Professional Documents
Culture Documents
ANai - 300 - 00124 - EN - Best Practice For Analyzing Compendia Water Samples
ANai - 300 - 00124 - EN - Best Practice For Analyzing Compendia Water Samples
application note
Sievers DUCT vials are cleaned and certified to less Figure 1: Method Conditions
than 10 ppb TOC, eliminating the need for any vial pre-
Set the acid and oxidizer flow rates for compendia
rinse steps. Best practices for sampling are:
water samples below 500 ppb TOC to 1.0 µL/min acid
1. Do not rinse vials prior to use. and 0.0 µL/min oxidizer. These flow rates ensure
complete carbon oxidation in the UV reactor while
2. To avoid contamination, do not touch the inside
avoiding over-oxidation of the sample.
of the vial, cap, or septa.
An additional method consideration is the use of the
3. Fill the DUCT vial once, completely full, with no
optional Inorganic Carbon Remover (ICR). If the IC
headspace and avoid turbulent flow when
measurement is approximately ten times greater in
filling the vial.
magnitude or more than the TOC measurement, use
4. Immediately cap the vial. of the ICR is recommended as best practice to
5. Do not reuse DUCT vials. improve TOC measurement stability and precision. 4,5
use of conductivity verification and system suitability compendia and non-compendia testing (e.g., cleaning
standards on a practical and defensible frequency while validation samples) may require additional steps to
simultaneously meeting minimum requirements of minimize carry-over between the two sample types.
USP <643> and <645>. These steps are unique to the types of non-
TOC system suitability standards verify the relative compendia samples being analyzed. For more
recovery capability of the instrument at the compendia information on such precautions please see Tech Tip
upper limit of 500 ppb TOC. System suitability ensures UPW 07-10. Instrumentation should be maintained
8
suitable recovery of TOC, providing confidence in the and verified per our operating and maintenance
ability to analyze unknown water samples. The manual for optimal performance.
conductivity verification standard verifies the accuracy
of conductivity measurements taken on the M9. conclusion
Conductivity verification standards performed at a Implementing proper sampling technique, method
justified frequency ensure accurate measurements of conditions, and justified verification frequencies will
unknown water samples while meeting compendia ensure a high degree of confidence in TOC and
requirements. conductivity measurements using the Sievers M9
To limit variation as much as possible, it is Analyzer. Execution of best practices outlined in this
recommended to use Sievers standards and vials from document will allow for simultaneous analysis of TOC
SUEZ as consistent and certified consumables. It is best and conductivity while meeting compendia specified
practice to use the Sievers system suitability and requirements.
conductivity verification standards shown in Table 1. References
When using the consumables in Table 1, companies 1. USP <645> Water Conductivity. Retrieved February 14, 2019 from
https://hmc.usp.org/sites/default/files/documents/HMC/GCs-
can take advantage of SUEZ’s OOS investigation Pdfs/c645.pdf
support. In the event of a sample, system suitability or 2. Sievers Lean Lab: Simultaneous Stage 1 Conductivity and TOC Lab
verification failure, the SUEZ quality team will Testing of Pharmaceutical Water (300 40030). Retrieved February 14,
2019 from
troubleshoot and thoroughly investigate inhouse https://www.suezwatertechnologies.com/kcpguest/documents/Applic
variables and onsite instrument performance, ation%20Notes_Cust/Americas/English/ANai_300_40030_EN.pdf
discussing the findings in the failure analysis report.
Find a contact near you by visiting www.sieversinstruments.com and clicking on “Contact Us”.
*Trademark of SUEZ, may be registered in one or more countries.
©2021 SUEZ. All rights reserved. 300 00124 EN Rev. C