Professional Documents
Culture Documents
Textbook of Clinical Echocardiography Endocardiography 5th Edition Ebook PDF
Textbook of Clinical Echocardiography Endocardiography 5th Edition Ebook PDF
/
Echocardiography Review Guide: Companion to the Textbook of Clinical
Echocardiography, Second Edition
9
Catherine Otto, Rebecca Schwaegler, and Rosario Freeman
i r
The Practice of Clinical Echocardiography, Fourth Edition
9
h
Catherine Otto
ta
Practical Echocardiography Series
r/
Series Editor: Catherine Otto
e
Volumes Included in This Series:
s
Advanced Approaches in Echocardiography
/r u
Linda Gillam and Catherine Otto
Intraoperative Echocardiography
.t c
Donald Oxorn
a
Echocardiography in Heart Failure
k
Martin St. John Sutton and Susan Wiegers
/: /
Echocardiography in Congenital Heart Disease
s
Mark Lewin and Karen Stout
tt p
h
xiv
Glossary
/
LVH = left ventricular hypertrophy R o = radius of microbubble
LVID = left ventricular internal dimension ROA = regurgitant orifice area
9
LVOT = left ventricular outflow tract RPA = right pulmonary artery
RSPV = right superior pulmonary vein
9
M-mode = motion display (depth versus time) RSV = regurgitant stroke volume
r
MAC = mitral annular calcification
i
RV = right ventricle
MI = myocardial infarction RVE = right ventricular enlargement
MR = mitral regurgitation
h
RVH = right ventricular hypertrophy
MS = mitral stenosis RVol =regurgitant volume
a
MV = mitral valve RVOT = right ventricular outflow tract
t
MVA = mitral valve area
r/
MVL = mitral valve leaflet s = second
MVR = mitral valve replacement SAM = systolic anterior motion
SC = subcostal
e
n = number of subjects SEE = standard error of the estimate
NBTE = nonbacterial thrombotic endocarditis SPPA = spatial peak pulse average
s
NCC = noncoronary cusp SPTA = spatial peak temporal average
/r u
ΔP = pressure gradient SSN = suprasternal notch
P = pressure ST = septal thickness
PA = pulmonary artery STJ = sinotubular junction
PAP = pulmonary artery pressure STVL = septal tricuspid valve leaflet
.t c
PCI = percutaneous coronary intervention SV = stroke volume or sample volume
PDA = patent ductus arteriosus or posterior (depends on context)
descending artery (depends on context) SVC = superior vena cava
a
PE = pericardial effusion T½ = pressure half-time
PEP = preejection period TD = thermodilution
k
PET = positron-emission tomography
/: /
TEE = transesophageal echocardiography
PISA = proximal isovelocity surface area TGA = transposition of the great arteries
PLAX = parasternal long-axis TGC = time-gain compensation
PM = papillary muscle Th = wall thickness
s
PMVL = posterior mitral valve leaflet TL = true lumen
post = posterior (or inferior-lateral) ventricular wall TN = true negatives
tt p
PR = pulmonic regurgitation TOF = tetralogy of Fallot
PRF = pulse repetition frequency TP = true positives
PRFR = peak rapid filling rate TPV = time to peak velocity
PS = pulmonic stenosis
h
TR = tricuspid regurgitation
PSAX = parasternal short-axis TS = tricuspid stenosis
PV = pulmonary vein TSV = total stroke volume
PVC = premature ventricular contraction TTE = transthoracic echocardiography
PVD = pulmonary vein diastolic velocity TV = tricuspid valve
PVR = pulmonary vascular resistance
PVD = pulmonary vein diastolic velocity v = velocity
PWT = posterior wall thickness V = volume or velocity (depends on context)
VAS = ventriculo-atrial septum
Q = volume flow rate Veg = vegetation
Q p = pulmonic volume flow rate Vmax = maximum velocity
Q s = systemic volume flow rate VSD = ventricular septal defect
r = correlation coefficient VTI = velocity-time integral
R = ventricular radius WPW = Wolff-Parkinson-White syndrome
RFR = regurgitant instantaneous flow rate
RA = right atrium Z = acoustic impedance
xv
Glossary
/
cm-mg-s2
μ mu Micro-
Resistance dyne · s · cm-5 Measure of vascular
9
π pi Mathematical constant resistance
(approx. 3.14)
9
Time s Second
ρ rho Tissue density ms Millisecond (1/1000 s)
i r
σ sigma Wall stress μs Microsecond
τ tau Time constant of Ultrasound W/cm2 Where watt (W) =
h
ventricular relaxation intensity mW/cm2 joule per second and
joule = m2 · kg · s-2
a
(unit of energy)
r/ t
Velocity (v) m/s Meters per second
cm/s Centimeters per second
UNITS OF MEASURE Velocity- cm Integral of the Doppler
e
time velocity curve (cm/s)
Variable Unit Definition integral over time (s), in units
s
(VTI) of cm
Amplitude dB Decibels = a logarithmic
/r u
scale describing Volume cm3 Cubic centimeters
the amplitude mL Milliliter, 1 mL = 1 cm3
(“loudness”) of the L Liter = 1000 mL
sound wave Volume Rate of volume flow
.t c
Angle degrees Degree = (π/180)rad. flow rate across a valve or in
Example: intercept (Q) cardiac output
angle L/min L/min = liters per minute
a
mL/s mL/s = milliliters per
Area cm2 Square centimeters.
second
A 2D measurement
k
(e.g., end-systolic Wall stress dyne/cm2 Units of meridional or
/: /
area) or a calculated circumferential wall
value (e.g., continuity stress
equation valve area) kdyn/cm2 Kilodynes per cm2
s
Frequency Hz Hertz (cycles per kPa Kilopascals where
(f) second) 1 kPa = 10 kdyn/cm2
tt p
kHz Kilohertz = 1000 Hz
MHz Megahertz =
1,000,000 Hz
h
Length cm Centimeter (1/100 m)
mm Millimeter (1/1000 m or
1/10 cm)
KEY EQUATIONS
/
Ultrasound Physics
f = cycles/s = Hz
9
Frequency
Wavelength λ = c / f = 1.54/f (MHz)
υ = c × Δf/ [2FT (cosθ)]
9
Doppler equation
Bernoulli equation ΔP = 4V 2
i r
LV Imaging
Stroke volume SV = EDV − ESV
EF( % ) = (SV / EDV) × 100 %
h
Ejection fraction
Wall stress σ = PR/2Th
a
Doppler Ventricular Function
t
Stroke volume SV = CSA × VTI
dP/dt = 32 mm Hg / time from 1 to 3 m/s of MR CW jet(sec)
r/
Rate of pressure rise
Myocardial performance index MPI = (IVRT + IVCT) / SEP
Pulmonary Pressures and Resistance
PAPsystolic = 4(VTR )2 + RAP
e
Pulmonary systolic pressure
PAP (when PS is present) PAPsystolic = [4(VTR )2 + RAP] − Δ PRV − PA
PAPmean = Mean Δ PRV − RA + RAP
s
Mean PA pressure
Diastolic PA pressure PAPdiastolic = 4(VPR )2 + RAP
/r u
Pulmonary vascular resistance PVR ≅ 10(VTR )/VTIRVOT
Aortic Stenosis
Maximum pressure gradient (integrate over ejection Δ Pmax = 4(Vmax )2
period for mean gradient)
AVA(cm2 ) = [π(LVOTD / 2)2 × VTILVOT ] / VTIAS-Jet
.t c
Continuity equation valve area
Simplified continuity equation AVA(cm2 ) = [π(LVOTD / 2)2 × VLVOT ] / VAS-Jet
Velocity ratio Velocity ratio = VLVOT /VAS-Jet
Mitral Stenosis
a
Pressure half-time valve area MVADoppler = 220 / T½
Aortic Regurgitation
k
Total stroke volume TSV = SVLVOT = (CSALVOT × VTILVOT )
/: /
Forward stroke volume FSV = SVMA = (CSAMA × VTIMA )
Regurgitant volume RVol = TSV − FSV
Regurgitant orifice area ROA = RSV / VTIAR
s
Mitral Regurgitation
Total stroke volume TSV = SVMA = (CSAMA × VTIMA )
tt p
(or 2D or 3D LV stroke volume)
Forward stroke volume FSV = SVLVOT = (CSALVOT × VTILVOT )
Regurgitant volume RVol = TSV − FSV
Regurgitant orifice area ROA = RSV/VTIAR
h
PISA method
Regurgitant flow rate RFR = 2πr2 × Valiasing
Orifice area (maximum) ROAmax = RFR / VMR
Regurgitant volume RV = ROA × VTIMR
Aortic Dilation
Predicted sinus diameter
Children (<18 years): Predicted sinus dimension = 1.02 + (0.98 BSA)
Adults (18-40 years): Predicted sinus dimension = 0.97 + (1.12 BSA)
Adults (>40 years): Predicted sinus dimension = 1.92 + (0.74 BSA)
Ratio = Measured maximum diameter / Predicted maximum diameter
Pulmonary (Q p) to Systemic (Q s) Shunt Ratio
Q p : Q s = [CSAPA × VTIPA ] / [CSALVOT × VTILVOT ]
xvii
1 Principles of Echocardiographic Image
Acquisition and Doppler Analysis
n understanding of the basic principles of is referred to the Suggested Reading at the end of
A ultrasound imaging and Doppler echocardiog-
raphy is essential both during data acquisition
the chapter for more information on these subjects.
Because the details of image processing, artifact for-
and for correct interpretation of the ultrasound infor- mation, and Doppler physics become more mean-
mation. Although, at times, current instruments pro- ingful with experience, some readers may choose to
vide instantaneous images so clear and detailed that return to this chapter after reading other sections of
it seems as if we can “see” the heart and blood flow this book and after participating in some echocardio-
directly, in actuality, we always are looking at images graphic examinations.
and flow data generated by complex analyses of ultra-
sound waves reflected and backscattered from the
patient’s body. Knowledge of the strengths, and more ULTRASOUND WAVES
importantly, the limitations, of this technique is critical
for correct clinical diagnosis and patient management. Sound waves are mechanical vibrations that induce
On the one hand, echocardiography can be used for alternate refraction and compression of any physical
decision making with a high degree of accuracy in a medium through which they pass (Fig. 1-1). Like other
variety of clinical settings. On the other hand, if an waves, sound waves are described in terms of (Table 1-1):
ultrasound artifact is mistaken for an anatomic abnor-
n requency: cycles per second, or hertz (Hz)
F
mality, a patient might undergo needless, expensive,
n Velocity of propagation
and potentially risky other diagnostic tests or thera-
n Wavelength: millimeters (mm)
peutic interventions.
n Amplitude: decibels (dB)
In this chapter, a brief (and necessarily simplified)
overview of the basic principles of cardiac ultrasound Frequency ( f ) is the number of ultrasound waves in a
imaging and flow analysis is presented. The reader 1-second interval. The units of measurement are hertz,
1
2
Chapter 1 | Principles of Echocardiographic Image Acquisition and Doppler Analysis
abbreviated Hz, which simply means cycles per second. each type of tissue. For example, the velocity of propa-
A frequency of 1000 cycles/s is 1 kilohertz (KHz), and gation in bone is much faster (about 3000 m/s) than
1 million cycles/s is 1 megahertz (MHz). Humans can in lung tissue (about 700 m/s). However, the velocity
hear sound waves with frequencies between 20 Hz and of propagation in soft tissues, including myocardium,
20 kHz; frequencies higher than this range are termed valves, blood vessels, and blood is relatively uniform,
ultrasound. Diagnostic medical ultrasound typically uses averaging about 1540 m/s.
transducers with a frequency between 1 and 20 MHz. Wavelength is the distance from peak to peak of an
The speed that a sound wave moves through the ultrasound wave. Wavelength can be calculated by
body, called the velocity of propagation (c), is different for dividing the frequency ( f in Hz) by the propagation
velocity (c in m/s).
/
Wavelength Propagation velocity (m/s) λ=c f (1-1)
λ
Since the propagation velocity in the heart is con-
stant at 1540 m/s, the wavelength for any transducer
frequency can be calculated (Fig. 1-2) as:
Amplitude (dB)
// /
λ (mm) = [1540m s f (Hz)] 1000 mm/m
or as:
/
λ (mm) = 1.54 f
For example, the wavelength emitted by a 5 MHz
transducer can be calculated as:
1s / /
cycles/s = Hz λ = 1540 m s ÷ 5,000,000 cycle s = 0.000308 m
Figure 1–1 Schematic diagram of an ultrasound wave. = 0.308 mm
Frequency (f) The number of cycles per Transducer frequencies Different transducer frequencies
second in an ultrasound are measured in MHz are used for specific clinical
wave: (1,000,000 cycles/s). applications because the
f = cycles/s = Hz Doppler signal frequencies transmitted frequency affects
are measured in KHz (1000 ultrasound tissue penetration,
cycles/s). image resolution, and the
Doppler signal.
Velocity of The speed that ultrasound The average velocity of The velocity of propagation is
propagation (c) travels through tissue ultrasound in soft tissue is similar in different soft tissues
about 1540 m/s. (blood, myocardium, liver, fat,
etc.) but is much lower in lung
and much higher in bone.
Wavelength (λ) The distance between Wavelength is shorter Image resolution is greatest
ultrasound waves: with a higher-frequency (about 1 mm) with a
λ = c/f = 1.54/f (MHz) transducer and longer shorter wavelength (higher
with a lower-frequency frequency).
transducer. Depth of tissue penetration
is greatest with a longer
wavelength (lower frequency).
Amplitude (dB) Height of the ultrasound A log scale is used for dB. A very wide range of amplitudes
wave or “loudness” On the dB scale, 80 dB can be displayed using a gray-
measured in decibels represents a 10,000- scale display for both imaging
(dB) fold and 40 dB indicates and spectral Doppler.
a 100-fold increase in
amplitude.
3
Principles of Echocardiographic Image Acquisition and Doppler Analysis | Chapter 1
Wavelength is important in diagnostic applications scale. Decibels (dB) are familiar to all of us as the stan-
for at least two reasons: dard description of the loudness of a sound. Decibels
are logarithmic units based on a ratio of the measured
n I mage resolution is no greater than 1 to 2 wave-
amplitude (A2) to a reference amplitude (A1) such that:
lengths (typically about 1 mm). /
n The depth of penetration of the ultrasound dB = 20 log (A2 A1 ) (1-3)
wave into the body is directly related to wave-
Thus, a ratio of 1000 to 1 is:
length; shorter wavelengths penetrate a shorter
distance than longer wavelengths. 20 × log (1000) = 20 × 3 = 60 dB
0.5
Penetration (cm)
1.0
10
.5
1.5
0 0
0 1 2.5 3.5 5 7.5 10 15 20
Transducer frequency (MHz)
10
2
1
0
0 6 20 40 60 80 100
Decibels
4
Chapter 1 | Principles of Echocardiographic Image Acquisition and Doppler Analysis
Acoustic A characteristic of each Lung has a low density and Ultrasound is reflected from
impedance (Z) tissue defined by slow propagation velocity, boundaries between
tissue density (r) and whereas bone has a high tissues with differences in
propagation of velocity density and fast propagation acoustic impedance (e.g.,
(c) as: velocity. Soft tissues have blood versus myocardium).
Z=r×c smaller differences in
tissue density and acoustic
impedance.
Reflection Return of ultrasound signal Reflection is used to generate Reflection is greatest when
to the transducer from a 2D cardiac images. the ultrasound beam is
smooth tissue boundary perpendicular to the tissue
interface.
Scattering Radiation of ultrasound in The change in frequency The amplitude of scattered
multiple directions from of signals scattered from signals is 100 to 1000
a small structure, such moving blood cells is the times less than reflected
as blood cells basis of Doppler ultrasound. signals.
Refraction Deflection of ultrasound Refraction is used in transducer Refraction in tissues results in
waves from a straight design to focus the double image artifacts.
path because of ultrasound beam.
differences in acoustic
impedance
Attenuation Loss in signal strength Attenuation is frequency A lower-frequency transducer
due to absorption of dependent with greater may be needed for apical
ultrasound energy by attenuation (less penetration) views or in larger patients
tissues at higher frequencies. on transthoracic imaging.
Resolution The smallest resolvable Resolution has three Axial resolution is most precise
distance between two dimensions: along the length (as small as 1 mm), so
specular reflectors on an of the beam (axial), lateral imaging measurements are
ultrasound image across the image (azimuthal) best made along the length
and in the elevational plane. of the ultrasound beam.
5
Principles of Echocardiographic Image Acquisition and Doppler Analysis | Chapter 1
act as specular, or “mirrorlike,” reflectors. The amount Although experimental studies show differences in
of ultrasound reflected is constant for a given interface, backscattering with changes in hematocrit, variation
although the amount received back at the transducer over the clinical range has little effect on the Dop-
varies with angle because (like light reflected from a pler signal. Similarly, the size of red blood cells and
mirror) the angle of incidence and reflection is equal. the compressibility of blood cells and plasma do not
Thus, optimal return of reflected ultrasound occurs change significantly. Thus, the primary determinant
at a perpendicular angle (90°). Remembering this fact of scattering is transducer frequency.
is crucial for obtaining diagnostic ultrasound images. Scattering also occurs within tissues, such as the
It also accounts for ultrasound “dropout” in a two- myocardium, from interference of backscattered sig-
dimensional (2D) or three-dimensional (3D) image nals from tissue interfaces smaller than the ultrasound
when too little or no reflected ultrasound reaches the wavelength. Tissue scattering results in a pattern of
transducer resulting from a parallel alignment between speckles; tissue motion can be measured by tracking
the ultrasound beam and tissue interface. these speckles from frame to frame, as discussed in
Chapter 4.
Scattering
Scattering of the ultrasound signal, instead of reflec-
Refraction
tion, occurs with small structures, such as red blood Ultrasound waves can be refracted—deflected from a
cells suspended in fluid, because the radius of the cell straight path—as they pass through a medium with
(about 4 µm) is smaller than the wavelength of the a different acoustic impedance. Refraction of an
ultrasound signal. Unlike a reflected beam, scattered ultrasound beam is analogous to refraction of light
waves as they pass through a curved glass lens (e.g.,
prescription eyeglasses). Refraction allows enhanced
image quality by using acoustic “lenses” to focus the
ultrasound beam. However, refraction also occurs in
unplanned ways during image formation, resulting in
Transducers ultrasound artifacts, most notably the “double-image”
artifact.
Scattering from
moving blood cells Attenuation
Attenuation is the loss of signal strength as ultrasound
interacts with tissue. As ultrasound penetrates into the
Reflection
Specular body, signal strength is progressively attenuated because
reflector of absorption of the ultrasound energy by conversion
to heat, as well as by reflection and scattering. The
degree of attenuation is related to several factors,
Refraction including the:
Attenuation
the depth of ultrasound penetration. Refraction, a change in direction of the The attenuation coefficient (α) for each tissue is
ultrasound wave, results in imaging artifacts. related the decrease in ultrasound intensity (measured
6
Chapter 1 | Principles of Echocardiographic Image Acquisition and Doppler Analysis
in dB) from one point (I1) to a second point (I2) sepa- and expands, generating an ultrasound wave. The fre-
rated by a distance (l) as described by the equation: quency that a transducer emits depends on the nature
and thickness of the piezoelectric material.
I2 = I1 e − 2αl (1-5)
Conversely, when an ultrasound wave strikes the
The attenuation coefficient for air is very high (about piezoelectric crystal, an electric current is generated.
1000×) compared to soft tissue so that any air between Thus, the crystal can serve both as a “receiver” and
the transducer and heart results in substantial signal as a “transmitter.” Basically, the ultrasound transducer
attenuation. This is avoided on transthoracic exami- transmits a brief burst of ultrasound and then switches
nations by use of a water-soluble gel to form an airless to the “receive mode” to await the reflected ultrasound
contact between the transducer and the skin; on trans- signals from the intracardiac acoustic interfaces. This
esophageal echocardiography (TEE) examination, atten- cycle is repeated temporally and spatially to generate
uation is avoided by maintaining close contact between ultrasound images. Image formation is based on the
the transducer and esophageal wall. The air-filled lungs time delay between ultrasound transmission and return
are avoided by careful patient positioning and the use of of the reflected signal. Deeper structures have a longer
acoustic “windows” that allow access of the ultrasound time of flight than shallower structures, with the exact
beam to the cardiac structures without intervening lung depth calculated based on the speed of sound in blood
tissue. Other intrathoracic air (e.g., pneumomediasti- and the time interval between the transmitted burst of
num, residual air after cardiac surgery) also results in ultrasound and return of the reflected signal.
poor ultrasound tissue penetration because of attenua- The burst, or pulse, of ultrasound generated by the
tion, resulting in suboptimal image quality. piezoelectric crystal is very brief, typically 1 to 6 µs,
The power output of the transducer is directly related because a short pulse length results in improved axial
to the overall degree of attenuation. However, an increase (along the length of the beam) resolution. Damping
in power output may cause thermal and mechanical material is used to control the ring-down time of the
bioeffects as discussed in Bioeffects and Safety, p. 27. crystal and, hence, the pulse length. Pulse length also
Overall attenuation is frequency-dependent such is determined by frequency because a shorter time
that lower ultrasound frequencies penetrate deeper is needed for the same number of cycles at higher
into the body than higher frequencies. The depth of frequencies. The number of ultrasound pulses per
penetration for adequate imaging tends to be limited second is called the pulse repetition frequency, or PRF.
to approximately 200 wavelengths. This translates The total time interval from pulse to pulse is called
roughly into a penetration depth of 30 cm for a 1-MHz the cycle length, with the percent of the cycle length
transducer, 6 cm for a 5-MHz transducer, and 1.5 cm used for ultrasound transmission called the duty factor.
for a 20-MHz transducer, although diagnostic images Ultrasound imaging has a duty factor of about 1%
at depths greater than these postulated limits can be compared to 5% for pulsed Doppler and 100% for
obtained with state-of-the-art equipment. Thus, attenu- continuous-wave (CW) Doppler. The duty factor is a
ation, as much as resolution, dictates the need for a par-
ticular transducer frequency in a specific clinical setting.
For example, visualization of distal structures from the Transducer Ultrasound
Pulse
apical approach in a large adult patient often requires
a low-frequency transducer. From a TEE approach, the Damping Acoustic Pulse
same structures can be imaged (at better resolution) with material lens length
a higher-frequency transducer. The effects of attenua-
Cable
tion are minimized on displayed images by using differ-
ent gain settings at each depth, an instrument control
called time-gain (or depth-gain) compensation.
λ
TRANSDUCERS Piezoelectric Impedance
crystal matching
Piezoelectric Crystal Figure 1–5 Schematic diagram of an ultrasound transducer. The
Ultrasound transducers use a piezoelectric crystal both piezoelectric crystal both produces and receives ultrasound signals, with
the electric input-output transmitted to the instrument via the cable. Damp-
to generate and to receive ultrasound waves (Fig. 1-5). ing material allows a short pulse length (improved resolution). The shape
A piezoelectric crystal is a material (such as quartz or a of the piezoelectric crystal, an acoustic lens, or electronic focusing (with a
titanate ceramic) with the property that an applied elec- phased-array transducer) are used to modify beam geometry. The material
tric current results in alignment of polarized particles of the transducer surface provides impedance matching with the skin. The
ultrasound pulse length for 2D imaging is short (1-6ms), typically consist-
perpendicular to the face of the crystal with consequent ing of two wavelengths (λ). “Ring down”—the decrease in frequency and
expansion of crystal size. When an alternating electric amplitude in the pulse—depends on damping and determines bandwidth
current is applied, the crystal alternately compresses (the range of frequencies in the signal).
7
Principles of Echocardiographic Image Acquisition and Doppler Analysis | Chapter 1
key element in the patient’s total ultrasound exposure which allows better resolution of structures distant
as discussed in Bioeffects and Safety, p. 27. from the transducer. The stated frequency of a trans-
The range of frequencies contained in the pulse is ducer represents the center frequency of the pulse.
described as its frequency bandwidth. A wider bandwidth
allows better axial resolution because of the ability
of the system to produce a narrow pulse. Transducer Types of Transducers
bandwidth also affects the range of frequencies that The simplest type of ultrasound transducer is based
can be detected by the system with a wider bandwidth, on a single piezoelectric crystal (Table 1-3). Alternate
Type Transducer characteristics Transthoracic (adult and pediatric) Each transducer type is
and configuration Nonimaging CW Doppler optimized for a specific
Most cardiac transducers 3D echocardiography clinical application.
use a phased array of TEE More than one transducer may be
piezoelectric crystals. Intracardiac needed for a full examination.
Transmission The central frequency Transducer frequencies A higher-frequency transducer
frequency emitted by the vary from 2.5 MHz for provides improved resolution
transducer transthoracic echo to 20 MHz but less penetration.
for intravascular imaging. Doppler signals are optimal at a
lower transducer frequency
than used for imaging.
Power output The amount of ultrasound An increase in transmitted Excessive power output
energy emitted by the power increases the may result in bioeffects
transducer amplitude of the reflected measured by the mechanical
ultrasound signals. and thermal indexes.
Bandwidth The range of frequencies Bandwidth is determined by A wider bandwidth allows
in the ultrasound pulse transducer design. improved axial resolution for
structures distant from the
transducer.
Pulse (or burst) The length of the A higher-frequency signal can A shorter pulse length improves
length transmitted ultrasound be transmitted in a shorter axial resolution.
signal pulse length compared to a
lower-frequency signal.
Pulse The number of The PRF decreases as imaging PRF affects image resolution and
repetition transmission-receive (or Doppler) depth increases frame rate (particularly with
frequency cycles per second because of the time needed color Doppler).
(PRF) for the signal to travel from
and to the transducer.
Duty factor The percentage of time Ranges from about 1% for A higher duty factor means more
that ultrasound is imaging to 5% for pulsed tissue exposure to ultrasound.
transmitted Doppler to 100% for CW
Doppler
Focal depth Beam shape and Structures close to the The length and site of a
focusing are used to transducer are best visualized transducer’s focal zone
optimize ultrasound with a short focal depth, is primarily determined
resolution at a specific distant structures with a long by transducer design, but
distance from the focal depth. adjustment during the exam
transducer. may be possible.
Aperture The surface of the A small nonimaging CW A larger aperture allows a more
transducer face Doppler transducer allows focused beam.
where ultrasound optimal positioning and A smaller aperture allows
is transmitted and angulation of the ultrasound improved transducer
received beam. angulation on TTE imaging.
8
Chapter 1 | Principles of Echocardiographic Image Acquisition and Doppler Analysis
pulsed transmission and reception periods allow transducer depends on transducer frequency (higher-
repeated sampling along a single line, with the sam- frequency transducers are smaller) and beam focusing,
pling rate limited only by the time delay needed for as well as the intended clinical use, for example, trans-
return of the reflected ultrasound wave from the depth thoracic versus TEE imaging.
of interest. An example of using the transducer for
simple transmission-reception along a single line is an
A-mode (amplitude versus depth) or M-mode (depth Beam Shape and Focusing
versus time) cardiac recording when a high sampling An unfocused ultrasound beam is shaped like the light
rate is desirable. from a flashlight, with a tubular beam for a short dis-
Formation of more complex images uses an array tance that then diverges into a broad cone of light (Fig.
of ultrasound crystals arranged to provide a 2D tomo- 1-6). Even with current focused transducers, ultrasound
graphic or 3D volumetric data set of signals. Each beams have a 3D shape that affects measurement accu-
element in the transducer array can be controlled racy and contributes to imaging artifacts. Beam shape
electronically both to direct the ultrasound beam and size depend on several factors, including:
across the region of interest and to focus the transmit-
n ransducer frequency
T
ted and received signals. Echocardiographic imaging
n Distance from the transducer
uses a sector scanning format with the ultrasound signal
n Aperture size and shape
originating from a single location (the narrow end of
n Beam focusing
the sector), resulting in a fanlike shape of the image.
Sector scanning is optimal for cardiac applications Aperture size and shape and beam focusing can be
because it allows a fast frame rate to show cardiac manipulated in the design of the transducer, but the
motion and a small transducer size (aperture or “foot- effects of frequency and depth are inherent to ultra-
print”) to fit into the narrow acoustic windows used for sound physics. For an unfocused beam, the initial seg-
echocardiography. Three-dimensional imaging trans- ment of the beam is columnar in shape (near field Fn)
ducers are discussed in Chapter 4. with a length dependent on the diameter D of the
Most transducers can provide simultaneous imag- transducer face and wavelength (λ):
ing and Doppler analysis, for example, 2D-imaging /
Fn = D2 4λ (1-6)
and a superimposed color Doppler display. Quantita-
tive Doppler velocity data are recorded with the image For a 3.5-MHz transducer with a 5-mm diameter
“frozen” or with only intermittent image updates, with aperture, this corresponds to a columnar length of 1.4
the ultrasound crystals used to optimize the Doppler cm. Beyond this region, the ultrasound beam diverges
signal. Although CW Doppler signals can be obtained (far field), with the angle of divergence θ determined as:
using two elements of combined transducer, use of a /
sin θ = 1.22λ D (1-7)
dedicated nonimaging transducer with two separate
crystals (with one crystal continuously transmitting This equation indicates a divergence angle of 6°
and the other continuously receiving the ultrasound beyond the near field, resulting in an ultrasound beam
waves) is recommended when accurate high-velocity width of about 4.4 cm at a depth of 20 cm for this 3.5-
recordings are needed. The final configuration of a MHz transducer. With a 10-mm diameter aperture, Fn
Divergence
angle
9
Principles of Echocardiographic Image Acquisition and Doppler Analysis | Chapter 1
10
10
5 mm aperture
0 0
0 5 10
Transducer frequency (MHz)
Figure 1–8 Transducer beam side lobes. Top: This diagram shows that
side lobes occur at the points where the distances traversed by the ultrasound
Side pulse from each edge of the crystal face differ by exactly one wavelength. The
lobe 1 Side
distance from the left edge of the crystal (P1) to the position of side lobe 1 is
lobe 2
exactly one wavelength (λ) longer than the distance from the extreme right
edge of the crystal (P2) to the position of side lobe 1. Bottom: The beam inten-
sity plot formed by sweeping along an arc at focal length F. (From Geiser EA:
θ2 θ1 0 θ1 θ2 Echocardiography: physics and instrumentation. In Skorton DJ, Schelbert AR,
Wolf GL, Brundage BH [eds]: Marcus Cardiac Imaging, 2nd ed. Philadelphia:
Angular position WB Saunders, 1996, p 280. Used with permission.)
10
Chapter 1 | Principles of Echocardiographic Image Acquisition and Doppler Analysis
Acoustic
lens
Lateral
Slice thickness
(elevational)
Axial
Transducer frequency
Resolution
Transducer bandwidth
Image resolution occurs for each of three dimensions Pulse length
(Fig. 1-9):
Lateral Resolution
n xial resolution along the length of the ultrasound
A
beam Transducer frequency
n Lateral resolution side to side across the 2D image Beam width (focusing) at each depth*
n Elevational resolution or thickness of the tomo- Aperture (width) of transducer
graphic slice
Bandwidth
Of these three, axial resolution is most precise, so
Side and grating lobe levels
quantitative measurements are made most reliably using
data derived from a perpendicular alignment between Elevational Resolution
the ultrasound beam and structure of interest. Axial Transducer frequency
resolution depends on the transducer frequency, band-
width, and pulse length but is independent of depth Beam width in elevational plane
(Table 1-4). Determination of the smallest resolvable *Most important.
distance between two specular reflectors with ultrasound
is complex but is typically about twice the transmitted
wavelength; higher-frequency (shorter-wavelength) beam width at each depth. In the focal region where
transducers have greater axial resolution. For example, beam width is narrow, lateral resolution may approach
with a 3.5 MHz transducer, axial resolution is about axial resolution, and a point target will appear as a
1 mm, versus 0.5 mm with a 7.5 MHz transducer. A point on the 2D image. At greater depths, beam width
wider bandwidth also improves resolution by allow- diverges so a point target results in a reflected signal
ing a shorter pulse, thus avoiding overlap between the as wide as the width of the beam, which accounts for
reflected ultrasound signals from two adjacent reflectors. “blurring” of images in the far field. If the 2D image
Lateral resolution varies with the depth of the spec- is examined carefully, progressive widening of the
ular reflector from the transducer, primarily related to echo signals from similar targets along the length of
11
Principles of Echocardiographic Image Acquisition and Doppler Analysis | Chapter 1
M-Mode
Historically, cardiac ultrasound began with a single-
crystal transducer display of the amplitude (A) of
reflected ultrasound versus depth on an oscilloscope
screen. This A-mode display may still be shown on
the 2D image screen to aid the examiner in optimal
adjustment of the instrument controls. Repeated pulse
transmission-and-receive cycles allow rapid updating
of the amplitude-versus-depth information so that
rapidly moving structures, such as the aortic or mitral
valve leaflets, can be identified by their characteristic
timing and pattern of motion (Fig. 1-11).
With the time dimension shown explicitly on the
horizontal axis and each amplitude signal along the
length of the ultrasound beam converted to a corre-
sponding gray-scale level, a motion (M) mode display is
produced. M-mode data are shown on the video moni-
tor either “scrolling” or “sweeping” across the screen
at 50 to 100 mm/s. Two-dimensional (2D) imaging
Figure 1–10 Beam width effect on 2D imaging. 2D echocardio- allows guidance of the M-mode beam to ensure an
graphic view of the LV from an apical approach. The effect of beam width appropriate angle between the M line and the struc-
can be appreciated by comparing the length of reflections from point targets tures of interest.
near and at greater distances from the transducer as shown by the arrows. Because only a single “line of sight” is included in
an M-mode tracing, the pulse repetition frequency
(PRF) of the transmission-and-receive cycle is lim-
ited only by the time needed for the ultrasound beam
the ultrasound beam can be appreciated (Fig. 1-10). to travel to the maximum depth of interest and back
Erroneous interpretations occur when the effects of to the transducer. Even a depth of 20 cm requires
beam width are not recognized. For example, beam only 0.26 ms (given a speed of propagation of 1540
width artifact from a strong specular reflector may m/s), allowing a PRF up to 3850 times per second.
appear to be an abnormal linear structure. Other In actual practice, sampling rates of about 1800
factors that affect lateral resolution are transducer times per second are used. This extremely high sam-
frequency, aperture, bandwidth, and side and grating pling rate is valuable for accurate evaluation of rapid
lobe levels. normal intracardiac motion such as valve opening
Resolution in the elevational plane is more difficult and closing. In addition, continuously moving struc-
to recognize on the 2D image but is equally impor- tures, such as the ventricular endocardium, may be
tant in clinical diagnosis. The thickness of the tomo- identified more accurately when motion versus time,
graphic plane varies across the 2D image, depending as well as depth, is displayed clearly on the M-mode
on transducer design and focusing, both of which recording. Other examples of rapid intracardiac
affect beam width in the elevational plane at each motion best demonstrated with M-mode imaging
depth. In general, cardiac ultrasound images have a include the high-frequency fluttering of the anterior
“thickness” of approximately 3 to 10 mm depend- mitral leaflet in patients with aortic regurgitation and
ing on depth and the specific transducer used. The the rapid oscillating motion of valvular vegetations.
tomographic image generated by the instrument, in
effect, includes reflected and backscattered signals
from this entire thickness. Strong reflectors adjacent Two-Dimensional Echocardiography
to the image plane may appear to be “in” the image
Image Production
plane because of elevational beam width. Even more
distant strong reflectors may appear superimposed on A 2D echocardiographic image is generated from the
the tomographic plane because of side lobes in the data obtained by electronically “sweeping” the ultra-
elevational plane. For example, a linear echo in the sound beam across the tomographic plane. For each
aortic lumen from an adjacent calcified atheroma may scan line, short pulses (or bursts of ultrasound) are
look like a dissection flap. These principles of ultra- emitted at a PRF determined by the time needed for
sound imaging also apply to 3D echocardiography ultrasound to travel to and from the maximum image
(see Chapter 4). depth. The pulse repetition period is the total time
12
Chapter 1 | Principles of Echocardiographic Image Acquisition and Doppler Analysis
Ao Ao
Depth
Ao
LA LA
Distance
Time
Figure 1–11 3D, 2D, M-mode, and A-mode recordings of aortic valve motion. This illustration shows the following: the relationship between the 3D and
2D long-axis image of the aortic valve (left), which shows distance in both the vertical and horizontal direction; M-mode recording of aortic root (Ao), left atrium
(LA), and aortic valve motion, which shows depth-versus-time (middle); and A-mode recording (right), which shows depth only (with motion seen on the video
screen). Spatial relationships are best shown with 3D or 2D imaging, but temporal resolution is higher with M-mode and A-mode imaging.
from pulse to pulse, including the length of the ultra- time-gain compensation (TGC), filtering (to reduce
sound signal plus the time interval between signals. noise), compression, and rectification. Envelope
Because a finite time is needed for each scan line detection generates a bright spot for each signal
of data (depending on the depth of interest), the time along the scan line, which then undergoes analog-
needed to acquire all the data for one image frame is to-digital scan conversion, since the original polar
directly related to the number of scan lines and the coordinate data must be fit to a rectangular matrix
imaging depth. Thus, PRF is lower at greater imag- with appropriate interpolation for missing matrix
ing depths and higher at shallow depths. In addition, elements. This image is subject to further “postpro-
there is a tradeoff between scan line density and cessing” to enhance the visual appreciation of tomo-
image frame rate (the number of images per sec- graphic anatomy and is displayed in “real time”
ond). For cardiac applications, a high frame rate (≥30 (nearly simultaneous with data acquisition) on the
frames per second) is desirable for accurate display monitor screen.
of cardiac motion. This frame rate allows 33 ms per Although standard ultrasound imaging is based
frame or 128 scan lines per 2D image at a displayed on reflection of the fundamental transmitted fre-
depth of 20 cm. quency from tissue interfaces, tissue harmonic imaging
The reflected ultrasound signals for each scan line (THI) instead is based on the harmonic frequency
are received by the piezoelectric crystal and a small energy generated as the ultrasound signal propagates
electric signal generated with: through the tissues. These harmonic frequencies
result from the nonlinear effects of the interaction of
n mplitude proportional to incident angle and
A ultrasound with tissue and with the key properties:
acoustic impedance
n Timing proportional to distance from the n armonic signal strength increases with depth
H
transducer of propagation.
n Harmonic frequencies are maximal at typical
This signal undergoes complex manipulation cardiac imaging depths.
to form the final image displayed on the monitor. n Stronger fundamental frequencies produce
Typical processing includes signal amplification, stronger harmonics.
13
Principles of Echocardiographic Image Acquisition and Doppler Analysis | Chapter 1
border tracing for calculation of ejection fraction, Other typical instrument controls include prepro-
reduces measurement variability, and results in visu- cessing and postprocessing settings that change the
alization of more myocardial segments during stress appearance of the displayed image. Image quality and
echocardiography. However, although THI improves resolution also depend on scan-line density and other
lateral resolution by 20-50%, it reduces axial resolu- factors (see Table 1-4). Scan-line density (or frame rate
tion by 40 to 100%. Thus, valves and other planar or both) can be increased by using a lower depth set-
objects may appear thicker with harmonic, compared ting or by narrowing the sector to less than the stan-
to fundamental, frequency imaging, so that caution is dard 60° wide image.
needed when diagnosing valve abnormalities or mak-
ing measurements of chamber or vessel size.
Imaging Artifacts
Imaging artifacts include (1) extraneous ultrasound
Instrument Settings
signals that result in the appearance of “structures”
Many of the elements in the process of image for- that are not actually present (at least at that location),
mation are features of a particular transducer and (2) failure to visualize structures that are present, and
instrument that cannot be modified by the operator. (3) an image of a structure that differs in size or shape
However, for each patient and echocardiographic or both from its actual appearance. Obviously, rec-
view, optimal image quality depends on transducer ognition of image artifacts is important for both the
selection and instrument settings. Standard imaging individual performing the study and the individual
controls available in most ultrasound systems include: interpreting the echocardiographic data (Table 1-5).
The most common image “artifact” is suboptimal
n ower output: This control adjusts the total ultra-
P image quality resulting from poor ultrasound tissue pen-
sound energy delivered by the transducer in the etration related to the patient’s body habitus with inter-
transmitted bursts; higher power outputs result position of high attenuation tissues (e.g., lung or bone)
in higher-amplitude reflected signals (see Bioef- or an increased distance (e.g., adipose tissue) between
fects and Safety, p. 27.). the transducer and cardiac structures. While, strictly
n Gain: Adjusts the displayed amplitude of the speaking, poor image quality is not an “artifact,” a low
received signals, similar to the volume control in signal-to-noise ratio makes accurate diagnosis difficult
an audio system. and precludes quantitative measurements. In many
14
Chapter 1 | Principles of Echocardiographic Image Acquisition and Doppler Analysis
Suboptimal image Poor ultrasound tissue penetration Body habitus (obesity, lung disease)
quality Postcardiac surgery
Acoustic shadowing Reflection of entire ultrasound signal by a strong Prosthetic valve
specular reflector Calcification
Reverberations Reverberation between two strong parallel reflectors Prosthetic valve
Beam width Superimposition of structures within the beam Aortic valve “in” LA
profile (including side lobes) into a single Atheroma “in” aortic lumen
tomographic image
Lateral resolution Displayed width of a point target varies with depth Excessive width of calcified mass or
prosthetic valve
Refraction Deviation of ultrasound signal from a straight path Double aortic valve or LV image in
along the scan line short-axis view
Range ambiguity Echo from previous pulse reaches transducer on Second, deeper heart image
next cycle
Electronic Instrument specific Variable
processing
LA
Transducer
Ao
RVOT
A B
Parallel
Reverberations strong reflectors Figure 1–15 Example of beam width artifact. Apparent “mass” at-
tached to the aortic valve on this off-axis TEE view is the noncoronary cusp
of the aortic valve seen “en face.” Imaging in other planes demonstrated a
normal trileaflet aortic valve.
A Ultrasound
artifacts
B
of the aortic valve or LV, where a second valve or LV
Figure 1–14 Reverberation artifacts result from the interaction of ul- is “seen” medial to and partly overlapping the actual
trasound with two parallel strong reflectors. The transmitted ultrasound
beam (red with down arrow) is reflected from the first reflector and returns valve or LV. The explanation for this appearance
to the transducer (red with up arrow) resulting in an ultrasound signal that is that the transmitted ultrasound beam is deviated
corresponds to the correct depth of the reflector. However, ultrasound from a straight path (the scan line) by refraction as it
signals also reflect back and forth between the two strong reflectors, with passes through a tissue near the transducer. When this
some signals returning to the transducer after two (A), three (B), or more
reverberation cycles. The longer time from transmission to reception of
refracted beam is reflected back to the transducer by a
these late-returning signals results in their display on the ultrasound im- tissue interface, the reflected signal is assumed to have
age at points distal to the actual reflector. In clinical imaging, reverberation originated from the scan line of the transmitted pulse
artifacts can either appear as a single linear signal distal to the actual object (Fig. 1-16) and thus is displayed on the image in the
or as a band of signals obscuring distal structures (see Fig. 1-13) because wrong location.
of multiple parallel reflectors.
Range ambiguity occurs when echo signals from an
earlier pulse cycle reach the transducer on the next
tomographic section corresponding to the main beam “listen cycle” for that scan line, resulting in deep struc-
(Fig. 1-15). tures appearing closer to the transducer than their
The second type of beam width artifact is a con- actual location. The appearance of an anatomically
sequence of varying lateral resolution at different unexpected echo within a cardiac chamber often is
imaging depths. A point target appears as a line whose due to range ambiguity, as can be demonstrated by the
length depends on the beam characteristics at that disappearance or a change in position of this artifact
depth and the amplitude of the reflected signal. For when the depth setting (and PRF) is changed. Another
example, the struts on a prosthetic valve can appear type of range ambiguity is the appearance of an
much longer than their actual dimension because of apparent second heart, deeper than the actual heart—
poor lateral resolution. Sometimes beam width arti- a double image on the vertical axis. This type of range
facts can be mistaken for abnormal structures such as a ambiguity results from echoes being re-reflected by a
valvular vegetation, an intracardiac mass, or an aortic structure close to the transducer (such as a rib), being
dissection flap. re-reflected by the cardiac structures and thus received
The appearance of a side-by-side double image at the transducer at a time twice normal. This artifact
results from ultrasound refraction as it passes through can be eliminated (or obscured) by decreasing the
a tissue proximal to the structure of interest. This depth setting or adjusting the transducer position to a
artifact often is seen in parasternal short-axis views better acoustic window.
16
Chapter 1 | Principles of Echocardiographic Image Acquisition and Doppler Analysis
Doppler effect The change in frequency of A higher velocity corresponds Ultrasound systems display
ultrasound scattered from to a higher Doppler velocity, which is calculated
a moving target frequency shift, ranging using the Doppler equation,
v = c × ∆F / [2 FT (cos θ)] from 1 to 20 kHz for based on transducer
intracardiac flow velocities. frequency and the Doppler
shift, assuming cos θ
equals 1.
Intercept angle The angle (θ) between the When the ultrasound beam is Velocity is underestimated
direction of blood flow parallel to the direction of when the intercept angle is
and the ultrasound beam blood flow (0° or 180°), cos not parallel. This can lead
θ is 1 and can be ignored in to errors in hemodynamic
the Doppler equation. measurements.
CW Doppler Continuous ultrasound CW Doppler allows CW Doppler is used to
transmission with measurements of high- measure high velocities
reception of Doppler velocity signals but does in valve stenosis and
signals from the entire not localize the depth of regurgitation.
length of the ultrasound origin of the signal.
beam
Pulsed Doppler Pulsed ultrasound Pulsed Doppler samples Pulsed Doppler is used to
transmission with velocities from a specific record low-velocity signals
timing of reception site but can only measure at a specific site, such as
determining depth of the velocity over a limited LV outflow velocity or LV
backscattered signal range. inflow velocity.
PRF The number of pulses The PRF is limited by the time The maximum velocity
transmitted per second needed for ultrasound to measurable with pulsed
reach and return from the Doppler is about 1 m/s at 6
depth of interest. cm depth.
PRF determines the maximum
velocity that can be
unambiguously measured.
Nyquist limit The maximum frequency The Nyquist limit is displayed The greater the depth,
shift (or velocity) as the top and bottom of the lower the maximum
measurable with pulsed the velocity range with the velocity measurable with
Doppler equal to ½ PRF baseline centered. pulsed Doppler
Signal aliasing The phenomenon that With aliasing of the LV Aliasing can result in
the direction of flow for outflow signal, the peak inaccurate velocity
frequency shifts greater of the velocity curve is measurements if not
than the Nyquist limit “cut off” and appears recognized.
cannot be determined as flow in the opposite
direction.
Sample volume The intracardiac location Sample volume depth Sample volume depth and
where the pulsed Doppler is determined by the length are adjusted to
signal originated time interval between record the flow of interest.
transmission and
reception.
Sample volume length is
determined by the duration
of the receive cycle.
Spectral analysis Method used to display Spectral analysis is used The velocity scale, baseline
Doppler velocity data for both pulsed and CW position, and time scale
versus time, with gray Doppler. of the spectral display are
scale indicating amplitude adjusted for each Doppler
velocity signal.
18
Chapter 1 | Principles of Echocardiographic Image Acquisition and Doppler Analysis
Doppler Equation θ
F1 F2 F1 F2
FT
FS
F1 = F2 F1 < F2
Figure 1–17 The Doppler effect. A stationary scatterer (left) scatters ultra- T
sound symmetrically in all directions with a wavelength identical to the trans-
mitted wavelength and with the same frequency in all directions (no Doppler
shift). A moving scatterer (right) also scatters ultrasound symmetrically in Figure 1–18 The Doppler equation. The velocity V of blood flow can be
all directions. However, the frequency will be higher when the scatterer is calculated from the speed of sound in blood c, transducer frequency FT,
moving toward the transducer (F2) than when it is moving away from the backscattered frequency FS, and the cosine of the angle θ between the
transducer (F1) because of the movement of the scatterer resulting in waves ultrasound beam and direction of blood flow.
closer together in advance of and farther apart behind the moving object.
APEX AV
Figure 1–19 Examples of pulsed (left) and CW (right) spectral Doppler displays. LV outflow recorded from an apical approach is shown in the standard
format. The baseline has been moved from the middle of the vertical axis to display the antegrade flow signal. Velocities toward the transducer are shown
above and velocities away from the transducer below the baseline. The velocity range is determined by the Nyquist limit (½ PRF) with pulsed Doppler echo.
Velocities are shown in shades of gray corresponding to the amplitude (dB) of the signal. Note the “envelope” of flow with pulsed Doppler because flow is
sampled at a specific intracardiac location with relatively uniform blood flow velocities. With CW Doppler, the curve is “filled in” due to multiple blood flow
velocities along the entire length of the ultrasound beam.
velocity (or frequency shift) component. Thus, each but, signal quality may be poorer, angulation is more
time point on the spectral display shows: difficult, and the 2D image may distract the operator
from optimizing the flow signal instead of the anatomic
n lood flow direction
B
image (which may not coincide).
n Velocity (or frequency shift)
Careful technique yields a Doppler spectral signal
n Signal amplitude
that has a smooth contour with a well-defined edge
Each of these components is displayed at 4-ms and maximum velocity, as well as with clearly defined
intervals (or 250 times per second) simultaneous with onset and end of flow. The audible signal is tonal
data acquisition. and smooth. A CW Doppler velocity curve is “filled
in” because lower-velocity signals proximal and distal
to the point of maximum velocity also are recorded.
Continuous-Wave Doppler Ultrasound
Note that while the maximum frequency shift depends
CW Doppler uses two ultrasound crystals; one con- on the intercept angle between the Doppler beam and
tinuously transmits and one continuously receives the the flow of interest, amplitude (gray-scale intensity),
ultrasound signal. The major advantage of CW Dop- shape, and audible quality are less dependent on inter-
pler is that very high-frequency shifts (velocities) can be cept angle. Thus a “good quality” Doppler signal may
measured accurately because sampling is continuous. be recorded at a nonparallel intercept angle, resulting
The potential disadvantage of CW Doppler is that sig- in underestimation of flow velocity. The empirical
nals from the entire length of the ultrasound beam are method to ensure a parallel intercept angle is to exam-
recorded simultaneously. However, even with overlap ine the flow of interest from multiple windows with
of flow data, a given signal often is characteristic in transducer angulation both in the plane of view and in
timing, shape, and direction, allowing correct identifi- the elevational plane to discover the highest-frequency
cation of the origin of the signal. In some cases, other shift. The highest value found is then assumed to rep-
methods (e.g., 2D echo, color, pulsed Doppler) must be resent a parallel intercept angle.
used to determine the depth of origin of the Doppler
signal.
CW Doppler optimally is performed with a dedi- Pulsed Doppler Ultrasound
cated, nonimaging transducer with two crystals. This Pulsed Doppler echocardiography allows sampling
type of transducer has a high signal-to-noise ratio and of blood flow velocities from a specific intracardiac
a small footprint, allowing it to fit into small acoustic depth. A pulse of ultrasound is transmitted, and then,
windows (e.g., between ribs) and to be angled to obtain after a time interval determined by the depth of inter-
a parallel intercept angle between the ultrasound beam est, the transducer briefly “samples” the backscattered
and the direction of blood flow. Use of a simultane- signals. This transducer cycle of transmit-wait-receive
ous imaging transducer may be helpful in some cases is repeated at an interval termed the pulse repetition
20
Chapter 1 | Principles of Echocardiographic Image Acquisition and Doppler Analysis
...
Cycle length
9/
9
Time
r
PRF = cycles/s
i
Figure 1–20 Pulsed Doppler ultrasound. The pulsed Doppler transducer
h
goes through a repetitive cycle of transmission of an ultrasound pulse at the
transducer frequency (FT), a waiting period determined by the time needed
a
for the signal to travel to and from the depth of interest, and a receive phase
when the backscattered signals are sampled. The travel-time duration de-
t
termines sample volume depth. The duration of the receive phase deter-
r/
mines sample volume.
e
frequency (PRF) (Fig. 1-20). Because the “wait” inter-
s
val is determined by the depth of interest—the time
it takes ultrasound to travel to and from this depth—
/r u
each transducer cycle is longer for increasing depths.
Thus, the PRF also is depth dependent, being high at
shallow depths and low for more distant sites.
The pulsed Doppler depth of interest is called the Figure 1–21 Principle of signal aliasing. This schematic diagram shows
.t c
sample volume because signals from a small volume of how sampling at a constant interval (shown by vertical gray lines with a red
blood are sampled, with the width and height of this dot where the waveform is sampled) results in ambiguity in the measured
sound wave frequency. Sampling at twice the frequency of the wavelength,
volume dependent on beam geometry. The length of as shown at the top, correctly measures the sound wave frequency. As the
a
the sample volume can be varied by adjusting the length sound wave frequency increases from top to bottom, intermittent sampling
of the transducer “receive” interval. Typically, a sample results in apparent frequencies that are lower and in the opposite direction
k
volume length of 3 mm is used to balance range resolu- of the actual sound waveform.
/: /
tion and signal quality, but a longer (5-10 mm) or shorter
(1-2 mm) sample volume may be useful in specific cases.
Because pulsed Doppler echo repeatedly samples the reverse channel, then back to the forward channel,
s
the returning signal, there is a maximum limit to the and so on. Occasionally, the shape of the waveform
frequency shift (or velocity) that can be measured can be discerned, but more often only an undiffer-
tt p
unambiguously. A waveform must be sampled at least entiated band of velocity signals can be appreciated.
twice in each cycle for accurate determination of wave- Both nonlaminar disturbed flow and aliased laminar
length. This phenomenon of ambiguity in the speed, high-velocity flow will appear (and sound) similar on
direction, or speed and direction of the sampled signal
h
spectral analysis. Methods that can be used to resolve
is known as signal aliasing (Fig. 1-21). In order for the aliasing include:
frequency of an ultrasound waveform to be correctly
n Using CW Doppler ultrasound
identified, it must be sampled at least twice per wave-
n Increasing the PRF to the maximum for that
length. Thus, the maximum detectable frequency shift
depth (the Nyquist limit)
(the Nyquist limit) is one half the PRF.
n Increasing the number of sample volumes (high-
If the velocity of interest exceeds the Nyquist limit
PRF Doppler)
by a small degree, signal aliasing is seen with the sig-
n Using a lower frequency transducer
nal cut off at the edge of the display and the “top”
n Shifting the baseline to the edge of the display
of the waveform appearing in the reverse channel
(Fig. 1-22). In these cases, baseline shift (in effect, an CW Doppler is the most reliable approach to
electronic “cut and paste”) restores the expected veloc- resolving aliasing for very high velocities. The other
ity curve and allows calculation of maximum veloc- approaches are useful when the aliased velocity
ity. When velocities further exceed the Nyquist limit, exceeds the Nyquist limit by a modest degree (e.g., ≤
repeat “wraparound” of the signal occurs first into twice the Nyquist limit).
21
Principles of Echocardiographic Image Acquisition and Doppler Analysis | Chapter 1
9/
i r 9
a h
Figure 1–22 Pulsed Doppler signal aliasing. The velocity of LV outflow recorded from an apical approach exceeds the Nyquist limit so that aliasing occurs
t
(left) with the appearance of the peak of the outflow curve in the reverse channel (arrow). This degree of aliasing can be resolved by shifting the baseline
r/
(right), in effect an electronic “cut and paste” of the spectral display.
e
a second sample volume (SV2 ) twice as far away (i.e., the
s
SV2 depth of interest) will return to the transducer during
the “receive” phase (albeit one cycle later). This record-
/r u
ing of the signal of interest at a higher PRF allows mea-
surement of higher velocities without signal aliasing (Fig.
SV1 1-24). An even higher PRF can be achieved by using
additional (three or four) proximal sample volumes. Of
.t c
course, the limitation of this approach is range ambigu-
ity. The spectral analysis now includes signals from each
of the sample volume depths and, as with CW Doppler,
a
T the origin of the signal of interest must be determined
based on ancillary data. However, high-PRF Doppler is
k
useful for evaluation of velocities just above the aliasing
/: /
Figure 1–23 High-pulse-repetition frequency (PRF) Doppler ultra- limit of conventional pulsed Doppler. Often, the high
sound. High-PRF Doppler is based on the concept that with a given sam- PRF mode is automatically enabled when the Doppler
ple volume depth (SV1), some ultrasound will penetrate beyond that depth. velocity range is increased.
s
Backscattered signals from exactly twice the set depth (SV2 ) will return to
the transducer (T) during the receive phase of the next cycle. Thus, signals
from both sample volume depths are recorded simultaneously.
tt p
Doppler Velocity Instrument Controls
Pulsed and CW Doppler instrument controls typically
High-PRF Doppler is the deliberate use of range
include:
h
ambiguity to increase the maximum velocity that can be
measured with pulsed Doppler echo (Fig. 1-23). When n Power output—adjusts the amount of electrical
the transducer sends out a pulse, backscattered signals energy transmitted to the transducer
from the entire length of the ultrasound beam return to n Receiver gain—changes the degree of amplifica-
the transducer. Range resolution is achieved by sampling tion of returning signals
only those signals in the short time interval correspond- n “Wall” or high-pass filters—eliminate low-
ing to the depth of interest. However, signals from exactly frequency Doppler shifts that result from motion
twice as far away as the sample volume will reach the of myocardium and valves (allowing only the
transducer during the “receive” phase of the next cycle. higher frequencies to pass the filter)
Thus, signals from “harmonics” at 2×, 3×, 4×, and so n Baseline shift—moves the zero line toward the
on from the sample volume depth have the potential of top or bottom of the display
being analyzed. Usually signal strength is low and there n Velocity range—increases or decreases the scale
are few moving scatterers at these depths, so this range (within the limits for each Doppler modality)
ambiguity can be ignored. If, instead, the sample volume n Dynamic range—compresses the signal ampli-
is placed purposely at one-half the depth of interest, tude into shades of gray
22
Chapter 1 | Principles of Echocardiographic Image Acquisition and Doppler Analysis
3.2 m/s
9/
i r 9
a h
Figure 1–24 Example of HPRF Doppler. In this case, LV outflow velocity exceeds twice the Nyquist limit so that aliasing persists even after baseline shift
t
(left). The “wrapped around” peak velocity is clearly seen (arrow). With high-PRF Doppler, the maximum velocity is resolved in this patient with a subaortic
r/
membrane (right).
e
TABLE 1-7 Doppler Ultrasound Artifacts
n Sample volume depth
s
Artifact Result
n Sample volume length
The number of sample volumes (high pulse
/r u
n Nonparallel Underestimation of velocity
repetition frequency Doppler echo) intercept angle
Each of the three major Doppler modalities may Aliasing Inability to measure maximum
be integrated with 2D imaging. However, while color velocity
.t c
Doppler flow imaging is nearly always conjoined with Range ambiguity Doppler signals from more
2D imaging, pulsed Doppler signal quality is opti- than one depth along the
mized when the 2D image is “frozen,” and CW Dop- ultrasound beam are recorded
a
pler is optimized using a dedicated, small-footprint Beam width Overlap of Doppler signals from
transducer with no 2D imaging. adjacent flows
/: / k
Mirror image Spectral display shows
Doppler Velocity Data Artifacts unidirectional flow both above
and below the baseline.
Many Doppler artifacts are related to ultrasound phys-
Electronic Bandlike interference signal
s
ics and beam geometry, analogous to those seen with interference obscures Doppler flow.
2D imaging. Others are specific to Doppler echocar-
tt p
diography (Table 1-7). Transit-time effect Change in the velocity of the
Clinically, the most important potential artifact ultrasound wave as it passes
through a moving medium
is velocity underestimation resulting from a nonparallel results in slight overestimation
intercept angle between the ultrasound beam and the
h
of Doppler shifts.
direction of blood flow (Fig. 1-25). Velocity underesti-
mation can occur with either pulsed or CW Doppler
techniques and is of most concern when measuring
high-velocity jets due to valve stenosis, regurgitation,
or other intracardiac abnormalities.
With pulsed Doppler echo, signal aliasing limits the signals from twice (or three times) the depth of the
maximum measurable velocity. If the examiner recog- sample volume will be received in the next “receive”
nizes that aliasing has occurred, appropriate steps can phase and may be misinterpreted as originating
be taken to resolve the velocity data if needed. Aliasing from the set sample volume depth. For example, in
can be due to nonlaminar disturbed flow, as well as an apical four-chamber view, placement of a sample
high-velocity laminar flow. volume in the LV apex at half the distance to the
Range ambiguity is inherent to CW Doppler but can mitral annulus results in a spectral display showing
occur with pulsed Doppler as well. With a sample the inflow signal across the mitral valve from the
volume positioned close to the transducer, strong “second” sample volume depth. This phenomenon
23
Principles of Echocardiographic Image Acquisition and Doppler Analysis | Chapter 1
Cosine
only 6% at a 20° angle to 50% at a 60° angle. At
0 a perpendicular (90°) intercept angle, no blood
/
flow velocities are recorded.
40
9
– 0.5
9
20
i r
20 degrees
6% error
0 –1
h
0 20 40 60 80 100 120 140 160 180
a
Intercept angle (degrees)
r/ t
AR
s e
/r u
LV inflow
.t c
k a
/: /
Figure 1–27 Mirror-image Doppler artifact. A mirror-image Doppler arti-
Figure 1–26 Doppler beam width artifact. This CW Doppler recording fact with apparent weaker flow signals in the reverse channel (arrow).
s
from an apical window shows superimposed signals of aortic regurgitation
(AR) and LV inflow in diastole because the width of the ultrasound beam
tt p
encompasses both flows. Bright lines caused by motion of prosthetic mitral or gain of the instrument. Interrogation of a flow sig-
valve disks also are seen (short arrow).
nal from a near-perpendicular angle also can result in
flow signals on both sides of the baseline.
of range ambiguity is used constructively in the
h
Electronic interference appears as a band of signals
high-PRF Doppler mode. across the spectral display that may obscure the flow
Beam width (and side or grating lobes) affects the signals. These artifacts are the result of inadequate
Doppler signal, as occurs with 2D imaging, resulting shielding of other electric instruments in the examina-
in superimposition of spatially adjacent flow signals tion environment and are particularly common during
on the spectral display. For example, LV outflow and studies in the intensive care unit, interventional proce-
inflow may be seen on the same recording, especially dure areas, or operating room.
with CW Doppler. Similarly, the LV inflow signal The transit time effect is the change in propagation
may be seen superimposed on the aortic regurgitant speed that occurs as an ultrasound wave passes through
jet (Fig. l-26). a moving medium, such as blood. This phenomenon
A mirror-image artifact is common with spectral anal- is separate from the Doppler effect (which affects the
ysis, appearing as a symmetric signal of somewhat backscattered signal) and is the basis of volume flow
less intensity than the actual flow signal in the oppo- measurement with a transit-time flow probe. On the
site flow direction (Fig. 1-27). Mirroring often can be spectral display, the transit-time effect may result in a
reduced or eliminated by decreasing the power output slight broadening of the velocity range at a given time
24
Chapter 1 | Principles of Echocardiographic Image Acquisition and Doppler Analysis
Pulsed Doppler Color Flow point (“blurring” on the vertical axis), which poten-
Ultrasound Imaging tially can result in slight overestimation of velocity.
/
scan line
of pulsed Doppler echocardiography. However, rather
than one sample volume depth along the ultrasound
9
beam, multiple sample volumes are evaluated along
each sampling line (Fig. 1-28). By combining data
9
from adjacent lines, a 2D image of intracardiac flow
i r
is generated.
Sample Along each scan line, a pulse of ultrasound is trans-
h
volume mitted, and then the backscattered signals are received
from each depth along that scan line (Table 1-8). In
a
order to calculate accurate velocity data, several bursts
t
Figure 1–28 Color Doppler flow imaging. With pulsed Doppler, the along each scan line are used—typically eight—which
r/
sample volume depth is determined by the time needed for ultrasound to is known as the burst length (Fig. 1-29). The PRF, as for
travel to and from the depth of interest (left). With color flow imaging, mul- conventional pulsed Doppler, is determined by the
tiple sample volume “gates” along each scan line are interrogated, with this
maximum depth of the Doppler signals. Signals from
e
process repeated for scan lines across the 2D image (right).
the eight sampling bursts at each position are analyzed
s
/r u
TABLE 1-8 Color Doppler Flow Imaging
Definition Examples Clinical Implications
.t c
Sampling line Doppler data is displayed Instead of sampling A greater number of sampling
from multiple sampling backscattered signals from lines results in denser Doppler
lines across the 2D one depth (as in pulsed data but a slower frame rate.
image. Doppler), signals from
a
multiple depths along the
beam are analyzed.
k
Burst length The number of Mean velocity is estimated A greater number of bursts
/: /
ultrasound bursts from the average of the results in more accurate mean
along each sampling backscattered signals from velocity estimates but a slower
line each burst. frame rate.
s
Sector scan The width of the A greater sector width requires A narrower sector scan allows a
width displayed 2D and color more sampling lines or less greater sampling line density
tt p
image dense velocity data. and faster frame rate.
Sector scan The depth of the The maximum depth of the The minimum depth needed to
depth displayed color sector scan determines PRF display the flow of interest
h
Doppler image (as with pulsed Doppler) and provides the optimal color
the Nyquist limit. display.
Color scale Color display of Doppler Most systems use shades The color scale can be adjusted
velocity and flow of red for flow toward the by shifting the baseline and
direction transducer and blue for flow adjusting the maximum
away from the transducer. velocity displayed (within the
Nyquist limit).
Variance The degree of variability Variance typically is A variance display highlights flow
in the mean velocity displayed as a green scale disturbances and high-velocity
estimate at each depth superimposed on the red- flow, but even normal flows
along a sampling line blue velocity scale. Variance will be displayed as showing
can be turned on or off. variance if velocity exceeds
the Nyquist limit.
25
Principles of Echocardiographic Image Acquisition and Doppler Analysis | Chapter 1
/
to exclude low-velocity signals from the color flow dis-
play. In addition, many instruments allow variation in
9
the assignment of a returning signal to 2D or Doppler
display (depending on signal strength). One approach
9
to optimizing the color flow display is to reduce the 2D
i r
gain because the instrument does not display flow data
Figure 1–29 Color Doppler flow imaging burst length. Along each color on top of “structures,” even when the 2D signal is due
h
Doppler scan line, several (typically eight) bursts of ultrasound are transmit- to excessive gain.
ted and received to allow adequate velocity resolution. Perhaps the most important technical factor in
a
color flow imaging is optimization of frame rate. Color
t
flow frame rate depends on sector width, depth, pulse
r/
to obtain mean velocity estimates for each depth along repetition frequency, and the number of samples per
the scan line. Velocities are displayed using a color sector line. The examiner optimizes frame rate by
scale showing flow toward the transducer in red and focusing on the flow of interest, narrowing the sec-
e
flow away from the transducer in blue, with the shade tor, and decreasing the depth as much as possible
s
of color indicating velocity up to the Nyquist limit. (Fig. 1-30). When frame rate remains inadequate for
The option of displaying “variance” allows an addi- timing flow abnormalities, a color M-line through
/r u
tional color (usually green) to be added to indicate that the area of interest may be helpful, for example in
there was variability in the estimated mean velocity assessment of aortic regurgitation.
for the eight bursts along that sample line, indicating
either a flow disturbance or aliasing of a high-velocity
.t c
Color Doppler Imaging Artifacts
signal. This process is repeated for each adjacent scan
line across the image plane. Because each of these pro- Color flow artifacts again relate to the physics of 2D
cesses takes a finite amount of time depending on the and Doppler flow image generation (Table 1-9). Shad-
a
speed of sound in tissue, the rapidity with which this owing may be prominent distal to strong reflectors with
image can be updated (the frame rate) depends on a absence of both 2D and flow data within the acoustic
k
combination of these factors. shadow.
/: /
Ghosting is the appearance of brief (usually one or
two frames) large color patterns that overlay anatomic
Color Doppler Instrument Controls structures and do not correspond to underlying flow
s
The color flow display is dependent on each specific patterns. This artifact is caused by strong moving
ultrasound instrument to some extent. However, many reflectors (such as prosthetic valve disks). Typically, this
tt p
parameters are adjustable by the operator, so an opti- artifact is solid red or blue and is inconsistent from beat
mal examination requires careful attention to instru- to beat.
ment settings. Color Doppler gain settings have a dramatic effect
on the color flow image. Extensive gain settings result
h
The color flow map usually can be varied in terms of:
in a uniform speckled pattern across the 2D image
n Color scale (assignment of colors to direction
plane resulting from random background noise. Con-
and velocity)
versely, too low a gain setting results in a smaller dis-
n Velocity range (within the Nyquist limit at that
played flow area than is actually present, an effect
depth)
colloquially known as “dial-a-jet.” Most experienced
n Zero baseline position on the color scale
echocardiographers recommend setting the gain level
n Addition of variance to the color scale
just below the level of random background noise to
The specific color scale used is a matter of personal optimize the flow signal.
preference, with the diagnostic goal being to opti- As for any Doppler technique, the intercept angle
mize the display and recognition of abnormal flow between the ultrasound beam and direction of blood
patterns. flow for each scan line affects the color display in terms
The velocity range of the color flow map is determined of both direction and velocity. Thus a uniform flow
by the Nyquist limit, and as for conventional pulsed velocity traversing the image plane may appear red
Doppler, the range can be altered by shifting the zero (toward the transducer) at one side of the sector and
Another random document with
no related content on Scribd:
DANCE ON STILTS AT THE GIRLS’ UNYAGO, NIUCHI
I see increasing reason to believe that the view formed some time
back as to the origin of the Makonde bush is the correct one. I have
no doubt that it is not a natural product, but the result of human
occupation. Those parts of the high country where man—as a very
slight amount of practice enables the eye to perceive at once—has not
yet penetrated with axe and hoe, are still occupied by a splendid
timber forest quite able to sustain a comparison with our mixed
forests in Germany. But wherever man has once built his hut or tilled
his field, this horrible bush springs up. Every phase of this process
may be seen in the course of a couple of hours’ walk along the main
road. From the bush to right or left, one hears the sound of the axe—
not from one spot only, but from several directions at once. A few
steps further on, we can see what is taking place. The brush has been
cut down and piled up in heaps to the height of a yard or more,
between which the trunks of the large trees stand up like the last
pillars of a magnificent ruined building. These, too, present a
melancholy spectacle: the destructive Makonde have ringed them—
cut a broad strip of bark all round to ensure their dying off—and also
piled up pyramids of brush round them. Father and son, mother and
son-in-law, are chopping away perseveringly in the background—too
busy, almost, to look round at the white stranger, who usually excites
so much interest. If you pass by the same place a week later, the piles
of brushwood have disappeared and a thick layer of ashes has taken
the place of the green forest. The large trees stretch their
smouldering trunks and branches in dumb accusation to heaven—if
they have not already fallen and been more or less reduced to ashes,
perhaps only showing as a white stripe on the dark ground.
This work of destruction is carried out by the Makonde alike on the
virgin forest and on the bush which has sprung up on sites already
cultivated and deserted. In the second case they are saved the trouble
of burning the large trees, these being entirely absent in the
secondary bush.
After burning this piece of forest ground and loosening it with the
hoe, the native sows his corn and plants his vegetables. All over the
country, he goes in for bed-culture, which requires, and, in fact,
receives, the most careful attention. Weeds are nowhere tolerated in
the south of German East Africa. The crops may fail on the plains,
where droughts are frequent, but never on the plateau with its
abundant rains and heavy dews. Its fortunate inhabitants even have
the satisfaction of seeing the proud Wayao and Wamakua working
for them as labourers, driven by hunger to serve where they were
accustomed to rule.
But the light, sandy soil is soon exhausted, and would yield no
harvest the second year if cultivated twice running. This fact has
been familiar to the native for ages; consequently he provides in
time, and, while his crop is growing, prepares the next plot with axe
and firebrand. Next year he plants this with his various crops and
lets the first piece lie fallow. For a short time it remains waste and
desolate; then nature steps in to repair the destruction wrought by
man; a thousand new growths spring out of the exhausted soil, and
even the old stumps put forth fresh shoots. Next year the new growth
is up to one’s knees, and in a few years more it is that terrible,
impenetrable bush, which maintains its position till the black
occupier of the land has made the round of all the available sites and
come back to his starting point.
The Makonde are, body and soul, so to speak, one with this bush.
According to my Yao informants, indeed, their name means nothing
else but “bush people.” Their own tradition says that they have been
settled up here for a very long time, but to my surprise they laid great
stress on an original immigration. Their old homes were in the
south-east, near Mikindani and the mouth of the Rovuma, whence
their peaceful forefathers were driven by the continual raids of the
Sakalavas from Madagascar and the warlike Shirazis[47] of the coast,
to take refuge on the almost inaccessible plateau. I have studied
African ethnology for twenty years, but the fact that changes of
population in this apparently quiet and peaceable corner of the earth
could have been occasioned by outside enterprises taking place on
the high seas, was completely new to me. It is, no doubt, however,
correct.
The charming tribal legend of the Makonde—besides informing us
of other interesting matters—explains why they have to live in the
thickest of the bush and a long way from the edge of the plateau,
instead of making their permanent homes beside the purling brooks
and springs of the low country.
“The place where the tribe originated is Mahuta, on the southern
side of the plateau towards the Rovuma, where of old time there was
nothing but thick bush. Out of this bush came a man who never
washed himself or shaved his head, and who ate and drank but little.
He went out and made a human figure from the wood of a tree
growing in the open country, which he took home to his abode in the
bush and there set it upright. In the night this image came to life and
was a woman. The man and woman went down together to the
Rovuma to wash themselves. Here the woman gave birth to a still-
born child. They left that place and passed over the high land into the
valley of the Mbemkuru, where the woman had another child, which
was also born dead. Then they returned to the high bush country of
Mahuta, where the third child was born, which lived and grew up. In
course of time, the couple had many more children, and called
themselves Wamatanda. These were the ancestral stock of the
Makonde, also called Wamakonde,[48] i.e., aborigines. Their
forefather, the man from the bush, gave his children the command to
bury their dead upright, in memory of the mother of their race who
was cut out of wood and awoke to life when standing upright. He also
warned them against settling in the valleys and near large streams,
for sickness and death dwelt there. They were to make it a rule to
have their huts at least an hour’s walk from the nearest watering-
place; then their children would thrive and escape illness.”
The explanation of the name Makonde given by my informants is
somewhat different from that contained in the above legend, which I
extract from a little book (small, but packed with information), by
Pater Adams, entitled Lindi und sein Hinterland. Otherwise, my
results agree exactly with the statements of the legend. Washing?
Hapana—there is no such thing. Why should they do so? As it is, the
supply of water scarcely suffices for cooking and drinking; other
people do not wash, so why should the Makonde distinguish himself
by such needless eccentricity? As for shaving the head, the short,
woolly crop scarcely needs it,[49] so the second ancestral precept is
likewise easy enough to follow. Beyond this, however, there is
nothing ridiculous in the ancestor’s advice. I have obtained from
various local artists a fairly large number of figures carved in wood,
ranging from fifteen to twenty-three inches in height, and
representing women belonging to the great group of the Mavia,
Makonde, and Matambwe tribes. The carving is remarkably well
done and renders the female type with great accuracy, especially the
keloid ornamentation, to be described later on. As to the object and
meaning of their works the sculptors either could or (more probably)
would tell me nothing, and I was forced to content myself with the
scanty information vouchsafed by one man, who said that the figures
were merely intended to represent the nembo—the artificial
deformations of pelele, ear-discs, and keloids. The legend recorded
by Pater Adams places these figures in a new light. They must surely
be more than mere dolls; and we may even venture to assume that
they are—though the majority of present-day Makonde are probably
unaware of the fact—representations of the tribal ancestress.
The references in the legend to the descent from Mahuta to the
Rovuma, and to a journey across the highlands into the Mbekuru
valley, undoubtedly indicate the previous history of the tribe, the
travels of the ancestral pair typifying the migrations of their
descendants. The descent to the neighbouring Rovuma valley, with
its extraordinary fertility and great abundance of game, is intelligible
at a glance—but the crossing of the Lukuledi depression, the ascent
to the Rondo Plateau and the descent to the Mbemkuru, also lie
within the bounds of probability, for all these districts have exactly
the same character as the extreme south. Now, however, comes a
point of especial interest for our bacteriological age. The primitive
Makonde did not enjoy their lives in the marshy river-valleys.
Disease raged among them, and many died. It was only after they
had returned to their original home near Mahuta, that the health
conditions of these people improved. We are very apt to think of the
African as a stupid person whose ignorance of nature is only equalled
by his fear of it, and who looks on all mishaps as caused by evil
spirits and malignant natural powers. It is much more correct to
assume in this case that the people very early learnt to distinguish
districts infested with malaria from those where it is absent.
This knowledge is crystallized in the
ancestral warning against settling in the
valleys and near the great waters, the
dwelling-places of disease and death. At the
same time, for security against the hostile
Mavia south of the Rovuma, it was enacted
that every settlement must be not less than a
certain distance from the southern edge of the
plateau. Such in fact is their mode of life at the
present day. It is not such a bad one, and
certainly they are both safer and more
comfortable than the Makua, the recent
intruders from the south, who have made USUAL METHOD OF
good their footing on the western edge of the CLOSING HUT-DOOR
plateau, extending over a fairly wide belt of
country. Neither Makua nor Makonde show in their dwellings
anything of the size and comeliness of the Yao houses in the plain,
especially at Masasi, Chingulungulu and Zuza’s. Jumbe Chauro, a
Makonde hamlet not far from Newala, on the road to Mahuta, is the
most important settlement of the tribe I have yet seen, and has fairly
spacious huts. But how slovenly is their construction compared with
the palatial residences of the elephant-hunters living in the plain.
The roofs are still more untidy than in the general run of huts during
the dry season, the walls show here and there the scanty beginnings
or the lamentable remains of the mud plastering, and the interior is a
veritable dog-kennel; dirt, dust and disorder everywhere. A few huts
only show any attempt at division into rooms, and this consists
merely of very roughly-made bamboo partitions. In one point alone
have I noticed any indication of progress—in the method of fastening
the door. Houses all over the south are secured in a simple but
ingenious manner. The door consists of a set of stout pieces of wood
or bamboo, tied with bark-string to two cross-pieces, and moving in
two grooves round one of the door-posts, so as to open inwards. If
the owner wishes to leave home, he takes two logs as thick as a man’s
upper arm and about a yard long. One of these is placed obliquely
against the middle of the door from the inside, so as to form an angle
of from 60° to 75° with the ground. He then places the second piece
horizontally across the first, pressing it downward with all his might.
It is kept in place by two strong posts planted in the ground a few
inches inside the door. This fastening is absolutely safe, but of course
cannot be applied to both doors at once, otherwise how could the
owner leave or enter his house? I have not yet succeeded in finding
out how the back door is fastened.