Professional Documents
Culture Documents
Dr Hmwe Nantthar
Professor
Department of Engineering Chemistry
Yangon Technological University
26.12.2022
1
(1) Colorimetry
are made
• Colorimetric measurements using
a white light source
pass
through a colour filter (or) alternative wavelength selection device
pass
• incident light through a currette containing a coloured
chemical compound in solution
• the loss of light or absorption α to the concentration of
compound
2
colorimetry measure visible region of the electromagnetic
spectrum (e.g. 370-780 nm)
3
d Thickness
Concentration of absorbing
C
sample, C moles/dm3
I0 = Ia + It
A = log I0 /It = εcd
4
Beer-Lambert’s Law
A = εcd
A αC
A = absorbance of the coloured solution
C = concentration of the substance in solution (mol/dm3)
ε = molar extinction coefficient
d = path length
A = log 1/T
= log I0 / It
A = absorbance (A) or optical density OD
5
Figure 9.1 Schematic working of a calorimeter
6
7
do not absorb (or)
In colorimetry coloured solution very low absorbance
visible region
8
9
10
2. Potentiometry
measured
potential electrochemical cell (no current flow)
potential α concentration of same component of the
analyte
one is reference electrode
two half cells other indicator electrode
15
Fig 9.4 Endpoint curves of potentiometric titrations (a) potential Vs
volume of titrant (b) first differential plot (c) second differential plot
16
❖ Direct plot of the potential E against the volume of titrant gives
the fig. 9.4 (a).
The endpoint of the titration is where there is an abrupt change in
the potential value at the indicator electrode.
Determination of pH of a solution
The pH observed through the ion selective electrode (ISE) is
directly proportional to the concentration of the H+ ion concentration
of the analyte solution.
E 0 ̶ E 0
pH = G SCE ̶ E cell
0.0591
18
Other Applications
19
Advantages of Potentiometric Titrations
To determine
• Turbid
• Coloured
• Very dilute solutions
Advantages
• Very sharp end point
• Ease of automation
• A large number of commercial rigs are available
• Accurate methods for many electroanalytical processes
20
(3) Conductometry
application chemical and biochemical studies
conductometer measuring the conductivity of an ionic solution
migrate
anion to the anode
cation to the cathode
source high frequency
Solution Conductance
to measure conductance of ionic solution rather than
its resistance
21
Ohm’s Law
V = IR
V = potential difference (V)
I = electric current (A)
R = resistance (Ω)
R = ρ l/A
ρ = proportionality constant (specific resistance)
Ƙ = 1/ρ
Ƙ = specific conductance
22
Molar Conductance
μ0+ and μ0- are the ionic mobilities of cation and anion
Where mobilities are expressed in terms of m/second under
unit potential gradient.
1 Faraday = 96500 Coulomb
u0± = λ0 ± = λ0 ±
F 96500 24
Applications of Conductometry
Conductometry used direct measurement in titration
Uses; (1) monitoring surface water
(2) waterworks
(3) water desalination plants
(4) preparation of ultra pure water
Conductivity detection used precipitation titration
Applications
▪ degree of dissociation of weak electrolyte
▪ Ionic product of water
▪ solubility's and solubility product of sparingly soluble salts and
▪ acid-base and precipitation titrations
electrical conductance depend (i) number of ions
(ii) mobility of ions
25
Conductometric titration curve
27
Fig. 9.7 Titration of weak acid Vs NaOH
28
Consider the titration of solution of a weak acid, such as acetic
acid CH3COOH, using a solution of strong base, NaOH. The weak acids,
as well as other weak electrolytes, are dissociated into very small
extent and they exist in solution essentially in form of the neutral acid
molecules. When a solution of NaOH is added the reaction occurs the
undissociated molecules of acetic acid are transformed into dissociated
molecules of potassium acetate. The changes are accompained by
increase in conductivity of the solution.
29
Fig. 9.8 Conductometric titration curve for the hydrochloric acid-
acetic acid mixture titrated using solution of sodium hydroxide
30
The method of conductometric titration is adapted to
the estimation of mixtures of acids of differing strengths.
When a mixture of strong and weak acid is titrated a plot of
conductance against alkali added takes form as Fig 9.8 (c). The
conductometric titration curve is a combination of the
diagrams obtained during the titration of strong and weak
acid respectively, where the first end point correspond to a
neutralisation of the strong acid present in the sample and
the second one is associated with a neutralisation of the weak
acid in the solution under investigation. The volume of the
alkali consumed by the latter is given by a difference of the
respective volumes.
31
32
(4) Flame Photometry
flame photometry atomic emission method
detect metal salts Na, K, Li, Ca and Ba
Instrumentation
consists (1) burner
(2) nebuliser
(3) monochromator
(4) detector
(5) recorder
flame photometry sample and inexpensiveused
for clinical biological environmental
analysis
33
Fig. 9.9 Flow diagram of a flame photometer
34
Applications
used to determine metals in solution (alkali and
alkaline earth) in the given samples
➢ Like described for atomic absorption.
➢ Qualitative determination is also possible as each element emits its
own characteristic line spectrum.
Disadvantages
➢ very sensitive to change
➢ spectral interference
➢ self-absorption
➢ affect the precision of the measurement
35
36
(5) Atomic Absorption Spectroscopy (AAS)
measured absorption of optical radiation by atoms in the
gaseous state
atoms absorb visible and ultraviolet region (excited state)
absorbed
wavelength of light electrons go to excited state
measure
characteristics of the sample
37
Fig 9.10 Component of an atomic absorption spectrophotometer
38
Three basic components for every AA spectrophotometer
( i) Light Source
39
(ii) Sample Cell
❖ burner system (flame AAS)
❖ electrically heated furnace or platform
❖ aligned in the optical path of the spectrophotometer
40
Applications
✓ quantitative determination of metals at trance levels (0.1 to 100
ppm)
✓ used to Beer-Lambert’s Law
✓ Standard curve (absorbance Vs concentration)
✓ Plot give the linear range
✓ To know the concentration from the plot
Disadvantages
Sample must be solution at least volatile
needed
individual source lamp and filters for each elements
every metal own characteristic absorption
41
Fig Atomic absorption spectroscopy
42
(6) Flurometry
measure amount of fluorescent radiation
produced
to monochromatic radiation
Principle
supplied
adequate energy electrons excited to higher energy
levels of the atom fall different lower energy states
back
emission
before reading to the actual ground state
energy
Applications
benzene
limited quantitative
estimations
fused benzene ring system
43
❖ inorganic metals (to form complexes with the ligands)
44
Figure Fluorometry
45
Figure Fluorometry 46
Figure: Fluorometry
47
(7) Ultraviolet Visible Spectroscopy (UV/Vis)
Principles
200- 700 nm
Apply Beer-Lambert Law
A = εbc
48
(i) Electronic Transitions
51
Fig 9.12 Possible electronic transitions in atom or molecule
52
1. σ σ* Transitions
53
2. n σ* Transitions
▪ Saturated compounds containing atoms with lone pairs (non-
bonding electrons) are capable of n σ* transitions.
▪ These transitions usually need less energy than σ σ*
transitions.
▪ Wavelength is in the range 150-250 nm.
▪ The number of organic functional groups with n σ* peaks
in the UV region is small.
54
3. n π* and π π* Transitions
55
Charge-Transfer Absorption
❖ Many inorganic species show charge-transfer absorption and are
called charge-transfer complex.
❖ eg., one is electron donating properties and another is electron
acceptor properties
electron from donor
absorption of radiation involves
transfer orbital associated with acceptor
56
Instrumentation
Two types of instruments
(1) single beam
(2) double beam (widely used)
57
Fig 9.13 A single beam UV/VIS Spectrophotometer
Ratio Output
59
Sources of Visible Radiation
60
Wavelength Selector (Monochromator)
61
Applications
Disadvantages
❑ should be in solution
❑ interference sometimes
❑ difficult to analyses (mixture)
❑ disturbances
62
Figure UV/VIS Spectrophotometer
63
(8) Infrared Spectroscopy (IR)
Region wave number range (cm-1)
wave number = 1/ wavelength(cm)
Near 1200-4000
Middle 4000-200
Far 200-10
most useful IR region 4000-670 cm-1
Theory of Infrared Absorption
molecule absorbs electromagnetic radiation in the form of
stretching, bending, etc
(a) Molecular Rotations
Rotational levels quantized
gives line spectra
liquid or solid line broaden (due to molecular
collisions and other interactions)
64
(b) Molecular vibrations
stretching and bending
stretching symmetric
asymmetric
65
Fig 9.15 Symmetric and aymmetric stretching vibrations
66
Fig 9.15 Symmetric and aymmetric stretching vibrations
67
Bending: Change in angle between two bonds.
(1) Rocking
(2) Scissoring
(3) Wagging
(4) Twisting
68
Fig 9.16 Bending vibrations of a molecule
69
Fig Bending vibrations of a molecule
70
71
72
Vibrational Coupling
If the vibrating bonds are joined to a single, central atom
(1) Strong coupling of stretching vibrations occurs when there is a
common atom between the two vibrating bonds
(2) Coupling of bending vibrations occurs when there is a common
bond between the vibrating groups
(3) Coupling between a stretching vibration and a bending vibration
occurs if the stretching bond is one side of an angle varied by
bending vibration
(4) Coupling is greatest when the coupled groups have approximately
equal energies
(5) No coupling is seen between groups separated by two or more
bonds
73
Instrumentation
A double beam IR spectrophotometer consists of the following
IR source
grating monochromator
thermocouple detector
cells made of either sodium, chloride or potassium bromide
materials, etc.
74
Applications:
75
76
77
78
(9) Microwave Spectroscopy
This technique is an extension to IR spectroscopy.
The microwave region lies at the far infrared region of the
electromagnetic spectrum.
Its absorption by molecules give rise to changes in the rotational
energies of the molecules.
In IR spectroscopy, the molecule are subjected for changes in
vibrational energies; the energy required for making changes at
rotational levels is lesser than for vibrational levels.
Through the principles are same as that of IR, the instrumentation is
slightly different in that it requires samples in gaseous state for the
analysis.
79
Its application are limited to smaller and simpler molecules since,
larger molecules will have interactions between the rotational
energy levels within the molecule through various bonds they
have.
In certain cases this technique can be considered as a good
alternate to IR spectroscopy.
Besides qualitative analysis, this technique can be applied for
conformational analysis of simpler compounds.
80
81
82
83
(10) Nuclear Magnetic Resonance Spectroscopy (NMR)
Principles
UV, IR and microwave techniques utilise the absorption of light
in electronic states as well as in molecular vibrational and
rotational levels.
In NMR, the absorption by substances occurs in the radio
frequency region of the electromagnetic spectrum resulting in
changes in the orientation of the spinning nuclei by applying a
magnetic field.
The nuclei of the atoms bonded to each other in molecules spin
on an axis like a top.
Since nuclei are positively charged, this spin will create a small
magnetic field around it.
84
If an external magnetic field applied to these nuclei, then the spin
of the nuclei will align to the magnetic field deviating from the
original axis of spin.
If radio waves are applied to the system, the nuclei will absorb this
energy and re-align back to their original axis of spin.
The nature of the compounds and the presence of various
functional groups and their environment will give.
Since this technique mostly measures the spinning of the hydrogen
nuclei (almost all the organic compounds contain hydrogen
atoms).
It is referred as Proton Magnetic Resonance (PMR) spectroscopy.
85
Instrumentation
• Powerful magnet
• Radiofrequency signal generator
• Amplifier
• Detector, etc,.
86
87
88
89
Applications
90
Disadvantages
91
92
Figure; NMR Spectroscopy 93
(11) Electron Spin Resonance Spectroscopy (ESR)
Electrons always have a spin and thus have a magnetic moment.
This technique is of high value when it comes to the compounds
which contain odd electrons.
Which have paramagnetic behaviour, if it is unpaired then it can
align with the applied magnetic field and the feasibility of
getting ESR spectra is higher.
Thus, the principle and the instrumentation are much similar to
that of NMR technique.
It is also referred as Electron Magnetic Resonance.
94
Figure: Electron Spin Resonance Spectroscopy (ESR) 95
(12) Chromatography
96
Based on the mobile phase,
97
98
Figure : HPLC
99
Figure : HPLC 100
Gas Liquid Chromatography (GLC)
In gas-liquid chromatography, it is the interaction between the
gaseous sample (mobile phase) and standard liquid (stationary
phase), which causes the separation of different molecular
constituents.
The stationary phase is either polar or non-polar liquid.
106
107
Figure; GC chromatography 108
109
(13) Differential Thermal Analysis (DTA)
measure temperature difference between a sample and a
reference material as a function of temperature as
they are heated or cooled or kept at a constant
temperature (isothermal).
110
Figure; DTA
111
112
Figure; DTA 113
114
115
(14) Differential Scanning Calorimeter (DSC)
• This technique is more or less similar to DTA except that it
measures the amount of heat absorbed or released by a sample as
it is heated or cooled or kept at constant temperature (isothermal).
• It is the detection of liquid crystal.
• This technique is applied to detect the heat of melting and melting
point of thermoplastic polymers.
• It provides information regarding the molecular weight and
structural differences between very similar materials.
• The instrumentation is exactly similar to that of DTA except for the
different in the manipulation of the results.
116
➢It Physical properties include mass, temperature, enthalpy,
dimension, dynamic characteristics and others.
➢ Its application is finding the purity, integrity, crystallinity, thermal
stability of the chemical substances under study.
➢It is used to determine the composition of complex mixtures.
➢It is applied in wide fields, including polymer, glass, ceramics,
metals, explosives, semiconductor, medicine, etc.
117
Applications:
• characteristic temperature identification
• melting and crystallisation behaviour
• purity
• polymorphism
• specific heat capacity
• heat of reaction
• oxidative stability
• glass transitions
• heat of melting and crystallisations
• compatibility
• solid liquid ratio
• Reaction behaviour
• Reaction kinetics
• Thermal stability 118
9.6.11 Thermogravimetric Analysis (TGA)
• The change in sample weight is measured while the sample is
heated or cooled at a constant rate and so, the technique is
known as thermogravimetric analysis (TGA).
• This technique is effective for quantitative analysis of thermal
reactions that are accompanied by mass change such as
evaporation, decomposition, gas absorption, desorption and
dehydration under normal as well under different gaseous
environment as desired.
119
Figure; TGA 120
Figure; TGA 121
Figure; TGA 122
123
124
125
Figure; TGA 126
Figure; TGA 127
9.6.12 Cyclic Voltametry
Ability
(1) Information on the thermodynamics of redox processes
(2) Kinetics of heterogeneous electron-transfer reactions
(3) On coupled chemical reactions or adsorption processes
128
129
130
131
In cyclic voltametry have always three electrode functions.
132
133
134
135
136
(1) Absorbance of light by a solution of substance depends on
i. path length ii. concentration of solution
iii. wavelength of incident light iv. all of these
(4) The most widely used method for endpoint determination in titrimetric
analysis is
i. Spectrophotometry ii. Potentiometry iii. Amperometry
iv. conductometry
137
(5) In microwave spectrocsopy, samples must be in
i. liquid state ii. Solid state iii. Gaseous state iv. Solution
138
(9) In vltraviolet visible spectroscopy, the concentration of an analyte in
solution can be determined by applying
i. Nernst equation ii. Ohm’s law iii. Beer-Lambert’ law
iv. none of these
139
(13) Which method(s) can be used for the end point determination of
acid and base reaction?
i. potentiometry ii. conductometry
iii. spectrophotometry iv. all of these
(14) The hollow cathode lamps are used as the light source of ----------.
i. Flame photometer ii. Atomic absorption
spectrophotometer iii. UV-Vis spectrophotometer iv. none of these
(23) In GLC, using a polar stationary phase, the elution times of polar
compounds from a mixture are _______ those of non-polar
compounds.
i. longer than ii. shorter than iii. the same as iv. all of
these
143
1. Construct the electrochemical cell for potentiometry.
2. Write the disadvantages of the following
i. Atomic Absorption Spectroscopy ii. IR spectroscopy
3.Explain the vibrational coupling occurred in molecules by the
absorption of IR radiation.
4. Write the instrumentation and applications of NMR spectroscopy.
5. Write a note on microwave spectroscopy.
6. Explain the different types of vibration of a molecule with suitable
representations.
7. Write the applications and disadvantages of IR spectroscopy.
8. What is cyclic voltametry? Write the electrodes and their functions
in cyclic voltametry.
9. Write the instrumentation of IR spectroscopy.
10. What are the types of electronic transition and molecular
vibrations?
11. What is chromatography? Write two differences between HPLC
and GLC techniques.
12. Write a note on the thermogravimetric analysis. 144