Battery Recovery
(PC Battery)
See following images to identify the PS2 touch screen connector on SECO M515 and SECO
M585.
B. Unplug the touch screen the touch screen adapter and plug the supplied PS2 keyboard
adapter to the connector, taking care to line PIN 1 of the adapter to PIN 1 of the PC board
connector.
C. The PS2 Keyboard must be connected to the PS2 connector of the supplied adapter.
D. Switch the equipment on and press repetitively the Del key on the keyboard during the boot.
The equipment will enter in BIOS setup mode,
E. In Basic configuration page, set proper time and date. In “ADVANCED CMOS
CONFIGURATION” be sure to enable “Ignore Keyboard error”
Properly configure BIOS should appear as in the following image (Note that entry positions may
change slightly according to the PC board model)
Important Note page 3/3
(PC Battery)
F. Press ESC to return on the main page, then select “Save CMOS and exit”.
The Pulsar 2 software should start normally.
G. Switch the equipment off, disconnect the keyboard adapter and re-connect the touch screen
paying attention to properly align PIN 1.
A. connect an USB keyboard to the “Remote” connector on the rear panel using the dedicated
adapter.
B. Switch off the equipment and then on again, during Power-on, hit the “Del” key repetitively
to get to the Password request screen.
C. Press “Enter” (do not input any password). The system will enter “System bios setup”.
D. Set Time and Date, hit F10 “Save CMOS and exit”