Analytical
Balances
Models: VE-204 and VE-303
Content
1) Introduction
2) Technical specifications
3) Installation diagram
4) Operation of equipment
a) Preparation
b) Operation
b.1) VE-04
b.2) VE-03
Please read fully the manual before using the balance. This balance has electronics
multifunctions using a system MCS- S4 patented by our company has heavy, ct,
ounces, grams, etc.. Including automatic calibration, adjustment of stability this
balance has RS232C port to connect a computer and printing.
Proprietary Medicinal Products, Industry, etc. Offering fast, efficient and accurate
weighing. Dressy
12
2. Technical specifications
Note: This balance light when the power button (ON) is pressed. The panel
only shows activation in the display. The connection of electric power must be
disconnected if the balance is not used in a certain time (5 days approx.) If you
use it daily there isn’t need to disconnect, you have to press the (ON) button
located in the display. jorge
12
12
3. Installation diagram Erick josheep
1) Display
2) Weighing platform
3) Platform
3
1
4) Interface 5
4
5) Electric connection
CAL
On
UNI
OF
T
COU
IN
T
ASD TA PR
Sensitivity R T Screw and cover
Levelling feet
12
4. Operation of equipment
a) Preparation
Unwrapping perfectly the equipment, extracts all the accessories and free
pieces of the packing of polystyrene foams
Place And properly assemble the equipment as indicated in the instructions
Place the balance in the table proper working, away from vibrations, light of
the Sun, air, wind and unfavourable conditions that alter the perfect work of
the balance
Environment -Temperature 20 °C +- 5 ° C
Relative -humidity 50% -75%
Suitable -Voltage 110V + - 11 V, 50 HZ Erick Joshep
b) b)Operation
For the balance VE-204 press "ON" and is show in the display of form
descendant 30 minutes, wait the time displayed and after finished the time
can press the button "ON"
At Turn its balance displayed on the display "0.0000g" means that your
balance was successfully calibrated and can work with the team. Note:
When in the heating process by pressing the "tare" you can open the
equipment without calibrate the balance
Press gently "OFF" (Switched off) and on the display will show "OFF"
The scale in protection way of screen "Screen Protection", later presses "TAR".
Place the container in the plate of the tray with the weight.
Before operating the Balance check that the bubble is positioned in the middle
of the circle, and this is achieved adjusted their legs thereof.
12
Press the "TAR" tare button, the display will show zeroes. The tare is completed
Press "TAR" tara again and the display will change showing zeroes Tara is ready
Salvador
Connect and turn "ON", for recommendation one asks to leave her an hour
ignited before using it (in case that the display does not show anything).
Gently press the "ON" button and the display on Czech "the
function display, after 2 seconds, the balance model appear and
then light up and then its heavy function.
Press the "OFF" button and the display will disappear. If the balance is not
used a certain time, please disconnect the unit from the power supply.
"TAR". Place the container on the pan of the balance and the weight keeps
the following.
Press "TAR" tare, the display changes and showing zeros on the display. The
tare is ready.
When you remove the weight, a negative value is shown on the display
12
Press "TAR" again, the display will mark zeros.
If the accuracy of the reading needs to be 0.1mg, check at the end of the
display and it indicates the unit of measure.
Press the button "ERY" until the display shows the following figure
And every value is explained example; "g" means grams, "ct" means value of
heavy so called Carat, "ozt" ounce troy, "GN" grain , "oz" ounce, "N" Newton,
"Ib" pound, etc. The unit will appear by means of RNG.
The integration time is divided as follows. INT-0 fast, INT-1 very short, INT-2
Short, INT-3 long. The selection of the integration time is the same as RNG.
Like the adjustment knob of time INT, there are 4 other ways to provide this to
adjust sensitivity.
ASD-0 higher. ASD-1 High, ASD-2 Medium, Low ASD-3. ASD-0 is used in test
equipment should not be used in user.
12
The selection of the sensitivity is the same as RNG.
Normally, your balance should be calibrated after long unused or stored, with a
sudden movement, or a different environment that you work, is when
calibration for accurate measurements of heavy is recommended.
Remove all objects that are on the weighing platform, press "CAL" is displayed
until the "CAL int" display, then automatically unppshed and the balance will go
to calibration mode automatically. When the display shows "0.0000g" means
that the calibration was successful. The balance will go to normal working
mode.
Remove all objects on the weighing platform, press "CAL" until is displayed on
the display "CRL-E", the balance will go the way of external calibration, it is
displayed flashing in the "CAL-200" display . Place the external weight (200 g)
in the center of the plate, then the balance will go to calibration status. After a
few seconds, it will be displayed on the display "200.000g". Remove the
calibration weight and the display indicates "0.000g". Balance Calibrated and
ready for use. Note: It is recommended 2 times this procedure to ensure
perfect calibration of equipment. The calibration sequence as shown below.
12
g. Output settings PRT Marco /mauricio
PRT and 4 hold modes will be shown on the display PRT-0: The indefinite mode
output setting time. PRT press while the weight shown in the output interface.
Attention: Press the button quickly or possibly the following display mode
h. Parts count
The balance has a counting function. The average is: 5. I0, 25 and 50 with a
total of four averages. COU press and hold, the display circulated as follows:
If functions is Required Heavy press when the display shows “COU-00” and
indicate a wait state "_" at least "0.000g" is displayed on the display. If the
percentage of the required value is five press when the display shows COU-5,
and place 5 items on the plate and so on depending 5,10, 25 and 50.
12
i. Weighing, Tare, add objects, reading of deviations
Susan /narciso
After selecting the mode (all modes can be used for weighing, after
Weighing
Put the container on the plate, the weight is shown on the display.
Press TAR and shows zero, that's tara. Despite putting objects in the
Tare
container or tray weighing weight value shown is the net weight of the
object only.
Place objects to weigh on the weighing plate one by one with the TARE
weight Cumulative
mode and cleaned one by one. Remove all objects that have been
weighed, the absolute value of the total weight is to be displayed.
Apply INT-0 mode and put the tray or container on the plate, then press
parts count
the TARE button. Add the objects to be weighed (either liquid or solid)
the container one by one continuously reading weighing obtaining
quickly. When added objects reach "0" on the right side of the display
of balance and the value required by the user. When you add all
objects so with seal, the net weight of each object can be measured by
Put the weight or sample on the pan and press TARE, then remove the
deviations Reading of
12
5. Maintenance and troubleshooting solution pepe
The scales themselves are equipment to be used in delicate, have always clean
the platform. Wipe with a soft cloth, never use agents or solutions that can
damage the equipment.
Númb
er Problem Case Solution
-It is not on -Check and turn
The display does not -Switch is not lit -Press' "ON"
1
turn -Interference -Check Switch and
connection
-Immediately
-Overloaded equipment
reduce weight
Medium shows -Calibration memory is
-Recalibration with
2 Superior Display damaged
proper procedure
Power -The dish was placed
-Remove the plate
incorrectly
and reinstall
-No plate Installed
Medium shows -Properly position
3 -Dish improperly installed
On lower display the weighing plate
-Air flow -Check the glass
The weighing is not -The worktable is not components are
4 stable, is in constant stable closed
motion -Integration time is very -Change Long Time
short Integration
-No zero before weighing
frame
-Press button tare
-The scale is not red, or the
The result is not -Recalibrate
5 calibration weight is not
correct -Change electrical
correct
connection
The line voltage is
incorrect
12
Númb
er Problem Case Solution
6 Display -Interference -Come back to
It remains in a certain -Erroneous line of connect the balance
digit or Indicates a Voltage and check connection
-change The line
meaningless code voltage
7 Display is in standby -The position of the -Put Less Sensitivity
balance is not correct -Improve the work
(Vibration and draught) environment
- very high sensibility
8 El Display shows "CRL - There is an object on -Remove The
Err" the platform that does object
not allow calibration -Clean And
-Press CTRL Calibrating
Q Display shows Err-1. Err- -Instan Interference --Turn Off and turn
2 - A critical error in the on the balance
balance -Bring a Technical
Service
10 The measure of - It is not calibrated -Balance calibrate
weighed in the right -The Number of
side of the display is not internal calibration
the sample memory has been
deleted
Octavio
12
6. Interface Ruben
The information that adopts the Interface Refers to a program (9-core RS-232C)
,comes with 2 Interface RS232C universal connections which can be connected
to a microprocessor or possibly any printer Pin and corresponding signals are:
7. RS232C
12
1 2 3 4 5 6 7 8 9 IO 11 12 13 14 15 16
*/Space
Space
Unt 1
Unt 2
Type
Data
Data
Data
Data
Data
Data
Data
Data
+/-
CR
LF
8. Accessories extras Dressy/jorge