15.AP Reprocessoring Ver 2
15.AP Reprocessoring Ver 2
PENTAX
Reprocessing
of flexible endoscopes
Pre-cleaning
After the procedure, keep the endoscope connected to
the Processor
& Suction unit
Using moistened
gauze, wipe the
external portion of
Insertion Tube
Pre-cleaning
Set the Lever of the water bottle to
“Drain”
position
Leakage Test
Remove Suction valve , Air/Water valve and rubber
inlet
•Distal Cap
(if applicable)
Leakage Test
Attach the soaking cap to prevent fluid invasion to
electrical pin
DO NOT attach
the RED
ETO venting cap
!!!
Leakage Test
Perform Dry leakage test
Move deflection
wheels to angulate
distal end
Leakage Test
Perform Wet leakage test (Recommended)
Using cylinder
cleaning brush,
brush the Inlet area
Using cylinder
cleaning brush,
brush both the
Suction & Air/Water
cylinder
Using cylinder
cleaning brush,
brush both the
Suction & Air/Water
valve
the Cleaning
Accessories
*Irrigation Tube
Soaking (Disinfection)
Manual method
Transfer the scope to a dedicated container with
Disinfectant
Soaking (Disinfection)
Manual method
Immerse the scope in a dedicated container with
Disinfectant
Follow recommended
exposure parameters
(time & temperature)
for disinfectant
according to
manufacture
Soaking (Disinfection)
Manual method
Transfer the scope to a dedicated sink with sterile water
Soaking (Disinfection)
Manual method
Immerse the scope in a dedicated sink with sterile water
Re-Connect all the
Cleaning
Accessories
Soaking (Disinfection)
Automatic method using AER machine
30
Lifecare
Post-Cleaning
Remove the scope from the sink
Post-Cleaning
Remove the water within the inner channel
Using Syringe,
pump air into all
channel to remove
water
Post-Cleaning
Hang dry the scope in a well ventilated, dust-free
cabinet.
THANK YOU
PENTAX