Professional Documents
Culture Documents
Micromass Quattro API Operators Guide
Micromass Quattro API Operators Guide
Trademarks
Waters and Micromass are registered trademarks of Waters Corporation, and
MassLynx, Quattro, and “THE SCIENCE OF WHAT’S POSSIBLE.” are
trademarks of Waters Corporation.
PEEK is a trademark of Victrex Corporation.
Rheodyne is a registered trademark of Rheodyne, L.P.
Swagelok is a registered trademark of Swagelok Inc.
Upchurch is a registered trademark of Scivex, Inc.
Windows is a registered trademark of Microsoft Corporation.
Other registered trademarks or trademarks are the sole property of their
owners.
ii
Customer comments
Waters’ Technical Communications department invites you to tell us of any
errors you encounter in this document or to suggest ideas for otherwise
improving it. Please help us better understand what you expect from our
documentation so that we can continuously improve its accuracy and
usability.
We seriously consider every customer comment we receive. You can reach us
at tech_comm@waters.com.
Contacting Waters
®
Contact Waters with enhancement requests or technical questions regarding
the use, transportation, removal, or disposal of any Waters product. You can
reach us via the Internet, telephone, or conventional mail.
Safety considerations
Some reagents and samples used with Waters instruments and devices can
pose chemical, biological, and radiological hazards. You must know the
potentially hazardous effects of all substances you work with. Always follow
iii
Good Laboratory Practice, and consult your organization’s safety
representative for guidance.
Warning:
• To confirm the integrity of the source exhaust system, renew
the source O-rings at intervals not exceeding one year.
• To avoid chemical degradation of the source O-rings, which can
withstand exposure only to certain solvents, see Appendix B to
determine whether any solvents you use that are not listed are
chemically compatible with the composition of the O-rings.
Never let the nitrogen supply pressure fall below 690 kPa (6.9 bar, 100 psi)
during analyses that require flammable solvents. Connect to the LC output
with a gas-fail connector to stop the LC solvent if the nitrogen supply fails.
iv
High temperature hazard
Source enclosure
assembly
v
High voltage hazard
Warning:
• To avoid electric shock, do not remove the mass spectrometer’s
protective panels. The components they cover are not
user-serviceable.
• To avoid nonlethal electric shock when the instrument is in Operate
mode, avoid touching the areas marked with the high voltage
warning symbol. To touch those areas, first put the instrument in
Standby mode.
High voltage
connectors
vi
Hazards associated with removing an instrument from service
When you remove the instrument from use to repair or dispose of it, you must
decontaminate all of its vacuum areas. These are the areas in which you can
expect to encounter the highest levels of contamination:
• Source interior
• Waste tubing
• Exhaust system
• Rotary pump oil (where applicable)
The need to decontaminate other vacuum areas of the instrument depends on
the kinds of samples the instrument analyzed and their levels of
concentration. Do not dispose of the instrument or return it to Waters for
repair until the authority responsible for approving its removal from the
premises specifies the extent of decontamination required and the level of
residual contamination permissible. That authority must also prescribe the
method of decontamination to be used and the appropriate protection for
personnel undertaking the decontamination process.
You must handle items such as syringes, fused silica lines, and borosilicate
tips used to carry sample into the source area in accordance with laboratory
procedures for contaminated vessels and sharps. To avoid contamination by
carcinogenic, toxic, or biohazardous substances, you must wear
chemical-resistant gloves when handling or disposing of used oil.
Safety advisories
Consult Appendix A for a comprehensive list of warning and caution
advisories.
vii
Operating this instrument
When operating this instrument, follow standard quality-control (QC)
procedures and the guidelines presented in this section.
Applicable symbols
Symbol Definition
Manufacturer
viii
Intended use of the Waters Quattro micro API
The Waters Quattro micro API is CE-marked according to the
European Union In Vitro Diagnostic Device Directive 98/79/EC.
Waters designed the Quattro micro API for use as a research tool to deliver
authenticated exact-mass measurement on all sample types.
The Waters Quattro micro API can be used for general in vitro diagnostic
applications. However, only professionally trained and qualified laboratory
personnel can use the instrument for those purposes.
Calibrating
To calibrate LC systems, follow acceptable calibration methods using at least
five standards to generate a standard curve. The concentration range for
standards must include the entire range of QC samples, typical specimens,
and atypical specimens.
When calibrating mass spectrometers, consult the calibration section of the
operator’s guide for the instrument you are calibrating. In cases where an
overview and maintenance guide, not operator’s guide, accompanies the
instrument, consult the instrument’s online Help system for calibration
instructions.
Quality-control
Routinely run three QC samples that represent subnormal, normal, and
above-normal levels of a compound. Ensure that QC sample results fall within
an acceptable range, and evaluate precision from day to day and run to run.
Data collected when QC samples are out of range might not be valid. Do not
report these data until you are certain that the instrument performs
satisfactorily.
ix
When analyzing samples from a complex matrix such as soil, tissue,
serum/plasma, whole blood, and other sources, note that the matrix
components can adversely affect LC/MS results, enhancing or suppressing
ionization. To minimize these matrix effects, Waters recommends you adopt
the following measures:
• Prior to the instrumental analysis, use appropriate sample
pretreatment such as protein precipitation, liquid/liquid extraction
(LLE), or solid phase extraction (SPE) to remove matrix interferences.
• Whenever possible, verify method accuracy and precision using
matrix-matched calibrators and QC samples.
• Use one or more internal standard compounds, preferably isotopically
labeled analytes.
ISM classification
x
EC authorized representative
Telephone: +44-161-946-2400
Fax: +44-161-946-2480
Contact: Quality manager
xi
xii
Table of Contents
Copyright notice ................................................................................................... ii
Trademarks ............................................................................................................ ii
Customer comments ............................................................................................ iii
Table of Contents xv
Analog data .................................................................................................... 4-15
xx Table of Contents
1 Instrument Description
Contents
Topic Page
Overview 1-2
Sample inlet 1-3
Vacuum system 1-3
Data system 1-3
MassLynx software 1-4
Theory and principles of operation 1-4
Front panel controls, indicators and connections 1-12
Rear panel connections 1-16
1-1
Overview
Removable
Center Panel
TM
The Quattro micro API is a high performance triple quadrupole mass
spectrometer designed for routine LC/MS/MS operation.
The instrument may be coupled to the following liquid introduction systems:
• HPLC system, to provide molecular weight information from an LC run
or to perform target analysis and quantification.
• Syringe pump, for analysis of precious, low-concentration compounds.
Sample ionization takes place in the source at atmospheric pressure. These
ions are sampled through a series of orifices into the first quadrupole where
they are filtered according to their mass to charge ratio (m/z). The mass
separated ions then pass into the hexapole collision cell where they either
undergo collision induced decomposition (CID) or pass unhindered to the
second quadrupole. The fragment ions are then mass analyzed by the second
quadrupole.
The transmitted ions are finally detected by a conversion dynode, phosphor
and photomultiplier detection system. The output signal is amplified,
digitized and presented to the data system.
Vacuum system
An external rotary pump and an internal split-flow turbomolecular pump
combine to create a vacuum. The turbomolecular pump evacuates the
analyzer and ion transfer region.
The system monitors turbomolecular pump speed and continuously measures
the vacuum with a built-in Pirani gauge. In the event of leaks, electrical
failure, or vacuum pump failure a loss of vacuum will occur. The Pirani gauge
also acts as a switch, discontinuing instrument operation if it senses a vacuum
loss.
An easy-access vacuum isolation valve enables routine source maintenance to
be performed without breaking vacuum.
Data system
The data system collects information from the mass analyzer. The data
system consists of:
• An embedded PC
• An external workstation
• The MassLynx™ software
The workstation-based data system, incorporating MassLynx software,
controls the mass spectrometer and, if applicable, the HPLC system,
autosampler, divert valve or injector valve. The workstation uses the
Windows® graphical environment with color graphics, and provides for full
user interaction with either the keyboard or mouse. MassLynx provides full
control of the system including setting up and running selected HPLC
systems, tuning, acquiring data, and data processing. MassLynx instrument
MassLynx software
MassLynx software, a Windows-based application, enables the following
operations:
• Configuring the Quattro micro API.
• Creating inlet and MS methods that define operating parameters for a
run.
• Tuning and calibrating the Quattro micro API.
• Running samples.
• Monitoring the run.
• Acquiring data.
Refer to the MassLynx User’s Guide and Help for more information on
installing and using MassLynx software.
MS operating modes
Ion optics
MS operating modes
The MS1 mode, in which MS1 is used as the mass filter, is the most common
and most sensitive method of performing MS analysis. This is directly
analogous to using a single quadrupole mass spectrometer.
The MS2 mode of operation is used, with collision gas present, when switching
rapidly between MS and MS/MS operation. It also provides a useful tool for
instrument tuning and calibration prior to MS/MS analysis, and for fault
diagnosis.
Typical application:
• Structural elucidation.
– Complementary or confirmatory information (for daughter scan
data).
MRM mode
This mode is the MS/MS equivalent of SIR (Selected Ion Recording). As both
MS1 and MS2 are static, this allows greater dwell time on the ions of interest
and therefore better sensitivity compared to scanning MS/MS.
Typical application:
• Rapid screening of dirty samples for known analytes:
– Drug metabolite and pharmacokinetic studies.
– Environmental, for example pesticide and herbicide analysis.
– Forensic or toxicology, for example screening for target drugs in
sport.
Example:
Monitor the transition (specific fragmentation reaction) m/z 609 to 195 for
reserpine in ElectroSpray positive ion LC/MS/MS mode.
MRM does not produce a spectrum as only one transition is monitored. As in
SIR, a chromatogram is produced.
Time Time
LC/MRM LC/MS
High specificity Low specificity
Good signal/noise Poor signal/noise
Collision Cell
RF only
MS1 (pass all masses)
MS2
scanning scanning
Front panel
Nebulizer Gas
Syringe Pump
Fused Silica Guide
Reset Switch
CID valve
The CID Gas valve is a fifteen-turn valve. The collision gas flow increases as
the valve is turned counterclockwise.
Caution: To prevent damage to the CID Gas valve, take care not to
over-tighten when turning the supply off.
HPLC Probe
Pump
Sample Loop
Sample Waste
Syringe
®
The divert/injection valve is an electrically driven Rheodyne injector that
may be used in several ways depending on the plumbing arrangement:
• As an injection valve, with the needle port and sample loop fitted.
• As a divert valve, to switch the flow of solvent during a LC run.
• As a switching valve, for example to switch between a LC system and a
syringe pump containing calibrant.
Control of the valve is primarily from the data system. The two switches
marked Load and Inject enable the user to override control of the valve when
making loop injections at the instrument.
For details of the use of the valve as a divert valve see page 4-21.
Vacuum
Operate
Vacuum LED
Operate LED
Analog channels
Four analog channel inputs are available, for acquiring simultaneous data
such as a UV detector output.
The input differential voltage must not exceed one volt. Analogue data is
20
processed by a 12-bit ADC with a gain ranging up to 2×10 counts.
If the input cable is only a two-wire assembly, then the negative pole of each
channel may need to be grounded.
Mux interface
This 9-way D-type connector enables interfacing to the MUX control unit.
Events
CE int (capillary electrophoresis interlock)
This connector enables interfacing with a capillary electrophoresis power
supply so that the instrument is safely interlocked against high voltages.
GF (gas fail)
If the desolvation gas flow decreases to less than 4% of its full-scale range, the
instrument generates a signal that can be used to stop solvent flowing into the
source by connecting it to the Stop Flow of the HPLC system.
FC (FractionLynx control)
A 100 mV analog output signal is provided to allow a trigger signal for an
external fraction collection device. The optional FractionLynx software must
be purchased for this.
PC link
This RJ45 connector links the instrument to the data system using the
network cable supplied.
Contents
Topic Page
Starting the Quattro micro API 2-2
Installing the ESI (ElectroSpray) probe 2-5
Setting up the syringe pump 2-6
Setting up the Quattro micro API 2-7
2-1
Starting the Quattro micro API
On/off switch
On/Off
Switch
ESI probe
Probe
High Voltage Lead
Desolvation Lead
Gas
Knurled
Thumbscrews
(2)
Probe
Nebulizer
Gas Flange
CID
Valve Probe Adjuster
Plate
Probe
Adjuster
Syringe pump
Syringe
Needle Port
Capillary
Syringe Stop
Adjuster
3. Screw the Rheodyne 9013 needle port fitting into the PEEK™ union,
and tighten it so that it does not leak.
Warning: Clip the ground cable (with plug-in clip), located at the
front panel, lower right, onto the syringe needle.
6. Connect the other end of the capillary to the inlet on the ESI probe with
an Upchurch Scientific nut, ferrule, and PTFE tubing.
3. Click on the default page to invoke the tune page. The example tune
page, shown in the figure “ESI tune page” on page 2-9, uses ions from a
solution of PPG1000, reserpine and PA ß Cyclodextrin.
4. Select Options from the menu bar, then Pump. The rotary pump starts
to evacuate the detector. In about 20 minutes, the instrument is
sufficiently evacuated to enable operation, and the Vacuum indicator on
the front panel shows green.
5. To view the actual values for instrument parameters select Readbacks
from the Options menu, then Always On.
6. Enter the suggested initial reference solution values from the table
titled “Tune page initial reference solution values” on page 2-9 in the
corresponding tune page fields. These settings are intended as starting
points only. Optimum values may vary between instruments.
Caution:
• Failure to flow desolvation gas during ESI operation can cause
heat damage to the source.
• To avoid contaminating source components, always specify a
desolvation gas flow rate that exceeds 100 L/h.
Tip: To avoid nitrogen leakage, fully tighten the probe locking ring.
8. Click the Analyser tab.
9. Enter the parameter values listed in the following table, dependent on
whether tuning for MS1 or MS2.
Analyser page
9. Set the syringe flow rate to 10 µL/min., and click on the tune page
menu bar.
10. On the tune page, set Desolvation Gas to 150 L/h.
11. Check for leaks at the probe and syringe fittings.
12. Monitor for mass peaks. The peaks should appear at approximately the
mass values entered on the ES+ Source tab.
13. Increase values in the Gain fields until mass peaks become clearly
visible.
Caution:
• To avoid contaminating source components, always specify a
desolvation gas flow rate that exceeds 100 L/h.
• An optimum signal must be obtained before the instrument can
successfully be calibrated.
The source is now ready for ElectroSpray use. Refer to Chapter 3 for
further information.
Caution:
• Before restarting nitrogen flow following its interruption, the
API gas flow must be stopped from the tune page. Restarting
the nitrogen while the API gas is flowing can damage the flow
meter.
• Do not start the liquid flow until the gas flow and probe heater
are switched on with the probe inserted.
4. Ensure that the desolvation gas tube is connected at the front panel.
5. Set Source Temp to 130 °C.
6. Set Corona to 2 µA and Sample Cone to 50 V.
7. Allow the source temperature to reach 130 °C.
Warning:
• The source enclosure and parts of the probe adjustment flange
may reach high temperatures when in use.
• Switch off the liquid flow and allow the probe to cool to less
than 100 °C before removing it from the source.
Probe position
Turn the probe flange adjuster to optimize the signal. Spray should be
approximately midway between the corona pin and the sample cone.
Corona current
Corona current can have a significant effect on sensitivity. The corona current
required depends upon the polarity of the compound and the polarity of the
analytical mobile phase. Optimization should be performed in the presence of
the analytical mobile phase.
Probe temperature
For maximum sensitivity, the APCI probe temperature must be optimized as
follows, ensuring that the analytical mobile phase is used during optimization:
Starting at 650 °C, reduce APCI Probe Temp in 50 °C decrements,
allowing time for the temperature to stabilize before taking a reading.
It is possible to set APCI Probe Temp too low for the mobile phase. This often
results in significant tailing of chromatographic peaks.
Desolvation gas
In most circumstances the desolvation gas flow has little effect on signal
intensity. However, in some situations, it can affect chemical background
noise levels. Adjusting desolvation gas can suppress chemical background
noise.
Cone gas
Set the cone gas flow to minimize formation of solvent adducts. The typical
value is about 50 L/h.
Contents
Topic Page
Overview 3-2
The tune page 3-2
Printing tune information 3-2
Experimental record 3-3
Saving and restoring tune parameter settings 3-4
Modifying the peak display 3-5
Changing the display 3-8
AutoTune 3-10
Ion mode 3-13
Scope parameters 3-13
Gas controls 3-14
Ramp controls 3-15
Resetting the zero level 3-17
Controlling readbacks 3-17
Changing tune parameter settings 3-18
Source voltages 3-18
3-1
Overview
For the highest mass accuracy, the instrument should be tuned and calibrated
using a suitable reference compound before sample data are acquired.
• Consult the relevant section of this manual for information concerning
source tuning procedures in the chosen mode of operation.
• Adjust the tuning parameters in the Source and Analyser menus to
optimize peak shape and intensity at unit mass resolution.
• Care should be taken to optimize the value of the collision energy. Note
that, in Daughter and Parent modes, Collision and Exit are interactive
parameters.
3-2 Tuning
Tune page
Experimental record
Tuning parameters are stored with every data file as part of the experimental
record. The tuning parameters for a particular data file can be viewed or
printed from the data browser, see the MassLynx NT User’s Guide for more
information.
To save the current tune parameters with the existing file name:
2. Enter a new file name, or select an existing file from the displayed list.
3-4 Tuning
3. Press Save.
4. If the selected file already exists on disk, a warning is displayed. Press
Yes to overwrite the existing information or No to enter a different file
name.
1. Press , or choose Open from the tune page File menu, the Open dialog
box is opened.
2. Select the required tuning parameter file, either by typing its name or
by selecting from the list.
3. Press Open.
Either:
1. Click and drag the mouse within the bounds of the axis to draw a
“rubber band” around the region of interest.
2. Release the button.
This range is redisplayed to fill the window. The mass displayed in the
Mass box is the mass at the centre of the window.
This operation can be repeated as often as required.
Or:
1. Enter a value in the Mass box for the required peak and press Return.
This becomes the default, so, if the range is altered with the mouse and
is pressed twice, Mass returns to this value.
Or:
1. Position the cursor at the top of the peak window, just below the line
showing the gain.
2. When appears, click the left mouse button and drag until the
required mass is displayed in the Mass box and at the top of the window.
3-6 Tuning
This becomes the default, so if the range is altered with the mouse and
is pressed twice Mass returns to this value.
Or:
1. Position the cursor at the top of the peak window, just below the line
showing the gain.
2. When appears, click the left mouse button and drag until the
required mass is displayed in the Mass box and at the top of the window.
This becomes the default, so if the range is altered with the mouse and
is pressed twice Mass returns to this value.
Either:
1. Press the left mouse button at one end of the region of interest and,
without releasing the button, drag the mouse horizontally to the other
end.
As the mouse is dragged a “rubber band” stretches out to indicate the
selected range.
Do not go beyond the bounds of the axis.
2. Release the mouse button to re-display the selected range filling the
current window.
This operation can be repeated as often as required.
Or:
1. Enter a value in the Span box for the required peak and press Return.
This becomes the default, so if the range is altered with the mouse and
is pressed twice Span returns to this value.
Either:
1. Double-click on the line above the peak which shows the gain, to double
the gain applied to that peak.
2. Double-click below the peak to halve the gain.
Or:
1. Press the left mouse button at one end of the region of interest and,
without releasing the button, drag the mouse vertically to the other end.
As the mouse is dragged, a marquee indicates the selected range.
Do not go beyond the bounds of the axis.
2. Release the mouse button to re-display the selected range filling the
current window.
Or:
1. Enter a value in the Gain box for the required peak and press Return.
The display area for each peak can be individually changed, e.g. the peak color
for peak 1 can be red, for peak 2 green, etc.
3-8 Tuning
Customize plot appearance
To change the color of the background and traces and to change the
number of traces displayed:
• Select Customise, Plot Appearance from the Peak Display pop-up menu,
see the figure “Peak display pop-up menu” on page 3-8.
• The Customise Plot Appearance dialog box is invoked, see figure below.
• Use to change the number, or enter a new value in the Visible Traces
box, within the range 2 to 20.
• If more than one trace is displayed, the older traces can be displayed in a
different shade to the new ones:
Trace
From the Peak Display pop-up menu, either:
Select the Trace, Outline option to display the peak outline only.
Or:
Select the Trace, Fill option to fill the trace with the trace fill colour.
Or:
Select the Trace, Min/Max option to show the minimum and maximum
data points only.
The option selected has a tick next to it.
Intensity
To select intensity:
1. From the Peak Display pop-up menu, select either Intensity,
Relative Intensity, or Intensity, Absolute Intensity as required.
2. Select Intensity, Normalise Data to display normalized data.
The options selected each have a tick next to them.
Grid
The Peak Display pop-up menu, options allow vertical and horizontal grid
lines to be independently displayed or hidden.
Selected options have ticks next to them. Selecting an option a second time
deselects the option.
AutoTune
MassLynx can automatically tune the mass spectrometer in both APcI and
ElectroSpray ionization modes. AutoTune ramps the settings for the tuning
parameters until they are optimized to give the best intensity, resolution and
peak shape.
3-10 Tuning
To run AutoTune:
1. Press on the tune page to turn on the API gas, and select Operate.
2. Choose AutoTune from the tune page Options menu, the AutoTune
dialog box is invoked.
AutoTune 3-11
AutoTune Setup dialog box
3-12 Tuning
• Ion energy tuning
• High and low mass resolution tuning
– The final four of these steps represent the implementation of the
ESP/APcI AutoTune algorithm. This involves changing key
parameters, one at a time, to maximize the intensity of a reference
peak with respect to that parameter. At present, ESP/APcI
Autotuning is carried out with respect to a single user-specified
reference peak.
When AutoTune has finished, it displays a status window to say that
AutoTune has been successfully completed.
Press OK to return to the tune page.
The tuning parameters determined by AutoTune are saved to the
current tune parameter file.
Ion mode
Select the required ionization mode from the Tune Page Ion Mode menu. The
selected mode has a tick next to it.
Scope parameters
The Scope Setup dialog box is invoked by pressing , or selecting the Tune
Page Options, Scope Parameters menu.
Scan Time(s) and Inter Scan Delay (s) control the speed with which the tune
peak display is updated.
Gas controls
3-14 Tuning
Ramp controls
Two values of cone voltage are defined at two particular masses. These
values define a gradient for the cone voltage which is then extrapolated
to cover the full mass range.
2. Make any changes required and press OK to exit.
1. Press , or choose Use Cone Ramp from the tune page Ramps menu.
A tick mark appears next to the menu item if the cone voltage ramp is
selected.
1. Press , or choose Use Collision Energy Ramp from the tune page
Ramps menu.
3-16 Tuning
Resetting the zero level
Controlling readbacks
The Readbacks dialog box is invoked by selecting the Tune Page Options,
Readbacks menu.
There are three options for displaying system readbacks on the tune page:
• Readbacks displayed continuously.
• Readbacks hidden.
• Readbacks displayed only when differing from their defined values by
more than 10%.
A number of the readbacks are for diagnostic purposes only, their function
being to confirm a voltage is present. The acceptable variation between the set
value and the readback value varies depending on the particular tune
parameter. If concerned about any reading, contact the local service office for
advice.
Source voltages
The following table lists the various components of Quattro micro API’s ion
optical system. The name in the table’s first column is the name used
throughout this manual to describe the component. When appropriate, the
second column shows the term used in the current MassLynx NT release.
Source voltages
Tune page
ESI+ve ESI-ve APcI+ve APcI-ve
name
ElectroSpray Capillary +3.0 (kV) -3.0 (kV) Not applicable
Probe
APcI Discharge Corona Not applicable 2 µA 2 µA
Pin
3-18 Tuning
Source voltages (Continued)
Tune page
ESI+ve ESI-ve APcI+ve APcI-ve
name
Sample Cone Cone +50 (V) -50 (V) +50 (V) -50 (V)
Extraction Cone Extractor +3 (V) -3 (V) +3 (V) -3 (V)
Hexapole RF Lens +0.2 (V) -0.2 (V) +0.2 (V) -0.2 (V)
The voltages shown are typical for an instrument in good condition. The
polarities given are those actually applied to the electrodes. Only positive
values need be entered via the tune page.
Contents
Topic Page
Starting an acquisition 4-2
Monitoring an acquisition 4-8
Instrument data thresholds 4-10
System manager 4-16
Stopping an acquisition 4-16
The function list editor 4-16
4-1
Starting an acquisition
There are two ways of starting an acquisition:
• A single sample acquisition from the tune page
• A multiple sample acquisition from the MassLynx top level screen
Multiple samples
The MassLynx default page contains a sample list editor for defining multiple
samples which may be used together to perform a quantitative analysis. The
list of samples is set up using a spreadsheet style editor, which can be tailored
to suit different requirements.
3. Check the Acquire Sample Data, Auto Process Samples and Auto
Quantify Samples boxes as required.
4. Enter values in the Run, From Sample and To Sample boxes.
The default is all samples in the list.
5. Check the Priority and/or Night Time Process boxes as required. See the
MassLynx NT User’s Guide for details.
6. Press OK.
To run a process after each sample in the sample list has been acquired:
Format the sample list to display the Process column and enter the
name of the process to be run for each of the samples.
For the process to automatically operate on the data file which has just been
acquired:
Leave unchecked Use Acquired File as Default on the System tab of the
MassLynx Options dialog box.
The MassLynx Options dialog box is accessed by choosing Options from the
MassLynx Tools menu.
Print Quantify Reports produces hard copies of the results of integration and
quantification.
Export Results to LIMS produces a text file containing the quantification
results details for use with LIMS systems. If this box is checked, the LIMS
Export Browse button becomes enabled. Press Browse, select a file, or enter
the name of a new file, and press Save.
The Project field displays the project into which data are acquired.
To change the project into which data are acquired, the acquisition should be
canceled and a new project created by choosing Project Wizard, or an existing
project opened by choosing Open Project, from the MassLynx top level File
menu.
From Sample and To Sample set the range of samples in the sample list which
is analyzed.
Instrument data thresholding allows the user to specify the type of data to
acquire and write to disk, and the type of data to discard and not write to disk.
Limiting the amount of data stored on disk can be particularly desirable when
acquiring continuum data and doing long LC runs.
MaxEnt
The MaxEnt algorithm needs to measure noise accurately within a data file.
For this reason Ion Counting Threshold should be set to zero when acquiring
data to be analyzed using MaxEnt.
Profile data
The controls for profile data allow control of the amount of data collected
during a continuum data acquisition.
Baseline Level is used to lift or drop the baseline to see more or less of the
noise by positioning of the baseline above zero. Baseline Level is typically set
to a value of 0.
It is possible to use a negative baseline. This reduces the noise seen and acts
as a form of thresholding to be applied to 1/16 amu type samples. This takes
place after ion counting and therefore has a less significant effect than Ion
Counting Threshold.
SIR data
SIR Baseline Level sets the position of the SIR baseline above zero. The
baseline level is typically set to 0. Increasing the value causes the baseline to
appear higher.
Typical
Typical
Baseline Ion count Typical peak Typical saving on
background
level threshold profile intensity .DAT file
noise
size
0 0 0 0
1 0 0
2 0 0
5 0
0 10 4% 8%
Typical
Typical
Baseline Ion count Typical peak Typical saving on
background
level threshold profile intensity .DAT file
noise
size
0 20 11% 10%
0 40 37% 61%
0 60 66% 69%
Analog data
Select the number of samples to acquire per second from the drop-down list.
Stopping an acquisition
Introduction
The Function List Editor is used to set up the function(s) that the mass
spectrometer uses to scan the instrument during an acquisition. A function
list can be a mixture of different scanning techniques that can be arranged to
run either sequentially or concurrently during an acquisition.
The figure above shows a simple function list containing only one function: a
centroided mode full scan, between 500 and 1500 amu using ES+ ionization.
Immediately above the function bar display is a time scale that shows from
when the function is active, and for how long it runs. In this case, the function
starts after 5 minutes and then runs for 35 minutes, terminating after a total
elapsed time of 40 minutes.
A more complicated function list, with four SIR functions each running
sequentially for 5 minutes, is shown in the figure “Function list with four SIR
functions” on page 4-18.
Entering a new a value in Total Run Time and pressing sets the maximum
retention time for the experiment. The ratio of the functions defined is
maintained. For example, if two functions are defined one from 0 to 5 minutes
Removing a function
To remove a function:
1. Select the function in the function list.
2. Press , choose Delete from the Edit menu, or press Del on the
3. When asked to confirm the deletion, select Yes.
To change the order of functions with same start and end time:
1. Highlight the required function.
Analog channels
Up to four channels of analog data can be acquired, which are stored with the
data acquired from the mass spectrometer. Analog channels are typically used
to collect data from external units such as UV detectors, which must be
connected to the user input/output PCB as described on page 1-16.
Mass (m/z)
Start Mass and End Mass specify the masses at which the scan starts and
stops. Start Mass must be lower than End Mass.
Start Time and End Time specify the retention time, in minutes, during which
this function becomes active, and data are acquired.
Cone voltage
When Use Tune Page is checked, the cone voltage set on the tune page at the
start of the acquisition is used.
The cone voltage value cannot be altered during acquisition by typing new
values into the tune page, since the new values are not downloaded during
acquisition. This can only be done by acquiring from the tune page.
The four parameters define a gradient for the cone voltage which is then
extrapolated to cover the full mass range of the function.
Method
Ionization Mode specifies the ionization mode and polarity to be used during
acquisition.
Data specifies the type of data to be collected and stored on disk. There are
three options:
• Centroid stores data as centroided, intensity and mass assigned peaks.
Data are stored for every scan.
• Continuum. The signal received by the interface electronics is stored
regularly to give an analog intensity picture of the data being acquired.
Data are not centroided into peaks, but are stored for every scan.
Due to the fact that data are acquired to disk at all times, even when no
peaks are being acquired, data files tend to be significantly larger than
centroided ones.
It is possible, however, to set a threshold below which the data are not
stored. Depending on the nature of the data acquired, this can greatly
reduce these effects. The threshold can be set so that data considered to
be “noise” can be discarded, thus improving data acquisition speed and
APcI probe
Probe Temp, in degrees centigrade, is enabled when Ionization Mode is set to
API.
When Use Tune Page Settings is selected, the APcI probe temperature set on
the tune page at the start of the acquisition is used. This control is enabled
when the Ionization Mode is set to API.
The APcI probe temperature value cannot be altered by typing new values
into the tune page during the acquisition, since the new values are not
downloaded during the acquisition. This can only be done by acquiring from
the tune page.
To enter a mass:
1. Type suitable values into the Mass, Dwell and Cone boxes.
2. Press Add.
Dwell specifies the length of time in seconds for which the highlighted mass is
monitored.
Method
Inter Channel Delay specifies the time, in seconds, between finishing
monitoring the highlighted mass and starting monitoring the next mass in the
function.
Repeats is only relevant for experiments having more than one function and
specifies the number of repeats of the function.
Span specifies a small mass window applied centrally about the highlighted
mass. During acquisition this range is scanned over the specified Dwell time.
A span of zero can be set to simply “sit on” the specified mass.
Retention window
Start and End together specify the retention time, in minutes, during which
this function is active.
Many of the fields in the MS/MS editors are similar to those in the Full Scan
function editor. Only fields which differ significantly are described below.
The four parameters define values of collision energy for two particular
masses. This collision energy gradient is then extrapolated to cover the full
mass range of the function.
MS to MSMS page
Switch criteria
MS/MS scanning commences:
• If TIC is selected, and the TIC of the spectrum rises above the specified
Threshold.
• If Intensity is selected, and the intensity of the largest peak rises above
the specified Threshold.
MSMS to MS page
When MS/MS functions have been generated, they are carried out in parallel
until the conditions for switching to MS are satisfied.
When all MS/MS functions have stopped, the MS survey function is again
carried out.
Switch method
If the MS/MS to MS switch method is Default, the MS/MS function stops when
the MS/MS to MS switch criteria are met.
Monitoring acquisitions
When an acquisition is started, the Function Switching Status dialog box is
invoked, showing the precursors currently running.
Contents
Topic Page
Introduction 5-2
Overview 5-2
ElectroSpray 5-5
Atmospheric pressure chemical ionization (APcI) 5-29
5-1
Introduction
This chapter is divided into three sections:
• A brief general overview of the calibration process.
• A complete mass calibration of Quattro micro API using ElectroSpray
ionization with a mixture of sodium iodide and rubidium iodide as the
reference compound.
• A complete mass calibration of Quattro micro API using atmospheric
pressure chemical ionization (APcI) with PEG as the reference
compound.
See Chapter 8, for details of calibration solutions and their preparation.
Overview
MassLynx NT allows a fully automated mass calibration to be performed,
which covers the instrument for static and scanning modes of acquisition over
a variety of mass ranges and scanning speeds.
A mass spectrum of a reference compound (a calibration file) is acquired and
matched against a table of the expected masses of the peaks in the reference
compound which are stored as a reference file. The mass differences between
the reference peaks and calibration peaks are the calibration points. A
calibration curve is fitted through the calibration points.
The vertical distance of each calibration point from the curve is calculated.
This distance represents the remaining (or residual) mass difference after
calibration.
Mass calibration
Overview 5-3
Calibration types
Each quadrupole analyser requires up to three calibration curves:
• A static calibration is used to “park” the analyser accurately on a specific
mass of interest (for example in tuning, SIR and MRM).
• A scanning calibration enables peaks acquired in a scanning acquisition
to be mass measured accurately.
• A scan speed compensation calibration compensates for “lag time” in the
system when the instrument is scanned rapidly.
A separate mass spectrum of the reference compound is acquired for each
selected calibration type.
Quattro micro API requires these three calibrations for both MS1 and MS2,
thereby generating a maximum of six calibration curves. The following table
shows which types of calibration are necessary for particular types of
experiment.
Calibration requirements
Calibration required
Experiment
MS1 MS2
MS All -
SIR Static -
MS/MS All All
MRM Static Static
Introduction
When a calibration is completed, it is possible to acquire data over any mass
range within the calibrated range. It is therefore sensible to calibrate over a
wide mass range.
With a mixture of sodium iodide and rubidium iodide, calibration over the
instrument’s full mass range is achievable.
ElectroSpray 5-5
To introduce the reference compound:
1. Fill the syringe with the reference solution. See page 2-6.
2. Couple the syringe to the ElectroSpray Probe with fused silica tubing.
3. Set the pump to a flow rate of 10 µL/min.
Tuning
Before beginning calibration, and with reference solution admitted into the
source:
1. Set Multiplier to 650 V.
2. Adjust source parameters to optimize peak intensity and shape.
3. Set the resolution and ion energy parameters for unit mass resolution on
MS1 and MS2.
2. In the Profile Data section set Baseline Level to 0 and Points per Dalton
to 16.
3. Set Ion Counting Threshold and Spike Removal as appropriate, see
page 4-10.
4. Select OK to save the parameters.
ElectroSpray 5-7
Calibration options
To access the Calibration Options dialog box:
1. Select Calibration then Calibrate Instrument on the tune page.
Selecting parameters
Several parameters need to be set before a calibration is started. Default
parameters are set when the software is initially loaded which usually give a
suitable calibration, but under some conditions these may need to be adjusted.
ElectroSpray 5-9
This dialog box defines limits which the calibration must attain before the
instrument is successfully calibrated. Two user parameters can be set:
• Missed Reference Peaks sets the maximum number of consecutive peaks
that are not matched when comparing the reference spectrum and the
acquired calibration spectrum. The calibration fails if this number is
exceeded. The default value for this parameter, 2, is suitable in most
cases.
• Maximum Std Deviation is set to a default of 0.20. During calibration
the difference between the measured mass in the acquired calibration
file and the true mass in the reference file is taken for each pair of
matched peaks. If the standard deviation of the set of mass differences
exceeds the set value, the calibration fails. Reducing the value of the
standard deviation gives a more stringent limit. Increasing the standard
deviation means that the requirement is easier to meet, but this may
allow incorrect peak matching. Values greater than 0.20 should not be
used unless exceptional conditions are found.
Apply Span Correction should always be selected. This allows different mass
ranges to be scanned, within the calibrated range, without affecting mass
assignment.
Check Acquisition Calibration Ranges causes warning messages to be
displayed if an attempt is made to acquire data outside of the calibrated range
for mass and scan speed. It is advisable to leave selected.
The Peak Match parameters determine the limits within which the acquired
data must lie for the software to recognize the calibration masses and result in
a successful calibration. The default values are shown in the figure above.
Increasing the Peak window and Initial error gives a greater chance of
incorrect peak matching. All peaks in the acquired spectrum below the
Intensity threshold value (measured as a percentage of the most intense peak
in the spectrum) are not used in the calibration procedure.
The Polynomial order of the curve has values from 1 to 5 as the available
options:
• A polynomial order of 1 should not be used.
• An order of 2 is suitable for wide mass ranges at the high end of the
mass scale, and for calibrating with widely spaced reference peaks.
Sodium iodide in particular has widely-spaced peaks (150 amu apart),
ElectroSpray 5-11
and horse heart myoglobin is used to calibrate higher up the mass scale,
so this is the recommended polynomial order for these calibrations.
• An order of 3 fits a cubic curve to the calibration.
• A fourth order is used for calibrations which include the lower end of the
mass scale, with closely spaced reference peaks. This is suitable for
calibrations with PEG which extend below 300 amu.
• A fifth order fit rarely has any benefit over a fourth order fit.
Performing a calibration
Three types of calibration are available with MassLynx:
• Static Calibration
• Scanning Calibration
• Scan Speed Compensation
ElectroSpray 5-13
These are selected on the Automatic Calibration dialog box which is invoked
by selecting Start from the Calibrate dialog box.
Acquisition parameters
Selecting Acquisition Parameters in the Automatic Calibration dialog box
invokes the Calibration Acquisition Setup dialog box, where the mass ranges,
scan speeds and acquisition mode are set. When this box is first accessed it
ElectroSpray 5-15
contains default parameters relevant to the chosen reference file. These
default parameters show the limits of scan range and scan speed for the
currently-selected instrument and calibration parameters.
The upper area contains the Acquisition Parameters where mass range, run
time and data type are set.
When the instrument is fully calibrated, any mass range or scan speed is
allowed within the upper and lower limits dictated by the calibrations.
Select the nairb.ref file.
The solution described in Chapter 8 is suitable for use with this reference file.
If compatible reference solutions and reference files are used, then simply
selecting Default is sufficient action, no parameters need be entered
manually.
Run Duration sets the time spent acquiring data for each part of the
calibration. The time set must allow a minimum of three scans to be acquired
at the slowest scan speed used. Data are not acquired if the Run Duration is
ElectroSpray 5-17
Starting the calibration process
Once all of the data have been acquired each data file is combined to give a
single spectrum which is then compared against the reference spectrum to
form a calibration. This process takes place in the same order as shown in the
table above. If the full calibration dialog box is open, a constantly updated
status message for the calibration is displayed.
If, when the process is completed, the calibration statistics meet with the
requirements specified by the selected calibration parameters, a successful
calibration message is displayed. A calibration report is then printed showing
a calibration curve for each of the calibration processes.
For the acquisition to be effective, it must be saved under a suitable file name.
Click Browse to select the calibration data file (for example STATMS1,
SCNMS1, FASTMS1, STATMS2 etc.). The selected file must be from the
appropriate project.
ElectroSpray 5-19
Clicking on OK repeats the calibration procedure for that particular file and
displays a calibration report.
Calibration report
Calibration failure
If the calibration statistics do not meet the requirements, a message is
displayed describing at what point, and why, the calibration failed. This
message also states where the attempted calibration data can be viewed so
that the exact cause of failure can be determined.
ElectroSpray 5-21
Too many consecutive peaks missed
If the number of consecutive peaks which are not found exceeds the Missed
Reference Peaks parameter set in the Automatic Calibration Check dialog
box, the calibration fails. Peaks may be missed for the following reasons:
• The reference solution is running out so that the less intense peaks are
not detected.
• Multiplier is too low so that the less intense peaks are not detected.
• An incorrect ionization mode is selected. Check that the data have been
acquired with Ion Mode set to ES+.
Note that it is possible to calibrate in negative ion mode ElectroSpray
using the naineg.ref reference file with a suitable reference solution.
• Intensity threshold, set in the Calibration Parameters dialog box, is too
high. Peaks are present in the acquired calibration file but are ignored
because they are below the threshold level.
• Either Initial error or Peak window, set in the Calibration Parameters
dialog box, is too small. The calibration peaks lie outside the limits set
by these parameters.
• Maximum Std Deviation, set in the Automatic Calibration Check dialog
box, has been exceeded.
• The wrong reference file has been selected. Check that the correct file
(nairb.ref in this case) is selected in the Calibrate dialog box.
In the case of too many consecutive peaks missed:
• Check the data in the on-screen calibration report to see if the missed
peaks are present in the acquired calibration file.
• If the peaks are not present, the first three reasons (as explained above
in No peaks) are likely causes.
• If the peaks are present in the data, but are not recognized during
calibration then the latter four (Too many consecutive peaks missed) are
likely reasons.
Incorrect calibration
If the suggested calibration parameters are used, and providing that good
calibration data have been acquired, then the instrument should be calibrated
correctly. However in some circumstances it is possible to meet the calibration
criteria without matching the correct peaks. This situation is unusual, but it
is always sensible to examine the on-screen calibration report to check that
the correct peaks have been matched. These errors may occur when the
following parameters are set:
• Intensity threshold set to 0.
• Initial error too high (»2.0).
• Peak window too high (»1.5).
• Maximum Std Deviation too high (»0.2).
If the acquired spectrum looks like the reference spectrum and all of the
expected peaks are highlighted, the calibration is OK.
An alternative cause of incorrect calibration is from contamination or
background peaks. If a contamination or background peak lies within one of
the peak matching windows, and is more intense than the reference peak in
that window, then the wrong peak is selected. Under some conditions this may
happen with PEG. There are two ways to counter this:
• If the reference peak is closer to the centre of the peak window, the peak
window can be narrowed until the contamination peak is excluded. Take
care to ensure that no other reference peak is excluded.
• If the reference peak is not closer to the centre of the peak window, or if
by reducing the window other reference peaks are excluded, then the
calibration can be edited manually.
ElectroSpray 5-23
Manual editing of peak matching
If an incorrect peak has been matched in the calibration process, this peak can
be excluded manually from within the on-screen calibration report.
To manually edit Using the mouse, place the cursor over the peak in the
reference spectrum and click with the right mouse button.
Place the cursor over the peak in the acquired spectrum and click with
the right mouse button.
The peak is excluded and is no longer highlighted.
If the true reference peak is present, this can be included in the calibration by
the same procedure:
Place the cursor over the required peak and click with the right mouse
button.
The peak is matched with the closest peak in the reference spectrum.
Manually editing one peak does not affect the other matched peaks in the
calibration.
Verification
Once a full instrument calibration is in place it is not always necessary to
repeat the full calibration procedure when the instrument is next used.
Instead a calibration verification can be performed. (There is no benefit in
verifying each calibration individually, re-calibration is just as quick.)
ElectroSpray 5-25
8. Set Scan From, Scan To, Run Duration, Data Type, Scan Time and Inter
Scan Delay to agree with the acquisition parameters that are to be used
for data acquisition.
With only the Scanning Calibration selected, all of the other options in
this dialog box are unavailable.
9. Select OK to return to the previous dialog box, and OK again to start the
verification procedure.
A scanning acquisition is now performed. When the acquisition is complete,
the data are combined to give a single spectrum which is compared against
the reference file. A calibration curve is drawn and a report printed in a
similar way to when the original calibration was performed.
Unlike the original calibration procedure, the instrument calibration is not
changed and the report that is printed is a verification report.
ElectroSpray 5-27
The spectrum shown in the figure below demonstrates how the PEG spectrum
can be dominated by doubly-charged ions (in this case [M+2NH4]2+) if the
wrong conditions are chosen. In this case, the concentration of ammonium
acetate in the reference solution is too high (5 mmol ammonium acetate is the
maximum that should be used) and Cone is too low.
Introduction
This section describes a complete mass calibration of Quattro micro API using
atmospheric pressure chemical ionization. The procedures described should be
followed only after reading “ElectroSpray” on page 5-5, describing the
automated calibration with ElectroSpray ionization.
Due to the high flow rates used with APcI, the residence time of an injection of
reference solution in the source is too short to allow a fully automated
calibration, and the procedure therefore has to be carried out in several steps.
The recommended reference compound for APcI is a solution of polyethylene
glycol (PEG) containing ammonium acetate. See Chapter 8, for advice on
preparing the reference solution. The figure “Typical APcI spectrum of PEG”
on page 5-30 shows a typical PEG + NH4+ spectrum.
With PEG, the possible calibration range is dependent upon the molecular
weight distribution of the PEGs used in the reference solution. For this
example PEG grades from PEG 200 to PEG 1000 are used.
Calibration options
To access the calibration options, click on Calibrate from the tune page.
Static calibration
The upper area contains the Acquisition Parameters where mass range, run
time and data type are set. When the instrument is fully calibrated, any mass
range or scan speed is allowed within the upper and lower limits dictated by
the calibrations. It is therefore sensible to calibrate over a wide mass range.
Since the pegnh4.ref reference file has peaks from m/z 89 to m/z 2017, it is
possible to calibrate over this mass range. A calibration effective up to
1000 amu is sufficient for the majority of applications with APcI. The
following example shows a setup to achieve this.
Run Duration sets the time spent acquiring data for the static calibration. The
time set must allow chance to inject a volume of reference solution and acquire
several scans.
Data Type allows a choice of Centroided, Continuum or MCA data to be
acquired. For APcI, while either Continuum or Centroided data may be used,
Continuum is recommended.
The lower area in the Calibration Acquisition Setup dialog box contains the
Scan Parameters.
Scanning calibration
The recommended scan speed for the Scanning calibration is 100 amu/s.
1. Set Scan From to 50 amu and Scan To to 1050 amu.
2. Set Scan Time to 10 s and Inter Scan Delay to 0.1 s.
3. Select Continuum as the Data Type.
Although Continuum is recommended, Centroided data may be used.
4. Set Run Duration to 2.0 minutes.
This allows time to start the acquisition, inject the reference solution
and acquire several scans. With a solvent flow rate of 200 µL/min and a
50 µL loop in line, an injection of reference solution lasts approximately
15 s, allowing at least one full scan of useful data to be acquired.
5. Choose a filename for the data.
The filename SCNMS1, the name used during an automatic calibration,
is valid.
6. Start the acquisition and inject the reference solution.
Calibration failure
When calibration is performed manually there is no warning message to show
that the calibration has not met the set criteria. This must be judged by
viewing the on-screen calibration report and examining the matched peaks
and statistics associated with the report. There are a number of reasons for a
calibration to fail:
No peaks
If the acquired calibration data file contains no peaks the calibration has
failed. This may be due to:
• Lack of reference compound.
• Wrong scans or wrong data file being used for the calibration.
• No flow of solvent into the source.
• Multiplier set too low.
Incorrect calibration
If the suggested calibration parameters are used, and providing that good
calibration data have been acquired, the instrument normally calibrates
correctly. However, in some circumstances it is possible to meet the
calibration criteria without matching the correct peaks.
This situation is unusual, but it is always wise to examine the on-screen
calibration report to check that the correct peaks have been matched. These
errors may occur when the following parameters are set:
• Intensity threshold set to 0
• Initial error too high (>2.0)
• Peak window too high (>1.5)
• Maximum Std Deviation too high >0.2).
If the acquired spectrum looks like the reference spectrum and all of the
expected peaks are highlighted then the calibration is OK.
An alternative cause of calibration failure is from contamination or
background peaks. If a contamination or background peak lies within one of
the peak matching windows, and is more intense than the reference peak in
that window, then the wrong peak is selected. Under some conditions this may
happen with PEG. There are two ways to counter this:
• If the reference peak is closer to the centre of the peak window, the peak
window can be narrowed until the contamination peak is excluded. Take
care to ensure that no other reference peak is excluded.
• If the reference peak is not closer to the centre of the peak window, or if
by reducing the window other reference peaks are excluded, the
calibration can be edited manually.
Contents
Topic Page
Maintenance schedule 6-2
Safety and handling 6-3
Routine maintenance 6-4
Replacing parts 6-28
6-1
Maintenance schedule
The following table lists periodic maintenance schedules to be followed to
ensure optimum performance.
The maintenance frequencies shown apply to instruments that normally
receive moderate use.
Maintenance schedule
Warning: Never open the instrument’s top or side panels to access power
supplies or other components. The instrument does not contain
user-serviceable parts.
Caution:
• Never disconnect an electrical assembly while the system is plugged
in. This could damage electrical parts. Also, turn off the power and
wait 10 seconds before disconnecting an assembly.
• Do not touch integrated circuit chips or other circuit board
components. Static electric charge can damage electronic
components.
Maintenance equipment
Routine parts cleaning requires the following equipment:
• An ultrasonic bath with a minimum chamber size of
12 inches × 6 inches × 4 inches (300 mm × 150 mm × 100 mm).
• Glass vessels, approximately 4 inches (100 mm) in diameter and
4.25 inches (120 mm) high.
• A 500-mL graduated cylinder (to use when cleaning the hexapole
assembly).
Caution:
• Failure to routinely gas-ballast the rotary pump shortens oil life and
consequently shortens pump life.
• Do not vent the instrument when the rotary pump is gas-ballasting.
• Do not gas-ballast the rotary pump while Quattro micro API is in the
Operate mode.
• Never gas-ballast the rotary pump for more than 2 hours.
3. When the oil is clear and has drained back to the rotary pump, return
the gas-ballast control to its normal position.
4. Open the vacuum system isolation valve.
Required materials
• Rubber gloves
• Flat-blade screwdriver
• Container to catch used oil
• Funnel
• Vacuum oil (use only Ultragrade 19 or Inland Q45 (Edwards 45) vacuum
pump oil)
Procedure
Required materials
The following materials are required to clean the source components:
• Lint-free cotton or powder-free nitrile gloves
• 6-inch (150-mm) or needle-nose pliers
• 2.5-mm hex wrench
• 6-mm hex wrench
• Small, flat-blade screwdriver
• Large, flat-blade screwdriver
Spare parts
The following spare parts may be required when cleaning source components:
• Ion block D-ring (AS035)
• Viton O-ring (AS214)
• Extraction cone O-ring
• Sample cone O-ring
3. When the probe has cooled, stop the nitrogen flow by deselecting on
the toolbar or by choosing Gas from the Gas menu.
4. Stop the liquid flow, and disconnect the LC line from the probe.
5. Click Press For Standby on the MassLynx tune page.
10. Unscrew the probe’s two knurled thumbscrews and retract it from the
source.
Probe lead
Probe adjuster
plate
Source enclosure
door
11. Remove the center section panel from the front of the instrument,
pulling it away from the instrument.
Removable
Center Panel
12. Unfasten the source enclosure door’s securing clips and open the door.
Caution:
• Wear lint-free cotton or powder-free nitrile gloves for the rest of
the cleaning procedure.
• Take care not to scratch the highly polished cone orifice
surfaces.
13. If using an APcI probe, carefully remove the corona discharge needle.
Source enclosure
securing screw
Gas cone
PTFE tubing
14. Remove the PTFE tube attached to the cone gas cone.
15. Remove the two screws that secure the cone retainer, using the small,
flat-blade screwdriver.
6-mm Hex
Screw Isolation
Valve
Cone
Retainer
16. Remove the cone gas cone, O-ring, and sample cone from the ion block.
Isolation
Valve
Sample
Cone
Cone Gas
Cone
Cone
Retainer Clip
17. Carefully separate the sample cone from the gas cone, then remove the
sample cone O-ring.
18. Remove the baffle plate, and set all pieces aside.
Exploded diagram
PEEK connector block
Isolation valve
PTFE ring
O-rings
O-rings
Extraction cone
21. Place the ion block on a flat surface, and remove any O-rings.
22. Use a screwdriver to remove the hold-down screw from the PEEK
extraction cone retainer.
Hold-Down
Screw PEEK Extraction
Extraction
Cone Retainer
Cone
23. Place the shaft of a small, flat-blade screwdriver in the ion block relief,
and carefully pry the insulator O-ring away from the ion block.
24. Take care not to damage the ion block surface and insulator O-ring.
25. Grasp the extraction cone pin with the needle-nose pliers, then lift the
extraction cone from the ion block.
26. Insert the small, flat-blade screwdriver under the inner edge of
polymeric seal ring and carefully pry the seal ring and O-ring out of the
ion block.
27. Take care not to damage the seal and O-ring on the ion block.
28. Remove the D-shaped O-ring from the front of the ion block.
29. Remove the ion block plug and seal using a flat-blade screwdriver.
30. Remove the O-ring from around the sample cone orifice, taking care not
to scratch the ion block surface.
Warning: Use extreme care when working with formic acid. Use a
fume hood and appropriate protective equipment.
4. If the sample cone contains debris, place a drop of formic acid on its
orifice.
5. Place the sample cone, cone gas cone, and extraction cone in a beaker
with methanol:water (1:1).
6. If the parts are obviously fouled, use a mixture of 45% methanol, 45%
water, and 10% formic acid.
7. Expose all parts to ultrasound for about 30 minutes.
8. If formic acid was used in the cleaning solution:
a. Rinse the parts, immersing them in a beaker of water and setting
the beaker in an ultrasonic bath for about 20 minutes to remove
formic acid from them.
b. Displace the water by immersing the parts in a beaker of methanol
and setting the beaker in the ultrasonic bath for 10 minutes.
9. Carefully remove the parts from the beaker and blow-dry them, using
inert, oil-free gas.
Alternatively. the parts may be placed on lint-free towels and allowed to
air dry. Wipe off any water spots with a lint-free cloth.
Pumping block
O-rings
PEEK ion block
support
Hexapole
assembly
2. Remove any O-rings that remain stuck to the surface of the pumping
block.
Rear
Support
Ring
Alignment
Notch
Required materials
The following materials are needed to clean the corona discharge needle:
• Lint-free cotton or powder-free nitrile gloves
• Lapping film
• HPLC-grade methanol
• Lint-free tissue
Procedure
2. When the probe has cooled, stop the nitrogen flow by deselecting on
the toolbar or by choosing Gas from the Gas menu.
3. Click Press For Standby on the MassLynx tune page.
4. Ensure that the icon changes from green to red. This means all high
voltages are turned off, as well as the APcI probe heater.
5. Remove the center panel from the front of the instrument.
6. Disconnect the nebulizer gas and the probe electrical connection.
Probe lead
Probe adjuster
plate
Source enclosure
door
9. Unfasten the source enclosure door’s securing clips, and open the door.
10. Remove the corona discharge needle from the source, pulling it straight
out.
Source enclosure
Source enclosure
securing screw
Gas cone
PTFE tubing
11. Clean and sharpen the tip of the needle with the lapping film, then wipe
it clean with a methanol-saturated tissue. Replace the needle if it is
deformed or otherwise damaged.
12. Reinstall the needle with the tip pointing toward the sample cone apex.
Warning: The source can be hot; to avoid burns, take great care
while working with this component.
2. Unfasten the source enclosure door’s securing clips and open the door
(see the figure “Probe and source connections” on page 6-10).
6. Use a hex (Allen) key to remove the four bolts securing the source
enclosure door glass retaining clips to the source enclosure door.
7. Remove the four source enclosure door glass retaining clips from the
source enclosure door.
8. Remove the source enclosure door glass from the source enclosure door.
9. Carefully remove the source enclosure door glass O-ring from the source
enclosure door.
10. Carefully remove the probe adjustment flange O-ring from the probe
adjustment flange.
Probe adjustment
flange O-ring
The nitrogen exhaust waste bottle in the nitrogen exhaust line must be
emptied before it is completely full.
To laboratory
exhaust port
From instrument
exhaust connection
®
Tip: To avoid damage to the instrument, Snoop (or equivalent) leak
detector liquid must only be used only for the purpose described in the
following step; it must not be used on any other part of the instrument.
9. Use Snoop (or equivalent) leak detector liquid to ensure that there are
no leaks at the instrument exhaust and laboratory exhaust system line
connections.
Replacing parts
Required materials
• 3 mm hex wrench.
• 1.5 mm hex wrench.
• Flat-blade screwdriver
• Needle-nose pliers
Procedure
Cartridge Heaters
4. Loosen the two set screws that secure the heater cartridges in the ion
block, using the 1.5-mm hex wrench.
5. Gently slide the heater cartridges out of the ion block using the
needle-nose pliers.
6. Slide the new heater cartridges into the ion block with the needle-nose
pliers. Secure them with two hex-head set screws and the 1.5-mm hex
wrench.
7. Position the two heater cartridge ring tags onto the PEEK block
terminals with the bent portion of their shafts extending into the
pumping block
8. Tighten the two terminal block screws with a flat-blade screwdriver.
9. Refit the ion block cover plate, and secure with the four hex screws.
10. Close the source enclosure door and fasten the door’s securing clips.
11. Install the probe.
Stainless Steel
Capillary
Probe
Tip
Required materials
• Flat-blade screwdriver
• 1.5-mm hex wrench
• 1/4-inch (6-mm) wrench
• 5/16-inch wrench
• 7/16-inch wrench
• Loupe (magnifying glass)
• Needle-nose pliers
Procedure
Caution: All work done on the probe should be carried out on a clean
work bench.
17. Reconnect the LC line, turn on the fluid flow, and check the probe for
liquid leaks.
clicking on the toolbar (or choose Gas from the Gas menu) on the
tune page.
20. Check the probe tip for nitrogen leaks. If a leak is found, replace the
probe tip assembly and its O-ring.
21. Refit the probe end-cover, and secure it with the two slotted screws.
Tighten the set screw to clamp the LC union in place.
Required materials
• 1/4-inch (6-mm) open-end wrench
• Loupe (magnifying glass)
Procedure
Caution: All work done on the probe should be carried out on a clean
work bench.
Required materials
1 -m hex wrench or flat-blade screwdriver, depending on the type of
screw that secures the probe tip.
Procedure
Caution: All work done on the probe should be carried out on a clean
work bench.
Required materials
• 1.5-mm hex wrench
• 5/16-inch open-end wrench
• 7/16-inch open-end wrenches (2-off)
• Ceramic capillary cutter (from the tools kit)
• Butane lighter or match
Spare parts
• GVF004 ferrules (2-off)
• Fused silica
• Filter pad
Procedure
Caution: All work done on the probe should be carried out on a clean
work bench.
13. Tighten the adapter nut until the capillary is snug in the fitting, then
gently pull on the capillary to ensure it stays in place
14. Feed the sample capillary through the probe, and gently tighten the
probe adapter nut with the 5/16-inch wrench. The capillary must
protrude 1.0 mm from the nebulizing tube.
15. Refit the probe end-cover and retaining screws.
16. Tighten the set screws in the probe end cover with the 1.5-mm hex
wrench to clamp the filter in place.
17. Refit the probe tip and heater assembly.
This chapter describes how to troubleshoot the Quattro micro API with
the help of recommended troubleshooting procedures.
Contents
Topic Page
Spare parts 7-2
Safety and handling 7-2
System troubleshooting 7-3
Component hardware troubleshooting 7-4
High noise levels in MRM analyses 7-13
Electronic noise 7-14
Contacting Waters 7-15
7-1
Spare parts
Parts not included in this document are not recommended for replacement by
the customer.
7-2 Troubleshooting
System troubleshooting
There are a few basic steps for performing system troubleshooting:
• Examine the system, checking the simple things first. Is something
obvious causing the problem (for example, is the instrument and its
cables improperly connected, is there any leakage of fluid, vacuum or
gas?)
• Compare current system operation with the way the system operated
before the problem started. To help identify normal operating
conditions:
– Record a map of the LC system (tubing and electrical connections).
– Keep a daily log.
– Run test samples regularly. Check the instrument performance
with known samples, preferably the ones used for instrument
acceptance.
This illustrates the importance of keeping track of system
parameters and the performance during normal operation.
Troubleshooting is easier if the typical conditions when the system
is operating correctly are known.
For example, are the system tuning parameters similar to those
when a test species was previously run? Are the lens settings
required for optimum sensitivity higher than those previously
obtained? If extreme values have to be used to achieve good results,
this implies that some part of the system requires attention.
• Methodically check and eliminate possible causes to identify the system
parameter that is atypical.
• Refer to the troubleshooting information in the following sections. These
tables enable possible causes of a symptom to be identified and
suggested corrective actions to be taken.
If it is determined that there is a problem related to another system
component (for example HPLC, autosampler, UV detector) refer to the
appropriate operator’s manual.
7-4 Troubleshooting
No peaks on the tune page (no ion beam) (Continued)
7-6 Troubleshooting
Unusually high LC back pressure
Unusually high LC back pressure
Insufficient vacuum
7-8 Troubleshooting
Leaking nitrogen
Warning: To confirm the integrity of the source exhaust system, if
any of the O-rings identified on page 6-30 are renewed, you must
perform a source pressure test as described in the Waters Source
Pressure Test Unit Operator’s Guide.
Leaking nitrogen
7-10 Troubleshooting
Ion mode fault
Drop-down menu options are grayed out, or instrument spontaneously
switches probe type.
Ripple
Peaks and baseline appear to vary cyclically in intensity.
Ripple
7-12 Troubleshooting
IEEE communication errors
IEEE communication errors
Electronic noise
Corrective action
Check that the valleys of peak-peak noise, when ion energies are fully
negative, just touch the baseline. Increase Ion Counting Threshold to suit.
7-14 Troubleshooting
Contacting Waters
Many problems with the Quattro micro API can be easily corrected by the
user. However, if this is not the case, it is necessary to contact Waters.
When contacting Waters, have the following information available:
• Nature of the symptom.
• Quattro micro API serial number.
• Details of flow rate, mobile phases and sample concentrations.
• Details of gas cell operating pressure.
• Tune page settings.
• Software version update reference.
Contents
Topic Page
Overview 8-2
Editing a reference file 8-2
Positive ion 8-3
Negative ion 8-6
8-1
Overview
Calibration reference files consist of two columns of numbers separated by any
number of spaces or TAB characters. The first column contains the reference
peak masses and the second column contains the reference peak intensities.
The reference files listed in this chapter have all ion intensities set to 100%.
Actual ion intensities are not, of course, all 100%, but the calibration software
does not take account of the ion intensities and this is a convenient way to
store the reference files in the required format. However, if required, realistic
intensity values can be entered to improve the appearance of the reference
spectra.
Most samples can be purchased from the Sigma chemical company. To order,
contact Sigma at http://www.sigma.sial.com. This site contains a list of
worldwide Sigma offices, many with local toll-free numbers.
To read the currently selected reference file into the Notepad text editor:
28
+ 606.419 21
+ 808.222 13
+ 1304.969
616.177 20+ 848.583 12+ 1413.633
26
+ 652.989 18
+ 942.758 10
+ 1696.158
25
+ 679.068 17
+ 998.155 9
+ 1884.508
22
+ 771.531 14
+ 1211.829
Polyethylene glycol
PEG + NH4+
Reference Files: PEGH1000.REF, PEGH2000.REF
The purpose of the rubidium iodide is to obtain a peak at m/z 85 (85Rb+) with
an intensity of about 10% of the base peak at m/z 173. Rubidium iodide has
the advantage that no rubidium clusters are formed which may complicate the
85 87
spectrum. Note that rubidium has two isotopes ( Rb and Rb) in the ratio
2.59:1, giving peaks at m/z 85 and 87.
Use reference file NAIRB.REF.
Contents
Topic Page
Warning symbols A-2
Caution symbol A-5
Warnings that apply to all Waters instruments A-6
Electrical and handling symbols A-11
A-1
Warning symbols
Warning symbols alert you to the risk of death, injury, or seriously adverse
physiological reactions associated with an instrument’s use or misuse. Heed
all warnings when you install, repair, and operate Waters instruments.
Waters assumes no liability for the failure of those who install, repair, or
operate its instruments to comply with any safety precaution.
Specific warnings
The following warnings can appear in the user manuals of particular
instruments and on labels affixed to them or their component parts.
Burst warning
This warning applies to Waters instruments fitted with nonmetallic tubing.
This warning applies to certain instruments when they are in Operate mode.
Caution symbol
The caution symbol signifies that an instrument’s use or misuse can damage
the instrument or compromise a sample’s integrity. The following symbol and
its associated statement are typical of the kind that alert you to the risk of
damaging the instrument or sample.
Important: Toute modification sur cette unité n’ayant pas été expressément
approuvée par l’autorité responsable de la conformité à la réglementation peut
annuler le droit de l’utilisateur à exploiter l’équipement.
注意:未經有關法規認證部門允許對本設備進行的改變或修改,可能會使使用者喪失操作該設
備的權利。
注意:未经有关法规认证部门明确允许对本设备进行的改变或改装,可能会使使用者丧失操
作该设备的合法性。
注意:規制機関から明確な承認を受けずに本装置の変更や改造を行うと、本装置のユー
ザーとしての承認が無効になる可能性があります。
警告:當在有壓力的情況下使用聚合物管線時,小心注意以下幾點。
• 當接近有壓力的聚合物管線時一定要戴防護眼鏡。
• 熄滅附近所有的火焰。
• 不要使用已經被壓癟或嚴重彎曲管線。
• 不要在非金屬管線中使用四氫呋喃或濃硝酸或濃硫酸。
• 要了解使用二氯甲烷及二甲基亞楓會導致非金屬管線膨脹,大大降低管線的耐壓能力。
警告:圧力のかかったポリマーチューブを扱うときは、注意してください。
• 加圧されたポリマーチューブの付近では、必ず保護メガネを着用してください。
• 近くにある火を消してください。
• 著しく変形した、または折れ曲がったチューブは使用しないでください。
• 非金属チューブには、テトラヒドロフラン(THF)や高濃度の硝酸または硫酸などを流
さないでください。
• 塩化メチレンやジメチルスルホキシドは、非金属チューブの膨張を引き起こす場合が
あり、その場合、チューブは極めて低い圧力で破裂します。
Attention: L’utilisateur doit être informé que si le matériel est utilisé d’une
façon non spécifiée par le fabricant, la protection assurée par le matériel risque
d’être défectueuses.
警告:使用者必須非常清楚如果設備不是按照製造廠商指定的方式使用,那麼該設備所提供
的保護將被消弱。
警告:使用者必须非常清楚如果设备不是按照制造厂商指定的方式使用,那么该设备所提供
的保护将被削弱。
警告: ユーザーは、製造元により指定されていない方法で機器を使用すると、機器が提供
している保証が無効になる可能性があることに注意して下さい。
Electrical symbols
These can appear in instrument user manuals and on the instrument’s front
or rear panels.
Electrical power on
Standby
Direct current
Alternating current
Fuse
Keep upright!
Keep dry!
Fragile!
Use no hooks!
Contents
Topic Page
Preventing contamination B-2
Items exposed to solvent B-2
Solvents used to prepare mobile phases B-3
B-1
Preventing contamination
For information on preventing contamination, refer to Controlling
Contamination in LC/MS Systems (part number 715001307). Visit
www.waters.com.
Item Material
Corona discharge pin mounting PEEK (Polyetheretherketone)
contact
Gas exhaust port Aluminium
Gas tubes FEP (Fluorinated ethylene
propylene)
Ion block Stainless steel
Ion block support PEEK (Polyetheretherketone)
Isolation valve Gold-plated aluminium/bronze
O-rings Viton or
®
PTFE (Polytetrafluoroethylene)-
encapsulated Viton
Probe adjuster bellows PTFE
(Polytetrafluoroethylene)/Viton
Probe adjuster assembly Anodized aluminium, glass filled
acetal, and stainless steel
Probe shaft PEEK (Polyetheretherketone)
Push-in gas fittings Nickel/brass
Source enclosure Alochromed aluminium
Item Material
Source enclosure view port Toughened plate glass
Waste bottle Polypropylene
.
Contents
Topic Page
Access C-2
Location C-2
Dimensions and clearances C-4
Weights C-6
Operating temperature C-7
Operating humidity C-7
Shipping and storage temperature C-7
C-1
Access
Doors through which the instrument is to be moved should be a minimum of
24 inches (0.6 m) wide. Elevators and corridors must be wide enough to allow
corners to be negotiated. Special arrangements may be required if access to
the laboratory is via a staircase.
Location
Vacuum pump
The rotary pump should be installed beneath, or behind, the Quattro micro
API, within 5 feet (1.5 m) of the rear of the chassis. The rotary pump is fitted
with a 6.5 feet (2 m) power cable which plugs into the rear of the instrument’s
chassis. It is recommended that the rotary pump is elevated 6 to 8 inches(15 to
20 cm) above the floor to aid in routine maintenance, such as changing pump
oil.
If the rotary pump is sited under the instrument bench, an access slot may
need to be drilled in the bench top for routing of vacuum tubing from the
Data system
The data system must be located within 16 feet (5 m) of the mass spectrometer
to allow connection of the communication cables. The data system power
cables are approximately 6.5 feet (2 m) in length. A typical layout for the data
system is shown in the figure “Data system” on page C-5.
LC system
Be sure to allow adequate space to the left of the instrument for the HPLC
system. Refer to the associated user manuals for individual space
requirements.
Location C-3
Dimensions and clearances
Horizontal clearances
Vertical clearances
Height
23 in. (572 mm)
Length
34.6 in. (880 mm)
Width
15.4 in.(390 mm).
Weights
The bench must be able to support the total weight of the system:
Instrument: 253 lb (115 kg)
Data system (computer, monitor and optional printer): 110 lb (50 kg)
Total Weight: 363 lb (165 kg)
Operating temperature
The ambient temperature range for instrument operation is 15 to 30°C (59 to
86°F), with short-term variation (1.5 h) not exceeding 2°C (4°F). The ambient
temperature range for rotary pump operation is 15 to 40°C (59 to 104°F).
The maximum overall heat dissipation into the room is 3.0 kW. This figure
does not take into account ancillary equipment such as LC systems.
Air-conditioning may have to be installed, or upgraded, to accommodate the
additional heat load into the room.
Operating humidity
The relative humidity range for instrument operation is 20 to 80%. It is
recommended that the instrument be situated in a draft-free position in an
air-conditioned laboratory free from excessive amounts of dust.
Contents
Topic Page
Installation D-2
Electrical safety D-4
D-1
Installation
It is recommended that the electrical installation in the laboratory include a
wall-mounted emergency isolator switch of the correct rating, which is
protected by a correctly-rated circuit breaker. The isolator should be clearly
marked “Mass Spectrometer” and ideally positioned with clear access at all
times for emergency switch-off, without risk of slipping or tripping. Ideally,
the isolator should be capable of being locked in the OFF position to prevent
unauthorized personnel from switching the instrument on. The Quattro micro
API must have an earthed power supply that satisfies the voltage and current
specifications for that country, detailed below:
Canada:
The user must supply a fused supply, rated at 15 A.
UK:
The cable is fused at 13 A.
Japan:
A single phase, 3-wire 200 V 5% phase/neutral supply, rated between 13
and16 A must be supplied. Additionally, a transformer must be ordered
from Waters to boost the input voltage.
Australia and New Zealand:
The equipment has been designed in compliance with the international safety
standard IEC1010-1 (EN61010-1). To be fully effective, the building
installation must comply with AS 3000: Electrical Installations for
Australia/New Zealand.
Rest of World:
The user must supply a fused supply, rated between 13 and 16 A.
USA EUROPE UK
NEMA L6-15P, 15 A 2-pin Europlug 3-pin UK plug, fused at
conforming to CEE7 13 A, BS1633
standards
For countries with a 60 Hz AC. supply (such as the USA), if the supply voltage
is less than 208 V AC., it is recommended that a step-up transformer is used
to boost the voltage to 208 to 230 V AC.
The mains supply must have at least a ±10% brown out capability, i.e. 10% for
0.3 seconds and transient ≤20ms to half mains voltage.
Appropriate sockets to mate with the plugs shown must be provided.
If a plug is required other than that shown, the user must supply the plug and
the appropriate socket.
The instrument is shipped with a 8 feet(2.5 m) power cable that must be
plugged into the rear of the chassis. The Quattro micro API power ON/OFF
switch is located on the front panel.
Line frequency
50 Hz, 47 to 53 Hz.
60 Hz, 57 to 63 Hz.
Installation D-3
Fuse rating
10 A, 250 V AC.
Electrical safety
The Waters Quattro micro API conforms to European standard
EN61010-1:2001, Safety requirements for electrical equipment for
measurement, control, and laboratory use - Part 1: General requirements.
The instrument is categorized as Pollution Category 1 and Installation
Category 2.
Index-1
G Negative ion 8-6
Grid 3-10
O
H Operate LED 1-15
handling symbols A-12
Herbicides 1-10 P
Hexapole assembly 6-18 Parent ion 1-8
PC link 1-17
I Peptides 1-7
Instrument data thresholds 4-10 Pesticides 1-10
intended use ix Pharmacokinetic studies 1-10
Intensity 3-10 Pirani gauge 1-3
Ion block 6-13 Precursor ion 1-8
Ion Counting Threshold 4-12 preventing contamination B-2
Product ion 1-6
M Profile data 4-11
Mass flow controllers 1-12 Profile data - spike removal 4-14
Mass Measure parameters 5-12 purpose and audience viii
mass spectrometer shock hazard A-4
MassLynx 1-4 R
materials of construction B-1 Readbacks 3-17
MaxEnt 4-11 Rear panel connections 1-16
mobile phases, solvents used in Reserpine 1-7, 1-9, 1-10
preparation of B-3 Rotary pump - gas ballasting 6-4
Mode 3-13 Rotary pump oil - changing 6-6
Monitoring acquisitions 4-39 Rotary pump oil - checking 6-5
MRM 1-10 Routine maintenance 6-4
MRM function 4-32 Rubidium iodide 8-5
MS operating modes 1-5
MS to MS/MS switching 4-35 S
MS/MS operating modes 1-6 safety advisories A-1
MS/MS scanning functions 4-30 Sample inlet 1-3
MS/MS to MS switching 4-37 Sample Inlet 1-3
MS1 mode 1-5 Scan Report window 4-8
MS2 mode 1-5 Scan speed compensation calibration
Multiple Reaction Monitoring 1-10 5-4
Mux Interface 1-17 Scanning calibration 5-4
Scope 3-13
N SIR data 4-12
Nebulizer gas 1-12 SIR function 4-28
Index-2
Sodium iodide 8-5 Z
Software 1-4 Zero level 3-17
Solvent delay 4-21
solvents I
compatible B-1
exposure of SQ detector
components to B-2
use in mobile phases B-3
Source enclosure 6-12
Source voltages 3-18
Spectrum - real-time update 4-9
Starting 2-2
Static calibration 5-4
Status display 1-15
Survey function 4-33
symbols
caution A-5
electrical A-11
handling A-12
warning A-2
System Manager 4-16
T
Toxicology 1-10
Trace 3-10
Tune page 3-2
Tune Parameter settings - changing
3-18
Tuning 3-2
Turbomolecular pump 1-3
U
UV detector 1-16
V
Vacuum LED 1-15
Vacuum system 1-3
W
warning symbols A-2, A-6
Index-3
Index-4