Professional Documents
Culture Documents
NEXUS 605RS/RSB
Rockwell (Brinell) hardness tester
Changes in products and/or product specifica-
tions can emerge due to new technologies and
continuous development.
FOR
NEXUS 605RS/RSB
The product meets the applicable DIN, ISO-EN, ASTM and JIS standards.
Before using this instrument, read through this user manual carefully to use the product properly.
After reading, keep the manual in an easy-to-access place for referencing whenever needed.
Read the “Safety Cautions” and “User’s Guide” thoroughly before usage, and properly operate the product.
In order to use the product safely, do not ignore visual cautions, warnings or safety symbols.
Please refer to the limited guarantee and responsibility issues in the INNOVATEST® guarantees conditions.
www.innovatest-europe.com 1
2 www.innovatest-europe.com
Make sure these signs and symbols are well understood before reading this manual or operating the instrument.
WARNING!
Ignoring this information and mishandling of the
equipment can lead to severe bodily injuries and
material damage.
- o not modify this equipment. Doing so can cause fire and/or electric shock.
D
- Do not fabricate anything with the power cord, forcibly bend it, twist, or pull it, doing so can cause fires and/or
electric shock.
- Do not disassemble this equipment. Doing so can cause electric shock. Contact our technical service department for
checks, adjustments, or repairs inside the equipment.
- Do not operate the equipment at a voltage other than the power voltage that is indicated. Doing so can cause fires.
- Do not place a vase, or any container holding water, near this equipment.
- Fires can occur if water gets inside the equipment.
- If, by any water does get inside the equipment, turn off the power to the equipment’s main unit, pull the plug out of
the socket, and call our company’s technical service department.
- Always install grounding. If, by any chance a component gets damaged without grounding, electric shock can occur.
- Do not operate the equipment if there is ever any abnormality, such as signs of smoke, unusual smell.
- In such a case, turn off the power to the equipment’s main unit, pull the plug out of the socket, and call our company’s
technical service department.
- Pull out the plug from the socket when exchanging a fuse. Not doing so can cause electric shock.
- Do not insert or unplug the plug with wet hands. Doing so can cause electric shock.
- Do not install the equipment on an unstable base. Doing so can affect the proper working and cause the equipment
to fall down and/or cause accidents and injuries.
- Do not scratch or forcibly twist a cable.
- Continued use of a cable after it is scratched or twisted can cause fire and/or electric shock.
- Do not place the equipment in a high temperature and humidity location.
- When cleaning or checking the equipment, make sure the plug is pulled out of the socket for safety.
- Do not use fuses and illumination lamps other than those specified. Doing so can cause fire.
- Do not connect special attachments to this equipment that are not “INNOVATEST®” standard attachments, doing so
can cause poor performance or defects.
- Do not use batteries other than the specified ones.
- Do not block the halogen lamp area and ventilation. Blocking the ventilation can cause heat to accumulate inside the
instrument, which in turn, can generate fire.
- Never open any of the panel or cover other than when unpacking or assembling.
- High voltages existing in the instrument may cause electrical shocks to personnel.
- Grounding, the power supply of the instrument must be grounded.
- For the power cable, always use the power cable with 3-pins ground wire provided.
www.innovatest-europe.com 3
Introduction
Thank you for buying the NEXUS 605RS/RSB closed loop, force feedback, Rockwell, Superficial Rockwell hardness tester.
The NEXUS 605RS/RSB is a Rockwell (Brinell) hardness tester with test forces ranging from 2.5Kgf to 187.5Kgf (depending on
model), Closed Loop, Load Cell, Force Feedback technology for a reliable fast measurement procedure.
The NEXUS 605RS/RSB, Rockwell (Brinell) hardness testers, combine a practical design with universal specifications. State-of-
the-art closed loop, load cell technology and a - Patent Pending - force feedback system, assure fast test results at the highest
possible accuracy.
The INNOVATEST® Instruments design team has used recently developed sensor technology to create a new industry standard
for Rockwell (Brinell) hardness testing machines, eliminating weights and complex mechanical control systems.
The closed loop system applies force, calculates, filters and controls digital data in a 32-bits embedded CPU system.
The NEXUS 605RS/RSB is perfectly suited for educational purposes, research institutes and light industry.
During the assembly process the NEXUS 605RS/RSB will be configured to your particular technical requirements or even
painted in your company style colours.
The NEXUS 605RS/RSB is designed in The Netherlands, Europe, and meets or exceeds all applicable EN, ISO and ASTM
standards.
4 www.innovatest-europe.com
Test procedure control
Traditional hardness testing systems use a “dead weight” mechanical design or inaccurate spring force mechanism to apply
the test force. Such systems are lacking test control, as there is no feed-back on the actual applied force.
The Closed Loop technology with a force feedback system, as applied in the NEXUS 605RS/RSB , constantly measures and
controls the applied force on the tester’s indentor and tested surface. Consequently, this superior control system offers an almost
unlimited selection of test loads and test rates for virtually any test condition imaginable.
s
m
ck
ste
ba
sy
d
ht
fee
ig
we
e
orc
da
de
pf
on
loo
m
m
sed
Co
Clo
Force
Time
Elimination of overshoot due to sophisticated algorithms detecting contact with the indentor and the tested object’s surface.
The application and removal of the test force is fully automatic, as well as the positioning of the indentor and the pre-determined
objective. The result is an absolute vibration free operation while reducing the operator’s workload to a minimum.
ste
ba
sy
eed
ht
ig
we
ef
orc
ad
de
pf
on
loo
m
m
sed
Co
Clo
Force
Time
www.innovatest-europe.com 5
Contents
Introduction 4
2. Main unit 11
2.1 Hardness tester body 11
7. Calibration procedures 51
7.1 Force Calibration 51
7.2 Maintenance advice 51
8. Technical specifications 52
6 www.innovatest-europe.com
9. Standard delivery and accessoires 53
9.1 Standard delivery 53
9.2 Optional accessoires 53
9.3 Technical drawings 54
www.innovatest-europe.com 7
1. Unpacking & checking contents
This section describes unpacking conditions and standard accessories.
1.1 Unpacking
When you receive the product, unpack it and check if all components are included.
NOTE:
While unpacking the unit, use no extreme force to remove the packing materials. The accessories box is placed next to the
instrument and separated by soft materials to avoid damage to the tester. REMOVE the accessory box first! STORE the packing
material in a safe place as you might need it again. Never ship or transport the tester without the correct
packing materials or with the transportation safety inside the tester not being in the correct position.
This could cause severe damage to the testers load application system.
The standard components and accessories differ with system types within the NEXUS 605RS/RSB, but each complete shipment
includes at least:
NOTE 1:
While lifting the tester, do not touch sensitive parts like the indentor or illumination source;
NOTE 2:
Note that the actual packing conditions may differ from those shown above.
8 www.innovatest-europe.com
1.2 Accessory Box (or installed on your tester)
- V-anvil
- Flat anvil 60mm
- Brinell microscope (NEXUS 605RSB only)
- Power cable
- Four adjustable feet
- INNOVATEST® certificate
- Installation and user manual
- Pre-installed (plug and play)
NOTE 1:
Some components or parts may be packaged separately and may not be included in the accessory box or may have been installed on the
hardness tester.
NOTE 2:
Special options may be set in the places which are left blank in the accessory box.
www.innovatest-europe.com 9
1.3 Standard and optional Accessories/features
10 www.innovatest-europe.com
2. Main unit
This section describes the individual parts of the main tester body
1. Front panel
2. 6.5” LCD touch panel 1 2
3. USB port
4. Adjustable clamping attachment
5. Large round workpiece stage
6. Spindle protection cover
7. Elevator spindle screw
8. Tester house
9. Emergency switch
3
10. Adjustable vibration damper
11. Power switch
10
11
www.innovatest-europe.com 11
3. Installation of the hardness tester
This section describes the correct way of installing your NEXUS 605RS/RSB hardness tester.
Install the hardness tester system in the conditions shown below in order to make full use of its functions and performances and
to avoid defects or inaccuracy of any kind.
VIBRATION
- Install the hardness tester in a place free from floor vibrations.
- Install the hardness tester on a rigid stand securely fixed (Do not install it on a to high or infirm stand.)
- Do not install the hardness tester in a place where many people come and go.
- Do not install the hardness tester near equipment generating vibrations.
- Install the hardness tester on the first floor of a building if possible.
- Do not install the hardness tester in a building near a road with heavy traffic.
- Do not install the hardness tester in a building where a crane, vibration tester or similar equipment is installed.
NOTE:
Vibrations transmitted to the hardness tester result in inaccurate measurement in particular in low force application procedures.
WORKBENCH
Place the hardness tester on a horizontal work bench (such as a cabinet, desk or table).
The installation area needed is not less than 500 mm in width, 600 mm in depth and 850 mm in height.
The work bench should be rigid and withstand a load over 150kg.
NOTE 1:
Do not apply any impacts to the hardness tester when moving it.
NOTE 2:
When carrying the tester, be careful not to bring your hands, etc. into contact with the indentor.
12 www.innovatest-europe.com
3.2 Leveling
a) Install the four adjustable vibration dampers which are packed in the accessory suitcase.
b) The NEXUS 605RS/RSB system is equipped with a unique load application system which is not sensitive to
incorrect levelling. Nevertheless, to eliminate possible wear and tear on the tester’s mechanical structure we advise
to make sure that your tester is correctly levelled.
This section describes the correct way of getting your tester powered up
Assure yourself that your tester’s power requirement is in compliance with your power supply.
The tester’s power details can be found on the back plate mentioning VOLTAGE requirements.
NOTE 1:
Connecting your tester to an incorrect main power supply could cause severe damage to the instrument and could cause fire.
NOTE 2:
At power on, the switch will light up.
NOTE 3:
Turn the emergency switch gently clockwise to unlock the switch.
Even when the power switch is light up, you need to assure that the emergency switch is not activated.
www.innovatest-europe.com 13
4. Preparing NEXUS 605RS/RSB for operation
The NEXUS 605RS/RSB is a complex mechanical/electronic equipment which requires professional installation from well
trained staff. We strongly recommend to opt for installation and training by INNOVATEST® service personnel.
Defect or damage caused by wrong operation or unsuitable installation are not covered by any warranty.
Make sure that the tester is calibrated (forces & test result) by INNOVATEST® or any authorized service partner at your location.
14 www.innovatest-europe.com
5. Using the NEXUS 605RS/RSB
5.1 Power on
Switch on the tester using the ON/OFF mains power switch on the back panel.
NOTE 1:
Make sure that the emergency switch is unlocked. When the emergency switch is locked, the tester will not power on, even after the power
switch is switched on.
Turn the emergency switch clockwise gently to unlock the emergency switch.
If the screen does not show any information within 15 seconds, this means that the battery of the time of day clock needs to be replaced. In
that case, the time and date will not be reliable.
After switching on the power of the tester, a splash screen will appear as long as the machine is homing. During homing,
the load motor will move the indenter upward to the home position. The splash screen shows the INNOVATEST logo, the
name of the tester type and the firmware version number.
A progress bar indicates the time that is left before the splash
screen is removed, meaning that the homing sequence is
completed. Wait until the splash screen disappears. To prepare
the machine for use, a popup tells you to turn down the elevator
spindle until the indenter is not touching the surface and to press
OK to zero the depth sensor. To turn the elevator spindle down,
move the spindle wheel counter clockwise.
www.innovatest-europe.com 15
After pressing OK on this popup, the machine is ready
to use.
NOTE:
You can zero the depth sensor at any time by touching the screen area that is showing the depth sensor bar graph under the text “READY”
and keeping this area pressed for more than 2 seconds. The same popup appears telling you to turn the elevator spindle down and to press
OK to zero the depth sensor.
When the main screen is displayed, the tester is ready to use. The concept behind the screen layout is to give direct access
to functions from within the user interface, without having to browse menus to find the desired feature. Therefore, the
screen is divided into area’s that function as touch-sensitive buttons. Some of these screen areas have a double function:
When pressing a button briefly, the “normal” function is executed. When pressing a button for more than 2 seconds, the
“alternative” function is executed. When pressing a screen area that has a touch sensitive button function, the background
colour of this button changes.
16 www.innovatest-europe.com
“Alternative Function”
“Normal” Function (when pressed briefly) (when pressed for more than 2 seconds)
3: Measurement data & statistics: To select the Save measurement: To save the result of a hardness
desired hardness scale and force for the indentation. test into the measurement list, when automatic saving of
measurements is disabled.
www.innovatest-europe.com 17
“Alternative Function”
“Normal” Function (when presseded briefly) (when pressed for more than 2 seconds)
6: Time/date: To show the current date and time in the Time/Date setup: European or American format
chosen format. To choose between the European or American date and
time notations.
10 : Settings: To enter the settings menu that gives access to Information screen : This screen shows the hardness tester
the settings such as operating mode, touch calibration, type and the firmware version number.
firmware upgrade, language selection.
Export measurements:
To export the stored measurements to an USB stick. X
12 :
Dwell time and progress indication: Shape correction setup : Convex or off
For Rockwell scales: To set the shape correction type and size.
The preload dwell time, main load dwell time and This function is only available when a hardness scale is
recovery load dwell time are shown individually. chosen for which a shape correction is relevant (Vickers,
13 : For non-Rockwell scales: Rockwell). Shape correction is not relevant for the Knoop
The dwell time is shown. The progress of the indentation and Brinell scales.
is shown using a graphical representation of the indenter
and the surface of the sample under test.
18 www.innovatest-europe.com
“Alternative Function”
“Normal” Function (when pressed briefly) (when pressed for more than 2 seconds)
14 : Manual start/stop : Zero depth sensor: To zero the depth sensor when the
The graph shows the depth sensor displacement when the elevator spindle is moved down and the indenter is not
indenter is pushed upward while touching the sample under touching the sample under test.
test and moving the elevator spindle upward. When the
depth sensor reaches the OK range that is indicated on the
bar graph, the indentation cycle starts.
The Rockwell depth sensor information/Brinell diameter entry fields are shown below:
When the tester supports non-Rockwell scales, the diameters D1 and D2 have to be entered in the diameter entry field.
The hardness is calculated using the entered diameters. When a Rockwell scale is selected, the indenter displacement (Ind.
displ.) and the depth are shown. The “indenter displacement” value is the distance the indenter moved relative to the zero
position of the depth sensor. The zero position is when the indenter is not touching anything. The indenter displacement
value is shown when a new indentation is started and not when a measurement from the measurement list is selected. The
indenter displacement shows the distance that the indenter has moved relative to the zero position of the depth sensor.
The ”depth” value is the difference between the indenter displacement value during preload dwelling and the indenter
displacement value during recovery dwelling. This is the actual Rockwell depth used to calculate the Rockwell hardness value.
www.innovatest-europe.com 19
5.3.2. Limit entry & limit setup
When a hardness value has been calculated or a measurement is selected in the measurement list, the background colour
of the measurement data & statistics field is either blue, red or green. When the background colour is blue, this means that
both the upper and lower limit field is set to inactive and showing “---“. When the background colour is green, this means that
the hardness value is within the limits. When the colour is red, this means that the hardness value is outside of the limits. The
hardness values in the measurement list will also be coloured depending on whether the value is within or outside the limits.
The limit entry fields are used to enter the upper and lower limit individually:
The alternative function of the limit entry fields allows setting the corresponding limit active or inactive by pressing either the upper
limit or lower limit entry field for more than 2 seconds. The following screen will appear when setting the upper limit entry field:
NOTE:
To be able to enter a limit value, first set this limit active.
When a limit is set to inactive, the text “---“ is shown in
the corresponding limit field.
20 www.innovatest-europe.com
When the upper limit is set active and the upper limit entry field is pressed briefly, a keypad appears, allowing entering the
desired limit value:
When the lower limit is set active, press the lower limit
entry field briefly to enter a new lower limit value.
NOTE:
When entering the limits, the device will not allow
entering f.e. a lower limit value that is higher than the
upper limit value or vice versa.
Here are some examples of limit settings and how they affect the way that the hardness value and measurements are displayed:
www.innovatest-europe.com 21
Example 2: Only the upper limit is active.
All hardness values below the upper limit are marked
green. All hardness values above the upper limit are
marked red.
22 www.innovatest-europe.com
Example 4 : Both limits are active.
All values that are within the limits are marked green.
All values that are out of the limits are marked red.
NOTE:
Up to a maximum of 50 measurements can be saved.
www.innovatest-europe.com 23
The measurement list can be browsed by pressing the
up or down button. You can also select a measurement
from the list simply by pressing the desired list item.
When selecting a measurement, the hardness value and
Rockwell depth/Brinell diameters of the corresponding
measurement will be shown. At the same time, the
background colour of the measurement data & statistics
field is updated to indicate if the hardness of the
selected measurement is within or outside the limits.
The device is able to take a snapshot of all the relevant settings at that moment and store these settings as a program with a
user-definable name. This way you can define your own predefined sets of settings and easily switch between them. The load
program button (see picture below) is used to recall a stored program. The “---“ below the load button icon is an indication that
currently no program is loaded, or that a loaded program has been overruled by changing one of the settings.
24 www.innovatest-europe.com
A window will appear to confirm the delete action:
To delete all the programs from the list, press the delete
button for more than 2 seconds.
www.innovatest-europe.com 25
5.3.5. Save a program
NOTE:
Up to a maximum of 50 programs can be stored.
When pressing the save button, a window will appear with an automatically generated program name:
26 www.innovatest-europe.com
After entering a new program name and pressing the
OK button to accept the changed name and pressing
OK to save the program, the current settings are saved
using the entered name. In the save program button, the
number of saved programs is shown.
When suddenly reversing the elevator spindle movement before the OK range is reached, the text “UP” will change to
“DOWN”, indicating that you need to move the elevator spindle down until the text “READY” appears, before you can start
over again to start a new indentation cycle.
www.innovatest-europe.com 27
To start an indentation, make sure that you keep
moving the elevator spindle upward, preferably without
any interruption and at a constant speed, until the bar
graph reaches the OK range. Stop moving the elevator
spindle upward when the bar graph is somewhere in
the middle of the OK range.
The text “STOP” appears and the background colour of
the bar graph changes to red. At the same moment the
indentation cycle starts.
During the indentation cycle, the status bar shows the different phases of the indentation process. When pressing the bar graph
while the text “STOP” is shown, the indentation cycle is stopped immediately, the load motor is retracted and the text “DOWN”
appears. The elevator spindle needs to be moved down until the indenter is not touching the sample under test anymore. At
that moment the tester will be ready for a new indentation and shows the text “READY”.
The measurement and statistics field shows either the current measured hardness value or the hardness of a stored
measurement after it is selected in the measurement list. When the limits are active, the colour of the measurement and statistics
field indicates whether the hardness value is within the limits or not.
Additionally the statistics data, which is calculated over all saved measurements, is shown.
1. Hardness value (calculated after an indentation was completed, or belonging to a stored measurement selected in
the measurement list). When the hardness value shows “---“, this means that a new measurement is initiated, but no
hardness is calculated yet.
Press the measurement and statistics field briefly to select the desired hardness scale and force. The following screen will appear:
28 www.innovatest-europe.com
Depending on the tester type and force range, the
appropriate scales and forces are enabled or disabled.
Press the desired hardness scale in the left column and
then press the desired force from the table at the right
side of the screen.Press the OK button to confirm your
choice or press the CANCEL button to discard any
changes.
www.innovatest-europe.com 29
• In automatic mode, the hardness value will be saved automatically to the measurement list without any user interaction.
• In manual mode, the hardness value is saved when the measurement and statistics field is being pressed for more
than 2 seconds. A popup appears with the question: “Add measurement to the list?”. Press YES to save the measurement.
5.3.9. Conversions
The device can store and show three user-selectable conversions of the current hardness value into three other hardness scales.
The current conversions are shown in the conversions field:
The conversion selection screen closes and the new conversion is shown in the conversions field.
30 www.innovatest-europe.com
When the hardness value cannot be converted into the selected conversion scale, the converted hardness value is displayed as
“---“. This happens when the hardness value is outside the range of the chosen conversion scale or when a new measurement
is not finished yet.
NOTE:
When one or more conversions are chosen for a certain scale and the scale is changed, it is possible that some of the conversions are not
relevant anymore for the new chosen scale. In that case, the irrelevant conversions are reset to “No conversion”.
The time/date button shows the current time of day and date in the current selected notation.
The time and date is set by pressing the time/date button briefly. The following screen appears:
www.innovatest-europe.com 31
The notation can be chosen with the alternative function of the time/date button by keeping this button pressed for more than
2 seconds. The following screen appears:
The delete button is used to delete a saved measurement or all saved measurements at once from the measurement list:
The light control button is used to the set the light level. The current light level in percent is shown in the button.
32 www.innovatest-europe.com
When the light control button is touched briefly, a window with a slider appears :
The settings button is used to change the operating mode, calibrate the touch screen, upgrade firmware, change the language
or to enter the calibration screens.
In the settings button, the current operating mode is shown. The default operating mode, that is set at power on, is “Operator”.
When pressing the setup button briefly, the settings screen appears:
www.innovatest-europe.com 33
The settings screen gives access to the following functions:
• Operating mode:
To change the operating mode. Service technicians use this function.
• Calibrate the touch screen:
To re-calibrate the touch screen.
• Upgrade firmware :
To install new firmware using an USB stick.
• Language selection:
To select another language. All texts are displayed in the chosen language.
When pressing the setup button and keeping it pressed for more than 2 seconds, the information screen appears:
34 www.innovatest-europe.com
This screen shows the tester type and the firmware version.
When pressing the “Calibrate touch screen” button on the settings screen, the following screen appears:
Press the circle at the top left corner of the screen. The following screen appears:
www.innovatest-europe.com 35
Press the circle at the bottom right corner of the screen. The following screen appears:
5.3.13.2. Language
To change the language, press the language button on the settings screen briefly, until the language of your choice appears.
The device remembers your choice, even after power off.
The status bar at the bottom of the start screen is used to show:
36 www.innovatest-europe.com
5.3.15. Dwell time and progress indication & shape correction setup
The dwell time and progress are combined in the dwell time and progress indication field:
The upper part of the field shows the dwell time value(s) in seconds.
For Rockwell scales, three dwell times are shown: the preload dwell time, the main load dwell time and the recovery load
dwell time.
For non-Rockwell scales, one dwell time is shown: the main load dwell time.
When an indentation cycle starts and the desired force for the particular indentation phase is reached, the corresponding
dwell time value will start counting down to zero. When the value reaches zero, the next phase of the indentation cycle starts.
When the indentation cycle is completed, the dwell time values reset back to their start values. The lower part of the field shows
a triangular shape representing the indenter and a rectangular shape representing the specimen surface. When an indentation
is started, the progress indication will show the position of the indenter while moving towards the specimen surface. When
the specimen surface is reached, the progress indication changes to a bar graph that represents the measured force until the
selected force is reached :
To set the dwell time, press the dwell time and progress indication field briefly. The following screen will appear:
www.innovatest-europe.com 37
For Rockwell scales, the preload, main load and
recovery dwell time can be adjusted individually. First,
select the dwell time that you want to adjust by pressing
the appropriate checkbox. Then keep pressing the slider
while moving the finger horizontally over the slider to
adjust the dwell time. For fine adjustments use the the +
and – buttons to increase and decrease the dwell time
in steps of 1 second. Press OK to save the changed
dwell time and leave the screen. The device will
remember the saved change, even after power off.
Press the CANCEL button to discard any change.
To set the shape correction, keep the dwell time and progress indication field press for more than 2 seconds. The following
screen appears:
38 www.innovatest-europe.com
5.3.16. Export measurements
All measurements in the measurement list, conversions and statistics can be exported to a file on an USB stick. The USB
connector for the stick is situated at the right side of the LCD screen.
NOTE:
After inserting the USB stick, the device needs several seconds to detect the stick. When the stick is not detected, re-insert the stick.
The USB stick has to be formatted for the FAT file system. Other file systems will not work.
When the measurements are saved to an USB stick, the measurements will be saved to a file called “report.txt” in the root
folder of the USB stick. When this file already exists, a popup message appears: “Overwrite file ?” When “Yes” is pressed,
the file is overwritten. After the file is written, a pop message appears: “File was written”. Press the OK button to proceed.
INNOVATEST
Nexus 605
TEST RESULT LIST
OPERATOR : ................
PART NR. : ................
DATE : 11/11/2016
TIME : 14:15:28
HARDNESS SCALE : HRA
CONVERSION 1 : None
:
CONVERSION 2 : None
:
CONVERSION 3 : None
:
FORCE : 60000.000g
DWELL TIME : 3s – 3s - 3s
MEASUREMENTS : 3
AVG : 83.43
MAX : 99.11
MIN : 75.50
S.DEV : 11.08
RANGE : 23.61
SINGLE VALUES
1. 99.11 HRA
2. 75.50 HRA
3. 75.70 HRA
www.innovatest-europe.com 39
5.4 Data output
USB PORT
www.innovatest-europe.com 40
Windows will try to automatically install the drivers and show the progress in a balloon.
Open the “Start Menu” and click with the right mouse button on “Computer” and select “Manage”
www.innovatest-europe.com 41
Click on “Device Manager”
Click with the right mouse button on “NXP LPC18xx VCOM” and select “Update Driver Software…”
42 www.innovatest-europe.com
Navigate to the directory with the drivers for “NXP LPC18xx VCOM”
Click on “Next”
www.innovatest-europe.com 43
Write down which COM port has been assigned to the LPC USB VCom Port in the Device Manager
44 www.innovatest-europe.com
Adjust the Port configuration to match the following settings:
The COM port can be different for each PC. Use the COM port which you wrote down earlier.
www.innovatest-europe.com 45
Output data from NEXUS 605RS/RSB to Excel with Data Redirector
- If you want each measurement in a new row, select “Add value in new row”
- If you want each value in a new column, select “Add value in new column”
- If you also want a timestamp with each measurement, select the box “Add timestamp”
After each measurement the selected cell in Excel will be updated with the measured value.
46 www.innovatest-europe.com
“Add value in new column”
www.innovatest-europe.com 47
“Add value in new column, Add timestamp”
If you only want to use the value without the scale (e.g. HRC), you can add a row below with a formula:
=LEFT(A1, LEN(A1)-4)
A1 is the cell which contains the measured value. -4 is the number of characters you wish to strip.
48 www.innovatest-europe.com
6. Performing a hardness test
2. Check that the correct indenter type (ball or diamond cone) is installed for the desired scale.
3. The adjustable clamping attachment is used to eliminate the play in the elevator spindle. When the elevator spindle
is adjusted so that the manual start bar graph indicates that the depth sensor is in the OK range, the clamping
attachment is clamped against the test table with a certain force. The clamping attachment should be adjusted so
that the clamping force is high enough to eliminate the play in the elevator spindle at the moment that the depth
sensor is in the OK range. When clamping, you feel an increasing resistance of the spindle wheel. This resistance
increases up to a maximum when the depth sensor moves into the OK range. When feeling no resistance while
the depth sensor moves into the OK range or when the resistance makes it impossible to reach the OK range, the
clamping attachment needs to be adjusted accordingly.
4. Select a hardness scale (and the desired force for non-Rockwell scales).
6. When relevant for the selected hardness scale, set the shape correction to the correct shape and shape
diameter.
In case a Rockwell scale is selected, the tester applies the preload force, then the main load force and finally the
recovery load. Each force application is followed by a corresponding dwell time during which the force is maintained.
When the indentation cycle is completed, the Rockwell depth and the calculated hardness are displayed, while the
text “DOWN” is displayed and the status bar shows: “Turn the spindle down”. Before starting a new indentation
cycle, move the elevator spindle downward.
In case a non-Rockwell scale is selected, the tester applies the main load, followed by the main load dwell time.
When the indentation cycle is completed, the text “DOWN” is displayed and the status bar shows: “Turn the
spindle down”. Before starting a new indentation cycle, move the elevator spindle downward.
Use the appropriate tool (Brinell microscope) to measure the horizontal and vertical diameter of the indentation and
enter the diameters in the diameter entry fields. The tester will calculate and show the hardness using the entered
diameters.
49 www.innovatest-europe.com
If you want the hardness measurements to be saved to the measurement list automatically, check the
“Automatic save measurement” option in the settings screen. Any new measurement will then be saved to the
list automatically up to a maximum of 50 measurements.
If you want to save the measurement manually, uncheck the “Automatic save measurement” option in the
settings screen. The measurement is saved when touching the measurement and statistics field and keeping
this field touched for more than 2 seconds. Saving is only possible when the hardness value is valid, thus not
showing “---“.
www.innovatest-europe.com 50
7. Calibration procedures
The NEXUS 605RS/RSB series is equipped with a highly sensitive and accurate load cell that gives feedback on the applied
force.
It is unlikely that you to need to adjust any settings of your NEXUS 605RS/RSB in the normal product live.
If your tester requires calibration please contact your supplier or the INNOVATEST service center directly.
Calibration of the NEXUS 605RS/RSB testers should be done by an experienced person only.
In the event that the force application of your tester needs to be recalibrated, please seek advice from your supplier or contact
the service department of INNOVATEST directly.
To calibrate the applicable force of your NEXUS 605RS/RSB series hardness tester, specific equipment is required (range of
calibrated precision load cells).
Do not clean your tester with chemical solutions such as “thinner” or white spirit as this can cause permanent damage to the
painting, metallic surfaces or plastics.
Keep the elevator spindle clean and do NOT lubricate the spindle with normal oil. Use sewing machine oil and CLEAN THE
SPINDLE THOROUGHLY after lubrication and make sure as little as possible oil is left on the spindle. Repeat the cleaning
procedure after a few days to ensure no oil will dry up on the spindle surface.
In case any part needs replacement, use genuine INNOVATEST parts to avoid disturbing of the performance of your tester.
51 www.innovatest-europe.com
8. Technical specifications
www.innovatest-europe.com 52
9. Standard delivery and accessories
• V-anvil
• Flat anvil 60mm
• Brinell microscope (NEXUS 605RSB only)
• Power cable
• Four adjustable feet
• Installation and user manual
• Pre-installed (plug and play)
53 www.innovatest-europe.com
9.3 Technical drawings
This SERVICE menu allows access to important machine settings. Pending the password, certain levels of the machi-
170
Max. 250
855
400
42
Spindle Dia.
220 540
190
ne settings can be reached. Your distributor has access to the password that is required for calibration and adjustments.
54 www.innovatest-europe.com
10. Periodical inspection
www.innovatest-europe.com 55
11. Trouble shooting
This section shows errors which may occur in the NEXUS 605RS/RSB series, their possible causes,
and countermeasures to be taken against such errors.
56 www.innovatest-europe.com
12. Replacement of fuse
Disconnect electric power cable from backside of main body after turn off the machine and disconnect remove power line.
www.innovatest-europe.com 57
Represented by:
Represented by: