Professional Documents
Culture Documents
Ap4498-Bulletin - Flight - 2nd - Stage - Micro - Adjustment Bulletin
Ap4498-Bulletin - Flight - 2nd - Stage - Micro - Adjustment Bulletin
AP4498-Bulletin 1
7/05/2010
FLIGHT Second Stage Regulator Technical Service Bulletin
FLIGHT regulator technical service 5. Locate the three kidney shaped pegs of the AT71
Bulletin Locking Ring Tool into the slots of the Inner Locking
WARNING: Compressed air can be highly Ring (3). Unscrew the Inner Locking Ring (3). Remove
explosive and is dangerous if misused. the Friction Ring (4) and the Diaphragm (5) from
inside of the Case (6).
Ensure the cylinder valve is opened slowly.
Use eye and ear personal protective
equipment when performing any tests
involving compressed air.
CAUTION: DO NOT use any tools to CAUTION: Take care not to damage or
remove the Hose. The connection has been distort the Lever (15) when removing the
designed to be operated by hand only. Venturi Lever (10). DO NOT depress the
tab of the Venturi Lever too far, as this will
result in permanent deformation.
4. Remove the Purge Clamp Ring (1) squeeze at the
positions shown on the photo below and turn anti-
NOTE: The Flight second stage regulator
clockwise. Separate the Purge Clamp Ring (1) from
maintenance manual (AP4498) shows the opening
the Front Cover Assy (2) and the Case (6).
effort being set with the Front Cover Assy (2) and
the Purge Clamp Ring (1) removed. It is possible to
set the opening effort with only the Diaphragm (5),
Flight Friction Ring (4) and the Inner Locking Ring
(3) installed. However to acheive a more accurate
setting the Front Cover Assy (2) and the Purge
Clamp Ring (1) should be re-attached prior to
adjustment. This is because the opening effort may
be slightly affected by the front cover reassembly, due
to the tight tolerances of the compact second stage.
1
FLIGHT Second Stage Regulator Technical Service Bulletin
2
FLIGHT Second Stage Regulator Technical Service Bulletin
12. If the opening effort requires further adjustment re- 17. Slide the Sliding Locking Ring (19) back onto the Inlet
engage the hexagonal bit of the Flight Micro Adjuster Fitting (18) half way, without engaging the notches
Tool and repeat step 11. on the Case (6). Turn the Sliding Locking Ring (19)
anti-clockwise to unscrew the Inlet Fitting (18). If the
Inlet Fitting is too tight unscrew using tool AT77 Flight
NOTE: A 2.5mm Allen key can be used to adjust the Venturi Combination Tool.
Micro Adjuster Screw (20) if a Flight Micro Adjuster
Tool (AT76) is not available. The open end of the case
must be blanked off to obtain a reading on the test
bench.
14. Observe for any bubbles arising from the 19. Carefully insert a new Rubber Seating (26) into the
submerged regulator over a one minute period. The end of the Shuttle Valve (25).
recommended time is necessary due to slower bubble
formation that occurs in smaller leaks. Bubbles
indicate a leak, which requires the system to be
disassembled at the source to check sealing surfaces,
assembly sequence and component positioning in
order to correct the problem(s).
3
FLIGHT Second Stage Regulator Technical Service Bulletin
4
FLIGHT Second Stage Regulator Technical Service Bulletin