayed in MC3. cedure Using MC3 on page 84.
5. Do the possible adjustment and the As
Found calibration or return to the Instru-
ment window.
Hint!
This procedure can be used as a reference when calibrating any
kind of indicators/recorders. Adapt the input signal connections/
settings according to the instruments input signal.
When calibrating a temperature instrument with a detachable sen-
sor (the input signal is simulated with MC3), use this example as
the source for the input signal part.
Next
MC3s Support for Instrument Adjustment on page 96
93
Calibration
Preparations Calibration
1. Connect the ET sections terminals 1. Move to Calibration Mode and select the
marked OUTPUT to the input of the instrument to be calibrated.
switch. 2. Start the calibration as is presented in
2. Connect the switch contact to the switch chapter A Calibration Procedure Us-
terminals in the E section. ing MC3 on page 84. MC3 does the
Prescan if it is enabled. During the
3. Test the connections in Basic Mode if
needed. To quickly configure the Basic prescan, MC3 searches for approxi-
Modes windows, go to Calibration mate values for the actuating and
Mode, select the instrument to be cali- deactuating point. This speeds up the
brated and the Function Key C/Cali- final test without sacrificing accuracy.
brate but immediately return to Basic
The prescan test is done only once in a
Mode. calibration.
· By default, the Prescan is set to on.
If you do not want MC3 to perform a
prescan, disable it (commands
D/MENU and 2 /Prescan ). When
Prescan is set to off, MC3 uses the
whole scan range also during the
T / C IN T . R J 2 -w x m t r actual test.
4 -w m e a s + 2 4 V 1
T / C O R E X T
W IR E S O N L Y
S E N S O R M E A S U R E & S IM U L A T E O U T P U T M E A S U R E
E T E
In p u t
0 6 .1 0 .2 0 0 0 1 0 :2 9
V o lta g e [E T : G e n .]
by decreasing the input signal until the
5 .0 2 3
T e s t in g s w itc h lim its ( 4 /5 )
V
switch deactuates. MC3s screen dis-
plays the obtained data as the test ad-
O u tp u t S w itc h [E : M e a s .]
8 .0 3 1 0 A c tu a te
-------
vances.
D e a c tu a te
7 .0 0 0
5 .5 0 0
4 .5 0 0
3 .0 0 0
P a u s e
M U L T I F U N C T I O N C A L I B R A T O R
94
Calibrating an Instrument
· If for some reason the calibration 4. Save or reject the results as is de-
cannot be done automatically (be- scribed in chapter A Calibration Pro-
cause the input signal is measured, cedure Using MC3 on page 84.
not generated/simulated), you will 5. Either do another calibration run or re-
have to change the input signal turn to the Instrument window.
manually. The slower you raise the
input signal the more accurate the
observed actuation point is. Be
equally careful when decreasing the
input signal in order to get the
deactuation point.
Notes.
Selecting Calibration Method Manual when the input signal is gen-
erated/simulated with MC3 means that each pair of actuation/
deactuation points have to be accepted manually.
If the prescan stops and displays an error message, try widening
the scan range. It is also advisable to have the scan range sym-
metrical in relation to the actuation/deactuation points.
Hint!
This procedure also suits other types of switches. How to calibrate
other types of switches may be adapted by combining the informa-
tion of this example with the input signal information from one of the
following examples:
· Pressure switches. See the example Pressure Transmit-
ters on page 88.
· Temperature switches with a fixed sensor. See the example
Temperature Sensors on page 90.
· Temperature switches with a detachable sensor. See the
example Temperature Indicators and Recorders on page
92.
Next
MC3s Support for Instrument Adjustment on page 96
95
Calibration
96
Maintaining MC3s Instrument Database
P o s itio n N a m e
Whichever way you start creating
the instrument, you will first have C a lib r a te d
B a s ic
S e le c t M E N U
M o d e
97
Calibration
to select the Input Quantity and the Output Quantity. These selec-
tions cannot be changed afterwards.
S E L E C T IN S T R U M E N T IN P U T
S E L E C T IN S T R U M E N T O U T P U T
V o lta g e
V o lta g e
L o w V o lta g e
L o w V o lta g e
C u rre n t
C u rre n t
R e s is ta n c e
R e s is ta n c e
P re s s u re
P re s s u re
T e m p e ra tu re
T e m p e ra tu re
T C T e m p e ra tu re
T C T e m p e ra tu re
R T D T e m p e ra tu re
R T D T e m p e ra tu re
F re q u e n c y
F re q u e n c y
V a lu e
V a lu e
S w itc h
Note.
Only the following fields may be edited if the instrument data is
downloaded from a calibration software (valid for Beamexs calibra-
tion software QD3 and QM6):
· Input Span and Output Span.
mum error. E r r o r C a lc . M e th o d % o f s p a n
R e je c t if > 0 .5 0
A d ju s t if > 0 .3 0
D o n o t A d ju s t if < 0 .1 0
A d ju s t to < 0 .1 0
N e x t
S a v e E d it M E N U
p a g e
98
Maintaining MC3s Instrument Database
Range
All other quantities except Value P re s s u re T y p e G a u g e
99
Calibration
100
Maintaining MC3s Instrument Database
Deleting Instruments
If you want to free some memory or otherwise do not need certain
instrument data anymore, you may remove instruments from MC3s
memory as follows:
To remove all instruments (and their calibration data), open the
Position List and press D/Menu, 6/Delete All Instruments (see
the leftmost picture below).
3 0 .0 1 .2 0 0 1 1 3 :5 0 3 0 .0 1 .2 0 0 1 1 3 :5 1
C re a te N e w I N S T R U M E N TV i e w
P O S IT IO N /D E V IC E ID 1 1 2 -T T -0 0 3 .1 In s tr u m e n t
In s tr u m e n t D e ta ils
1 0 1 -X I-0 0 1 .1 In ta k e T e m p e r a tu r e o n F e e d 1
1 1 2 -T T -0 0 3 .1 D e le te
1 1 2 -T T -0 0 7 .1 In s tr u m e n t
P T 1 0 1 .6 T r a n s fe r F u n c tio n L in e a r
V ie w
P T 1 1 2 .1 2 C a l. P o in ts 5 ¯ C a lib r a tio n
P T 1 1 5 .1 5 -1 R e s u lts
P T 1 1 5 .1 5 -2
P T 1 1 2 .1 6
T T 1 1 2 .0 8 IN P U T 0 .0 0 ... 1 0 0 .0 0 ° C
T T 1 1 2 .1 0
M e th o d S im u la te d
T T 1 1 2 .1 2
S e n s o r T y p e P t1 0 0 a 3 8 5
D e v ic e ID
T T 1 1 2 .0 7 D e le te A ll
P o s itio n N a m e In s tr u m e n ts O U T P U T 4 .0 0 0 0 ... 2 0 .0 0 0 0 m A
In ta k e T e m p e ra tu re o n F e e d 1 M e th o d M e a s u re d
C a lib r a te d C o n tr o lle r
N o t C a lib r a te d S e ttin g s
B a s ic C lo s e C lo s e
S e le c t M E N U B a c k E d it C a lib r a te M E N U
M o d e M E N U M E N U
101
Calibration
3 0 .0 1 .2 0 0 1 1 4 :3 4 3 0 .0 1 .2 0 0 1 1 4 :3 5
I N S T R U M E N TV i e w In p u t R T D T e m p e r a t u Zr e e r [ oE T : S im u l]
In s tr u m e n t P P t r1 e 0 s 0 s a u 3r e 8 5
1 1 2 -T T -0 0 3 .1
D e ta ils
In ta k e T e m p e r a tu r e o n F e e d 1 0 .0 0 ° M C o ( Id T u S l e9 0 )
D e le te P re s c a n
In s tr u m e n t [O n ]
T r a n s fe r F u n c tio n L in e a r O u tp u t C u rre n t [E : M e a s ]
V ie w
C a l. P o in ts 5 ¯ C a lib r a tio n
R e s u lts
3 .9 7 6 2 m A
V ie w
C a lib r a tio n
E rro r
IN P U T 0 .0 0 ... 1 0 0 .0 0 ° C 1 .0 0 - 0 . 1 5 R% e o s fu sl t p s a n
M e th o d S im u la te d +
S e n s o r T y p e P t1 0 0 a 3 8 5
0
O u tp u t 4 .0 0 0 0 ... 2 0 .0 0 0 0 m A
M e th o d M e a s u re d
-
0 % 5 0 % 1 0 0 %
C lo s e C lo s e
B a c k E d it C a lib r a te M E N U B a c k S ta rt A d ju s t M E N U
M E N U M E N U
102
Viewing Calibration Results
1 2
N e x t
B a c k M E N U
P a g e
103
Calibration
(Empty)
104
Appendixes
Appendix 1:
Technical Data ................... 106
Appendix 2:
Index .................................... 112
Part_X1_K-_ja_T-tyypin_termoparien_speksit_korjattu.p65
105 27.09.2004, 10:34
Appendix 1, Technical Data
Appendix 1,
Technical Data
Contents,
MC3 General Specifications 107
Sections 108
Pressure Sections .............................. 108
Temperature Section (ET) .................. 109
RTD Measurement
and Simulation ............................. 109
Thermocouple Measurement
and Simulation ............................. 110
Reference Junction (RJ) .............. 110
Electrical Section (E) ......................... 111
106
Part_X1_K-_ja_T-tyypin_termoparien_speksit_korjattu.p65
106 27.09.2004, 10:34
General
General
Display 96 x 72 mm (3.78" x 2.83"), 320 x 240 pixels, back lit LCD
Weight 1.7 - 2.3 kg (3.7 - 5.1 lbs)
Dimensions 245 mm (9.6") x 192 mm (7.5") x 74 mm (2.9") (d/w/h)
Case protection IP65 (dust and water proof)
Keyboard Membrane protected individual keys
Battery type Rechargeable NiMH, 4000 mAh, 7.2V DC
Battery operation Average 10 hours
Charger supply 100...240 VAC, 50-60 Hz
Operating temperature -10...50°C (14...122°F)
Specifications valid 15...35°C (59...95°F)
Temperature coefficient < ±0.001% RDG / °C outside of 15...35°C (59...95°F)
Storage temperature -20 to 60°C (-4 to 140°F)
Humidity 0 to 80% R.H. non condensing
Measurement sample rate 2.5 / second
Warranty for MC3 2 years as standard, battery pack 1 year
107
Part_X1_K-_ja_T-tyypin_termoparien_speksit_korjattu.p65
107 27.09.2004, 10:34
Appendix 1, Technical Data
Sections
Pressure Sections
MC3 can include the barometric option (PRB) and one of the other pressure inputs listed
below.
(2 (1
Module Unit Range Resolution Uncertainty (±)
PRB kPa a 80 to 120 0.01 1 mbar
mbar a 800 to 1200 0.1
psi a 11.6 to 17.4 0.001
PR2C kPa -100 to 200 0.001 0.05% FS (1 mbar)
bar -1 to 2 0.00001
psi -14.5 to 30 0.0001
PR20C kPa -100 to 2000 0.01 0.05% FS (10 mbar)
bar -1 to 20 0.0001
psi -14.5 to 300 0.001
PR160 MPa 0 to 16 0.0001 0.05% FS (80 mbar)
bar 0 to 160 0.001
psi 0 to 2400 0.01
1)
Uncertainty includes reference standard uncertainty, hysteresis, non-linearity, repeatability and typical long term stability for men-
tioned period. (k=2)
2)
Every pressure input’s range may be displayed also in absolute pressure if the Barometric option (PRB) is installed.
Supports the following pressure units as standard: Pa, hPa, kPa, MPa, mbar, bar, lbf/ft2, psi, gf/cm2, kgf/cm2, kgf/m2, kp/cm2, at,
mmH2O, cmH2O, mH2O, iwc, ftH2O, mmHg, cmHg, mHg, inHg, mmHg(0°C), inHg(0°C), mmH2O(4°C), inH2O(4°C), ftH2O(4°C),
inH2O(60°F), mmH2O(68°F), inH2O(68°F), ftH2O(68°F), torr, atm.
Pressure inputs 20 bar and below, pressure connection G 1/8 (ISO 228/1) 60° internal cone.
PR160 input, pressure connection G 1/8 (ISO 228/1) female.
Wetted parts AISI316 stainless steel, Hastelloy, Nitrile rubber.
The maximum overpressure is twice the range.
108
Part_X1_K-_ja_T-tyypin_termoparien_speksit_korjattu.p65
108 27.09.2004, 10:34
Sections
1)
Uncertainty includes reference standard uncertainty, hysteresis, nonlinearity, repeatability and typical long term stability for men-
tioned period. (k=2)
2)
Load effect < 5 mV/mA. Maximum output current 5 mA.
3)
Load effect < 100 mV/mA. Maximum output current 10 mA (0 .. 10 V), 3 mA (10 .. 12 V).
4)
Maximum load impedance 400 ohm.
5)
Amplitude range 0 .. 12 Vpp. Amplitude setting accuracy up to 5 kHz ±(200 mV + 5% of set value). Waveforms: Square wave
(positive / symmetric) and sinewave (above 40 Hz).
6)
Pulse generation frequency range 0.1 ... 1000 Hz. Amplitude setting 0 ... 12 Vpp.
7)
Valid with measurement current 0.2 ... 5 mA (1 ... 1000 ohm), 0.1 .. 1mA (1 ... 4 kohm). Ohm/RTD simulation speed 1 ms.
8)
Whichever is greater.
9)
Specification valid with 4 wire connection. In 3 wire connection add 10 mohm.
10)
Bias current < 10 nA.
109
Part_X1_K-_ja_T-tyypin_termoparien_speksit_korjattu.p65
109 27.09.2004, 10:34
Appendix 1, Technical Data
Resolution 0.01°C.
With internal reference junction (RJ) add 0.1°C uncertainty.
Thermocouple types U and L (DIN 43710) as well as C 3) (ASTM E 988 - 96), G 3) (ASTM E 1751 - 95e1) and D 3) (ASTM E 988 - 96)
also available as standard.
Other thermocouple types are available as option.
1)
Uncertainty includes reference standard uncertainty, hysteresis, nonlinearity, repeatability and typical long term stability for men-
tioned period. (k=2)
2)
IEC 584, NIST MN 175, BS 4937, ANSI MC96.1
3)
±0.02 % of thermovoltage + 4 µV
110
Part_X1_K-_ja_T-tyypin_termoparien_speksit_korjattu.p65
110 27.09.2004, 10:34
Sections
1)
Uncertainty includes reference standard uncertainty, hysteresis, nonlinearity, repeatability and typical long term stability for men-
tioned period. (k=2)
2)
Impedance >1 Mohm
3)
Impedance < 7.5 ohm
4)
Impedance > 1 Mohm. Frequency measurement minimum amplitude 0.5 Vpp (< 5 kHz),
1Vpp (5...50 kHz). Pulse counting minimum amplitude 0.5 Vpp (pulse length > 100 µs),
1 Vpp (pulse length 100 µs...10 µs).
Trigger level range -1…+15 V.
5)
Maximum load impedance 800 ohm
111
Part_X1_K-_ja_T-tyypin_termoparien_speksit_korjattu.p65
111 27.09.2004, 10:34
Appendix 2, Index
C E
Calibration Editing fields 19
Appendix 2, Adjustment 78
As Found 78
Electrical limit switch
calibration 94
Index
As Left 79 Electrical section (E) 6
Automatic 86 Error situations in thermo-
Deleting the results 103 couple measurement 71
Examples 87 Examples of Instrument
Generally 76 Calibration 87
Instrument adjustment 96 External Reference
Instrument Window 83 Junction 69
Maintaining the Instrument
2 Database 97 F
2-wire system 73 Manual 86
Phases 77 Firmware
Procedure, defined 84 Basic Mode 15
3 Editing fields 19
Required sections 80
3-wire system 72 Saving the results 85 Function Keys 17
Selecting the General description 14
Menus 17
4 Instrument 82
Calibration Results Section information 19
4-wire system 72 Status bar 16
Deleting 103
Viewing 102 User Interface 16
A Calibrator configuration, Frequency generation 51
see Settings 62 Frequency
Adding Instruments 97
Carrying case 13 measurement 40
Adjustment 78, 96
Case 13 Function Keys 17
As Found calibration 78
As Left calibration 79 Charger 11
Automatic calibration 86 Compensation loop 73 G
Auto-Off Delay 63 Computer Interface 5 Generating (see also
Configuring the Calibrator, Simulating)
B see Settings 62 Changing the generated
Connections 4 signal 46
Bag 13 Computer Interface 5 Current 47
Basic Mode, defined 29 Contrast 7 Frequencies 51
Battery Pack 12 Creating Instruments 97 Pulses 52
Binary Signal Detection 37 Current generation 47 Using the E section 47
Brightness 7 Current measurement 33 Using the ET section 47,
49, 51, 52
D Voltages 49
Deleting Instruments 101
Display
Brightness 7
Contrast 7
112
Index
I M P
Info Key 8, 19 Maintenance, see Parallel functions
Instrument adjustment Settings 62 in MC3 74
support 96 Manual calibration 86 Pressure
Instrument Database 97 MC3 recalibration 26 Naming conventions 67
Calibration Instructions Measuring Pressure Types 66
Page 101 Current 33 Square rooting
Calibration Settings Frequencies 40 instruments 67
Page 100 General description 31 Pressure input, zeroing 32
Deleting Instruments 101 Limit Switch test 38 Pressure measurement 32
General Data Page 98 Pressure 32 Pressure sensor/transmit-
Instrument Input Page 99 Pulses 41 ter calibration 88
Instrument Output Resistance 36 Pulse generation 52
Page 99 Switch State Detection 37 Pulse measurement 41
Instrument Window 83 Temperatures (RTD
Internal Reference sensor) 42 R
Junction 68 Temperatures (T/C) 43
Using the E section 33, Ramping 58
K 34, 40, 41 Recalibration 26
Using the ET section 34, Reference Junction
Keyboard, defined 8 External 43, 54, 69
36, 42, 43
Voltages 34 Internal 7, 43, 54, 68
L Menus 17 Modes 69
Limit Switch test 38 Reference Junction
Block 7
Line Frequency 63 N Required sections when
Neck support strap 13 calibrating
an instrument 80
O Resistance
Measurement 36
Operational Sections 4 Resistance simulation 53
Options 21 RTD sensor simulation 53
113
Appendix 2, Index
S T
Safety 22 Temperature indicator
Saving calibration calibration 92
results 85 Temperature Measurement
Selecting the instrument to RTD sensor 42
be calibrated 82 Thermocouple 43
Service 26 Temperature recorder
Settings calibration 92
Auto-Off Delay 63 Temperature
Date Format 63 section (ET) 6
Display Backlight Test Diode, measurement
Brightness 63 parallel to a 73
Language 63 Thermocouple
Line Frequency 63 simulation 54
Temperature Scale 63 Troubleshooting thermo-
Temperature Unit 63 couple measurement 71
Time and Date 64
Time Format 63 U
Volume setting 63
Simulating (see also Unpacking 3
Generating) User Interface 16
Resistance 53
RTD sensors 53 W
Thermocouples 54 Warnings 23
Simultaneous measure- Viewing 7
ment/generation/ Viewing Calibration
simulation 74 Results 102
Special Generations 56 Voltage generation 49
Ramping 58 Voltage measurement 34
Stepping 57 Wrist strap 13
Stand 13
Starting MC3 14, 28 Z
Status bar 16
Stepping 57 Zeroing a
Support 13 pressure input 32
Support for instrument
adjustment 96
Switch State Sensing 37
114