You are on page 1of 5

Journal of Radiotherapy in

Practice
Percentage depth dose fragmentation for
investigating and assessing the photon
cambridge.org/jrp beam dosimetry quality
Mohamed Bencheikh, Abdelmajid Maghnouj and Jaouad Tajmouati
Original Article LISTA Laboratory, Physics Department, Faculty of Sciences Dhar El-Mahraz, University of Sidi Mohamed Ben
Cite this article: Bencheikh M, Maghnouj A, Abdellah, Fez, Morocco
Tajmouati J. (2019) Percentage depth dose
fragmentation for investigating and assessing Abstract
the photon beam dosimetry quality. Journal
of Radiotherapy in Practice 18: 280–284. Aim: The purpose of this study is to introduce a new approach to assess the dosimetry quality
doi: 10.1017/S1460396918000687 of photon beam with energy and irradiation field size. This approach is based on percentage
depth dose (PDD) fragmentation for investigating the dosimetry quality. Materials and
Received: 9 August 2018
methods: For the investigation of the dosimetry quality of 6 and 18 MV photon beams, we
Revised: 20 October 2018
Accepted: 3 November 2018 have proceeded to fragment the PDD at different field sizes. This approach checks the overall
First published online: 4 December 2018 PDD and is not restricted to the exponential decay regions, as per the International Atomic
Energy Agency Technical Reports Series No 398 and the American Association of Physicist in
Key words: Medicine Task Group 51 recommendations. Results and discussion: The 6 MV photon beam
dosimetry quality; fragmentation; photon
beam energy; quality index; radiotherapy
deposited more energy in the target volume than the 18 MV photon beam. The dose delivered
efficiency by the 6 MV beam is greater by a factor of 1·5 than that delivered by the 18 MV beam in the
build-up region and the dose delivered by the 6 MV beam is greater by a factor of 2·6 than
Author for correspondence: that delivered by the 18 MV beam in the electronic equilibrium and the exponential decay
Mohamed Bencheikh, Physics Department,
Faculty of Sciences Dhar El-Mahraz, University of
regions. Conclusion: The dose measured at different points of the beam is higher for 6 MV
Sidi Mohamed Ben Abdellah, Fez, 30000, than for 18 MV photon beam. Therefore, the 6 MV beam is more dosimetrically efficient than
Morocco. E-mail: bc.mohamed@gmail.com the 18 MV beam. Using the proposed approach, we can assess the dosimetry quality by taking
into account overall PDD not only in the exponential decay region but also in the field.

Introduction
Periodically checking the beam quality is an integral part of quality assurance management in
a radiotherapy department.1,2 The use of an appropriate quality control (QC) methods to
adequately assess the photon beam quality can be a challenge for medical physicists, because to
verify the dosimetry quality the procedure that has been followed must be as accurate as
possible. On some occasions, after assessing the beam quality, minor calibrations can be made
without introducing any functioning perturbation of the Linac head.3 The required QC
programmes have been published by different international or national authorities that
recommend a fixed period of time in checking the dosimetry output quality based on absorbed
dose measurements for the Linac head calibration. The International Atomic Energy Agency
(IAEA) and the American Association of Physicist in Medicine (AAPM) recommend assessing
the beam quality using the percentage depth dose (PDD) parameter.4,5
With reference to PDD, both the international bodies recommend not assessing the beam
quality before the depth of 10 cm, whatever be the photon beam energy, in order to avoid
measuring electron contamination even though this is an integral part of the incident photon
beam. The beam quality index is an essential parameter used in beam quality assessment in
radiotherapy treatment, but it should be considered in overall PDD variation.6 Earlier, Otto
had worked on the determination of a quality index for photon beams of arbitrary field sizes in
treatments where the reference calibration conditions could not be met; however, the accuracy
of this index was not rigorously checked.7
This study aims to find a new approach for the investigation and evaluation of dosimetry
quality of the incident photon beam based on PDD sub-divisions or PDD fragmentation by
examining all the dosimetric properties of PDD sub-regions. This new approach allows to
check the beam dosimetric quality at any depth and of any field size by including all dosi-
metric particularities of all PDD sub-regions starting from the skin dose to exit dose. We
consider that it is necessary to include the dosimetry measurements due to electron con-
© Cambridge University Press 2018. tamination and region of low photon energy in the beam quality procedure to ensure accuracy
of the treatment. Electron contamination and low photon energy are always present in the
incident photon beam and if not measured, it is considered as a procedural weakness.
This study is performed based on previous work, in which we studied the photon beam
quality and the impact of particle contamination on beam quality, when the flattening filter is
removed from the Linac head.8–10 For any new Linac technological development, the

https://doi.org/10.1017/S1460396918000687 Published online by Cambridge University Press


Journal of Radiotherapy in Practice 281

evaluation procedure of the beam quality should be reviewed in To demonstrate the PDD dependence on beam energy,
order to assure a high degree of accuracy in dosimetry evaluation. Figure 2 gives the PDD variation in two photon beam energies of
6 and 18 MV for field size of 10 × 10 cm2.
It is obvious from Figure 2 that PDD varies with photon beam
Materials and Methods energy for the same irradiation field size. The depth of maximum
The dosimetry quality assessment is crucial to ensure radio- dose of 6 MV photon beam is at 1·5 cm and the depth of max-
therapy quality and accuracy. The IAEA TRS 398 recommends imum dose of 18 MV photon beam is at 3 cm.
assessing the beam quality index using tissue phantom ratio
(TPR) as a quotient of PDD at a depth of 20 cm to 10 cm for field PDD fragmentation
size of 10 × 10 cm2. However, the AAPM Task Group 51 The PDD fragmentation is aimed to investigate the dosimetric
recommends assessing the beam quality as per PDD at a depth of particularities in each PDD fragment of the PDD curve from the
10 cm by using a lead slab at the exit window of Linac head of skin dose (depth = 0 mm) to exit dose (depth = 300 mm). This
thickness x which varied with photon beam energy to remove approach included the dosimetric properties of all PDD sub-
electron contamination.4,5 regions and it was not restricted to the dosimetric properties of
These protocols allow the evaluation of the beam quality at a one sub-region of the exponential decay region, as recommended
depth of 10 cm whatever be the photon beam energy and aim to by IAEA TRS 398 and AAPM TG 51. To accurately investigate
avoid electron contamination and low photon energy. This the overall dosimetric properties of the PDD, the PDD curve was
avoidance can have negative impact on the accuracy if there is an sub-divided into seven sub-regions. In agreement with TRS 398,
abnormal functioning of a beam modifier in the Linac head. In the PDD sub-region was measured from 10 to 20 cm. Table 1
fact, electron contamination and low photon beam energy are an presents the depth intervals of seven PDD sub-regions for both
integral part of the photon beam and are always present in the photon beam energies, 6 and 18 MV.
incident beam. We have introduced an investigation parameter Q based on
the PDD method that was used in earlier studies.12,13 The PDD
Measurements of PDD method is expressed as an empirical formula relating the TPR and
In this study, dose measurements are measured for two photon PDD as follows:
beam energies, 6 and 18 MV, produced by Varian Clinac 2100 TPR20;10 = 1:2661; PDD20;10 = 0:0595 (1)
(Varian, Palo Alto, CA, USA). These beam energies are more
commonly used in clinical practice. The measurements were where,
performed in a water phantom of volume 40 × 40 × 40 cm3, as PDD20
recommended by the Swiss Society of Radiobiology and Medical PDD20;10 = (2)
PDD10
Physics (SSRMP).11
The PTW 30013 chamber (Physikalisch Technische Werk- In Equation (2) we have generalised for seven PDD sub-
stätten (PTW) Freiburg, Germany) was used for the PDD mea- regions instead of one region as required in TRS 398, where the
surements of both 6 and 18 MV photon beams. MEPHYSTO bordering depths are 20 cm and 10 cm. Equation (3) calculates the
software (PTW, Freiburg, Germany) was used to drive the ion investigation parameter Q using the formula:
chamber for data acquisition at increments of 2·5 mm in depth. PDDd2
All PDD measurements were carried out under conditions of Qd2 ;d1 = (3)
PDDd1
temperature of 20°C and the pressure of 101·3 Pa and humidity of
50%. The uncertainty of PDD measurements was found to be where d1 and d2 are the depths bordering the PDD sub-regions
<2% and this included all the uncertainties of experimentation and d1 < d2.
and the uncertainty of the measurements of the devices used. After determining the parameter Q for each PDD sub-region,
According to IAEA TRS 389, the PDD measurements are normalisation of Q is introduced to standardise the parameter
measures and the PDD curve is formed by three regions: build-up variation with each photon beam energy and to conceal the
dose region, electronic equilibrium region and exponential decay dependence on the beam energy. The normalised Q is symbolised
region. Figure 1 illustrates the PDD sub-regions of 6 MV photon by Q and it is defined as a quotient of Q of each PDD sub-region
beam for field size of 10 × 10 cm2. to Q of overall PDD. The Q is expressed in Equation (4) as

120 6 MV 18 MV
120
Build-up region
100
100

80 80
PDD (%)

PDD (%°)

Exponentiel decay region


60 60

40 40

20 Electronic equilibrium region 20

0 0
0 50 100 150 200 250 300 350 0 50 100 150 200 250 300 350
Depth d (mm) Depth (mm)

Figure 1. PDD variation as a function of depth for 6 MV photon beam. Figure 2. PDD variation as a function of depth for two photon beam energies.

https://doi.org/10.1017/S1460396918000687 Published online by Cambridge University Press


282 Mohamed Bencheikh et al.

follows: Figure 3 shows Q variation as a function of irradiation field


size in two sub-regions of the build-up dose region.
Qd2 ;d1 From Figure 3, the normalised Q is above 1 for both photon
Q= (4)
Q0;300 beam energies of 6 and 18 MV. It means that the photon beam
delivers the majority of its energy in the build-up dose region. So,
where Qd2 ;d1 for a PDD sub-region ranges from d1 to d2; Q0;300 is in case A, the Q value decreased rapidly with field size, whereas in
the overall PDD curve. case B, the Q value decreased slowly with field size.
Here we have demonstrated that the build-up dose varies with
beam energy and field size. The photons of the lower energy
Results and Discussion (6 MV) deposit more energy in this region than the photons of a
Build-up dose quality analysis high energy (18 MV). This dependence on photon beam energy
should be included in the evaluation procedure of the beam
The build-up dose region ranges from 0 mm to depth of max- quality.
imum dose. In this region, the PDD is fragmented into two sub-
regions, as shown in Table 1; therefore, the dosimetry quality is Electronic equilibrium quality analysis
investigated by calculation of Q on two depth intervals A and B
(Table 1). The electronic equilibrium region ranges on bordering regions of
maximum dose. Figure 4 shows the variation in Q in the elec-
Table 1. Depth intervals of each PDD sub-region tronic equilibrium region as a function of field size for photon
beam energies of 6 and 18 MV.
Photon beam PDD region in the TRS Depths bordering the PDD sub-
It can be observed from Figure 4 that in the electronic equi-
energy 398 region
librium region, the Q decreased slightly in field size for 6 MV
6 MV Build-up of dose A: d1 = 0 mm and d2 = 15 mm photon beam for small field but it is approximately constant for
B: d1 = 5 mm and d2 = 15 mm field size greater than 12 × 12 cm2, but the Q value increased with
Electronic equilibrium A: d1 = 12·5 mm and field size for 18 MV photon beam (Figure 4). These discrepancies
d2 = 17·5 mm can be used as a basis to assess the photon beam quality in the
B: d1 = 10 mm and d2 = 20 mm electronic equilibrium region. Again, the dosimetry quality has
shown that it varied with photon beam energy and also with field
Exponential decay A: d1 = 50 mm and d2 = 100 mm
B: d1 = 100 mm and size. It varied from one sub-region of PDD to another (Figures 3
d2 = 200 mm and 4).
C: d1 = 200 mm and
d2 = 300 mm
Exponential decay quality analysis
Overall PDD d1 = 0 mm and d2 = 300 mm
The exponential decay region ranges from depth of maximum
18 MV Build-up of dose A: d1 = 0 mm and d2 = 30 mm dose to exit dose (depth of 300 mm). It is characterised by
B: d1 = 10 mm and d2 = 30 mm exponential photon beam attenuation with depth. The beam
quality index is calculated on this PDD sub-region as recom-
Electronic equilibrium A: d1 = 25 mm and d2 = 35 mm
B: d1 = 22·5 mm and mended by IAEA TRS 398 protocol.12 In our study, we have
d2 = 37·5 mm determined the parameter Q for three sub-regions in the expo-
nential decay region.
Exponential decay A: d1 = 50 mm and d2 = 100 mm To understand the dosimetric properties in the exponential
B: d1 = 100 mm and
d2 = 200 mm decay region with field size and the overall PDD for both photon
C: d1 = 200 mm and beam energies 6 and 18 MV, the parameter Q is determined for
d2 = 300 mm three sub-regions of the PDD.
Figure 5 presents the Q variation in the exponential decay
Overall PDD d1 = 0 mm and d2 = 300 mm
region with field size.

(a) 6 MV 18 MV (b) 6 MV 18 MV
3.5
Normalized investigation parameter

6
Normalized investigation parameter

3
5
2.5
4
2
3
1.5
2
1
1 0.5

0 0
0 10 20 30 0 10 20 30
Side of square field (cm) Side of square field (cm)
Figure 3. Q variation for build-up dose region as a function of field size.

https://doi.org/10.1017/S1460396918000687 Published online by Cambridge University Press


Journal of Radiotherapy in Practice 283

(a) 6 MV 18 MV (b) 6 MV 18 MV
3 3

Normalized investigation parameter

Normalized investigation parameter


2.5 2.5

2 2

1.5 1.5

1 1

0.5 0.5

0 0
0 10 20 30 0 10 20 30
Side of square field (cm) Side of square field (cm)
Figure 4. Q variation for electronic equilibrium region as a function of field size.

(a) 6 MV 18 MV
(b) 6 MV 18 MV
1.8
1.8

Normalized investigation parameter


Normalized investigation parameter

1.6
1.6
1.4 1.4

1.2 1.2

1 1
0.8 0.8
0.6 0.6
0.4 0.4
0.2 0.2
0 0
0 10 20 30 0 10 20 30
Side of square field (cm) Side of square field (cm)

(c) 6 MV 18 MV
2.5
Normalized investigation parameter

1.5

0.5

0
0 5 10 15 20 25 30
Side of square field (cm)
Figure 5. Q variation for exponential decay region as a function of field size.

It can be seen from Figure 5, the dosimetry quality varied with and at any sub-regions of exponential decay region. For example,
depth in the exponential decay region and is not limited to middle the 6 MV photon beam has Q value >1 and the 18 MV photon
regions, as recommended by IAEA TRS 398 protocol. Q is located beam has Q < 1 (Figures 5a–5c).
above to 1 for 6 MV and below 1 for 18 MV.
As a generalisation of our approach, for a photon beam with
energy <6 MV, we found a Q value >1 and for a beam with an
Conclusion
energy >18 MV, the Q value is <1 in the overall exponential
decay region for field size less than 15 × 15 cm2. However, IAEA The findings of this work are based on an experimental study of
TRS 398 and AAPM TG 51 limit the beam quality determination beam quality by introducing a new approach based on frag-
on field size of 10 × 10 cm2.12 On the basis of these results, we mentation of PDD curves into seven sub-regions. On the basis of
were able to make a distinction between photon beams used in the proposed approach, we can assess the dosimetry quality by
radiotherapy treatment at any field size smaller than 15 × 15 cm2 taking into account overall PDD not only in the exponential

https://doi.org/10.1017/S1460396918000687 Published online by Cambridge University Press


284 Mohamed Bencheikh et al.

decay region but also in the field. This new approach of the beam 6. Klein E E, Hanley J, Bayouth J et al. AAPM Task Group 142 Report:
quality assessment is more accurate than in current calibration quality assurance of medical accelerators. Med Phys 2009; 36: 4197–4212.
methods as recommended by the IAEA and AAPM.10 7. Otto A S. Determination of the quality index (Q) for photon beams at
arbitrary field sizes. Med Phys 2009; 39 (9): 4168–4172.
According to this new approach, the 6 MV photon beam is
8. Bencheikh M, Maghnouj A, Tajmouati J. Dosimetry Investigation and
more dosimetric efficient than an 18 MV photon beam.
Evaluation for Removing Flattening Filter Configuration of Linac: Monte
Carlo Study. Moscow University Physics Bulletin 2017; 72(6): 640–646.
Acknowledgements. The authors would like to thank Varian Medical Sys- 9. Bencheikh M, Maghnouj A, Tajmouati J. Study of photon beam dosimetry
tems who provided us with the Varian Clinac 2100 dosimetry data and quality for removing flattening filter Linac configuration. Ann Univ
afforded the opportunity to study the Varian linear accelerator technology and
Craiova Phys 2017; 27: 50–60.
to contribute to its future development. 10. Bencheikh M, Maghnouj A, Tajmouati J. Energetic properties’ investiga-
tion of removing flattening filter at phantom surface: Monte Carlo study
using BEAMnrc code, DOSXYZnrc code and BEAMDP code. PEPAN
References Letters 2017; 14 (No. 6): 921–930.
1. Technical Reports Series TRS No. 430. Commissioning and quality 11. Swiss Society of Radiobiology and Medical Physics (SSRMP). Dosimétrie
assurance of computerized planning systems for radiation treatment of des faisceaux de photons de haute énergie l’aide de chambres
cancer. Vienna: International Atomic Energy Agency, 2004. d’ionisation, Recommandation N° 8, 2000, Bern-Wabern, Swiss. ISBN
2. IAEA-TECDOC-1540. Specification and acceptance testing of radio- 3-908125-26-X.
therapy treatment planning systems. Vienna: International Atomic Energy 12. Lye J E, Butler D J, Oliver C P et al. Comparison between the TRS-398
Agency, 2007. code of practice and the TG-51 dosimetry protocol for flattening filter
3. Uddin M T. Quality control of modern linear accelerator: dose stability free beams. Phys Med Biol 2016; 61 (14): 362–372.
long and short-term. Proceedings of IPAC 2012, New Orleans, Louisiana, 13. Silvia V C, Francisco C H. Comparison of IPSM 1990 photon dosimetry
USA, 2012: 2660–2662. code of practice with IAEA TRS-398 and AAPM TG-51. J Appl Clin Med
4. Technical Reports Series TRS No. 398. Absorbed dose determination in Phys 2008; 10 (1): 136–146.
external beam radiotherapy: an international code of practice for 14. Followill D S, Tailor R C, Tello V M, Hanson W F. An empirical
dosimetry based on standards of absorbed dose to water. Vienna: relationship for determining photon beam quality in TG-21 from a ratio
International Atomic Energy Agency, 2004. of percent depth doses. Med Phys 1998; 25: 1202–1205.
5. Almond P R, Biggs P J, Coursey B M et al. AAPM’s TG-51 protocol for 15. Hugo P. Determination of the beam quality index of high-energy photon
clinical reference dosimetry of high-energy photon and electron beams. beams under nonstandard reference conditions. Med Phys 2012; 39 (9):
Med Phys 1999; 26 (9): 1847–1869. 5513–5519.

https://doi.org/10.1017/S1460396918000687 Published online by Cambridge University Press

You might also like