You are on page 1of 3

Water Technologies & Solutions

application note

best practice for analyzing compendia water samples on the


Sievers* M9 Total Organic Carbon (TOC) Analyzer with conductivity
This document is intended to provide best practice guidance for both TOC and conductivity testing on the
guidance on how to measure simultaneous TOC and M9 is a minimum of four repetitions with one reject.
conductivity of compendia water samples on the Stage 1 conductivity testing uses raw conductivity and
Sievers M9 TOC Analyzer with conductivity. temperature data so selecting a compensating
algorithm is not required. The M9 will give a raw
sampling conductivity value, a temperature value, and a
temperature compensated value. For compendia
The critical components for successful implementation water samples, the raw conductivity value and
of compendia water testing on the M9 with conductivity temperature should be reported. Acceptance criteria
are: should be determined by Table 1 in USP <645>, Stage
1. Good sampling technique 1 – Temperature and Conductivity Requirements.
Based on the measured temperature, the
2. Use of a Dual Use Conductivity and TOC (DUCT) corresponding conductivity value is the limit for
vial compendia water samples. 1

According to USP <645>, “Stage 1 conductivity may be


performed offline in a suitable container.” A suitable
1

container for simultaneous testing is one that does not


contribute either TOC or conductivity when in contact
with the sample. Sievers DUCT vials, caps, and septa
have demonstrated that with good sampling technique,
there is no significant TOC or conductivity contribution
for up to five days.
2,3

Sievers DUCT vials are cleaned and certified to less Figure 1: Method Conditions
than 10 ppb TOC, eliminating the need for any vial pre-
Set the acid and oxidizer flow rates for compendia
rinse steps. Best practices for sampling are:
water samples below 500 ppb TOC to 1.0 µL/min acid
1. Do not rinse vials prior to use. and 0.0 µL/min oxidizer. These flow rates ensure
complete carbon oxidation in the UV reactor while
2. To avoid contamination, do not touch the inside
avoiding over-oxidation of the sample.
of the vial, cap, or septa.
An additional method consideration is the use of the
3. Fill the DUCT vial once, completely full, with no
optional Inorganic Carbon Remover (ICR). If the IC
headspace and avoid turbulent flow when
measurement is approximately ten times greater in
filling the vial.
magnitude or more than the TOC measurement, use
4. Immediately cap the vial. of the ICR is recommended as best practice to
5. Do not reuse DUCT vials. improve TOC measurement stability and precision. 4,5

In the event of a Stage I conductivity failure, proceed


instrument method conditions to Stage II testing per USP <645>.

When analyzing compendia water samples on the M9, it


is important to evaluate multiple repetitions of the
determining frequency of verification
sample for statistical robustness and for additional Challenging the instrument and method with known
measurement stability confidence. Best practice standards provides confidence in each measurement.
Verify the nominal performance of the method on the Table 1: Best Practice Consumables
instrument by using TOC system suitability and Product Description Part Number
conductivity verification standards on a frequency
determined via risk assessment. Standards passing the System Suitability Set STD 31004-01
acceptance criteria give a high degree of assurance that
100µS/cm KCl Conductivity STD 77050-01
the analysis of unknown water samples complies with
Verification
the compendia requirements for pharmaceutical grade
bulk water. 1,6
DUCT Vials for Dual TOC and HMI 77500-01
Conductivity Analysis
The compendia specify the periodic use of conductivity
verification and system suitability standards as a
method verification, though the frequency of Lastly, it is best practice to always leave the
verification is not specified. Frequency should be instrument with water in the flow path. Following the
determined by assessing the risk and potential impact last standard, a syringe flush should be performed
unique to each process. Risk management requires using DI or MilliQ water. This will purge the
clearly defining and assessing all variables and their instrument of all previous sample and replace it with
impact to the process. Factors to consider include water.
frequency of use, sample time, risk of system suitability
or verification failures, likelihood of out-of- troubleshooting/maintenance
specification (OOS) results, and/or time constraints. As Due to the method sensitivity and the low-level
specific frequencies are not prescribed by USP <643> acceptance criteria for offline Stage 1 conductivity
and <645>, it is the responsibility of each facility to have testing, many users have a dedicated Sievers M9 TOC
robust processes and procedures that manage risk Analyzer with conductivity for offline TOC and
unique to the process. It is important to establish the conductivity measurements. An M9 used for both
7

use of conductivity verification and system suitability compendia and non-compendia testing (e.g., cleaning
standards on a practical and defensible frequency while validation samples) may require additional steps to
simultaneously meeting minimum requirements of minimize carry-over between the two sample types.
USP <643> and <645>. These steps are unique to the types of non-
TOC system suitability standards verify the relative compendia samples being analyzed. For more
recovery capability of the instrument at the compendia information on such precautions please see Tech Tip
upper limit of 500 ppb TOC. System suitability ensures UPW 07-10. Instrumentation should be maintained
8

suitable recovery of TOC, providing confidence in the and verified per our operating and maintenance
ability to analyze unknown water samples. The manual for optimal performance.
conductivity verification standard verifies the accuracy
of conductivity measurements taken on the M9. conclusion
Conductivity verification standards performed at a Implementing proper sampling technique, method
justified frequency ensure accurate measurements of conditions, and justified verification frequencies will
unknown water samples while meeting compendia ensure a high degree of confidence in TOC and
requirements. conductivity measurements using the Sievers M9
To limit variation as much as possible, it is Analyzer. Execution of best practices outlined in this
recommended to use Sievers standards and vials from document will allow for simultaneous analysis of TOC
SUEZ as consistent and certified consumables. It is best and conductivity while meeting compendia specified
practice to use the Sievers system suitability and requirements.
conductivity verification standards shown in Table 1. References
When using the consumables in Table 1, companies 1. USP <645> Water Conductivity. Retrieved February 14, 2019 from
https://hmc.usp.org/sites/default/files/documents/HMC/GCs-
can take advantage of SUEZ’s OOS investigation Pdfs/c645.pdf
support. In the event of a sample, system suitability or 2. Sievers Lean Lab: Simultaneous Stage 1 Conductivity and TOC Lab
verification failure, the SUEZ quality team will Testing of Pharmaceutical Water (300 40030). Retrieved February 14,
2019 from
troubleshoot and thoroughly investigate inhouse https://www.suezwatertechnologies.com/kcpguest/documents/Applic
variables and onsite instrument performance, ation%20Notes_Cust/Americas/English/ANai_300_40030_EN.pdf
discussing the findings in the failure analysis report.

300 00124 EN Rev. C


3. DUCT Vial Performance and Stability (300 00297). Retrieved February 14,
2019 from
https://www.suezwatertechnologies.com/kcpguest/documents/Technica
l%20Bulletins_Cust/Americas/English/TBai_300_00297_EN.pdf
4. Reserve Sample Bottles for Conductivity and TOC (300 00299). Retrieved
February 14, 2019 from
https://www.suezwatertechnologies.com/kcpguest/documents/Technica
l%20Bulletins_Cust/Americas/English/TBai_300_00299_EN.pdf
5. Sievers Inorganic Carbon Remover (ICR) (300 00109). Retrieved February
14, 2019 from
https://www.suezwatertechnologies.com/kcpguest/documents/Applicati
on%20Notes_Cust/Americas/English/ANai_300_00109_EN.pdf
6. USP <643> Total Organic Carbon. Retrieved July 25, 2019 from
https://hmc.usp.org/sites/default/files/documents/HMC/GCs-
Pdfs/c643.pdf
7. Low Level Linearity Conductivity Study on the Sievers M9 TOC Analyzer
(300 00339). Retrieved February 14, 2019 from
https://www.suezwatertechnologies.com/kcpguest/documents/Applicati
on%20Notes_Cust/Americas/English/ANai_300_00339_EN.pdf
8. UPW 07-10 Multiple Products Biological Contamination (800 19025)
Retrieved February 14, 2019 from
https://www.suezwatertechnologies.com/kcpguest/documents/Technica
l%20Bulletins_Cust/Americas/English/ai_UPW_07-10_EN.pdf

Find a contact near you by visiting www.sieversinstruments.com and clicking on “Contact Us”.
*Trademark of SUEZ, may be registered in one or more countries.
©2021 SUEZ. All rights reserved. 300 00124 EN Rev. C

You might also like