Professional Documents
Culture Documents
Operation Guide
www.delta-c.com
I032r01 (Water Cut Monitor Operation Guide) Page 1 of 37
Table of Contents
Section 1 – Introduction ……………………………………………………………………………… 4
1.1 Capacitance Measurement 4
1.2 Field Calibration 4
1.3 Multi-Curve Calibration 4
1.4 Sample Collection 4
1.5 Sample Analysis 5
1.6 Communication 5
1.7 Hazardous Areas Classification 5
1.8 General Specifications 5
1.9 Referenced Industry Standards 5
www.delta-c.com
I032r01 (Water Cut Monitor Operation Guide) Page 2 of 37
4.4 Alarm Relay Configuration 10
4.4.1 Alarm Pick-up 10
4.4.2 Alarm Drop-out 10
4.4.3 Alarm Delay 10
4.5 Measurement Units 10
4.5.1 Set Temperature Units 10
4.5.2 Set Water Volume (BS&W) Units 10
4.6 Measurement Output Averaging 10
4.6.1 Set Water Volume (BS&W) Averaging 10
4.6.2 Set Analog Output Averaging 11
4.6.3 Set Temperature Averaging 11
Section 6 – Appendices………………………………………………………………………………. 29
6.1 General Specifications 29
6.2 Wiring Diagram 30
6.3 RS-485 Modbus Configuration 31
6.4 V200UI Download & Installation 34
6.5 Product Markings 36
6.6 Industry Standards 37
www.delta-c.com
I032r01 (Water Cut Monitor Operation Guide) Page 3 of 37
SECTION 1 - INTRODUCTION
The Delta C Water Cut Monitor is an online capacitance measurement device which monitors
changes in the water content present in crude oils, condensates and refined petroleum fluids.
This guide will explain the basic principles of operation, communication, configuration and field
calibration for all current models of the Delta C Water Cut Monitor.
Please Note:
This Operation Guide is intended to be read in conjunction with the V200 User Interface Guide.
The following introductory points explain key considerations and product features:
Note: Delta C factory (bench scale) calibration is only possible if the fluid to be measured is
commercially manufactured to specification and made available to the Delta C factory.
www.delta-c.com
I032r01 (Water Cut Monitor Operation Guide) Page 4 of 37
1.5 Sample Analysis
Analysis of samples drawn during field calibration is used to establish the baseline capacitance of
the process fluid at any given water content. Therefore, accurate and repeatable sample analysis
is critical for reliable field calibration. There are several sample analysis methods recommended
by industry groups. Such methods include centrifuge, titration or distillation. The user must select
the analysis method that is most appropriate for the end user’s range and accuracy requirements.
Please refer to Appendix 6.6 for a list of sample analysis standards.
1.6 Communication
The Delta C will report online water volume using a number of output signals. Outputs signals
include High Water Alarm Relay, 4-20mA Analog Output, RS-232 Serial and RS-485 Modbus.
The end user may elect to integrate as many or as few output signals as desired and all outputs
listed are supplied as standard with the Delta C Water Cut Monitor.
Please note that Delta C Technologies does not re-distribute published industry standards due to
copyright restrictions. The user may obtain the referenced standards directly from the publisher or
their authorized distributors.
www.delta-c.com
I032r01 (Water Cut Monitor Operation Guide) Page 5 of 37
SECTION 2 - OPERATING CONSIDERATIONS
The Delta C Water Cut Monitor is an extremely flexible device offering water in oil measurement
under a variety of process conditions. For optimum results, the user must be aware of certain
operating considerations that may influence the overall calibration strategy and measurement
performance.
www.delta-c.com
I032r01 (Water Cut Monitor Operation Guide) Page 6 of 37
2.5 Fluid Density
The characteristic of fluid density is commonly used by industry stakeholders to distinguish
between two or more fluid types or compositions. While distinct fluid densities may equate to
distinct fluid dielectric values, the relationship between density and dielectric is not always
predictable. In fact, two fluids of similar density may exhibit different dielectric properties, while in
other cases the opposite may be true.
For this reason, the user should consider fluid density to be a potentially useful property for
automated calibration curve selection, but caution must be exercised as density may not always
be used for this purpose.
www.delta-c.com
I032r01 (Water Cut Monitor Operation Guide) Page 7 of 37
SECTION 3 – COMMUNICATIONS OVERVIEW
The Delta C Water Cut Monitor is designed to support several methods of communication. The
user may choose to utilize any one or combination of the signals as required.
Note: Previous to the release of the V200UI, the communication application “HyperTerminal”
(bundled with Windows 95 to XP operating systems) was the standard Delta C user interface tool.
As HyperTerminal has been superseded by the V200UI, HyperTerminal communication will not be
addressed in this guide.
www.delta-c.com
I032r01 (Water Cut Monitor Operation Guide) Page 8 of 37
SECTION 4 – BASIC CONFIGURATION
The user must select the appropriate COM port and press the Save button in the Configuration
Toolbar. Once the correct COM Port has been selected the V200 User Interface will begin
communication with the Water Cut Monitor. The COM Port setting is stored in the V200 User
Interface on the user’s laptop computer. Once the correct COM Port is selected, it should not
need to be set again, regardless of which Water Cut Monitor the user wishes to communicate
with.
The remaining COM Port values displayed in the Configuration Window (Baud Rate, Parity, Data
Bits and Stop Bits) are not user selectable and cannot be manipulated.
Operational parameters that could cause the instrument to stop responding to the master requests
are not permitted via Modbus commands but are modified via the RS-232 terminal access. This
includes such items as Modbus Mode, Baud Rate and Asynchronous Character parameters.
There is also a restriction on the character length so that if the RTU mode is chosen, the character
length cannot be set to seven bits. Physical access is via a two-wire RS-485 port.
4.3.1 Set Zero Scale (4mA) permits the operator to set the water volume at which the
Delta C will send a 4mA output signal. This is normally set to 0, or the minimum water
content anticipated in the process.
www.delta-c.com
I032r01 (Water Cut Monitor Operation Guide) Page 9 of 37
4.3.2 Set Full Scale (20mA) permits the operator to set the water volume at which the
Delta C will send a 20mA output signal. This is normally set to represent the maximum
water content anticipated in the process.
4.3.3 Trim Zero Scale permits the operator to trim the 4mA output to match a local display,
with a lower limit of 0mA.
4.3.4 Trim Full Scale permits the operator to trim the 20mA output to match a local display,
with an upper limit of approximately 24mA.
4.4.1 Alarm Pick-up defines the maximum water content permitted before the alarm is
triggered. If the water content measured by the Water Cut Monitor exceeds the alarm pick-
up value, the alarm relay will be de-energized (closed).
4.4.2 Alarm Drop-out defines the minimum water content required to re-energize (open)
the alarm relay.
4.4.3 Alarm Delay defines the number of seconds the water content is permitted to remain
above the alarm pick-up value before the alarm relay will de-energized (closed). The user
may enter any value in one second increments to avoid tripping the alarm on transient
conditions.
4.5.1 Set Temperature Units to display process fluid temperature in either ˚F or ˚C.
4.5.2 Set Water Volume (BS&W Units) to display water volume in either % or ppm.
4.6.1 Set Water Volume (BS&W) Averaging instructs the Delta C to output the water
volume as a running average. For example, if the BS&W Average value is set to 10 the
water volume output will be a calculated average of the last ten measurements. The
maximum averaging value permitted is 64. The normal factory setting for this value is 5.
www.delta-c.com
I032r01 (Water Cut Monitor Operation Guide) Page 10 of 37
4.6.2 Set Analog Output Averaging instructs the Delta C to output the analog output as a
running average, similar to the BS&W average. The maximum averaging value permitted
is 64. The normal factory setting for this value is 5.
4.6.3 Set Temperature Averaging instructs the Delta C to output the temperature output
as a running average, similar to the BS&W average. The maximum averaging value
permitted is 256. The normal factory setting for this value is 25.
Field calibration of the Delta C Water Cut Monitor is required to define the baseline dielectric and
related capacitance of the process fluid being measured in the field. Prior to attempting field
calibration, the user must assess the available field resources such as available sampling and
analysis methods, as well as process control capabilities for fluid temperature and water volume.
Once determined, the field technician must work with the end user to identify achievable
calibration and measurement objectives given the resources and capabilities available.
The field calibration technician is encouraged to review the following four calibration methods with
the end user in order to select the method that is most appropriate for the end user’s application
and measurement requirements. While doing so, some basic questions will help guide the
method selection. These questions are:
1. Is it possible to control the water content of the fluid circulating through the Delta C
during calibration such that a “dry” or 0% water condition can be achieved?
This condition would be required to perform Calculated Temperature Compensation,
(described in section 5.2 below) which is a requirement of the Advanced calibration
methods.
2. Is it possible to control the temperature of the fluid circulating through the Delta C so
that the temperature may be raised or lowered by a minimum of 25˚C?
This condition is also required to perform Calculated Temperature Compensation
required for the Advanced calibration methods.
3. Is it possible to control the water content of the fluid circulating through the Delta C
during calibration such that more than one distinct water contents may be maintained
and observed?
This condition is required to perform either the Two Point or Three Point calibration
methods.
www.delta-c.com
I032r01 (Water Cut Monitor Operation Guide) Page 11 of 37
5.1.1 Single Point (CS) Basic Method requires a single fluid sample point. The sample
point is assigned a value of capacitance, water and temperature. The relationships
between changes in water (slope) and temperature (TC) versus changes in capacitance
are estimated and defined by the user. As capacitance changes with process conditions,
water content is calculated by the Delta C and output to the user based on the estimated
slope and TC. This method is commonly used for measurement where the water volume
ranges from 0 to 2%.
www.delta-c.com
I032r01 (Water Cut Monitor Operation Guide) Page 12 of 37
5.1.2 Two Point (CF) Basic Method requires two distinct fluid sample points. The sample
points are assigned values of capacitance, water and temperature. The relationship
between changes in water (slope) and capacitance is calculated by the Delta C using the
two sample points. The relationship between changes in temperature (TC) and
capacitance is estimated and defined by the user. As capacitance changes with process
conditions, water content is calculated by the Delta C and output to the user based on the
calculated slope and estimated TC. This method is commonly used for measurement
where the water volume ranges from 0 to 2% and where greater slope accuracy is desired.
Limitations: Two sample points with different water contents are required.
www.delta-c.com
I032r01 (Water Cut Monitor Operation Guide) Page 13 of 37
5.1.3 Two Point (CN) Advanced Method requires two distinct fluid sample points, one of
which must be dry (containing 0% water), and the ability to manipulate the temperature of
the dry fluid sample by 25˚C as it circulates through the Delta C measurement probe. The
sample points are assigned values of capacitance, water and temperature. The
relationship between changes in water (slope) and temperature (TC) versus changes in
capacitance are calculated by the Delta C. As capacitance changes with process
conditions, water content is calculated and output to the user based on the calculated slope
and TC. This method is commonly used for measurement where the water volume ranges
from 0 to 10% and where greater slope and TC accuracy are desired.
Limitations: Two distinct fluid samples required (one dry, one wet). Temperature
control required.
www.delta-c.com
I032r01 (Water Cut Monitor Operation Guide) Page 14 of 37
5.1.4 Three Point (CW) Advanced Method requires three distinct fluid sample points, one
of which must be dry (containing 0% water), and the ability to manipulate the temperature
of the dry fluid sample by 25˚C as it circulates through the Delta C measurement probe.
The sample points are assigned values of capacitance, water and temperature. The
relationship between changes in water (curve) and temperature (TC) versus changes in
capacitance are calculated by the Delta C. As capacitance changes with process
conditions, water content is calculated and output to the user based on the calculated curve
and TC. This method is commonly used for measurement where the water volume ranges
from 0 to 50% and where greater curve and TC accuracy are desired.
Limitations: Three distinct fluid samples required (one dry, two wet). Temperature
control required.
www.delta-c.com
I032r01 (Water Cut Monitor Operation Guide) Page 15 of 37
5.2 Temperature Compensation (TC)
Changes in fluid temperature influence fluid dielectric values and related capacitance, therefore,
temperature compensation is required to ensure water volume output from the Delta C is
temperature corrected. Temperature compensation is either Estimated or Calibrated and is
dictated by the calibration procedure chosen by the user as described in Section 5.1 above.
5.2.1 Estimated TC is used with the calibration methods CS and CF. The estimated TC
value is related to fluid density and measurement units. Based on Delta C’s experience,
the following (TC) values may be used to provide reasonable temperature compensation
where the process fluid temperature does not vary significantly:
The TC values above are entered in the Configuration Window of the V200UI prior to
performing the One Point and Two Point Basic Calibration procedures.
5.2.2 Calculated TC is used with the Two Point and Three Point Advanced Calibration
methods. This requires heating a dry fluid sample (containing 0% water) as it circulates
through the Delta C, and defining the capacitance of the fluid at two temperature points a
minimum of 25°C apart. During the procedure, the Delta C will calculate the temperature
compensation values and store them in the data set of the current calibration curve
number. To perform Calculated Temperature Compensation, proceed as follows:
1. Provide a means to circulate dry oil through the Delta C measurement probe and a
means to heat the oil and begin circulation of the oil at ambient temperature.
2. From the V200UI Production Window Data Display, monitor stability of the R Value
(capacitance) and temperature. When the capacitance and temperature are stable,
temperature point #1 will be defined using the V200UI ASCII Window as follows:
At the prompt:
[1 ] Command (? for help) >
Enter CC
The Delta C will ask:
Measure Point #1 now (Y/N)?
Enter Y
The Delta C will ask:
Measure Point #2 now (Y/N)?
Enter N
The Delta C will ask:
Done (Y/N/esc)?
Enter Y
The capacitance of temperature point #1 is now defined in the Delta C.
www.delta-c.com
I032r01 (Water Cut Monitor Operation Guide) Page 16 of 37
3. Apply heat to raise the fluid temperature by 25˚C or more and wait until the temperature
and capacitance have stabilized as observed in the V200UI Production Window. A fast
and steady fluid circulation in a closed loop may be used to increase the temperature.
When the capacitance and temperature are stable, temperature point #2 will be defined
using the V200UI ASCII Window as follows:
At the prompt:
[1 ] Command (? for help) >
Enter CC
The Delta C will ask:
Measure Point #1 now (Y/N)?
Enter N
The Delta C will ask:
Measure Point #2 now (Y/N)?
Enter Y
The Delta C will ask:
Done (Y/N/esc)?
Enter Y
The capacitance of temperature point # 2 is now defined in the Delta C and
the Calculated Temperature Compensation procedure is complete.
5.3 Slope
The Slope defines the linear relationship between changes in water content and changes in
capacitance. While it is generally more accurate to calculate the slope using two distinct sample
points, the One Point Basics Calibration method permits the user to enter an estimated slope,
which may provide reasonable accuracy within a relatively narrow measurement range. The
following slopes have been found to be generally acceptable by many users:
The Slopes above are entered in the Configuration Window of the V200UI prior to performing the
One Point Basic Calibration procedure.
www.delta-c.com
I032r01 (Water Cut Monitor Operation Guide) Page 17 of 37
5.4.1 Single Point (CS) Basic Calibration
Before beginning the Single Point (CS) Basic Calibration, refer to Sections 5.2.1 and 5.3
and enter suitable Temperature Compensation and Slope values in the appropriate fields in
the V200UI Configuration Window, then proceed as follows:
www.delta-c.com
I032r01 (Water Cut Monitor Operation Guide) Page 18 of 37
4. Prepare to draw a physical fluid sample from the pipeline. When ready, begin drawing
the sample and click Drawing Sample Now.
5. Accept and note the system generated calibration name, or rename the calibration as
desired. Click Continue.
6. If the sample can be analyzed immediately on site, or if the water content is already
known, select The water content is known and click Continue. If the sample must be
analyzed off-site proceed directly to step 9 below.
www.delta-c.com
I032r01 (Water Cut Monitor Operation Guide) Page 19 of 37
7. Enter the water content of the sample and click Continue.
8. At the One Point Calibration Successful window click Exit. The One Point Calibration
procedure is now complete.
9. If the sample must be analyzed off-site select The sample needs to be analyzed and
click Continue.
www.delta-c.com
I032r01 (Water Cut Monitor Operation Guide) Page 20 of 37
10. Click Exit to complete the first phase of the One Point Calibration.
11. Once the sample has been analyzed and the water content is known, re-connect to the
Water Cut Monitor, launch the V200UI and from the Production Window toolbar click
One Point Calibration. At the introduction window click Continue.
13. Compare the Reference Point name to the Calibration name saved in step 5 above. If
the two names match click Continue. If the two names do not match click Browse to
retrieve the appropriate calibration file, then click Continue.
www.delta-c.com
I032r01 (Water Cut Monitor Operation Guide) Page 21 of 37
14. Enter the water content of the sample and click Continue then click Exit at the One
Point Calibration Successful window. This completes the One Point (CS) Calibration
procedure.
1. With a fluid mixture circulating through the Delta C measurement probe, observe the
process for stability using the V200UI Production Window. This mixture will be referred
to as sample fluid #1. When the process is stable, define the capacitance of sample
point #1 in the Delta C through the V200UI ASCII Window as follows:
At the prompt:
[1 ] Command (? for help) >
Enter CF
The Delta C will ask:
Measure point #1 now (Y/N)?
Enter Y
Immediately draw a physical sample of fluid #1 for analysis.
The Delta C will ask:
Water Content #1 (%)?
Hit the enter key
The Delta C will ask:
Measure point #2 now (Y/N)?
Enter N
The Delta C will ask:
Water Content #2 (%)?
Hit the enter key
The Delta C will ask:
Temperature Compensation (%/C, SIGN REQUIRED!)?
www.delta-c.com
I032r01 (Water Cut Monitor Operation Guide) Page 22 of 37
Enter a TC Value as determined in Section 5.2.1. A common TC value would
be -0.03 (note this is a negative value).
The Delta C will ask:
Done (Y/N/esc)?
Enter Y
The capacitance of sample point #1 has now been defined in the Delta C.
2. When the analysis of sample #1 is complete and the water content is known, define the
water content of sample point #1 in the Delta C through the V200UI ASCII Window as
follows:
At the prompt:
[1F] Command (? for help) >
Enter CF
The Delta C will ask:
Measure point #1 now (Y/N)?
Enter N
The Delta C will ask:
Water Content #1 (%)?
Enter the analyzed water content of sample #1
The Delta C will ask:
Measure point #2 now (Y/N)?
Enter N
The Delta C will ask:
Water Content #2 (%)?
Hit the enter key
The Delta C will ask:
Temperature Compensation (%/C, SIGN REQUIRED!)?
Hit the enter key
The Delta C will ask:
Done (Y/N/esc)?
Enter Y
The water content of sample point #1 has now been defined in the Delta C.
3. With a fluid mixture circulating through the Delta C measurement probe which is
distinctly different in water content from fluid sample #1, observe the process for stability
using the V200UI Production Window. This mixture will be referred to as sample fluid
#2. When the process is stable, define the capacitance of sample point #2 in the Delta
C through the V200UI ASCII Window as follows:
At the prompt:
[1F] Command (? for help) >
Enter CF
The Delta C will ask:
Measure point #1 now (Y/N)?
Enter N
www.delta-c.com
I032r01 (Water Cut Monitor Operation Guide) Page 23 of 37
The Delta C will ask:
Water Content #1 (%)?
Hit the enter key
The Delta C will ask:
Measure point #2 now (Y/N)?
Enter Y
Immediately draw a physical sample of fluid #2 for analysis.
The Delta C will ask:
Water Content #2 (%)?
Hit the enter key
The Delta C will ask:
Temperature Compensation (%/C, SIGN REQUIRED!)?
Hit the enter key
The Delta C will ask:
Done (Y/N/esc)?
Enter Y
The capacitance of sample point #2 has now been defined in the Delta C.
4. When the analysis of sample #2 is complete and the water content is known, define the
water content of sample point #2 in the Delta C through the V200UI ASCII Window as
follows:
At the prompt:
[1F] Command (? for help) >
Enter CF
The Delta C will ask:
Measure point #1 now (Y/N)?
Enter N
The Delta C will ask:
Water Content #1 (%)?
Hit the enter key
The Delta C will ask:
Measure point #2 now (Y/N)?
Enter N
The Delta C will ask:
Water Content #2 (%)?
Enter the analyzed water content of sample #2
The Delta C will ask:
Temperature Compensation (%/C, SIGN REQUIRED!)?
Hit the enter key
The Delta C will ask:
Done (Y/N/esc)?
Enter Y
The water content of sample point #2 has now been defined in the Delta C
and the Two Point (CF) Calibration procedure is complete.
www.delta-c.com
I032r01 (Water Cut Monitor Operation Guide) Page 24 of 37
5.4.3 Two Point (CN) Advanced Calibration
Before beginning the Two Point (CN) Advanced Calibration, the user must complete the
Calculated Temperature Compensation procedure described in Section 5.2.2. By
performing the temperature compensation procedure, the user will have defined sample
point #1 as 0% water. The remaining steps below are required to define sample point #2
and are performed with the V200UI ASCII Window:
1. Since sample point #1 has already been defined, the next step is to define sample point
#2. A mixture of oil and water that is in approximately the highest range of the
anticipated span will be required. Circulate the mixture through the Delta C
measurement probe, observing the process for stability using the V200UI Production
Window.
2. When the process is stable, define the capacitance of sample point #2 in the Delta C
through the V200UI ASCII Window as follows:
At the prompt:
[1N ] Command (? for help) >
Enter CN
The Delta C will ask:
Measure point #2 now (Y/N)?
Enter Y
Immediately draw a physical sample of fluid #2 for analysis.
The Delta C will ask:
Water Content #2? (%)?
Hit the enter key
The Delta C will ask:
Done (Y/N/esc)?
Enter Y
The capacitance of sample point #2 has now been defined in the Delta C.
3. When the analysis of sample #2 is complete and the water content is known, define the
water content of sample point #2 in the Delta C through the V200UI ASCII Window as
follows:
At the prompt:
[1N ] Command (? for help) >
Enter CN
The Delta C will ask:
Measure point #2 now (Y/N)?
Enter N
The Delta C will ask:
Water Content #2? (%)?
Enter the analyzed water content of sample #2
The Delta C will ask:
Done (Y/N/esc)?
www.delta-c.com
I032r01 (Water Cut Monitor Operation Guide) Page 25 of 37
Enter Y
The water content of sample point #2 has now been defined in the Delta C and the
Two Point (CN) Calibration procedure is complete.
Note: If a full range calibration of 0 to 50% is desired, the recommended target water
content for sample points # 2 & 3 would be 20% and 40% respectively. Narrower spacing
of the sample points is permitted, however greatest accuracy will be achieved if the sample
points are evenly distributed across the desired measurement range. For example, if the
operating range is only expected to be 0 – 30%, then suggested sample points #2 & 3
would be 15% and 30% respectively.
1. Since sample point #1 has already been defined during the temperature compensation
procedure, the next step is to define sample point #2. A mixture of oil and water that is
approximately mid-range of the anticipated span will be required. Circulate the mixture
through the Delta C measurement probe, observing the process for stability using the
V200UI Production Window.
2. When the process is stable, define the capacitance of sample point #2 in the Delta C
through the V200UI ASCII Window as follows:
At the prompt:
[1W ] Command (? for help) >
Enter CW
The Delta C will ask:
Measure point #2 now (Y/N)?
Enter Y
Immediately draw a physical sample of fluid #2 for analysis.
The Delta C will ask:
Water Content #2? (%)?
Hit the enter key
The Delta C will ask:
Measure Point #3 now (Y/N)?
Enter N
The Delta C will ask:
Water Content #3? (%)?
Hit the enter key
www.delta-c.com
I032r01 (Water Cut Monitor Operation Guide) Page 26 of 37
The Delta C will ask:
Done (Y/N/esc)?
Enter Y
The capacitance of sample point #2 has now been defined in the Delta C.
3. When the analysis of sample #2 is complete and the water content is known, define the
water content of sample point #2 in the Delta C through the V200UI ASCII Window as
follows:
At the prompt:
[1W ] Command (? for help) >
Enter CW
The Delta C will ask:
Measure point #2 now (Y/N)?
Enter N
The Delta C will ask:
Water Content #2? (%)?
Enter the water content of sample #2
The Delta C will ask:
Measure Point #3 now (Y/N)?
Enter N
The Delta C will ask:
Water Content #3? (%)?
Hit the enter key
The Delta C will ask:
Done (Y/N/esc)?
Enter Y
The water content of sample point #2 has now been defined in the Delta C.
4. Since sample point #1 and #2 have now been defined, the next step is to define sample
point #3. A mixture of oil and water that is in approximately the highest range of the
anticipated span will be required. Circulate the mixture through the Delta C
measurement probe, observing the process for stability using the V200UI Production
Window.
5. When the process is stable, define the capacitance of sample point #3 in the Delta C
through the V200UI ASCII Window as follows:
At the prompt:
[1W ] Command (? for help) >
Enter CW
The Delta C will ask:
Measure point #2 now (Y/N)?
Enter N
www.delta-c.com
I032r01 (Water Cut Monitor Operation Guide) Page 27 of 37
The Delta C will ask:
Water Content #2? (%)?
Hit the enter key
The Delta C will ask:
Measure Point #3 now (Y/N)?
Enter Y
Immediately draw a physical sample of fluid # for analysis.
The Delta C will ask:
Water Content #3? (%)?
Hit the enter key
The Delta C will ask:
Done (Y/N/esc)?
Enter Y
The capacitance of sample point #3 has now been defined in the Delta C.
6. When the analysis of sample #3 is complete and the water content is known, define the
water content of sample point #3 in the Delta C through the V200UI ASCII Window as
follows:
At the prompt:
[1W ] Command (? for help) >
Enter CW
The Delta C will ask:
Measure point #2 now (Y/N)?
Enter N
The Delta C will ask:
Water Content #2? (%)?
Hit the enter key
The Delta C will ask:
Measure Point #3 now (Y/N)?
Enter N
The Delta C will ask:
Water Content #3? (%)?
Enter the water content of sample #3
The Delta C will ask:
Done (Y/N/esc)?
Enter Y
The water content of sample point #3 has now been defined in the Delta C
and the Three Point (CW) Calibration is complete.
www.delta-c.com
I032r01 (Water Cut Monitor Operation Guide) Page 28 of 37
SECTION 6 - APPENDICES
General
Operating Principal: advanced digital capacitance
Water in Oil Range: 0 to 50% standard
(50 to 80% optional)
Resolution: 0.001%
Repeatability: up to 0.01%
Precision/Accuracy: up to 0.01%
Temp. Variation: compensated
Temp. Stability: 0.0015% water/˚C
Temp. Comp. Range: 0˚C to 260˚C
Mechanical
Wetted Parts: 316SS standard
(Monel 400, Inconel 625 & Duplex 2205 optional)
O-ring Material: Viton A standard
(Nitrile, PTFE & Kalrez optional)
Max. Fluid Temp.: 135˚C standard
(260˚C optional)
Max. Ambient Temp.: 85˚C
Transmitter Mounting: integral with sensor
Corrosion Protection: NACE MR-0175/ISO 15156 2009, sour service compliant
Hazardous Area Class: CSA Class 1, Div.1, Gr. C&D (North America)
Ex II 2G, EEx d IIB T4, Zone1(Europe)
Ingress Protection: IP66 compliant
Electrical
Power Supply: 24 VDC @ 150mA floating
Electrical Connections: 1 x M20, 1 x 1” NPT & 1 x ¾” NPT
Humidity Protection: conformal coated CB
Analog: 4-20mA DC, isolated, self-powered
RS-232: 9600 baud, max. 100m
RS-485: 9600 baud, max. 1500m
Maximum Load: 600 ohm load to control room
Isolation Voltage: 500 volts peak
Relay Contacts: SPST-NO 2A @ 24 VDC, non-inductive
Adjustable Delay: one second increments
www.delta-c.com
I032r01 (Water Cut Monitor Operation Guide) Page 29 of 37
Appendix 6.2 – Wiring Diagram
www.delta-c.com
I032r01 (Water Cut Monitor Operation Guide) Page 30 of 37
Appendix 6.3 – RS-485 Modbus Configuration
This write process takes somewhat less than 250 msec. and we recommend that the
master be set for a timeout value of 500 msec., minimum.
The flash memory is capable of accepting approximately one million writes in its lifetime.
Writing should therefore be minimized to preserve flash memory.
6.3.3 Modbus configuration
The Modbus communication parameters are set up using the ASCII window in the V200
User Interface.
Begin by entering the command “CM” (for Configure Modbus) at the Command prompt.
The response is as follows and the example shows the factory default values:
At this point type in the desired slave address followed by a carriage return, next item
shown:
www.delta-c.com
I032r01 (Water Cut Monitor Operation Guide) Page 31 of 37
For this selection and the following, current value is marked with an asterisk, type in the
desired setting from the suggested choices or just carriage return if the current value is
acceptable.
The Modbus – Float Order indicates the order of the bytes that make up floating point
numbers. Normal means that the higher order byte is sent first and reverse means that the
lowest order byte is sent first, followed by the three remaining bytes.
Note: For users running BSW software version BSW200ar or higher, a response delay
prompt is included. The selection is as follows:
Modbus – TxDelay
TxDelay = 3
Set TxDelay (0..9)?
The numbers 0 – 9 represent the delay in character time, that is, 2 will result in a delay of
two character lengths. This delay is valid for both the ASCII and RTU mode.
Done (Y/N/esc)?
If settings are OK, type Y, otherwise N or carriage return to begin the settings procedure
again.
www.delta-c.com
I032r01 (Water Cut Monitor Operation Guide) Page 32 of 37
6.3.4 Modbus address assignments
Modbus
Name Data Data type Units R/W Command Value/units
Address
Water content 0000 Single float Eng R 03
Temperature 0002 Single float Eng R 03
R Value 0004 Single float Eng R 03
Density 0006 Single float Eng R 03 Kg/m3
Note: Only curves that are fully calibrated are allowed to be selected. If the master
attempts to select an un-calibrated curve, response will be a message indicating the current
curve. Also, a new un-calibrated analyzer will have curve 1 as default and will ask with data
even though it is un-calibrated.
Attempting writes of invalid values or addresses out of range will result in a response of an
appropriate error message.
As noted earlier, the flash memory is capable of accepting approximately one million writes
in its lifetime. Writing should therefore be minimized to preserve flash memory.
www.delta-c.com
I032r01 (Water Cut Monitor Operation Guide) Page 33 of 37
Appendix 6.4 – V200UI Download and Installation
The Delta C V200 User Interface software is generally shipped on a USB flash drive with each
order. The flash drive is shipped inside the Water Cut Monitor documentation binder. The
V200UI software is also available for download from the Delta C FTP site.
To install the V200UI application, simply copy the contents of the USB flash drive to your laptop
computer, open the installation folder V200UI-InstallPack (XPVistaWin7 x.x.x) and run the
application setup.exe as shown below.
When the installation is complete the Delta C V200UI application shortcut (shown below) will have
been added to the desktop of the laptop computer. You are now ready to communicate with the
Delta C via the V200 User interface.
If you do not have access to the USB Flash Drive that was shipped with the product, you may
download the V200UI installation files from the Delta C public FTP site.
To access the site, you must have an FTP Client such as FileZilla (http://filezilla-project.org/)
installed on your computer. Once you have installed the FTP Client software, you may then
access the Delta C FTP site using the following information.
The password and contents of the site may change without notice. To acquire a valid password
for login, or if you have any questions please contact support@delta-c.com.
www.delta-c.com
I032r01 (Water Cut Monitor Operation Guide) Page 34 of 37
Once you have connected to the FTP site, you will see two file folders similar to those shown
below:
Select and copy the entire folder named V200UI-InstallPack (XPVistaWin7 x.x.x) to the laptop.
This may take several minutes. Once download is complete, follow the installation instructions
above.
Note: If you are running Windows 2000, please contact Delta C for appropriate installation files.
www.delta-c.com
I032r01 (Water Cut Monitor Operation Guide) Page 35 of 37
Appendix 6.5 – Product Markings
www.delta-c.com
I032r01 (Water Cut Monitor Operation Guide) Page 36 of 37
Appendix 6.6 – Reference Standards for Measurement, Sampling and Analysis
www.delta-c.com
I032r01 (Water Cut Monitor Operation Guide) Page 37 of 37