Professional Documents
Culture Documents
ST-200
ST-200CC Arterial Blood Gas Electrolyte Analyzer Service Manual
Contents
1 Introduction …... 5
1.1 The Analyzer ….. 5
1.2 Operational ….. 6
1.3 Cautions and Hazards ….. 6
1.4 Required tools ….. 7
1.5 Terms ….. 8
2 Product Description ….. 9
2.1 Front Assembly Components ….. 11
2.2 Fluid Circuit parts ….. 11
2.3 Electrode Stacking ….. 13
2.4 Electrode Housing ….. 13
2.5 Mechanical Assemblies ….. 15
2.5.1 Valve Module ….. 15
2.5.2 Sampler ….. 15
2.5.3 Pump motor ….. 16
2.5.4 Flip Thermal System ….. 17
2.6 Electronic Assemblies ….. 17
2.6.1 Power Distribution ….. 17
2.6.2 Spare parts-Description/part number ….. 17
2.6.3 Mother board ….. 18
2.6.4 Pre-Amplifier Board ….. 20
2.6.5 Motor Driver board ….. 21
2.6.6 RFID Board ….. 21
2.6.7 Ultrasonic Bubble Detector Board ….. 22
2.6.8 Display Assembly ….. 23
2.6.9 Printer Assembly ….. 23
2.6.10 Peripheral boards ….. 24
2.6.11 AC/DC Switching Adapter ….. 25
2.7 Rear Panel ….. 26
3 Replacement procedures ….. 27
1
ST-200CC Arterial Blood Gas Electrolyte Analyzer Service Manual
2
ST-200CC Arterial Blood Gas Electrolyte Analyzer Service Manual
3
ST-200CC Arterial Blood Gas Electrolyte Analyzer Service Manual
4
ST-200CC Arterial Blood Gas Electrolyte Analyzer Service Manual
1. Introduction
This manual contains information and procedures used in servicing the
ST-200CC ABGE Analyzer. The intent is to supplement the information provided in
the ST-200CC Instructions and the Advanced User Applications for Use Manual.
This manual does not supercede any current performance specification claims,
expendable maintenance procedures, or warranty criteria as outlined in the
reference manual.
Environmental
Ambient Temperature : 5° - 28° C (41 ° - 80 ° F)
Ambient Humidity : 5% - 85%
Energy consumption
Power Consumption : ~240 watts peak
Power requirements : 100/115 V AC, 50-60 Hz or 220 V AC, 50-60 Hz,
0.75 Amp.
Disturbances on the power line may be caused by several factors such as:
Starting and stop any of electric motors.
Atmospheric phenomena.
Connection and disconnection of high line loads.
Generator and square waves inverters.
Physical Dimensions
Weight : 8 kg (19 lbs.)
Dimensions : 15" W (38 cm) x 12" H (30 cm) x 7" D (18 cm)
Regulatory Compliance
The analyzer is tested and certified by International Certifications (CE)
Location
Place on a stable, stationary surface.
~8 cm ( 3 in) clearance to a wall or adjacent instrumentation
Not intended for operating room use
Not intended for use on vibrating ( helicopter, etc.) surface
Placement on a cart is acceptable if the cart is stationary during operation.
5
ST-200CC Arterial Blood Gas Electrolyte Analyzer Service Manual
1.2 Operational
Any shutdown period will require a 2-point calibration and verification by control
material upon restarting operation.
The analyzer utilizes 3 control boards. One is Mother board which consists
MSP430F6779 Microcontroller. Mother board gives proper supply to each and
every component mounted on PCB by using Voltage Regulators. Mother board
takes +12 VDC power supply as input.
The second one is Pre Amplifier board which converts all absorbance readings to
an analog signals to digital signals for each channel and processing of the signals
is done by an analog and microcontroller mounted on the Mother board. It takes
+5 VDC from the voltage regulators which are present on Mother board.
The third one is Motor Driver board which receives +12 VDC input voltage from
the power supply assembly and supplies voltage to Pump motor in one direction
and to Pinch valves for switching.
These 3 control boards communicates with the male and female connectors.
The display is 7 inch high definition lit liquid crystal (LCD) with capacitive touch
and 1024x552 pixel resolution. It communicates via the OTG connection with the
controller and takes +5 VDC input power supply.
NOTE: The circuit boards use fine pitch surface mount technology. Field repair of
these circuit boards is not possible. DO NOT attempt to repair any components
mounted on the board.
There are NO user serviceable assemblies inside the analyzer. Only a trained,
authorized service representative should remove the spare parts of the Analyzer.
6
ST-200CC Arterial Blood Gas Electrolyte Analyzer Service Manual
WARNING: Internal surfaces may have sharp edges. Care should be taken
to avoid cuts and scrapes when accessing internal assemblies. Used tubing,
bio-sensors, electrodes, reagents, controls, etc. may be disposed by
normal laboratory waste procedures.
7
ST-200CC Arterial Blood Gas Electrolyte Analyzer Service Manual
1.5 Terms
8
ST-200CC Arterial Blood Gas Electrolyte Analyzer Service Manual
2. Product Description
The ST-200CC Arterial Blood Gas Electrolyte Analyzer is an automated
microprocessor controlled analyzer. It is intended for in-vitro diagnostics use in
the quantitative determination of sodium, potassium, ionized calcium, lithium, pH,
chloride, pCO2, pO2, Hct, Hb, HCO-3, BE, BE-B, BE-ECF, TCO2, Ag_Na, Ag_K, SO2,
O2Ct, SBC, FiO2, A, AaDO2, a/A, Glu (Glucose), Lac (Lactate) in Whole Blood,
Electrolytes in Serum/Plasma/CSF and Diluted Urine. Measurements obtained by
this device are used for the diagnosis, monitoring, and treatment of diseases
involving blood gas, metabolites, Hematocrit and electrolyte imbalance.
To obtain accurate results, the ST-200CC Arterial Blood Gas Electrolyte
analyzer must be operated with specially packaged calibrant.
The instrument is fitted with the flow system of Electrode housing which
ensures a low level of carry over even with limited sample volume.
The peristaltic pump guarantees a high level of precision in the aspirated
volume.
The instrument is having 7 inch high definition LCD with capacitive touch and
it guides user through all the aspects of analyzer operation, including analyzer
setup, information entry and display and setting.
The test results are shown directly in the major units and are printed on
heat-sensitive paper.
Measured parameters:
pH, pCO2, pO2, Hct, Na+, K+, Cl-, iCa++, Li+
Glucose and Lactate are optional.
Hct Whole blood measurements of the packed red cell volume of a blood
sample are used to distinguish normal from abnormal states, such as
anemia and erythrocytosis (an increase in the number of red cells).
9
ST-200CC Arterial Blood Gas Electrolyte Analyzer Service Manual
Calculated parameters:
HCO3-, TCO2, SBC, SO2, O2Ct, pO2%, BE, BE_B, BE_ECF, AG_Na, AG_K, A, AaDO2,
a/A
The Sample:
Lithium heparin or Sodium heparin whole blood samples from syringes, open
tubes, and capillary tubes can be used on the ST-200 CC Arterial Blood Gas
Analyzer. The minimum sample size for analysis is 140 μL.
Suitable Anticogulants:
Lithium heparin and Sodium heparin are the acceptable anticoagulant for use
with the analyzer.
EDTA, citrate, oxalate and sodium fluoride ARE NOT acceptable for use.
Depending on the amount of heparin used in the collection syringe and
whether it is filled to capacity with blood, heparin concentrations of result
may vary.
Liquid or dry heparin when present in excess may cause errors. Ensure blood
collection devices are filled per manufacturer instructions.
CAUTION:
ST-200CC Arterial Blood Gas Electrolyte Analyzer users should take careful
note of these considerations when establishing reference intervals and
interpreting results.
10
ST-200CC Arterial Blood Gas Electrolyte Analyzer Service Manual
8
1
2
9
10
3
11
12
13
5
6
14
7
The following figure shows the main parts constituting the fluidic circuit for
aspiration of a sample from the sample probe.
a) Valve module
b) Sampler
c) Electrode housing
d) Pump motor
11
ST-200CC Arterial Blood Gas Electrolyte Analyzer Service Manual
12
ST-200CC Arterial Blood Gas Electrolyte Analyzer Service Manual
4. Slide the stacked electrode set in Reference Housing. Push Down firmly until
the sample detector snaps into its position. Insert the Reference lock (SM lock)
to complete the stacking.
5. Connect the sample tube at the top of Bubble Detector.
Note:
Electrode parameters
13
ST-200CC Arterial Blood Gas Electrolyte Analyzer Service Manual
Reference Seal
Reference Membrane
Removal of electrodes:
Slide the electrode stack up at least one inch by gently pushing up on the
bottom electrode pin. This process will remove all electrodes from the reference
housing.
14
ST-200CC Arterial Blood Gas Electrolyte Analyzer Service Manual
B
C
2.5.2 Sampler
Sampler or sample probe is used to intake sample, which consist of a hole at
the bottom of the probe. It has a self probe wiping facility with three different
aspiration modes (sample tube, sample syringe and capillary).
15
ST-200CC Arterial Blood Gas Electrolyte Analyzer Service Manual
Figure 7 Sampler
The pump tube has two stoppers. One stopper (red) is connected at the base
of the electrode housing and the other stopper (blue) is connected to the reagent
pack socket i.e. to the waste.
Pump module is used to generate vacuum in pump tube for aspirating liquid.
Pump module consist of four rollers.
16
ST-200CC Arterial Blood Gas Electrolyte Analyzer Service Manual
WASTE
CAL-C
CAL-B
CAL-A
17
ST-200CC Arterial Blood Gas Electrolyte Analyzer Service Manual
Main Functions:
All the electronic circuits processing the analogical and digital signals and
driving the motors and sensors on boards are controlled by the Mother board.
It distributes proper power supply to each and every component mounted on
PCB by using Voltage Regulators. +12 VDC power is delivered to the Mother board
via J27 Connector/Cable from DC Adapter.
TAB OTG RFID Connector
Pre-amplifier Connector
RS232
TAB Power Printer
Connector Connector Connector
USB
Connector-1
USB
Microcontroller
Connector-2
(24-bit)
USB LIS
Temperature Connector
Sensor Connector J-TAG
Buzzer Connector
Bubble Detector
Connector Jumper
Optical Sensor Connector
Connector
12 VDC
Power Socket
Pump Motor
Connector
Heater 1 Heater 3
Motor Driver
Heater 2 Card
18
ST-200CC Arterial Blood Gas Electrolyte Analyzer Service Manual
Specification of Fuses:
Mother board consist of 6 fuses to protect IC’s on the mother board when
fluctuations are occurred.
TP 6(12V) TP 4(5V)
TP 1(2.5V)
19
ST-200CC Arterial Blood Gas Electrolyte Analyzer Service Manual
Main Functions:
Pre amplifier is used to measure the voltages near every channel (electrode).
It receives the sensor output and performs the A/D conversion. The digitized
signal is then passed to the microcontroller for processing.
20
ST-200CC Arterial Blood Gas Electrolyte Analyzer Service Manual
Main Functions:
Motor driver board is used to operate the pump motor and pinch valves for
rotating and switching by supplying +12VDC.
The second pins voltages of Heater 1, Heater 2 & Heater 3 is from +2.5 VDC
to +5 VDC when pinch valve is in rest condition and 0 VDC when pinch valve is in
work condition while switching.
Main Functions:
After giving input +5 VDC, the LED 3 (power, Red LED) which presents on the
circuit glows and when label on the reagent pack is detected, the LED 6 (protocol,
Orange LED) glows.
21
ST-200CC Arterial Blood Gas Electrolyte Analyzer Service Manual
Main Functions:
Ultrasonic Bubble Detector board communicates between Bubble Detector
and micro-controller bidirectionally. It passes signal to micro-controller while the
sound waves are transmitted from the transmitter sensor to receiver sensor when
the Bubble Detector flow path is filled with liquid and the Red LED on the PCB will
glow.
22
ST-200CC Arterial Blood Gas Electrolyte Analyzer Service Manual
Main functions:
It guides user through all aspects of analyzer operation, including
analyzer setup, information entry and display, and testing. The screen is
powered-on by pressing the power button of the touchscreen panel.
Power button
23
ST-200CC Arterial Blood Gas Electrolyte Analyzer Service Manual
Note:
Do not use force to pull out the paper from the printer.
Never pull the paper in the reverse direction of printing.
The Feed button is only active when the printer detects paper in the printer.
Printer mechanism should have thermal paper roll loaded compulsory to increase
life of printer even printer is not in use.
Main functions:
SMPS/LIS USB board: It receives +12V DC input voltage from power supply
assembly (DC Adapter) through the port and distributes to exhaust fan and
Mother board/all components of the analyzer.
USB type A board: It consists two USB ports and connects to the Mother board
for connectivity of Keyboard, Mouse, Bar code reader etc.
24
ST-200CC Arterial Blood Gas Electrolyte Analyzer Service Manual
J-Tag connectivity board: It has 14 pin FRC connector. It is used for the software
upgradation and connectivity between analyzer and the PC.
Exhaust fan: It receives +12V DC from the SMPS/LIS USB board. It is used for the
protection of components in the analyzer from excess heat.
Exhaust Fan
25
ST-200CC Arterial Blood Gas Electrolyte Analyzer Service Manual
Air Vent
USB Ports
LIS USB Port Power ON/OFF Switch
J-Tag port RS232 Serial port
Power Socket
26
ST-200CC Arterial Blood Gas Electrolyte Analyzer Service Manual
3. Replacement Procedures
WARNING: Removal of the analyzer rear cover exposes the line voltage
wiring. The Power must be turned off and the Power supply disconnected
before any assembly replacement procedure is performed.
1. Turn the unit OFF, disconnect the power adapter, and have the back of
the unit facing you.
2. Remove the 7 panel mounting screws.
3. Slightly spread the top sides and remove the cover.
4. Lift the cover off the unit. Take care not to damage any cables or wiring
during removal.
27
ST-200CC Arterial Blood Gas Electrolyte Analyzer Service Manual
2 6
3
8
WARNING: The Power must be turned off and the Power Adapter disconnected
before changing the Mother board.
28
ST-200CC Arterial Blood Gas Electrolyte Analyzer Service Manual
Pre Amplifier
board
29
ST-200CC Arterial Blood Gas Electrolyte Analyzer Service Manual
WARNING: The Power must be turned off and the Power Adapter disconnected
before changing the Pre Amplifier board.
NOTE: Make sure that the pinch valves connectors are connected correctly to its
own positions as shown in figure.
WARNING: The Power must be turned off and the Power Adapter disconnected
before changing the Motor Driver board.
Motor driver
board
Pinch Valves
V1 V2 V3
(V1, V2 & V3)
Connectors
WARNING: The Power must be turned off and the Power Adapter disconnected
before changing the RFID board.
30
ST-200CC Arterial Blood Gas Electrolyte Analyzer Service Manual
WARNING: The Power must be turned off and the Power Adapter disconnected
before changing the Printer assembly.
31
ST-200CC Arterial Blood Gas Electrolyte Analyzer Service Manual
WARNING: The Power must be turned off and the Power Adapter disconnected
before changing the Printer assembly.
WARNING: The Power must be turned off and the Power Adapter disconnected
before changing the Printer assembly.
32
ST-200CC Arterial Blood Gas Electrolyte Analyzer Service Manual
1. Remove the Reference housing and electrodes set from the analyzer.
2. Remove the analyzer rear cover.
3. Disconnect the channel cables from the Pre Amplifier board.
4. Remove four mounting screws and remove the plate.
5. Install the Pogo plate/Electrode pin plate by reversing the above steps.
WARNING: The Power must be turned off and the Power Adapter disconnected
before changing the Printer assembly.
33
ST-200CC Arterial Blood Gas Electrolyte Analyzer Service Manual
Lock
1. Remove Pump tube from the bottom of the Reference Housing & waste
pipe.
2. Remove the front cover, rollers and pump tubing from the pump module
base.
1. Clean the module base: It is fixed to the chassis through a lock at the top
of base as shown in the figure. We need to clean the module base and
then apply the grease to inner surroundings softly.
2. Apply grease to rollers: Take a four rollers set and apply grease gently to
rollers. The rollers set consists a grove at back side and a knob on the
front side. Insert grove side of the rollers set to the module base knob.
3. Apply grease to new pump tubing: Take a new pump tubing and apply
grease all around it and then insert it as per the direction by holding the
roller firmly as shown in the figure.
4. Module cover: Close the module with front cover. So that rollers and
pump tubing will lock. Ensure that the tag of pump tubing should be
above the front cover lock.
34
ST-200CC Arterial Blood Gas Electrolyte Analyzer Service Manual
35
ST-200CC Arterial Blood Gas Electrolyte Analyzer Service Manual
36
ST-200CC Arterial Blood Gas Electrolyte Analyzer Service Manual
37
ST-200CC Arterial Blood Gas Electrolyte Analyzer Service Manual
Note:
Improper alignment of probe seal and seal holder may cause flow errors.
38
ST-200CC Arterial Blood Gas Electrolyte Analyzer Service Manual
39
ST-200CC Arterial Blood Gas Electrolyte Analyzer Service Manual
4. Troubleshooting
This chapter assumes the reader is familiar with the operation of the Analyzer
and has read the user manuals.
Serum/Whole blood:
Parameter Measurement Range Reproducibility (CV%)
Na+ 20 - 250 mmol/L <1.0 %
K+ 0.20 - 40.0 mmol/L <2.0%
Cl- 25 - 200 mmol/L <1.0%
iCa++ 1.0 - 20.0 mmol/L <2.0%
Li+ 0.2 - 5 mmol/L <2.0%
pH 6-8 ± 0.03
pO2 10 - 750 mmHg <3.0%
pCO2 5 - 120 mmHg <3.0 %
TCO2 5 - 50 mmol <3.0%
Hct 10 - 70 % <3.0%
HCO3- 5 - 50 mmol <3.0%
Urine:
Parameter Measurement Range Reproducibility (CV%)
Na+ 25 - 1000 mmol/L <5.0 %
K+ 1 - 500 mmol/L <5.0%
Cl- 25 - 500 mmol/L <5.0%
40
ST-200CC Arterial Blood Gas Electrolyte Analyzer Service Manual
The standard electrical potential (E′) and slope (S) are determined by
measuring the electrical potentials of the ion-selective electrode in two
calibration solutions that have known concentrations of the measuring ions at
different levels. This process is called two-point calibration. Once the E′ and S are
determined, the unknown concentration of a sample can be determined by
measuring the electric potential of the electrode in a sample.
Definition of pCO2
The partial pressure (tension) of carbon dioxide in solution is defined as
the partial pressure of carbon dioxide in the gas phase in equilibrium with the
blood.
41
ST-200CC Arterial Blood Gas Electrolyte Analyzer Service Manual
The pH of this internal solution varies with the pCO2 according to the
Henderson-Hasselbalch equation.
42
ST-200CC Arterial Blood Gas Electrolyte Analyzer Service Manual
4.2.1.5 Glucose
Glucose measurement is based on the level of H2O2 produced during the
enzymatic reaction between glucose and oxygen molecules in the presence of
the glucose oxidase enzyme. The reaction is described by the following
equation:
The current generated by the flow of electrons at the surface of the strip is
proportional to the glucose concentration of the sample.
4.2.1.6 Lactate
Lactate measurement is based on the level of H2O2 produced during the
enzymatic reaction between lactate and oxygen molecules in the presence of
the lactate oxidase enzyme.
The reaction is described by the following equation:
Lactate O2 LactateOxi
dase
Pyruvate Acid H 2O2
Equation 6
The current generated by the flow of electrons at the surface of the strip is
proportional to the lactate concentration of the sample.
43
ST-200CC Arterial Blood Gas Electrolyte Analyzer Service Manual
U
5.49 1011 Y 0.071
T 37
9.72 10 9
Y 2.30
Y= e[3.88 x In(pO2)]
pH pK log
HCO3
PCO2 Equation 11
where pH and pCO2 are measured.
pK = 6.091
α = 0.0307 = solubility coefficient of CO2 in plasma at 37 °C
CAUTION:
The ST-200 CC ABG Electrolyte Analyzer provides an estimation of
hemoglobin only from normal hematocrit values citing the specific normal
adult male/female range. In cases of abnormal blood composition, e.g., red
cell dyscrasia or hemoglobinopathies or in cases of disease states, e.g.,
anemia, repeat testing by conventional laboratory methods is indicated.
44
ST-200CC Arterial Blood Gas Electrolyte Analyzer Service Manual
NOTE:
The hemoglobin calculation is an estimation based on a normal mean
corpuscular hemoglobin concentration of 33.3% and a nominal male Hct of 39 to
49% or female Hct of 35 to 45%. Hemoglobin estimations made from samples
with Red cell dyscrasia or hemoglobinopathies may vary significantly from
hemoglobin measured by cyanmethemoglobin method. The estimated
hemoglobin may vary significantly in cases of abnormal blood composition or
disease states such as anemia in which abnormal values may not be reported.
These conditions should warrant repeat testing by conventional laboratory
methods.
45
ST-200CC Arterial Blood Gas Electrolyte Analyzer Service Manual
NOTE:
The equation for calculating oxygen saturation assumes a normal shape
and position of the patient's oxygen dissociation curve.
A
% FIO2
B.P. 0.045T 2 0.84T 16.5 * * PCO2 % FIO2 1 % FIO2 / 100
100 100 0 .8
Equation 21
Where T = patient temperature
B.P. = barometric pressure
%FIO2 = fraction inspired oxygen, as a percent
** Temperature corrected gas value
46
ST-200CC Arterial Blood Gas Electrolyte Analyzer Service Manual
NOTE:
For capillary samples, AaDO2 results have an asterisk (*). AaDO2 results are
dependent on how the samples are drawn and handled, thus care must be
taken when interpreting these calculated results.
47
ST-200CC Arterial Blood Gas Electrolyte Analyzer Service Manual
used for measured values between pH 7.2 and 7.6. Between pH 6.9 and 7.2 and
between pH 7.6 and 8.0, modified forms of the equation are used. Normalized
ionized calcium values for samples with pH outside the range pH 6.9 to pH 8.0 are
not displayed.
Depending upon your requirement you can change settings. Once settings
have been changed it will not change until you made any changes or 3 VDC
battery is removed on motherboard.
CAUTION: Care should be taken while using a syringe to flush the flow
path. A syringe may exert enough force to rupture a membrane or break
the seal between flow through sensors, allowing fluids to splash onto the
surfaces of the flow cell components.
In troubleshooting Flow Path Error is the main error mainly it will cause due to
Protein deposition.
48
ST-200CC Arterial Blood Gas Electrolyte Analyzer Service Manual
1) Electrode housing
2) Bubble Detector
3) Sampler i.e. probe mechanism
4) Reagent pack &
5) Valve module.
Perform priming if you found that liquid is not coming then check Reagent
Pack consumption.
If it shows Solution left as 0%, then change the pack. Make a note that
sometimes if Solution left shows <15% also then there will be a chance of empty
Reagent Pack.
Check the LED on the Ultrasonic Bubble Detector. It will glow when the liquid
is passed through the bubble detector. Refer topic 2.6.7 (Page no: 20)
49
ST-200CC Arterial Blood Gas Electrolyte Analyzer Service Manual
Remove Sample tube from Bubble detector & remove the Pump tube from
the bottom of Reference housing and directly connect Sample tube to pump
tube i.e we Bypassed Reference Housing.
If liquid is coming out then remove electrode housing and remove the stack
of electrodes along with sample detector from the reference housing and
flush it with distilled water 2 times or daily cleaner gently. Don’t forget to
flush air several times after distilled water or daily cleaner.
Insert all stacked electrodes into the reference housing after cleaning inner
silicon seals individually and repeats above procedure with air flush. Connect
the electrode housing in the instrument.
If still liquid is not coming then check the sampler seal by doing below procedure.
Still if you are not able to see the liquid in sample tube then clean the Valve
module tubing by using VALVE TEST option in menu.
50
ST-200CC Arterial Blood Gas Electrolyte Analyzer Service Manual
First remove CAL-A, CAL-B & CAL-C tubes from thermal system.
Take out whole valve tubing from valves (V1, V2 & V3) slots.
Take syringe with DI water.
Flush whole tubing with DI water in order to clear all the salt formation in
flow path.
Finally flush whole tubing with air and then reinsert in respective positions at
valves flow path tube slots.
Note:
Pinch valve tubing should be totally inserted in the valve towards inner side.
Partial or incomplete inserted tube can let solutions mix up.
B
C
The valves should control standards flow by using the close or open
movements of lock pins.
If getting standards milli volts equally or ‘Air in CAL-B’ error then do the
VALVE TEST by using below steps.
By performing this test, we can identify the salt blockage in the tubing and
also we can identify the functioning of individual valves.
Note:
At an ideal state A (V1), Air (V2) and output of V2-Air (V3) slots are in
open condition and obviously B (V1), A/B (V2) & C (V3) slots are in close
condition.
51
ST-200CC Arterial Blood Gas Electrolyte Analyzer Service Manual
CAL-A flow:
Check Cal A flow through valves as per its respective functionality. It activates
Cal A flow and provides 300 seconds time duration to check its operation. When
the time gets completed, it deactivates automatically or the user can off it by
press the Stop button.
52
ST-200CC Arterial Blood Gas Electrolyte Analyzer Service Manual
the reasons are either may valves V1 or V2 are not functioning or there is a
blockage in CAL-A flow path. Hence, flush the tubing with DI water, Air and
check the pinch valves connectors (heater 1, heater 2 & heater 3) on Motor
driver card.
CAL-B flow:
Check Cal B flow through valves as per its respective functionality. It activates
Cal B flow and provides 300 seconds time duration to check its operation. When
the time gets completed, it deactivates automatically or the user can off it by
press the Stop button.
53
ST-200CC Arterial Blood Gas Electrolyte Analyzer Service Manual
CAL-A/B IN flow:
Check Cal A/B IN flow through valves as per its respective functionality. It
activates Cal A/B IN flow and provides 300 seconds time duration to check its
operation. When the time gets completed, it deactivates automatically or the
user can off it by press the Stop button.
Air flow:
Check Air flow through valves as per its respective functionality. It activates
Air flow and provides 300 seconds time duration to check its operation. When the
time gets completed, it deactivates automatically or the user can off it by press
the Stop button.
54
ST-200CC Arterial Blood Gas Electrolyte Analyzer Service Manual
Observe the lock pins movements of V1, V2 & V3. There is no movements
and are in ideal states.
CAL-A/B/Air flow:
Check Cal A/B/Air flow through valves as per its respective functionality. It
activates Cal A/B/Air flow and provides 300 seconds time duration to check its
operation. When the time gets completed, it deactivates automatically or the
user can off it by press the Stop button.
55
ST-200CC Arterial Blood Gas Electrolyte Analyzer Service Manual
Observe the lock pins movements of V1, V2 & V3. There is no movements and are
in ideal states.
CAL-C Flow:
Check Cal C flow through valves as per its respective functionality. It activates
Cal C flow and provides 300 seconds time duration to check its operation. When
the time gets completed, it deactivates automatically or the user can off it by
press the Stop button.
Observe the lock pin movement of V3.
V3 lock closes the output of V2 and opens CAL-C slot.
56
ST-200CC Arterial Blood Gas Electrolyte Analyzer Service Manual
If the CAL-C solution does not comes out and syringe goes reverse flow then
the reason is valve V3 is not functioning or there is a blockage in CAL-C flow
path. Hence, flush the tubing with DI water, Air and check the pinch valve
connector (heater 3) on Motor driver card.
Caution:
Do not use the spring needle or plain needle to clean the deposited protein
from electrode blocks as it contains different parameters.
After doing every step please calibrate the instrument two times with print
mV is ON.
57
ST-200CC Arterial Blood Gas Electrolyte Analyzer Service Manual
5. Error messages:
If you are getting all Cal B, Cal C & Cal A mV’s as ± 460 mV’s or equal mV’s Error
then look at left down corner of the reference slightly so that you can see the
reference hole having any bubbles. If bubbles are present in the pin hole,
knock it or slightly shake up side down the reference to remove the bubbles
from the hole. If bubbles are completely removed then calibrate the
instrument it will work.
Also check any salt coming from silver pin side. If yes clean it & try to place
the reference in the instrument & check mV’s.
See reference IFS (Inner Filling Solution) is leak from any side. If yes don't use it.
After trying all the above steps and then check reference Pogo pin. And make
sure that it should be tight.
Then also problem remain same, then Reinsert Pre amplifier board on the
mother board, and make sure that no loose connections on the Pre amplifier
board. Reinsert the connections on the Pre amp.
58
ST-200CC Arterial Blood Gas Electrolyte Analyzer Service Manual
59
ST-200CC Arterial Blood Gas Electrolyte Analyzer Service Manual
60
ST-200CC Arterial Blood Gas Electrolyte Analyzer Service Manual
6. SOFTWARE UP-DATE:
STEP 1: Download and install FET-Pro430 Lite Software from below link. Flash
Programmers for TI's MSP430 MCU uses TI's FET Adapter.
http://www.elprotronic.com/download.html
STEP 3: Now connect JTAG 14 PIN male connector to motherboard female pin and
USB connection to PC (Computer) respectively.
STEP 4: After connecting the device, switch on the instrument and open the FETPRO
Lite application by double-clicking on the shortcut key. Ensure to observe the red &
green led glows.
STEP 5: After opening the application you need to click on open source code sub
menu and browse the path for .HEX file provided by us.
STEP 6: Select respective controller GROUP (MSP430F6xx) and the controller specified
MSP430F6779.
STEP 7: Now click on the Verify security fuse and wait for 5-6 seconds. You will find
the verify security fuse as passed. Then, click on the auto program (AUTO PROG.)
menu and wait for 2-3 minutes. After finishing the software update it will show
passed and completed syntax.
61
ST-200CC Arterial Blood Gas Electrolyte Analyzer Service Manual
62
ST-200CC Arterial Blood Gas Electrolyte Analyzer Service Manual
63
ST-200CC Arterial Blood Gas Electrolyte Analyzer Service Manual
64
Manufactured by:
Sensa Core Medical Instrumenta on Pvt. Ltd.
Plot No.3, Export Promo on Industrial Park, Pashamylaram,
Sangareddy (Dist), Hyderabad- 502307, Telangana, INDIA.
www.sensacore.com, Tel: +08455 223400.
service@sensacore.com