Professional Documents
Culture Documents
Introduction
Electrolytic conductivity measures a solution’s ability to
carry an electric current.
A current is produced by applying a potential between two
inert metallic electrodes (e.g., platinum) inserted into the
solution being tested.
The change of electrolyte conductivity can indicate the
change of the concentration of the total ions,
Defined as conductometric Analysis
2
Principles
Ohm’s Law states that the current I (amperes) flowing in a conductor
is directly proportional to the applied electromotive force E (volts) and
inversely proportional to the resistance R (ohms) of the coductor :
= E / R
The reciprocal of the resistance is termed the conductance; this is
measured in reciprocal ohms (-1) (Omega-1),
The resistance of a sample of homogenous material, length L and
cross-section area a is given by :
R = Rho . L / a
Where is a characteristic property of the material termed the resistivity (specific
resistance). 3
The reciprocal of resistivity is the conductivity, Kappa ( specific
conductance)
Which is the conductance of a one meter cube of substance and has the units -1 m-
4
The molar conductivity ( m ) (Lambda) is the conductivity per unit
= / C
= 1000 / C
• Where C is the conc expressed in terms of normality of the solution, unit of is -1 cm2 eq-1.
5
The conductivity of electrolyte solutions is equal to the sum of
the conductivities of each type of ion present.
6
The equivalent conductance of salts or ions increases as the
concentration decreases.
At infinite dilution there are no disturbing effects on the mobilities
of the ions other than variations in solvent and temperature, and the
equivalent conductance reaches its maximum value.
The equation may be rewritten as:
0 = +0 + -0
Where 0 = the equivalent conductance of the electrolyte at infinite dilution.
+0 = the limiting ionic equivalent conductance of the cation at infinite dilution.
-0 = the limiting ionic equivalent conductance of the anion at infinite dilution.
7
Conc.of NaCl 0.1 0.01 0.001 infinite dilution
Λ 106.7 118.5 123.7 126.4
Increasing direction
8
Table: Values of the limiting ionic molar conductivities for some ions in water at 25 C
α = Λ / Λo
For a weakly ionized substance, Λ varies much more markedly with
conc. b/c the degree of ionization, α, varies strongly with conc.
10
The equivalent conductance, however, must approach a
constant finite value at infinite dilution, Λ0, which again
corresponds to the sum of the limiting ionic conductance’s.
11
The measurement of conductivity
12
The specific conductivity is the conductance of a cube of
material, 1 cm on each edge.
Since it is difficult to build a cell with well defined
geometrical parameters A & l, any cell should be calibrated
with a solution of exactly known specific conductance,
The cell factor or cell constant, c
c =R =l/A= length/electrode area
The cell constant, c, with unit’s m-1, is determined by
measuring the resistance of a cell filled with a solution of
known conductivity, which is invariably KCl (0.002M).
13
Thus if c=1, the observed resistance is equal
numerically to the reciprocal of the specific
conductance of the solution used .
Once c has been determined for a cell, then the
measurement of the resistance of any other solution
will provide values of Λ or .
14
Conductometric Titrations
Rate of change of conductance as a function of added titrant used to
determine the equivalence point.
Note that:
16
Applications...
17
Applications of the conductometric titrations
1- Strong acid with a strong base :
The conductance first falls, due to the replacement of the hydrogen ions ( 350) by the added
cation ions ( 40 - 80) and then, after the equivalence point has been reached, rapidly rises with
further additions of strong alkali due to the large value of the hydroxyl ions ( 198 ). The two
branches of the curve are straight lines provided that the volume of the reagent added is
negligible, and their intersection gives the end point.
BA salt
ml of the base
e.p. 18
2- Strong acid with a weak base :
The titration of a strong acid with a moderately weak base may be illustrated by the
neutralization of dilute sulfuric acid by dilute ammonia solution. The first branch of the
graph reflects the disappearance of hydrogen ions during the neutralization, but after
the end point has been reached the graph becomes almost horizontal, since the excess
aqueous ammonia is not significantly ionized in the presence of ammonium sulfate.
ml of base
e.p.
19
NaOH
3- Weak acid with a strong base:
The shape of the curve will depend upon the conc. Acetic acid
Conductance
and ionization of the acid. Thus the neutralization 0.
1
of the acetic acid with NaOH solution, the salt which 0. M
0 1
is formed during the first part of the titration tends M
0.
to repress ionization of acetic acid still present so 00
1
M
that its conductance decreases. The rising salt conc
(w/c is ionized) will, however, tend to produce an ml of base
increase in the conductance. AC- + H2O = HAC +
OH-
In consequence of these opposing influences the
titration curves may have minima, the position of
Conductance
which will depend upon the conc. and strength of the
weak acid. As the titration proceeds, a somewhat
indefinite break will occur at the end point, and the
graph will become linear after all the acid has been
neutralized.
Generally titrations of weak and moderately strong
acids with a strong base are suitable for ml of base
conductimetric techniques. 20
4- Weak acids with weak bases:
Titration of a weak acid with a weak base
can be readily carried out, and frequently it
is preferable to use this procedure rather
than use a strong base.
The neutralization curve up to the
conductanc
equivalence point is similar to that
obtained with NaOH since both Na &
NH3
amm. acetates are strong electrolytes, after
the e.p. an excess of aqueous ammonia
solution has little effect upon the
conductance as its ionisation is depressed
by the ammonia salt present in the soln.
The advantages over the use of strong ml of base
alkali are that the e.p. is easier to detect.
21
5-Mixture of a strong acid & a weak acid with a NaOH
HCl
strong base :
Upon adding a strong base to a mixture of a strong
acid and a weak acid (e.g. HCl & CH3COOH), the
conductance falls until the strong acid is neutralised, CH 3COOH
the rises as the weak acid is converted to its salt and
finally rises more steeply as excess alkali is
introduced.
Note that;
1- Increasing ionisation of the weak acid results in a
rounding-off at the first e.p. (HCl)
2- hydrolysis of the salt of the weak acid causes a
rounding-off at the second e.p. (CH3COOH)
HCl
3- Extrapolation of the straight portions of the three
branches leads to definite location of the end points.
22
Coulometry
23
Coulometric Methods of Analysis
Measures quantity of electricity or the charge generated by a
reaction at an electrode.
Coulometry is a term derived from “Coulomb” that units
used for measuring quantity of electricity
Coulometric analysis is an application of Faraday’s First
Law of Electrolysis which states that:
The current or charge passed through an electrode in a redox
reaction is proportional to the moles of the reaction’s
reactants and products.
24
Faraday’s law:
Total charge, Q, in coulombs passed during electrolysis is related to
the absolute amount of analyte:
Q nFN
QM
m
Fn
Q = It (constant current)
Q = ∫ I dt (controlled potential)
equations
26
Two types of coulometric methods
27
1. Controlled-Potential Coulometry
28
Background current
31
Fundamental requirements of a coulometric titration are:
32
Advantages of coulometric titrations:
2- Unstable reagents such as Br2, Ag+ and Ti3+ can be used safely as standards
(They are generated and consumed immediately)
3- The method is ideally applied for micro and semi-micro scale titrations
without the well known difficulties involved in the standardization and
storage of very dilute solutions.
34
* Using a constant current does present two important
experimental problems that must be solved if accurate results are
to be obtained.
First, as electrolysis occurs the analyte's concentration and,
therefore, the current due to its oxidation or reduction steadily
decreases.
To maintain a constant current the cell potential must change until
35
Second problem is the need for a method of determining
when the analyte has been exhaustively electrolyzed.
In controlled‑potential coulometry this is signaled by a decrease in the
36
Methods for detection of end points in coulometric titrations:
37
Application of Coulometric Method
Used for the determination of inorganic cations and anions,
including trace metals and halides,
It also can be applied to the quantitative analysis of organic
compounds, like trichloroacetic acid
Cl3CCOO–(aq) + H3O+(aq) + 2e– ↔ Cl2HCCOO–(aq) + Cl–(aq) + H2O(l)
m = QM
nF
m = = 9.87 x 10-2 g
39
2. Suppose that the ff cell, producing a constant
voltage of 1.02V, is used to operate a light bulb
with a resistance of 2.8Ω.
Zn(s)|Zn2+(aq)||Cu2+(aq)|Cu(s)
How many hours are required for 5.0g of Zn to be
consumed? (MWZn=65.4)
Solution
The anode rxn is: Zn(s) → Zn 2+ + 2e-
Q= nFN = 2×96485×5/65.4 =1.48×104C
The current flowing via the circuit is:
I=E/R =1.02V/2.8Ω = 0.364A
Time: t = Q/I = 1.48×104C/0.364A
= 4.06×104s = 11.3h
40
3. A 0.3619-g sample of tetrachloropicolinic acid,
C6HNO2Cl4 (MWt= 261), is dissolved in distilled
water, transferred to a 1000-mL volumetric flask,
and diluted to volume. An exhaustive controlled-
potential electrolysis of a 10.00-mL portion of this
solution at a spongy silver cathode requires 5.374 C
of charge. What is the value of n (moles of elecrons)
for this reduction reaction?
Solution
The 10.00-mL portion of sample contains 3.619 mg,
i.e N = 1.39 × 10–5 mol of tetrachloropicolinic acid.
41
42
To determine the purity of a sample of Na 2S2O3, a
sample is titrated coulometrically using I – as a mediator
and I3– as the titrant. A sample weighing 0.1342 g is
transferred to a 100-mL volumetric flask and diluted to
volume with distilled water. A 10.00-mL portion is
transferred to an electrochemical cell along with 25 mL
of 1 M KI, 75 mL of a pH 7.0 phosphate buffer, and
several drops of a starch indicator solution. Electrolysis
at a constant current of 36.45 mA requires 221.8 s to
reach the starch indicator endpoint. Determine the
sample’s purity.
43
44
45
Polarography
46
VOLTAMMETRY
47
VOLTAMMETRY…
General principles:
Id = kC
Where Id = Diffusion current,
k = constant,
C = conc. of analyte
48
A simple classification of voltammetric methods is
based on the electrode type.
Polarography - voltammetric technique that uses
dropping mecrury electrode (DME) as working
electrode.
49
Polarography
A form of voltammetry using a dropping mercury electrode.
50
Instrumentation
It is usually composed of three electrodes:
51
R
SCE DME
P
V
Counter
Electrode Fig. Polarographic apparatus
DME - Renewable surface
53
Advantages of dropping mercuric electrode
The surface area can be calculated from the weight of the drop.
Disadvantages of dropping mercuric electrode
55
Supporting electrolyte:
It is a solution of indifferent electrolyte ( base or ground electrolyte)
which by itself don’t oxidized or reduced at working electrode at the
selected potential range.
It raises the conductivity of the solution and carry the bulk of the
current so prevent the migration of electro-active materials (sample
material to be determined) to the working electrode.
It may control or (buffer) the level of hydrogen ion activity in
solution.
Atypical current-voltage curve (polarogram) is shown in the following figure:
From A to B, practically no current will pass through the cell.
At B, when the potential of the working electrode (DME) is equal to the deposition
potential of the cadmium ions with respect to metallic cadmium electrode, the current
suddenly increases and the working electrode becomes depolarized by the cadmium
ions, which are then discharged upon the electrode surface to form metallic cadmium;
consequently a rapid increase in the current flowing through the cell occur.
At point C, the current no longer increase linearly with the applied potential but
approaches a steady limiting value.
At point D no increase in current is observed at higher cathode potential unless a
second compound ,able to depolarize the working electrode , present in the solution.
Decomposition potential
rrent
Current (microampers)
C Limiting cu
D
Diffusion current
si d ual c urrent
Re B
A
Half-wave potential Applied potential (volts)
Note that:
The distance between B and C usually known as Polarographic wave.
diffusion from the main bulk of solution always exceeds the number of cadmium ions
which reacted and deposited on the cathode surface.
At the point C the rate of supply of cadmium ions from the main bulk of the
solution to the working electrode has become equal to the rate of their
deposition and no further increase in current passing through the cell is expected
with increasing the potential.
The polarogram is characterized by different parameters :
1-Residual current: at small applied potential only a residual current flow in the cell which
caused by the presence of impurities in the supporting electrolyte or sample.
2-Limiting current: is the region on the polarogram in which the current after increased
sharply , becomes essentially independent on the applied potential .
3-Diffusion current: it is the difference between limiting and residual currents . It is directly
proportional to the concentration of the reactive substance.
4- Half wave potential (E1/2) : is the potential at which the current is equal to one half the
diffusion current. It is a characteristic of the nature of the reactive material, independent on
concentration, depend on pH of the medium, type of supporting electrolyte and type of
electrode.
5- Decomposition potential: is the potential at which the substance begin to oxidize or reduce.
Diffusion current
When the potential applied to a polarographic cell exceeds the decomposition potential of an
electrolyte species, its concentration at the surface of the electrode (d.m.e.) is immediately
diminished.
A conc gradient is therefore established and more of that species diffused from the bulk
solution to the electrode surface (Fick’s law of diffusion) .
The resulting current flow is proportional to the rate of diffusion
i ( C - Co) or i = K (C - Co)
where C and Co are the conc of the electro-active species in the bulk solution and at the surface of
electrode respectively.
By increasing the applied voltage, reduction occurs more rapidly and (Co) becomes zero and the
conc gradient reaches a maximum.
At this point, the rate of the diffusion and the current flowing in the cell reach to their limiting
values.
i = KC Figure
Ilkovic Equation:
It is the equation that relates the various parameters that determine the
magnitude of the diffusion current obtained with a dropping mercury electrode
All the positively charged ions present in the solution will be attracted to the negative
working electrode by:
- Electrical force; due to the attraction of oppositely charged bodies to each other.
- Diffusive force; arising from the concentration gradient produced at the electrode
surface.
The total current passing through the cell can be regarded as the sum of these two
factors.
the following reaction will occur during the reduction process at the surface of
cathode
Cd+2 + 2e + Hg0 == Cd0(Hg0)
Potential applied on the working electrode is usually swept over (i.e. scan)
a pre-defined range of applied potential
0.001 M Cd2+ in 0.1 M KNO3 supporting electrolyte
Electrode become more and more
reducing and capable of reducing Cd 2+
All Cd2+ around the electrode has
Cd 2+
+ 2e Cd-
already been reduced. Current at
Current starts to be registered at the the electrode becomes limited by
i (A)
electrode the diffusion rate of Cd2+ from the
Working electrode is E½ bulk solution to the electrode.
no yet capable of Thus, current stops rising and
reducing Cd2+ levels off at a plateauid
Current at the working
only small residual
electrode continue to rise as
current flow through
the electrode become more
the electrode Base line
reducing and more Cd2+ around
the electrode are being of residual
reduced. Diffusion of Cd2+ current
does not limit the current yet
Useful also for stability studies and study the connection between
polarographic properties and physiological activities of drugs.
B-Quantitative application
Polarography ( voltammetry) can be used for determination of the
conc. of the electro-active compounds by direct measurements of the
diffusion current.
It can be used in combination with other technique e.g. HPLC to
determine the concentration of substances presents in minute
amounts especially in biological fluids.
66
Voltammetric analysis
Analyte selectivity is provided by the applied potential on the working electrode.
Pt working Ag counter
electrode at -1.0 electrode at
V vs SCE 0.0 V
X M of PbCl2
0.1M KCl
-1.0 V vs SCE Concentration gradient created
Pb2+ + 2e- Pb
between the surrounding of the
electrode and the bulk solution
K+
Pb2+ Pb2+ K+
K+ Pb 2+
Pb2+
K+ K+ Pb 2+
K+
K+ K+
Pb 2+
K
+
Pb2+
Pb2+ K+ Pb2+
K+ Pb2+ K+
K+
K+
Pb 2+
K +
K+ Pb2+ Pb 2+ K+ Pb2+
K+Pb2+ migrate to
the electrode
K+
K + via diffusion Pb2+
Pb 2+
Pb2+ K+
Pb2+
Pb2+
K+ Pb2+ Pb2+
K +
K+
K+ K+