application note
and Beverage Industry was product contamination from After determining cellular density, a Sievers* M9 TOC
microorganisms. A robust cleaning process is critical
1
Cell Density Raw TOC Std. Dev. Filtered TOC Std. Dev. Corrected TOC Std. Dev. %
(cells/ml) (ppb) (ppb) (ppb) (ppb) (ppb) (ppb) RSD
10,000
trend across the cellular densities measured. Based on
the linear fit trend line equation shown in Figure 3, the y = 0.0002x + 182.72
Level of Detection (LOD), defined as 3 times the noise, 1,000 R² = 0.9981
TOC (ppb)
ell-and-tissue-analysis/protocols/dapi-imaging-protocol.html
shows calculations for theoretical microbial TOC yield. 3. Phillips, Rob, and Ron Milo. “A Feeling for the Numbers in Biology.”
Proceedings of the National Academy of Sciences 106, no. 51
Based on the referenced carbon atoms per cell, the
(December 22, 2009): 21465.
theoretical TOC concentration for 5.8E+07 cells/mL is https://doi.org/10.1073/pnas.0907732106.
11.6 ppm.
Find a contact near you by visiting www.sieversinstruments.com and clicking on “Contact Us.” 300 40049 EN Rev. A
*Trademark of SUEZ, may be registered in one or more countries.