Professional Documents
Culture Documents
ML C 650 c65mm1091
ML C 650 c65mm1091
(10) If water droplet and track pattern is not acceptable, glass must be retreated with Surface Seal.
NOTE: The Surface Seal treatment and the measurement procedure are not necessary for the cockpit side
windows.
NOTE: Operators may use Water Droplet Contact Angle Measurement to verify, in a less subjective
method, the condition of the Surface Seal treatment.
END OF TASK
4. Water Droplet Contact Angle Measurement
A. Procedure.
NOTE: If not familiar with the techniques required for performing this measurement, review the video included
in the testing kit.
NOTE: Move the airplane into a hangar or, if necessary to perform outside, to a dry protected area with the
windshield not warm to the touch. Airplane must be nearly level because the angle of the glass surface
affects drop length.
NOTE: Ensure that the appropriate water droplet contact angle measurement kit is available.
(1) Water Droplet Contact Angle Measurement.
(a) Clean the windshield with the supplied windshield glass cleaner and cleaning cloths to remove dirt and grime.
(b) Thoroughly clean the locations (refer to Figure 605) at which the droplet contact angle is to be measured using the
solvent and cleaning cloths supplied in the test kit.
(c) Obtain a pair of new disposable gloves and wear them until the measuring steps are completed to ensure that no
fingerprints or oils are deposited on the measurement locations.
(d) Install a fresh reservoir tip on the calibrated pipetter.
(e) Fill the reservoir of the calibrated pipetter with deionized water from the supply container.
(f) At the location to be tested, place a 10 μl (microliter) drop on the windshield with the calibrated pipetter using the
following procedure:
1 Expel a drop from the calibrated pipetter and touch some surface to remove the drop from the tip of the pipetter.
This ensures that the next drop that is released will be of correct size.
2 Expel a 10 μl drop from the calibrated pipetter.
3 Gently place the drop on the surface at the location to be tested by touching the drop to the surface and then
withdrawing the pipetter tip vertically to prevent smearing or otherwise distorting the drop on the glass surface.
NOTE: If the 10 μl drop will not stick to the windshield, but runs off, it may be understood that the
contact angle for that spot exceeds 100 degrees.
NOTE: The drop dimension must be measured within one minute since the evaporation of water
will affect the accuracy.
(g) The Macroscope 25 optical instrument with the reticle length scale in millimeters should be used. (Ignore the lower
3/16 inch scale.) Focus on the reticle with the upper knurled ring.
(h) Position the macroscope on the glass surface over the drop, and adjust the focus (lower knurled ring) until the drop is
clearly seen. If necessary, use and properly adjust the portable light source which attaches to the Macroscope.
Position the macroscope so that the major axis (length) of the drop can be precisely measured. Align the reticle scale
so that the left edge of the drop as viewed in the eyepiece corresponds to the 0 reading. Measure the drop dimension
by determining which reticle marking corresponds to the right edge of the drop.
NOTE: If measurement exceeds five mm, estimate the actual length.
(i) Record the length measurement and location letter on the supplied note pad.
(j) Remove the water drop by wiping dry with a cleaning cloth.
(k) Repeat Steps (f) through (j) to total five measurements at each of the four specified locations.
NOTE: If a single measurement is very different from the others, discard the measurement and make
an additional substitute measurement.
NOTE: If fewer measurements are used, a high degree of confidence in the average value is not
achieved.
(l) Average the five readings at each location to determine the average drop length.
(m) Using the appropriate chart, convert the average drop length to contact angle.
NOTE: Because the angle of the windshield is different at the inboard and outboard locations, the
droplet will be longer when the surface of the glass is more nearly vertical. The correct
conversion chart must be used to determine contact angle.
(n) Record the contact angle for each location and the flight hours in the airplane log book.
(o) If contact angle is less than 50°, refurbish the windshield coating using the PPG manufactured kit.
DROP DIMENSION VS CONTACT ANGLE - DROP DIMENSION VS CONTACT ANGLE -
LOCATIONS A and C LOCATIONS B and D
Drop Length (Millimeter) Contact Angle Drop Length (Millimeter) Contact Angle
3.20 101° 3.20 100°
3.25 100° 3.25 99°
3.30 99° 3.30 98°
3.35 97° 3.35 97°
3.40 96° 3.40 96°
3.45 95° 3.45 94°
3.50 93° 3.50 93°
3.55 92° 3.55 92°
3.60 91° 3.60 91°
3.65 89° 3.65 90°
3.70 88° 3.70 88°
3.75 87° 3.75 87°
3.80 85° 3.80 86°
3.85 84° 3.85 85°
3.90 83° 3.90 84°
3.95 81° 3.95 83°
4.00 80° 4.00 81°
4.05 79° 4.05 80°
4.10 77° 4.10 79°
4.15 76° 4.15 78°
4.20 75° 4.20 77°
4.25 73° 4.25 75°
4.30 72° 4.30 74°
4.35 71° 4.35 73°
4.40 69° 4.40 72°
4.45 68° 4.45 71°
4.50 67° 4.50 70°
4.55 65° 4.55 68°
4.60 64° 4.60 67°
4.65 63° 4.65 66°
(3) If delamination is found in the critical vision area, it is necessary to replace the windshield or window before a subsequent
flight.
END OF TASK
7. Optical Prism Inspection
A. Description
(1) This optical inspection procedure will find voids and cracks in the area of the fastener holes of the acrylic windows without
the removal of the edge retainers or their related fasteners.
(2) Do an inspection for cracks that start at the fastener hole(s) and go to adjacent fastener holes, into the viewable area, or to
the edge of the window.
NOTE: An optical prism can be purchased or locally fabricated. Refer to Figure 606 for information on how
to make the optical prism.
B. Preparation
(1) Remove the windshield deicing ducts.
NOTE: If the AWR P-17 prism is used, the removal of the windshield deicing ducts is not required.
CAUTION: The use of cleaning materials not shown in the equipment table below can cause crazing of the acrylic
windows.
(2) Use aliphatic naphtha followed by a solution of liquid soap and water solution to thoroughly clean dust and unwanted
material from the window. Clean the acrylic window area a minimum of six to eight inches from the fastener holes.
C. Inspection Procedure with a 70-Degree Prism (Refer to Figure 607).
(1) Apply the couplant to the 70-degree face of the prism and the inspection area of the window.
NOTE: The inspections are done from the outside surface of the windows.
(2) Put the prism on the window, refer to Figure 607.
(3) Use the light source to add light at an angle of 30 to 60 degrees from the vertical of the prism and examine the fastener
holes.
NOTE: To get a clear view of both the top and the bottom surfaces of the fastener hole, move the prism
toward, or away from the fastener.
(a) The image of an undamaged hole will show as a cylinder that is not transparent.
(b) The image of a fastener hole with a crack that extends from one surface of the material under inspection into the hole
will show as a reflection. The reflection is not transparent and it extends from the fastener hole as in View A-A of
Figure 607.
(c) The image of a crack from one fastener hole to an adjacent fastener hole will show as an irregular surface that is not
transparent. View B-B of Figure 607 shows a crack from hole to hole.
D. Procedure Using an AWR P-17 Prism (Refer to Figure 608).
(1) Put the prism to the window as shown in Figure 608, and with the light source to add light at an angle of 30 to 60 degrees
from the vertical of the prism, examine the fastener holes.
E. After the inspection is completed, remove the couplant from the window with aliphatic naphtha followed by a weak soap and
water solution.
TASK 56-11-03-221
8. Glass Windshield and Side Windows Hump or Fillet Seals Detailed Inspection
A. Complete a Detailed Inspection of the Glass Windshield and Side Windows Hump or Fillet Seals.
(1) Inspect the windshield hump seal for cracks.
(2) If cracks are detected, repair immediately to prevent moisture entry and delamination of the windshield. Refer to Chapter
56, Electric Heated Glass Windshield - Maintenance Practices, Hump Seal Repair.
(3) Also check for security of installation, cleanliness, and other damage.
(4) Integrity of the hump seal and its bond to the glass surface is critical to the life of the windshield. If debonding is noted, the
depth of the debonded hump seal must be monitored closely.
CAUTION: Do not use a metal ruler, feeler gage or other rigid tool to measure depth of debonding. A rigid tool
could actually cause more debonding.
(5) Measure depth of debonding by inserting a 0.50 inch wide strip of paper into the gap between the hump seal and the glass
surface.
(6) If paper can be inserted 0.20 inch or more, the hump seal must be repaired immediately. Refer to Chapter 56, Electric
Heated Glass Windshield - Maintenance Practices, Hump Seal Repair.
(7) Lesser debonding must be monitored closely and inspected frequently to ensure repair is performed before permanent
damage, from moisture entering the plies, to the windshield occurs.
(8) Inspect the side window fillet seals for cracks, security of installation and other damage.
END OF TASK