Acclaim®
Mixed-Mode WCX-1
Product Manual for the Acclaim Mixed-Mode WCX-1 Column Page 1 of 28
Product Manual
for
Acclaim Mixed-Mode WCX-1, 5µm, Analytical column, 2.1 x 150mm (P/N 068371)
Acclaim Mixed-Mode WCX-1, 5µm, Analytical column, 4.6 x 150mm (P/N 068353)
Acclaim Mixed-Mode WCX-1, 5µm, Analytical column, 4.6 x 250mm (P/N 068352)
Acclaim Mixed-Mode WCX-1, 5µm, Guard column, 4.3 x 10mm (P/N 068354)
Acclaim Mixed-Mode WCX-1, 5µm, Guard column, 3.0 x 10mm (P/N 071911)
Acclaim Mixed-Mode WCX-1, 5µm, Guard column, 4.6 x 10mm (P/N 069705)
TABLE OF CONTENTS
SECTION 1 – INTRODUCTION
The Acclaim Mixed-Mode WCX-1 column is based on a new mixed-mode silica-based packing material that incorporates
both hydrophobic and weak cation-exchange properties. Unlike traditional reversed-phase stationary phases, the new packing
features an alkyl long chain with an ionizable terminus, and demonstrates great potentials for separating a wide range of
cationic compounds containing mixtures, including pharmaceuticals, food & beverage, chemical, and more.
Ion exchange columns are used to separate ionic or ionizable compounds such as proteins, nucleic acids, inorganic ions,
small organic acids, etc. Because most conventional ion-exchange stationary phases provide inadequate hydrophobic
retention for neutral molecules, they have limited applications in small molecules separations.
Ion pairing chromatography is a method for separating ionic or ionizable compounds on a conventional RP medium, which
requires hydrophobic ionic compounds, typically comprised of an alkyl chain with an ionizable terminus, are added to the
mobile phase. Generally, retention of neutral analytes is nearly unaffected, while analytes with charges complementary to the
ion pairing reagent are retained for a longer period of time and analytes with the same charge as the ion pairing reagent are
retained for a shorter period of time. Limitations of ion pairing chromatography include long column equilibration times and
the quantity of solvent and time needed to elute the ion pairing reagent from the column.
Mixed mode chromatography combines aspects of ion exchange chromatography and conventional reversed-phase
chromatography. A mixed-mode stationary phase has both hydrophobic and ion-exchange properties (see figure 1). These
two strong interactions of the phase with analytes allow for controlling retention of ionizable and neutral molecules
independently. As a result, many application challenges involving hydrophilic ionizable compounds that are difficult for C18
columns, can be easily tackled on a mixed-mode column.
O- O- O- O- O-
O CO C O C O CO C
Silica Gel
Figure 1
Acclaim Mixed-Mode WCX-1
1.2. Features
Adjustable selectivity (Figures 2-4)
Selectivity complementary to RP columns (Figure 6)
Orthogonal Selectivity (Figures 7-8)
Ideal selectivity for separating basic molecules (Figures 10-14)
Multimode separation mechanism: reversed-phase, cation-exchange, anion-exclusion, and HILIC (Figures 5, 16 and 17)
Simultaneous separation of acidic, neutral and basic compounds (Figure 9)
! Degas the aqueous component of the mobile phase and then add the solvent component. Avoid excessive
purging or degassing of mobile phases containing solvents, if possible, since the volatile solvent can be 'boiled'
NOTE off from the solution.
2.2.2. Solvents
The solvents used must be free from ionic and UV-absorbing impurities. Use of ultrahigh purity solvents, HPLC grade, will
usually ensure that your chromatography is not affected by impurities in the solvent.
! Due to various reasons, such as difference of LC systems, mobile phases, oven temperature control, etc, you may
observe slightly different retention time for the iodide peak from that in the report.
NOTE
100
min
-20
0.00 1.00 2.00 3.00 4.00 5.00 6.42
QA Results:
Analyte Parameter Specification Results
Naphthalene Efficiency >=10,800 Passed
Naphthalene Asymmetry 0.95-1.32 Passed
Naphthalene Retention Time 4.8-5.8 Passed
Pressure <=1320 Passed
Production Reference:
Datasource: QAR
Directory: Acclaim\Mixed-Mode_WCX
Sequence: 068353_MMWCX_46X150MM
Sample No: 1
Chromeleon® Dionex Corp. 1994-2008
Serial No.: 4
220
mAU WVL:254 nm
1
100
min
-20
0.0 1.3 2.5 3.8 5.0 6.3 7.5 8.8 10.0 12.0
QA Results:
Analyte Parameter Specification Results
Naphthalene Efficiency >=16,200 Passed
Naphthalene Asymmetry 0.95-1.32 Passed
Naphthalene Retention Time 8.0-9.6 Passed
Pressure <=1980 1377
Production Reference:
Datasource: QAR
Directory: Acclaim\Mixed-Mode_WCX
Sequence: 068352_MMWcx_46X250m m
Sample No: 1
Chromeleon® Dionex Corp. 1994-2008
1
2
AU
1
2 40 mM buffer
3
0 6 12
Minutes
Figure 2
Adjustable Selectivity - Ionic Strength Effect
AU
1 2
50% MeCN (v/v )
3
0 5 10
Minutes
Figure 3
Adjustable Selectivity - Organic Modifier Effect
1 Switch “On”
3
pH6.5
AU
3 2 Switch “Off”
1
pH2.8
0 6 12
Minutes
Figure 4
Adjustable Selectivity - pH Effect
2
HILIC Mode
AU
1
RP Mode
0 5 10
Minutes
Figure 5
RP Mode vs. HILIC Mode
1. Wash the column with 20 mM ammonium (or sodium) acetate buffer, pH5/ acetonitrile v/v 50/50 for 3 column
volumes at the recommended flow rate (see Table in Section 1.3).
2. Wash the column with 100 mM ammonium (or sodium) acetate buffer, pH5/ acetonitrile v/v 50/50 for 20 column
volumes at a flow rate between 0.2 to 1 mL/min (to remove strongly retained cationic species).
3. Wash the column with 0.1% oxalic acid in D.I. water for 20 column volumes at a flow rate between 0.2 to 1 mL/min
(to remove metal contamination).
4. Wash the column with 20 mM ammonium (or sodium) acetate buffer, pH5/ acetonitrile v/v 50/50 for 3 column
volumes at a flow rate between 0.2 to 1 mL/min.
5. Wash the column with 20 mM ammonium (or sodium) acetate buffer, pH5/ acetonitrile v/v 25/75 for 20 column
volumes at a flow rate between 0.5 to 1 mL/min (to remove strongly retained hydrophobic compounds).
6. Equilibrate the column with the mobile phase. Note: before any injection is made, the column should be equilibrated
with a mobile phase for at least 30 column volumes.
! An ammonium acetate buffer can be used instead of a phosphate buffer if a LC-MS application is intended.
NOTE If above treatment fails to improve the column performance, replace it with a new one.
5.4. What factors should I consider for method development using this column?
There are three main factors that affect column selectivity: mobile phase ionic strength, mobile phase pH, and mobile phase
organic composition. You can optimize your separation by changing one, two, or all three factors.
5.11. Can I use this column to separate a mixture of basic, acidic, and neutral molecules?
Yes. As shown in Figure 9, the Acclaim Mixed-Mode WCX-1 separates a mixture of basic, neutral, and acidic molecules in a
single, with excellent peak shape and resolution. It provides greater flexibility for application method development compared
to both conventional reversed-phase and ion-exchange columns.
5.12. Do I need a guard cartridge with an Acclaim Mixed-Mode WCX-1 analytical column?
Yes. It is highly recommended to use guard cartridges with an Acclaim Mixed-Mode WCX-1 analytical column. The guard
cartridge protects the more expensive analytical column by trapping highly retained components and particulates from the
mobile phase or the sample.
Column: 5 µm
Dimension: 150x4.6 mm
Mobile Phase: 50/50 v/v MeCN/10 mM (total) sodium phosphate buffer, pH6.5
Temperature: 30 °C
Flow Rate: 1 mL/min
Inj. Volume: 5 µL
Detection: UV (215 nm)
Peaks: µg/mL
1. Benzoic acid 200
2. Naphthalene 50
3. Benzyl amine 300
1
Mixed–Mode WCX-1
2
AU 1,3
Acclaim 120 C8
0 6 12
Minutes
Figure 6
WCX vs. RP – Orthogonal Selectivity
Column: 5 µm
Dimension: 150x4.6 mm
Mobile Phase: 60/40 v/v MeCN/20 mM sodium phosphate buffer, pH6.5
Temperature: 30 °C
Flow Rate: 1 mL/min
Inj. Volume: 5 µL
Detection: UV (215 nm)
Peaks: µg/mL
1. Benzoic acid 200
2. Naphthalene 50
3. Benzyl amine 300
Mixed–Mode WCX-1
2
1
3
AU
2
Mixed–Mode WAX-1
3
1
0 4 8
Minutes
Figure 7
WCX vs. WAX – Orthogonal Selectivity
Column: 5 µm
Dimension: 4.6x150 mm
Mobile Phase: 60/40 v/v MeCN/sodium phosphate, pH6.5
40 mM for Mixed-Mode WCX-1
10 mM for Mixed-Mode WAX-1
Temperature: 30 °C
Flow Rate: 1 mL/min
Inj. Volume: 2 µL
Detection: UV (215 nm)
Sample: Trimipramine Maleate (250 ppm)
Peaks:
1. Maleate 2. Trimipramine
1
Mixed–Mode WCX-1
AU 2
Mixed–Mode WAX-1
0 5 10
Minutes
Figure 8
WCX vs. WAX – Orthogonal Selectivity
SECTION 7 – APPLICATIONS
7
AU
2
5
4 6
0 4 8
Minutes
Figure 9
Simultaneous Separation of Acidic, Neutral and Basic Pharmaceuticals
1
3
AU
0 7.5 15
Minutes
Figure 10
Separation of Catecholamines
AU 1 4 5 6
2
3
0 10 20
Minutes
Figure 11
Separation of Antidepressants
1
3
40 mM
mV 1 3
30 mM
1 3 20 mM
0 7.5 15
Minutes
Figure 12
Analysis of Quaternary Amines
1
2
5 mM
2
1
mV 10 mM
2
20 mM
0 4 8
Minutes
Figure 13
Analysis Tris HCl Salt
1
40 mM
2
1
mV
20 mM
1
2
10 mM
0 6 12
Minutes
Figure 14
Analysis of Alkyl Phosphonium Salt
Sample preparation:
1. Finely grind a 1500 mg tablet
2. Mix 0.2 g material in 7.5 mL D.I. water and 2.5 mL MeOH
3. Sonicate for 20 min
4. Dilute 20 times
5. Filter through 0.1 µm membrane filter
mV
0 4 8
Minutes
Figure 15
Analysis of Glucosamine Tablet
* Total concentration
2
15 mM*
1
mV 2
10 mM*
1
2
5 mM*
0 4 8
Minutes
Figure 16
Analysis of NaCl – Ionic Strength Effect
1
2
pH3.5
1
2
pH3.8
1
mV
2
pH4.1
1
2
pH4.4
0 2.5 5
Minutes
Figure 17
Analysis of CaCl2 – pH Effect