Professional Documents
Culture Documents
Solvent mixtures
Method number 2
Application Air Analysis
Analytical principle Gas chromatography (activated charcoal/ternary
mixture) (Double capillary technique)
Completed in October 2013
Summary
This measurement method is suitable for monitoring the Occupational Exposure
Limits (OELs) of solvent components at workplaces. Sampling is performed by
drawing air through a tube containing activated charcoal using a suitable sampling
pump. Then the adsorbed solvents are extracted with a ternary mixture consisting
of dichloromethane, carbon disulphide and methanol. The sample solution is ana-
lysed by means of gas chromatography using two capillary columns of different po-
larity with the help of flame ionisation detectors (FID). Quantitative evaluation is
based on a multiple-point calibration with an internal standard.
Solvent mixtures
Method number 2
Application Air Analysis
Analytical principle Gas chromatography (activated charcoal/ternary
mixture) (Double capillary technique)
Completed in October 2013
Contents
The MAK Collection for Occupational Health and Safety 2016, Vol 1, No 3
10.1002/3527600418.am0lmix002e1816, Downloaded from https://onlinelibrary.wiley.com/doi/10.1002/3527600418.am0lmix002e1816 by Readcube (Labtiva Inc.), Wiley Online Library on [30/01/2024]. See the Terms and Conditions (https://onlinelibrary.wiley.com/terms-and-conditions) on Wiley Online Library for rules of use; OA articles are governed by the applicable Creative Commons License
Solvent mixture Meth.-No. 2 2141
1 General principles
A defined air volume from the breathing zone is drawn through a charcoal tube (B
type) by a sampling pump. The solvent components occurring in a gaseous state in
the workplace air are adsorbed onto the activated charcoal layer. Then the adsorbed
solvents are extracted with a ternary mixture consisting of dichloromethane, carbon
disulphide and methanol. The sample solution is analysed by means of gas chroma-
tography using two capillary columns of different polarity with the help of flame
ionisation detectors (FID). Quantitative evaluation is based on a multiple-point
calibration with an internal standard.
2.1 Equipment
• Charcoal tubes type B, collection layer of 300 mg, control layer of 600 mg (e.g.
from Dräger Safety, 23558 Lübeck, Germany)
• Pump, suitable for a flow rate of 0.333 L/min (e.g. model LFS-113DC, Gilian Inc,
USA)
• Gas chromatograph with two flame ionisation detectors (FID), autosampler,
split/splitless injector and data evaluation system
• Non-polar capillary column (e.g. Rtx-5: 60 m, 0.25 μm film thickness and 0.25
mm ID, from Restek, 61348 Bad Homburg, Germany)
• Polar capillary column (e.g. HP-Innowax: 60 m, 0.25 μm film thickness and 0.25
mm ID, from Agilent, 76337 Waldbronn, Germany)
• Flow meter (e.g. GFM model, Analyt-MTC, 79379 Müllheim, Germany)
• Screw-capped vials with septum, nominal volume 20 mL
• Dispensette, 1 to 10 mL, adjustable in increments of 0.2 mL
• Volumetric flasks, nominal volume 2, 5, 10, 100 and 1 000 mL
• Microlitre syringes, 5, 10, 25, 50, 100, 250, 500 and 1 000 μL
• Disposable syringes made of polypropylene (PP), 2 mL
• Disposable filter (e.g. PTFE syringe filter, pore size 0.45 μm)
• Autosampler vials, 1.5 mL with PP screw caps and PTFE-coated butyl rubber septa
2.2 Chemicals
The substances to be determined should have the highest possible purity (see also
Table 5 in the Appendix).
• 1,2,3-Trimethylbenzene, p.a., purity ≥ 93.7% (e.g. Fluka)
• n-Propylbenzene, p.a., purity ≥ 99.8% (e.g. Fluka)
2.3 Solutions
Desorption solution: Ternary mixture: CH2Cl2/CS2/MeOH (60/35/5)
The MAK Collection for Occupational Health and Safety 2016, Vol 1, No 3
10.1002/3527600418.am0lmix002e1816, Downloaded from https://onlinelibrary.wiley.com/doi/10.1002/3527600418.am0lmix002e1816 by Readcube (Labtiva Inc.), Wiley Online Library on [30/01/2024]. See the Terms and Conditions (https://onlinelibrary.wiley.com/terms-and-conditions) on Wiley Online Library for rules of use; OA articles are governed by the applicable Creative Commons License
Solvent mixture Meth.-No. 2 2143
tive substances into a 2 mL volumetric flask, into which approx. 500 μL of de-
sorption solution have already been placed. The volumetric flask is then filled to
the mark with desorption solution and shaken.
Table 2 Dosing scheme for the preparation of the calibration standards of the 6-point working
calibration and corresponding air concentrations*
tion range and it must have been prepared independently from the calibration solu-
tions described above.
Table 4 Dosing scheme for the calibration standards of the equidistant 10-point working cali-
bration and corresponding air concentrations
The MAK Collection for Occupational Health and Safety 2016, Vol 1, No 3
10.1002/3527600418.am0lmix002e1816, Downloaded from https://onlinelibrary.wiley.com/doi/10.1002/3527600418.am0lmix002e1816 by Readcube (Labtiva Inc.), Wiley Online Library on [30/01/2024]. See the Terms and Conditions (https://onlinelibrary.wiley.com/terms-and-conditions) on Wiley Online Library for rules of use; OA articles are governed by the applicable Creative Commons License
Solvent mixture Meth.-No. 2 2145
3.1 Sampling
A charcoal tube (B type) is opened and connected to a flow-regulated pump for
sampling. Following this, workplace air is drawn through the charcoal tube at a flow
rate of 20 L/h (0.333 L/min). It must be ensured that an air sample volume of 40 L
is not exceeded. If the sampling is to be carried out over a longer time period (up to
8 h), then the flow rate must be adjusted accordingly. Sampling can be carried out
at a relative humidity of up to 75% without any limitations; in the case of a greater
relative humidity two tubes connected in series are to be used. After sampling, the
loaded charcoal tubes are sealed with the plastic caps that are supplied with them
and unambiguously labelled. No breakthrough at concentrations of up to twice the
Occupational Exposure Limits occurs, if compliance with these parameters is en-
sured.
5 Analytical determination
1 μL of the sample solution is injected into the gas chromatograph in each case
using the autosampler, and split onto two columns of different polarity. The mean
value of the measurement data from both columns is used for the evaluation in
each case. If the deviation of the data from both separation columns is less than
20%, the mean value is used for the evaluation. In the case of a greator deviation, it
must be determined on the basis of different criteria, which value should be used
for the evaluation, e.g.:
• If the signal in a chromatogram of a substance to be determined is overlapped by
that of another substance, then the result from the other chromatogram is used
for the evaluation.
• If the signal of the internal standard in a chromatogram is overlapped by that of
another substance, then the results from the other chromatogram are used for
the evaluation.
If the measured concentrations are above the calibration range, then a suitable dilu-
tion must be prepared and the analysis must be carried out again.
The MAK Collection for Occupational Health and Safety 2016, Vol 1, No 3
10.1002/3527600418.am0lmix002e1816, Downloaded from https://onlinelibrary.wiley.com/doi/10.1002/3527600418.am0lmix002e1816 by Readcube (Labtiva Inc.), Wiley Online Library on [30/01/2024]. See the Terms and Conditions (https://onlinelibrary.wiley.com/terms-and-conditions) on Wiley Online Library for rules of use; OA articles are governed by the applicable Creative Commons License
Solvent mixture Meth.-No. 2 2147
6 Calibration
The calibration solutions described in Section 2.4 are used to obtain a calibration
function. The internal standard method is used for calibration. Each calibration
standard is injected into the gas chromatograph three times and analysed in the
same manner as the sample solution (determination in triplicate). The calibration
must be performed anew if the analytical conditions change or the quality control
results indicate that this is necessary.
Based on the peak areas obtained the corresponding mass X in μg is obtained from
the calibration curve. The corresponding mass concentration (ρ) of a component is
calculated according to Equation (1) as follows:
X
ρ¼ ð1Þ
V $η
273 þ ta 1013
ρ0 ¼ ρ $ $ ð2Þ
293 pa
where:
ρ is the mass concentration of a solvent component in mg/m3
ρ0 is the mass concentration of the component in mg/m3 at 20 °C and 1013 hPa
X is the mass of the component in the analytical sample in μg
V is the air sample volume in litres
η is the recovery
ta is the temperature during sampling in °C
pa is the air pressure during sampling in hPa
The characteristics of the method were calculated as stipulated in EN 482 [1] and
EN 1076 [2]. Test gases were produced on a dynamic test gas facility in a concen-
tration range of the respective limits of quantification up to twice the Occupational
Exposure Limits. These test gases were humidified to give a relative humidity from
30 to 75%. Samples were taken from each of these concentrations at room tempera-
ture (air sample volume 40 mL), prepared according to the Sections 3.2, 4 and 5
and analysed.
8.1 Precision
Six charcoal tubes for each of the three different concentrations were loaded at a
relative humidity of approx. 50% for the purpose of determining the precision. The
resulting relative standard deviations are shown in Table 5 (Appendix).
8.2 Recovery
The recoveries of the selected solvent components were determined for concentra-
tions from 0.5 to twice the Occupational Exposure Limits using a dynamic test gas
facility. The values were determined for six sample tubes in each case at a relative
humidity of approx. 50%. In contrast, determination of the recoveries and relative
standard deviations close to the limits of quantification was carried out by direct
application of the substances onto the charcoal tubes (liquid spiking). Then 40 L of
air were drawn through the tubes at a flow rate of approx. 20 L/h (0.333 L/min).
The samples were prepared and analysed as described in the Sections 3.2, 4 and 5.
The results are shown in Table 6 (Appendix).
The MAK Collection for Occupational Health and Safety 2016, Vol 1, No 3
10.1002/3527600418.am0lmix002e1816, Downloaded from https://onlinelibrary.wiley.com/doi/10.1002/3527600418.am0lmix002e1816 by Readcube (Labtiva Inc.), Wiley Online Library on [30/01/2024]. See the Terms and Conditions (https://onlinelibrary.wiley.com/terms-and-conditions) on Wiley Online Library for rules of use; OA articles are governed by the applicable Creative Commons License
Solvent mixture Meth.-No. 2 2149
relative humidity from 75 to 80% with concentrations of half and twice the Occupa-
tional Exposure Limits of the selected substances and sealed with plastic caps. After
a storage period of 7 and 14 days at room temperature six sample tubes were ana-
lysed in each case. During the investigation period no significant losses could be
detected.
8.6 Interferences
Interferences from overlapping signals should be corrected by the use of two se-
paration columns arranged in parallel of different polarity. Experience of using this
equipment has shown that, apart from a few exceptions, at least one chromatogram
without signal interference is obtained and can be evaluated. Table 7 lists the reten-
tion times of the investigated substances for the columns used (for chromato-
graphic conditions see Section 4).
Appendix
Table 5 (Continued)
The MAK Collection for Occupational Health and Safety 2016, Vol 1, No 3
10.1002/3527600418.am0lmix002e1816, Downloaded from https://onlinelibrary.wiley.com/doi/10.1002/3527600418.am0lmix002e1816 by Readcube (Labtiva Inc.), Wiley Online Library on [30/01/2024]. See the Terms and Conditions (https://onlinelibrary.wiley.com/terms-and-conditions) on Wiley Online Library for rules of use; OA articles are governed by the applicable Creative Commons License
Solvent mixture Meth.-No. 2 2151
Table 5 (Continued)
Table 5 (Continued)
Table 6 Recoveries, rel. standard deviations, and limits of quantification for n = 6 determina-
tions
The MAK Collection for Occupational Health and Safety 2016, Vol 1, No 3
10.1002/3527600418.am0lmix002e1816, Downloaded from https://onlinelibrary.wiley.com/doi/10.1002/3527600418.am0lmix002e1816 by Readcube (Labtiva Inc.), Wiley Online Library on [30/01/2024]. See the Terms and Conditions (https://onlinelibrary.wiley.com/terms-and-conditions) on Wiley Online Library for rules of use; OA articles are governed by the applicable Creative Commons License
Solvent mixture Meth.-No. 2 2153
Table 6 (Continued)
Table 7 Retention times (RT) of the investigated solvents on the polar (HP-INNOWAX) and
non-polar (Rtx-5) separation column
Rtx-5 HP-INNOWAX
RT [min] Substance RT [min] Substance
6.00 Ethanol* 4.70 n-Pentane
6.60 2-Propanol 5.00 n-Hexane
6.72 n-Pentane 5.60 n-Heptane
7.90 1-Propanol 5.93 Cyclohexane
9.43 n-Hexane 6.18 2-Methylheptane
9.43 2-Butanol 6.44 Methylcyclohexane
9.93 Ethyl acetate 6.70 n-Octane
10.67 2-Methoxy-1-propanol 7.33 Methyl acetate
(isobutanol)
10.88 Tetrahydrofuran 8.11 Tetrahydrofuran
12.55 Isopropyl acetate 8.79 Ethyl acetate
12.69 1-Butanol 9.00 n-Nonane
13.13 Cyclohexane 9.17 Isopropyl acetate
13.31 1-Methoxy-2-propanol 10.00 2-Propanol
15.00 n-Heptane 10.30 Propyl acetate
Table 7 (Continued)
Rtx-5 HP-INNOWAX
RT [min] Substance RT [min] Substance
15.47 1,4-Dioxane 12.91 n-Decane
15.65 Propyl acetate 14.11 Isobutyl acetate
16.80 Methylcyclohexane 14.53 2-Butanol
18.80 2-Methylheptane (ISTD) 15.21 1-Propanol
19.08 Isobutyl acetate 15.54 Toluene
19.10 Toluene 16.54 1,4-Dioxane
20.65 n-Octane 16.96 n-Butyl acetate
21.18 n-Butyl acetate 17.58 2-Methoxy-1-propanol
(isobutanol)
23.58 1-Methoxy-2-propyl acetate 17.62 n-Undecane (ISTD)
23.95 Ethylbenzene 19.29 1-Methoxy-2-propanol
24.34 m-Xylene 19.47 Ethylbenzene
24.41 p-Xylene 19.80 p-Xylene
25.05 Cyclohexanol 19.90 1-Butanol
25.30 2-Butoxyethanol 20.07 m-Xylene
25.44 o-Xylene 21.37 Cumene
25.44 n-Nonane 21.86 o-Xylene
26.76 Cumene 22.76 n-Propylbenzene
28.04 n-Propylbenzene 23.12 1-Methoxy-2-propyl acetate
28.63 1,3,5-Trimethylbenzene 24.07 1,3,5-Trimethylbenzene
29.52 1,2,4-Trimethylbenzene 25.41 1,2,4-Trimethylbenzene
29.69 n-Decane 27.29 1,2,3-Trimethylbenzene
30.82 1,2,3-Trimethylbenzene 28.99 2-Butoxyethanol
32.42 2-Butoxyethyl acetate 29.01 Cyclohexanol
33.12 n-Undecane 30.53 2-Butoxyethyl acetate
*Ethanol can only be determined on the Rtx-5 column and methyl acetate only on the HP-Innowax column,
as they overlap with the extraction agent on the respective other column.
References
1 EN 482 (2012) Workplace atmospheres – General requirements for the performance of proce-
dures for the measurement of chemical agents. Beuth Verlag, Berlin, Germany
2 EN 1076 (2010) Workplace exposure – Procedures for measuring gases and vapours using
pumped samplers – Requirements and test methods. Beuth Verlag, Berlin, Germany
The MAK Collection for Occupational Health and Safety 2016, Vol 1, No 3
10.1002/3527600418.am0lmix002e1816, Downloaded from https://onlinelibrary.wiley.com/doi/10.1002/3527600418.am0lmix002e1816 by Readcube (Labtiva Inc.), Wiley Online Library on [30/01/2024]. See the Terms and Conditions (https://onlinelibrary.wiley.com/terms-and-conditions) on Wiley Online Library for rules of use; OA articles are governed by the applicable Creative Commons License
Solvent mixture Meth.-No. 2 2155
3 DIN 32645 (2008) Chemical analysis – Decision limit, detection limit and determination limit
under repeatability conditions – Terms, methods, evaluation. Beuth Verlag, Berlin, Germany
4 Heckmann P, Breuer D (2013) IFA method for calculating expanded measurement uncer-
tainty, http://www.dguv.de/ifa/index-2.jsp